whole blood, serum, or plasma. This test consists of two specimens or kits are handled.

components, an IgG component and an IgM component. In  Do not use the test if the pouch is damaged.
the IgG component, anti-human IgG is coated in test line  Handle all specimens as if they contain infectious
region 1 of the test. During testing, the specimen reacts agents. Observe established precautions against
with Dengue antigen-coated particles in the test strip. The microbiological hazards throughout testing and
ONE STEP Anti-Dengue (IgM & IgG) Tri-line Test mixture then migrates upward on the membrane follow the standard procedures for proper
(Whole Blood/Serum/Plasma) chromatographically by capillary action and reacts with the disposal of specimens.
For In-Vitro diagnostic and professional use only anti-human IgG in test line region 1. If the specimen  Wear protective clothing such as laboratory coats,
contains IgG antibodies to Dengue, a colored line will disposable gloves and eye protection when
Store at 2-8C appear in test line region 1. In the IgM component, anti- specimens are being tested.
ligand is coated in test line region 2 of the test. During  The used test should be discarded according to
testing, the specimen reacts with ligand anti-human IgM. local regulations.
INTENDED USE Dengue IgM antibodies, if present in the specimen, reacts  Humidity and temperature can adversely affect
The ATLAS ONE STEP Anti Dengue (IgM & IgG) Tri Line Test is with the ligand anti-human IgM and the Dengue antigen- results.
a rapid immunochromatographic assay for the qualitative coated particles in the test strip, and this complex is
detection of antibodies (IgM & IgG) TO DENGUE VIRUS (DV) in captured by the anti-ligand, forming a colored line in test STORAGE AND STABILITY
human whole blood , serum or plasma as an aid in the line region 2.  The kit should be stored at 2-30°C until the expiry
diagnosis of primary and secondary Dengue infections. Therefore, if the specimen contains Dengue IgG antibodies, date printed on the sealed pouch.
a colored line will appear in test line region 1. If the  The test must remain in the sealed pouch until use.
INTRODUCTION specimen contains Dengue IgM antibodies, a colored line  Do not freeze.
Dengue is a flavivirus, transmitted by Aedes aegypti and will appear in test line region 2. If the specimen does not  Care should be taken to protect the components
Aedes albopictus mosquitoes.1 It is widely distributed contain Dengue antibodies, no colored line will appear in of the kit from contamination. Do not use if there
throughout the tropical and subtropical areas of the either of the test line regions, indicating a negative result. is evidence of microbial contamination or
world,1 and causes up to 100 million infections annually.2 To serve as a procedural control, a colored line will always precipitation. Biological contamination of
Classic Dengue infection is characterized by a sudden onset change from blue to red in the control line region, dispensing equipments, containers or reagents
of fever, intense headache, myalgia, arthralgia and rash. indicating that the proper volume of specimen has been can lead to false results.
Primary Dengue infection causes IgM antibodies to added and membrane wicking has occurred.
increase to a detectable level in 3 to 5 days after the onset SPECIMEN COLLECTION AND PREPARATION
of fever. IgM antibodies generally persist for 30 to 90 MATERIALS
 The Dengue Rapid Test Device (Whole
days.3 Most Dengue patients in endemic regions have  Test cards/ individually foil pouched with a Blood/Serum/Plasma) can be performed using whole
secondary infections,4 resulting in high levels of specific desiccant blood, serum, or plasma.
IgG antibodies prior to or simultaneous with IgM  Sample diluents  To collect Fingerstick Whole Blood Specimens:
response.5 Therefore, the detection of specific anti-Dengue  Dropper  Wash the patient’s hand with soap and warm water
IgM and IgG antibodies can also help to distinguish  Package Insert or clean with an alcohol swab. Allow to dry.
between primary and secondary infections. MATERIALS REQUIRED BUT NOT PROVIDED  Massage the hand without touching the puncture
The Dengue Rapid Test Device (Whole  Timer site by rubbing down the hand towards the fingertip
Blood/Serum/Plasma) is a rapid test that utilizes a  Lancets of the middle or ring finger.
combination of Dengue antigen coated colored particles for  Specimen collection container  Puncture the skin with a sterile lancet. Wipe away
the detection of IgG and IgM Dengue antibodies in human  Centrifuge the first sign of blood.
whole blood, serum, or plasma.
 Gently rub the hand from wrist to palm to finger to
PRECAUTIONS form a rounded drop of blood over the puncture site.
