Chapter 11 Control of Scatter Radiation

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

CHAPTER 14

CONTROL OF SCATTER RADIATION

PRODUCTION OF SCATTER RADIATION  Reduced Scatter Radiation: lowers


radiographic optical density
Two Types of X-rays Responsible for the Optical o Effect: increased radiographic
Density & Contrast on a Radiograph technique to increase OD
 X-rays that pass through the patient without
interacting Patient Thickness
 X-rays that are scattered within the patient  Increased Thickness: increases scattered
through Compton interaction radiation
o Rationale: more x-rays undergo
Remnant X-rays multiple scattering
 X-rays that exit from the patient  Compression Paddle: used to reduce scatter
radiation to the image receptor
Image-Forming X-rays
 X-rays that exit & interact with the image Compression of anatomy improves spatial
receptor resolution & contrast resolution & lowers
patient dose!
Proper Collimation
 Effects: Compression is particularly important during
o Less scatter radiation mammography!
o Reduces patient dose
o Improves contrast resolution CONTROL OF SCATTER RADIATION

Three Factors Contribute to Increased Scatter Contrast


Radiation  One of the most important characteristics of
 kVp, Field Size & Patient Thickness image quality
 The visible difference between light & dark
kVp areas on an image
 Increased kVp: increases scattered radiation  The degree of difference in OD between
o Rationale: the relative number that areas of a radiographic image
undergo Compton interaction
increases Contrast Resolution
o Result: reduced image contrast  The ability to image & distinguish soft
 Decreased kVp: decreases scattered tissues
radiation
o Results: Effect of Scatter Radiation on Image Contrast
 Minimum scatter radiation  High Contrast: use of only transmitted &
 Improved image contrast unattenuated x-rays
 Increased patient dose o Appearance: sharp image
 Due to increased mAs  No Contrast: use of only scattered x-rays
o Appearance: dull gray image
63

Approximately 1% of x-rays incident on the  Moderate Contrast: use of both transmitted


Page

patient reach the image receptor! & scattered x-rays

Field Size Devices That Reduced Scattered Radiation


 Increased Field Size: increases scatter  Beam Restrictors & Grids
radiation

STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO


CHAPTER 14
CONTROL OF SCATTER RADIATION

Beam Restrictors/ Collimators Grid


 Three Types: aperture diaphragm, cones or  Device used to reduce the intensity of scatter
cylinders & variable-aperture collimator radiation in the remnant x-ray beam
 Principal Function: to improve image
Aperture Diaphragm contrast
 Simplest of all beam-restricting devices  Position: between the patient & IR
 Fixed-Lead Opening: for fixed IR size &  Grid Strips: radiopaque material
constant SID o Size: 50 μm wide
 Applications: trauma & dental radiography  Interspace: radiolucent material
o Size: 350 μm wide
Cones & Cylinders  Three Important Dimensions:
 Modification of the aperture-diaphragm o Grid Strip Thickness (T)
 Cylinders: most commonly used o Interspace Material Width (D)
 Restricted Useful Beam: circular shape o Grid Height (h)
 Advantages:
o Reduce scattered radiation Gustav Bucky (1913)
o Improve image contrast  He demonstrated the technique for reducing
 Disadvantage: the amount of scatter radiation that reaches
o Difficulty in alignment of tube, cone the IR
& IR
 Causes: cone cutting Grid Surface X-ray Absorption
 % x-ray absorption: width of grid strip ÷
Variable-Aperture Collimator (width of grid strip + width of interspace) x
 Most commonly used 100

Off-Focus Radiation Grid Ratio


 X-rays produced in the anode but not at the  The height of the grid divided by the
focal spot interspace width
 Formula: grid ratio = h/D
Collimation reduces patient dose & improves o h = the height of the lead strips
contrast resolution! o D = the distance between lead strips
 High Ratio Grid:
Positive-Beam Limiting (PBL) Devices o Advantage Over Low Ratio Grid:
 Features of radiographic collimators that More effective in cleaning up scatter
automatically adjusts the radiation field to radiation
the size of the IR  Rationale: angle of deviation
is smaller
Under no circumstances should the x-ray o Disadvantage: increases patient dose
beam exceed the size of the IR!  General Radiography: 8:1 to 10:1
 Mammography: 4:1 to 5:1
64

