Professional Documents
Culture Documents
ClinChemLabMed2024 X
ClinChemLabMed2024 X
net/publication/377066680
CITATIONS READS
0 129
10 authors, including:
All content following this page was uploaded by Giuseppe Lippi on 05 January 2024.
Giuseppe Lippi*, Laura Pighi, Elisa Paviati, Davide Demonte, Simone De Nitto, Matteo Gelati,
Martina Montagnana, Giorgio Gandini, Brandon M. Henry and Gian Luca Salvagno
irreversible myocardial ischemia. In the history of AMI around 0.8 mL each. The first aliquot was used for patient testing,
diagnostics, the way in which patients with suspected whereas the other aliquots were anonymized and then employed for the
analytical study. The entire study was performed always using the same
myocardial ischemia have been triaged has been changed
calibration curve and lot of reagents.
tremendously with the introduction of many different bio-
markers and diagnostic algorithms [3]. Nevertheless, some
milestones have been eventually defined, including the use Analytical assay performance
of cardiac troponin I (cTnI) or T (cTnT) measured with highly
The calculation of the LOB and LOD was performed as follows: [LOB]=
sensitive (hs) immunoassays and incorporated into the
[mean] + 1.645 × [standard deviation (SD)] of 20 repetitions of dilution
so-called “fast-track algorithms” [4, 5]. The guidelines recently
buffer; [LOD]=[LOB] + 1.645 × [SD] of 20 repetitions of a routine plasma
published by the European Society of Cardiology (ESC) at sample with the lowest measurable cTnI value (i.e., around 0.3 ng/L, as
the end of 2023 emphasize several important aspects in derived from the LOD studies), as recommended by Armbruster and Pry
this context [6]. Apart from cardiac troponins being the [8]. For the calculation of the LOB, we have planned a further series of
only biomarkers at the core of the diagnostic reasoning, experiments in which the dilution buffer has been replaced with the
Mindray “Calibrator 0”, as expressly recommended by the manufac-
especially in patients with non-ST-elevation myocardial
turer. The functional sensitivity of the immunoassay was defined as the
infarction (STEMI), the guidelines present new concepts for lowest cTnI value that could be measured with less than 10 % impreci-
serial tests to be used in patients with cardiac troponin sion. Specifically, the value was calculated by measuring six consecutive
levels below the 99th percentile upper reference limit (URL) 1:2 scalar dilutions in dilution buffer (i.e. from 1:2 to 1:64) of a routine
of a population of healthy individuals. In these patients, the plasma sample with baseline cTnI value of 12.6 ng/L. All dilutions were
tested in 10 replicates, with calculation of SD of each dilution and rela-
ESC advocates the use of rapid “rule-in” and “rule-out”
tive imprecision, finally expressed as coefficient of variation (CV %).
algorithms, consisting of baseline measurement of cardiac
troponin and a further assessment 1 or 2 h later, with calcu-
lation of an “absolute” diagnostic delta threshold. These Linearity of the assay
algorithms, validated in large multicenter studies using many
currently available hs immunoassays, require the calculation The assay linearity was assessed by serially diluting a routine plasma
sample with a low value of cTnI (i.e. 0.8 ng/L) with another routine
of method-specific cutoffs derived from analytical and clinical
plasma sample displaying a high value of cTnI (i.e. 9,726.9 ng/mL) at
studies, such that the validation of the analytical performance fixed ratios (0.1:9.9; 0.5:9.5; 1:9; 2:8; 3:7; 4:6; 5:5; 6:4; 7:3, 8:2; 9:1). All serial
of each new immunoassay is an essential prerequisite for its dilutions were measured in duplicate, the theoretical values of cTnI
introduction in the clinical practice. Therefore, the current were estimated from the measured values of undiluted specimens, and
study was designed to evaluate the analytical performance of the linearity was finally reported as Spearman’s correlation coefficient.
the new Mindray hs-cTnI immunoassay on the Mindray
CL-1200i platform. Imprecision studies
cTnI values were expressed as median, interquartile range (IQR) and 99th correlation coefficient (r) of 1.00 (95 % CI, 1.00–1.00; p<0.001)
percentile URL. (Figure 1).
Method comparison
Imprecision studies
The comparison study was performed with the residual plasma
collected from patients admitted to the ED of the University Hospitals of
The results of the imprecision study using three routine
Verona for suspected myocardial ischemia. All samples were immedi-
ately assayed with the local routine immunoassay (Roche hs-cTnT; plasma samples with low (14–17 ng/L), medium (85–92 ng/L)
Roche diagnostics, Basel, Switzerland), and then with the Beckman and high (1,635–1,754 ng/L) cTnI concentrations are sum-
Coulter Access hs-TnI on Access 2 analyzer and Mindray hs-cTnI on marized in Table 1. Specifically, the intra-assay imprecision
CL-1200i. The analytical characteristics of the Beckman Coulter Access was comprised between 1.1 and 1.3 %, the inter-assay
hs-TnI immunoassay have been reported in detail elsewhere [11]. In
imprecision ranged between 5.5 and 8.1 %, while the total
brief, the method is a paramagnetic particle-based chemiluminescent
immunoassay characterized by LOB, LOD and functional sensitivity of
imprecision was 5.6–8.2 %, respectively.
