CALIBRATION CERTIFICATE
[Namo and Address of Customer [Page Number Tort
ULR Number [0250723000007 62F
IMangal Electrical Industries Pvt. Ltd, [Certificate Number [2502/0151/08
[Job Request Number & Date [202315109 & 31/01/2023,
Piot No. E-40 to E-46 & E-46A, [Date of Receipt foavoar2023
[Shree Khatu Shyam Ji Industrial Area, [Calibration Dato joavozr2023
Reengus, Sikar, Rajasthan - 332 404 [Recommended due on ary
(as per customer request)
(Certificate Issue Date josrozr2023
Description & Identification of UUC
Name [Measuring Tape Condition on Receipt oK
Maker [Calibration
Model Freemans cr | SENG Performed At oat
[Calibration Gaanve
Range 0.5000 mm iD. No. |MEIPLIMT-01 arena, Dimension
ce Tmm [Location _[—
Environment Condition
Calibration Procedure Number
[Temperature / Humidity
[20 = °C /50 + 10%RH
ES/SOPINI2T based on OIML R-36-1 and NABL-120-1(A)
Dols of Standard Master Equipment Used:
Nomencature See. Range | Cerificate No. | Vali [ Callvated By
Wessurbg Beale 8 Tape] ase o-tooomm | 22000041361 | zarcarzoze | NABL Cer No, Co-2105
MECHANICAL CALIBRATION - DIMENSION
CALIBRATION RESULTS
Woninal Value on USTs] Avg Value on Standard in tror Expanded Uncaranty
mm rom mm a
200 200065 D088
00 400.188 “0158
oo 600265 “0295
ooo soot 031
1000 1000.54 “054 121 pm mete
2o0 200724 “07%
sooo soon a82 “0392
«oa oons21 “0921
5000 500.058 1054
"Rorrarks=
1. The reported expanded uncertainty is at coverage factor k=2 which corresponds fo @ coverage probability of approximately
99% fora normal astnbution
2 Certificates shal nt be roproduced except infu,
3. Results relaie only to the tems calibrated.
4 Readings are taken without adjustment and readin
without The written epproval of Excellent Services
198 are average of 3 sets of readings.
5. All standard equipments used for callration are traceable to National Standards.
6 UG: Unt Under Caloation.
Engineer-Calibration
Vee ingh
(Format No: QR/M/05, Rev. 00),CALIBRATION CERTIFICATE
a a RASS oT CUSTOM Page Nor Ta
ee name aS TOTOROTTERF
tanga lotic industios Pt Lic Gerfeate Hana 0001510
ob Request nurtor& Date —|p0euo1s 108 OTROS
Plot No. E-40 to E-46 & E-46A, |Date of Receipt }os/02/2023
Shree Kat Shyam J indus Area, Garaton Dao osarozs
Foongu, Sar ojstan 32 40a Recomm Sa nT
as pr customer request)
Gate ame Da SaaS
Description & Identification of UUC
Sanionon
tame [gta Frequency Meter ect ox
aie airaton
ake lerown arn. _[oEs-2oaneora0ans earn ta
ange 70-1000 #2 Do eatraton Frequenay
ne THE [osaton
Tram Coan Gaibraon Provedare Wambo?
