N-Terminal Pro-Brain Natriuretic Peptide Monoclonal - Dimension - Rev H DXDCM 09017fe980338ff4-1705666298857
N-Terminal Pro-Brain Natriuretic Peptide Monoclonal - Dimension - Rev H DXDCM 09017fe980338ff4-1705666298857
NT-proBNP (pg/mL)
CHF Population – All
NYHA Functional Class 15,000
Intercept Correlation
0.9 Comparative Method Slope pg/mL Coefficient ng
Dimension® PBNP 1.02 -5 0.99 122
0.8 f. Model equation by Passing-Bablok linear regression is: [results for Dimension® system] = slope x [comparative
Sensitivity (Dimension® (NT-proBNP)
0.6 Specificity
HIL interference
0.5 The NT-proBNP method was evaluated for interference according to CLSI/NCCLS EP7-A2.26 Bias is the difference
in the results between the control sample (without the interferent) and the test sample (contains the interferent)
0.4 expressed in percent. Bias exceeding 10% is considered interference.
Messbereich 10 – 30000 pg/ml Männer <55 Jahre 55 – 64 Jahre 65 – 74 Jahre ≥75 Jahre
Kalibrationsmaterial Dimension® PBNP-Kalibrator, Art.- Nr. RC423A Mittelwert 29.7 37.9 76.0 414.3
Kalibrierschema Level 1 n = 4, Level 2, 3 und 5 n = 3, Level 4 n = 2 SA 42.5 41.6 – 889.0
Einheiten pg/ml Medianwert 13.4 13.4 76.0 111.8
Typische Kalibrator-Level 0, 300, 1500, 12000, 30000 pg/ml 95. Perzentil 134.4 114.3 76.0 1475.7
Kalibrationshäufigkeit Alle 30 Tage mit derselben Charge % <125 pg/ml 95 % 100 % 100 % –
Eine neue Kalibration ist • Für jede neue Charge von Flex®-Reagenzkassetten % <450 pg/ml – – – 77 %
erforderlich • Nach größeren Wartungs- oder Servicemaßnahmen, falls die Ergebnisse der N 76 6 1 62
Qualitätskontrolle dies nahelegen
• Nach Maßgabe der Qualitätskontrollverfahren des Labors
• Nach Maßgabe behördlicher Vorschriften Frauen <55 Jahre 55 – 64 Jahre 65 – 74 Jahre ≥75 Jahre
Startkoeffizienten C0 -2826.0 Mittelwert 34.2 39.9 84.1 291.0
C1 16014.0 SA 34.0 51.6 58.9 639.7
C2 -2.2 Medianwert 23.7 18.8 59.5 131.3
C3 18293.0 95. Perzentil 75.8 163.1 202.6 1080.7
C4 0.5 % <125 pg/ml 99 % 89 % 86 % –
% <450 pg/ml – – – 88 %
Kalibrations-Skaler Siehe Etikett auf dem Umkarton. Vor der Kalibration müssen N 87 9 7 60
chargenspezifische Flex®-Skaler in den Kalibrationsbildschirm A, B, C und D
eingegeben werden.
Erkrankungsgruppe der Studie Testauswertung
Es wurden Blutproben von 227 Patienten gewonnen, bei denen eine kongestive Herzinsuffizienz (CHF) Die Receiver Operator Characteristics (ROC)-Kurve zeigt die klinische Sensitivität und Spezifität bei verschiedenen
diagnostiziert wurde. Die Population umfasste 69 Frauen und 158 Männer. Die Zahlenangaben und die Cutoffs für die 227 Patienten mit CHF und 308 Personen ohne CHF. Im Folgenden sehen Sie die ROC-Kurve
Funktionsklassen der New York Heart Association (NYHA) finden Sie im Folgenden. Jedes Labor sollte einen für den Dimension® NT-proBNP-Test. Der Bereich unter der Kurve (AUK) für den Dimension® NT-proBNP-Test
Referenzbereich ermitteln, der der zu untersuchenden Patientenpopulation entspricht. Darüber hinaus sollte sich beträgt 0.921 mit einem 95%-Konfidenzintervall von 0.898 bis 0.943.
das Labor über die aktuell in seiner Einrichtung praktizierte Untersuchungsmethodik bei CHF im Klaren sein.
