01 Lectura Caracteristicas Lab

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 30

Analytical Techniques

Basic Principles and


CHAPTER

and Intrumentation
Practices
Alan H. B. Wu
Eileen Carreiro-Lewandowski
Roberta A. Martindale

C H A P T E R O U T L I N E

■ UNITS OF MEASURE ■ LABORATORY MATHEMATICS AND


Electronic Reporting of Results CALCULATIONS
■ REAGENTS Significant Figures
Chemicals Logarithms
Reference Materials Concentration
Water Specifications Dilutions
Solution Properties Water of Hydration
■ CLINICAL LABORATORY SUPPLIES Graphing and Beer’s Law
Thermometers/Temperature ■ SPECIMEN CONSIDERATIONS
Glassware and Plasticware Types of Samples
Desiccators and Desiccants Sample Processing
Balances Sample Variables
■ BASIC SEPARATION TECHNIQUES Chain of Custody
Centrifugation ■ REFERENCES
Filtration
Dialysis

T he primary purpose of a clinical chemistry labora-


tory is to facilitate the correct performance of ana-
lytic procedures that yield accurate and precise infor-
International d’Unités (SI), adopted internationally in
1960, is preferred in scientific literature and clinical
laboratories and is the only system used in many coun-
mation, aiding patient diagnosis and treatment. The tries. This system was devised to provide the global sci-
achievement of reliable results requires that the clinical entific community a uniform method of describing
laboratory scientist be able to correctly use basic sup- physical quantities. The SI system units (referred to as
plies and equipment and possess an understanding of SI units) are based on the metric system. Several sub-
fundamental concepts critical to any analytic proce- classifications exist within the SI system, one of which
dure. The topics in this chapter include units of meas- is the basic unit. There are seven basic units (Table 1-1),
ure, basic laboratory supplies, and introductory labora- with length (meter), mass (kilogram), and quantity of a
tory mathematics, plus a brief discussion of specimen substance (mole) being the units most frequently en-
collection and processing. countered. Another set of SI-recognized units is termed
derived units. A derived unit, as the name implies, is a
derivative or a mathematical function of one of the
UNITS OF MEASURE
basic units. An example of an SI-derived unit is meters
Any meaningful quantitative laboratory result consists of per second (m/s), used to express velocity. However,
two components. The first component represents the ac- some non-SI units are so widely used that they have be-
tual value, and the second is a label identifying the units come acceptable for use with SI basic or SI-derived
of the expression. The number describes the numeric units (Table 1-1). These include certain long-standing
value, whereas the unit defines the physical quantity or units such as hour, minute, day, gram, liter, and plane
dimension, such as mass, length, time, or volume.1 angles expressed as degrees. These units, although
Although several systems of units have traditionally widely used, cannot technically be categorized as either
been used by various scientific divisions, the Système basic or derived SI units.

3
4 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

TABLE 1-1 SI UNITS


BASE QUANTITY NAME SYMBOL
Length Meter m
Mass Kilogram kg
Time Second s
Electric current Ampere A
Thermodynamic temperature Kelvin K
Amount of substance Mole mol
Luminous intensity Candela cd
SELECTED DERIVED
Frequency Hertz Hz
Force Newton N
Celsius temperature Degree Celsius °C
Catalytic activity Katal kat
SELECTED ACCEPTED NON-SI
Minute (time) (60s) min
Hour (3,600s) h
Day (86,400s) d
3 3 3
Liter (volume) (1 dm 10 m ) L
10
Angstrom (0.1 nm 10 m) Å

The SI uses standard prefixes that, when added to a


given unit, can indicate decimal fractions of multiples of TABLE 1-2 PREFIXES USED WITH SI UNITS
that unit (Table 1-2). For example, 0.001 liter can be ex-
FACTOR PREFIX SYMBOL
pressed using the prefix milli, becoming 1 milliliter,
18
which is then written as 1 mL. Note that the SI term for 10 atto a
15
mass is kilogram; it is the only basic unit that contains a 10 femto f
prefix as part of its naming convention. Generally, the 10 12
pico p
standard prefixes for mass use the term gram rather than 9
10 nano n
kilogram.
6
Reporting of laboratory results is often expressed in 10 micro
3
terms of substance concentration (e.g., moles) or the 10 milli m
mass of a substance (e.g., mg/dL, g/dL, g/L, mEq/L, and 10 2
centi c
IU) rather than in SI units. These familiar and tradi- 1
10 deci d
tional units can cause confusion during interpretation.
1
It has been recommended that analytes be reported 10 deka da
2
using moles of solute per volume of solution (substance 10 hecto h
concentration) and that the liter be used as the refer- 10 3
kilo k
ence volume.2 Appendix D, Conversion of Traditional
104 mega M
Units to SI Units for Common Clinical Chemistry
9
Analytes, lists both reference and SI units together with 10 giga G
12
the conversion factor from traditional to SI units for 10 tera T
common analytes. As with other areas of industry, the 1015 peta P
laboratory and the rest of medicine is moving toward 18
10 exa E
adopting universal standards promoted by the
International Organization for Standards, often re- Prefixes are used to indicate a subunit or multiple of a basic SI unit.
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 5

ferred to as ISO. This group develops standards of prac- edge of chemicals, standards, solutions, buffers, and water
tice, definitions, and guidelines that can be adopted by requirements is necessary.
everyone in a given field, providing for more uniform ter-
minology and less confusion. As with any change, this Chemicals
process can take time and clinical laboratory scientists
must be familiar with the many terms currently in use. Analytic chemicals exist in varying grades of purity: ana-
lytic reagent (AR); ultrapure, chemically pure (CP);
United States Pharmacopeia (USP); National Formulary
Electronic Reporting of Results (NF); and technical or commercial grade.3 A committee
Electronic transmission of laboratory data and the more of the American Chemical Society (ACS; www.acs.org)
routine use of an electronic medical record, coding, established specifications for AR grade chemicals, and
billing, and other data management systems have caused chemical manufacturers will either meet or exceed these
much debate regarding appropriate standards needed in requirements. Labels on reagents state the actual impuri-
terms of both reporting guidelines and safeguards to en- ties for each chemical lot or list the maximum allowable
sure privacy of the data and records. Complicating matters impurities. The labels should be clearly printed with the
is that there are many different data management systems percentage of impurities present and either the initials
in use by health care agencies that all use laboratory infor- AR or ACS or the term For laboratory use or ACS
mation. For example, the Logical Observation Identifiers Standard-Grade Reference Materials. Chemicals of this
Names and Codes (LOINC) system, International category are suitable for use in most analytic laboratory
Federation of Clinical Chemistry/International Union of procedures. Ultrapure chemicals have been put through
Pure and Applied Chemistry (IFCC/IUPAC), American additional purification steps for use in specific proce-
Society for Testing and Materials (ASTM), Health Level 7 dures such as chromatography, atomic absorption, im-
(HL7), and Systematized Nomenclature of Medicine, munoassays, molecular diagnostics, standardization, or
Reference Technology (SNOMED RT) are databases that other techniques that require extremely pure chemicals.
use their own coding systems for laboratory observations. These reagents may carry designations of HPLC (high-
There are also additional proprietary systems in use, performance liquid chromatography) or chromato-
adding to the confusion. While no uniform standard ex- graphic (see later) on their labels.
ists, there is agreement in most systems that each test Because USP and NF grade chemicals are used to man-
name should be clearly identified and should have its ufacture drugs, the limitations established for this group of
unique code and should include the value and the unit or chemicals are based only on the criterion of not being in-
the appropriate abbreviation. Reference ranges, for most jurious to individuals. Chemicals in this group may be
clinical chemistry tests, are not universal and in those in- pure enough for use in most chemical procedures; how-
stances need to be included for proper interpretation. ever, it should be recognized that the purity standards are
not based on the needs of the laboratory and, therefore,
may or may not meet all assay requirements.
REAGENTS
Reagent designations of CP or pure grade indicate that
In today’s highly automated laboratory, there seems to be the impurity limitations are not stated and that prepara-
little need for reagent preparation by the clinical labora- tion of these chemicals is not uniform. Melting point
tory scientist. Most instrument manufacturers also make analysis is often used to ascertain the acceptable purity
the reagents, usually in a readily available “kit” form (i.e., range. It is not recommended that clinical laboratories
all necessary reagents and respective storage containers use these chemicals for reagent preparation unless fur-
are prepackaged as a unit) or requiring only the addition ther purification or a reagent blank is included.
of water or buffer to the prepackaged reagent components. Technical or commercial grade reagents are used prima-
A heightened awareness of the hazards of certain chemi- rily in manufacturing and should never be used in the
cals and numerous regulatory agency requirements have clinical laboratory.
caused clinical chemistry laboratories to readily eliminate Organic reagents also have varying grades of purity
massive stocks of chemicals and opt instead for the ease of that differ from those used to classify inorganic reagents.
using prepared reagents. Periodically, especially in hospi- These grades include a practical grade with some impu-
tal laboratories involved in research and development, rities; chemically pure, which approaches the purity level
biotechnology applications, specialized analyses, or of reagent grade chemicals; spectroscopic (spectrally
method validation, the laboratorian may still face prepar- pure) and chromatographic (minimum purity of 99% de-
ing various reagents or solutions. As a result of reagent termined by gas chromatography) grade organic
deterioration, supply and demand, or the institution of reagents, with purity levels attained by their respective
cost-containment programs, the decision may be made to procedures; and reagent grade (ACS), which is certified
prepare reagents in-house. Therefore, a thorough knowl- to contain impurities below certain levels established by
6 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

the ACS. As in any analytic method, the desired organic ondary standards will list the SRM or primary standard
reagent purity is dictated by the particular application. used for comparison. This information may be needed
Other than the purity aspects of the chemicals, laws re- during laboratory accreditation processes.
lated to the Occupational Safety and Health Administration
(OSHA)4 require manufacturers to clearly indicate the
Water Specifications11
lot number, plus any physical or biologic health hazard
and precautions needed for the safe use and storage of Water is the most frequently used reagent in the labora-
any chemical. A manufacturer is required to provide tory. Because tap water is unsuitable for laboratory ap-
technical data sheets for each chemical manufactured on plications, most procedures, including reagent and stan-
a document called a Material Safety Data Sheet (MSDS). dard preparation, use water that has been substantially
A more detailed discussion of this topic may be found in purified. Water solely purified by distillation results in
Chapter 3. distilled water; water purified by ion exchange produces
deionized water. Reverse osmosis, which pumps water
across a semipermeable membrane, produces RO water.
Reference Materials
Water can also be purified by ultrafiltration, ultraviolet
Unlike other areas of chemistry, clinical chemistry is in- light, sterilization, or ozone treatment. Laboratory re-
volved in the analysis of biochemical byproducts found quirements generally call for reagent grade water that,
in biological fluids, such as serum, plasma, or urine, mak- according to the Clinical and Laboratory Standards
ing purification and a known exact composition of the Institute, is classified into one of six categories based on
material almost impossible. For this reason, traditionally the specifications needed for its use rather than the
defined standards used in analytical chemistry standards method of purification or preparation.12 These categories
do not readily apply in clinical chemistry. include clinical laboratory reagent water (CLRW), spe-
Recall that a primary standard is a highly purified cial reagent water (SRW), instrument feed water, water
chemical that can be measured directly to produce a sub- supplied by method manufacturer, autoclave and wash
stance of exact known concentration and purity. The ACS water, and commercially bottled purified water.
purity tolerances for primary standards are 100 0.02%. Laboratories need to assess whether the water meets the
Because most biologic constituents are unavailable within specifications needed for its application. Most water pa-
these limitations, the National Institute of Standards and rameters include at least microbiological count, pH, re-
Technology (NIST; http://ts.nist.gov)-certified standard sistivity, silicate, particulate matter, and organics. Each
reference materials (SRMs) are used instead of ACS pri- category has a specific acceptable limit. A long-held con-
mary standard materials.5–9 vention for categorizing water purity was based on three
The NIST developed certified reference materials types, I through III, with type I water having the most
(CRMs/SRMs) for use in clinical chemistry laboratories. stringent requirements and generally suitable for routine
They are assigned a value after careful analysis, using laboratory use.
state-of-the-art methods and equipment. The chemical Prefiltration can remove particulate matter for munici-
composition of these substances is then certified; however, pal water supplies before any additional treatments.
they may not possess the purity equivalent of a primary Filtration cartridges are composed of glass; cotton; acti-
standard. Because each substance has been characterized vated charcoal, which removes organic materials and chlo-
for certain chemical or physical properties, it can be used rine; and submicron filters ( 0.2 mm), which remove any
in place of an ACS primary standard in clinical work and substances larger than the filter’s pores, including bacteria.
is often used to verify calibration or accuracy/bias assess- The use of these filters depends on the quality of the mu-
ments. Many manufacturers use an NIST SRM when pro- nicipal water and the other purification methods used. For
ducing calibrator and standard materials and, in this way, example, hard water (containing calcium, iron, and other
these materials are considered “traceable to NIST” and dissolved elements) may require prefiltration with a glass
may meet certain accreditation requirements. There are or cotton filter rather than activated charcoal or submi-
SRMs for a number of routine analytes, hormones, drugs, cron filters, which quickly become clogged and are ex-
and blood gases, with others being added.10 pensive to operate. The submicron filter may be better
A secondary standard is a substance of lower purity suited after distillation, deionization, or reverse osmosis
with concentration determined by comparison with a treatment.
primary standard. The secondary standard depends not Distilled water has been purified to remove almost all
only on its composition, which cannot be directly deter- organic materials, using a technique of distillation much
mined, but also on the analytic reference method. Once like that found in organic chemistry laboratory distilla-
again, because physiologic primary standards are gener- tion experiments in which water is boiled and vaporized.
ally unavailable, clinical chemists do not by definition The vapor rises and enters into the coil of a condenser, a
have “true” secondary standards. Manufacturers of sec- glass tube that contains a glass coil. Cool water surrounds
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 7

