Elsevier. Finite Element Analysis of Mechanical Behavior of Human Dysplastic Hip Joints A Systematic Review

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 10

Osteoarthritis and Cartilage xxx (2016) 1e10

Review

Finite element analysis of mechanical behavior of human dysplastic


hip joints: a systematic review
B. Vafaeian y, D. Zonoobi z, M. Mabee z, A.R. Hareendranathan z, M. El-Rich y x,
S. Adeeb y a, J.L. Jaremko z *
y Department of Civil and Environmental Engineering, University of Alberta, 7-203 Donadeo Innovation Centre for Engineering, 9211-116 Street, Edmonton,
Alberta, T6G 1H9, Canada
z Department of Radiology and Diagnostic Imaging, University of Alberta, 2A2.41 WMC, 8440-112 Street, Edmonton, Alberta, T6G 2B7, Canada
x Department of Mechanical Engineering at Khalifa University (UAE), United Arab Emirates

a r t i c l e i n f o s u m m a r y

Article history: Developmental dysplasia of the hip (DDH) is a common condition predisposing to osteoarthritis (OA).
Received 8 April 2016 Especially since DDH is best identified and treated in infancy before bones ossify, there is surprisingly a
Accepted 28 October 2016 near-complete absence of literature examining mechanical behavior of infant dysplastic hips. We sought
to identify current practice in finite element modeling (FEM) of DDH, to inform future modeling of infant
Keywords: dysplastic hips. We performed multi-database systematic review using PRISMA criteria. Abstracts
Human dysplastic hip
(n ¼ 126) fulfilling inclusion criteria were screened for methodological quality, and results were analyzed
Finite element modeling
and summarized for eligible articles (n ¼ 12). The majority of the studies modeled human adult
Infant hip dysplasia
Systematic review
dysplastic hips. Two studies focused on etiology of DDH through simulating mechanobiological growth
of prenatal hips; we found no FEM-based studies in infants or children. Finite element models used
either patient-specific geometry or idealized average geometry. Diversities in choice of material prop-
erties, boundary conditions, and loading scenarios were found in the finite-element models. FEM of adult
dysplastic hips demonstrated generally smaller cartilage contact area in dysplastic hips than in normal
joints. Contact pressure (CP) may be higher or lower in dysplastic hips depending on joint geometry and
mechanical contribution of labrum (Lb). FEM of mechanobiological growth of prenatal hip joints revealed
evidence for effects of the joint mechanical environment on formation of coxa valga, asymmetrically
shallow acetabulum and malformed femoral head associated with DDH. Future modeling informed by
the results of this review may yield valuable insights into optimal treatment of DDH, and into how and
why OA develops early in DDH.
© 2016 Osteoarthritis Research Society International. Published by Elsevier Ltd. All rights reserved.

Introduction cartilage contact pressure (CP), subluxation3e5, and OA3,5e7. In


addition to age, trauma, activity, weight, and genetics, OA in DDH
Developmental dysplasia of the hip (DDH) is a common relates to anatomic hip abnormalities5e7. Since the mechanical
anatomic deformity leading to hip dysfunction and osteoarthritis environment strongly affects bone growth and development6,8e13,
(OA). Present in 1e3/1000 live births1, DDH accounts for one-third understanding hip mechanical behavior can provide insight into
of hip replacement surgeries in patients under 60 years old2. Un- how premature OA occurs6,7,14, and help optimize treatment10,15e18.
treated dysplastic hips are associated with mechanical instability, Hip mechanical behavior has been studied through experi-
limited mobility, muscle imbalance, abnormal joint load, increased mental measurements, theoretical models, and computational
methods including multibody dynamics (MBD), discrete element
* Address correspondence and reprint requests to: J.L. Jaremko, Department of analysis (DEA), and finite element modeling (FEM)19e21. Experi-
Radiology and Diagnostic Imaging, University of Alberta, 2A2.41 WMC, 8440-112 mental studies14,22e27 measure hip CP directly using sensors, but
Street, Edmonton, Alberta, T6G 2B7, Canada. Fax: 1-780-4073853.
these are invasive and it is difficult to maintain physiological con-
E-mail addresses: vafaeian@ualberta.ca (B. Vafaeian), zonoobi@ualberta.ca
(D. Zonoobi), mmabee@ualberta.ca (M. Mabee), hareendr@ualberta.ca ditions during measurement20,21. Theoretical models6,14,28e33 esti-
(A.R. Hareendranathan), elrich@ualberta.ca (M. El-Rich), adeeb@ualberta.ca mate articular surface CP with fundamental methodological
(S. Adeeb), jjaremko@ualberta.ca (J.L. Jaremko). simplifications limiting their predictive validity20,34. MBD35,36 can
a
Website: http://sameradeeb.srv.ualberta.ca.

http://dx.doi.org/10.1016/j.joca.2016.10.023
1063-4584/© 2016 Osteoarthritis Research Society International. Published by Elsevier Ltd. All rights reserved.

Please cite this article in press as: Vafaeian B, et al., Finite element analysis of mechanical behavior of human dysplastic hip joints: a systematic
review, Osteoarthritis and Cartilage (2016), http://dx.doi.org/10.1016/j.joca.2016.10.023
2 B. Vafaeian et al. / Osteoarthritis and Cartilage xxx (2016) 1e10

only estimate joint reaction force (JRF), not CP. DEA34,37e41 and
FEM19,20 are both capable of simulating articular surface CP, contact
area, and JRF. Although more computationally intensive, FEM can
also model the sliding contact mechanism inside the hip joint,
cartilage deformations in all directions, bone deformations, mate-
rial anisotropy, and stresses inside tissue layers19,20.
Mechanical behavior of adult normal hips has been well studied
through FEM, with generally good agreement with experimental
outcomes19e21,27. In contrast, FEM-based studies on dysplastic hips
have not been specifically reviewed. A dedicated review on this
complex topic can clarify the diverse methods in use to study DDH
from prenatal initiation to adult morbidity. Notably we could find
no published FEM of infant (post-birth) dysplastic hips. Since the
most important time to diagnose and treat DDH to prevent OA is in
infancy, this is an important gap in the literature. The objective of
this systematic review was to determine the state-of-the-art in FEM
of human dysplastic hips, with a view to making informed rec-
ommendations for future FEM of infant dysplastic hips.

Materials and methods

Search strategy

We performed a systematic review, searching PubMed, Medline


and Elsevier (ScienceDirect) for relevant peer-reviewed articles
with English-language abstracts published from 1946 to June 25th,
2016 based on these keywords: (“dysplasia” OR “hip dysplasia” OR
“dysplastic hip”) AND (“finite element” OR “computational” OR
“computer simulation” OR “mechanics”). Reference lists of relevant
publications were also reviewed to avoid missing relevant articles.

Selection criteria

Fig. 1. Study selection flow chart.