eprare Ray Zs £0 TSETEU AH | ESSOPETIT based on15 12a o}.2010 NRSC ADEA
DatalsofStandard/Mestr Equipment Used
Tomei Sr Nurbor | — Carats Number [Vaal —— Galbated By
Tih Product Calter aeere01 | Fuereiososoce-coot [oeean25 | WABL Cot. No. C5
ELECTHOTECHNICAL CAUBRATION
CALIBRATION RESULTS
Galbraon Points 00, 0, 00H
‘Average Value on Stanardin | Average Vale ov UUGT® | —Emarin Sepandad Uncen
Hz Hz Hz + (%)
100, 99 a 0.58
500 40 1 oa
080 os 2 008
—
1 Tm wpe pun! wets set orp fact ed wth cempach cvtign peROY of ppm
2. Catior sal ro tepid ecptn tet Th mite pele Eclat Sanne
2 hosts ton too fons ented
{Renna ant ener od doe a wget Sst fra
4 Alone pat dr caiireten eae hare! Suna
EnginorCatraton Autionzed
6
Ratah@&h Kr. Singh Sarva) Brijesh Kuma\Singh
ae ‘UR |= (Technical Manager)
(Format No.: QR/ET/O1 Rev. 00)CALIBRATION CERTIFICATE
is aR Ce age Robe Tar
OLR Norber cc aSeTOOOOTTEAE
hang lets ndusie Pvt. ceriteate ora snr
eb Requot Nona Dio —osanTernTEOTETERD
Plt Ns. E4010 £488 £404, at of Rese ea
sre Kt Shyam inti clbston Dt osca
Rena, Sar, Rajan 252408 Recorinended a3
(as per customer request) panielanas
cet sve Dao [S523
Descipon &lencaion of UC
Name Digital Ampere Meter AC Condition on Receipt OK
Mako! —[Soeider eee
ake (RRS ESCTT Sno, anezaasea Calan Poorest tab
ange [o-sooa es: 64) ano, [= Calibrator Parenter Curent
ce OA [maior ==
Enaronont Condon Gallatin Procedre Numer
Tampa TH 52°C S5~TORA | —ESISOPETEY tsa on Tapa 9) 2070 and NGL
Ont of Standard/ ster Equipment Used
aT SeMambor | Coniesto ta.“ ] Vay aE
Mae Pewuacatvar | __sooraor | FHOVET00082327" | cgaanzs | WABL Cet No. cOaTEG
LECTROTECRIGAL Ca HRATION
cAUpRATION RESULTS
rg agai on Sanaa Ear SSRI TRAY
Temal Value | Roqured | Wessured value | Conveid hs =
| vate) (hoa)
‘009 1 ‘eos — sons “082 O18
2000 3 som | amar 087 ons
eone 8 cor | anza 123 os
amore:
1. The reported expanded uncertainty is at coverage factor k=2 which corresponds to @ coverage probability of approximately
95% fora normal cstrbution
2. Cattiicates shal not be reproduced excoptin ful, without The writen approval of Excelent Serves.
3. Results relate ony tothe tems caltrate.
4. Readings are takon without adjustment and readings are avorage ofS sets of readings.
5. All standard equipmonts used for calibration are traceable to National Standards.
6. Affced the Excolont Calibration Sticker’Seal onthe Instrument
7. WUC: Unit Under Calitration.
Engineer-Calibration
cs
(Format No.: QR/ET/O1 Rev. 0)CALIBRATION CERTIFICATE
Name and Address of Customer Page Number ror
[ULR Number [eczss72s000001765F
IMangal Electrical Industries Pvt. Ltd. [Cerificate Number 230210181112
[dob Requost Number & Date [2028/0151/12 & 31/01/2003
Plot No, E-40 to E-48 & E-48A, Date of Receipt [osvoa/e023
[Shree Khatu Shyam Ji Industral Area, [Calibration Date fosroare0as
R = jue on
feengus, Sikar, Rajasthan - 332 404 Recommended due iaaeceeda
(as per customer request)
[Cerificate Issue Date [osroara025
Description & Identification of UUC
Name Digital Ampere Meter AC Condition on Receipt 0K
Maker |Schneider Electrie7
Meet eee erie. (Sr.No. [aetezatais2 Calibration Performed At Lab
Range }0-600A (FSD: 6A) id. No. a Calibration Parameter ‘Current
Lc. 041A [Location |-
Environment Condition Calibration Procedure Number
[Temperature 7Humicity 254°C /38- 76V0RH ES/SOPIETIOI based on 1S:1248(part-0):2019 and NABL 120-4
Detalls of Standard / Master Equipment Used
Nomenclature Sr Number Cerificate No. Validity Calibrated By
Mult Product Calibrator 3967801 FUGIET/08062022- |” gayoeaoza NABL Gert. No. CC-2739
‘ELECTRO-TECHNIGAL CALIBRATION
CALIBRATION RESULTS
vue “Avg. value on Standard Error Expanded Uncertainty
NominalValue| Required | Measured Value | — Converted
(A) Value (A) (A) (Ato A) ” zm)
700.0 1 1.0048 100.48 “0.48 015
300.0 3 3.0091 300.91 091 018
600.0 6 sore 601.18 1.18 018
Romarks:-
1. The reported expanded uncertainty is at coverage actor ke2 which corresponds toa coverage probably of aporoximatoly
9556 fora normal distribution
2. Cotcaes shall not be reproduced excep in ul, without The wrlten approval of Excelent Services.
5, Results relate only tothe items calibrated.