CHF-Population – Alle
<55 Jahre 55 – 64 Jahre 65 – 74 Jahre ≥75 Jahre 1
Mittelwert 2400.1 3120.7 4167.4 4579.3
SA 3855.9 8809.1 8550.7 8721.2
Medianwert 660.5 533.3 1506.7 2513.4 0.9
95. Perzentil 10961.2 16817.2 10889.3 11398.6
% >125 pg/ml 79 % 73 % 90 % – 0.8
0.6
CHF-Population – Männer
<55 Jahre 55 – 64 Jahre 65 – 74 Jahre ≥75 Jahre
Mittelwert 2748.2 3475.1 4427.8 5773.0 0.5
SA 4179.3 10332.0 9669.3 10238.5
Medianwert 1148.2 821.7 1273.4 3086.6 0.4
95. Perzentil 12127.1 16817.2 10889.3 14118.8
% >125 pg/ml 77 % 64 % 89 % – 0.3
% >450 pg/ml – – – 92 %
N 34 32 45 47 0.2
CHF-Population – Frauen 0.1
<55 Jahre 55 – 64 Jahre 65 – 74 Jahre ≥75 Jahre
Mittelwert 1554.7 2453.7 3435.2 2028.9
SA 2886.2 5024.5 4175.0 2581.3 0
Medianwert 257.8 492.9 1828.4 1077.1 0 0.5 1
95. Perzentil 10961.2 20855.8 13287.1 8891.2
% >125 pg/ml 86 % 53 % 94 % – 1-Spezifität (Dimension® NT-proBNP)
% >450 pg/ml – – – 91 % Im Folgenden finden Sie einen Boxplot der Population der klinischen Studie. Die empfohlenen klinischen
N 14 17 16 22 Grenzwerte, unter Ausschluss chronischer Herzinsuffizienz, liegen bei 125 pg/ml für Patienten unter 75 Jahren
und 450 pg/ml für Patienten ab 75 Jahren. Drei Proben aus der Erkrankungsgruppe mit Werten oberhalb des
Messbereichs sind in der Grafik nicht mit erfasst.
CHF-Population – Alle
NYHA-Funktionsklasse Dimension® NT-proBNP-Werte für Referenzgruppe und Erkrankungsgruppe
Alle CHF NYHA I NYHA II NYHA III NYHA IV
Mittelwert 3693.0 1492.9 2124.5 5423.2 7134.5
25,000
SA 7911.8 2260.2 2946.1 10692.5 11900.1
Medianwert 1422.0 659.2 1077.1 2636.5 2754.5
5. Perzentil 59.3 47.9 29.4 117.4 76.4
20,000
95. Perzentil 11398.6 5748.8 8344.4 13287.1 20855.8
% > Cutoff 85 % 76 % 83 % 90 % 90 %
NT-proBNP (pg/ml)
CHF-Population – Männer,
NYHA-Funktionsklasse 5,000
Alle CHF NYHA I NYHA II NYHA III NYHA IV
Mittelwert 4273.6 1706.9 2582.3 6273.3 9300.3
SA 9109.3 2242.9 3238.0 12099.8 15619.8 0
Medianwert 1710.7 938.6 1273.4 3269.9 2717.1
5. Perzentil 60.3 57.8 33.1 122.1 76.4
95. Perzentil 13600.1 5748.8 8344.4 15686.3 63515.3 Referenzgruppe Erkrankungsgruppe
% > Cutoff 86 % 81 % 83 % 93 % 88 %
Minimum 21.7 21.7 27.3 60.3 76.4 Sensitivität und Spezifität im Vergleich mit Alter und Geschlecht
Maximum 70025.3 10613.6 16817.2 70025.3 63515.3
N 158 41 48 53 16 Die folgenden Tabellen zeigen die klinische Sensitivität und Spezifität des Dimension® NT-proBNP-Tests bei
einem Cutoff von 125 pg/ml für Patienten unter 75 Jahren und 450 pg/ml für Patienten ab 75 Jahren.
Isosorbide dinitrate 6.0 mg/dl 254 µmol/l Sensibilité fonctionnelle : ≤30 pg/ml
Lidocaïne 1.2 mg/dl 51.2 µmol/l Elle représente la plus faible concentration d'analyte qui puisse être mesurée avec une précision
Lisinopryl 16 µg/ml 36.2 µmol/l intra-laboratoire observée de 20% CV.
Lithium 2.2 mg/dl 3.2 mmol/l
Lovastatine 16 µg/ml 39.6 µmol/l Fabriqué sous licence de Roche Diagnostics.
US 5786163 et équivalents étrangers
Application du brevet américain : 09/890442 et équivalents étrangers
Explication des symboles : Voir le tableau ci-contre.
Bibliographie : Voir le tableau ci-contre.
Siemens Healthcare Diagnostics Inc. Dimension® et Flex® sont des marques commerciales de Siemens Healthcare Diagnostics.
500 GBC Drive ©2009 Siemens Healthcare Diagnostics
Newark, DE 19714 USA Tous droits réservés.