Titles and abstracts of all potentially relevant articles were
reviewed. Article selection was performed by an engineer/research
associate (BV) and confirmed in consensus with a radiologist/
(protrusio, femoroacetabular impingement44,46), we focused on the
biomedical engineer (JJ). Full-text articles were included when they
results obtained in dysplastic hips. All studies contained FEM of
(1) demonstrated mechanical behavior of the human dysplastic hip
either adult hips or simplified prenatal hips. We found no studies
through FEM of full or partial three-dimensional (3D) geometries of
regarding post-birth pediatric hips.
natural and/or post-acetabular-osteotomy dysplastic hip joints;
and/or (2) employed FEM to study formation of human dysplastic
hip morphology. We excluded articles related to total hip arthro- Discussion
plasty, and those in which hip mechanics were investigated by
methods other than FEM, i.e., theoretical models, MBD, and DEA. The following sections summarize methods and results from the
However, noting that the closest alternative to FEM is DEA and to reviewed studies modeling dysplastic hips.
capture merits of DEA in this area, we included a brief review on
applications of DEA for studying human dysplastic hips at the end Geometry
of this paper.
To better understand potential biases, we stratified studies by Hip FEM requires computational geometries for the pelvis and
data source (patient-specific or virtual) and structures modeled. proximal femur (PF), from either (1) patient-specific volumetric
The summary measures reviewed were differences in model out- data, usually from computed tomography (CT) scan, or (2) idealized
puts between normal and dysplastic hips. computer-aided design (CAD) models (Table I). Patient-specific CT
data for both normal and dysplastic hips were employed in several
Results studies42,43,45. Others had only access to either normal17 or
dysplastic15,47,48 patient-specific CT data and generated models of
We found 76 unique articles. 29 merited full-text review, and 12 dysplastic or close-to-normal hips by deforming the acetabular rim
met eligibility criteria (Fig. 1). or changing morphological parameters of the acetabulum through
The included studies, published between 2004 and 2016, fell virtual osteotomy. One study employed patient-specific CT data of
into three groups. The first group42e46 compared mechanical dysplastic hips pre- and post-osteotomy16.
behavior of dysplastic vs normal hips (Tables IeIII). The second CT-based models demonstrate bony anatomy well, but gener-
group compared the behavior of dysplastic hips before and after ating accurate geometries of soft tissues (cartilage and ligaments) is
physical16 or virtual15,17,46e48 acetabular osteotomy (Tables IeIII). difficult since these structures are poorly seen on CT images. Some
The third group8,11 investigated DDH etiology by mechanobio- studies42,43,45,48 used CT arthrography (CTA), which is invasive but
logical prediction of patterns of prenatal hip growth (Table IV). clearly demonstrates articular cartilage (AC) boundaries, to directly
When an article also included FEM of joints with other pathology model cartilage thickness (0.5e2.8 mm45, 0.8e2.0 mm42,43). Non-

Please cite this article in press as: Vafaeian B, et al., Finite element analysis of mechanical behavior of human dysplastic hip joints: a systematic
review, Osteoarthritis and Cartilage (2016), http://dx.doi.org/10.1016/j.joca.2016.10.023
B. Vafaeian et al. / Osteoarthritis and Cartilage xxx (2016) 1e10 3

Table I
Subjects and FEM specifications of adult hip joints

Study Simulated hips FEM

Normal Dysplastic Other Geometry Parts Bone material Soft tissue material

PB PF AC Lb CL Cortical (ILE) Cancellous (ILE) AC Lb

Russell (2006)45 1 11 N Patient-specific (CTA) Y Y Y N N E ¼ 2 GPa, n ¼ 0.3 E ¼ 120 MPa, ILE (E ¼ 12 MPa, NA
(Subchondral bone) n ¼ 0.3 n ¼ 0.42)
Zhao (2010) 17
1 3* 3 VPPAO Patient-specific (CT) Y Y Y Y N E ¼ 17 GPa, n ¼ 0.3 E ¼ 70 MPa, ILE (E ¼ 15 MPa, ILE (E ¼ 15 MPa,
n ¼ 0.2 n ¼ 0.45) n ¼ 0.45)
Henak (2011)43 1 1 N Patient-specific (CTA) Y Y Y Y N E ¼ 17 GPa, n ¼ 0.29 NA Neo-Hookean IHE Transversely IHEy
(K ¼ 1359 MPa,
G ¼ 13.6 MPa)
Zou (2013)15 N 5 24 VPPAO Patient-specific (CT) Y Y Y N Y E ¼ 17 GPa, n ¼ 0.3E ¼ 70 MPa, ILE (E ¼ 15 MPa, NA
n ¼ 0.2 n ¼ 0.45)
Henak (2014) 42
10 10 N Patient-specific (CTA) Y Y Y Y N E ¼ 17 GPa, n ¼ 0.29 NA Neo-Hookean IHE Transversely IHEy
(K ¼ 1359 MPa,
G ¼ 13.6 MPa)
Ike (2015)16 N 13 13 PRAO Patient-specific (CT) Y Y Y N N E: CT density-based ILE (E ¼ 10.35 MPa, NA
values, n ¼ 0.4 n ¼ 0.4)
Liu (2015) 47
N 4 4 VPPAO Patient-specific (CT) Y Y Y N N E ¼ 17 GPa, n ¼ 0.3 NA ILE (E ¼ 15 MPa, NA
n ¼ 0.45)
Liu (2016) 48
N 10z 40z VPPAO Patient-specific (CTA) Y Y Y N N E ¼ 17 GPa, n ¼ 0.3 NA ILE (E ¼ 15 MPa, NA
n ¼ 0.45)
Chegini (2009)44 4 4 NA CAD-based Yx Y Y Y N Rigid ILE (E ¼ 12 MPa, ILE (E ¼ 20 MPa,
n ¼ 0.45) n ¼ 0.4)
Liechti (2015)46 1 1 NA CAD-based Yx Y Y Y N Rigid ILE (E ¼ 12 MPa, ILE (E ¼ 20 MPa,
n ¼ 0.45) n ¼ 0.4)
Y: Yes, N: No, NA: not applicable, VPPAO: virtual post-periacetabular osteotomy, PRAO: post-rotational acetabular osteotomy, K: bulk modulus, G: shear modulus.
*
Dysplastic hip models were generated by deforming the acetabular rim of the normal hip model.
y
See ref. 43 for details.
z
The models (10 dysplastic and 40 VPPAO) were replicated but with constant cartilage thicknesses instead of the patient-specific cartilage thicknesses.
x
Includes only the acetabular cup.

Table II
Analysis conditions and finite element analysis specifications of adult hip joints

Study Analysis conditions Finite element analysis


(subjects' body weight)
Loading method Fixed boundary conditions Joint contact Direct output Software

Russell (2006)45 Gait JRF at femoral head centroid Medial pelvic wall Y CP and areay,z ABAQUS
(51e90 kgf)
Zhao (2010)17 One leg stance Vertical load on superior aspect Distal end of PF Y JRF MSC.Marc
(74 kgf) of sacrum Von-Mises stressy,z,x
Abductor force
Henak (2011)43 Gait (heel strike and mid- JRF by displacement control Pubis and the Sacro-iliac joints Y CP and areay,x NIKE3D
stance) technique Distal end of PF* Lb load support
Stair descent and ascent (heel Deflection of the Lb
strike)
(Female: 66 kgf; Male: 87 kgf)
Zou (2013)15 One leg stance JRF on pelvic bone Distal end of PF Y CP and areay ABAQUS
(70 kgf) Von-Mises stressy
42
Henak (2014) Gait (heel strike and mid- JRF by displacement control Pubis and the Sacro-iliac joints Y CP and areay,x NIKE3D
stance) technique Distal end of PF* Lb load support
Stair descent and ascent (heel
strike)
(70 ± 13.9 kgf)
Ike (2015)16 Nonphysiological load Vertical load on superior aspect Distal end of PF N Von-Mises stressy ANSYS
(Load: 183.5 kgf) of sacrum
47
Liu (2015) One leg stance JRF at femoral head centroid Pubis and top surface of pelvis Y CP and areay ABAQUS
(66.3 kgf) Distal end of PF*
Liu (2016)48 One leg stance JRF at femoral head centroid Pubis and top surface of pelvis Y CP and areay ABAQUS
(66.3 kgf) Distal end of PF*
Chegini (2009)44 Gait JRF at femoral head centroid Acetabulum Y CPy,x ABAQUS
Stance to sit Von-Mises stressy
(85 kgf)
Liechti (2015)46 Gait JRF at femoral head centroid Acetabulum Y CPy ABAQUS
Stance to sit Von-Mises stressy
(71.5 kgf)
*
Fixed or limited displacement only in lateral-medial and anterior-posterior directions.
y
Acetabular cartilage.
z
Femoral cartilage.
x
Lb.

