Cnr.

Barlinka & Muscat Road ,Saxenburg Park 1


Blackheath,
Cape Town. PO Box 1333, Kuils River.
Ph: 021 9057041/2 Fax: 021 9057038. Laboratory Number:

Pressure Laboratory. 271


Certificate of Calibration Page 1of 1
Calibration performed using measuring equipment traceable to National Measurement Standards
Pressure Laboratory Standards: Serial Number: 14468 Calibration Procedure: PL -1.
Certificate Number:
Customer Details: Calibration Details: P8901-2
Calibration for: BC Gas Installation Description: Pressure Gauge
Address: 10 Pastorale Make: Rhomberg
Street Durbanville Serial Number: Ref P8901-2
Business Park 7550
Measuring Range: Unit of Measure: kPa
Minimum: 0
Condition on receipt: Satisfactory Maximum: 1000
Calibration performed at: Rhomberg lab Resolution: 20
Type of Calibration: Full Date of Calibration: 19/07/2023
Requested recalibration date: Jul-24

Calibration Results.

As received Values.
Calibration Test Calibration Results.
Positions. Only applicable if the Mean Measured Value. Mean error from Test
Increasing Pressure gauge required Increasing Pressure Position.
adjustment. Increasing Pressure
kPa kPa kPa kPa
0 na 0 0
200 na 192 -8
400 na 392 -8
600 na 592 -8
800 na 792 -8
1000 na 992 -8
Decreasing Pressure Decreasing Pressure Decreasing Pressure Decreasing Pressure
800 na 800 0
600 na 600 0
400 na 400 0
200 na 200 0
0 na 0 0

Measurement Uncertainty @ 95% confidence level expressed as a percentage of full scale +/- 0,57%
Temperature at the time of calibration: 21˚C
Remarks:

Calibrated by: M.Lubando

A.A Barnes
Technical Signatories: A.A Barnes
This document is digitally signed with an Advanced Electronic Signature in accordance with the regulations and compliant with the Electronic
Communications and Transactions Act, 2002 This certificate is issued in accordance with SANS ISO/IEC 17025 and the policies of Rhomberg
Instruments. The measurement results recorded on this certificate were correct at the time of calibration and relate only to the item calibrated.
The subsequent accuracy will depend on such factors such as care, handling and frequency of use. It is recommended that recalibration be
undertaken at an interval that will ensure that the instrument remains within the desired limits. The reported expanded uncertainty is based on
a standard uncertainty multiplied by a coverage factor k=2 providing a level of confidence of approximately 95%.The uncertainty of
measurement has been estimated in accordance with the principles defined in the GUM, Guide to Uncertainty of Measurement, ISO
Geneva1993. The applicant hereby indemnifies, holds harmless and absolves Rhomberg Instruments and the Pressure laboratory, from any
damage whatsoever and any legal liability in the event of a mistake in the services performed for the applicant. This certificate may not be
reproduced other than in full.
End of Certificate
doc: R PL- 5 Rev: 5a January 2015