0.14, 0.34 and 1.35 ng/L, respectively, and intra-assay, inter-assay and
overall imprecision of 2.2–2.9 %, 4.6–5.4 % and 5.4–6.1 %, respectively
[11]. The concordance of cTnI values measured with Mindray hs-cTnI
Calculation of the 99th percentile upper
and Access hs-TnI was assessed with Spearman’s correlation and Pass-
ing and Bablok regression, while the absolute difference between values reference limit (URL)
(i.e. percentage bias) was estimated from Bland and Altman plots.
The 99th percentile URL was calculated using residual
Statistical analysis plasma collected from 246 healthy blood donors (149 men
and 97 women; mean age 42 ± 14 years, range 18–66 years).
The statistical analysis was performed with Analyse-it (Analyse-it The median cTnI value was 1.2 ng/L (IQR, 0.6–1.9 ng/L),
Software Ltd, Leeds, UK). The entire study, based on pre-existing whilst the 99th percentile URL was 9.2 ng/L. As predictable,
plasma samples referred for diagnostic testing, was conducted under the median cTnI value was lower in women (0.8 ng/L; IQR,
the ISO 15189:2012 accreditation in accordance with the Declaration of 0.4–1.3 ng/L) compared to men (1.5 ng/L; IQR, 0.9–2.3 ng/L;
Helsinki and under the conditions of relevant local legislation. The
p<0.001), and so was the 99th percentile URL (i.e., 4.3 vs.
study was cleared by the local Institutional Review Board (University
Hospital of Verona, Verona, Italy; SOPAV2, protocol number: 971CESC,
12.3 ng/L). Both values were lower than those indicated by
date of approval: 25 July, 2016). the manufacturer and reported by Li et al. in their evalu-
ation (i.e., 15.3 ng/L in women and 31.3 ng/L in men,
respectively) [7].
Results
Analytical assay performance
Table : With-in run, between-run and total imprecision of the Mindray hs-TnI immunoassay.
Method comparison important factors that can influence the short- and long-
term prognosis of patients with myocardial ischemia. The
The final population included in the method comparison development and introduction of accurate and efficient
study consisted of 265 subjects (166 men and 99 women; cardiac biomarkers such as cardiac troponins into clinical
mean age: 73 ± 14 years) admitted to the local ED for sus- pathways have virtually revolutionized the diagnostic
pected AMI, 166 (62.6 %) displaying cTnI values above the approach in patients with suspected AMI, allowing for rapid
99th percentile URL defined by Mindray (i.e., 24.2 ng/L) and rule-out and more efficient diagnosis. The introduction of hs
176 (66.4 %) with cTnI values above the 99th percentile URL immunoassays was a further step forward in the diagnostic
defined by Beckman Coulter (i.e., 17.5 ng/L), respectively. The reasoning of patients with suspected myocardial infarction
Spearman’s correlation between Mindray hs-cTnI and and enabled the introduction of fast-track algorithms [4, 5],
Access hs-TnI was 0.97 (95 % CI, 0.97–0.98; p<0.001), whilst the which allow a near definitive diagnosis of myocardial
Passing & Bablok linear fit was [Mindray hs-cTnI]=[Access ischemia to be made within 1–2 h of hospital admission. To
hs-TnI × 1.09] − [0.98]. The Bland and Altman plot analysis is be truly efficient, these diagnostic algorithms, which are
shown in Figure 2, revealing a cumulative percentage bias of essentially based on serial measurement of cardiac tropo-
7.2 % (95 % CI, 2.6–11.9 %; p=0.002) between paired patient nins, require a thorough clinical validation aimed at
samples measured with the two immunoassays. reporting the main analytical characteristics (LOB, LOD,
functional sensitivity) for allowing development of assay-
specific algorithms. To this end, we designed this study to
Discussion evaluate and validate the analytical performance of the
novel Mindray hs-cTnI immunoassay.
AMI remains the leading cause of preventable death and The results of our study show that this fully automated
disability worldwide. Today, there is no question that timely method has optimal analytical performance, with LOD, LOB
diagnosis and early revascularization therapy are the most and functional sensitivity of 0.32, 0.35 and 0.35 ng/L,
respectively, and with optimal linearity over a wide range of Research ethics: The study was cleared by the local
clinically significant measurable values (i.e. r=1.00 between Institutional Review Board (University Hospital of Verona,
0.8 and 9,727 ng/L). Notably, we were unable to replicate the Verona, Italy; SOPAV2, protocol number: 971CESC, date of
values found in the earlier study by Li et al. using sample approval: 25 July, 2016).
buffer [7], which reported values of LOD and LOB corre- Informed consent: Not applicable.
sponding to 0.07, 0.21 ng/L, whilst we found a lower value of Author contributions: The authors have accepted respon-
functional sensitivity (i.e., 0.35 vs. 0.51 ng/L). We could only sibility for the entire content of this manuscript and
approximate that LOD with 20 replicates of the Mindray approved its submission.