Please cite this article in press as: Vafaeian B, et al., Finite element analysis of mechanical behavior of human dysplastic hip joints: a systematic
review, Osteoarthritis and Cartilage (2016), http://dx.doi.org/10.1016/j.joca.2016.10.023
4 B. Vafaeian et al. / Osteoarthritis and Cartilage xxx (2016) 1e10

Table III
Simulation results of adult hip joints

Study Significant model outputs Significant findings

Circumstances Output Normal Dysplastic Ratio*

Russell (2006)45 Gait Peak CP (MPa) 1.8 3.6e9.9 2.0e5.7 Bone irregularities in dysplastic hip models led to localized elevated
Contact area (mm2) 2265 215 0.1 cartilage CP. In dysplastic hip models, peak CP was not coincident
with maximum JRF during gait cycles as a consequence of
incongruous surfaces of dysplastic hips.
Zhao (2010)17 One-leg stance Peak svon (MPa) 15.2 22.7e30.7 1.5e2.0 Decrease in CE and vertical centre anterior (VCA) angles (from 25
to 10 ) of the simulated hips led to increase in the cartilage stress
and altered the uniform stress distribution of the normal joint to
stress concentration in the acetabular edge in the dysplastic joint
models..
Henak (2011)43 Gait Peak CP (MPa) 6e14 6e14 NR Due to shallow acetabulum in the dysplastic hip model, the Lb
Stair descent Average CP (MPa) 1.1e1.3 0.5e1.2 0.4e1.2 significantly contributed to the joint stability (lateral stability in
Stair ascent Contact area (mm2) 680e800 350e440 0.4e0.7 particular) and load transfer in the dysplastic hip model in
Lb load support 1e2% 4e11% 2e11 comparison with the normal hip model. Consequently, cartilage CP
did not substantially increase in the dysplastic hip model.
Zou (2013)15 One-leg stance Average CP (MPa) 3.59e6.07y 3.9e6.2 1.0e1.1 Average cartilage stresses (pressure and von-Mises) and contact
Contact Area (mm2) 319e531y 304e492 0.9e1.0 area were not linearly and monotonically correlated with CE angle
Average svon (MPa) 1.4e2.2y 1.5e2.4 1.0e1.1 (10 e45 ). However, patient-specific optimal CE angles minimizing
the cartilage stresses and maximizing the contact areas were found
for the dysplastic hip models.
Henak (2014)42 Gait Peak CP (MPa) 5.0e15.0 2.5e16.2 0.5e1.1 The load supported by the Lb (especially on the lateral side) was
Stair descent Average CP (MPa) 0.5e1.9 0.1e2.1 0.2e1.3 significantly higher in the dysplastic hip models than in normal hip
Stair ascent Lb load support 2.6e3.3% 8.2e10.3% 2.8e4.0 models. CPs in the dysplastic hip models were not elevated
compared with the normal hip models because the Lb provided
additional contact surface in the dysplastic hip models.
Ike (2015)16 Non-physiological Peak svon (MPa) 1.7e4.0y 1.2e6.5 0.7e1.8 Moderate correlations between von Mises stress in the joint and
morphological parameters, CE angle (R ¼ 0.65), acetabular head
index (R ¼ 0.60), acetabular angle (R ¼ 0.48), and acetabular roof
angle (R ¼ 0.57), were observed.
Liu (2015)47 One-leg stance Peak CP (MPa) 3.8e6.3y 5.5e18.9 1.5e3.6 Acetabular inclination and version, femoral head coverage ratio and
Contact Area (mm2) 834e1497y 503e1005 0.5e0.7 CE angle were morphological parameters engaged in the status of
CP and area of normal and dysplastic hips.
Liu (2016)48 One-leg stance Peak CP (MPa) 4.5e10y 5.5e15 1.1e2.9 Peak CP and contact area predicted by models using constant
Contact Area (mm2) 480e1198y 425e1166 0.6e1.0 thickness cartilage respectively had moderate (R ¼ 0.63) and strong
(R ¼ 0.87) correlations with their corresponding values resulted
from models employing varying-thickness cartilage obtained from
patient-specific data. An optimal range of CE angle minimizing the
peak CP and maximizing the contact area was found to be 23.5
e38.9 .
Chegini (2009)44 Gait Peak CP (MPa) 2.4e3.4 6.1e9.9 1.8e4.1 Activities with high load (gait) rather than with large range of
Stance to sit Peak CP (MPa) 3.3e3.7 3.5e3.6 0.9e1.1 motion (stance to sit) led to higher pressure and von Mises stress in
the dysplastic hip models. An optimal range of CE angle to minimize
the CP and von Mises stress was found to be 20 e30 .
Liechti (2015)46 Gait Peak CP (MPa) 1.4 2.0 1.4 During gait, the area of peak CP in the normal hip model was
Peak svon (MPa) 0.7 1.4 2.0 centered in the acetabular roof whereas this area shifted towards
Stance to sit Peak CP (MPa) 1.3 1.5 1.2 the lateral face of the acetabular cartilage in the dysplastic hip
Peak svon (MPa) 0.8 1.0 1.3 model. During stance to sit, the area of peak CP in both the normal
and the dysplastic hip models was located on the posterior aspect of
the lunate surface.

svon: Von-Mises stress in cartilage or the underneath cortical shell, NR: not reported, VCA: vertical center-anterior margin.
*
Ratio of the simulated values corresponding with dysplastic hips to those of normal hips.
y
The models of VPPAO and PRAO leading to minimized CP and maximized contact area are listed as normal hips.

arthrographic studies15e17,47,48 assumed constant AC thickness greater (by 2e11 times) in dysplastic than normal hips. Not
(1.8e2 mm), simplifying FEM. However, Anderson et al.49 demon- including the Lb in FEM of dysplastic hips therefore notably affects
strated that use of constant-thickness cartilage in normal hip FEM its simulated mechanical behavior42e44.
overestimated cartilage CP and underestimated contact areas. In Defining capsular ligaments (CLs) on CT images is as difficult as
dysplastic hips, Liu et al.48 found 21 ± 19% and 17 ± 14% error in defining the Lb. Only one study (15) incorporated these ligaments
simulated CP and area respectively when replacing patient-specific in FEM, however, in simple fashion with ligaments modeled as
cartilage thickness with a constant-thickness cartilage layer. discrete spring elements, with no report of the functionality and
Therefore, utilizing patient-specific cartilage thickness likely en- effects of modeling these ligaments. The stabilizing contribution of
hances simulation accuracy49; especially in dysplastic hips in which the hip capsule depends on the joint angular position and con-
cartilage thickness is greater and more variable than in normal gruency51e53, and is greater in dysplastic hips51. The circumstances
hips50. in which including the capsule may affect FEM results were not
The cartilage-labrum (Lb) boundary is usually not visible even in discussed in the reviewed studies, and require further investigation
arthrograms. Studies either determine this based on morphology and experimental validation.
and reader judgment17,42,43, or simply disregard the Lb15,16,45,47,48. To investigate the effect of wide variations in hip geometry on
Several studies that included the Lb found its mechanical contri- mechanical behavior, some studies used CAD-based rather than
bution in hip stabilization and load transfer from cartilage was patient-specific models44,46. Averaged anatomic measurements led

Please cite this article in press as: Vafaeian B, et al., Finite element analysis of mechanical behavior of human dysplastic hip joints: a systematic
review, Osteoarthritis and Cartilage (2016), http://dx.doi.org/10.1016/j.joca.2016.10.023
B. Vafaeian et al. / Osteoarthritis and Cartilage xxx (2016) 1e10 5