PRINCIPLE  For professional in vitro diagnostic use only. Do
The Dengue Rapid Test Device (Whole  Add the Fingerstick Whole Blood specimen to the
not use after the expiration date. test device by using a dropper or micropipette
Blood/Serum/Plasma) is a qualitative membrane-based  Do not eat, drink or smoke in the area where the
immunoassay for the detection of Dengue antibodies in measuring 10uL. The dropper provided with the test
dispenses approximately 10uL in one drop even if blood to the specimen well (S) of the test device, then region (C) changes from blue to red. No line appears in
more blood is aspirated in the dropper. add 3 drops of buffer (approximately 90 µL) and start the test line regions 1 or 2 (T1 or T2).
 Separate serum or plasma from blood as soon as timer. See illustration below.
possible to avoid hemolysis. Use only clear, non- 3. Wait for the colored line(s) to appear. Read results at 10
hemolyzed specimens. minutes. Do not interpret the result after 20 minutes.
 Testing should be performed immediately after
specimen collection. Do not leave the specimens at
room temperature for prolonged periods. Serum and
INVALID: Control line (C) is still
plasma specimens may be stored at 2-8°C for up to 3
completely or partially Blue, and fails to completely
days. For long-term storage, specimens should be kept
change from blue to red. Insufficient buffer volume or
below -20°C. Whole blood collected by venipuncture
incorrect procedural techniques are the most likely
should be stored at 2-8°C if the test is to be run within 2
reasons for control line failure. Review the procedure and
days of collection. Do not freeze whole blood specimens.
repeat the procedure with a new test device. If the
Whole blood collected by fingerstick should be tested INTERPRETATION OF RESULTS problem persists, discontinue using the test kit
immediately.
immediately and contact your local distributor.
 Bring specimens to room temperature prior to testing.
Frozen specimens must be completely thawed and IgG POSITIVE:* The colored line in the control line
mixed well prior to testing. Specimens should not be region (C) changes from blue to red, and a colored line QUALITY CONTROL
frozen and thawed repeatedly. appears in test line region 1 (T1). The result is positive for An internal procedural control is included in the test. A
 If specimens are to be shipped, they should be packed Dengue virus specific-IgG and is probably indicative of colored line changes from blue to red in the control line
in compliance with local regulations covering the secondary Dengue infection. region (C), confirming sufficient buffer volume and
transportation of etiologic agents. adequate membrane wicking.
Control standards are not supplied with this kit; however, it
PROCEDURE is recommended that positive and negative controls be
IgM POSITIVE:* The colored line in the control tested as a good laboratory practice to confirm the test
Allow the test device, specimen, buffer, and/or controls to
line region (C) changes from blue to red, and a colored procedure and to verify proper test performance.
reach room temperature (15-30°C) prior to testing.
line appears in test line region 2 (T2). The result is positive
1. Bring the pouch to room temperature before
for Dengue virus specific-IgM antibodies and is indicative
opening.Remove the test device from the sealed PERFORMANCE CHARACTERISTICS
of primary Dengue infection.
pouch and use it as soon as possible. Clinical Sensitivity, Specificity and Accuracy
2. Place the test device on a clean and level surface. The Dengue Rapid Test Device (Whole
For Serum or Plasma Specimens: Blood/Serum/Plasma) has been evaluated with specimens
Hold the dropper vertically, draw the specimen up to the IgG AND IgM POSITIVE:* The colored line in the obtained from a population of symptomatic and
Fill Line (approximately 5 μL), and transfer the specimen control line region (C) changes from blue to red, and two asymptomatic individuals. Results were confirmed by a
to the specimen well (S) of the test device, then add 3 colored lines should appear in test line regions 1 and 2 (T1 leading commercial Dengue ELISA test.
drops of buffer (approximately 90 µL) and start the timer. and T2). The color intensities of the lines do not have to
See illustration below. Avoid trapping air bubbles in the match. The result is positive for IgG & IgM antibodies and Dengue Rapid Test Device vs. ELISA
specimen well (S). is indicative of secondary Dengue infection. Dengue Result Ig M IgG
For Whole Blood (Venipuncture/Fingerstick) Specimens: *NOTE: The intensity of the color in the test line region(s) Infection
To use a dropper: Hold the dropper vertically, draw the (T1 and/or T2) will vary depending on the concentration of Primary Positive 14 0
specimen 0.5-1 cm above the Fill Line, and transfer 1 Dengue antibodies in the specimen. Therefore, any shade of Infection
drop of whole blood (approximately 10 µL) to the color in the test line region(s) (T1 and/or T2) should be Negative 3 17
specimen well (S) of the test device, then add 3 drops of considered positive. Total 17 17
buffer (approximately 90 uL) and start the timer. See Relative 82.4% 0%
illustration below. Sensitivity
To use a micropipette: Pipet and dispense 10 µL of whole NEGATIVE: The colored line in the control line
Secondary Positive 39 55 1.The Dengue Rapid Test Device (Whole
Infection Blood/Serum/Plasma) is for in vitro diagnostic use only. REFERENCES
Negative 16 0 The test should be used for the detection of Dengue 1. Guyader, M., Emerman, M., Sonigo, P., et al.