“Calibrator 0”, which always yielded cTnI values <0.1 ng/L in Competing interests: The instrumentation and reagents
our hands. One possible explanation for the difference used in this study were provided by Mindray, but the
observed between the LOD and LOB, is that our instrument company played no role in the study design; in the collection,
was a Mindray CL-1200i, set by the manufacturer with a analysis, and interpretation of data; in the writing of the
detection limit of 0.1 ng/L. Conversely, Li et al. conducted a report; or in the decision to submit the report for publication.
pilot study with the Mindray CL-8000i [7], which could have Research funding: None declared.
been characterized by extended range of reportable values, Data availability: Data will be available upon reasonable
even if the analytical methodology was essentially identical. request to the corresponding author.
Another possible interpretation is that Li et al. [7] may have
used a different sample matrix (e.g., the lowest calibrator)
for calculating the LOD. In its instructions for use, the References
manufacturer reported values of LOD, LOB and functional
1. Tsao CW, Aday AW, Almarzooq ZI, Anderson CAM, Arora P, Avery CL,
sensitivity of 0.5, 0.7 and 2.4 ng/L, which are however higher
et al. Heart disease and stroke statistics-2023 update: a report from the
than those found in our analytical evaluation. American Heart Association. Circulation 2023;147:e93–621.
Regarding imprecision, the manufacturer states that the 2. De Luca G, Suryapranata H, Ottervanger JP, Antman EM. Time delay to
intra-assay and inter-assay imprecision in routine samples treatment and mortality in primary angioplasty for acute myocardial
ranges between 1.2 and 2.9 % and 2.0–2.9 %, which are lower infarction: every minute of delay counts. Circulation 2004;109:1223–5.
3. Cervellin G, Mattiuzzi C, Bovo C, Lippi G. Diagnostic algorithms for
than the values found in our study (Table 1). The overall
acute coronary syndrome-is one better than another? Ann Transl
imprecision found by Li et al. on two routine serum samples Med 2016;4:193.
with cTnI values of 42 and 9,656 ng/L was between 2.8 and 4. Krintus M, Panteghini M. Laboratory-related issues in the
3.5 %, which is also lower than the total imprecision calcu- measurement of cardiac troponins with highly sensitive assays. Clin
lated on our data (i.e., 5.6–8.2 %). We thus agree with Clerico Chem Lab Med 2020;58:1773–83.
et al. [5], who suggested that Li et al. may have under- 5. Clerico A, Zaninotto M, Plebani M. Rapid rule-in and rule-out protocols
of acute myocardial infarction using hs-cTnI and hs-cTnT methods. Clin
estimated the imprecision of this novel hs-cTnI immuno-
Chem Lab Med 2023 Sep 25. https://doi.org/10.1515/cclm-2023-1010
assay, and that even the LOD and LOB are perhaps too [Epub ahead of print].
optimistic, consistently lower than those of many other 6. Byrne RA, Rossello X, Coughlan JJ, Barbato E, Berry C, Chieffo A, et al.
commercially available cTnI immunoassays. In contrast, the 2023 ESC guidelines for the management of acute coronary syndromes.
findings of our study are more closely aligned with those Eur Heart J 2023;44:3720–826.
7. Li L, Shu X, Zhang L, Xu A, Yang J, Jing Y, et al. Evaluation of the analytical
shown by Clerico et al. for the vast majority of commercially
and clinical performance of a new high-sensitivity cardiac troponin I
available cardiac troponin immunoassays [5]. assay: hs-cTnI (CLIA) assay. Clin Chem Lab Med 2023 Sep 26. https://
Although an excellent correlation was found with doi.org/10.1515/cclm-2023-0529 [Epub ahead of print].
another commercial hs-cTnI method (Access hs-TnI), we 8. Armbruster DA, Pry T. Limit of blank, limit of detection and limit of
observed a statistically significant bias of around 7 % between quantitation. Clin Biochem Rev 2008;29(1 Suppl):S49–52.
paired patient samples measured with the two methods, thus 9. Krouwer JS, Rabinowitz R. How to improve estimates of imprecision.
Clin Chem 1984;30:290–2.
confirming the compelling need for improving the standard-
10. Clinical and Laboratory Standards Institute. Defining, establishing, and
ization of cTnI immunoassays [12]. verifying reference intervals in the clinical laboratory, 3rd ed. Wayne,
In conclusion, although we were unable to confirm the PA, USA: Document EP28-A3C. CLSI; 2010.
otherwise very optimistic analytical characteristics reported 11. Lippi G, Ferrari A, Gandini G, Gelati M, Lo Cascio C, Salvagno GL.
by Li et al. [7], our evaluation of the novel Mindray hs-cTnI Analytical evaluation of the new Beckman Coulter Access high
sensitivity cardiac troponin I immunoassay. Clin Chem Lab Med 2017;
immunoassay on CL-1200i showed that the overall analytical
56:157–61.
performance is comparable to that of other commercially 12. Clerico A, Zaninotto M, Passino C, Padoan A, Migliardi M, Plebani M.
available cTnI techniques, making it a viable alternative to High-sensitivity methods for cardiac troponins: the mission is not over
other methods. yet. Adv Clin Chem 2021;103:215–52.