Table IV
FEM methodology and simulation results of prenatal hip joint growth

Study FEM specifications Significant findings

Model components Material (ILE) Boundary conditions Contact Stress output


(CAD-based) and load

Shefelbine PF (3D)* Bonez: Boundary conditions: N Hydrostatic Normal hip JRF loading history led to a convex shape of the
(2004)8 E ¼ 500 MPa, fixed distal end of PF pressure growth front, whereas dysplastic hip loading history resulted in
n ¼ 0.2 Load: varying angle JRF Octahedral formation of coxa valga.
Cartilagex: with arbitrary shear
G ¼ 2 MPa, magnitude
n ¼ 0.49
Giorgi PF Acetabular Cartilage: Boundary conditions: Y Hydrostatic Symmetric prenatal hip rotation within normal ranges
(2015)11 cup (2D)y E ¼ 1.1 MPa, Fixed acetabulum, pressure promoted normal formation of acetabulum, whereas reduced-
n ¼ 0.49 arbitrary distal- range rotations led to prediction of shallow acetabulum.
proximal displacement Asymmetric rotations simulated formation of asymmetrically
of distal end of the PF, shallow acetabulum and malformed femoral head associated
planar rotation of the with DDH.
femoral head
*
A 7-month fetal PF with a non-ossified head located above a thin cartilaginous growth region supported by a bony diaphysis.
y
Idealized 2D geometries of a non-ossified PF, and an acetabular cup associated with a fetus at 11th gestational week.
z
E ¼ 500 MPa distally, linearly decreasing toward the cartilaginous femoral head.
x
Young's modulus can be calculated as 5.96 MPa.

to idealized ball-and-socket geometries of adult hips: hemispher- In a CAD-based prenatal femur model8, the bony part was
ical femoral head and acetabulum. Articular surfaces were modeled modeled as an elastic deformable material since simulation of
as portions of smooth spherical surfaces, including horseshoe- mechanical stress in the growth front within the diaphysis was
shaped acetabular cartilage. Reducing the lateral center-edge (CE) required (Table IV).
angle introduced dysplasia into these models. Although conve- Unlike bone, cartilage exhibits time- and rate-dependent
nient, this likely does not lead to reliable quantitative prediction of behavior under physiological loads. Nevertheless, these complex
hip mechanical behavior. In comparison with patient-specific effects can be disregarded if external load on the cartilage is
models of normal hips, CPs were underestimated by 25e50%49 instantaneous19,56e58, such as with loading time less than 150 ms56
from CAD-based FEM, due at least partly to contact area over- or loading cycles less than 1 s (gait, stair descent/ascent, stance-to-
estimation44,46,49. This error is likely greater in dysplastic hips, sit, onset of one-leg stance)57,58. Since cartilage is reportedly
where shape irregularities cause incongruent articulation and incompressible under instantaneous load59, this simplified FEM of
locally increased CP45, not accounted for by CAD-based models adult hips allowing use of either nearly incompressible ILE15e17,44e48
using simplified/smoothed geometry44,46. These stress concentra- or incompressible isotropic hyperelastic (IHE) models42,43 for
tions at the lateral acetabular roof may be important in OA devel- cartilage (Table I), with parameters determined based on fast-
opment. Use of patient-specific geometry is therefore preferred, as loading experimental conditions. Cartilage in the prenatal hip
it limits the risk of bias from anatomic assumptions. models8,11 was modeled as a nearly incompressible ILE material
The studies modeling prenatal hips (Table IV) utilized CAD- (Table IV). Disregarding time and rate properties of the cartilage may
based modeling of either a non-ossified fetal PF8, or an idealized not be associated with physiological loading conditions inside the
hip joint including non-ossified femoral and acetabular uterus, however, it satisfies the minimum FEM requirements for
components11. cross-comparison purposes between the simulation results.
The acetabular Lb consists primarily of type 1 collagen fiber
Mechanical properties of materials bundles extending around the acetabular rim60, and is therefore
highly anisotropic, having different mechanical response to loads
Deformations and stresses inside the hip depend on material parallel vs perpendicular to the fibers. Assuming any direction
properties of its components including bone, cartilage and liga- except circumferential has similar properties, the Lb can still be
ments. Adult patient-specific FEM15e17,42,43,45,47,48 all assumed considered transversely isotropic. Moreover, time- and rate-
homogeneous isotropic linear-elastic (ILE) materials for cortical dependent properties can be neglected under fast loading rates.
and cancellous bone. This simplifying assumption has been shown Thus, the Lb has been modeled as incompressible transversely IHE
to be appropriate for analyzing hip mechanical behavior54,55, and material42,43, or as a nearly incompressible isotropic material with
requires only two mechanical constants, Young's modulus E and linear elastic properties measured along the fibers17,44,46. Isotropic
Poisson's ratio n. For cortical bone most studies used E ¼ 17 GPa, labral material assumption reportedly overestimates the load
while those that modeled cancellous bone separately used supported by the Lb by 2e11%43, since an isotropic Lb is unrealis-
E ¼ 70e120 MPa (Table I). One study16 used location-dependent tically stiff perpendicular to the fibers compared with the circum-
values of E based on CT data. Because cortical bone is much ferential direction.
stiffer than cancellous bone and forms a sandwich shell over it55, Another simplification of material properties in adult hips was
some consider cancellous bone to have only a minor effect on employing the soft tissue material properties of the normal hips for
CP27,49. Therefore, some FEM studies disregarded cancellous dysplastic hips. As noted by Henak et al.42, cartilage may be stiffer
bone42,43,47,48. Some CAD-based models use rigid material for and softer, respectively, in dysplastic than normal hips.
bone44,46, further simplifying FEM. One of these groups showed
this assumption to have no effect on predicted cartilage stresses, Contact mechanism
since bone is much stiffer than AC44. However, the assumption
that bone is rigid may not be appropriate in patient-specific In a hip joint, the convex femoral articular surface slides against
models, where use of rigid bones led to a significant increase in the concave acetabulum. Articular friction is demonstrated to be
predicted CP49. negligible (frictional coefficient of 0.002e0.02) when synovial fluid

Please cite this article in press as: Vafaeian B, et al., Finite element analysis of mechanical behavior of human dysplastic hip joints: a systematic
review, Osteoarthritis and Cartilage (2016), http://dx.doi.org/10.1016/j.joca.2016.10.023
6 B. Vafaeian et al. / Osteoarthritis and Cartilage xxx (2016) 1e10