Total Filtration
Page

 Equivalent: 2.5 mm Al Grid Frequency


 Inherent (0.5 mm Al) + Added Filtration  The number of grid strips per centimeter
(2.0 mm Al)  High Frequency: less distinct grid lines on a
radiograph
 Range: 25-45 lines/centimeter
STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO
CHAPTER 14
CONTROL OF SCATTER RADIATION

 Formula: grid frequency = 10,000 μm/cm ÷  Symbol: k


(T+D) μm/linepair  Formula: k = image contrast with grid ÷
o T = grid strips thickness/width image contrast without grid
o D = interspace width  Most Grids: k=1.5-2.5
 k=1: no improvement
The use of high frequency grids requires high  Use of Grid: double the image contrast
radiographic technique & results in higher
patient radiation dose! The contrast improvement factor is higher for
high-ratio grid!
Interspace Material
 Purpose: to maintain a precise separation Bucky/Grid Factor
between the delicate lead strip of the grid  Ratio of incident radiation to transmitted
 Composition: radiation through a grid
o Aluminum (Al)  Ratio of patient dose with & without a grid
 Advantages Over Fiber:  Formulas:
 High atomic number o B = incident remnant x-rays ÷
 Produces less visible transmitted image-forming x-rays
grid lines o B = patient dose with grid ÷ patient
 Nonhygroscopic: does dose without grid
not absorb moisture  Purposes:
 Easier to manufacture o To measure how much of an increase
 Disadvantages Over Fiber: in technique will be required
 Increases absorption compared with nongrid exposure
of primary beam o Indicates how large an increase in
 Results: higher mAs patient dose will accompany the use
& higher patient dose of a particular grid
o Plastic fiber  Higher Grid Ratio: higher Bucky factor
 More preferred than Al o Rationale: penetration of scatter
radiation becomes less likely
Grid Strip  Increasing kVp: increases Bucky factor
 Characteristics: thin & high absorption o Rationale: more scatter radiation is
properties produced & it has more difficult time
 Composition: lead (Pb) of penetration
o Advantages:
 Easy to shape As the Bucky factor increases, radiographic
 Inexpensive technique & patient dose increase
 High atomic number proportionately!
 High mass density
GRID TYPES
65
GRID PERFORMANCE
Parallel Grid
Page

Contrast Improvement Factor  Simplest type


 Ratio of radiographic contrast with a grid to  It cleans up scatter radiation in one direction
that without a grid  Grid Strips: parallel
 Purpose: measures improvement in image  Advantage: easiest to manufacture
quality when grids are used  Disadvantage: grid cutoff
STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO
CHAPTER 14
CONTROL OF SCATTER RADIATION

o Occur at: short SID & large area IR  Side of the grid should face
 Optical Density: decreases toward the edge the x-ray tube
of IR  Characteristics: same with parallel
o Exception: exhibit no grid cut off
Grid Cutoff
 The undesirable absorption of primary x-ray Grid Lines
by the grid  The images made when primary x-rays are
 Cause: improper grid position absorbed within the grid strips
 Parallel Grid: most common  Visibility: directly related to the width of the
 Distance To Grid Cutoff: SID ÷ Grid ratio grid strips

