Download as pdf or txt
Download as pdf or txt
You are on page 1of 10

Exp Physiol 91.

6 pp 967–976 967

Experimental Physiology

Effect of frequency and pulse duration on human muscle


fatigue during repetitive electrical stimulation
Trisha Kesar and Stuart Binder-Macleod
Department of Physical Therapy, 301 McKinly Laboratory, University of Delaware, Newark, DE 19716, USA

Different combinations of stimulation frequency and intensity can generate a targeted force
during functional electrical stimulation (FES). This study compared isometric performance and
muscle fatigue during repetitive stimulation with three different combinations of frequency and
pulse duration that produced the same initial peak forces: protocol 1 used long pulse duration
(fixed at 600 μs) and 11.5 ± 1.2 Hz (low frequency); protocol 2 used 30 Hz (medium frequency)
and medium pulse duration (150 ± 21 μs); and protocol 3 used 60 Hz (high frequency) and
short pulse duration (131 ± 24 μs). Twenty and 60 Hz pre- and postfatigue testing trains were
delivered at the pulse duration used by the fatiguing trains and at 600 μs pulse duration. The
percentage decline in peak force between the first and last fatiguing train of each protocol was
the measure of muscle performance. The declines in peak force of the 60 Hz testing trains were
used to measure muscle fatigue. The 20 Hz:60 Hz peak force ratio was used as a measure of low-
frequency fatigue. The results showed that protocol 1 produced the least decline in peak force
in response to the fatiguing trains, as well as the least muscle fatigue and low-frequency fatigue
when the pulse duration was maintained at the level used by the fatiguing trains. Interestingly,
protocol 2 produced the least muscle fatigue, and there were no differences in the levels of low-
frequency fatigue across protocols when a comparable motor unit population was tested using
600 μs pulse duration. The results suggest that if the frequency and intensity are kept constant
during FES, using the lowest frequency and longest pulse duration may maximize performance.
(Received 13 March 2006; accepted after revision 22 June 2006; first published online 14 August 2006)
Corresponding author S. Binder-Macleod: Department of Physical Therapy, 301 McKinly Laboratory, University of
Delaware, Newark, DE 19716, USA. Email: sbinder@udel.edu

Functional electrical stimulation (FES) uses electrical During FES, however, skeletal muscles fatigue more
stimulation to generate functional movements in rapidly during repetitive stimulation than during
individuals with upper motor neurone paresis (Liberson volitional contractions (Marsolais & Edwards, 1988;
et al. 1961; Kralj et al. 1988). Functional electrical Riener, 1999). Rapid fatigue during FES is thought to
stimulation can help individuals with paralysis caused by result from the differences in motor unit recruitment
upper motor neurone dysfunctions, such as spinal cord order, higher activation frequencies and imprecise control
injury, cerebral palsy and stroke, to regain the ability of muscle force during FES compared to volitional
to stand (Triolo et al. 1996), walk (Agarwal et al. 2003; contractions (Peckham & Knutson, 2005). The problem
Johnston et al. 2005) and grasp objects (Mangold et al. of muscle fatigue is further compounded by the fact that
2005). Functional electrical stimulation has proven to paralysed muscles show greater fatigability than healthy
be effective for lower and upper extremity rehabilitation muscle (Gerrits et al. 1999, 2003). Muscle fatigue is an
in individuals with hemiplegia following stroke (Ring & important factor limiting the clinical use of FES (Riener,
Rosenthal, 2005), and for providing exercise alternatives 1999). During repetitive electrical stimulation, stimulation
for tetraplegic individuals through FES-induced rowing frequency and intensity are two primary parameters
(Davoodi et al. 2002; Wheeler et al. 2002) and cycling that can be modulated to control skeletal muscle force.
(Gfohler & Lugner, 2000; Hunt et al. 2004). For effective Although numerous combinations of frequency and
task performance during FES, it is necessary to maintain intensity can be used to generate the required muscle force
the level of muscle force required for generation of the during FES, most clinical FES systems use the minimum
FES-elicited movement. frequency that can generate a fused tetanic contraction


C 2006 The Authors. Journal compilation 
C 2006 The Physiological Society DOI: 10.1113/expphysiol.2006.033886
968 T. Kesar and S. Binder-Macleod Exp Physiol 91.6 pp 967–976