becomes pressurized at the onset of loading and remains pressur- loading conditions to the upper sacrum preserved patient-specific
ized for a prolonged duration61e63. Accordingly, the reviewed effects of the joint geometry on the predicted JRF.
studies accounting for the sliding mechanism of the joint assumed The prenatal studies employed both direct and displacement
a frictionless contact mechanism11,15,17,42e48. This allows the small control loading (Table IV). Applying fixed displacement boundary
tangential (shear) component of the JRF to be neglected. One study, conditions at the distal femur, Shefelbine et al.8 used a dynamically
however, modeled AC as one entity without considering the sliding varying-angle JRF vector distributed symmetrically on the carti-
mechanism16. Ignoring the joint sliding contact mechanism over- laginous femoral head. The force angles assumed JRF angles in
estimates cartilage stresses and does not permit evaluation of joint normal (25 , 40 , 50 ) and dysplastic (60 , 85 , 105 ) joints. The
CP. The studies that included the Lb also assumed frictionless two-dimensional (2D) model by Giorgi et al.11 had an immobilized
cartilage-Lb contact17,42e44,46. acetabular cup and a freely rotating femur. The femur was proxi-
mally displaced towards the acetabulum for contact force initiation,
Boundary conditions and loading and dynamically rotated in-plane according to reported ranges of
motion from early to late gestation (±40 to ±5 respectively).
To analyze hip behavior, each bone should be in a state of force Symmetric and asymmetric rotations were tested. Contact forces in
equilibrium. Commonly the adult hip is analyzed assuming static both studies had an arbitrary magnitude, adequate for cross-
loads and immobilized bones. Generally42e48, the pelvic bone (PB) comparison but not necessarily physiologic.
was immobilized by applying fixed-displacement boundary con-
ditions (i.e., no motion), and a load exerted on the PF already ori-
ented within the hip according to the analysis conditions (Table II). Finite element analysis, outputs and results
Boundary conditions may also be applied to the femur to fix or limit
its medial-lateral or antero-posterior displacements42,43,47,48. Hip FEM can be categorized as a contact problem since loads are
Typically, the femur is initially mobile in distal-proximal direction, transferred by the contact mechanism of the joint. In the reviewed
and becomes immobilized and statically stable in the acetabular studies, FEM offered a choice of model outputs for adult hips.
socket once femoral-acetabular articular contact is completely Cartilage CP (distributed form of the JRF on the surfaces in contact)
established. Other studies15e17 applied fixed-displacement and von Mises stress (indicator of internal distortion energy inside
boundary conditions for the distal femur and applied load to the the tissues) are parameters that reflect the status of the load
PB to analyze natural anatomic posture, e.g., one-leg stance. transfer mechanism and mechanical stresses inside the joint
The reviewed studies investigated adult hip mechanical respectively. High cartilage CP may lead to OA when accumulated
behavior under different physiological (gait, stair descent/ascent, over years6,14 and elevated von Mises stress may be associated with
stance-to-sit, one-leg stance) and non-physiological loading sce- cartilage damage44.
narios (Table II). Most studies consider the JRF vector obtained The reviewed studies on adult hip joints show widely varying
experimentally23,64 to be the loading input for the FEM. JRF is the ranges of CP and area, and von Mises stress in normal and
balancing force transmitted at articular surfaces due to ligament dysplastic hips under a variety of loading scenarios (Table III). Each
and muscle forces, and body weight. JRF can be used to avoid dif- study performed mesh sensitivity analysis to minimize FEM error
ficulties in locating, distributing and orienting individual muscle and produce accurate results. Most of the studies (except Refs. 16
forces in FEM, as long as analysis results near the joint (i.e., ace- and 48) validated their FEM by showing that simulated CP and
tabulum and femoral head including soft tissues) are the focus of contact area were comparable with or in range of available
FEM. JRF can be applied either under load control or displacement measured experimental data. The numeric ranges of simulated
control. Under load control, JRF was applied at the femoral head values are not directly comparable with each other because the
centroid44e48 or at the upper pelvis15. Applying JRF at the femoral models are associated with different joint geometry, material
head centroid requires kinematic coupling of femoral head finite- properties, and loading vectors, methods and scenarios. Fortu-
element nodes to the centroid, i.e., assumption of a rigid femoral nately, the diversity of model design reduces the risk of bias in
head, which induces a potentially undesirable approximation in the results across studies. Qualitative comparisons between the FEM
results49. Applying JRF to the upper hemi-pelvis involves kinematic results of dysplastic vs normal hips were performed.
coupling of upper pelvis finite-element nodes to a single point such Simulating gait cycles, some researchers44e46 found peak CP to
that the line of action of the force, applied at the reference point, be 1.4e5.7 times higher in dysplastic hips than normal joints. In
passes through the femoral head centroid, allowing the femoral contrast, Henak et al.42,43 found peak CP were actually larger in
head model to remain deformable. Two reviewed studies employed normal hip models than dysplastic hips, because dysplastic hip
displacement-control loading42,43, in which the PF was pushed geometry shifted part of the JRF towards the Lb, engaging part of
against the acetabulum along the line of action of JRF until the the labral surface to contact femoral head cartilage. The extra labral
desired force was achieved. Since the displacement vector (as contact area helped redistribute JRF, sparing cartilage by increasing
boundary condition) was applied to the femoral shaft, the load on the Lb. According to Henak et al.42, other studies predict
deformable property of the femoral head was preserved. A meth- elevated CP during gait in dysplastic hips because they disregard
odological limitation of employing JRF in the hip models is that the either the Lb45 or patient-specific joint geometry44,46. The fact that
JRF vectors in the reviewed studies were based on in vivo mea- the superior Lb was most affected by shift of JRF in the dysplastic
surement data from instrumented implants within hips after total hip models may explain the high clinical prevalence of superior or
hip arthroplasty, scaled by patient weight. These estimates are not anterosuperior labral tears and adjacent cartilage damage. Cartilage
necessarily equivalent to the JRF one might actually observe in contact area during gait was reportedly smaller in dysplastic than
patient-specific models, especially in dysplastic hips. normal hips, whether or not the Lb was included in FEM43,45.
Rather than loading along the line of action of JRF, some studies Groups investigating mechanical behavior of dysplastic hips
instead applied a vertical load to the upper sacrum16,17, to assess under one-leg stance conditions15,17,47,48 simulated acetabular
one-leg stance and non-physiological loading65. For one-leg osteotomies to create close-to-normal hip models out of dysplastic
stance17, gluteus medius abductor muscle force was also applied hips. They reported that maximizing contact area in dysplastic hips
to the PB and the greater trochanter to maintain static force equi- via acetabular reorientation generally decreased CP and von Mises
librium. Unlike the direct utilization of JRF in FEM, applying the stress. However, minimized peak CP may not be always concurrent

Please cite this article in press as: Vafaeian B, et al., Finite element analysis of mechanical behavior of human dysplastic hip joints: a systematic
review, Osteoarthritis and Cartilage (2016), http://dx.doi.org/10.1016/j.joca.2016.10.023
B. Vafaeian et al. / Osteoarthritis and Cartilage xxx (2016) 1e10 7

with the maximized contact area48, due to stress concentrations at connecting the acetabulum and femoral head34,40. Ligaments are
patient-specific bone irregularities45. represented by springs. Neglecting mechanical stresses inside
Two studies investigating stance-to-sit conditions44,46 both re- bones, DEA estimates joint CP from forces in compressive springs
ported higher CP and von Mises stress in dysplastic hips, accentu- distributed on the joint surfaces. This does not accurately model the
ated by activities with high load (gait) rather than with large range joint sliding contact mechanism, but can still demonstrate patterns
of motion (stance-to-sit). of contact area and CP distribution regardless of absolute magni-
The reviewed prenatal studies recruited other FEM-simulated tudes. This makes DEA appropriate for studies comparing me-
stress components (Table IV). Shefelbine et al.8 evaluated the chanical behavior of normal vs dysplastic hips, and pre- vs post-
mechanobiological theory that growth is accelerated by intermit- operative joints34,40 (Table V).
tent octahedral shear stress and inhibited by intermittent hydro-
static pressure66. Giorgi et al.11 assessed the theory that growth is
accelerated by dynamic hydrostatic pressure and inhibited by static Recommendations for FEM of infant hips
pressure67. Growth in both studies was simulated through
computational expansion of finite elements with rates calculated We found no FEM-based studies in infants or children. Prenatal
based on the growth theories. These simulations provided insights FEM-based studies focus on DDH etiology. Pathological conse-
on the effect of the mechanical environment on cartilage growth quences of the dysplastic mechanical environment in infant hips,
within limited and idealized assumptions. and optimal methods of DDH correction in children, have not, to
Shefelbine et al.8 found that in normal hips, hydrostatic pressure our knowledge, been modeled yet. Adult studies are likely not
was higher peripherally and octahedral shear stress was highest directly applicable to infant hips, since unlike in adults most of the
centrally, promoting a convex shape of the growth front. In dysplastic infant/prenatal hip is composed of cartilage, including the femoral
hips, octahedral shear stress was increased on the medial side of the head and the pubic rami. Given the extent to which small changes
growth region, imbalancing the growth front toward coxa valga. in cartilage thickness affect results in patient-specific adult hip
Giorgi et al.11 showed that asymmetric rotations of the femoral models48, the proportionately thicker infant hip AC likely alters
head inside the acetabulum prenatally led to asymmetric joint infant hip mechanical behaviour compared to adults. It is also un-
loading which eventually opened the acetabulum toward the di- clear to what extent infant cartilage mechanical properties differ
rection of the generated JRF, giving the asymmetrically shallow from adult cartilage77. Patient-specific computational geometries of
acetabulum and malformed femoral head of DDH. Such asymmetric infant hips are challenging to produce via CT which poorly depicts
rotations of a prenatal hip joint might result from fetal breech cartilage, and CT is not practical for this indication due to radiation
position or increased joint laxity. dose concerns in infants. Magnetic resonance imaging (MRI) shows
clearer boundaries of soft tissue structures78,79, without ionizing
radiation, but infants may require sedation, and spatial resolution is
Limitations of, and alternatives to, FEM lower than on CT. 3D ultrasound is an emerging modality offering
high spatial resolution with no ionizing radiation. Since early re-
The reviewed studies show that FEM provides a versatile tool to sults show promising geometric accuracy80, this may be useful in
assess etiology and effects of DDH. Nevertheless, FEM is time- future patient-specific modeling.
intensive and computationally expensive, especially with detailed Loading scenarios in adult hip FEM such as walking, climbing
finite-element meshes, nonlinear material properties and non- stairs or one-leg standing are obviously not relevant in infants,
linear contact mechanism formulation20. This can paralyze the where more important loading scenarios include positioning
method if analysis time is a critical factor, such as in large popu- within a swaddle, sling, car seat or Pavlik harness. In-utero loading
lation studies37e39,68e70, osteotomy surgical planning71,72 or real- scenarios are also uniquely challenging. Verbruggen81 evaluated
time intraoperative feedback73e75. Trading strict accuracy for fetal movements, and reaction force of fetal kicking against the
faster modeling in these settings, DEA can be regarded as the uterine wall. Unlike in adults, infant load scenarios may involve
closest alternative to FEM76. DEA (or rigid body-spring modeling) prolonged slow loading, as with corrective harnesses worn 24 h/
considers bones as rigid bodies, and AC as one-dimensional springs day. According to Hjelmstedt et al.12,13, long-duration moderate