4. Readings are takon without adjustment and readings are average of 6 sets of readings.
5. Al standard equipments used for callration are traceable Natona Standard
6. Afixed the Excelent Calibration Sticke/Seal onthe Intumant
7. WG: Unit Under Caran.
Engineer-Calibration i Authorized Signa
Se)
ae
Rate she Kr. Singh (Blsaieuk 5) Brijesh Kuma Singh
(Format No. QR/ET/O1 Re. 00)
AN AS
Re
(Technical Makiager)Coser
CALIBRATION CERTIFICATE
Rami and rane of Cuatomr Page Naor Ta
OLR Nurer oxsSTASOOOOOTERF
Mangal Eocesl industries Pt, Li, eriteate Narbar oasis
lob Request Nuber& Daa [2021713 ESTOTROES
lone. 4010 £48 848A, Bate ef Recent panaroz3
Stee Kat Sys J dual Ae, caibraton Date psarnzs
Reena, Ska, Rojan 22404 Recommended don
|(as per customer request) josvociades
Gest ese Oste [D507 2025
Description & ientiicaion of UC
heme [isa Ampere itor AG Candin on Rei ox
ata” —Sehraaer ocr
key [Rehr EIGCH=T yo, aetozaer7aa Calbraton Peon A a
Range [06004 80:64) fa.No. [= Calraton Parner Curent
ke OFA locaton [=
Enviroment Gondlion Callbraton Procedure Number
FemperatrerRumiaiy [2524°C/35-7SRRA_| — ESSOPIETAT based on1S12spa o} 2070 ad NAGL DET
Details of Standard | Master Equipment Used
Tomenciare Se Wuber —[-—Coresteo. [Vai Caibama ey
Mt Product Caterer | __sesreor | FUOIL08082022TGepgmeas | NABL Cot No. COZTEO
LECTRO-TECIRIGAL CATERATION
CALIBRATION RESULTS
WE Tiga on Standard a Expanded Urea
Woninal aloo] Roqued | Wansured Valo | Convertod
(A) Value (A) A) (Ato A) “a =O
‘000 1 7.088 10088 “088 76
2000 a a.cone 20o8e 08 os
e000 8 ote eo1.02 02 os
Remarks
1. The reported expanded uncertainty is at coverage feciork=2 which corresponds to a coverage probabilty of epproximetely
95% fora normal dstrbution
2. Conifcates shal not be reproduced except in ful, without The witen approval of Excellent Services.
3. Resuits relate only tothe items calibrated.
4. Readings are taken without adjustment and readings are average of $ sets of readings.
5. All stndard equipments used for caloraton ere traceable to National Standards.
6. Aired the Excellent Calsation Sticker!Seal on the Insiument
7, WUC: Unit Under Calibration
Engineer-Calibration
PR: ee ee
(Format no. QR/ET/03 Res. 00)
Brijesh Kumar\singh
(Technical Manager)CALIBRATION CERTIFICATE
Ra an ars CT age ber Tar
OLR Nabe SG 207TEUOOOOTTF
tanga lectialindusties Pt Li carta Nonber soars
ib Resunt nba Dale —pOOUOTET 4 T STRESS
oto, 4010 £488 484, Bate Reset oanarczs
Sie Kat Syam J aust a, cabrton Dats osoartzs
eens, Shar, Rajostan 382404 Reconmended doe on
(as per customer request) renee
ceria ue ba ——SOTAOS
Grecian & enifteation of WUE
fone [Baar Ape Motor AS Condon on Rca ox
iaral —Seneidr ooo?
scar [SRRRarEOGCT ye Jotezaraso CaliatonPoomodA tab
ange [ston 30: 6a) No. [= Caran Parameter caret
ce OA [loeton [=
Envronment Condon Calbraton Procedure Number
ampere Huma fs 4°01 S5- TERA | ESSOPTETI bated on Spo) 2010 and ABLAZE
of Standard Master Equipment Used
SEED LA TET Sa
Vas Peat Caltrar | _seere0r | FURTETINEOE2022— Tsgcsonzs | NABL Gen No. CO-TH0
ELESTROTECHRICAL CAUBEATION
cAUpRaTiON RESULTS
WE Tigre on Sana a Eepanaed UnSTA
oninalVale | Requved | Weasured Val | Convene im a
(A) Value (A) 4A) (Ato A) a
1099 1 7008s 10035 “28s 098
x00 3 aoe | nee 06 o1s
eo00 8 eo | _oone0 099 ous
Remarks:-
1. The reported expanded uncertainty Isat coverage factor k=2 which coiresponds to @ coverage probabil of approximately
95% for a normal distribution.