RF423A
Frammento N-terminale del propeptide natriuretico cerebrale NT-proBNP + CrO2-Ab + CrO2-Ab-NT-proBNP- F(ab’)2-ALP + CrO2-
F(ab’)2-ALP Ab + F(ab’)2-ALP
Uso previsto: Il test PBNP utilizzato sul sistema di chimica clinica Dimension® con il modulo
lavaggio
immunodosaggio eterogeneo è un test diagnostico in vitro finalizzato alla determinazione quantitativa
CrO2-Ab-NT-proBNP-F(ab’)2-ALP + CrO2-Ab-NT-proBNP-F(ab’)2-ALP + CrO2-Ab
del frammento N-terminale propeptide natriuretico cerebrale (NT-proBNP) in siero o plasma umani. Nei
CrO2-Ab + F(ab’)2-ALP (trasferito nella cuvette)
pazienti con sospetta insufficienza cardiaca congestizia (CHF), le misurazioni di NT-proBNP vengono
utilizzate come supporto alla diagnosi e valutazione della gravità del caso. Il test è inoltre indicato per la CrO2-Ab-NT-proBNP-F(ab’)2-ALP
stratificazione del rischio dei pazienti con sindrome coronarica acuta e insufficienza cardiaca. FADP FAD + Pi
Riassunto: La disfunzione ventricolare sinistra può manifestarsi come parte della coronaropatia, FAD
dell’ipertensione arteriosa, della valvulopatia e della miocardiopatia primaria. Se la disfunzione ventricolare apo D-amminoacido ossidasi olo D-amminoacido ossidasi
sinistra non viene curata ed è progressiva, la probabilità di mortalità è elevata e la causa potrebbe essere olo D-amminoacido ossidasi
la morte cardiaca improvvisa. D-prolina +O2 H2O2
L'insufficienza cardiaca cronica è una sindrome clinica causata dalla compromissione della funzionalità HRP
della pompa del cuore. In base alla sintomatologia, la gravità dell'insufficienza cardiaca viene classificata H2O2 + DCHBS + 4-AAP Colore (assorbimento a 510 nm)
secondo classi da I a IV dalla New York Heart Association (NYHA). Per la diagnosi della disfunzione
ventricolare sinistra si utilizzano test clinici e tecniche di imaging.1 Reagenti
Nel controllo della funzione del sistema cardiovascolare è stata riconosciuta l'importanza dei peptidi Pozzettia Forma Componente Concentrazioneb Origine
natriuretici. I primi studi effettuati rivelano che i peptidi natriuretici possono essere utilizzati per 1, 2 Compresse c FADP 0.056 mg/ml
diagnosticare problemi clinici associati alla disfunzione ventricolare sinistra.2 I seguenti peptidi natriuretici D-prolina 68.8 mg/ml
trovano riscontro nella letteratura: peptide natriuretico atriale (ANP), peptide natriuretico cerebrale (BNP) 3, 4 Compressec Apo-D-amminoacido ossidasi 0.23 mg/ml Rene porcino o
e peptide natriuretico di tipo C (CNP).3,4 ricombinante, E. coli
I peptidi ANP e BNP hanno proprietà natriuretiche e diuretiche. Quali antagonisti del sistema renina- Perossidasi di rafano 0.19 mg/ml Rafano
angiotensina-aldosterone, infuiscono sull'equilibrio degli elettroliti e dei fluidi nell'organismo.5,6 Nei soggetti DCHBS 3.9 mg/ml
affetti da disfunzione ventricolare sinistra, le concentrazioni di BNP presenti nel siero e nel plasma 4-AAP 0.94 mg/ml
aumentano alla stregua delle concentrazioni del proBNP, pro-ormone biologicamente inattivo. Il ProBNP 5 Vuoto
è costituito da 108 amminoacidi. Viene secreto principalmente dal ventricolo sinistro del cuore e, nel corso 6 Compressec Anticorpo-CrO2 1.3 mg/mld Ovina
di questo processo, si suddivide in BNP fisiologicamente attivo (amminoacidi 77 – 108) e frammento 7 Liquida Tampone e stabilizzanti
N-terminale di proBNP (amminoacidi 1 – 76).4 8 Liquida d
Anticorpo-fosfatasi alcalina Tampone Ovina
e stabilizzanti
Gli studi indicano che il frammento NT-proBNP può essere utilizzato in applicazioni diagnostiche a. I pozzetti sono numerati consecutivamente a partire dall'estremità larga della cartuccia.
e prognostiche.7-13 La concentrazione di NT-proBNP nel plasma è un indicatore prognostico per la b. Valore nominale nella cartuccia idratata.
disfunzione ventricolare sinistra. È inoltre un utile strumento per l’attribuzione della sintomatologia a cause c. Le compresse contengono eccipienti, tamponi e stabilizzanti.
cardiache o di diversa natura. d. La titolazione dell'anticorpo e l'attività del coniugato variano da un lotto all'altro.