Table V
DEA-based studies regarding human dysplastic hips

Study Modeling focus Significant findings

Genda (1995)37 Evaluating CP in normal and acetabular Even distribution of CP over the joint surface and low peak CP (averaged 2.6 MPa) in normal joints
dysplastic hip joints (loading scenarios: (CE > 20 ). Concentration of CP on anterolateral edge of acetabulum and increasing peak CP (2.5
heel strike and toe-off phases of gait) e17 MPa) with reduction of CE angle in dysplastic joints (1.75 < CE < 19.8 ).
Tsumura (1998)41 Simulating 3D CP distribution in normal The peak CP in dysplastic hips (15 < CE < 7 ) was located at the edge of the acetabulum. The peak CP
and dysplastic hips (loading scenario: values in dysplastic hips (2.83e5.31 MPa) were higher in normal hips (1.83e2.51 MPa) with
one-leg stance phase of gait) 24 < CE < 38 .
Rab (2007)70 Investigating effects of different Progressive anterolateral subluxation associated with hip dysplasia occurred in the hip models with
acetabular rotations on hip subluxation CE < 20 . The subluxation was followed by dislocation in the models with CE < 0 . Models with CE  30
(loading scenario: one-leg stance) did not lead to lateral subluxation.
Armiger (2009)69 Calculating pre- and post-operative CP Calculated peak CP values in preoperative dysplastic hips (1.9e7.7 MPa) were reduced (1.4e3.2 MPa) by
in hip joints of PAO patients (loading an average factor of 1.7 after PAO. Reduction of the peak CP was not proportional with increase of CE
scenarios: standing, gait, and sitting) angle.
Niknafs (2013)72 Evaluating the effects of cartilage Choice of cartilage thickness distribution (patient-specific vs constant), or linear/non-linear spring
thickness distribution and compressive models of cartilage did not affect the predicted optimal alignment of acetabulum.
spring models of cartilage on optimal
alignment of acetabulum in PAO
patients (loading scenarios: standing,
gait, and sitting)

PAO: periacetabular osteotomy.

Please cite this article in press as: Vafaeian B, et al., Finite element analysis of mechanical behavior of human dysplastic hip joints: a systematic
review, Osteoarthritis and Cartilage (2016), http://dx.doi.org/10.1016/j.joca.2016.10.023
8 B. Vafaeian et al. / Osteoarthritis and Cartilage xxx (2016) 1e10

loads can provoke DDH due to time-dependent deformations of References


cartilaginous femoral head and acetabulum. The rate- and time-
dependency of cartilage material properties may play a role in 1. Schwend RM, Shaw BA, Segal LS. Evaluation and treatment of
FEM of infant hips. developmental hip dysplasia in the newborn and infant.
Pediatr Clin North Am 2014;61:1095e107.
Summary 2. Furnes O, Lie SA, Espehaug B, Vollset SE, Engesaeter LB,
Havelin LI. Hip disease and the prognosis of total hip re-
FEM of adult dysplastic hips is most commonly performed using placements. J Bone Joint Surg 2001;83:579e86.
patient-specific CT-based geometry, ILE models of bone, cartilage 3. Maquet P. Biomechanics of hip dysplasia. Acta Orthop Belg
and Lb, omission of most muscles and ligaments, and assumptions 1999;65:302e14.
of rapid loading to simplify cartilage mechanical properties. Load 4. Gulati V, Eseonu K, Sayani J, Ismail N, Uzoigwe C,
conditions modeled simulate situations such as walking or stair- Choudhury MZ, et al. Developmental dysplasia of the hip in the
climbing. Joint CPs and stresses are simulated. Most models newborn: a systematic review. World J Orthop 2013;4:32e41.
demonstrate increased stress on cartilage and/or increased reliance 5. Harris-Hayes M, Royer NK. Relationship of acetabular
on the Lb to bear load in dysplastic hips, correlating well to clinical dysplasia and femoroacetabular impingement to hip osteoar-
findings of premature OA and early labral tearing in dysplastic hips. thritis: a focused review. PM R 2011;3. 1055.e1e1067.e1.
DEA offers a rapid alternative to FEM when computational time is 6. Hadley NA, Brown TD, Weinstein SL. The effects of contact
an issue. FEM of prenatal hips demonstrates ways in which in-utero pressure elevations and aseptic necrosis on the long-term
loading might produce DDH. FEM of infant dysplastic hips offers an outcome of congenital hip dislocation. J Orthop Res 1990;8:
opportunity to understand the stress environment and factors 504e13.
affecting it at a time when the dysplasia can still be corrected 7. Brand RA. Joint contact stress: a reasonable surrogate for
conservatively. Models in infants do not yet exist, and have special biological processes? Iowa Orthop J 2005;25:82e94.
challenges, including the need to obtain accurate cartilage geom- 8. Shefelbine SJ, Carter DR. Mechanobiological predictions of
etry, consideration of rate-dependent cartilage material properties, growth front morphology in developmental hip dysplasia.
and infant-relevant loading scenarios. J Orthop Res 2004;22:346e52.
Hip mechanical modeling through the human life cycle may 9. Siffert RS. Patterns of deformity of the developing hip. Clin
improve our understanding of the causes of DDH and of premature Orthop Relat Res 1981:14e29.
hip OA, and ultimately help us optimize DDH treatment to reduce 10. LeVeau BF, Bernhardt DB. Developmental biomechanics. Effect
its lifelong disability. of forces on the growth, development, and maintenance of the
human body. Phys Ther 1984;64:1874e82.
Author contributions 11. Giorgi M, Carriero A, Shefelbine SJ, Nowlan NC. Effects of
Author list: normal and abnormal loading conditions on morphogenesis of
BV: B. Vafaeian the prenatal hip joint: application to hip dysplasia. J Biomech
DZ: D. Zonoobi 2015;48:3390e7.
MM: M. Mabee 12. Hjelmstedt A, Asplund S. Congenital dislocation of the hip a
ARH: A. R. Hareendranathan biomechanical study in autopsy specimens. J Pediatr Orthop
ME: M. El-Rich 1983;3:491e7.
SA: S. Adeeb 13. Hjelmstedt A, Asplund S, Rauschning W. Cryodissection and
JJ: J. L. Jaremko cryosectioning in biomechanical studies on congenital dislo-
Contributions of authors (All authors should have made cation of the hip. Anat Clin 1982;4:13e21.
substantial contributions to all three of sections (1), (2) and (3)) 14. Michaeli DA, Murphy SB, Hipp JA. Comparison of predicted and
measured contact pressures in normal and dysplastic hips.
 (1) Conception and design: BV, DZ, MM, ARH, ME, SA, JJ Med Eng Phys 1997;19:180e6.
 (1) Analysis and interpretation of the data: BV, DZ, MM, ARH 15. Zou Z, Cha vez-Arreola A, Mandal P, Board TN, Alonso-
 (2) Drafting of the article: BV, JJ Rasgado T. Optimization of the position of the acetabulum in a
 (2) Critical revision of the article for important intellectual ganz periacetabular osteotomy by finite element analysis.
content: BV, DZ, MM, ARH, ME, SA, JJ J Orthop Res 2013;31:472e9.
 (3) Final approval of the article: BV, DZ, MM, ARH, ME, SA, JJ 16. Ike H, Inaba Y, Kobayashi N, Yukizawa Y, Hirat Y, Tomioka M,
 (.) Provision of study materials or patients: ME, SA, JJ et al. Effects of rotational acetabular osteotomy on the me-
 (.) Statistical expertise: JJ chanical stress within the hip joint in patients with develop-
 (.) Obtaining of funding: JJ mental dysplasia of the hip: a subject-specific finite element
 (.) Administrative, technical, or logistic support: DZ, ME, SA, JJ analysis. Bone Joint J 2015;97-B:492e7.
 (1) Collection and assembly of data: BV, JJ 17. Zhao X, Chosa E, Totoribe K, Deng G. Effect of periacetabular
 Guarantor of overall integrity of the work: JJ osteotomy for acetabular dysplasia clarified by three-
dimensional finite element analysis. J Orthop Sci 2010;15:
632e40.
Conflict of interest 18. Schmitt J, Meiforth J, Lengsfeld M. Development of a hybrid
None of the co-authors has any competing interests to disclose that finite element model for individual simulation of inter-
could potentially and inappropriately influence the study. trochanteric osteomies. Med Eng Phys 2001;23:529e39.
19. Stops A, Wilcox R, Jin Z. Computational modelling of the
Acknowledgements natural hip: a review of finite element and multibody simu-
lations. Comput Methods Biomech Biomed Engin 2011;15:
Unrestricted funding support for this work was provided by the 963e79.
Capital Health Endowed Chair in Diagnostic Imaging and the CIHR- 20. Bachtar F, Chen X, Hisada T. Finite element contact analysis of
HSC Young Investigator Award (NI15-004). the hip joint. Med Biol Eng Comput 2006;44:643e51.

