CO2 Incubator
(Water-coat)
User Manual
1
Ⅰ.Application Range:
II.Feature
water jacket structure , with air duct liner and fan forced convection formed to
raise the box temperature uniformity and the equilibrium concentration of CO2。
2.Imported first-class CO2 detected by the infrared waveguide patents and gold-
years.
3.Working indoor CO2 concentrations can range from 0 to 20% of the arbitrary,
and in order to prevent pollution, reduce the concentration of CO2 gas in the
4.When the door is open, automatically turn off fans and CO2 into the valve,
and stop heating.Savings of CO2 gas consumption, and isolated to reduce air
control which uses two probes, respectively, the door control box mild
2
temperature, so that studio temperature high precision, small fluctuations.
9.Box temperature, carbon dioxide concentration and the set of parameters, digital
display. Door heat, carbon dioxide into the air, high-water level, water shortage
10.Super warm, breathe and other protection features to ensure safe operation of
equipment
Ⅳ.Installation
1.The equipment should be placed in a dry, flat, non-toxic and hazardous gas areas.
And should avoid direct sunlight, there is a certain space around equipment to
3
facilitate maintenance.
2.Use of this equipment must be equipped with 99.9% of high purity gas cylinders,
and CO2 loaded valve (above components by users themselves). Cylinders should
be placed in the incubator near the incubator with silicone hoses and the rear "of
3.Incubator at ambient temperature (20 ~ 25) ℃ for the best. Incubator used to set
High Water
Door Pump2
Power Heat Overtemp Pump1
water Shortage Sterilize Light
Heat
Inner temperature
2.On the right of the panel is a set of keys, the functions are as follows:
1) Key : This key is a function key,after powered on, press this button can
enter the temperature setting state;Press this button for five seconds can enter
the other parameters setting state;When the instrument into the set state, the
it.
4)▲Key:The key is the addend key. When setting parameters, press the button
to plus 1 .
5)▼Key:The key is the subtraction key. When setting parameters, press the key
minus 1.
6)◄Key:The key is the left key. When setting parameters, the key can moved to
set data bit. In the boot state, click the button, you can see the indoor temperature.
Note: Please press the key" " after setting parameters, or the parameters can
not been written to memory.
Instructions:
1)When running , the temperature window (top row) shows the inside
temperature.Time window (bottom row) shows remaining time (when time set to
zero, the time window shows set temperature), when the remaining time reaches
0, it shows end, and will stop running.When it stopped running, press the key"
" to rerun.
2)Set inside temperature and run time as follows:①Press the key,the top
modify status →③press the key" " ,the bottom row digital display the set
flashing state → ②Use ▼、▲、◄ modify time status →③press the key"
"to write into memory, and adjust out the next parameter, and so on →④After
setup is complete,press the key for about 5 minutes, exit the setting state,
return to running.
4)Set the PID parameters as follows:Press the key for about 5 minutes,enter
the other parameter setting state first,then press ◄ key for about 5 minutes, enter
PID parameter setting state, the top row digital display each parameters symbols,
the bottom row digital display PID parameters,end-term use in flashing state →
②Use ▼、▲、◄ modify time status →③press the key" "to write into
memory, and adjust out the next parameter, and so on →④After setup is
complete,press the key for about 5 minutes, exit the setting state, return to
running.
6
the water and the box
1. Wiring diagram:
100Ω
Water Temp
COM
water shortage
high water
Main board
Water coat
1# 2# Fan Light Door Heat Steri
Pump Pump Heat lamp
1 2 3 4 5 6 7 8 9 10 11 12
Transformer
220VAC
2. Attention:
1) It will display ER1 if the inner probe is broken circuit or short circuit,ER2
for inner door broken circuit or short circuit,ER3 for short circuit or broken
7
circuit for the probe in the water.ER4 for machine beeping.
2) If the tested temp>setting temp + AL , the machine will beep to alarm
continuously,if the tested temp<setting temp+temp alarm value, the machine
will beep to alarm, press any key to mute.
VI.Order of operations of using
1.Irrigation: Connect one end of the rubber tube to the "inlet,outlet" valve
behind the device,and the other end connected to water, open the "inlet ,
outlet 'valve. Put water into the water jacket by the water pressure. When
injected water is more than 12 liters (the water level exceed the heating tubes), it
can be connected to power, continue to inject water and observe the water level
indicator (water shortage was likely to alarm) until the high water level
indicator---the green light on,stop inject water, turn off "into , outlet "valve .