La determinazione dell'NT-proBNP consente di identificare i soggetti con disfunzione ventricolare sinistra.
Eventuali variazioni delle concentrazioni di NT-proBNP possono essere utilizzate per valutare l'efficacia del Rischio e sicurezza:
trattamento nei pazienti con disfunzione ventricolare sinistra. Alcuni elementi indicano che, per via delle H317
sue funzioni, l'NT-proBNP è indicato per l'uso nella valutazione del rimodellamento vascolare contribuendo P280, P272, P302 + P352, P333 + P313, P501
alla definizione dei programmi di riabilitazione personalizzati.10-11 Avvertenza!
Fisher et al hanno rilevato che i pazienti affetti da insufficienza cardiaca congestizia con elevati valori di Può provocare una reazione allergica cutanea.
NT-proBNP presentavano un tasso di mortalità entro un anno pari al 53% rispetto all'11% nei pazienti con Indossare guanti/indumenti protettivi/Proteggere gli occhi/il viso. Gli indumenti da lavoro
valori inferiori.12 Lo studio Gusto IV, che ha coinvolto più di 6800 pazienti, ha riconosciuto nel frammento contaminati non devono essere portati fuori dal luogo di lavoro. IN CASO DI CONTATTO CON
NT-proBNP il principale predittore indipendente di mortalità entro un anno in pazienti con sindrome LA PELLE: lavare abbondantemente con acqua e sapone. In caso di irritazione o eruzione
coronarica acuta.13 della pelle: consultare un medico. Smaltire il prodotto e il contenitore in conformità con tutte le
disposizioni locali, regionali e nazionali.
Nelle sue linee guida, la European Society of Cardiology Task Force for the Diagnosis and Treatment
of Chronic Heart Failure (Task Force della Società Europea di Cardiologia sulla diagnosi e la cura Contiene: 2-Cloracetamide
dell'Insufficienza cardiaca cronica) sostiene che i peptidi natriuretici, compreso NT-proBNP "possono Le schede di sicurezza sono disponibili sul sito siemens.com/healthcare
rivelarsi più utili dal punto di vista clinico come test di esclusione a causa di valori predittivi negativi molto
alti e coerenti". Precauzioni: Contiene azoturo di sodio (<0.1%) come conservante. L’azoturo di sodio può reagire con le
tubazioni in rame o piombo nelle linee di scarico formando composti esplosivi. Provvedere allo smaltimento in
I livelli di NT-proBNP aumentano nei pazienti con angina instabile e in seguito a infarto miocardico.14,15 modo appropriato e secondo le normative locali.
Benché non patognomonici, diversi studi hanno indicato come le misurazioni di NT-proBNP forniscano
informazioni prognostiche per la stratificazione del rischio a breve e lungo termine in pazienti con angina Le cuvette usate contengono liquidi di origine umana; maneggiare con cura per evitarne il contatto con la pelle
instabile o infarto miocardico.15,16,17 o l'ingestione.
Principi del metodo: Il metodo PBNP è un immunodosaggio enzimatico monofase basato sul principio Per uso diagnostico in vitro
del "sandwich". Il campione viene incubato con particelle di biossido di cromo rivestite di anticorpi Preparazione del reagente: L'idratazione, la diluizione e la miscelazione vengono eseguite automaticamente dal
monoclonali, che riconoscono gli epitopi situati nel frammento N-terminale di proBNP, e un anticorpo sistema di chimica clinica Dimension®.
monoclonale marcato con reagente coniugato [fosfatasi alcalina (ALP)] specifico per un secondo epitopo
indipendente su NT-proBNP, per formare un sandwich costituito da particella/NT-proBNP/coniugato. Il Conservare a: 2 – 8 °C
coniugato non legato viene rimosso mediante separazione magnetica e lavaggio. Dopo la separazione e Scadenza: Per la data di scadenza delle singole cartucce reagenti ancora chiuse fare riferimento alla confezione.
il lavaggio, il sandwich particella/NT-proBNP/coniugato viene trasferito nella provetta in cui la fosfatasi I pozzetti sigillati sullo strumento sono stabili per 30 giorni.
alcalina legata (ALP) avvia un'amplificazione a cascata.* La fosfatasi alcalina defosfora la flavina adenina
dinucleotide fosfato (FADP) producendo FAD. La FAD si lega all'apo D-amminoacido ossidasi e la Stabilità pozzetto aperto: 3 giorni per i pozzetti da 1 a 4
converte in olo D-amminoacido ossidasi. Ciascuna molecola di olo-D-amminoacido ossidasi produce 6 giorni per i pozzetti 6 a 8
varie molecole di perossido di idrogeno (H2O2). H2O2 in presenza di perossidasi di Radice di Rafano (HRP)
converte l’acido 3, 5-dicloro-2-idrossibenzensolfonico (DCHBS) e la 4-amminoantipirina (4-AAP) in un Raccolta e manipolazione dei campioni: I tipi di campioni raccomandati sono campioni di siero o plasma
prodotto colorato assorbente a 510 nm. La variazione cromatica misurata è direttamente proporzionale alla (litio eparina e sodio eparina).