in the muscle being stimulated and vary the intensity intensity that can generate the targeted muscle force
to produce the desired force (Taylor et al. 1999; Weber while simultaneously minimizing muscle fatigue. It has
et al. 2005). The stimulation intensity is further increased been hypothesized, but not systematically tested, that the
as the muscle fatigues (Taylor et al. 1999; Weber et al. combination of the lowest frequency and highest intensity
2005). The rationale for using low frequencies in FES may minimize fatigue when used for repetitive stimulation
is based on the premise that higher frequencies cause (Binder-Macleod & Snyder-Mackler, 1993). No previous
greater fatigue (Bigland-Ritchie et al. 1979; Garland et al. study has attempted systematically to investigate the
1988). Although several previous studies have shown that combination of stimulation frequency and intensity that
fatigue is a function of the stimulation frequency or the can minimize fatigue while producing a targeted force level
number of pulses (Bigland-Ritchie et al. 1979; Marsden during repetitive stimulation.
et al. 1983; Garland et al. 1988; Binder-Macleod et al. The purpose of this study was to determine which
1998), these studies have often ignored the effect of combination of stimulation intensity and frequency
either the differences in initial peak force or stimulation produces the least decline in force during repetitive
intensities on muscle fatigue (Binder-Macleod et al. 1995; electrical stimulation, for the same initial peak force.
Russ et al. 2002b). Russ and colleagues (Russ et al. Both the amplitude and the duration of the stimulus
2002c) recently showed that increasing the frequency or pulses can be varied to modulate the stimulation intensity
number of pulses did not affect the amount of muscle during electrical stimulation. For this study, we used
fatigue produced during repetitive isometric contractions stimulus pulse duration (PD) to vary the intensity
if the initial force produced by the stimulation trains because it was easier to control and provided a more
was controlled. Another study showed that intermittent consistent force response from the muscle compared
high-frequency stimulation produced less fatigue than to stimulation amplitude (Grill & Mortimer, 1996).
low-frequency repetitive stimulation in able-bodied and Specifically, we compared the percentage decline in
spinal cord injured subjects (Matsunaga et al. 1999). Thus, quadriceps femoris isometric muscle force produced
previous literature does not provide conclusive evidence during repetitive stimulation with trains consisting of
about the isolated effect of stimulation frequency on three different combinations of stimulation frequencies
muscle fatigue. and pulse durations that produced the same initial
Only a few studies have investigated the relationship targeted peak force, as follows: protocol 1 used trains
between stimulation intensity and muscle fatigue. Binder- with 600 μs pulse duration (long pulse duration), and
Macleod et al. (1995) tested the rate and amount the frequency was varied for each subject to generate
of fatigue during repetitive stimulation of the human the targeted force (low frequency); protocol 2 used 30 Hz
quadriceps muscle with trains at stimulation amplitudes trains (medium frequency), and the pulse duration was
that produced 20, 50 and 80% maximum voluntary varied for each subject to generate the targeted force
isometric contraction (MVIC) forces, and found less (medium pulse duration); and protocol 3 used 60 Hz trains
decline in force of the fatiguing trains during repetitive (high frequency) and the pulse duration was varied for
stimulation at higher compared to lower stimulation each subject to generate the targeted force (short pulse
amplitudes. In contrast, Godfrey et al. (2002) recently duration). Please note that the terms ‘medium pulse
showed greater declines in force due to fatigue at duration’ and ‘short pulse duration’ were operational
high (supramaximal) compared to low (submaximal) definitions for the pulse durations used to generate the
stimulation intensities. Thus, we also do not know which targeted force using 30 and 60 Hz trains for protocols 2
stimulation intensity levels can help to minimize fatigue and 3, respectively. In addition, the frequency and pulse
when used for repetitive stimulation during FES. duration used during protocol 2 were similar to the
Since the stimulation frequency (Bigland-Ritchie et al. parameters commonly used in clinical FES systems (Taylor
1979; Marsden et al. 1983; Garland et al. 1988; Binder- et al. 1999; Donaldson et al. 2000; Johnston et al. 2005).
Macleod et al. 1998), intensity (Binder-Macleod et al.
1995; Godfrey et al. 2002) and force generated in response
to electrical stimulation (Russ et al. 2002a,b) can affect Methods
the amount of muscle fatigue produced during repetitive
Subjects
stimulation, it is difficult to isolate the effects of stimulation
frequency versus intensity on muscle fatigue while Twelve healthy individuals (6 males and 6 females) aged
controlling for the force generated in response to electrical 22–30 years participated in the study. The subjects had
stimulation. For FES applications, however, because the no history of lower extremity orthopaedic, neurological
targeted force is determined by the task requirements, it or vascular problems. The subjects were requested to
may not be important to isolate the effects of frequency refrain from strenuous exercise for at least 48 h before
versus intensity on muscle fatigue and performance, the testing sessions. The subjects signed informed consent
but to determine the combination of frequency and forms approved by the Human Subjects Review Board


C 2006 The Authors. Journal compilation 
C 2006 The Physiological Society
Exp Physiol 91.6 pp 967–976 Muscle fatigue during repetitive stimulation 969

of the University of Delaware. The study was performed was used. A custom-made switch was connected in series
according to the Declaration of Helsinki. with the stimulator to control the pulse duration. A
custom-written LabVIEW program was used for data
acquisition, and to control the timing and the duration of
Apparatus and set-up the pulses.
The subjects were seated on an electromechanical
force dynamometer (KinCom III 500-11, Chattecx Corp.,
Experimental procedure
Chattanooga, TN, USA) with the back supported, hips
flexed approximately to 75 degrees and knees flexed at Each subject participated in four sessions, with a minimum
90 deg (Fig. 1). Velcro straps were used to stabilize the of 48 h separating consecutive sessions. At the start of the
subjects’ upper trunk, waist and thigh. Each subject’s first session, subjects received an overview of the testing
ankle was stabilized with a strap placed approximately procedures, signed the informed consent form, and were
5 cm proximal to the lateral malleolus. The isometric force trained to perform the MVIC test (Fig. 2). They were
output of the quadriceps femoris muscle was recorded via seated on the KinCom, surface electrodes were attached
a force transducer placed against the anterior aspect of to the skin of the subjects’ thigh and tested for appropriate
the lower leg, 5 cm proximal to the lateral malleolus. The placement. Next, the MVIC force was recorded using the
subjects could see a representation of the force recorded burst superimposition technique (Snyder-Mackler et al.
by the KinCom force transducer on a display screen. 1994). During the MVIC test, the subjects attempted
Electrical stimulation was delivered via two to produce as much knee extension force as possible.
self-adhesive surface electrodes (Versa-Stim, During the maximal voluntary contraction, an electrical
76 mm × 127 mm, CONMED Corp., New York, NY, stimulation train (amplitude, 130 V; frequency, 100 Hz;
USA). The proximal electrode was placed over the upper pulse duration, 600 μs) was delivered to the quadriceps
thigh, covering the proximal portion of the rectus femoris femoris muscle. This stimulation train or ‘burst’ was
and vastus lateralis muscles. The distal electrode was superimposed on the volitional contraction to ensure
placed over the lower aspect of the thigh, covering the that the subjects were truly generating maximal force.
vastus medialis and distal portion of the rectus femoris. If the electrical stimulation train increased the force by
The same experimenter placed electrodes across sessions, ≤ 10%, the subject’s MVIC was recorded. If the electrical
and care was taken to maintain consistency in electrode stimulation train increased the subject’s force output by
placement. A Grass S8800 stimulator (Grass Instrument
Co., Quincy, MA, USA) with a SIU8T stimulus isolation
Session 1
unit was used to deliver the electrical stimulation. A MVIC testing
personal computer equipped with a PCI-6024E data > 48-hours interval
acquisition board and a PCI-6602 counter-timer board Sessions 2, 3, and 4
Set stimulation amplitude to generate 50% MVIC peak force
(using 60-Hz trains at 600-μs pulse duration)

