Standard Operating Procedure of Mindray BS-600M Biochemistry Analyzer
Standard Operating Procedure of Mindray BS-600M Biochemistry Analyzer
Revision record
Software
Version Overview of modifications Revised by Date
version
1
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/2
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
Contents
1. Purpose ......................................................................................................................................................... 3
2. Scope ............................................................................................................................................................ 3
3. Operating Environment ................................................................................................................................ 3
4. Safety Information........................................................................................................................................ 3
5. Check before Powering on ........................................................................................................................... 5
6. Powering on the System ............................................................................................................................... 6
7. Checking System Status ............................................................................................................................... 7
8. Tests Setup.................................................................................................................................................. 14
9. Calibration Setup ........................................................................................................................................ 16
10. QC Setup .................................................................................................................................................... 19
11. Loading Biochemical Reagents .................................................................................................................. 21
12. Loading special reagents ............................................................................................................................ 23
13. Calibration .................................................................................................................................................. 25
14. QC test........................................................................................................................................................ 27
15. Sample test ................................................................................................................................................. 29
16. Twin chemistry ........................................................................................................................................... 33
17. Maintenance ............................................................................................................................................... 35
18. Powering off ............................................................................................................................................... 38
19. Check after powering off ............................................................................................................................ 38
2
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/3
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
1. Purpose
Applicable to the daily operation, maintenance, troubleshooting, and learning of the system.
3. Operating Environment
3
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/4
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
unauthorized service person.
2) Spillage of reagent and sample on the instrument may lead to instrument failure and
electric shock.
3) Do not place samples and reagent on the instrument. If spillage occurs, switch off the
power immediately and contact your service department.
4.4 To prevent personal injury caused by moving parts
1) When the system is in operation, do not touch the moving parts of the system. These
moving parts include the sample probe, reagent probe, mixers, cuvette wash station,
rack supply push-in part, and rack storage push-in part.
2) Do not put your finger or hand into any open part when the system is in operation.
3) Do not place racks when the rack supply push-in part is moving forward. Do not
aspirate racks when the retrieval push-in part is moving.
4.5 To prevent personal injury caused by light source
1) When the system is working or standby, do not remove the lamp of the photometer in
case the light beam from the lamp may hurt your eyes.
2) Use the Replace Lamp command in the Maintenance function when replacing the lamp.
Ensure that the lamp is not touched before the countdown for lamp cooling ends, lest it
would be burnt.
4.6 Laser Hazards
1) Light sent by the bar code reader may hurt your eyes. Do not stare into the laser beam
radiated from the bar code reader when the system is in operation.
4.7 Dispose of Waste and Solid Waste
2) Some substances contained in reagent、controls、calibrators、concentrated wash solution
and waste are subject to regulations of contamination and disposal. Follow your local
emissions regulations and consult your reagent manufacturer or distributor.
3) The used solid waste such as reagent bottles and reaction cuvettes are subject to
contamination regulations. Dispose of the waste in accordance with local laws and
regulations.
4) When disposing of waste and solid waste, wear gloves and lab coat, if necessary,
goggles and mask.
4.8 Prevent fire and explosion
1) The alcohol is flammable. Exercise caution when using the alcohol near the analyzer to
4
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/5
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
avoid fire and explosion.
4.9 Cleaning the Analyzer
1) If reagent or sample drops accidentally on the device surface, clean the device surface in
accordance with laboratory safety regulations. If a lot of liquid enters the system,
contact our customer service department or your local distributor.
2) Do not use cleaning agents that may cause risks due to chemical reactions with
components or materials inside the equipment. Do not use strong acidic or alkaline
solutions to clean the equipment.
3) If you have questions about the compatibility of the wash solution with components or
materials in the system, consult our customer service department or your local
distributor.
5. Check before Powering on
6.1 After connecting correctly, the system to the power sockets, switch on the power in the
sequence presented below:
1) Turn on the circuit breaker.
6
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/7
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
7) When the operation unit (computer) is turned on, the operating software will run
automatically.
8) Enter the username and password in the Login window, and then select OK.
9) When the startup check is passed, the main screen shows. The startup procedure is
finished.
6.2 Auto startup Setup
The system allows you to auto awake the system every day by setting up the auto startup
time.
Select " " > System, select Auto Sleep/Awake, and then select Auto Awake to set up
the auto startup time. When the time is reached, the system will be started up automatically if it
is off.
7
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/8
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
Status Description
Biochemistry ISE / SDM status This indicator appears on the left of the status display area. If an
ISE module is installed, the ISE status appears.
The biochemistry status of both analyzing module includes:
Initialize, Wakeup Incubation, Standby, Running, Reagent Load,
Inventory, Stopped, Restore, Sleep, Maintenance, and
Shutdown. The status of the ISE module includes: Initialize,
Restoring, Stopped, Shutdown, Sleep, Wake Up, Standby, and
Running.