Crossed Grid Moving Grid


 It is fabricated by sandwiching two parallel  A Potter-Bucky diaphragm or Bucky
grids together  Grid Used: focused grids
 Grid Strips: perpendicular  Two Basic Types: reciprocating &
 Advantages: oscillating
o Not too difficult to manufacture  Disadvantages:
o Not excessively expensive o Requires bulky mechanism
 Disadvantages:  Result: subject to failure
o Grid cutoff o Increase distance between the patient
o Critical grid positioning & IR
o Tilt table techniques  Results: unwanted increase in
 Possible only if x-ray tube & magnification & image blur
the table are properly aligned (undetectable)
o Exposure technique required is o Introduce motion into cassette-
substantial holding device
 Result: higher patient dose  Result: additional image blur
 Advantages Over Parallel Grid: (undetectable)
o More efficient in cleaning up scatter
radiation Hollis E. Potter (1920)
o Higher contrast improvement factor  Idea: move the grid while the exposure is
than parallel grid with twice grid being made
ratio o Results:
o Advantage increases with increasing  Grid lines disappear
kVp  Increased radiographic
technique
The main disadvantage of parallel & crossed
grids is grid cutoff! Reciprocating Grid
 A moving grid that is motor-driven back &
forth several times during x-ray exposure
66
Focused Grid
 Purpose: to minimize grid cutoff
Page

 Grid Strips: parallel to primary x-ray path Oscillating Grid


 Disadvantages:  A moving grid that oscillates in a circular
o Difficult to manufacture fashion around the grid frame, coming to
o Geometric limitations rest after 20-30 seconds
 Intended focal distance
GRID PROBLEMS
STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO
CHAPTER 14
CONTROL OF SCATTER RADIATION

Central Ray GRID SELECTION


 The x-ray that travels along the center of the
useful x-ray beam Grid Selection
 Depends of Three Interrelated Factors:
Off-Level Grid o kVp
 Only problem in parallel & crossed grid o Degree of cleanup
 Central Ray: not perpendicular to the grid o Patient dose
 Cause: improperly positioned x-ray tube  Focused Grid: most commonly used
 Results: o Considerations: must have properly
o Grid cutoff across image adjusted
o Underexposed  SID indicator
o Light image (low OD)  STD indicator
 Collimators
Off-Center Grid  Higher Ratio Grid:
 Problem in focused grid o For high kVp technique
 Central Ray: perpendicular to the grid o Increases cleanup
 Causes: o Increases patient dose
o Improperly positioned x-ray tube
o Grid is shifted laterally (lateral In general, grid ratios up to 8:1 are
decentering) satisfactory at tube potentials below 90 kVp.
 Results: Grid ratios above 8:1 are used when kVp
o Partial grid cutoff across image exceeds 90 kVp!
o Underexposed
o Light image (low OD) Patient Dose
 Moving Grid: 15% more radiation to patient
Off-Focus Grid than stationary grid
 Problem in focused grid  High- kVp & High-ratio Grid: lower patient
 Cause: improper focal distance dose than low-kVp & low-grid rati
 Result:
o Grid cutoff toward edge of image Grid Selection Factors
 Proper Focal Distance: more important 1. Patient dose increases with increasing
with high ratio grid ratio
o Rationale: less positioning latitude 2. High-ratio grids are used for high-
than low ratio grid kVp
3. Patient dose at high kVp is less than
Upside-Down Grid that at low kVp
 Problem in focused grid Air-Gap Technique
 Cause: improperly positioned grid  An alternative to the use of radiographic
 Result: severe/complete grid cutoff toward grids
67

edge of image
 IR Distance: 10-15 cm from the patient
 Advantages:
Page

Off-Center, Off Focus Grid


o Reduces scatter radiation
 Most common improper grid position o Enhances image contrast
 Result: grid cutoff on one side of image  Disadvantages:
o Image magnification
o Focal-spot blur
STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO
CHAPTER 14
CONTROL OF SCATTER RADIATION

 Applications:
o Chest radiography
 From 180 to 300 cm SID
 Results: little magnification
& sharper image
o Cerebral angiography

APPROXIMATE CHANGE IN
RADIOGRAPHIC TECHNIQUE FOR
STANDARD GRIDS
Grid Ratio mAs Increase kVp Increase
No grid 1x 0
5:1 2x + 8-10
8:1 4x + 13-15
12:1 5x + 20-25
16:1 6x + 30-40

68
Page

STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO

You might also like