MEASUREMENT UNCERTAINTY CALCULATIONS ( FOR MASS OF ZINC COATING)

Sources of Uncertainty:
1.Uncertainty due to Repeatability of Test (UA)
2. Uncertainty due to the Analytical Balance (UB1)
3. Uncertainty due to vernier calipers (UB2)
4. Combined Uncertainty (Uc)
5.Expanded Uncertainty (Uexp)
Calculation of Type 'A" Uncertainty, U A which is due to Repeatability Test (Random Error)

Std.
Reading Mean (X) Deviation (Xi-X)² Uncertainty
Deviation
Sr no
σ - √∑ (Xi-
(Xi) Value (Xi-X) UA
X)² n-1
1 1117.95 1209.8500 -91.9000 8445.61000000 74.4198 33.2815443
2 1164.07 -91.9000 8445.61000000
3 1205.15 -4.7000 22.09000000
4 1308.00 98.1500 9633.42250000
5 1254.08 44.2300 1956.29290000
Caculation of Type "B" Uncertainty which is due to Systematic Errors - U B
C4 a UB1 = Uncertainty associated with CRM used. (CRM 116/04 )
UB1 = 0.0003 (from Calibration Certificate at 95% CL with a coverage factor, k=2)
The Standard Uncertainty =UB1/2 (Assuming normal distribution)

UB1 = 0.00015
Degree of Freedom is = ∞
b. UB2 = Uncertainty due to the Instrument used (SPECTROLAB LAV M11)
UB2 = 0.0001 (Detection limit of Pb)
The Standard Uncertainty = UB2/√3 (Assuming Rectangular distribution)
UB2 = 0.000058
Degree of Freedom is = ∞
UC = √ (UA)² + (UB1)² + (UB2)²
33.28154429
Expanded Uncertainty, UEXP = k x UC = 66.563089
The reported expanded Uncertainty in measurement is stated as the standard Uncertainty in measurment multiplied by coverage factor,k=2 at 95%CL
Remark: The measured Uncertainty calculated for mass of Zinc Coating at 1209.85 is ± 66.56gms
MEASUREMENT UNCERTAINTY CALCULATIONS ( FOR CARBON IN CARBON STEEL)
Sources of Uncertainty:
1.Uncertainty due to Repeatability of Test (UA)
2. Uncertainty due to the Analytical Balance (UB1)
3. Uncertainty due to vernier calipers (UB2)
4. Combined Uncertainty (Uc)
5.Expanded Uncertainty (Uexp)
Calculation of Type 'A" Uncertainty, U A which is due to Repeatability Test (Random Error)

Reading Mean (X) Deviation (Xi-X)² Std. Deviation Uncertainty


Sr no
σ - √∑ (Xi-X)² n-
(Xi) Value (Xi-X) UA
1
1 0.143 0.1484 0.0053548387 2.86742976067E-05 0.02213074541 0.003974799
2 0.163 -0.014645161 0.00021448074922
3 0.195 -0.046645161 2.17577107180E-03
4 0.169 -0.020645161 0.000426222684703
5 0.128 0.0203548387 4.14319458897E-04
6 0.127 0.0213548387 0.000456029136316
7 0.136 0.0123548387 1.52642039542E-04
8 0.150 -0.002 2.706555671176E-06
9 0.156 -0.008 5.84484911550E-05
10 0.150 -0.002 2.706555671176E-06
11 0.170 -0.022 4.68513007284E-04
12 0.111 0.037 0.001395383975026
13 0.149 -0.001 4.16233090531E-07
14 0.115 0.0333548387 0.001112545265349
15 0.177 -0.177 3.13290000000E-02
16 0.119 0.0293548387 0.000861706555671
17 0.128 0.0203548387 4.14319458897E-04
18 0.143 0.0053548387 2.867429760666E-05
19 0.164 -0.015645161 2.44771071800E-04
20 0.121 0.0273548387 0.000748287200832
21 0.148 0.0003548387 1.25910509886E-07
22 0.144 0.0043548387 1.89646201873E-05
23 0.164 -0.015645161 2.44771071800E-04
24 0.146 0.0023548387 5.545265348595E-06
25 0.131 0.0173548387 3.01190426639E-04
26 0.183 -0.034645161 0.001200287200832
27 0.119 0.0293548387 8.61706555671E-04
28 0.153 -0.005 2.157752341311E-05
29 0.188 -0.040 1.57173881374E-03
30 0.168 -0.020 0.000385932362123
31 0.141 0.007 5.40936524454E-05
Degree of freedom=(n-1)=31-1 30.000
Caculation of Type "B" Uncertainty which is due to Systematic Errors - U B
UB1 = Uncertainty associated with the Analitycal Balance used:
0.36mg (from Calibration Certificate at 95% CL with a coverage factor, k=2)
The Standard Uncertainty =UB1/2 (Assuming normal distribution)
UB1 = 0.36/2 0.18 mg ~ 0.00018 g
UB2 = Uncertainty associated with the vernier caliper used used:
0.16μm ~ 00016mm (from Calibration Certificate at 95% CL with a coverage factor, k=2)
UB2 = 0.00016/2 0.00008mm ~ 0.000000008 m (Assuming normal distribution)
UC = √ (UA)² + (UB1)² + (UB2)²
0.00018
Expanded Uncertainty, UEXP = k x UC = 0.000360
The reported expanded Uncertainty in measurement is stated as the standard Uncertainty in measurment multiplied by coverage fact
Remark: The measured Uncertainty calculated for mass of Zinc Coating at 1209.85 is ± 66.56gms