this condensing coil, lowering the temperature of the used immediately, so storage is discouraged because the
water vapor. The water vapor returns to a liquid state, resistivity changes. Depending on the application, CLRW
which is then collected. Many impurities do not rise in water should be stored in a manner that reduces any
the water vapor, remaining in the boiling apparatus. The chemical or bacterial contamination and for short periods.
water collected after condensation has less contamina- Testing procedures to determine the quality of reagent
tion. Because laboratories use thousands of liters of water grade water include measurements of resistance, pH,
each day, stills are used instead of small condensing ap- colony counts (for assessing bacterial contamination) on
paratuses; however, the principles are basically the same. selective and nonselective media for the detection of co-
Water may be distilled more than once, with each distil- liforms, chlorine, ammonia, nitrate or nitrite, iron, hard-
lation cycle removing additional impurities. ness, phosphate, sodium, silica, carbon dioxide, chemi-
Deionized water has some or all ions removed, al- cal oxygen demand (COD), and metal detection. Some
though organic material may still be present, so it is nei- accreditation agencies13 recommend that laboratories
ther pure nor sterile. Generally, deionized water is puri- document culture growth, pH, and specific resistance on
fied from previously treated water, such as prefiltered or water used in reagent preparation. Resistance is meas-
distilled water. Deionized water is produced using either ured because pure water, devoid of ions, is a poor con-
an anion or a cation exchange resin, followed by replace- ductor of electricity. The relationship of water purity to
ment of the removed particles with hydroxyl or hydro- resistance is linear. Generally, as purity increases, so
gen ions. The ions that are anticipated to be removed does resistance. This one measurement does not suffice
from the water will dictate the type of ion exchange resin for determination of true water purity because a nonionic
to be used. One column cannot service all ions present in contaminant may be present that has little effect on re-
water. A combination of several resins will produce dif- sistance. Note that reagent water meeting specifications
ferent grades of deionized water. A two-bed system uses from other organizations, such as the ASTM, may not be
an anion resin followed by a cation resin. The different equivalent to those established for each type by CLSI and
resins may be in separate columns or in the same col- care should be taken to meet the assay procedural re-
umn. This process is excellent in removing dissolved quirements for water type requirements.
ionized solids and dissolved gases.
Reverse osmosis is a process that uses pressure to Solution Properties
force water through a semipermeable membrane, pro-
In clinical chemistry, substances found in biologic fluids
ducing water that reflects a filtered product of the origi-
are measured (e.g., serum, plasma, urine, and spinal
nal water. It does not remove dissolved gases. Reverse os-
fluid). A substance that is dissolved in a liquid is called a
mosis may be used as a pretreatment of water.
solute; in laboratory science, these biologic solutes are
Ultrafiltration and nanofiltration, like distillation, are
also known as analytes. The liquid in which the solute is
excellent in removing particulate matter, microorgan-
dissolved—in this instance, a biologic fluid—is the sol-
isms, and any pyrogens or endotoxins. Ultraviolet oxida-
vent. Together they represent a solution. Any chemical
tion (removes some trace organic material) or steriliza-
or biologic solution is described by its basic properties,
tion processes (uses specific wavelengths), together with
including concentration, saturation, colligative proper-
ozone treatment, can destroy bacteria but may leave be-
ties, redox potential, conductivity, density, pH, and ionic
hind residual products. These techniques are often used
strength.
after other purification processes have been used.
Production of reagent grade water largely depends on Concentration
the condition of the feed water. Generally, reagent grade Analyte concentration in solution can be expressed in
water can be obtained by initially filtering it to remove many ways. Routinely, concentration is expressed as per-
particulate matter, followed by reverse osmosis, deioniza- cent solution, molarity, molality, or normality and, be-
tion, and a 0.2-mm filter or more restrictive filtration cause these non-SI expressions are so widely used, they
process. Type III/autoclave wash water is acceptable for will be discussed here. Note that the SI expression for the
glassware washing but not for analysis or reagent prepara- amount of a substance is the mole.
tion. Traditionally, type II water was acceptable for most Percent solutions is expressed as equal parts per hun-
analytic requirements, including reagent, quality control, dred or the amount of solute per 100 total units of solu-
and standard preparation, while type I water was used for tion. Three expressions of percent solutions are weight
test methods requiring minimum interference, such as per weight (w/w), volume per volume (v/v), and, most
trace metal, iron, and enzyme analyses. Use with HPLC commonly, weight per volume (w/v). For v/v solutions,
may require less than a 0.2-mm final filtration step and fall it is recommended that grams per deciliter (g/dL) be used
into the SRW category. Some molecular diagnostic or instead of percent or % (v/v).
mass spectrophotometric techniques may require special Molarity is expressed as the number of moles per 1 L
reagent grade water; some reagent grade water should be of solution. One mole of a substance equals its gram mo-
8 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

lecular weight (gmw). The SI representation for the tradi- Colligative Properties
tional molar concentration is moles of solute per volume The behavior of particles in solution demonstrates four
of solution, with the volume of the solution given in repeatable properties based only on the relative number
liters. The SI expression for concentration should be rep- of each kind of molecule present. The properties of os-
resented as moles per liter (mol/L), millimoles per liter motic pressure, vapor pressure, freezing point, and boil-
(mmol/L), micromoles per liter ( mol/L), and nanomoles ing point are called colligative properties. Vapor pressure
per liter (nmol/L). The familiar concentration term mo- is the pressure at which the liquid solvent is in equilib-
larity has not been adopted by the SI as an expression of rium with the water vapor. Freezing point is the temper-
concentration. ature at which the vapor pressures of the solid and liquid
Molality represents the amount of solute per 1 kg of phases are the same. Boiling point is the temperature
solvent. Molality is sometimes confused with molarity; at which the vapor pressure of the solvent reaches one at-
however, it can be easily distinguished from molarity be- mosphere.
cause molality is always expressed in terms of weight per Osmotic pressure is the pressure that opposes osmo-
weight or moles per kilogram and describes moles per sis when a solvent flows through a semipermeable mem-
1,000 g (1 kg) of solvent. The preferred expression for brane to establish equilibrium between compartments of
molality is moles per kilogram (mol/kg). differing concentration. The osmotic pressure of a dilute
Normality is the least likely of the four concentration solution is proportional to the concentration of the mol-
expressions to be encountered in clinical laboratories, ecules in solution. The expression for concentration is
but it is often used in chemical titrations and chemical the osmole. One osmole of a substance equals the molar-
reagent classification. It is defined as the number of gram ity multiplied by the number of particles at dissociation.
equivalent weights per 1 L of solution. An equivalent When a solute is dissolved in a solvent, these colligative
weight is equal to the gmw of a substance divided by its properties change in a predictable manner for each os-
valence. The valence is the number of units that can mole of substance present; the freezing point is lowered
combine with or replace 1 mole of hydrogen ions for by 1.86°C, the boiling point is raised by 0.52°C, the
acids, hydroxyl ions for bases, and the number of elec- vapor pressure is lowered by 0.3 mm Hg or torr, and the
trons exchanged in oxidation reduction reactions. It is osmotic pressure is increased by a factor of 1.7 104
the number of atoms/elements that can combine for a mm Hg or torr. In the clinical setting, freezing point and
particular compound; therefore, the equivalent weight is vapor pressure depression are measured as a function of
the gram combining weight of a material. Normality is osmolality.
always equal to or greater than the molarity of that com-
pound. Normality was previously used for reporting elec- Redox Potential
trolyte values, such as sodium [Na ], potassium [K ], Redox potential, or oxidation-reduction potential, is a
and chloride [Cl ] expressed as milliequivalents per liter measure of the ability of a solution to accept or donate
(mEq/L); however, this convention has been replaced with electrons. Substances that donate electrons are called re-
the more familiar units of millimoles per liter (mmol/L). ducing agents; those that accept electrons are considered
Solution saturation gives little specific information oxidizing agents. The pneumonic—LEO (lose electrons
about the concentration of solutes in a solution. oxidized) the lion says GER (gain electrons reduced)—
Temperature, as well as the presence of other ions, can in- may prove useful when trying to recall the relationship
fluence the solubility constant for a solute in a given solu- between reducing/oxidizing agents and redox potential.
tion and thus affect the saturation. Routine terms in the Conductivity
clinical laboratory that describe the extent of saturation Conductivity is a measure of how well electricity passes
are dilute, concentrated, saturated, and supersaturated. A di- through a solution. A solution’s conductivity quality de-
lute solution is one in which there is relatively little solute. pends principally on the number of respective charges of
In contrast, a concentrated solution has a large quantity of the ions present. Resistivity, the reciprocal of conductiv-
solute in solution. A solution in which there is an excess ity, is a measure of a substance’s resistance to the passage
of undissolved solute particles is a saturated solution. As of electrical current. The primary application of resistiv-
the name implies, a supersaturated solution has an even ity in the clinical laboratory is for assessing the purity of
greater concentration of undissolved solute particles than water. Resistivity or resistance is expressed as ohms and
a saturated solution of the same substance. Because of the conductivity is expressed as ohms 1 or mho.
greater concentration of solute particles, a supersaturated
solution is thermodynamically unstable. The addition of a pH and Buffers
crystal of solute or mechanical agitation disturbs the su- Buffers are weak acids or bases and their related salts that,
persaturated solution, resulting in crystallization of any as a result of their dissociation characteristics, minimize
excess material out of solution. An example is seen when changes in the hydrogen ion concentration. Hydrogen ion
measuring serum osmolality by freezing point depression. concentration is often expressed as pH. A lowercase p in
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 9

front of certain letters or abbreviations operationally When the ratio of [A ] to [HA] is 1, the pH equals the
means the “negative logarithm of” or “inverse log of” that pK and the buffer has its greatest buffering capacity. The
substance. In keeping with this convention, the term pH dissociation constant Ka, and therefore the pKa, remains
represents the negative or inverse log of the hydrogen ion the same for a given substance. Any changes in pH are
concentration. Mathematically, pH is expressed as solely due to the ratio of base/salt [A ] concentration to
weak acid [HA] concentration.
pH log
([H1 ]) Ionic strength is another important aspect of buffers,
particularly in separation techniques. Ionic strength is
pH log[H ] (Eq. 1-1) the concentration or activity of ions in a solution or
where [H ] equals the concentration of hydrogen ions in buffer. It is defined14 as follows:
moles per liter. I 1
⁄2 Σ CiZi2 or
The pH scale ranges from 0 to 14 and is a convenient
way to express hydrogen ion concentration. Σ {(Ci) (Zi)2}
(Eq. 1-8)
A buffer’s capacity to minimize changes in pH is re- 2
lated to the dissociation characteristics of the weak acid where Ci is the concentration of the ion, Zi is the charge
or base in the presence of its respective salt. Unlike a of the ion, and Σ is the sum of the quantity (Ci)(Zi)2 for
strong acid or base, which dissociates almost completely, each ion present. In mixtures of substances, the degree of
the dissociation constant for a weak acid or base solution dissociation must be considered. Increasing ionic
tends to be very small, meaning little dissociation occurs. strength increases the ionic cloud surrounding a com-
The ionization of acetic acid (CH3COOH), a weak pound and decreases the rate of particle migration. It can
acid, can be illustrated as follows: also promote compound dissociation into ions effectively
[HA] ↔ [A ] [H ] increasing the solubility of some salts, along with
changes in current, which can also affect electrophoretic
[CH3COOH]↔ [CH3COO ] [H ] (Eq. 1-2) separation.
where HA is a weak acid, A is a conjugate base, and
[H ] represents hydrogen ions. CLINICAL LABORATORY SUPPLIES
Note that the dissociation constant, Ka, for a weak Many different supplies are required in today’s medical
acid may be calculated using the following equation: laboratory; however, several items are common to most
[A ][H ] facilities, including thermometers, pipets, flasks,
Ka (Eq. 1-3) beakers, burets, desiccators, and filtering material. The
[HA]
following is a brief discussion of the composition and
Rearrangement of this equation reveals general use of these supplies.
[HA]
[H ] Ka (Eq. 1-4)
[A ] Thermometers/Temperature
Taking the log of each quantity and then multiplying by The predominant practice for temperature measurement
minus 1 ( 1), the equation can be rewritten as uses the Celsius (°C) scale; however, Fahrenheit (°F) and
Kelvin (°K) scales are also used.15,16 The SI designation
[HA]
log[H ] log Ka log (Eq. 1-5) for temperature is the Kelvin scale. Appendix C (on the
[A ]
companion website) lists the various conversion formu-
By convention, lower case p means “negative log of”; las between each scale.
therefore, log[H ] may be written as pH, and Ka may All analytic reactions occur at an optimal temperature.
be written as pKa. The equation now becomes Some laboratory procedures, such as enzyme determina-
tions, require precise temperature control, whereas others
[HA]
pH pKa log (Eq. 1-6) work well over a wide range of temperatures. Reactions
[A ]
that are temperature dependent use some type of heat-
Eliminating the minus sign in front of the log of the ing/cooling cell, heating/cooling block, or water/ice bath
[HA] to provide the correct temperature environment.
quantity results in an equation known as the
[A ] Laboratory refrigerator temperatures are often critical and
Henderson-Hasselbalch equation, which mathematically need periodic verification. Thermometers are either an in-
describes the dissociation characteristics of weak acids and tegral part of an instrument or need to be placed in the de-
bases and the effect on pH: vice for temperature maintenance. The three major types
of thermometers discussed include liquid-in-glass, elec-
[A ]
pH pKa log (Eq. 1-7) tronic thermometer or thermistor probe, and digital
[HA]
10 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