concentrazione di NT-proBNP presente nel campione del paziente. Raccogliere il campione di siero o plasma utilizzando le procedure raccomandate.18 Per l'uso del dispositivo di
*Tecnologia su licenza di London Biotechnology, Ltd., Londra, Regno Unito. raccolta dei campioni e l'analisi, seguire le istruzioni fornite col dispositivo.19
I risultati dei test di confronto eseguiti sul sistema Dimension® per i campioni di siero rispetto ai campioni di
plasma con litio eparina o sodio eparina sono i seguenti:
Intercetta
Campione comparativo Pendenza pg/ml Coefficiente di correlazione n
Litio eparina 1.02 1.7 0.991 63
Sodio eparina 1.04 -4.7 0.996 58
L’equazione del modello per la statistica di regressione lineare Passing-Bablok è: Intervallo di misura analitica (AMR): 10 – 30000 pg/ml
Risultati siero = [pendenza x (risultati litio eparina)] + intercetta È l’intervallo dei valori di analita che è possibile misurare direttamente dal campione senza alcuna diluizione
Risultati siero = [pendenza x (risultati sodio eparina)] + intercetta o pretrattamento e che non sia parte integrante del processo di analisi abituale.
Prima della centrifugazione, attendere la completa formazione del coagulo. Separare fisicamente il siero I campioni con risultati superiori a 30000 pg/ml devono essere diluiti e rianalizzati.
o il plasma dalle cellule nel più breve tempo possibile; entro un limite massimo di due ore dal momento della Diluizione manuale: D iluire con il calibratore PBNP Livello 1 o diluente per campioni Dimension® per ottenere
raccolta.20 I campioni devono essere privi di materiale corpuscolato. risultati compresi nell'intervallo refertabile. Immettere il fattore di diluizione. Ripetere
I campioni sono stabili per 3 giorni a una temperatura di 20 – 25 °C, per 3 giorni a 2 – 8 °C e per 12 mesi se l'analisi. La lettura che ne risulta è quella corretta per la diluizione.
congelati a -20 °C o a una temperatura inferiore.21 Evitare di congelare e scongelare ripetutamente i campioni. Autodiluizione (AD): Sconsigliata.
Accertarsi che i campioni dei pazienti, i calibratori e i controlli abbiano raggiunto la temperatura ambiente
(22 – 28 °C) prima del test. I campioni scongelati o contenenti precipitati devono essere centrifugati prima I campioni con risultati inferiori a 10 pg/ml devono essere refertati come "inferiore a 10 pg/ml" anziché con
di eseguire il test. il valore numerico.
Non utilizzare sangue raccolto in presenza di EDTA, citrato o ossalato e campioni di plasma congelato con Limiti della procedura
anticoagulanti insufficienti in quanto possono potenzialmente causare l'agglutinazione delle particelle di cromo. Il sistema di refertazione dello strumento contiene avvisi e commenti per fornire all'utente informazioni sugli errori
di analisi dello strumento, sullo stato dello strumento e potenziali errori nei risultati relativi al NT-proBNP. Per il
Procedura significato di avvisi e commenti nei referti fare riferimento alla Guida per l'operatore del sistema Dimension®.
Materiale fornito Tutti i referti contenenti avvisi e/o commenti vanno gestiti in base al manuale di procedura del proprio laboratorio
Cartuccia reagente PBNP Flex®, Num. cat. RF423A e non refertati.
Materiale necessario ma non fornito Le concentrazioni di peptidi natriuretici possono essere elevate nei pazienti affetti da infarto miocardico acuto,
Calibratore PBNP Dimension®, Num. cat. RC423A nei soggetti candidati alla dialisi renale e negli individui che sono già stati sottoposti alla dialisi renale.
Recipienti di reazione, Num. cat. RXV1A I campioni dei pazienti possono contenere anticorpi eterofilici che potrebbero reagire negli immunodosaggi dando
Lavaggio chimico, Num. cat. RD701 luogo a risultati falsamente elevati o ridotti. Il test è stato concepito in modo da ridurre al minimo l'interferenza da
Detergente di pulizia della sonda reagenti, Num. cat. RD702 parte degli anticorpi eterofilici.