Total 55 55 antibodies in whole blood, serum or plasma specimens Genome organization and transactivation of the
Relative 70.9% >99.0 only. Neither the quantitative value nor the rate of human immunodeficiency virus type 2. Nature,
Sensitivity % increase in Dengue antibody concentration can be 326:662-669. 1987.
Non-Dengue Positive 0 0 determined by this qualitative test. 2. Curran, J.W., Morgan. W.M., Hardy, A.M., et al. The
Infection 2. The Dengue Rapid Test Device (Whole epidemiology of AIDS: Current status and future
Negative 378 378 Blood/Serum/Plasma) will only indicate the presence of prospects. Science 229:1352-1357. 1985.
Total 378 378 Dengue antibodies in the specimen and should not be 3. Weber, J.N., Weiss, R.A., Roberts, C., et al. Human
Relative >99.0 >99.0 used as the sole criteria for the diagnosis of Dengue. immunodeficiency virus infection in two cohorts of
Specificity % % 3. In the early onset of fever, anti-Dengue IgM homosexual men: Neutralising sera and association
For the primary and secondary infection, the overall concentrations may be below detectable levels. For of anti-gag antibody with prognosis, Lancet 1:119-
sensitivity is 95.8%, the overall specificity is >99.0% and the primary infection, an IgM antibody-capture enzyme- 124. 1987.
overall accuracy is 99.3%. linked immunosorbent assay (MAC-ELISA) showed that 4. Clavel, F., Guetard. D., Brun-Vezinet, F., et al.
80% of the Dengue patients tested exhibited detectable Isolation of a new human retrovirus from West
Precision levels of IgM antibody by the fifth day after infection, African patient with AIDS. Science 233:343-346.
Intra-Assay and 99% of the patients tested IgM positive by day 10.5 1986.
Within-run precision has been determined by using 10 It is recommended that patients be tested within this
replicates of four specimens: a negative, an IgG positive, an time. Atlas Medical
IgM positive and an IgG/IgM dual positive. The specimens 4. For the secondary infection, a low molar fraction of anti- William James House,
were correctly identified >99% of the time. Dengue IgM and a high molar fraction of IgG that is Cowley Road, Cambridge, CB4 4WX, UK
broadly reactive to flaviviruses characterize the Tel: ++44 (0) 1223 858 910
Inter-Assay antibodies. 5 The IgM signal may be faint and the cross Fax: ++44 (0) 1223 858 524
Between-run precision has been determined by 10 reaction in the region of IgG line may appear. PPI898 A01
independent assays on the same four specimens: a 5. Serological cross-reactivity across the flavivirus group Rev A (02.11.2015)
negative, an IgG positive, an IgM positive and an IgG/IgM (Dengue 1, 2, 3 & 4, St. Louis encephalitis, West Nile
virus, Japanese encephalitis and yellow fever viruses) is Product For in-vitro
dual positive. Three different lots of the Dengue Rapid Test Reference No. diagnostic use.
Device (Whole Blood/Serum/Plasma) have been tested common.6,7,8 Positive results should be confirmed by
using these specimens. The specimens were correctly other means.
6. The continued presence or absence of antibodies cannot Caution. Store at 2 - 30°C.
identified >99% of the time.
be used to determine the success or failure of therapy.
Read product
EXPECTED VALUES 7. Results from immunosuppressed patients should be Number of tests
insert before
Primary Dengue infection is characterized by the presence interpreted with caution. in the pack.
use.
of detectable IgM antibodies 3-5 days after the onset of 8. As with all diagnostic tests, all results must be
interpreted together with other clinical information Lot (batch)
infection. Secondary Dengue infection is characterized by Manufacturer.
number.
the elevation of Dengue-specific IgG. In the majority of the available to the physician.
cases, this is accompanied by elevated levels of IgM.5 The 9. If the test result is negative and clinical symptoms persist, Manufacturer
additional testing using other clinical methods is Expiry date. telephone
Dengue Rapid Test Device (Whole Blood/Serum/Plasma) number.
has been compared with a leading commercial Dengue recommended. A negative result does not at any time
ELISA test, demonstrating sensitivity of 82.4% for IgM in preclude the possibility of Dengue infection. Manufacturer
fax number.
primary infection and >99.0% for IgG in secondary infection.