thermometer; however, several other types of temperature- tolerances of accuracy. Those that satisfy NIST specifica-
indicating devices are in use. Regardless of which is being tions20,21 are classified as Class A. Vessels holding or
used, all temperature reading devices must be calibrated to transferring liquid are designed either to contain (TC) or
ascertain accuracy. Liquid-in-glass thermometers use a to deliver (TD) a specified volume. As the names imply,
colored liquid (red or other colored material) or mercury the major difference is that TC devices do not deliver that
encased in plastic or glass material with a bulb at one end same volume when the liquid is transferred into a con-
and a graduated stem. They usually measure temperatures tainer, whereas the TD designation means that the lab-
between 20°C and 400°C. Partial immersion thermome- ware will deliver that amount.
ters are used for measuring temperatures in units such as Glassware used in the clinical laboratory usually falls
heating blocks and water baths and should be immersed to into one of the following categories: Kimax/Pyrex
the proper height as indicated by the continuous line (borosilicate), Corex (aluminosilicate), high silica, Vycor
etched on the thermometer stem. Total immersion ther- (acid and alkali resistant), low actinic (amber colored),
mometers are used for refrigeration applications, and sur- or flint (soda lime) glass used for disposable material.22
face thermometers may be needed to check temperatures Whenever possible, routinely used clinical chemistry
on flat surfaces, such as in an incubator or heating oven. glassware should consist of high thermal borosilicate or
Visual inspection of the liquid-in-glass thermometer aluminosilicate glass and meet the Class A tolerances rec-
should reveal a continuous line of liquid, free from sepa- ommended by the NIST. Glassware that does not meet
ration or gas bubbles. The accuracy range for a ther- Type A specifications may have twice the tolerance range
mometer used in clinical laboratories is determined by the despite its appearance being identical to a piece of Type
specific application but, generally, the accuracy range A glassware. The manufacturer is the best source of in-
should equal 50% of the desired temperature range re- formation about specific uses, limitations, and accuracy
quired by the procedure. specifications for glassware.
Liquid-in-glass thermometers should be calibrated Plasticware is beginning to replace glassware in the
against an NIST-certified or NIST-traceable thermometer laboratory setting. The unique high resistance to corro-
for critical laboratory applications.17 NIST has an SRM sion and breakage, as well as varying flexibility, has made
thermometer with various calibration points (0°, 25°, plasticware most appealing. Relatively inexpensive, it al-
30°, and 37°C) for use with liquid-in-glass thermome- lows most items to be completely disposable after each
ters. Gallium, another SRM, has a known melting point use. The major types of resins frequently used in the
and can also be used for thermometer verification. clinical chemistry laboratory are polystyrene, polyeth-
As automation advances and miniaturizes, the need for ylene, polypropylene, Tygon, Teflon, polycarbonate, and
an accurate, fast-reading electronic thermometer (thermis- polyvinyl chloride. Again, the individual manufacturer is
tor) has increased and is now routinely incorporated in the best source of information concerning the proper use
many devices. The advantages of a thermistor over the and limitations of any plastic material.
more traditional liquid-in-glass thermometers are size and In most laboratories, glass or plastic that is in direct
millisecond response time. One disadvantage can be the contact associated with biohazardous material is usually
initial cost, although the use of a thermistor probe at- disposable. If not, it must be decontaminated according
tached to an already owned volt–ohm meter (VOM) can to appropriate protocols. Should the need arise, however,
be cost effective. Similar to the liquid-in-glass thermome- cleaning of glass or plastic may require special tech-
ters, the thermistor can be calibrated against an SRM ther- niques. Immediately rinsing glass or plastic supplies after
mometer or the gallium melting point cell.18,19 When the use, followed by washing with a powder or liquid deter-
thermistor is calibrated against the gallium cell, it can be gent designed for cleaning laboratory supplies and sev-
used as a reference for any type thermometer. eral distilled water rinses, may be sufficient. Presoaking
glassware in soapy water is highly recommended when-
ever immediate cleaning is impractical. Many laborato-
Glassware and Plasticware
ries use automatic dishwashers and dryers for cleaning.
Until recently, laboratory supplies (e.g., pipets, flasks, Detergents and temperature levels should be compatible
beakers, and burets) consisted of some type of glass and with the material and the manufacturer’s recommenda-
could be correctly termed glassware. As plastic material tions. To ensure that all detergent has been removed
was refined and made available to manufacturers, plastic from the labware, multiple rinses with appropriate grade
has been increasingly used to make laboratory utensils. water is recommended. Check the pH of the final rinse
Before discussing general laboratory supplies, a brief sum- water and compare it with the initial pH of the prerinse
mary of the types and uses of glass and plastic commonly water. Detergent-contaminated water will have a more
seen today in laboratories is given. (See Appendices G, H, alkaline pH as compared with the pH of the appropriate
and I on the book’s companion website.) Regardless of grade water. Visual inspection should reveal spotless ves-
design, most laboratory supplies must satisfy certain sel walls. Any biologically contaminated labware should
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 11

be disposed of according to the precautions followed by Graduated cylinders are long, cylindrical tubes usu-
that laboratory. ally held upright by an octagonal or circular base. The
Some determinations, such as those used in assessing cylinder has calibration marks along its length and is
heavy metals or assays associated with molecular test- used to measure volumes of liquids. Graduated cylinders
ing, require scrupulously clean or disposable glassware. do not have the accuracy of volumetric glassware. The
Some applications may require plastic rather than glass sizes routinely used are 10, 25, 50, 100, 500, 1,000, and
because glass can absorb metal ions. Successful cleaning 2,000 mL.
solutions are acid dichromate and nitric acid. It is sug- All laboratory utensils should be Class A whenever
gested that disposable glass and plastic be used when- possible to maximize accuracy and precision and thus
ever possible. decrease calibration time. (Figure 1-8 illustrates repre-
Dirty pipets should be placed immediately in a con- sentative laboratory glassware.) Table 1-3 lists the Class
tainer of soapy water with the pipet tips up. The con- A tolerances for some commonly used volumes.
tainer should be long enough to allow the pipet tips to be
covered with solution. A specially designed pipet soaking Pipets
jar and washing/drying apparatus are recommended. For Pipets are glass or plastic utensils used to transfer liquids;
each final water rinse, fresh reagent grade water should they may be reusable or disposable. Although pipets may
be provided. If possible, designate a pipet container for transfer any volume, they are usually used for volumes of
final rinses only. Cleaning brushes are available to fit al- 20 mL or less; larger volumes are usually transferred or
most any size glassware and are recommended for any ar- dispensed using automated pipetting devices or jar-style
ticles that are washed routinely. pipetting apparatus. To minimize confusion, Table 1-4
Although plastic material is often easier to clean be- lists the classification scheme further described here.
cause of its nonwettable surface, it may not be appropri- Examples of pipets are found in Figure 1-1.
ate for some applications involving organic solvents or
autoclaving. Brushes or harsh abrasive cleaners should
not be used on plasticware. Acid rinses or washes are not
required. The initial cleaning procedure described in TABLE 1-3 CLASS A TOLERANCES
Appendix J (on the book’s companion Web site) can be SIZE (mL) TOLERANCES (mL)
adapted for plasticware as well. Ultrasonic cleaners can BURETS
help remove debris coating the surfaces of glass or plas-
5 0.01
ticware. Properly cleaned glassware should be com-
pletely dried before using. 10 0.02
25 0.03
Laboratory Vessels 50 0.05
Flasks, beakers, and graduated cylinders are used to hold
100 0.10
solutions. Volumetric and Erlenmeyer flasks are two types
of containers in general use in the clinical laboratory. PIPETS (TRANSFER)
A Class A volumetric flask is calibrated to hold one 0.5–2 0.006
exact volume of liquid (TC). The flask has a round, 3–5 0.01
lower portion with a flat bottom and a long, thin neck
10 0.02
with an etched calibration line. Volumetric flasks are
used to bring a given reagent to its final volume with the 15–25 0.03
prescribed diluent and should be Class A quality. When 50 0.05
bringing the bottom of the meniscus to the calibration VOLUMETRIC FLASKS
mark, a pipet should be used when adding the final drops
1–10 0.02
of diluent to ensure maximum control is maintained and
the calibration line is not missed. 25 0.03
Erlenmeyer flasks and Griffin beakers are designed 50 0.05
to hold different volumes rather than one exact amount. 100 0.08
Because Erlenmeyer flasks and Griffin beakers are often
200 0.10
used in reagent preparation, flask size, chemical inert-
ness, and thermal stability should be considered. The 250 0.12
Erlenmeyer flask has a wide bottom that gradually 500 0.20
evolves into a smaller, short neck. The Griffin beaker has 1,000 0.30
a flat bottom, straight sides, and an opening as wide as
2,000 0.50
the flat base, with a small spout in the lip.
12 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

TABLE 1-4 PIPET CLASSIFICATION Similar to many laboratory utensils, pipets are de-
signed to contain (TC) or to deliver (TD) a particular vol-
I. Design
A. To contain (TC) ume of liquid. The major difference is the amount of
B. To deliver (TD) liquid needed to wet the interior surface of the ware and
the amount of any residual liquid left in the pipet tip.
II. Drainage characteristics
Most manufacturers stamp TC or TD near the top of the
A. Blowout
B. Self-draining pipet to alert the user as to the type of pipet. Like other
TC-designated labware, a TC pipet holds or contains a
III. Type particular volume but does not dispense that exact vol-
A. Measuring or graduated
ume, whereas a TD pipet will dispense the volume indi-
1. Serologic
2. Mohr
cated. When using either pipet, the tip must be immersed
3. Bacteriologic in the liquid to be transferred to a level that will allow it
4. Ball, Kolmer, or Kahn to remain in solution after the volume of liquid has en-
5. Micropipet tered the pipet—without touching the vessel walls. The
B. Transfer pipet is held upright, not at an angle (Fig. 1-2). Using a
1. Volumetric pipet bulb or similar device, a slight suction is applied to
2. Ostwald-Folin the opposite end until the liquid enters the pipet and the
3. Pasteur pipets meniscus is brought above the desired graduation line
4. Automatic macropipets or micropipets (Fig. 1-3A), suction is then stopped. While the meniscus
level is held in place, the pipet tip is raised slightly out of
the solution and wiped with a laboratory tissue of any ad-
hering liquid. The liquid is allowed to drain until the bot-
tom of the meniscus touches the desired calibration mark

FIGURE 1-1. Laboratory glassware. FIGURE 1-2. Correct and incorrect pipet positions.
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 13

5 in 1/10
Meniscus Bottom of
Graduation line the meniscus

A B Total volume Major divisions


FIGURE 1-3. Pipetting technique. (A) Meniscus is brought above FIGURE 1-5. Volume indication of a pipet.
the desired graduation line. (B) Liquid is allowed to drain until the
bottom of the meniscus touches the desired calibration mark.
have graduations to the tip. It is a self-draining pipet, but
(Fig. 1-3B). With the pipet held in a vertical position and the tip should not be allowed to touch the vessel while
the tip against the side of the receiving vessel, the pipet the pipet is draining. A serologic pipet has graduation
contents are allowed to drain into the vessel (e.g., test marks to the tip and is generally a blowout pipet. A mi-
tube, cuvet, flask). A blowout pipet has a continuous cropipet is a pipet with a total holding volume of less than
etched ring or two small, close, continuous rings located 1 mL; it may be designed as either a Mohr or serologic
near the top of the pipet. This means that the last drop of pipet. Measuring pipets are used to transfer reagents and
to make dilutions and can be used to repeatedly to trans-
liquid should be expelled into the receiving vessel.
Without these markings, a pipet is self-draining, and the fer a particular solution.
user allows the contents of the pipet to drain by gravity. The next major category is the transfer pipets. These
pipets are designed to dispense one volume without fur-
The tip of the pipet should not be in contact with the ac-
cumulating fluid in the receiving vessel during drainage. ther subdivisions. The bulblike enlargement in the pipet
With the exception of the Mohr pipet, the tip should re- stem easily distinguishes the Ostwald-Folin and volumetric
main in contact with the side of the vessel for several sec- subgroups. Ostwald-Folin pipets are used with biologic
onds after the liquid has drained. The pipet is then re- fluids having a viscosity greater than that of water. They
moved. Various pipet bulbs are illustrated in Figure 1-4. are blowout pipets, indicated by two etched continuous
Measuring or graduated pipets are capable of dispens- rings at the top. The volumetric pipet is designed to dis-
ing several different volumes. Because the graduation lines pense or transfer aqueous solutions and is always self-
located on the pipet may vary, they should be indicated on draining. This type of pipet usually has the greatest de-
the top of each pipet. For example, a 5-mL pipet can be gree of accuracy and precision and should be used when
used to measure 5, 4, 3, 2, or 1 mL of liquid, with further diluting standards, calibrators, or quality-control mate-
graduations between each milliliter. The pipet is desig- rial. They should only be used once. Pasteur pipets do not
nated as 5 in 1/10 increments (Fig. 1-5) and could deliver have calibration marks and are used to transfer solutions
any volume in tenths of a milliliter, up to 5 mL. Another or biologic fluids without consideration of a specific vol-
ume. These pipets should not be used in any quantitative
pipet, such as a 1-mL pipet, may be designed to dispense
1 mL and have subdivisions of hundredths of a milliliter. analytic techniques.
The markings at the top of a measuring or graduated pipet The automatic pipet is the most routinely used pipet in
indicate the volume(s) it is designed to dispense. today’s clinical chemistry laboratory. Automatic and
The subgroups of measuring or graduated pipets are semiautomatic pipets have many advantages, including
Mohr, serologic, and micropipets. A Mohr pipet does not safety, stability, ease of use, increased precision, the abil-
ity to save time, and less cleaning required as a result of
the contaminated portions of the pipet (e.g., the tips)
often being disposable. Figure 1-6 illustrates many com-
mon automatic pipets. A pipet associated with only one
volume is termed a fixed volume, and models able to se-
lect different volumes are termed variable; however, only
one volume may be used at a time. The available range of
volumes is 1 L to 5,000 mL. The widest volume range
usually seen in a single pipet is 0 to 1 mL. A pipet with a
pipetting capability of less than 1 mL is considered a mi-
cropipet, and a pipet that dispenses greater than 1 mL is
called an automatic macropipet.
The term automatic, as used here, implies that the
mechanism that draws up and dispenses the liquid is an
integral part of the pipet. It may be a fully automated/
FIGURE 1-4. Types of pipet bulbs. self-operating, semiautomatic, or completely manually
14 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

FIGURE 1-6. (A) Fixed-volume, ultramicrodigital, air-displacement pipetter with tip ejector. (B) Fixed-volume
air-displacement pipet. (C) Digital electronic positive-displacement pipetter. (D) Syringe pipetter.

operated device. There are three general types of auto- not require a different tip for each use. Because of carry-
matic pipets: air-displacement, positive-displacement, and over concerns, rinsing and blotting between samples may
dispenser pipets. An air-displacement pipet relies on a pis- be required. Dispensers and dilutor/dispensers are auto-
ton for suction creation to draw the sample into a dispos- matic pipets that obtain the liquid from a common reser-
able tip that must be changed after each use. The piston voir and dispense it repeatedly. The dispensing pipets may
does not come in contact with the liquid. A positive- be bottle-top, motorized, handheld, or attached to a dilu-
displacement pipet operates by moving the piston in the tor. The dilutor often combines sampling and dispensing
pipet tip or barrel, much like a hypodermic syringe. It does functions. Figure 1-7 provides examples of different types

FIGURE 1-7. (A) Digital dilutor/dispenser. (B) Dispenser. (C) Dispenser.


CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 15

of automatic pipetting devices. These pipets should be These calibration techniques are time consuming and,
used according to the individual manufacturer’s direc- therefore, impractical for use in daily checks. It is rec-
tions. Many automated pipets use a wash between samples ommended that pipets be checked initially and subse-
to eliminate carry-over problems. However, to minimize quently three or four times per year, or as dictated by the
carry-over contamination with manual or semiautomatic laboratory’s accrediting agency. Many companies offer
pipets, careful wiping of the tip may remove any liquid calibration services; the one chosen should also satisfy
that adhered to the outside of the tip before dispensing any any accreditation requirements. A quick, daily check for
liquid. Care should be taken to ensure that the orifice of many larger-volume automatic pipetting devices involves
the pipet tip is not blotted, drawing sample from the tip. the use of volumetric flasks. For example, a bottle-top
Another precaution in using manually operated semiauto- dispenser that routinely delivers 2.5 mL of reagent may
matic pipets is to move the plunger in a continuous, slow be checked by dispensing four aliquots of the reagent
manner. into a 10-mL Class A volumetric flask. The bottom of the
Disposable, one-use pipet tips are designed for use meniscus should meet with the calibration line on the
with air-displacement pipets. The laboratory scientist volumetric flask.
should ensure that the pipet tip is seated snugly onto the
end of the pipet and free from any deformity. Plastic tips Burets
used on air-displacement pipets are likely to vary. A buret looks like a wide, long, graduated pipet with a
Different brands can be used for one particular pipet but stopcock at one end. A buret’s usual total volume ranges
they do not necessarily perform in an identical manner. from 25 mL to 100 mL of solution and is used to dis-
Plastic burrs may be present in the interior of the tip that pense a particular volume of liquid during a titration
cannot always be detected by the naked eye. A method (Fig. 1-8).
using a 0.1% solution of phenol red in distilled water has
been used to compare the reproducibility of different
brands of pipet tips.23 When using this method, the pipet
and the operator should remain the same so that varia-
tion is only a result of changes in the pipet tips.
Tips for positive-displacement pipets are made of
straight columns of glass or plastic. These tips must fit
snugly to avoid carry-over and can be used repeatedly
without being changed after each use. As previously men-
tioned, these devices may need to be rinsed and dried be-
tween samples to minimize carry-over.
Class A pipets, like all other Class A glassware, do not
need to be recalibrated by the laboratory. Automatic
pipetting devices, as well as non–Class A materials, do
need recalibration. A gravimetric method (see page 16)
can accomplish this task by delivering and weighing a so-
lution of known specific gravity, such as water. A cur-
rently calibrated analytic balance and at least Class 2
weights should be used. A pipet should be used only if it
is within 1.0% of the expected value.
Although gravimetric validation is the most desirable
method, pipet calibration may also be accomplished by
using photometric methods, particularly for automatic
pipetting devices. When a spectrophotometer is used, the
molar extinction coefficient of a compound, such as potas-
sium dichromate, is obtained. After an aliquot of diluent is
pipetted, the change in concentration will reflect the vol-
ume of the pipet. Another photometric technique used to
assess pipet accuracy compares the absorbances of dilu-
tions of potassium dichromate, or another colored liquid
with appropriate absorbance spectra, using Class A volu-
metric glassware versus equivalent dilutions made with
the pipetting device. FIGURE 1-8. Examples of laboratory glassware.
16 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

Syringes crystallization is removed from the compound, it is said


Syringes are sometimes used for transfer of small volumes to be anhydrous. Substances that take up water on expo-
(less than 500 L) in blood gas analysis or in separation sure to atmospheric conditions are called hygroscopic.
techniques such as chromatography or electrophoresis. Materials that are very hygroscopic can remove moisture
The syringes are glass and have fine barrels. The plunger from the air as well as from other materials. These mate-
is often made of a fine piece of wire. Tips are not used rials make excellent drying substances and are some-
when syringes are used for injection of sample into a gas times used as desiccants (drying agents) to keep other
chromatographic system. In electrophoresis work, how- chemicals from becoming hydrated. If these compounds
ever, disposable Teflon tips may be used. Expected inac- absorb enough water from the atmosphere to cause dis-
curacies of volumes less than 5 L is 2%, whereas for solution, they are called deliquescent substances.
greater volumes, the inaccuracy is approximately 1%. Desiccants are most effective when placed in a closed,
airtight chamber called a desiccator (Fig. 1-9). The des-
Desiccators and Desiccants iccant is placed below the perforated platform inside the
desiccator. Placing a fine film of lubricant on the rim of
Many compounds combine with water molecules to the cover seals a glass or plastic desiccator; it is correctly
form loose chemical crystals. The compound and the opened or sealed by slowly sliding the lid horizontally.
associated water are called a hydrate. When the water of The lubricated seal prevents the desiccator from being

GRAVIMETRIC PIPET CALIBRATION

Materials 8. If plastic tips are used, change the tip between


Pipet each dispensing. Repeat steps 1 to 6 for a mini-
10 to 20 pipet tips, if needed mum of nine additional times.
Balance capable of accuracy and resolution to 9. Obtain the average or mean of the weight of
0.1% of dispensed volumetric weight the water. Multiply the mean weight by the cor-
Weighing vessel large enough to hold volume of responding density of water at the given tem-
liquid perature and pressure. This may be obtained
Type I/CLRW water from the Handbook of Chemistry and Physics.6
Thermometer and barometer At 20°C, the density of water is 0.9982.
10. Determine the accuracy or the ability of the pipet
to dispense the expected (selected or stated) vol-
Procedure
ume according to the following formula:
1. Record the weight of the vessel. Record the tem-
perature of the water. It is recommended that Mean volume
100% (Eq. 1-9)
all materials be at room temperature. Obtain the Expected volume
barometric pressure. The manufacturer usually gives acceptable limita-
2. Place a small volume (0.5 mL) of the water into tions for a particular pipet, but they should not be
the container. To prevent effects from evapora- used if the value differs by more than 1.0% from
tion, it is desirable to loosely cover each con- the expected value.
tainer with a substance such as Parafilm. Avoid Precision can be indicated as the percent coeffi-
handling of the containers. cient of variation (%CV) or standard deviation (SD)
3. Weigh each container plus water to the nearest for a series of repetitive pipetting steps. A discussion
0.1 mg or set the balance to zero. of %CV and SD can be found in Chapter 4. The equa-
4. Using the pipet to be tested, draw up the speci- tions to calculate the SD and %CV are as follows:
fied amount. Carefully wipe the outside of the
tip. Care should be taken not to touch the end Σ (x x)2
SD n 1
of the tip; this will cause liquid to be wicked
out of the tip, introducing an inaccuracy as a re- SD
sult of technique. % CV 100 (Eq. 1-10)
x
5. Dispense the water into the weighed vessel.
Touch the tip to the side. Required imprecision is usually 1 SD. The %CV will
6. Record the weight of the vessel. vary with the expected volume of the pipet, but the
7. Subtract the weight obtained in step 3 from that smaller the %CV value, the greater is the precision.
obtained in step 6. Record the result. When n is large, the data are more statistically
valid.21,24
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 17

FIGURE 1-9. (A) Glass desiccator. (B) Glass desiccator with dry seal ring (no grease is necessary to seal cover
to body).

opened with a direct upward pull. The desiccator should the balance to the desired mass and places the material,
be opened slowly and with caution because the air pres- contained within a tared weighing vessel, on the sample
sure inside the desiccator could be below atmospheric pan. An optical scale allows the operator to visualize the
pressure. Desiccants containing indicators that signify mass of the substance. The weight range for certain ana-
desiccant exhaustion and that can be regenerated using lytic balances is from 0.01 mg to 160 g.
heat or dried in a microwave oven are particularly help- Electronic balances are single-pan balances that use an
ful. Many packaged materials will come with small pack- electromagnetic force to counterbalance the weighed sam-
ets of dessicant material that can be disposed of after the ple’s mass. Their measurements equal the accuracy and
packet is opened. Desiccants that produce dust should precision of any available mechanical balance, with the ad-
also be avoided, and these small packets of dessicants vantage of a fast response time (less than 10 seconds).
should not be confused with food condiments. In the lab-
oratory, desiccants are primarily used to prevent mois-
ture absorption by chemicals, gases, and instrument
components.

Balances
A properly operating balance is essential in producing
high-quality reagents and standards. However, because
many laboratories discontinued in-house reagent prepa-
ration, balances may no longer be as widely used.
Balances are classified according to their design, number
of pans (single or double), and whether they are me-
chanical or electronic or classified by operating ranges,
as determined by precision balances (readability 2 g),
analytic balances (readability 0.001 g), or microbal-
ances (readability 0.1 g; Fig. 1-10).
Analytic and electronic balances are currently the most
popular in the clinical laboratory. Analytic balances are re-
quired for the preparation of any primary standards. The
mechanical analytic balance is also known as a substitution
balance. It has a single pan enclosed by sliding transparent
doors, which minimize environmental influences on pan
movement. The pan is attached to a series of calibrated
weights that are counterbalanced by a single weight at the FIGURE 1-10. (A) Electronic top-loading balance. (B) Electronic an-
opposite end of a knife-edge fulcrum. The operator adjusts alytic balance with printer.
18 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

Test weights used for calibrating balances should be


selected from the appropriate ANSI/ASTM Classes 1
through 4.25 This system has replaced the former NIST
Class S standards used prior to 1993. Class 1 weights pro-
vide the greatest precision and should be used for cali-
brating high-precision analytic balances in the weight
range of 0.01 mg to 0.1 mg. Former NBS S standard
weights are equivalent to ASTM Class 2 (0.001–0.01 g),
and S-1 is equivalent to ASTM Class 3 (0.01–0.1 g). The
frequency of calibration is dictated by the accreditation/li-
censing guidelines for a specific laboratory. Balances
should be kept scrupulously clean and be located in an
area away from heavy traffic, large pieces of electrical
equipment, and open windows. A slab of marble sepa-
rated from its supporting surface by a flexible material is
sometimes placed under a balance to minimize any vibra-
tion interference that may occur. The level checkpoint
should always be corrected before weighing occurs.

BASIC SEPARATION TECHNIQUES


Contemporary modifications of filtration and dialysis use
a matrix-based fibrous material that provides a mecha-
nism of separation in many homogeneous immunoassays.
These materials may be coated with specific antibody-lig-
and to foster selection of specific materials or species.
Certain labels use magnetic particles in conjunction with
strong magnets to effect separation. Further discussion of
separation mechanisms used in immunoassays may be
found in Chapters 7 and 8. Basic universally used separa-
tion mechanisms, outside of those incorporated in im-
munoassay, are centrifugation, filtration, and dialysis.

Centrifugation FIGURE 1-11. (A) Benchtop centrifuge. (B) Swinging-bucket rotor.


(C) Fixed-head rotor.
Centrifugation is a process in which centrifugal force is
used to separate solid matter from a liquid suspension.
The centrifuge carries out this action. It consists of a head Appendix F on the book’s companion website. Centrifuge
or rotor, carriers, or shields (Fig. 1-11) that are attached classification is based on several criteria, including bench-
to the vertical shaft of a motor and enclosed in a metal top or floor model, refrigeration, rotor head (e.g., fixed,
covering. The centrifuge always has a lid and an on/off hematocrit, swinging-bucket, or angled; Fig. 1-11), or
switch; however, many models include a brake or a built- maximum speed attainable (i.e., ultracentrifuge).
in tachometer, which indicates speed, and some cen- Centrifuges are generally used to separate serum or
trifuges are refrigerated. Centrifugal force depends on plasma from the blood cells as the blood samples are
three variables: mass, speed, and radius. The speed is ex- being processed; to separate a supernatant from a precip-
pressed in revolutions per minute (rpm), and the cen- itate during an analytic reaction; to separate two immisci-
trifugal force generated is expressed in terms of relative ble liquids, such as a lipid-laden sample; or to expel air.
centrifugal force (RCF) or gravities (g). The speed of the Centrifuge care includes daily cleaning of any spills or
centrifuge is related to the RCF by the following equation: debris, such as blood or glass, and ensuring that the cen-
5 trifuge is properly balanced and free from any excessive
RCF 1.118 10 r (rpm)2 (Eq. 1-11)
vibrations. Balancing the centrifuge load is critical (Fig.
where 1.118 10 5 is a constant, determined from the 1-12). Many newer centrifuges will automatically de-
angular velocity, and r is the radius in centimeters, meas- crease their speed if the load is not evenly distributed, but
ured from the center of the centrifuge axis to the bottom more often, the centrifuge will shake and vibrate or make
of the test-tube shield. The RCF value also may be ob- more noise than expected. A centrifuge needs to be bal-
tained from a nomogram similar to that found in anced based on equalizing both the volume and weight
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 19

FIGURE 1-13. Methods of folding a filter paper. (A) In a fan. (B) In


fourths.