Detergente di pulizia della sonda campioni, Num. cat. RD703 Nonostante ciò non è possibile garantire la completa eliminazione di questa interferenza da tutti i campioni
Diluente campione, Num. cat. 791092901 di pazienti. Risultati del test incoerenti con il quadro clinico e l'anamnesi del paziente devono essere interpretati
con cautela.22,23
Prodotti di controllo qualità, ad esempio: La seguente precisione con 5 test indica la possibilità di un cattivo funzionamento del sistema:
MAS® CardioImmune® proBNP Liquid Assayed Cardiac Marker Control
Bio-Rad Liquichek™ Cardiac Markers Control LT Concentrazione SD
300 pg/ml >31 pg/ml
MAS® e CardioImmune® sono marchi registrati di Medical Analysis Systems, Inc. Camarillo, CA, USA. 1500 pg/ml >128 pg/ml
Liquichek™ è un marchio di Bio-Rad Laboratories, Irvine, California 92618, USA.
Valori attesi
Fasi del test
Le concentrazioni di NT-proBNP nel gruppo di riferimento vengono mostrate nelle tabelle riportate di seguito.
Il sistema Dimension® con il modulo immunodosaggio eterogeneo effettua automaticamente il campionamento, Ciascun laboratorio deve determinare il proprio intervallo di riferimento per il metodo PBNP eseguito sul sistema
l’erogazione del reagente, la miscelazione, l'analisi e la stampa dei risultati. Per ulteriori informazioni, fare Dimension®. Le soglie decisionali consigliate per queste distribuzioni sono riportate di seguito:
riferimento alla Guida per l'operatore di Dimension®.
Pazienti <75 anni: 125 pg/ml
Condizioni del test Pazienti ≥75 anni: 450 pg/ml
Recipiente di reazione Gruppo di studio di riferimento
Volume di campione 50 µl
Anticorpo-CrO2 25 µl La concentrazione di NT-proBNP è stata determinata in 308 individui non affetti da insufficienza cardiaca
Anticorpo-ALP 40 µl congestizia (163 donne e 145 uomini). Questa popolazione comprendeva individui apparentemente sani e
Volume del diluente 135 µl soggetti affetti da diabete, ipertensione e patologia polmonare. Le statistiche per le concentrazioni di NT-proBNP
Temperatura di incubazione 42 °C nel gruppo di studio di riferimento vengono mostrate nella tabella riportata di seguito.
Periodo di incubazione 6.4 minuti
Tutti <55 anni 55 – 64 anni 65 – 74 anni ≥75 anni
Cuvette Reazione Bianco Media 32.1 39.1 83.1 353.6
Volume trasferimento 40 µl 0 µl SD 38.2 46.2 54.6 775.8
Volume reagente FADP 30 µl 30 µl Mediana 17.9 18.8 59.5 125.6
Volume reagente APO 30 µl 30 µl 95° percentile 81.9 163.1 202.6 1372.8
Volume del diluente 350 µl 390 µl % <125 pg/ml 97% 93% 88% -
Temperatura 37 °C % <450 pg/ml - - - 83%
Lunghezza d'onda 510 e 700 nm N 163 15 8 122
Tipo di misurazione Cinetica bicromatica
Unità pg/ml Uomini <55 anni 55 – 64 anni 65 – 74 anni ≥75 anni
Calibrazione Media 29.7 37.9 76.0 414.3
Per informazioni sulla procedura di calibrazione generale, consultare la Guida per l'operatore di Dimension®. SD 42.5 41.6 - 889.0
Mediana 13.4 13.4 76.0 111.8
Il metodo PBNP richiede l’uso di scaler di calibrazione specifici per lotto. Vedere l’etichetta sulla confezione 95° percentile 134.4 114.3 76.0 1475.7
di PBNP Flex®. Prima della calibrazione, è necessario immettere gli scaler nella schemata di calibrazione. In % <125 pg/ml 95% 100% 100% -
caso contrario, saranno ottenuti risultati imprecisi. % <450 pg/ml - - - 77%
N 76 6 1 62
Intervallo di misura 10 – 30000 pg/ml
Materiale di calibrazione Calibratore PBNP Dimension®, Num. cat. RC423A
Schema di calibrazione Livello 1 n = 4, Livelli 2, 3 e 5 n = 3, Livello 4 n = 2 Donne <55 anni 55 – 64 anni 65 – 74 anni ≥75 anni
Unità pg/ml Media 34.2 39.9 84.1 291.0
Livelli di calibrazione tipici 0, 300, 1500, 12000, 30000 pg/ml SD 34.0 51.6 58.9 639.7
Frequenza di calibrazione Ogni 30 giorni per ciascun lotto Mediana 23.7 18.8 59.5 131.3
Occorre effettuare una nuova • Per ogni nuovo lotto di cartucce reagenti Flex® 95° percentile 75.8 163.1 202.6 1080.7
calibrazione • In seguito a manutenzione o riparazione importante, se indicato dai risultati % <125 pg/ml 99% 89% 86% -
del controllo qualità % <450 pg/ml - - - 88%
• Se indicato nelle procedure del controllo qualità del laboratorio N 87 9 7 60
• Quando richiesto in base alle normative in vigore
Coefficienti iniziali C0 -2826.0 Gruppo di studio della malattia
C1 16014.0
C2 -2.2 I campioni di sangue sono stati prelevati da 227 pazienti a cui era stata diagnosticata insufficienza cardiaca
C3 18293.0 congestizia. La popolazione comprendeva 69 donne e 158 uomini. Le statistiche descrittive e le classi funzionali
C4 0.5 della "New York Heart Association (NYHA)" vengono riportate di seguito. Ciascun laboratorio deve stabilire un
intervallo di riferimento che rappresenti la popolazione di pazienti da valutare. Inoltre, nei laboratori deve essere
Scaler di calibrazione Vedere l'etichetta sulla confezione. Prima della calibrazione, è necessario nota la prassi utilizzata presso il proprio istituto per la valutazione dell’insufficienza cardiaca congestizia.