Please cite this article in press as: Vafaeian B, et al., Finite element analysis of mechanical behavior of human dysplastic hip joints: a systematic
review, Osteoarthritis and Cartilage (2016), http://dx.doi.org/10.1016/j.joca.2016.10.023
B. Vafaeian et al. / Osteoarthritis and Cartilage xxx (2016) 1e10 9

21. Harris MD, Anderson AE, Henak CR, Ellis BJ, Peters CL, Weiss JA. 41. Tsumura H, Miura H, Iwamoto Y. Three-dimensional pressure
Finite element prediction of cartilage contact stresses in distribution of the human hip joint e comparison between
normal human hips. J Orthop Res 2012;30:1133e9. normal hips and dysplastic hips. Fukuoka Igaku Zasshi
22. Brown TD, Shaw DT. In vitro contact stress distributions in the 1998;89:109e18.
natural human hip. J Biomech 1983;16:373e84. 42. Henak CR, Abraham CL, Anderson AE, Maas SA, Ellis BJ,
23. Bergmann G, Deuretzabacher G, Heller M, Graichen F, Peters CL, et al. Patient-specific analysis of cartilage and
Rohlmann A, Strauss J, et al. Hip forces and gait patterns from labrum mechanics in human hips with acetabular dysplasia.
routine activities. J Biomech 2001;34:859e71. Osteoarthritis Cartilage 2014;22:210e7.
24. Afoke NY, Byers PD, Hutton WC. Contact pressures in the hu- 43. Henak CR, Ellis BJ, Harris MD, Anderson AE, Peters CL, Weiss JA.
man hip joint. J Bone Joint Surg Br 1987;69:536e41. Role of the acetabular labrum in load support across the hip
25. Krebs DE, Elbaum L, Riley PO, Hodge WA, Mann RW. Exercise joint. J Biomech 2011;44:2201e6.
and gait effects on in vivo hip contact pressures. Phys Ther 44. Chegini S, Beck M, Ferguson SJ. The effects of impingement and
1991;71:301e9. dysplasia on stress distributions in the hip joint during sitting
26. Von Eisenhart R, Adam C, Steinlechner M, Muller-Gerbl M, and walking: a finite element analysis. J Orthop Res 2009;27:
Eckstein F. Quantitative determination of joint incongruity and 195e201.
pressure distribution during simulated gait and cartilage 45. Russell ME, Shivanna KH, Grosland NM, Pedersen DR. Cartilage
thickness in the human hip joint. J Orthop Res 1999;17:532e9. contact pressure elevations in dysplastic hips: a chronic
27. Anderson AE, Ellis BJ, Maas SA, Peters CL, Weiss JA. Validation overload model. J Orthop Surg Res 2006;1:6.
of finite element predictions of cartilage contact pressure in 46. Liechti EF, Ferguson SJ, Tannast M. Protrusio acetabuli: joint
the human hip joint. J Biomech Eng 2008;130:51008. loading with severe pincer impingement and its theoretical
28. Ipavec M, Brand RA, Pedersen DR, Mav ci
c B, Kralj-Igli
c V, implications for surgical therapy. J Orthop Res 2015;33:106e13.
Igli
c A. Mathematical modelling of stress in the hip during gait. 47. Liu L, Ecker T, Xie L, Schumann S, Siebenrock K, Zheng G.
J Biomech 1999;32:1229e35. Biomechanical validation of computer assisted planning of
29. Pompe B, Daniel M, Sochor M, Vengust R, Kralj-Iglic V, Igli c A. periacetabular osteotomy: a preliminary study based on finite
Gradient of contact stress in normal and dysplastic human element analysis. Med Eng Phys 2015;37:1169e73.
hips. Med Eng Phys 2003;25:379e85. 48. Liu L, Ecker TM, Schumann S, Siebenrock K. Evaluation of
30. Daniel M, Antolic V, Iglic A, Kralj-Iglic V. Determination of constant thickness cartilage models vs. patient specific carti-
contact hip stress from nomograms based on mathematical lage models for an optimized computer-assisted planning of
model. Med Eng Phys 2001;23:347e57. periacetabular osteotomy. PLoS One 2016;11:1e12.
31. Brinckmann P, Frobin W, Hierholzer E. Stress on the articular 49. Anderson AE, Ellis BJ, Maas SA, Weiss JA. Effects of idealized
surface of the hip joint in healthy adults and persons with joint geometry on finite element predictions of cartilage con-
idiopathic osteoarthrosis of the hip joint. J Biomech 1981;14: tact stresses in the hip. J Biomech 2010;43:1351e7.
149e56. 50. Nishii T, Sugano N, Sato Y, Tanaka H, Miki H, Yoshikawa H.
32. Maxian TA, Brown TD, Weinstein SL. Chronic stress tolerance Three-dimensional distribution of acetabular cartilage thick-
levels for human articular cartilage: two nonuniform contact ness in patients with hip dysplasia: a fully automated
models applied to long-term follow-up of CDH. J Biomech computational analysis of MR imaging. Osteoarthritis Cartilage
1995;28:159e66. 2004;12:650e7.
33. Hipp JA, Sugano N, Millis MB, Murphy SB. Planning acetabular 51. Nepple JJ, Smith MV. Biomechanics of the hip capsule and
redirection osteotomies based on joint contact pressures. Clin capsule management strategies in hip arthroscopy. Sports Med
Orthop Relat Res 1999:134e43. Arthrosc 2015;23:164e8.
34. Abraham CL, Maas SA, Weiss JA, Ellis BJ, Peters CL, Anderson AE. 52. Martin HD, Savage A, Braly BA, Palmer IJ, Beall DP, Kelly B. The
A new discrete element analysis method for predicting hip joint function of the hip capsular ligaments: a quantitative report.
contact stresses. J Biomech 2011;72:181e204. Arthrosc e J Arthrosc Relat Surg 2008;24:188e95.
35. Ardila OJ, Divo EA, Moslehy FA, Rab GT, Kassab AJ, Price CT. 53. Aspden RM, Rudman KE, Meakin JR. A mechanism for
Mechanics of hip dysplasia reductions in infants using the balancing the human body on the hips. J Biomech 2006;39:
Pavlik harness: a physics-based computational model. 1757e9.
J Biomech 2013;46:1501e7. 54. Dalstra M, Huiskes R. Load transfer across the pelvic bone.
36. Huayamave V, Rose C, Serra S, Jones B, Divo E, Moslehy F, et al. Biomechanics 1995;28:715e24.
A patient-specific model of the biomechanics of hip reduction 55. Dalstra M, Huiskes R, van Erning L. Development and valida-
for neonatal developmental dysplasia of the hip: investigation tion of a three-dimensional finite element model of the pelvic
of strategies for low to severe grades of developmental bone. J Biomech Eng 1995;117:272e8.
dysplasia of the hip. J Biomech 2015;48:2026e33. 56. Shepherd DE, Seedhom BB. The “instantaneous” compressive
37. Genda E, Konishi N, Hasegawa Y, Miura T. A computer simu- modulus of human articular cartilage in joints of the lower
lation study of normal and abnormal hip joint contact pres- limb. Ann Rheum Dis 1999;38:124e32.
sure. Arch Orthop Trauma Surg 1995;114:202e6. 57. Park S, Hung C, Ateshian GA. Mechanical response of bovine
38. Genda E, Iwasaki N, Li G, MacWilliams BA, Barrance PJ, articular cartilage under dynamic unconfined compression
Chao EY. Normal hip joint contact pressure distribution in loading at physiological stress levels. Osteoarthritis Cartilage
single leg standing. J Biomech 2001;34:895e905. 2004;12:65e73.
39. Yoshida H, Faust A, Wilckens J, Kitagawa M, Fetto J, Chao EYS. 58. Henak CR, Kapron AL, Anderson AE, Ellis BJ, Maas SA, Weiss JA.
Three-dimensional dynamic hip contact area and pressure Specimen-specific predictions of contact stress under physio-
distribution during activities of daily living. J Biomech logical loading in the human hip: validation and sensitivity
2006;39:1996e2004. studies. Biomech Model Mechanobiol 2014;13:387e400.
40. Chao EYS, Volokh KY, Yoshida H, Shiba N, Ide T. Discrete 59. Wong M, Ponticiello M, Kovanen V, Jurvelin JS. Volumetric
element analysis in musculoskeletal biomechanics. Mol Cell changes of articular cartilage during stress relaxation in un-
Biomech 2010;7:175e92. confined compression. J Biomech 2000;33:1049e54.