Traditionally, filter paper was folded in a manner that al-


lowed it to fit into a funnel. In method A, round filter
paper is folded like a fan (Fig. 1-13A); in method B, the
paper is folded into fourths (Fig. 1-13B).
Filter papers differ in pore size and should be selected
FIGURE 1-12. Properly balanced centrifuge. Colored circles repre- according to separation needs and associated flow rate
sent counterbalanced positions for sample tubes. for given liquids. Filter paper should not be used when
using strong acids or bases. When the filter paper is
placed inside the funnel, the solution slowly drains
distribution across the centrifuge head. Many laborato- through the filter paper within the funnel and into a re-
ries will make up “balance” tubes that approximate rou- ceiving vessel. The liquid that passes through the filter
tinely used volumes and tube sizes, including the stopper paper is called the filtrate.
on phlebotomy tubes, which can be used to match those
needed from patient samples. A good rule of thumb is Dialysis
one of even placement and one of “opposition.” Exact po-
Dialysis is another method for separating macromole-
sitioning of tubes depends on the design of the centrifuge
cules from a solvent or smaller substances. It became
holders.
popular when used in conjunction with the Technicon
The centrifuge cover should remain closed until the
Autoanalyzer system in the 1970s. Basically, a solution is
centrifuge has come to a complete stop to avoid any
put into a bag or is contained on one side of a semiper-
aerosol contamination. It is recommended that the timer,
meable membrane. Larger molecules are retained within
brushes (if present), and speed be periodically checked.
the sack or on one side of the membrane, while smaller
The brushes, which are graphite bars attached to a re-
molecules and solvents diffuse out. This process is very
tainer spring, create an electrical contact in the motor.
slow. The use of columns that contain a gel material has
The specific manufacturer’s service manual should be
replaced manual dialysis separation in most analytic
consulted for details on how to change brushes and on
procedures.
lubrication requirements. The speed of a centrifuge is
easily checked using a tachometer or strobe light. The
LABORATORY MATHEMATICS AND
hole located in the lid of many centrifuges is designed for
CALCULATIONS
speed verification using these devices but may also rep-
resent an aerosol biohazard. Accreditation agencies re- Significant Figures
quire periodic verification of centrifuge speeds.
Significant figures are the minimum number of digits
needed to express a particular value in scientific notation
Filtration
without loss of accuracy. The number 814.2 has four sig-
Filtration can be used instead of centrifugation for the nificant figures, because in scientific notation it is written
separation of solids from liquids. However, paper filtra- as 8.142 102. The number 0.000641 has three signifi-
tion is only occasionally used in today’s laboratory and, cant figures, because the scientific notation expression for
therefore, only the basics are discussed here. Filter mate- this value is 6.41 10 4. The zeros are merely holding
rial is made of paper, cellulose and its derivatives, poly- decimal places and are not needed to properly express the
ester fibers, glass, and a variety of resin column materials. number in scientific notation.
20 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

Logarithms The logarithm of N (5.4) is equal to 0.7324, or 0.73.


The pH becomes
The base 10 logarithm (log) of a positive number N
greater than zero is equal to the exponent to which 10 pH 6 0.73 5.27 (Eq. 1-14)
must be raised to produce N. Therefore, it can be stated The same formula can be applied to obtain the hydro-
that N equals 10x, and the log of N is equal to x. The gen ion concentration of a solution when only the pH is
number N is the antilogarithm (antilog) of x. given. Using a pH of 5.27, the equation becomes
The logarithm of a number, which is written in deci-
mal format, consists of two parts: the character, or char- 5.27 x log N (Eq. 1-15)
acteristic, and the mantissa. The characteristic is the In this instance, the x term is always the next largest
number to the left of the decimal point in the log and is whole number. For this example, the next largest whole
derived from the exponent, and the mantissa is that number is 6. Substituting for x, the equation becomes
portion of the logarithm to the right of the decimal
point and is derived from the number itself. Although 5.27 6 log N (Eq. 1-16)
several approaches can be taken to determine the log, Multiply all the variables by 1:
one approach is to write the number in scientific nota-
tion. The number 1,424 expressed in scientific notation ( 1)(5.27) ( 1)(6) ( 1)(log N).
is 1.424 103, making the characteristic value a 3. The 5.27 6 log N (Eq. 1-17)
characteristic can also be determined by adding the
number of significant figures and then subtracting 1 Solve the equation for the unknown quantity by
from the sum. The mantissa is derived from a log table adding a positive 6 to both sides of the equal sign and the
or calculator having a log function for the remainder of equation becomes:
the number. For 1.424, a calculator would give a man- 6 5.27 log N
tissa of 0.1535. Certain calculators with a log function
do not require conversion to scientific notation. This 0.73 log N (Eq. 1-18)
would give a log value of 3.1535 (3 plus 0.1535). The result is 0.73, which is the antilogarithm value of N,
To determine the original number from a log value, which is 5.37, or 5.4.
the process is done in reverse. This process is termed
the antilogarithm. If given a log of 3.1535, most calcu- Antilog 0.73 N; N 5.37 5.4 (Eq. 1-19)
lators require that you enter this value, use an inverse The hydrogen ion concentration for a solution with a
function, and enter log, and the resulting value should pH of 5.27 is 5.4 10 6. Many scientific calculators
be 1.424 103. Consult the specific manufacturer’s di- have an inverse function or syntax that allows for more
rections to become acquainted with the proper use of direct calculation of inverse or negative logarithms. It is
these functions. important, however, to fully understand the proper use
Negative Logarithms of the many calculator functions available, keeping in
The characteristic of a log may be positive or negative, mind that the specific steps vary between manufacturers.
but the mantissa is always positive. In certain circum-
stances, the laboratory scientist must deal with inverse or Concentration
negative logs. Such is the case with pH or pKa. As previ-
A detailed description of each concentration term (e.g.,
ously stated, the pH of a solution is defined as minus the
molarity, normality) may be found at the beginning of
log of the hydrogen ion concentration. The following is a
this chapter. The following discussion focuses on the
convenient formula to determine the negative logarithm
basic mathematical expressions needed to prepare
when working with pH or pKa:
reagents of a stated concentration.
pH
x log N (Eq. 1-12) Percent Solution
pKa
A percent solution is determined in the same manner re-
where x is negative exponent base 10 expressed without gardless of whether weight/weight, volume/volume, or
the minus sign and N is the decimal portion of the scien- weight/volume units are used. Percent implies “parts per
tific notations expression. 100,” which is represented as percent (%) and is inde-
For example, if the hydrogen ion concentration of a pendent of the molecular weight of a substance.
solution is 5.4 10 6, then x 6 and N 5.4.
Substitute this information into Equation 1-12, and it
Example 1-1 Weight/Weight (w/w)
becomes:
To make up 100 g of a 5% aqueous solution of hy-
pH 6 log 5.4 (Eq. 1-13) drochloric acid (using 12 M HCl), multiply the total
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 21

amount by the percent expressed as a decimal. The cal- the amount of substance needed to yield a particular con-
culation becomes centration, initially decide what final concentration units
5 are needed. For molarity, the final units will be moles per
5% 0.050 (Eq. 1-20) liter (mol/L) or millimoles per milliliter (mmol/mL). The
100
second step is to consider the existing units and the rela-
Therefore, tionship they have to the final desired units. Essentially, try
0.050 100 g 5 g of 12 M HCl (Eq. 1-21)
to put as many units as possible into “like” terms and
arrange so that the same units cancel each other out, leav-
Another way of arriving at the answer is to set up a ing only those wanted in the final answer. To accomplish
ratio so that this, it is important to remember what units are used to de-
fine each concentration term. It is key to understand the re-
5 x
100 100 lationship among molarity (moles/liter), moles, and gmw.

x 5 (Eq. 1-22) Example 1-4


How many grams are needed to make 1 L of a 2 M solu-
Example 1-2 Weight/Volume (w/v) tion of HCl?
The most frequently used term for a percent solution is Step 1: Which units are needed in the final answer?
weight per volume, which is often expressed as grams Answer: Grams per liter (g/L).
per 100 mL of diluent. To make up 1,000 mL of a 10% Step 2: Assess other mass/volume terms used in the prob-
(w/v) solution of NaOH, use the preceding approach. lem. In this case, moles are also needed for the calcula-
The calculations become tion: How many grams are equal to 1 mole? The gmw of
HCl, which can be determined from the periodic table,
0.10 1000 100 g will be equal to 1 mole. For HCl, the gmw is 36.5, so the
(% expressed as a decimal) (total amount) equation may be written as

or 36.5 g HCl 2 mol 73 g HCl


(Eq. 1-25)
mol L L
10 x
100 1000 Cancel out like units, and the final units should be
grams per liter. In this example, 73 grams HCl per liter is
x 100 (Eq. 1-23) needed to make up a 2 M solution of HCl.
Therefore, add 100 g of 10% NaOH to a 1,000-mL vol-
umetric Class A flask and dilute to the calibration mark Example 1-5
with reagent grade water. A solution of NaOH is contained within a Class A 1-L
volumetric flask filled to the calibration mark. The con-
tent label reads 24 g of NaOH. Determine the molarity.
Example 1-3 Volume/Volume (v/v)
Step 1: What units are ultimately needed? Answer: Moles
Make up 50 mL of a 2% (v/v) concentrated hydrochloric
per liter (mol/L).
acid solution.
Step 2: The units that exist are grams and 1 L. NaOH may
0.02 50 1 mL be expressed as moles and grams. The gmw of NaOH is
calculated to equal 40 g/mol. Rearrange the equation so
or
that grams can be canceled and the remaining units re-
2 x flect those needed in the answer, which are mole/L.
100 50 Step 3: The equation becomes
x 1 (Eq. 1-24) 24 g NaOH 1 mol mol
0.6 (Eq. 1-26)
L 40 g NaOH L
Therefore, add 40 mL of water to a 50-mL Class A vol-
umetric flask, add 1 mL of concentrated HCl, mix, and By canceling out like units and performing the appro-
dilute up to the calibration mark with reagent grade priate calculations, the final answer of 0.6 M or 0.6 mol/L
water. Remember, always add acid to water! is derived.

Molarity Example 1-6


Molarity (M) is routinely expressed in units of moles per Make up 250 mL of a 4.8 M solution of HCl.
liter (mol/L) or sometimes millimoles per milliliter Step 1: Units needed? Answer: Grams (g).
(mmol/mL). Remember that 1 mol of a substance is equal Step 2: Determine the gmw of HCl (36.5 g), which is
to the gmw of that substance. When trying to determine needed to calculate the molarity.
22 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

Step 3: Set up the equation, cancel out like units, and Because 500 mL is equal to 0.5 L, the final equation
perform the appropriate calculations: could be written by substituting 0.5 L for 500 mL, elim-
inating the need to include the 1,000 mL/L conversion
36.5 g HCl 4.8 mol HCl 250 mL 1 L factor in the equation.
mol L 1,000 mL B. What is the normality of a 0.5 M solution of H2SO4?
43.8 g HCl (Eq. 1-27) Continuing with the previous approach, the final equa-
In a 250-mL volumetric flask, add 200 mL of reagent tion is
grade water. Add 43.8 g of HCl and mix. Dilute up to the 0.5 mol H2SO4 98 g H2SO4 1 Eq H2SO4
calibration mark with reagent grade water. L mol H2SO4 49 g H2SO4
Although there are various methods to calculate labo-
ratory mathematical problems, this technique of cancel- 0.5 mol H2SO4 98 g H2SO4 1 Eq H2SO4
ing like units can be used in most clinical chemistry sit- L mol H2SO4 49 g H2SO4
uations, regardless of whether the problem requests mo- 1 Eq/L 1N (Eq. 1-32)
larity, normality, or exchanging one concentration term
for another. However, it is necessary to recall the inter- When changing molarity into normality or vice versa, the
following conversion formula may be applied:
relationship between all the units in the expression.
M V N (Eq. 1-33)
Normality
Normality (N) is expressed as the number of equivalent where V is the valence of the compound. Using this for-
weights per liter (Eq/L) or milliequivalents per milliliter mula, Example 1-7.3 becomes
(mEq/mL). Equivalent weight is equal to gmw divided by 0.5 M 2 1N (Eq. 1-34)
the valence (V). Normality has often been used in acid-
base calculations because an equivalent weight of a sub- Example 1-8
stance is also equal to its combining weight. Another ad- What is the molarity of a 2.5 N solution of HCl? This
vantage in using equivalent weight is that an equivalent problem may be solved in several ways. One way is to use
weight of one substance is equal to the equivalent weight the stepwise approach in which existing units are ex-
of any other chemical. changed for units needed. The equation is
2.5 Eq HCl 36 g HCl 1 mol HCl
Example 1-7
L 1 Eq 36 g HCl
Give the equivalent weight, in grams, for each substance
listed below. 2.5 mol/L HCl (Eq. 1-35)

1. NaCl (gmw 58 g, valence 1) The second approach is to use the normality-to-mo-


larity conversion formula. The equation now becomes
58/1 58 g per equivalent weight (Eq. 1-28)
M V 2.5 N
2. HCl (gmw 36, valence 1)
V 1
36/1 36 g per equivalent weight (Eq. 1-29)
2.5 N
3. H2SO4 (gmw 98, valence 2) M 2.5 N (Eq. 1-36)
1
98/2 49 g per equivalent weight (Eq. 1-30) When the valence of a substance is 1, the molarity will
A. What is the normality of a 500-mL solution that con- equal the normality. As previously mentioned, normality
tains 7 g of H2SO4? The approach used to calculate mo- either equals or is greater than the molarity.
larity could be used to solve this problem as well. Specific Gravity
Step 1: Units needed? Answer: Normality expressed as Density is expressed as mass per unit volume. The spe-
equivalents per liter (Eq/L). cific gravity is the ratio of the density of a material when
Step 2: Units you have? Answer: Milliliters and grams. compared to the density of water at a given temperature.
Now determine how they are related to equivalents per The units for specific gravity are grams per milliliter.
liter. (Hint: There are 49 g per equivalent—see Equation Specific gravity is often used with very concentrated ma-
1–30 above.) terials, such as commercial acids (e.g., sulfuric and hy-
Step 3: Rearrange the equation so that like terms cancel drochloric acids).
out, leaving Eq/L. This equation is The density of a concentrated acid can also be ex-
7 g H2SO4 1 Eq 1,000 mL pressed in terms of an assay or percent purity. The actual
500 mL 49 g H2SO4 1L concentration is equal to the specific gravity multiplied
by the assay or percent purity value (expressed as a deci-
0.285 Eq/L 0.285 N (Eq. 1-31) mal) stated on the label of the container.
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 23