Muscle Potentiation

Determine frequency or pulse duration of


fatiguing protocol to be tested during session
Protocol #1 Protocol #2 Protocol #3
Freq: Varied PD: Varied PD: Varied
PD: 600-μs Freq: 30-Hz Freq: 60-Hz
5-minute rest
Repeat muscle potentiation

Pre-Fatigue Testing Trains


Rest time between trains = 10-second
60-Hz @ 20-Hz @ 60-Hz @ PD of 20-Hz @ PD of
600-μs 600-μs fatiguing train fatiguing train

Fatiguing Trains
176 fatiguing trains
Rest time: 700-ms
Protocol #1 Protocol #2 Protocol #3

Post-fatigue testing trains


Rest time between trains = 700-ms
60-Hz @ 20-Hz @ 60-Hz @ PD of 20-Hz @ PD of
600-μs 600-μs fatiguing train fatiguing train

Figure 1. Schematic representation of the experimental set-up Figure 2. Flow chart showing the experimental protocol for the
used for testing 4 sessions
F indicates the position of the force transducer. Please see text for details.


C 2006 The Authors. Journal compilation 
C 2006 The Physiological Society
970 T. Kesar and S. Binder-Macleod Exp Physiol 91.6 pp 967–976

> 10%, the MVIC test was repeated after a 10 min rest. Potentiation. Potentiation was repeated before the fatigue
If the subject failed to elicit a true MVIC within three test to prevent the effects of potentiation from interacting
repetitions, they were not tested on that day. For each with fatigue (Binder-Macleod et al. 2002).
subject, the MVIC value measured during the first session
was used to set the stimulation amplitude for all four Prefatigue testing trains. Sixty and 20 Hz trains were
sessions. delivered at the same pulse duration as that used for the
The remaining three sessions involved fatigue testing. fatiguing trains for that session and also at 600 μs pulse
The order of testing of the three fatigue protocols was duration. The prefatigue testing trains were delivered with
randomized across subjects. Only one protocol was tested a rest time of 10 s.
on each day. Each testing session consisted of ‘fatiguing
trains’ and pre- and postfatigue ‘testing trains’. The
Fatiguing trains. After the potentiation and prefatigue
‘fatiguing trains’, consisting of stimulation trains of three
testing trains had been delivered, 176 fatiguing trains were
different combinations of frequency and pulse duration
delivered at a rate of one train every second (train duration,
repetitively delivered at a rate of one train every second,
300 ms; rest time, 700 ms). Three different combinations
were used to fatigue the muscle and to assess the muscle’s
of frequency and pulse duration were used during the three
performance during the fatigue test. The ‘testing trains’,
different fatigue protocols, as follows:
consisting of stimulation trains of 60 and 20 Hz frequency
at two different pulse durations, were delivered before
(prefatigue) and after (postfatigue) the fatiguing trains. Protocol 1: Long pulse duration (600 μs) and low
The testing trains measured the decline in the force- frequency. The pulse duration of a 300 ms long train was
generating ability of the muscle. All fatiguing and testing fixed at 600 μs, and the frequency was set to produce 20%
trains were 300 ms long. The duty cycles used to fatigue MVIC peak force.
the muscles (300 ms long trains with 700 ms rest time)
were similar to the activation patterns recorded in the Protocol 2: Medium frequency (30 Hz) and medium pulse
quadriceps muscle during normal walking (Pierrynowski duration. The frequency of a 300 ms long train was fixed
& Morrison, 1985). Owing to differences in frequencies at 30 Hz, and the pulse duration was set to produce 20%
between the trains used for testing, however, the fatiguing MVIC peak force.
trains for each of the three protocols contained different
numbers of pulses. Protocol 3: High frequency (60 Hz) and short pulse
The following procedures were followed during the duration. The frequency of a 300 ms long train was fixed
three fatigue sessions (Fig. 2): at 60 Hz, and the pulse duration was set to produce 20%
MVIC peak force.
Set stimulation amplitude to generate 50% MVIC peak
Postfatigue testing trains. At the end of the fatiguing
force. During each testing session, after applying the
electrodes, the stimulation amplitude was set to produce trains, postfatigue testing trains were delivered in the same
50% of the subject’s MVIC force using 300 ms long, order as the prefatigue testing trains at a rate of one train
60 Hz trains with 600 μs pulse duration. The 50% every second to maintain the state of muscle fatigue.
MVIC amplitude was used because it allowed a range of
frequencies and pulse durations to be used to generate the Data analyses
target force of 20% MVIC.
The decline in force generated in response to the fatiguing
trains for each protocol was used as a measure of the
Potentiation. Eleven trains (770 ms train duration, 14 Hz ‘muscle’s performance’ or the muscle’s ability to maintain
frequency, 600 μs pulse duration) were delivered with a 5 s force output in response to the fatiguing trains. The
rest time between trains to potentiate the muscle (Binder- percentage declines in peak force between the first and last
Macleod et al. 2002). fatiguing train were calculated for each fatigue protocol.
Since muscle fatigue was the primary focus of our study,
it was important to define muscle fatigue in the context
Set frequency and pulse duration to generate 20% MVIC of this study. Muscle fatigue is a decline in the force-
target peak force. Either the stimulation pulse duration generating ability of the muscle as a result of recent activity
or frequency of a 300 ms long train was varied to generate (Edwards, 1981; Vollestad, 1997). We used the percentage
peak force equal to 20% of the subject’s MVIC. The first decline in peak force between pre- and postfatigue 60 Hz
train of each fatigue protocol generated ∼20% MVIC peak testing trains as a measure of the decline in the force-
force. generating ability (Vollestad, 1997). The decline in peak


C 2006 The Authors. Journal compilation 
C 2006 The Physiological Society
Exp Physiol 91.6 pp 967–976 Muscle fatigue during repetitive stimulation 971