immettere gli scaler specifici per il lotto Flex® nella schemata di calibrazione
A, B, C, e D. Popolazione CHF – Tutti
<55 anni 55 – 64 anni 65 – 74 anni ≥75 anni
Controllo qualità Media 2400.1 3120.7 4167.4 4579.3
Almeno una volta per ogni giorno di utilizzo, analizzare due livelli di un materiale di controllo qualità con SD 3855.9 8809.1 8550.7 8721.2
concentrazioni note di NT-proBNP. Seguire le procedure di controllo qualità interne del laboratorio se i risultati Mediana 660.5 533.3 1506.7 2513.4
ottenuti non rientrano nei limiti accettabili. 95° percentile 10961.2 16817.2 10889.3 11398.6
% >125 pg/ml 79% 73% 90% -
Risultati: Lo strumento calcola la concentrazione di NT-proBNP in pg/ml utilizzando lo schema di calcolo % >450 pg/ml - - - 91%
illustrato nella Guida per l'operatore di Dimension®. N 48 49 61 69
I risultati di questo test devono essere sempre interpretati alla luce della anamnesi del paziente, della
presentazione clinica e valutando contestualmente l'esito di altri accertamenti.
Popolazione CHF – Uomini Di seguito viene riportato un diagramma a scataola della popolazione dello studio clinico. Le soglie cliniche
<55 anni 55 – 64 anni 65 – 74 anni ≥75 anni raccomandate che consentono di escludere l'insufficienza cardiaca cronica sono di 125 pg/ml per i pazienti di
Media 2748.2 3475.1 4427.8 5773.0 età inferiore a 75 anni e di 450 pg/ml per i pazienti di età pari o superiore a 75 anni. Nel diagramma non sono
SD 4179.3 10332.0 9669.3 10238.5 rappresentati tre campioni del gruppo della malattia con valori superiori all’intervallo di misura.
Mediana 1148.2 821.7 1273.4 3086.6 Valori Dimension® NT-proBNP per i gruppi di studio di riferimento e della malattia
95° percentile 12127.1 16817.2 10889.3 14118.8
% >125 pg/ml 77% 64% 89% -
% >450 pg/ml - - - 92% 25,000
N 34 32 45 47
Popolazione CHF – Donne 20,000
<55 anni 55 – 64 anni 65 – 74 anni ≥75 anni
NT-proBNP (pg/ml)
Media 1554.7 2453.7 3435.2 2028.9
15,000
SD 2886.2 5024.5 4175.0 2581.3
Mediana 257.8 492.9 1828.4 1077.1
95° percentile 10961.2 20855.8 13287.1 8891.2 10,000
% >125 pg/ml 86% 53% 94% -
% >450 pg/ml - - - 91%
N 14 17 16 22 5,000
È stata verificata l'interferenza sul metodo NT-proBNP in base alle linee guida del CLSI/NCCLS EP7-A2.26 Il bias
0.7 è la differenza tra i risultati del campione di controllo (non contenente sostanze interferenti) e quelli del campione
di test (contenente sostanze interferenti), espressa in percentuale. Un errore sistematico superiore al 10% viene
0.6 considerato “interferenza”.