2. Cettiicates shall not be reproduced except in ful, without The written approval of Excellent Services.
3. Results relate only tothe items calorated
4. Readings are taken without adjustment and readings are average of 5 sets of readings.
5. All stonderd equipments used for calibration are traceable fo National Standards.
6. Affced the Excellent Calration Sticker’Seal onthe Instrument
7, UG: Unit Under Calbration
Engineer-Calibration
(Format No. GR/ET/02 Rew. 0)
Authorized Stgnat
Brijesh Kum
(Technical Marlager)CALIBRATION CERTIFICATE
[Name and Address of Customer [Page Number [rot
[OLR Number [oczss72s000001760F
[Mangal Electrical Industries Pvt. Ltd. (Gerificate Number lasozrorsins
ob Request Number & Date |2023/0161/16 & S1/01/2020
Plot No. E-40 to E-48 & E-46A, [Date of Receipt losroar2023
[Shree Khatu Shyam Ji Industrial Area, (Calibration Date josroaraozs
FReengus, Sikar, Rajasthan - 332 404 Recommended due on
fas per customerrequesty _|°%02/2024
(Certificate Issue Date losroara0es
Description & Identification of UUC
[Condition on
Name Digital Volt Meter AC Receipt oK
Make! Ty, (Calibration
Nake’ |sigmaysop.260|sr.no. Jauvis patemet Al Lab
; = (Calibration jaw
Range | 0-80v(FSD: 400) |iD. No ale Voltage
Lc OAV [Location ==
Environment Condition Calibration Procedure Number
[Temperature Humidity | 254°C 735- 76%RH ESISOPIETION based on 1S:1248(par-6):2010 and NABL “120-4
Details of Standard / Master Equipment Used:
Nomenclature ‘Sr. Number Certificate Number _[ Validity Calibrated By
Muli Product Calibrator 3967801 FLIG/ET/08082022-C00T | 9/08/2025 |_NABL Cert. No. CC-2739
ELECTRO-TECHNICAL CALIBRATION
CALIBRATION RESULTS
WUC “Average on Standard Error Expanded Uncertainty
Nominal Value | Required Value | Measured Value | Converted w w £(%)
Ww) w) w (Vtow)
10.0 50 50.315 10.08 0.315 -0.08 tz
40.0 200 200.53 40.11 0.53 01 0.23
80.0 400 400.87 80.17 -0.87 0.47 0.46
Remarks:-
1. The reported expanded uncertainty is at coverage factor ke2 which corresponds fo @ coverage probability of approximataly
95% fora normal distribution.
2. Comtfeates shal not be reproduced except in ful, without The writen approval of Excellent Services.
3. Results rolate only to the items calibrated
44. Readings are taken without adjustment and readings are average of sets of readings,
5. All standard equipments. used for caliration are traceable to National Standards.
6, Axed the Excolont Calloation Sticker/Seal onthe Instrument
7. WUC: Unit Under Cainration
Engineor-Calibration Authorized:
i
(Format No: QR/ET/O1 Re. 00)
Brijesh Kumar\singh
(Technical Marlager)CALIBRATION CERTIFICATE
[Name and Address of Customer Page Number iret
JOUR Number [eCaseras00000T Ter
lMangal Electrical industries Pvt. Lt. Cerifeate Number 2s02016116
[dob Request Number & Date [2020/0151/16 & S1/0T2025,
Plot No. €-40 to E45 & E-45A, Date of Receipt loswaroz3
[Shiee Knatu Shyam J industrial Area, (Caibration Date losvoarzozs
[Reengus, Skar, Rajasthan - 332 404 Recommended due on
(es percustomerrequesty [22/02/2024
(Certiicate Issue Date osa2035
Description & Wentification of UUC
[Condition on
[Name [Digital insulation Tester Sona oK
ake? calioraton
Maker |motwane /it-51 sco. Eabraton Lab
OaTEIsOOV E Calibration :
Range TAI500V& lipo, | MEIPLIDIT-01 [Eatbraton Resistance
ce. =
Calibration Procedure Number
[Temperature Humaiy — [254°C 735-75%RA ESISOPIETIOT based on IS'1248(par-0):2019 and NABL 1294
Details of Standard / Master Equipment Used
Nomenclature Sr_Number Certificate Number Valty Calbrated By
Standard Megohm Box To0801 ‘CaluiCaL/22-ve/00s —| — 08/06/2028 | NABL Cer. No. CO-Z2T6
Standard Megotm Box 710 corecur7aere22 | 1390272023 _| NABL Cert. No. CC-2287
ELECTRO-TECHNICAL CALIBRATION.