Table 1. The sex (M = male, F = female), age, and maximal voluntary isometric contraction (MVIC) force for individual subjects; the
frequency and pulse durations (PD) used for repetitive stimulation during the 3 fatigue protocols are also listed
Protocol 1 Protocol 2 Protocol 3
Subject Sex Age (years) MVIC (N) PD (μs) Frequency (Hz) PD (μs) Frequency (Hz) PD (μs) Frequency (Hz)
1 M 22 1012 600 11.8 148 30 138 60
2 M 26 881 600 11.7 178 30 164 60
3 M 27 1117 600 14 112 30 106 60
4 M 24 1231 600 10.8 182 30 166 60
5 M 24 1335 600 12 155 30 139 60
6 M 30 1277 600 13.1 180 30 78 60
7 F 23 449 600 10 123 30 121 60
8 F 24 798 600 10.6 148 30 137 60
9 F 23 775 600 12.4 147 30 128 60
10 F 22 969 600 10.6 141 30 129 60
11 F 27 549 600 10.4 147 30 135 60
12 F 26 1000 600 10.3 145 30 137 60
Mean — 24.8 949.4 600 11.5 150 30 131 60
S.D. — 2.4 276.2 — 1.2 22 — 24 —

force of the 60 Hz testing trains at the same pulse duration Results


as the fatiguing trains provided a measure of the amount
of muscle fatigue produced in the population of motor Data were collected from 12 healthy individuals (6 males
units recruited by the fatiguing trains during each fatiguing and 6 females). All 12 subjects were successfully able to
protocol. The decline in peak force of the 60 Hz testing complete the MVIC testing during the first session in one
trains at the 600 μs pulse duration provided a measure of or two attempts. Table 1 provides detailed information
the muscle fatigue produced within a comparable number about the subjects’ age, sex, MVIC forces, and the
of motor units as were recruited during the protocol stimulation parameters of the fatiguing trains. The
that used the longest pulse duration (i.e. protocol 1). In frequency used for repetitive stimulation during protocol 1
addition, the ratio of peak forces produced in response was 11.5 ± 1.2 Hz. The pulse durations used for repetitive
to 20 versus 60 Hz testing trains (20 Hz:60 Hz peak force stimulation during protocols 2 and 3 were 150 ± 22 and
ratio) was used as a measure of low-frequency fatigue 131 ± 24 μs, respectively. Student’s paired t test confirmed
(Vollestad, 1997; Russ & Binder-Macleod, 1999). The that the pulse durations used for the fatigue protocols
20 Hz:60 Hz peak force ratios were calculated at the were significantly different (P < 0.05). The peak forces
beginning (prefatigue) and end (postfatigue) of the fatigue produced in response to the fatiguing trains during the
protocols both for testing trains at the pulse duration of three fatigue protocols for a representative subject are
the fatiguing trains and at the 600 μs pulse duration. shown in Fig. 3. It is notable that though the first fatiguing
trains of the three protocols produced similar peak forces
(Fig. 3A), the last fatiguing train of protocol 1 produced
Statistical analyses the highest and the last fatiguing train of protocol 3
produced the lowest peak force for this subject (Fig. 3B).
The percentage decline in peak forces from the first to The repeated measures ANOVA showed no significant
last fatiguing trains, percentage decline in peak force difference in the initial peak force produced by the three
between pre- and postfatigue 60 Hz testing trains at fatigue protocols (F = 1.09; n.s.). The average peak forces
the pulse duration of the fatiguing trains, and the in response to the first fatiguing trains were 21.2 ± 1.7,
percentage decline in 60 Hz testing trains at 600 μs 21.5 ± 2.2 and 22.3 ± 2.2% MVIC for protocols 1, 2 and 3,
pulse duration were compared using one-way repeated respectively.
measures ANOVAs. Pairwise post hoc comparisons using
the least squared difference (LSD) were performed
Force responses to the fatiguing trains
only if the ANOVA showed significant differences. The
pre- and postfatigue 20 Hz:60 Hz peak force ratios for There were significant differences in percentage decline
the three fatigue protocols were compared using two- in peak force (F = 30.08; P < 0.01) between the first
way (protocol × fatigue) repeated measures ANOVAs. and last fatiguing trains among the three protocols
In addition, peak forces produced in response to the (Fig. 4). Protocol 1, consisting of fatiguing trains with long
first fatiguing train of each protocol were compared (600 μs) pulse duration and low frequency, produced the
using a repeated-measures one-way ANOVA to determine smallest percentage decline in peak force (31.3 ± 9.4%),
whether the fatigue protocols produced similar initial peak and protocol 3, consisting of fatiguing trains with high
forces. The significance level was set at P = 0.05. frequency (60 Hz) and short pulse duration, produced


C 2006 The Authors. Journal compilation 
C 2006 The Physiological Society
972 T. Kesar and S. Binder-Macleod Exp Physiol 91.6 pp 967–976

A 1st Train Protocol #1 B Last Train Protocol #1


Protocol #2 Protocol #2
Protocol #3 Protocol #3
Force (N)

120 120

Force (N)
60 60

0 0
0 200 400 600 0 200 400 600 800
Time (ms) Time (ms)

C Fatiguing Trains Protocol #1


Protocol #2
180 Protocol #3
Peak Force (N)

Figure 3. Raw force profiles of the first (A)


and last fatiguing trains (B) of a
90
representative subject for the 3 fatigue
protocols
The figure also shows peak forces produced in
0 response to each fatiguing train during the
0 50 100 150 3 fatigue protocols for a representative subject
Contraction # (C).

the largest percentage decline in peak force (51.3 ± 7.5%; smallest decline in peak force (21.5 ± 9.5%), protocol 2
Fig. 4). produced an intermediate decline (27.4 ± 8.2%), and
protocol 3 produced the largest decline (46.1 ± 6.7%) in
peak force in response to the 60 Hz testing train at the same
Force responses of testing trains pulse duration as the fatiguing trains (F = 77.23; P < 0.01;
Fig. 5). Average peak forces produced in response to the
Average peak forces produced in response to the 60 Hz testing trains at the 600 μs pulse duration for
60 Hz testing trains at the same pulse duration as the protocols 1, 2 and 3, respectively, were 486.6 ± 160.9,
fatiguing trains for protocols 1, 2 and 3, respectively, 491.7 ± 142.4 and 489.2 ± 151.8 N (prefatigue) and
were 481.5 ± 164.2, 238.7 ± 69.2 and 221.1 ± 78.6 N 362.6 ± 101.4, 416.4 ± 112.0 and 376.3 ± 114.1 N
(prefatigue) and 367.5 ± 102.8, 170.4 ± 44.3 and (postfatigue). Interestingly, for the 60 Hz testing trains at
117.1 ± 36.8 N (postfatigue). Protocol 1 produced the 600 μs pulse duration, protocol 2, consisting of medium
frequency and medium pulse duration fatiguing trains,