Example 1-9 It requires 25 mL of the 2 M solution to make up 500 mL


A. What is the actual weight of a supply of concentrated of a 0.1 M solution. This problem differs from the other
HCl whose label reads specific gravity 1.19 with an assay conversions in that it is actually a dilution of a stock so-
value of 37%? lution. A more involved discussion of dilution problems
follows.
1.19 g/mL 0.37 0.44 g/mL of HCl (Eq. 1-37)
B. What is the molarity of this stock solution? The final Dilutions
units desired are moles per liter (mol/L). The molarity of
the solution is A dilution represents the ratio of concentrated or stock
material to the total final volume of a solution and consists
0.44 g HCl 1 mol HCl 1,000 mL of the volume or weight of the concentrate plus the vol-
mL 35 g HCl L ume of the diluent, with the concentration units remain-
12.05 mol/L or 12 M (Eq. 1-38)
ing the same. This ratio of concentrated or stock solution
to the total solution volume equals the dilution factor.
Conversions Because a dilution is made by adding a more concentrated
To convert one unit into another, the same approach of substance to a diluent, the dilution is always less concen-
crossing out like units can be applied. In some instances, trated than the original substance. The relationship of the
a chemistry laboratory may report a given analyte using dilution factor to concentration is an inverse one; thus, the
two different concentration units—for example, calcium. dilution factor increases as the concentration decreases. To
The recommended SI unit for calcium is millimoles per determine the dilution factor, simply take the amount
liter. The better known and more traditional units are needed and divide by the stock concentration, leaving it in
milligrams per deciliter (mg/dL). Again, it is important to a reduced-fraction form.
understand the relationship between the units given and
those needed in the final answer. Example 1-12
What is the dilution factor needed to make a 100 mEq/L
Example 1-10 sodium solution from a 3,000 mEq/L stock solution? The
Convert 8.2 mg/dL calcium to millimoles per liter dilution factor becomes
(mmol/L). The gmw of calcium is 40 g. So, if there are 40 100 1
g per mol, then it follows that there are 40 mg per mmol. (Eq. 1-42)
3,000 30
The units wanted are mmol/L. The equation becomes
The dilution factor indicates that the ratio of stock
8.2 mg 1 dL 1,000 mL 1 mmol material is 1 part stock made to a total volume of 30. To
dL 100 mL L 40 mg actually make this dilution, 1 mL of stock is added to 29
2.05 mmol mL of diluent. Note that the dilution factor indicates
(Eq. 1-39) the parts per total amount; however, in making the di-
L
lution, the sum of the amount of the stock material plus
Once again, the systematic stepwise approach of deleting the amount of the diluent must equal the total volume
similar units can be used for this conversion problem. or dilution fraction denominator. The dilution factor
A frequently encountered conversion problem or, may be correctly written as either a fraction or a ratio.
more precisely, a dilution problem occurs when a weaker Confusion arises when distinction is not made between
concentration or different volume is needed than the a ratio and a dilution, which by its very nature is a ratio
stock substance available, but the concentration terms of stock to diluent. A ratio is always expressed using a
are the same. The following formula is used: colon; a dilution can be expressed as either a fraction or
V1 C1 V2 C2 (Eq. 1-40) a ratio.26 Many directions in the laboratory are given
orally. For example, making a “1-in-4” dilution means
This formula is useful only if the concentration and vol- adding one part stock to a total of four parts. That is,
ume units between the substances are the same and if one part of stock would be added “to” three parts of
three of four variables are known. diluent. The dilution factor would be 1/4. Analyses per-
formed on the diluted material would need to be multi-
Example 1-11 plied by 4 to get the final concentration. That is very
What volume is needed to make 500 mL of a 0.1 M solu- different from saying make a “1-to-4” dilution! In this
tion of Tris buffer from a solution of 2 M Tris buffer? instance, the dilution factor would be 1/5! It is impor-
tant during procedures that you fully understand the
V1 2M 0.1M 500 mL
meaning of these expressions. Sample dilutions should
(V1)(2M) (0.1M)(500 mL); (V1)(2M) 50; be made using either reagent grade water, saline, or
therefore V1 50/2 25 mL (Eq. 1-41) method-specific diluent using Class A glassware. The
24 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

sample and diluent should be thoroughly mixed before


1 Stock “part” or volume
use. It is not recommended that sample dilutions be
made in smaller-volume sample cups or holders. Any 2
total volume can be used as long as the fraction reduces
to give the dilution factor. 3
“Parts” or volumes of diluent
needed to make a 1/5 dilution
Example 1-13 4
If in the preceding example 150 mL of the 100 mEq/L 5
sodium solution was required, the dilution ratio of
stock to total volume must be maintained. Set up a ratio FIGURE 1-14. Simple dilution. Consider this diagram depicting a
between the desired total volume and the dilution fac- substance having a 1/5 dilution factor. The dilution factor represents
tor to determine the amount of stock needed. The equa- that 1 part of stock is needed of a total of 5 parts. To make this di-
lution, you would determine the volume of 1 “part,” usually the
tion becomes stock or patient sample. The remainder of the “parts” or total
1 x would constitute the amount of diluent needed, or four times the
volume used for the stock.
30 150
x 5 (Eq. 1-43)
parts of stock needed from the total volume or parts to get
Note that ⁄150 reduces to the dilution factor of ⁄30. To make
5 1
the number of diluent “parts” needed. Once the volume
up this solution, 5 mL of stock is added to 145 mL of the of each part, usually stock, is available, multiply the dilu-
appropriate diluent, making the ratio of stock volume to ent parts needed to obtain the correct volume.
diluent volume equal to 5⁄145. Recall that the dilution factor
includes the total volume of both stock plus diluent in Example 1-16
the denominator and differs from the amount of diluent You have a 10 g/dL stock of protein standard. You need
to stock that is needed. a 2 g/dL standard. You only have 0.200 mL of 10 g/dL
stock to use. The procedure requires 0.100 mL.
Example 1-14 Solution:
Many laboratory scientists like using (V1)(C1) (V2)(C2)
for simple dilution calculations. This is fine, as long as you 2 g/dL 1
Dilution factor (Eq. 1-45)
recall that you will need to subtract the stock volume from 10 gdL 5
the total volume final for the correct diluent volume. You will need 1 part or volume of stock of a total of
(V1)(C1) (V2)(C2) 5 parts or volumes. Subtracting 1 from 5 yields that
4 parts or volumes of diluent is needed (Fig. 1-14.) In
(x)(3,000) (150)(100) this instance, you need at least 0.100 mL for the proce-
x 5; 150 5 145 mL of diluent dure. You have 0.200 mL of stock. You can make the di-
should be added to 5 mL of stock (Eq. 1-44) lution in various ways, as seen in Example 1-17. This is
a personal decision and often reflects the volumes of
Simple Dilutions available pipets, etc.
When making a simple dilution, the laboratory scientist
must decide on the total volume desired and the amount Example 1-17
of stock to be used. There are several ways to make a 1/5 dilution having only
0.200 mL of stock and needing a total minimum volume
Example 1-15 of 0.100 mL.
A 1:10 (1⁄10) dilution of serum can be achieved by using
a) Add 0.050 mL stock (1 part) to 0.200 mL of diluent
any of the following approaches. A ratio of 1:9—one part
(4 parts 0.050 mL).
serum and nine parts diluent (saline):
b) Add 0.100 mL of stock (1 part) to 0.400 mL of dilu-
A. 100 L of serum added to 900 L of saline. ent (4 parts 0.100 mL).
B. 20 L of serum added to 180 L of saline. c) Add 0.200 mL of stock (1 part) to 0.800 mL of diluent
C. 1 mL of serum added to 9 mL of saline. (4 parts 0.200 mL).
D. 2 mL of serum added to 18 mL of saline.
The dilution factor is also used to determine the final
Note that the sum of the ratio of serum to diluent (1:9) concentration of a dilution by multiplying the original
needed to make up each dilution satisfies the dilution fac- concentration by the inverse of the dilution factor or the
tor (1:10 or 1⁄10) of stock material to total volume. When dilution factor denominator when it is expressed as a
thinking about the stock to diluent volume, subtract the fraction.
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 25

Example 1-18 tion of the 1:4 dilution will result in the next dilution
Determine the concentration of a 200 mol/mL human (1:8). To establish the dilution factor needed for subse-
chorionic gonadotropin (hCG) standard that was di- quent dilutions, it is helpful to solve the following equa-
luted 1/50. This value is obtained by multiplying the tion for (x):
original concentration, 200 mol/mL hCG, by the dilu-
Stock/preceding concentration (x)
tion factor, 1/50. The result is 4 mol/mL hCG. Quite
(final dilution factor) (Eq. 1-48)
often, the concentration of the original material is
needed. To make up these dilutions, three test tubes are labeled
1:2, 1:4, and 1:8, respectively. One milliliter of diluent
is added to each test tube. To make the primary dilution
Example 1-19
of 1:2, 1 mL of serum is added to test tube no. 1. The
A 1:2 dilution of serum with saline had a creatinine re-
solution is mixed, and 1 mL of the primary dilution is
sult of 8.6 mg/dL. Calculate the actual serum creatinine
removed and added to test tube no. 2. After mixing,
concentration.
this solution contains a 1:4 dilution. Then 1 mL of the
Dilution factor: 1/2
1:4 dilution from test tube no. 2 is added to test tube
Dilution result 8.6 mg/dL
no. 3. Mix, and the resultant dilution in this test tube
Because this result represents 1/2 of the concentration,
is 1:8, satisfying all of the previously established crite-
the actual serum creatinine value is
ria. Refer to Figure 1-15 for an illustration of this serial
2 8.6 17.2 17.2 mg/dL (Eq. 1-46) dilution.

Serial Dilutions Example 1-21


A serial dilution may be defined as multiple progressive Another type of serial dilution combines several dilution
dilutions ranging from more-concentrated solutions to factors that are not multiples of one another. In our pre-
less-concentrated solutions. Serial dilutions are ex- vious example, 1:2, 1:4, and 1:8 dilutions are all related
tremely useful when the volume of concentrate or dilu- to one another by a factor of 2. Consider the situation
ent is in short supply and needs to be minimized or a when 1:10, 1:20, 1:100, and 1:200 dilution factors are
number of dilutions are required, such as in determining required. There are several approaches to solving this
a titer. The volume of patient sample available to the lab- type of dilution problem. One method is to treat the 1:10
oratory may be small (e.g., pediatric samples), and a serial and 1:20 dilutions as one serial dilution problem, the
dilution may be needed to ensure that sufficient sample is 1:20 and 1:100 dilutions as a second serial dilution, and
available. The serial dilution is initially made in the same the 1:100 and 1:200 dilutions as the last serial dilution.
manner as a simple dilution. Subsequent dilutions will Another approach is to consider what dilution factor of
then be made from each preceding dilution. When a serial the concentrate is needed to yield the final dilution. In
dilution is made, certain criteria may need to be satisfied. this example, the initial dilution is 1:10, with subse-
The criteria vary with each situation but usually include quent dilutions of 1:20, 1:100, and 1:200. The first di-
such considerations as the total volume desired, the lution may be accomplished by adding 1 mL of stock to
amount of diluent or concentrate available, the dilution 9 mL of diluent. The total volume of solution is 10 mL.
factor, the final concentration needed, and the support Our initial dilution factor has been satisfied. In making
materials required. the remaining dilutions, 2 mL of diluent is added to each
test tube.
Example 1-20 Initial/preceding dilution (x) dilution needed
A sample is to be diluted 1:2, 1:4, and, finally, 1:8. It is
arbitrarily decided that the total volume for each dilution
is to be 1 mL. Note that the least common denominator
for these dilution factors is 2. Once the first dilution is
made (1/2), a 1:2 (least common denominator between 2
and 4) dilution of it will yield
1 1 1
/2 /2 /4
(initial dilution factor)(next dilution factor)
(final dilution factor) (Eq. 1-47)

By making a 1:2 dilution of the first dilution, the sec-


ond dilution factor of 1:4 is satisfied. Making a 1:2 dilu- FIGURE 1-15. Serial dilution.
26 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