SULTS
Range Value on Standard “Average Value on UUC Error Expanded Uncertainty
+0)
2.069 NO 2.02 Mo “0040 WO 37
@soov 20:21 Ma 20.1 Ma 0221 Ma 37
199,983 ma 199 ma 0.983 Ma 37,
99.98 GO, 8 Ga “0.98 GO 38
@ s000v 487.04 co 498 ca 0.96 Ga 60
1102985 Ta 4102 Ta -0.00885 Ta 60
Romarks:-
1. The reported expanded uncertainty is at coverage factor k=2 which corresponds to a coverage probabiliy of proximately
95% fora narmal aistnbution
2. Cericates shall not be reproduced except in ful, without The writen approval of Excallent Services.
3. Results relate only tothe items calbrated.
44. Readings are taken without adustment and readings are average of 5 sets of readings.
5. All siandard equipments used for callraton are traceable to National Standards,
6, Affxed the Excelent Calibration Sticken/Seal onthe Instrument
Enter olor storied
(Technical Manager)
(Format Na: OR/E1/04 Re. 00)CALIBRATION CERTIFICATE
Name and Address of Customer Page Number ra
[OLR Number lccaearas00000TTTOF
Mangal Electrical Industries Pvt Ltd. Ceniicate number aoaotsin7
Job Request Numbor & Date 20230161717 & STOTRODS
Plot No. E-40 to £45 & E-48A, Date of Receipt osvoaz028
Shee Khatu Shyam si Industal Area, Calivation Date osroaa023
Reengus, kar, Rejastnan- 922 404 Recommended due on a
(as per customer request) _ [0222
Cerificate Tue Dato [OS7020025
Description & identification of UUC
(Condon on
Name tal Turn Ratio Meter coed oK
Calibration
Make! Model stNo. | Performed At ea
T78808 Carton
Range aoteoes TUM |I0.No, — |MEIPLIOTRM-O1 eeeraen Ratio
LC. 0.004 & 0.01 Turn [Location |-—— [Cal. Range. Up to 110 Turn
Environment Gondtion Callbration Procedure Number
Temperature Homa 252 CT S00RA ESISOPIETT based on 1S:1246(par-0}:20T0 and NABL-T2O
Details of Standard / Master Equipment Used
Nomensiatre Sr. Number | Cortfeate Number Calibrated By
Rats Cairaton Standard C2000L16 ‘cxz2igearei01 | Taoare023 | ~RABL Cert No. CODE
ELECTRO-TECHNICAL CALIBRATION
CALIBRATION RESULTS
Range] — SetValue on Standardin | Average Value on UUG in | Error n | ~ Expanded Unceraialy
Tun Tum Ture 4)
0-10.000Tum tt 11.007 0.007 04
@ 72.06 008 oa
0-199, 09Tum “4 4409 0.09 04
110 110.14 014 o4
Remarks:-
1. The reperted expanded uncertainty is at coverage factor k=2 which corresponds to @ coverage probabiliy of approximately
195% for a normal cistibuton
2. Certfcates shall nt be reproduced except in ful, without The writen approval of Excellent Services.
3, Results relate onl to the itome calibrated.