Or directly to inject water into the overflow until the water overflow, then
release a little water.
2.Please inject pure water or distilled water to prevent scaling inside the water
jacket.
3.Compounding PH liquid and detergent:
Please insert the PH bottle and the detergent bottle.
A.Detergent-The copper sulfate solution
CuSO2 / 5H2O 0.7~1.4g
H2SO4 0.25ml
Musk grass powder (few)
Stilled Water 500ml
B.PH liquid phenol red liquid
According to the room temperature, the solution should be changed once every
two weeks.
Stilled Water 1000ml
0.2%phenol red liquid 3ml
NaHCO3Reference Table
RT(℃) 14 16 18 20 22 24 26
8
NaHCO3(g) 1.87 1.80 1.73 1.67 1.61 1.54 1.48
RT(℃) 28 30 32 34 36 37 38
1.Open the door, make the working chamber clean and put the shelves well.
2.Please inject pure water or distilled water to prevent scaling inside the water
jacket.
3.After installed the incubator and irrigated water, you can operate according to
1)Open the door, clean the studio, put away the shelf.
2)If need humidity, humidity tray will need to join two-thirds of the water, place
3)Open the power switch behind the control box, this time incubator should
4)Press the "sterilize" key, sterile the studio. (Note: when there are cultures inside,
do not press this key!!) After sterilization, press "sterilization" button again, close
5)First of all, setting the required temperature for heating of the incubator. (See
6)Wait until the temperature constant, you can open the door and put the culture
9)Open the switch of the CO2 gas cylinder, adjust the CO2 valve slowly so that
the second pressure gauge indicates about 0.025~0.05Mpa,if the pressure is too
large and the CO2 concentrations rise too fast, it will increase the fluctuation range.
If the pressure is too small and the CO2 concentrations rise too slowly, it will
cause an alarm. Open the regulator valve (clockwise to open) general, it was
opened before it was sent out. Concentration can be seen gradually increased
after 1min, it can reach 5.0% after 10min (when the set value is 5.0%).
1)Close the switch of the CO2 gas cylinder and the reducing valve.
2)Close the CO2 control power and the pump power to make the CO2 control
3)Open the door, remove the humidity tray. And to withstand the door switch by
hand to make the situation open the door to work in the incubator a few minutes
4)Close the door and continue heating about 10min, then turn off the power and
clean inside.
VIII.Attention
1. Equipment should be installed in a clean place where has small changes
in temperature as possible.
2. Before powering on should be fully familiar and understand the
component parts of the manual, to master the correct use.
10
4. When the Incubator has no water, it can not boot, so as not to damage
the internal parts.
12. In order to reduce temperature fluctuations and the uniformity after the
studio closed, so need to delay about 10min after closed (delay time can
be set according to actual by yourself), when it starts to heat up, please
do not frequently open the door!
13. Please exhaust the water in the water jacket when unused for long
periods .
IX.Troubleshooting
Phenomenon Reason Treatment
11
2.Box Temp
ER1 Bad sensor or open studio Check wiring or replace sensor.
Display
Bad sensor or open circuit
ER2 Check wiring or replace sensor.
within the door
Temperature sensor is
ER3 Check wiring or replace sensor.
broken or open circuit
the water level wiring short
ER4 Check wiring and measuring probe.
circuit or disconnect
Plug is not plugged in or
3.No Power Plugged in plug or connected line.
disconnected.
Fuse open Replace the same model fuse
4.Not Heating Set the temperature lower Adjust and set temperature
Door not properly closed or Close the door,replace or adjust the door
the door switch failure
switch.
Heater bad Replace the same model heater
﹕All the herein above should be done by a qualified person, turn off the power
to maintain!!
X.Storage and transportation
The incubator should be stored in excellent room whose relative humidity less
than 80%, non-corrosive gases . Transportation should be shock, moisture and has
12
other necessary protective measures. Can not be inverted transport, handle with
care.
XI.After Sales
Our company has implement three warranty (including repair, replace or
return) of the incubator for one year (except heating element ). Under
warranty any damaged or not working which is because of the quality of the
product ,we responsible for free repair or replace components of the
product ;After one year, in line with the spirit of customer service,we will
make every effort to provide users convenience. Note:If you dismantle the
CO2 test kits inside,this will cancel the Warranty Service
Packing List