Remaining Time This is displayed in status display area, indicating the remaining
time for the test.
Date and time This indicator appears on the upper right corner of the main
screen. It indicates the system date and time.
8
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/9
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
LIS connection status
This indicator appears on the left of the status display area. The
following information is indicated:
Login user This indicator appears in the middle of the status display area. It
indicates the user who logs in the system.
Module Status This icon shows the configured analyzing unit and its status.
Select this icon to display the System Status screen, on which
the detailed status of the analyzer and sample delivery module
are provided.
The module status includes:
Error – red: An error occurs. The analyzer icon appears in
red.
Warning – yellow: A warning occurs. The analyzer icon
appears in yellow.
Normal: The system is running normally.
: Start icon. Select it to display the Start Conditions window , on which you are allowed to
9
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/10
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
start new analysis or resume early testing.
: Emergency stop icon. Select it to stop all tests and other actions. To restore the system into
Standby status, execute the Home command.
: Online help icon. Select it to display the online help of the current window, where you will find
description of parameters and operations.
3) Function window
The function window contains options, buttons and other controls used to perform various functions of
the system.
4) Function buttons area
The function buttons area contains the following buttons used to access various function windows of the
system:
⚫ : used to order sample tests and control samples, and view samples and tests, sample
results and r ack status.
⚫ : used to load view reagent and consumable statuses and inventory check.
⚫ : used to define/edit controls and rules, request QC tests and recall QC results and
summary.
⚫ : Calibration Used to define calibrators, request calibration tests and view calibration
results and data.
⚫ : used to perform maintenance, view system statuses and view edit log and error log.
Select “ ” and the warning list and prompt list pops up.
View Error log
⚫ Select System.
⚫ Select Error Log
10
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/11
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
⚫ Select Error Details
⚫ Troubleshoot according to the recommended method
7.3 Checking Reagent/ Consumable Status
Select “ ”˃ Reagent Overview.
1) The reagent has the following three statuses
⚫ Sufficient: No color is marked for the
⚫ Low inventory: The background is orange.
⚫ Exhausted: The background is red.
2) The Consumable has the following three statuses
⚫ Sufficient: blueish green
⚫ Insufficient for inventory alarm limit: orange.
⚫ Exhausted: red.
View via Reagent Carousel
11
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/12
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
12
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/13
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
13
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/14
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
8. Tests Setup
14
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/15
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
1) Select“ ”-Test-New.
15
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/16
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
2) The user-defined chemistries are set with basic parameters and error detection limits based
on the H version provided by the manufacturer.
9. Calibration Setup
16
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/17
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
2) Select “New”, click "Add Manually" enter the calibrator name, expiration date, lot number
and Number of concentrations, and then click "Calibration chemistry". Select Tests, and then
click “Submit”.
17
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/18
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
4) To set up the dilution factors, select the calibrator and Test, click Dil. Setup, enter the
calibrator concentration, sample volume for analysis (1.5μl~25μl), Aspirated (1.5μl~25μl)
and diluent volume (75μl~ 200 μl), and then click OK. Sample volume and concentration
must be entered.
1) Select " " > "Tests" -Cal Para. Set up calibration math model, K factor and
replicates.
18
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/19
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
2) Select " " > "Select Calibrators". Select calibrators, then click OK.
10. QC Setup
19
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/20
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
2) Select “New,” enter the control name, sample Type, lot number, expiration date and
Manufacture, then select “QC Test”, Click Submit.
3) Choose a control, select “Rack” and “Position”, set up the mean and SD for the control,
and then click Submit.
4) Click QC Rules, set up Westgard multi-rule batch setup, and then click OK.
20
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/21
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
21
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/22
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
22
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/23
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
◼ Standby, Incubation or Sleep: Proceed to the next step.
◼ Running: Select “ ”˃ Reagent List. Select Reagent Stop to request
reagent load. When the countdown for reagent load becomes 0 and the system
status is Reagent Load, a message box pops up. Select OK, and then proceed to
the next step.
2) Remove the reagent carousel cover.
3) Place reagent in an idle position of the reagent carousel and open the rea gent cap.
4) Restore the reagent carousel cover. The system scans all reagent positions
automatically and read the following reagent information from the bar code
24
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/25
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
◼ Running: Select “ ”˃ Reagent List. Select Reagent Stop to request
reagent load. When the countdown for reagent load becomes 0 and the system
status is Reagent Load, a message box pops up. Select OK, and then proceed to
the next step.
2) Select “ ”˃ Reagent List ˃ Consumable
3) Select Physiological Saline in No. 80 position that you want to load.
4) Select Load.
5) Remove the reagent carousel cover.
6) Place the physiological saline in position 80W.
7) Restore the reagent carousel cover.
8) Enter the following information of physiological saline for sample blanks and
sample dilution:
Volume (100 be default)
Bottle type
9) Select Load.