Fatiguing Trains
75
*
*
% Decline in Peak Force

50
*

25

0
Protocol #1 Protocol #2 Protocol #3
Low-Freq
Low-Freq 30-Hz
Medium-Freq High-Freq
600-PD Medium-PD Short-PD

Figure 4. The percentage decline in peak force produced Figure 5. The percentage decline in peak forces between pre-
between the first and the last fatiguing trains during the and postfatigue 60 Hz testing trains at the same pulse duration
3 fatigue protocols as used for the fatiguing trains (left) and at 600 μs pulse
PD, pulse duration. ∗ Significant differences between protocols duration (right)
(P < 0.01). ∗ Significant difference between protocols (P < 0.01).


C 2006 The Authors. Journal compilation 
C 2006 The Physiological Society
Exp Physiol 91.6 pp 967–976 Muscle fatigue during repetitive stimulation 973

produced significantly smaller declines in peak force during muscle activation (Homsher, 1987; Baker et al.
(14.7 ± 7%) than protocol 1 (23.6 ± 8.3%) or protocol 3 1994; Frank et al. 1998; Hogan et al. 1998). Since protocol 1
(22.4 ± 10.3%; F = 6.40; P = 0.01; Fig. 5). used the fewest number of pulses and generated the
Average peak forces produced in response to the 20 Hz fewest action potentials, the least ATP utilization by the
testing trains at the pulse duration of the fatiguing trains Ca2+ -ATPase and Na+ –K+ -ATPase reactions also occurred
for protocols 1, 2 and 3, respectively, were 363.2 ± 117.7, during protocol 1 (Marsden et al. 1983; Frank et al. 1998).
175.0 ± 55.1 and 162.6 ± 53.9 N (prefatigue) and Since metabolic demand is related to muscle fatigue
224.7 ± 68.4, 73.6 ± 24.7 and 41.9 ± 17.1 N (postfatigue). (Cooke et al. 1988; Sahlin et al. 1998; Westerblad et al.
Average peak forces produced in response to the 20 Hz 1998), we believe that protocol 1 produced the least fatigue
testing trains at the 600 μs pulse duration for protocols 1, 2 in the motor unit population recruited by the fatiguing
and 3, respectively, were 366.2 ± 116.5, 375.7 ± 111.5 trains because the lowest ATP consumption and lowest
and 362.6 ± 110.1 N (prefatigue) and 221.4 ± 68.9, metabolic demand was placed on each muscle fibre during
253.0 ± 72.6 and 230.9 ± 70.3 N (postfatigue). For testing protocol 1 among the three protocols tested.
trains at the pulse duration of the fatiguing trains, the Recently, Godfrey et al. (2002) studied the effects of
two-way repeated measures ANOVA (protocol × fatigue) stimulation intensity on force production of thenar hand
showed significant effects of protocol (F = 32.00; muscles in patients with spinal cord injury. In contrast
P < 0.01) and fatigue (F = 308.93; P < 0.01) on the to our present findings, Godfrey et al. (2002) found
20 Hz:60 Hz peak force ratios (Fig. 6A). There was a greater fatigue during stimulation at supra- compared to
significant interaction between protocol and fatigue submaximal intensities. We believe that the differences
(F = 35.32; P < 0.01). There were no significant in the findings of Godfrey et al. (2002) versus our
differences in the prefatigue 20 Hz:60 Hz peak force ratios present study resulted from methodological differences.
among the three protocols. The postfatigue 20 Hz:60 Hz Since Godfrey et al. (2002) delivered both supra- and
peak force ratios showed significant differences among submaximal stimulation intensities at the same frequency
protocols (F = 50.7; P < 0.01). For testing trains at the (40 Hz), different force levels were generated at the two
pulse duration of the fatiguing trains, protocol 1 showed intensities. The higher forces generated at supra- versus
the largest postfatigue 20 Hz:60 Hz ratio (0.61 ± 0.07),
and protocol 3 showed the smallest ratio (0.37 ± 0.11;
Fig. 6A). For testing trains at 600 μs pulse duration, the
two-way ANOVA showed a significant effect of fatigue
(F = 66.95; P < 0.01), no significant effect of protocol
(F = 0.03; n.s.), and no significant interaction between the
effects of protocol and fatigue (F = 2.62; n.s.; Fig. 6B).

Discussion
Our study compared the performance and muscle
fatigue produced during repetitive stimulation using three
different combinations of frequencies and pulse durations,
and found that protocol 1 produced the least muscle
fatigue in the motor units recruited by the fatiguing trains
(see Fig. 5). Of the three protocols tested, protocol 1 used
the lowest frequency (11.5 ± 1.2 Hz) and highest level of
recruitment (600 μs pulse duration; Robinson & Snyder-
Mackler, 1995) to fatigue the muscle. Since protocol 1
recruited the most motor units, and all three protocols
generated the same target peak force, the least amount
of force was generated by each active muscle fibre during
protocol 1. Since ATP utilization by actin–myosin ATPase Figure 6. Ratios of peak forces produced in response to the pre-
is proportional to force generation (Boska, 1994; Potma and postfatigue 20 versus 60 Hz testing trains (20 Hz:60 Hz peak
et al. 1994), we suggest that protocol 1 resulted in the force ratio) at the same pulse duration as used for the fatiguing
least ATP utilization by the actin–myosin ATPase per trains (A) and at the 600 μs pulse duration (B) during the
3 fatigue protocols
muscle fibre. In addition to the actin–myosin ATPase, the ∗ Significant differences between fatigue protocols (P < 0.05). All
Ca2+ -ATPase and Na+ –K+ -ATPase reactions in response prefatigue 20 Hz:60 Hz peak force ratios were significantly different
to each action potential contribute to ATP utilization from the postfatigue ratios (P < 0.05).