Please cite this article in press as: Vafaeian B, et al., Finite element analysis of mechanical behavior of human dysplastic hip joints: a systematic
review, Osteoarthritis and Cartilage (2016), http://dx.doi.org/10.1016/j.joca.2016.10.023
10 B. Vafaeian et al. / Osteoarthritis and Cartilage xxx (2016) 1e10

60. Ferguson SJ, Bryant JT, Ito K. The material properties of the 72. Niknafs N, Murphy RJ, Armiger RS, Lepisto € J, Armand M.
bovine acetabular labrum. J Orthop Res 2001;19:887e96. Biomechanical factors in planning of periacetabular osteot-
61. Caligaris M, Ateshian GA. Effects of sustained interstitial fluid omy. Front Bioeng Biotechnol 2013;1:1e10.
pressurization under migrating contact area, and boundary 73. Lepisto€ J, Armand M, Armiger RS. Periacetabular osteotomy in
lubrication by synovial fluid, on cartilage friction. Osteoar- adult hip dysplasia e developing a computer aided real-time
thritis Cartilage 2008;16:1220e7. biomechanical guiding system (BGS). Suom Ortoped Trauma-
62. Schmidt TA, Sah RL. Effect of synovial fluid on boundary tol 2008;31:186e90.
lubrication of articular cartilage. Osteoarthritis Cartilage 74. Murphy RJ, Armiger RS, Lepisto € J, Mears SC, Taylor RH,
2007;15:35e47. Armand M. Development of a biomechanical guidance system
63. Adeeb SM, Sayed Ahmed EY, Matyas J, Hart DA, Frank CB, for periacetabular osteotomy. Int J Comput Assist Radiol Surg
Shrive NG. Congruency effects on load bearing in diarthrodial 2015;10:497e508.
joints. Comput Methods Biomech Biomed Engin 2004;7: 75. Murphy RJ, Armiger RS, Lepisto € J, Armand M. Clinical evalua-
147e57. tion of a biomechanical guidance system for periacetabular
64. Pedersen R, Brand RA, Davy T. Pelvic muscle and acetabular osteotomy. J Orthop Surg Res 2016:11e36.
contact forces during gait. J Biomech 1997;30:959e65. 76. Henak CR, Anderson AE, Weiss JA. Subject-specific analysis of
65. Macirowski T, Tepic S, Mann RW. Cartilage stresses in the joint contact mechanics: application to the study of osteo-
human hip joint. J Biomech Eng 1994;116:10e8. arthritis and surgical planning. J Biomech Eng 2013;135:
66. Carter DR, Wong M. The role of mechanical loading histories in 21003.
the development of diarthrodial joints. J Orthop Res 1988;6: 77. Gannon AR, Nagel T, Bell AP, Avery NC, Kelly DJ. Postnatal
804e16. changes to the mechanical properties of articular cartilage are
67. Giorgi M, Carriero A, Shefelbine SJ, Nowlan NC. Mechanobio- driven by the evolution of its collagen network. Eur Cells
logical simulations of prenatal joint morphogenesis. J Biomech Mater 2015;29:105e23.
2014;47:989e95. 78. Assassi L, Magnenat-Thalmann N. Assessment of cartilage
68. Tsumura H, Kaku N, Ikeda S, Torisu T. A computer simulation contact pressure and loading in the hip joint during split
of rotational acetabular osteotomy for dysplastic hip joint: posture. Int J Comput Assist Radiol Surg 2016;11(5):745e56.
does the optimal transposition of the acetabular fragment 79. Windhagen H, Chincisan A, Choi HF, Thorey F. Soft-tissue
exist? J Orthop Sci 2005;10:145e51. balance in short and straight stem total hip arthroplasty. Or-
69. Armiger RS, Armand M, Tallroth K, Lepisto € J, Simon C. Three- thopedics 2015;38:S14e20.
dimensional mechanical evaluation of joint contact pressure in 80. Diederichs C, Heath A, Hareendranathan AR, Zonoobi D,
12 periacetabular osteotomy patients with 10-year follow-up. Kuntze G, Dulai S, et al. Cross-modality validation of acetabular
Acta Orthop 2009;80:155e61. surface models using 3-D ultrasound versus magnetic reso-
70. Rab GT. Lateral acetabular rotation improves anterior hip nance imaging in normal and dysplastic infant hips. Ultra-
subluxation. Clin Orthop Relat Res 2007;456:170e5. sound Med Biol 2016;42:2308e14.
71. Armiger RS, Armand M, Lepisto J, Minhas D, Tallroth K, 81. Verbruggen SW, Loo JHW, Hayat TTA, Hajnal JV, Rutherford MA,
Mears SC, et al. Evaluation of a computerized measurement Phillips ATM, et al. Modeling the biomechanics of fetal move-
technique for joint alignment before and during periacetabular ments. Biomech Model Mechanobiol 2016;15(4):995e1004.
osteotomy. Comput Aided Surg 2007;12:215e24.

Please cite this article in press as: Vafaeian B, et al., Finite element analysis of mechanical behavior of human dysplastic hip joints: a systematic
review, Osteoarthritis and Cartilage (2016), http://dx.doi.org/10.1016/j.joca.2016.10.023

You might also like