44 Readings are taken without adjustment and readings are average of § sets of readings.
5 All standard equipments used for callration are traceable to National Standards.
6 Affxed the Excellent Calbration Sticker/Sea! onthe Instrument
7. WUC: Unit Under Calibration.
Engineer-Calibration
(Format to. QR/ET/10 Fe. 00)
(Technical Marlager)CALIBRATION CERTIFICATE
Name and Address of Customer Page Number rare
[ULR Number [oc2ss72s00000T TIF
IMangal Electrical Industries Pvt. Ltd. Cerificate Number s020151/18
[Job Request Number & Date _[2029/0161/18 & 317012028
Plot No. £40 to £46 & E-45A, Date of Receipt osrozrz0ze
[Shree Khatu Shyam Ji Industrial Area, [Calibration Date osoar0z3
Reengus, Sika, Rajasthan - 332 404 [Recommended due on ebapoae
(as per customer request)
[Cerificate Issue Date oeroar2025
tion & Mentification of UC
Name [Three Phase Power Analyzer Condition on Receipt OK
Make? Calibration Performed
Maker Yokogawa! W7333E [sco [carD18049e nf Lab
fa Calibration] Voltage Gurrent/ Frequency
ange: [As Pet, Resatts Hee co Parameter Power / Power Factor
CSAs perresurts Cocation [=
Environment Condition Gallbration Procedure Number
Temporaiwre/Humidty | 25 2 £0735- 76%RH_| ESSOP/ETHS based on 1S:1240 (par-/ 2076 and NABL-T2O-T
Details of Standard / Master Equipment Used
Nomenclature Sr.No. Certificate Number Valid Calibrated By
Three Phase Calibration Soures | EDPVIta09.076 | CalufCAL/a2-0a10s8 | ~Ta/a2/2003 | WABL Cen Ne: CODD
"ELECIRO-TECHNICAL CALIBRATION
CAUBRATION RESULTS
[Ac Wottage @ site
Phase Set Value on Standard] Average Value on WUC] Error] Expanded Uncertainty
(Vv) (Vy VY) £(%)
$3.00 07 DOr 023
R 1 150.00 160.11 On O23
i 300.00 30022 022 023
T 63.00 33.015 DOs 023
Y i 150.00, 150.08, 0.08 0.23,
i 300.00 300.18 079 023
1 53.00 e014 Dot 028
8 1 750.00 150.14 0.14 023
i 300.00 SOO A 023
IAC Current @ SoH
Phase (A) 4A) 4A) +(%)
7 2.00 Z0008 T0008 023
R i 10.00 10.004 0.004 0.23,
iT 20.00 z0.0t2 0012 023
1 2.00 Z00Ts D.00t8 023
y 1 7000 70.007 0.007 025
i 20.00 Boor O01 025
i 200 Z0015, T0015 023
8 i 70.00 70008 0.008 023
i 20.00 20.016 0.016 oy
Engineer-Calibration ‘Authorized STypa
Ratandphr Kr. Singh
(Format Na. QR/E1/03 Rev. 00)
(Technical ager)CALIBRATION CERTIFICATE
Calibration Date losroz2023 Page Number 2 of 2
pomereoeen lozi0z/2024 |ULR Number 1cc269723000001771F
(es per customer request)
‘CALIBRATION RE:
Frequency:
Frequency ‘Set Value on Standard ‘Avg. Value on UUG Error Expanded Uncertainty
(Hz) (Hz) (Hz) £0)
50.00 30.05, 0.03 0.14
Frequency 55.00, 35.04, 0.04 014
60.00 60.05 0.05) 026
Power:
Phase Set Value on Standard ‘Avg. Value on UC] Error Expanded Uncertainty
Ww (A) | (PF) | (watt) (Watt) (Watt) =(%)
T 3.00 | 2.00 | 1.0 | 126.0 126.08 0.08. 0.14
R T 150.00_[_10.00_| 1.0 | 1500.0 1501.73, 173 0.14
TH 300.00_[~20.00_| 1.0 | 6000.0 {007.96 7.96 0.26
T 63.00 | 2.00 [1.0 | 126.0 126.11 011 0.14
Y T 150.00 [10.00 _|_1.0 | 1500.0 1501.85 1.83 0.14
i "300.00 | 20.00 [7.0 | 6000.0 6008.18 8.16 0.26
i 63.00_| 2.00 | 1.0 | 1260 126.12 0.12 0.14
8 W 150.00 | 10.00 _| 4.0 | 16000 7502.62 2.62 0.14
ur 300.00 | 20.00 [1.0 [6000.0 6008.14 9.14 0.26
Power Factor:
Phase ‘Set Value on Standard ‘Avg. Value on UUG Error Expanded Uncertainty
(PF) (PF) (PF) +(PF)
Tag 0.500, 0.5002 0.0002 Oi
R tunity 7.0007 7.0003 0.0003. ont
lead 0.500) 5004 0.0004 O41
lag 0.500. 0.5000. 0.0000. ont
x unity 7,000, 7.0002, (0.0002 ont
lead 0.500. 0.5001 0.0007 Ott
Tag 0.500. 0.5001 (0.0001 ott
8 ‘unity 7,000 7.0002 .0002 ott
Tead 0.500, 7.5003 0003 ort
Remarks
1. The reported expanded uncertenty is at coverage factor k=2 which corresponds o a coverage probability of approximately