13. Calibration
25
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/26
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
4) Select Calibration Request Options
Full
Reagent blank
5) Select OK.
6) If you want to abort the calibration requests, select No Cal.
7) Calibration tests can be canceled only when they have not been started or are
interrupted.
2) You can select the filter icon “ ”to search for results within a certain period.
The screen shows all the calibrations requested on the day, including the following
information:
26
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/27
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
Reagent carousel
Test name
Result flag
Calibration status
Calibration date and time
Reagent Lot No.
Reagent Serial No.
Math model.
14. QC test
27
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/28
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
2) Select a control.
3) Input Rack ID and Position.
4) Choose a sample cup type to be used by the selected control.
5) Choose desired tests and panels in the test list.
You can select Test View
6) If you want to perform QC on certain reagent, select Assign Rgt. Select a reagent and
select OK on the popup window.
7) Select Submit.
14.2 Recalling QC Results
1) Select Sample Current Results or History Results.
The Current screen displays all incomplete patient samples and control samples, as
well as those ordered on the current day.
The History screen displays all patient samples and control samples ordered before
the current day.
2) Choose a result recall mode:
View samples
View tests
3) Select filter icon“ ” to filter the QC results. Enter the date range and select QC from
28
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/29
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
sample type
4) When recalling results by sample, choose a control in the left list. The right list displays
all results of the control. When recalling results by test, choose a test in the left list. The
right list displays all results of the test
5) Select Details to view result flags, alarm information, reaction curve and data. Select
Return to return to previous screen.
6) Choose the following buttons as needed:
Delete: to delete the QC results of the selected QC test.
1) Select “ ” ˃ S orders.
2) Select Routine or STAT
29
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/30
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
30
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/31
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
31
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/32
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
2) Enter the sample ID, bar code, position, patient name or patient ID to search for samples.
The system supports fuzzy query.
32
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/33
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
The retrieved sample information is automatically displayed on the screen.
3) Select OK. The samples matching the condition are displayed on the screen.
4) Select a function button to perform relevant operations.
Twin chemistries are run and calculated based on the same reagent. Results of the two twin
chemistries are calculated through the same test. Take the reagent HbA1c as an example. It can
be used for running two chemistries in the same test. The chemistry Hb is measured during the
former reaction period, while the chemistry HbA1c measured during the latter one. Finally,
HbA1c (%) can be calculated based on results of the two chemistries.
Similar to normal chemistries, twin chemistries can be run only when the following settings
are finished:
1) defining chemistries
2) assigning reagent position
3) setting up calibrator and calibration rule
4) setting up control and QC rule
16.1 Chemistry definition
Twin chemistries can be defined in the same way as normal chemistries. The following
parameters, however, must be set up differently for two twin chemistries:
1) Sample type
33
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/34
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
2) Normal sample volume, increased sample volume, and decreased sample volume
4) Prozone check
For setup of chemistry parameters, see 8.2
34
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/35
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
16.4 Setting up and requesting quality control
1) QC setup
Twin chemistries must be defined with the same control, and the QC setup of twin
chemistries is the same as that of normal chemistries.
For QC settings, see 10.1
2) Programming controls
Twin chemistries can be requested for quality control in the same way as normal chemistries.
When either of the twin chemistries is requested, the other twin will be requested automatically,
and finally both chemistries will be run for quality control. You are allowed to recall the QC
results and QC reaction curves of the two chemistries.
17. Maintenance
36
Controlled
Title: Standard Operating Procedure for Mindray BS-600M
state: Under Page:/37
Biochemistry Analyzer
control
No.: XX-SH-001 Version: 1.0 Revision: Revision 0 Effective date:
17.2 Weekly maintenance
17.2.1 Biochemistry Maintenance
Biochemistry Maintenance screen
3) Select “ ”> Maintenance, select Auto Perform, select the weekly special
wash and cuvette check procedure, and then select Execute.
4) After maintenance, select the maintenance data button on the Maintenance History
window to view cuvette check results.
5) Close the upper shielding cover of the analyzer.
17.2.4 Phase 7/8 Wash Probe
1) Select “ ”> Maintenance, select Auto Perform, select the phase 7/8 wash
probe procedure, and then select Execute.
2) After maintenance, select the maintenance data button on the Maintenance History
window to view phase 7/8 wash p robe results.
1) Select Exit “ ”on the lower left corner of the main screen.
2) Select Shutdown. The shutdown maintenance dialog box pops up. Select OK. If the
consumables for the shutdown maintenance procedure s are insufficient, the
corresponding dialog box pops up prompting which consumables are insufficient.
Select OK to return to the main screen of the operating software. After replenishing
the consumables, repeat step 1.
1) Remove the sample carousel cover and take out the calibrators, controls and samples.
2) Check the analyzer panel for stains. If yes, use clean soft cloth to wipe away the stains.
3) Check the high-concentration waste tank. If there is waste, empty the waste tank.
38