C 2006 The Authors. Journal compilation 
C 2006 The Physiological Society
974 T. Kesar and S. Binder-Macleod Exp Physiol 91.6 pp 967–976

submaximal intensities probably contributed to the greater The present findings could have implications for the
muscle fatigue observed by Godfrey et al. (2002) at the development of strategies for optimal activation of skeletal
supramaximal intensities (Russ et al. 2002b,c). However, muscle during FES. Functional electrical stimulation is
the differences between the results of our present study used to generate functional movements in patients with
and those of Godfrey et al. (2002) may also result from upper motor neurone lesions, such as spinal cord injury,
differences in the subject populations and the muscle hemiplegia following stroke, and cerebral palsy. This
tested. study represents the first step in a project whose long-
Protocol 1 also produced the least low-frequency term aim is to develop electrical stimulation strategies
fatigue in the motor units recruited by the fatiguing that can maximize FES performance. Initial testing on
trains (Fig. 6A). Low-frequency fatigue is the result of healthy individuals has helped us to identify hypotheses
impairment in excitation–contraction coupling that is that can then be tested on paralysed muscles using fewer
thought to result from increased levels of intracellular experimental sessions. Interestingly, consistent with out
Ca2+ during muscle activation (Westerblad et al. 1993; present findings, recent studies showed that for healthy
Chin & Allen, 1996; Chin et al. 1997). Intracellular Ca2+ subjects and for subjects with spinal cord injury, starting
concentrations have been shown to be directly related to at low frequencies and later switching to high frequencies
the stimulation frequency (Chin & Allen, 1996). The low produced better performance during repetitive non-
frequency used during protocol 1 would therefore result isometric contractions than stimulation using either a
in the lowest levels of intracellular Ca2+ and the least low- low or high frequency alone (Kebaetse & Binder-Macleod,
frequency fatigue among the three protocols (Westerblad 2004; Kebaetse et al. 2005). Starting repetitive stimulation
et al. 1993; Chin & Allen, 1996; Chin et al. 1997). with low frequencies produced less muscle fatigue, and
In contrast to our present findings, Matsunaga et al. switching to a higher stimulation frequency allowed the
(1999) showed smaller declines in force during repetitive stimulation to overcome the effects of low-frequency
stimulation at high- (100 Hz) versus low-frequency fatigue (Kebaetse & Binder-Macleod, 2004; Kebaetse et al.
(20 Hz) activation. Compared to our study, the fatigue 2005). Future studies will need to identify the best
protocols tested by Matsunaga et al. (1999) were of much frequency and intensity of the initial trains, as well as the
longer durations (60 min versus 180 s in our study), used strategies for modulation of frequency and intensity of
much shorter duty cycles (1:15, 1:30 and 1:60 versus 1:2.3 the subsequent trains, to maximize muscle performance
in our study), and produced smaller percentage declines during FES.
in peak force (22.3 ± 15.1% at 100 Hz versus 51.3 ± 7.5% An interesting finding of this study was the difference
at 60 Hz in our study). Since muscle performance is a in the responses to the testing trains when the pulse
function of the extent of force fatigue and low-frequency duration was maintained at the levels used to fatigue
fatigue produced by the fatiguing trains in the motor unit the muscle versus when a 600 μs pulse duration was
population recruited by the fatiguing trains, we believe that used (see Figs 5 and 6). During protocols 2 and 3, the
the results of Matsunaga et al. (1999) can be explained by postfatigue testing trains at 600 μs pulse duration activated
greater low-frequency fatigue and less muscle fatigue than previously unrecruited motor units. The responses to the
presently observed. That is, even if greater muscle fatigue testing trains at 600 μs pulse duration were therefore
was produced by the 100 Hz trains than the 20 Hz trains in the sum of the forces produced by the recruited and
their study (Matsunaga et al. 1999), low-frequency fatigue previously unrecruited motor units. Protocol 3 showed
may have markedly reduced the muscles’ responses to the the most and protocol 1 the least muscle fatigue and low-
20 Hz trains and therefore resulted in poorer performance frequency fatigue when the pulse duration was maintained
in response to the 20 Hz trains. at the level used to fatigue the muscle (Fig. 5). Thus,
The frequencies used during protocols 1, 2 and 3 in our during FES applications, if the frequency and pulse
study were on the low, middle and high ranges of the force– duration are held constant during repetitive stimulation,
frequency curves, respectively. Surprisingly, however, using the lowest frequency and longest pulse duration
the average pulse durations used during protocols 2 may maximize performance. However, Protocols 3 and
(150 ± 22 μs) and 3 (131 ± 24 μs), although significantly 1 showed comparable amounts of muscle fatigue and
different, only varied by 19 μs. A possible reason could protocol 2 showed the least muscle fatigue when tested at
be that the pulse durations used for protocols 2 and 3 the 600 μs pulse duration (Fig. 5). In addition, in response
were in the steep rising part of the force versus pulse to testing trains at 600 μs pulse duration, there were no
durations curves. Thus, our results showed that although differences in the overall levels of low-frequency fatigue
the difference in frequencies between protocols 2 and 3 among the three protocols (Fig. 6B). This is an important
was relatively large (30 versus 60 Hz), a relatively small finding because most FES systems used at present deliver
difference in pulse duration between protocol 3 and a constant frequency and increase the intensity to increase
protocol 2 enabled the targeted 20% MVIC force to be muscle force output as the muscle fatigues (Weingarden
reached for both the protocols. et al. 1997; Taylor et al. 1999; Weber et al. 2005). Thus,