Solve for (x). Therefore, 1 g of CuSO4 5H2O contains 0.64 g of CuSO4,


Using the dilution factors listed above and solving for so the equation becomes:
(x), the equations become
0.9 g CuSO4 needed
1:10 (x) 1:20, 0.64 CuSO4 in CuSO4 H2O
where (x) 2 (or 1 part stock to 1 part diluent) 1.41 g CuSO4 5H2O required (Eq. 1-52)
1:20 (x) 1:100,
where (x) 5 (or 1 part stock to 4 parts diluent) Graphing and Beer’s Law
1:100 (x) 1:200, The Beer-Lambert law (Beer’s law) mathematically es-
where (x) 2 (or 1 part stockto 1 part diluent) tablishes the relationship between concentration and ab-
(Eq. 1-49) sorbance in many photometric determinations. Beer’s
law is expressed as
In practice, the 1:10 dilution must be diluted by a fac-
tor of 2 to obtain a subsequent 1:20 dilution. Because the A abc (Eq. 1-53)
second tube already contains 2 mL of diluent, 2 mL of where A is absorbance; a is absorptivity constant for a
the 1:10 dilution should be added (1 part stock to 1 part particular compound at a given wavelength under speci-
diluent). In preparing the 1:100 dilution from this, a 1:5 fied conditions of temperature, pH, and so on; b is length
dilution factor of the 1:20 mixture is required (1 part of the light path; and c is concentration.
stock to 4 parts diluent). Because this tube already con- If a method follows Beer’s law, then absorbance is pro-
tains 2 mL, the volume of diluent in the tube is divided portional to concentration as long as the length of the
by its parts, which is 4; thus, 500 L, or 0.500 mL, of light path and the absorptivity of the absorbing species re-
stock should be added. The 1:200 dilution is prepared in mains unaltered during the analysis. In practice, however,
the same manner using a 1:2 dilution factor (1 part stock there are limits to the predictability of a linear response.
to 1 part diluent) and adding 2 mL of the 1:100 to the 2 Even in automated systems, adherence to Beer’s law is
mL of diluent already in the tube. often determined by checking the linearity of the test
method over a wide concentration range. The limits of
Water of Hydration linearity often represent the reportable range of an assay.
Some compounds are available in a hydrated form. To This term should not be confused with the reference
obtain a correct weight for these chemicals, the attached ranges associated with clinical significance of a test.
water molecule(s) must be included. Assays measuring absorbance generally obtain the con-
centration results by using a Beer’s law graph, known as a
Example 1-22 standard graph or curve. This graph is made by plotting
How much CuSO4 5H2O must be weighed to prepare 1 L absorbance versus the concentration of known standards
of 0.5 M CuSO4? When calculating the gmw of this sub- (Fig. 1-16). Because most photometric assays set the
stance, the water weight must be considered so that the initial absorbance to zero (0) using a reagent blank, the
gmw is 250 g rather than gmw of CuSO4 alone (160 g). initial data points are 0,0. Graphs should be labeled
Therefore, properly and the concentration units must be given.
The horizontal axis is referred to as the x-axis, whereas
250 g CuSO4 5H2O 0.5 mol 125 g/L (Eq. 1-50) the vertical line is the y-axis. It is not important which
mol 1L
Cancel out like terms to obtain the result of 125 g/L. A
reagent protocol often designates the use of an anhy- .4
drous form of a chemical; frequently, however, all that is
Absorbance

available is a hydrated form. .3

.2
Example 1-23
A procedure requires 0.9 g of CuSO4. All that is available .1
is CuSO4 5H2O. What weight of CuSO4 5H2O is 1 2 3 4 5
needed? Calculate the percentage of CuSO4 present in Concentration
CuSO4 5H2O. The percentage is
Unknown absorbance = .250
160 0.64, or 64% (Eq. 1-51)
Concentration from graph = 3.2
250 FIGURE 1-16. Standard curve.
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 27

variable (absorbance or concentration) is assigned to an tration directly without initially needing a standard
individual axis, but it is important that the values as- graph, as follows:
signed to them are uniformly distributed along the axis.
A abc
By convention, in the clinical laboratory, concentration
C A
is usually plotted on the x-axis. On a standard graph, (Eq. 1-57)
ab
only the standard and the associated absorbances are
plotted. When the absorptivity constant (a) is given in units of
Once a standard graph has been established, it is per- grams per liter (moles) through a 1-centimeter (cm) light
missible to run just one standard, or calibrator, as long path, the term molar absorptivity ( ) is used. Substitution
as the system remains the same. One-point calculation of for a and A for A produces the following Beer’s law
or calibration refers to the calculation of the compari- formula:
son of the known standard/calibrator concentration and
C A (Eq. 1-58)
its corresponding absorbance to the absorbance of the
unknown value according to the following ratio:
Units used to report enzyme activity traditionally have
Concentration of standard (Cs) included weight, time, and volume. In the early days
Absorbance of standard (As) of enzymology, method-specific units (e.g., King-
Concentration of unknown (Cu) Armstrong, Caraway) were all different and confusing. In
(Eq. 1-54)
Absorbance of unknown (Au) 1961, the Enzyme Commission of the International
Union of Biochemistry recommended using one unit, the
Solving for the concentration of the unknown, the equa- international unit (IU), for reporting enzyme activity.
tion becomes: The IU is defined as the amount of enzyme that will cat-
Cu (Au)(Cu) (Eq. 1-55) alyze 1 mol of substrate per minute per liter. These
As units were often expressed as units per liter (U/L). The
designations IU, U, or IU/L were adopted by many clini-
Example 1-24 cal laboratories to represent the IU. Although the report-
The biuret protein assay is very stable and follows Beer’s ing unit is the same, unless the analysis conditions are
law. Rather than make up a completely new standard identical, use of the IU does not standardize the actual
graph, one standard (6 g/dL) was assayed. The ab- enzyme activity and, therefore, results between different
sorbance of the standard was 0.400, and the absorbance methods of the same enzyme do not result in equivalent
of the unknown was 0.350. Determine the value of the activity of the enzyme. For example, an ALP performed
unknown in g/dL. at 37°C will catalyze more substrate than if it is run at
(0.350)(6 g/dL) lower temperature, such as 25°C, even though the unit of
Cu 5.25 g/DL (Eq. 1-56) expression, U/L, will be the same. The SI recommended
(0.400)
unit is the katal, which is expressed as moles per liter per
This method of calculation is acceptable as long as second. Whichever unit is used, calculation of the activ-
everything in the system, including the instrument and ity using Beer’s law requires inclusion of the dilution
lot of reagents, remains the same. If anything in the sys- and, depending on the reporting unit, possible conver-
tem changes, a new standard graph should be done. sion to the appropriate term (e.g., mol to mol, mL to L,
Verification of linearity and/or calibration is required minute to second, and temperature factors). Beer’s law
whenever a system changes or becomes unstable. for the IU now becomes:
Regulatory agencies often prescribe the condition of ver-
ification as well as the how often the linearity needs to be C ( A) 10 6 (TV) (Eq. 1-59)
checked. ( ) (b)(SV)
where TV is total volume of sample plus reagents in mL
Enzyme Calculations and SV is sample volume used in mL. The 10 6 converts
Another application of Beer’s law is the calculation of en- moles to nmol for the IU. If another unit of activity is
zyme assay results. When calculating enzyme results, the used, such as the katal, conversion into liters and sec-
rate of absorbance change is often monitored continu- onds would be needed, but the conversions to and from
ously during the reaction to give the difference in ab- micromoles are excluded.
sorbance, known as the delta absorbance, or A. Instead
of using a standard graph or a one-point calculation, the Example 1-25
molar absorptivity of the product is used. If the absorp- The A per minute for an enzyme reaction is 0.250. The
tivity constant and absorbance, in this case A, is given, product measured has a molar absorptivity of 12.2
Beer’s law can be used to calculate the enzyme concen- 103 at 425 nm at 30°C. The incubation and reaction
28 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

temperature are also kept at 30°C. The assay calls for 1 ily involves the outer area of the bottom of the foot (a
mL of reagent and 0.050 mL of sample. Give the en- heel stick), the fleshy part of the middle of the last pha-
zyme activity results in international units. lanx of the third or fourth (ring) finger (finger stick), or
Applying Beer’s law and the necessary conversion in- possibly the fleshy portion of the earlobe. A sharp lancet
formation, the equation becomes: is used to pierce the skin and a capillary tube (i.e., short,
narrow glass tube) is used for sample collection.
C (0.250)(10 6)(1.050 mL) 430 U (Eq. 1-60) Additional information regarding phlebotomy and skin
(12.2 103)(1)(0.050 mL)
puncture is available in Chapter 2.28
Note: b is usually given as 1 cm; because it is a constant, Analytic testing of blood involves the use of whole
it may not be considered in the calculation. blood, serum, or plasma. Whole blood, as the name im-
plies, uses both the liquid portion of the blood called
SPECIMEN CONSIDERATIONS plasma and the cellular components (red blood cells,
white blood cells, and platelets). This requires blood col-
The process of specimen collection, handling, and pro-
lection into a vessel containing an anticoagulant.
cessing remains one of the primary areas of preanalytic
Complete mixing of the blood immediately following
error. Careful attention to each phase is necessary to
venipuncture is necessary to ensure the anticoagulant can
ensure proper subsequent testing and reporting of
adequately inhibit the blood’s clotting factors. As whole
meaningful results. All accreditation agencies require
blood sits, the cells fall toward the bottom, leaving a clear
laboratories to clearly define and delineate the proce-
yellow supernate on top called plasma. If a tube does not
dures used for proper collection, transport, and process-
contain an anticoagulant, the blood’s clotting factors are
ing of patient samples and the steps used to minimize
active to form a clot incorporating the cells. The clot is
and detect any errors, along with the documentation of
encapsulated by the large protein fibrinogen. The remain-
the resolution of any errors. The Clinical Laboratory
ing liquid is called serum rather than plasma (Fig. 1-17).
Improvement Amendments Act of 1988 (CLIA 88)27
Most testing in the clinical chemistry laboratory is per-
specifies procedures for specimen submission and
formed on either plasma or serum. The major difference
proper handling, including the disposition of any speci-
between plasma and serum is that serum does not contain
men that does not meet the laboratories’ criteria of ac-
fibrinogen (i.e., there is less protein in serum than
ceptability, be documented.
plasma) and some potassium is released from platelets
(serum potassium is slightly higher in serum than in
Types of Samples plasma). It is important that serum samples be allowed to
Phlebotomy, or venipuncture, is the act of obtaining a completely clot ( 20 minutes) before being centrifuged.
blood sample from a vein using a needle attached to a sy- Centrifugation of the sample accelerates the process of
ringe or a stoppered evacuated tube. These tubes come in separating the plasma and cells. Specimens should be
different volume sizes; from pediatric sizes ( 150 L) to centrifuged for approximately 10 minutes at an RCF of
larger, 7-mL tubes. The most frequent site for venipunc- 1,000g to 2,000g but should avoid mechanical destruc-
ture is the antecubital of the arm. A tourniquet made of tion of red cells that can result in hemoglobin release,
pliable rubber tubing or a strip with Velcro at the end is called hemolysis.
wrapped around the arm, causing a cessation of blood Arterial blood samples measure blood gases (partial
flow and dilation of the veins, making them easier to de- pressures of oxygen and carbon dioxide) and pH.
tect. The gauge of the needle is inversely related to the Syringes are used instead of evacuated tubes because of
size of the needle; the larger the number, the smaller are the pressure in an arterial blood vessel. The radial,
the needle bore and length. An intravenous (IV) infusion brachial, and femoral arteries are the primary arterial
set, sometimes referred to as a butterfly because of the sites. Arterial punctures are more difficult to perform be-
appearance of the setup, is used whenever the veins are cause of inherent arterial pressure, difficulty in stopping
fragile, small, or hard to reach or find. The butterfly is at- bleeding afterward, and the undesirable development of
tached to a piece of tubing, which is then attached to ei- a hematoma, which cuts off the blood supply to the sur-
ther a hub or a tube. Sites adjacent to IV therapy should rounding tissue.29
be avoided; however, if both arms are involved in IV Continued metabolism may occur if the serum or
therapy and the IV cannot be discontinued for a short plasma remains in contact with the cells for any period.
time, a site below the IV site should be sought. The initial Evacuated tubes may incorporate plastic, gel-like mate-
sample drawn (5 mL) should be discarded because it is rial that serves as a barrier between the cells and the
most likely contaminated with IV fluid and only subse- plasma or serum and seals these compartments from one
quent sample tubes should be used for analytic purposes. another during centrifugation. Some gels can interfere
In addition to venipuncture, blood samples can be with certain analytes, notably trace metals, and drugs
collected using a skin puncture technique that customar- such as the tricyclic antidepressants.
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 29

Before separation... Separation

PLASMA
Anticoagulant present-plasma
Whole blood contains fibrinogen
(if anticoagulant SERUM
present) No anticoagulant present

CLOT
Formed encapsulating cells

A B
FIGURE 1-17. Blood sample.

Proper patient identification is the first step in sample the serum in a specified time interval and urine volume
collection. The importance of using the proper collection (often correcting for the surface area). The generic
tube, avoiding prolonged tourniquet application, draw- formula is:
ing tubes in the proper order, and proper labeling of
UV/P (Eq. 1-61)
tubes cannot be stressed strongly enough. Prolonged
tourniquet application causes a stasis of blood flow and where U represents the urine creatinine value in mg/dL,
an increase in hemoconcentration and anything bound V is urine volume per unit of time expressed in mL/min,
to proteins or the cells. Having patients open and close and P is the plasma or serum creatinine value in mg/dL.
their fist during phlebotomy is of no value and may cause This formula expresses the creatinine clearance value in
an increase in potassium and, therefore, should be mL/min (see Chapter 24).
avoided. Intravenous contamination should be consid- Other body fluids analyzed by the clinical chemistry
ered if a large increase occurs in the substances being in- laboratory include cerebrospinal fluid (CSF), paracentesis
fused, such as glucose, potassium, sodium, and chloride, fluids (pleural, pericardial, and peritoneal), and amniotic
with a decrease of other analytes such as urea and creati- fluids. The color and characteristics of the fluid before
nine. In addition, the proper antiseptic must be used. centrifugation should be noted for these samples. Before
Isopropyl alcohol wipes, for example, are used for clean- centrifugation, a laboratorian should also verify that the
ing and disinfecting the collection site; however, this is sample is designated for clinical chemistry analysis only
not the proper antiseptic for disinfecting the site when because a single fluid sample may be shared among sev-
drawing blood alcohol levels. eral departments (i.e., hematology or microbiology) and
Blood is not the only sample analyzed in the clinical centrifugation could invalidate certain tests in those areas.
chemistry laboratory. Urine is the next most common CSF is an ultrafiltrate of the plasma and will, ordi-
fluid for determination. Most quantitative analyses of narily, reflect the values seen in the plasma. For glucose
urine require a timed sample (usually 24 hours); a com- and protein analysis (total and specific proteins), it is
plete sample (all urine must be collected in the specified recommended that a blood sample be analyzed concur-
time) can be difficult because many timed samples are rently with the analysis of those analytes in the CSF.
collected by the patient in an outpatient situation. This will assist in determining the clinical utility of the
Creatinine analysis is often used to assess the complete- values obtained on the CSF sample. This is also true for
ness of a 24-hour urine sample because creatinine out- lactate dehydrogenase and protein assays requested on
put is relatively free from interference and is stable, with paracentesis fluids. All fluid samples should be handled
little change in output between individuals. The average immediately without delay between sample procure-
adult excretes 1 to 2 g of creatinine per 24 hours. Urine ment, transport, and analysis.
volume differs widely among individuals; however, a 4-L Amniotic fluid is used to assess fetal lung maturity
container is adequate (average output is 2 L). It should (L/S ratio), congenital diseases, hemolytic diseases, ge-
be noted that this analysis differs from the creatinine netic defects, and gestational age. The laboratory scien-
clearance test used to assess glomerular filtration rate, tist should verify the specific handling of this fluid with
which compares urine creatinine output with that in the manufacturer of the testing procedure(s).
30 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