9596 for normal distribution
2 Corticates shall nat be reproduced except in full without The writen approval of Excellent Services.
3. Results relate ony tothe items calibrated.
4. Readings are taken without adustmant and readings are average of & sets of readings.
5 All standard equipments used for caliration are traceable to National Standards
6 Affixed the Excellent Calbration Sticker/Soa! onthe Instrument
7. WUC: Unit Under Calibration.
Engineer-Calibration ‘Authorized
anlar: singh
(Format No QR/E1/03 Rew. 0)
Brijesh Kumay Singh
(Technical Maljager)CALIBRATION CERTIFICATE
Namo and Address of Customer [Page Number rer
[ULR Number [ocass7a300000T TTF
IMangal Electrical Industries Pvt. Ltd, (Cerificate Number 23021016119
jiob Request Number & Date _|2029/0151/19 & S1/012023,
Pict No, E-40 to E-46 & E-46A, Date of Receipt fosrozreoz3
[Shree Khatu Shyam Ji Industrial Area, [Calbration Date fosrozr2023
IReengus, Sikar, Rajasthan - 332 404 Recommended due on anemia
(as per customer request)
[Gerifcate Issue Date loaroara025
Description & Identification of UC
[Name __ [Transformer Winding Resistance Tester [Condition on Receipt OK
Makes Motweane: Sr.No. |A04299115, [Calibration Performed At Lab
Model XWRM-10A :
Range: As por Results |I0No. [= [Galbration Parameter Resistance
Lc. ‘As perResults [Location [—
Environment Condition Calibration Procedure Number
[Temperature /Humiaity 25 = 40736- TSGRA_| ESISOPIETIOT based on IS:1248\(part-9) - 2079 and NABL 120-4
Details of Standard / Master Equipment Used
Nomenclature Sr. Number Cerificate Number | Validity Calibrated By
Slandard Resistance Box 238 72783 (2022-2028) | 14/12/2023 | NABL Gert. No. CO-3385
ELECTRO-TECHNICAL CALIBRATION
CALIBRATION RESULTS
Current | Range | Value on Standard | Average Value on UUC Error Expanded Uncertainty
Rango + (%)
Bm 700029 mo TOOT mal “BOOTS ma 0.14
toa | 20m 10.0115 ma 10.001 ma 0.00015 ma 014
200m 100.069 ma 100.06 ma 0.008 ma 0.14
20 1.000402 0 1.0003 9000102 0 012
Zone 70.0015 ma TO.0TT ma 0.00965 ma O14
ta | 200m0 100.069 ma 100.04 ma 0.029 ma. ona
20 4.000402 0 4.0002 -0.000202 0 oz
200 10.00104 a 410.000 0 -0.00104 0.12
Zo0me 700.069 mir 100.05 mah 0.019 ma 0.14
tooma | 20 1000402 a 1.0005 0 0.000088 a 0.12
200 10,00104 a 10.002 0 0.00096 0 0.12
2000 92.9949 0 9999.0 0.0049 0 012
Romarks:-
1. The reported expanded uncertainty is at coverage factor k=2 which corresponds to a coverage probabil of approximately
98% for a normal distribution
2. Ceitiicates shal not be reproduced exceptin ful, without The writen approval of Excellent Services.
3. Results relate only tothe items calibrated.
4. Readings are taken without adjustment and readings are avorage of§ sate of readings.
5. All standard equipments used for calibration are traceable to National Stendards.
6. affixed the Exoslient Caliration Sticker / Seal on the instrument
7. WUC: Unit Under Calioration
Engineer-Calibration
aie og &
(Forme No. OR/ET/03 Rev. 00)
Authorized:
\
Brijesh Kumar\singh
(Technical Marlager)