CALIBRATION MANUAL
SAMBO Corporation
Tel : 02-866-3125 Fax : 02-866-3129 E-mail : sambo@sbcorp.co.kr Home Page : http://www.sbcorp.co.kr
#503 Kolon Techno Valley, 60-4 Kasan-Dong, Kumchon-Gu, Seoul, Korea
1. Operation Screen Display This display mode of Controller.
in operation screen
parameter
number
explanation for
parameter & setting
4) When you change parameter ,press key, then white input window
will display in bottom line.
when input window display,press a number and key,
it will go to desired parameter
Ex)When parameter 43'setting value want
change to 0.152221
After confirming cursor is located in [No.],
→ → →
→ →
setting value.
2/5
3. CALIBRATION
1) In operation screen, when you press key, below screen will display.
adjustment start.
③ When it arrives a standard pulse, Controller
stop calculating automatically.
3/5
☞ press key to go to former screen.
adjustment screen.
is inputted.
⑥ After doing 3 times a ①~⑤, when average value is in scope of ±0.25%,
adjustment is finished.
note) Adjustment is completed, you don’t need to press ,
4/5
4. Error revision
* Need error revision for after material feed
1) By changing parameter No.43, a error can be revised.
parameter Number
explanation for
parameter & setting
Ex) In case real quantity- 100.00 ton, indication value - 99.45 ton ,
5/5