Sample Processing problems, and put an action plan in place that contains
policies, procedures, or checkpoints throughout the
When samples arrive in the laboratory, the samples are sample’s journey to the laboratory scientist who is actu-
then processed. In the clinical chemistry laboratory, ally performing the test. Good communication with all
this means correctly matching the blood collection personnel involved helps ensure that whatever plans are
tube(s) with the appropriate analyte request and patient in place meet the needs of the laboratory and, ultimately,
identification labels. This is a particularly sensitive area the patient and physician. Most accreditation agencies
of preanalytic error. Bar code labels on primary sample require that laboratories consider all aspects of preana-
tubes are a popular means to detect errors and to mini- lytic variation as part of their quality assurance plans, in-
mize clerical errors at this point of the processing. In cluding effective problem solving and documentation.
some facilities, samples are numbered or entered into Physiologic variation refers to changes that occur
work lists or a second identification system that is use- within the body, such as cyclic changes (diurnal or cir-
ful during the analytic phase. The laboratory scientist cadian variation) or those resulting from exercise, diet,
must also ascertain if the sample is acceptable for fur- stress, gender, age, underlying medical conditions (e.g.,
ther processing. The criteria used depend on the test in- fever, asthma, obesity), drugs, or posture (Table 1-5).
volved but usually include volume considerations (i.e., Most samples are drawn on patients who are fasting (usu-
is there sufficient volume for testing needs?), use of ally overnight for at least 8 hours). Because overnight and
proper anticoagulants or preservatives, whether timing fasting are relative terms, however, the length of time and
is clearly indicated and appropriate for timed testing, what was consumed during that time should be deter-
and whether the specimen is intact and has been prop- mined before sample procurement for those tests most
erly transported (e.g., cooled or on ice, within a rea- affected by diet or fasting. Patient preparation for timed
sonable period, protected from light, in a tube that is samples or those requiring specific diets or other in-
properly capped). Unless a whole blood analysis is structions must be well written and verbally explained to
being performed, the sample is then centrifuged as pre- patients. Elderly patients often misunderstand or are
viously described and the serum or plasma should be overwhelmed by the directions given to them by physi-
separated from the cells. cian office personnel. A laboratory information tele-
Once processed, the laboratory scientist should note phone number listed on these printed directions can
the presence of any serum or plasma characteristics such often serve as an excellent reference for proper patient
as hemolysis and icterus (increased bilirubin pigment) or preparation.
the presence of turbidity often associated with lipemia Drugs can affect various analytes.30 It is important to
(increased lipids). Samples should be analyzed within ascertain what, if any, medications the patient is taking
4 hours; to minimize the effects of evaporation, samples that may interfere with the test. Unfortunately, many lab-
should be properly capped and kept away from areas of oratorians do not have either the time for or access to this
rapid airflow, light, and heat. If testing is to occur after information and the interest in this type of interference
that time, samples should be appropriately stored. For only arises when the physician questions a result. Some
most, this means refrigeration at 4°C for 8 hours. Many frequently encountered influences are smoking, which
analytes are stable at this temperature, with the excep- causes an increase in glucose as a result of the action of
tion of alkaline phosphatase (increases) and lactate de- nicotine; growth hormone; cortisol; cholesterol; triglyc-
hydrogenase (decreases as a result of temperature labile erides; and urea. High amounts or chronic consumption
fractions four and five). Samples may be frozen at 20°C of alcohol causes hypoglycemia, increased triglycerides,
and stored for longer periods without deleterious effects and an increase in the enzyme gamma-glutamyltrans-
on the results. Repeated cycles of freezing and thawing, ferase and other liver function tests. Intramuscular injec-
like those that occur in so-called frost-free freezers, tions increase the enzyme creatine kinase and the skeletal
should be avoided. muscle fraction of lactate dehydrogenase. Opiates, such as
morphine or meperidine, cause increases in liver and pan-
Sample Variables creatic enzymes, and oral contraceptives may affect many
Sample variables include physiologic considerations, analytic results. Many drugs affect liver function tests.
proper patient preparation, and problems in collection, Diuretics can cause decreased potassium and hypona-
transportation, processing, and storage. Although labora- tremia. Thiazide-type medications can cause hyper-
torians must include mechanisms to minimize the effect glycemia and prerenal azotemia secondary to the decrease
of these variables on testing and must document each pre- in blood volume. Postcollection variations are related to
analytic incident, it is often frustrating to try to control those factors discussed under specimen processing.
the variables that largely depend on individuals outside of Clerical errors are the most frequently encountered, fol-
the laboratory. The best course of action is to critically as- lowed by inadequate separation of cells from serum, im-
sess or predict the weak areas or links, identify potential proper storage, and collection.
CHAPTER 1 ■ BASIC PRINCIPLES AND PRACTICE 31

TABLE 1-5 VARIABLES AFFECTING SELECT CHEMISTRY ANALYTES


FACTOR EXAMPLES OF ANALYTES AFFECTED
Age Albumin, alkaline phosphatase (ALP) (↑ older), phosphorus (P), cholesterol
Gender (↑ Males): Albumin, ALP, creatine, Ca2 , uric acid, creatinine kinase (CK), aspartate amino-
transferase (AST), phosphate (PO4), BUN, Mg2 , bilirubin, cholesterol
(↑ Females): Fe, cholesterol, -globulins, -lipoproteins
Diurnal variation ↑ in AM: adrenocorticotropic hormone (ACTH), cortisol, Fe, aldosterone
↑ in PM: acid phosphatase, growth hormone, parathyroid hormone (PTH), thyroid-stimulating
hormone (TSH)
Day-to-day variation 20% for alanine aminotransferase (ALT), bilirubin, Fe, TSH, triglycerides
Recent food ingestion ↑ Glucose, insulin, triglycerides, gastrin, ionized Ca2
↓ chloride, phosphorus, potassium, amylase, ALP
Posture ↑ When standing: albumin, cholesterol, aldosterone, Ca2
Activity ↑ In ambulatory patients: CK
↑ With exercise: lactic acid, creatine, protein, CK, AST, lactate dehydrogenase (LD)
↓ With exercise: cholesterol, triglycerides
Stress ↑ ACTH, cortisol, catecholamines
Race Total protein (TP) ↑ (black), albumin ↓ (black); IgG 40% ↑, and IgA 20%↑ (black male vs.
white male); → CK/LD ↑ black males; ↑ cholesterol and triglycerides white 40 years old
(glucose-incidence diabetes in Asian, black, Native American, Hispanic)
Require fasting Fasting blood sugar, glucose tolerance test, triglycerides, lipid panel, gastrin, insulin,
aldosterone/renin
Require ICE Lactic acid, ammonia, blood gas (if not cooled ↓ pH, and po2)
(immediate cooling)
Hemolysis ↑ K , ammonia, PO4, Fe, Mg2 , ALT, AST, LD, ALP, catecholamines, CK (marked hemolysis)

Chain of Custody document receipt of the sample, the condition of the


When laboratory tests are likely linked to a crime or ac- sample at the time of receipt, and the date and time it was
cident, they become forensic in nature. In these cases, received. In some instances, one witness verifies the en-
documented specimen identification is required at each tire process and co-signs as the sample moves along. Any
phase of the process. Each facility has its own forms and analytic test could be used as part of legal testimony;
protocols; however, the patient, and usually a witness, therefore, the laboratory scientist should give each sam-
must identify the sample. It should have a tamper-proof ple—even without the documentation—the same atten-
seal. Any individual in contact with the sample must tion given to a forensic sample.

REFERENCES 4. Department of Labor and Occupational Safety and Health


1. National Institute of Standards and Technology (NIST). Reference Administration (OSHA). Occupational exposure to hazardous
on constants, units, and uncertainty. Washington, D.C.: U.S. chemicals in laboratories. Washington, D.C.: OSHA, 1990.
Department of Commerce, 1991/1993. (Accessed December 2007 (Federal Register, 29 CFR, Part 1910.1450.)
at http://physics.nist.gov.) [Adapted from Special Publications 5. National Institute of Standards and Technology (NIST). Standard
(SP) nos. 811 and 330.] reference materials. Washington, D.C.: U.S. Department of
2. National Committee for Clinical Laboratory Standards Commerce, 1991. (Publication no. 260.)
(NCCLS)/CLSI. The reference system for the clinical laboratory: 6. International Union of Pure and Allied Chemistry (IUPAC). In:
Criteria for development and credentialing of methods and mate- McNaught A, Wilkinson A, eds. Royal Society of Chemistry.
rials for harmonization and results; approved guideline. Wayne, Compendium of Chemical Terminology, 2nd ed. Cambridge, UK:
Pa.: NCCLS, 2000. (NCCLS Document no. NRSCL 13-P.) Blackwell Scientific, 1997.
3. American Chemical Society (ACS). Reagent chemicals. 9th ed. 7. Clinical and Laboratory Standards Institute (CLSI)/National
Washington, D.C.: ACS Press, 2000. Committee for Clinical Laboratory Standards (NCCLS). How to
32 PART 1 ■ BASIC PRINCIPLES AND PRACTICE OF CLINICAL CHEMISTRY

define and determine reference intervals in the clinical laboratory; 18. Bowers GN, Inman SR. The gallium melting-point standard. Clin
approved guideline. 2nd ed. Villanova, Pa.: 2000. (Publication no. Chem 1977;23:733.
C28-A2.) 19. Bowie L, Esters F, Bolin J, et al. Development of an aqueous tem-
8. International Organization for Standardization, European perature-indicated technique and its application to clinical labora-
Committee for Standardization (ISO). In vitro diagnostic medical tory instruments. Clin Chem 1976;22:449.
devices—measurement of quantities in samples of biological ori- 20. American Society for Testing and Materials (ASTM). Standards
gin—metrological traceability of values assigned to calibrators specification for volumetric (transfer) pipets. West Conshohocken,
and control material. Geneva, Switzerland: ISO, 2000. Pa.: ASTM, 1993:14.02:500–501. (Publication no. E969–83.)
(Publication no. ISO/TC 212/WG2 N65/EN 17511.) 21. American Society for Testing and Materials (ASTM). Calibration
9. Lide DH, ed. Handbook of chemistry and physics. 88th ed. Boca of volumetric flasks. West Conshohocken, Pa.: ASTM,
Raton, Fla.: CRC Press, 2008. 2003:14:04. (Publication no. E288–94.)
10. National Institute of Standards and Technology (NIST). Standard 22. Seamonds B. Basic laboratory principles and techniques. In: Kaplan
reference materials program. Washington, D.C.: U.S. Department L, Pesce A, Kazmierczak S, eds. Clinical chemistry. Theory, analy-
of Commerce. (Accessed January 2008 at http://ts.nist.gov/mea- sis, and correlation. 4th ed. St. Louis, Mo.: CV Mosby, 2003.
surementservices/referencematerials/CATALOG/.) 23. Bio-Rad Laboratories. Procedure for comparing precision of pipet
11. Carreiro-Lewandowski E. Basic principles and practice of clinical tips. Hercules, Colo.: Bio-Rad Laboratories, 1993. (Product
chemistry. In Bishop M, Duben-Engelkirk J, Fody P, eds. Clinical Information no. 81-0208.)
chemistry, principles, procedures, and correlations. 4th ed. 24. National Committee for Clinical Laboratory Standards (NCCLS).
Baltimore, Md.: Lippincott Williams & Wilkins, 2000. Determining performance of volumetric equipment. Villanova,
12. Clinical Laboratory Standards Institute/National Committee for Pa.: NCCLS, 1984. (NCCLS document no. 18-P.)
Clinical Laboratory Standards. Preparation and testing of reagent 25. American Society for Testing and Materials (ASTM). Standard
water in the clinical laboratory; approved guideline. 4th ed. specification for laboratory weights and precision mass standards.
Wayne, Pa.: 2006. (Publication no. C3-A4.) West Conshohocken, Pa.: ASTM, 2003:14.04. (Publication no.
13. College of American Pathologists (CAP). General laboratory E617–97.)
guidelines; water quality. Northfield, Ill.: CAP, July 1999. 26. Campbell JB, Campbell JM. Laboratory mathematics: Medical and
14. Skoog D, West D, Holler J, Crouch S, eds. Analytical chemistry: biological applications. 5th ed. St. Louis, Mo.: CV Mosby, 1997.
an introduction. 7th ed. Stamford, Conn.: Thompson/Brooks- 27. Centers for Disease Control and Prevention, Division of
Cole, 2000. Laboratory Systems. Clinical laboratory improvement amend-
15. National Committee for Clinical Laboratory Standards ments. CFR Part 493, Laboratory requirements. (Accessed May
(NCCLS)/Clinical Laboratory Standards Institute. Temperature 2007 at http://www.phppo.cdc.gov/clia/regs2/toc.asp.)
calibration of water baths, instruments, and temperature sensors. 28. Clinical Laboratory Standards Institute (CLSI)/National Committee
Villanova, Pa.: NCCLS, 1990. (Publication no. I2-A2.) for Clinical Laboratory Standards. Procedures and devices for the
16. National Institute of Standards and Technology (NIST). collection of diagnostic blood specimens by skin puncture. 5th ed.
Calibration uncertainties of liquid-in-glass thermometers over the Wayne, Pa.: CLSI, 2004. (Approved standard no. HO4-A5.)
range from 20°C to 400°C. Gaithersburg, Md.: U.S. Department 29. Clinical Laboratory Standards Institute (CLSI)/National
of Commerce, 2000. Committee for Clinical Laboratory Standards. Procedures for the
17. National Institute of Standards and Technology (NIST), Wise J. A handling and processing of blood specimens. 3rd ed. Wayne, Pa.:
procedure for the effective recalibration of liquid-in-glass ther- CLSI, 2004. (Publication no. H18-A3.)
mometers. Gaithersburg, Md.: NIST August 1991. (Publication 30. Young DS. Effects of drugs on clinical laboratory tests. 5th ed.
no. SP 819.) Washington, D.C.: AACC Press, 2000.

You might also like