C 2006 The Authors. Journal compilation 
C 2006 The Physiological Society
Exp Physiol 91.6 pp 967–976 Muscle fatigue during repetitive stimulation 975

for FES applications where intensity is increased as the Boska M (1994). ATP production rates as a function of force
muscle fatigues, a ‘medium’ frequency, similar to the level in the human gastrocnemius/soleus using 31 P MRS.
frequency used in protocol 2 of our study, may minimize Magn Reson Med 32, 1–10.
muscle fatigue. The combination of frequency and pulse Chin ER & Allen DG (1996). The role of elevations in
duration that minimizes muscle fatigue and/or maximizes intracellular [Ca2+ ] in the development of low frequency
fatigue in mouse single muscle fibres. J Physiol 491, 813–824.
performance during FES may depend on whether or not
Chin ER, Balnave CD & Allen DG (1997). Role of intracellular
modulation of frequency or intensity will be used as calcium and metabolites in low-frequency fatigue of mouse
strategies during repetitive stimulation. skeletal muscle. Am J Physiol 272, C550–C559.
Cooke R, Franks K, Luciani GB & Pate E (1988). The inhibition
of rabbit skeletal muscle contraction by hydrogen ions and
Conclusion phosphate. J Physiol 395, 77–97.
The present findings support the hypothesis that when Davoodi R, Andrews BJ, Wheeler GD & Lederer R (2002).
Development of an indoor rowing machine with manual FES
the same initial peak force is generated using different
controller for total body exercise in paraplegia. IEEE Trans
combinations of frequency and pulse duration, and when Neural Syst Rehabil Eng 10, 197–203.
the frequency and pulse duration are kept constant Donaldson N, Perkins TA, Fitzwater R, Wood DE & Middleton
throughout repetitive stimulation, repetitive stimulation F (2000). FES cycling may promote recovery of leg function
with a long pulse duration (600 μs) and low frequency after incomplete spinal cord injury. Spinal Cord 38, 680–682.
(11.5 ± 1.2 Hz; protocol 1) would maximize isometric Edwards RH (1981). Human muscle function and fatigue. Ciba
performance by minimizing muscle fatigue. Interestingly, Found Symp 82, 1–18.
repetitive stimulation with a medium frequency (30 Hz) Frank K, Bolck B, Bavendiek U & Schwinger RH (1998).
and medium pulse duration (150 ± 22 μs; protocol 2) Frequency dependent force generation correlates with
produced the least muscle fatigue when comparable sarcoplasmic calcium ATPase activity in human
motor unit populations were tested across protocols myocardium. Basic Res Cardiol 93, 405–411.
Garland SJ, Garner SH & McComas AJ (1988). Relationship
(i.e. at the 600 μs pulse duration). These findings may
between numbers and frequencies of stimuli in human
have important implications for the design of optimal muscle fatigue. J Appl Physiol 65, 89–93.
stimulation strategies to use during FES. Specifically, the Gerrits HL, De Haan A, Hopman MT, van Der Woude LH,
present results should help in the design of future studies Jones DA & Sargeant AJ (1999). Contractile properties of the
involving patient populations where complex stimulation quadriceps muscle in individuals with spinal cord injury.
strategies that modulate both the frequency and pulse Muscle Nerve 22, 1249–1256.
durations will be tested. Gerrits HL, Hopman MT, Offringa C, Engelen BG, Sargeant
AJ, Jones DA & Haan A (2003). Variability in fibre properties
in paralysed human quadriceps muscles and effects of
References training. Pflugers Arch 445, 734–740.
Gfohler M & Lugner P (2000). Cycling by means of functional
Agarwal S, Kobetic R, Nandurkar S & Marsolais EB (2003). electrical stimulation. IEEE Trans Rehabil Eng 8, 233–243.
Functional electrical stimulation for walking in paraplegia: Godfrey S, Butler JE, Griffin L & Thomas CK (2002).
17-year follow-up of 2 cases. J Spinal Cord Med 26, 86–91. Differential fatigue of paralyzed thenar muscles by stimuli of
Baker AJ, Brandes R, Schendel TM, Trocha SD, Miller RG & different intensities. Muscle Nerve 26, 122–131.
Weiner MW (1994). Energy use by contractile and Grill WM Jr & Mortimer JT (1996). The effect of stimulus pulse
noncontractile processes in skeletal muscle estimated by duration on selectivity of neural stimulation. IEEE Trans
31
P-NMR. Am J Physiol 266, C825–C831. Biomed Eng 43, 161–166.
Bigland-Ritchie B, Jones DA & Woods JJ (1979). Excitation Hogan MC, Ingham E & Kurdak SS (1998). Contraction
frequency and muscle fatigue: electrical responses during duration affects metabolic energy cost and fatigue in skeletal
human voluntary and stimulated contractions. Exp Neurol muscle. Am J Physiol 274, E397–E402.
64, 414–427. Homsher E (1987). Muscle enthalpy production and its
Binder-Macleod SA, Dean JC & Ding J (2002). Electrical relationship to actomyosin ATPase. Annu Rev Physiol 49,
stimulation factors in potentiation of human quadriceps 673–690.
femoris. Muscle Nerve 25, 271–279. Hunt KJ, Stone B, Negard NO, Schauer T, Fraser MH, Cathcart
Binder-Macleod SA, Halden EE & Jungles KA (1995). Effects of AJ, Ferrario C, Ward SA & Grant S (2004). Control strategies
stimulation intensity on the physiological responses of for integration of electric motor assist and functional
human motor units. Med Sci Sports Exerc 27, 556–565. electrical stimulation in paraplegic cycling: utility for
Binder-Macleod SA, Lee SC, Russ DW & Kucharski LJ (1998). exercise testing and mobile cycling. IEEE Trans Neural Syst
Effects of activation pattern on human skeletal muscle Rehabil Eng 12, 89–101.
fatigue. Muscle Nerve 21, 1145–1152. Johnston TE, Betz RR, Smith BT, Benda BJ, Mulcahey MJ,
Binder-Macleod SA & Snyder-Mackler L (1993). Muscle Davis R, Houdayer TP, Pontari MA, Barriskill A & Creasey
fatigue: clinical implications for fatigue assessment and GH (2005). Implantable FES system for upright mobility and
neuromuscular electrical stimulation. Phys Ther 73, bladder and bowel function for individuals with spinal cord
902–910. injury. Spinal Cord 43, 713–723.

