Professional Documents
Culture Documents
Autotitrator
Autotitrator
Types of Titration.
Acid-base Titrations
Redox Titrations.
Precipitation Titrations.
Complexometric Titrations.
1. Acid-Base Titration
● The strength of an acid can be determined using a standard
solution of a base. This process is called acidimetry.
● The titrations are named after the reagent that used as follows;
Permanganate Titrations
Dichromate Titrations
Iodimetric and Iodometric Titrations
Permanganate Titrations
● the potassium permanganate is used as an oxidizing agent.
l2 + 2e →
– 2l …………….
– (reduction)
2l → I2 + 2e ……………. (oxidation)
– –
Hg2+ + → Hg(SCN)2
2SCN–
Ag + 2CN → [Ag(CN)2]
+ – –
● Then carefully guide the titration vessel upward into the titration stand A and tighten the threaded ring (B) B.
To remove the titration vessel from the titration stand, loosen the threaded ring approximately a quarter to half turn
C, and carefully pull the titration vessel downwards out of the titration stand D.
5.Inserting a burette :
● burettes are equipped with RFID transponder chip on holder (visible by the small, black cover plate D).
● This chip is used for reading and writing properties such as titrate name, concentration or usable life.
Screw the burette and holder onto the titrant bottle (see operating instructions for burettes).
● Turn the arrestment knob in the opposite direction to the arrow.
● Slide the burette either from the left (as illustrated) or from the right over the withdrawal position
B onto the titrator:
● The recesses on the driver arm A must be parallel to the slide-on part of the burette.
The piston rod C must be in the home position.
● Turn the arrestment knob in the direction of the arrow to secure the burette .
● Place the dispensing tube into one of the openings in the titration stand .
Note:
When rinsing burettes and tubes, place tips from dispensing tubes in a waste container.
6.Connecting the measuring electrode and the compact stirrer :
● Insert the measuring electrode into the correct opening in the titration stand.1
● Use the electrode cable to connect the measuring electrode to the "SENSOR" socket on the rear of the titrator.
● Insert the stirrer into the correct opening in the titration stand.
● Connect the stirrer to the STIRRER / PUMP outlet of the titrator .Both the turning on and off, as well and the
stirrer speed are controlled by the software.
Procedure :
1. Turn on G20 from on / off button.
2. Connect G20 to CPU if it not connected using USB Cable .
3. From CPU Click on Start menu >> All Programs >> Labx>>Labx Titration another easy Way Click on
Labx icon on the Desktop .
4. The Labx software is now Open , Click on + sign beside titrators to find our Titrator .
● The following Page will appear ,you will find Labx and titrators method on the left to activate them Click on
+ sign Beside Titrator and Methods also you can find Methods shortcuts on the right , and titration graph on
the middle ,on the top titrator Components menu you can use it to control titrators components individually .
Titrator Components
Titration Graph
Methods shortcuts
Titration results
● before you run any method check the if connected Burette and sensor are suitable for
this method and check the dispensing tube if it Contain any bubbles or Air Dispense it
by Choosing Burette Symbol from titrator components menu on the top then choose
the volume then Click on Play Button .
● To run method go to Methods shortcuts and choose it by double click on it's icon the
following box will appear , if V (mL) Box is Active add your sample volume or Weight
then Click Start.
Notes
● The following message will appear on Titrator screen " add sample 1/1 " click ok to
start titration.
● After the titration is end the titrator will stop automatically and
give you results according to your method calculations .
● Clean your vessel , Stirrer , Dispensing tube , Sensor after Titration and keep your:
DP 5 Phototrode in Demi water.
DG 115 Potentiometric in Buffer 4 or 3M KCl solution .
Calibration
How to calibrate PH sensor
From Method icon, select M436 method “calibration DG 115-SC”, then press start.
Or from home screen , press Cal shortcut , then press start
Or from labx software press right click on calibration DG 115-SC method then select run
Then rinse the electrode with Demi water.
Put buffer PH 4, then ok
Then put buffer PH 7 then ok
Then put buffer PH 10, then OK
Wait unit the sensor finished updating results
From the results, if slope of electrode from - 55 to - 65
The calibration is Ok
Measuring Samples
How to analysis sulfuric acid sample
Open method icon , select sulfuric content (EP) method.
Or select method from home screen, select sulfuric shortcut.
Or from labx software press right click on sulfuric content (EP)
method then select run.
press start, insert weight of sample 0.5 to 1 g in sample size icon.
Then start, put sample beaker in titrator, then OK .
How to measure PH for any solution
Open labx software from desktop Pc
From Manual operations (upper task bar of labx)
Then select measure, select sensor type from table (PH; DG 115-SC;
PH, sensor, Pnp)
Then press on play icon
May be, you need to change time of stirring from table.
Take volume of sulfuric acid std solution in conical flask, then put
mixed indicator.
Then from manual icon, select burrette, then manual titration, then
press dispense drop by drop.
How to create new method from Labx software
Press right click on methods, select new method, then Gx, then Mettler.
Select any method will be comply with your application, for example
end point method or equivalence point.
Then change title to create new method.
Then change parameters in this method to adjust your method, for
example
Titrant name
Equation
Sensor type
Recognizing the Endpoints of Automated Titrations
Different Detection Principles
● The endpoint (EP) can be detected in an automated titration in a number of ways.
● Aside from the visual endpoint recognition (e.g., by the appearance of turbidity, a color change, or appearance
of a precipitate), a titration EP can also be detected by the automated monitoring of a change in a chemical or
physical property which happens once the reaction is complete.
● The table below shows that there are many different detection principles:
Potentiometric Principle
● In the potentiometric principle, the concentration-dependent potential (mV) of a solution is measured against a
reference potential, as shown in the table. So, a silver-silver chloride (Ag/AgCl) reference electrode is utilized
together with a measuring electrode (pH-sensitive glass membrane or metal ring).
● Typically, a combined sensor (electrode) including both measuring and reference electrodes is employed.
Step 1: Start the titration before the titrant is added.
Step 2: Addition of titrant. As the titration nears the endpoint see signs of the color change start to be observed.
At this point in an automatic titration, the sensor will detect an alteration in mV signal, and the titrator starts to
dose the titrant in smaller volumes and at a slower rate.
Step 3: The EP is reached with a faint pink color that corresponds with the inflection point in the titration curve.
Step 4: Titrating beyond the endpoint results in over titration, and the mV signal is fairly constant.
The steps above outline how to achieve the characteristic S-shaped titration curve observed when carrying out
an automated titration. It is not just acid-base titrations that can be converted; Figure 2 demonstrates how a
simple chloride titration can be converted.
Photometric Principle
● Titrations that utilize color indicators are still widely employed, e.g., in pharmacopeias. When manually
performed, the results literally depend on the eye of the beholder.
● Photometric titration it possible to replace this subjective determination of the equivalence point with an
objective process completely independent of the human eye.
● The basis of photometric indication is the change in intensity at a certain wavelength of a light beam passing
through a solution.
● In photometry, the transmission is the primary measured variable and is supplied by the light transmission
(mV or % transmission) of a colored or turbid solution which is measured with a photometric sensor.
● An example of how to convert an EDTA titration of manganese sulfate from manual titration to
automated titration is shown above, and the procedure remains the same.
● One benefit of automated titration is that a lower volume of chemicals is required, resulting in less
waste. The photo electrode is used at a wavelength of 610 nm with the same indicator, Eriochrome
Black TS.
THANK YOU
Best regards,
Ahmed Talaat Elnagar
Lab. Junior Chemist