C 2006 The Authors. Journal compilation 
C 2006 The Physiological Society
976 T. Kesar and S. Binder-Macleod Exp Physiol 91.6 pp 967–976

Kebaetse MB & Binder-Macleod SA (2004). Strategies that Russ DW, Vandenborne K & Binder-Macleod SA (2002b).
improve human skeletal muscle performance during Factors in fatigue during intermittent electrical stimulation
repetitive, non-isometric contractions. Pflugers Arch 448, of human skeletal muscle. J Appl Physiol 93, 469–478.
525–532. Russ DW, Vandenborne K, Walter GA, Elliott M &
Kebaetse MB, Lee SC, Johnston TE & Binder-Macleod SA Binder-Macleod SA (2002c). Effects of muscle activation on
(2005). Strategies that improve paralyzed human quadriceps fatigue and metabolism in human skeletal muscle. J Appl
femoris muscle performance during repetitive, nonisometric Physiol 92, 1978–1986.
contractions. Arch Phys Med Rehabil 86, 2157–2164. Sahlin K, Tonkonogi M & Soderlund K (1998). Energy supply
Kralj A, Bajd T & Turk R (1988). Enhancement of gait and muscle fatigue in humans. Acta Physiol Scand 162,
restoration in spinal injured patients by functional electrical 261–266.
stimulation. Clin Orthop Relat Res 233, 34–43. Snyder-Mackler L, De Luca PF, Williams PR, Eastlack ME &
Liberson WT, Holmquest HJ, Scot D & Dow M (1961). Bartolozzi AR 3rd (1994). Reflex inhibition of the quadriceps
Functional electrotherapy: stimulation of the peroneal nerve femoris muscle after injury or reconstruction of the anterior
synchronized with the swing phase of the gait of hemiplegic cruciate ligament. J Bone Joint Surg Am 76, 555–560.
patients. Arch Phys Med Rehabil 42, 101–105. Taylor PN, Burridge JH, Dunkerley AL, Wood DE, Norton JA,
Mangold S, Keller T, Curt A & Dietz V (2005). Transcutaneous Singleton C & Swain ID (1999). Clinical use of the Odstock
functional electrical stimulation for grasping in subjects with dropped foot stimulator: its effect on the speed and effort of
cervical spinal cord injury. Spinal Cord 43, 1–13. walking. Arch Phys Med Rehabil 80, 1577–1583.
Marsden CD, Meadows JC & Merton PA (1983). ‘Muscular Triolo RJ, Bieri C, Uhlir J, Kobetic R, Scheiner A & Marsolais
wisdom’ that minimizes fatigue during prolonged effort in EB (1996). Implanted functional neuromuscular stimulation
man: peak rates of motoneuron discharge and slowing of systems for individuals with cervical spinal cord injuries:
discharge during fatigue. Adv Neurol 39, 169–211. clinical case reports. Arch Phys Med Rehabil 77, 1119–1128.
Marsolais EB & Edwards BG (1988). Energy costs of walking Vollestad NK (1997). Measurement of human muscle fatigue.
and standing with functional neuromuscular stimulation J Neurosci Meth 74, 219–227.
and long leg braces. Arch Phys Med Rehabil 69, 243–249. Weber DJ, Stein RB, Chan KM, Loeb G, Richmond F, Rolf R,
Matsunaga T, Shimada Y & Sato K (1999). Muscle fatigue from James K & Chong SL (2005). BIONic WalkAide for correcting
intermittent stimulation with low and high frequency foot drop. IEEE Trans Neural Syst Rehabil Eng 13, 242–246.
electrical pulses. Arch Phys Med Rehabil 80, 48–53. Weingarden HP, Kizony R, Nathan R, Ohry A & Levy H (1997).
Peckham PH & Knutson JS (2005). Functional electrical Upper limb functional electrical stimulation for walker
stimulation for neuromuscular applications. Annu Rev ambulation in hemiplegia: a case report. Am J Phys Med
Biomed Eng 7, 327–360. Rehabil 76, 63–67.
Pierrynowski MR & Morrison JB (1985). Estimating the muscle Westerblad H, Allen DG, Bruton JD, Andrade FH &
forces generated in the human lower extremity when Lannergren J (1998). Mechanisms underlying the reduction
walking: a physiological solution. Math Biosci 75, of isometric force in skeletal muscle fatigue. Acta Physiol
43–68. Scand 162, 253–260.
Potma EJ, Stienen GJ, Barends JP & Elzinga G (1994). Westerblad H, Duty S & Allen DG (1993). Intracellular calcium
Myofibrillar ATPase activity and mechanical performance of concentration during low-frequency fatigue in isolated single
skinned fibres from rabbit psoas muscle. J Physiol 474, fibers of mouse skeletal muscle. J Appl Physiol 75, 382–388.
303–317. Wheeler GD, Andrews B, Lederer R, Davoodi R, Natho K,
Riener R (1999). Model-based development of neuroprosthesis Weiss C, Jeon J, Bhambhani Y & Steadward RD (2002).
for paraplegic patients. Philos Trans R Soc Lond B Biol Sci Functional electric stimulation-assisted rowing: increasing
354, 877–894. cardiovascular fitness through functional electric stimulation
Ring H & Rosenthal N (2005). Controlled study of rowing training in persons with spinal cord injury. Arch Phys
neuroprosthetic functional electrical stimulation in Med Rehabil 83, 1093–1099.
sub-acute post-stroke rehabilitation. J Rehabil Med 37,
32–36.
Robinson AJ & Snyder-Mackler L (1995). Clinical
Electrophysiology, Electrotherapy and Electrophysiologic
Testing. Williams & Wilkins, Baltimore. Acknowledgements
Russ DW & Binder-Macleod SA (1999). Variable-frequency
trains offset low-frequency fatigue in human skeletal muscle. The authors would like to thank R. Perumal and R. Maladen
Muscle Nerve 22, 874–882. for their helpful comments on an early draft of this manuscript
Russ DW, Elliott MA, Vandenborne K, Walter GA & and for the development of the software and hardware for data
Binder-Macleod SA (2002a). Metabolic costs of force acquisition. Also, we thank Li-Wei Chou for his assistance and
generation and maintenance of human skeletal muscle. suggestions for the design of the study. This work was supported
Am J Physiol Endocrinol Metab 282, E448–E457. by NIH grants no. HD36797 and no. HD38582.


C 2006 The Authors. Journal compilation 
C 2006 The Physiological Society

You might also like