0
0 0.5 1
1-Specificità (Dimension® NT-proBNP)
Sostanze non interferenti Effetto gancio
Le seguenti sostanze non interferiscono con il metodo NT-proBNP, se presenti nel plasma nelle concentrazioni Gli immunodosaggi monofase a sandwich sono sensibili all’"effetto gancio" ad alte dosi, dove un eccesso
indicate. Le imprecisioni sistematiche (bias) dovute a tali sostanze sono inferiori al 10% ad una concentrazione di antigene ostacola il simultaneo legame della cattura e del rilascio degli anticorpi a una singola molecola
di NT-proBNP di 125 pg/ml. dell'analita. 27 Tali campioni devono essere diluiti e rianalizzati prima di refertare i risultati (vedere le informazioni
sulla diluizione). Il metodo NT-proBNP non mostra alcun effetto gancio fino ad almeno 300,000 pg/ml.
Sostanza Concentrazione del test Unità S.I.
Acetaminofene 20 mg/dl 1324 µmol/l Cross-reattività
Allopurinolo 2.5 mg/dl 184 µmol/l Alle concentrazioni indicate, le sostanze seguenti non hanno alcun effetto significativo (inferiore al 1%)
Amiodarone 20 µg/ml 31 µmol/l se aggiunte a campioni contenenti tra 0 e circa 125 pg/ml di NT-proBNP.
Amikacina 8.0 mg/dl 137 µmol/l Sostanza Unità
Amiodipina besilato 4.0 µg/ml 7.1 µmol/l ANP28 3.1 µg/ml
Ampicillina 5.3 mg/dl 152 µmol/l NT-proANP1-30 (preproANP26-55) 3.5 µg/ml
Acido ascorbico 6.0 mg/dl 342 µmol/l NT-proANP31-67 (preproANP56-92) 1.0 ng/ml
Atenololo 1.0 mg/dl 37.6 µmol/l NT-proANP79-98 (preproANP104-123) 1.0 ng/ml
Caffeina 6.0 mg/dl 308 µmol/l BNP32 (Natrecor®) 3.5 µg/ml
Captopril 5.0 mg/dl 230 µmol/l CNP22 2.2 µg/ml
Carbamazepina 3.0 mg/dl 127 µmol/l DNP 1.0 ng/ml
Cloramfenicolo 5.0 mg/dl 155 µmol/l VNP 1.0 ng/ml
Clordiazepossido 1.0 mg/dl 33.3 µmol/l Adrenomedullina 1.0 ng/ml
Clorpromazina 0.20 mg/dl 6.27 µmol/l Aldosterone 0.6 ng/ml
Colesterolo 503 mg/dl 13 mmol/l Angiotensina I 0.6 ng/ml
Cimetidina 2.0 mg/dl 79.2 µmol/l Angiotensina II 0.6 ng/ml
Creatinina 30 mg/dl 2.7 mmol/l Angiotensina III 1.0 ng/ml
Ciclosporina A 4000 ng/ml 3.3 µmol/l Endotelina 20 pg/ml
Destrano 40 3000 mg/dl 750 µmol/l Renina 50 ng/ml
Diazepam 0.51 mg/dl 18 µmol/l Urodilatina 3.5 µg/ml
Diclofenac 60 µg/ml 188.6 µmol/l Arg-Vasopressina 1.0 ng/ml
Digitossina 60 ng/ml 78.6 nmol/l
Digossina 6.1 ng/ml 7.8 nmol/l Linearità di diluizione
Diltiazem 120 µg/ml 266.1 µmol/l La linearità di diluizione è stata valutata confrontando i valori osservati con quelli attesi ottenuti con il metodo
Dipiridamolo 30 µg/ml 59.4 µmol/l Dimension® PBNP utilizzando campioni di siero a elevata concentrazione di NT-proBNP diluiti con campioni
Eritromicina 6.0 mg/dl 81.6 µmol/l di siero a bassa concentrazione di NT-proBNP. Altri campioni di siero sono stati diluiti in modo lineare per
Etanolo 400 mg/dl 86.8 mmol/l confermare l’accuratezza della diluizione ai valori inferiori dell’intervallo di misura. Nella tabella che segue sono
Etosuccimide 25.0 mg/dl 1770 µmol/l riassunti i risultati degli studi:
Furosemide 6.0 mg/dl 181 µmol/l
Gentamicina 1.0 mg/dl 21 µmol/l Campione siero alto (pg/ml) 31883 1177
Eparina, sodio 3.0 U/ml 3000 U/l Campione siero basso (pg/ml) 12 21
Idralazina 20 µg/ml 101.7 µmol/l N 7 5
Idroclorotiazide 20 µg/ml 67.2 µmol/l Statistiche di regressione
Ibuprofene 50 mg/dl 2425 µmol/l Pendenza 1.002 1.014
Indometacina 16 µg/ml 36.9 µmol/l Intercetta -68 -14
Isosorbide dinitrato 6.0 mg/dl 254 µmol/l Coefficiente di correlazione 0.999 0.999
Lidocaina 1.2 mg/dl 51.2 µmol/l % recupero
Lisinopril 16 µg/ml 36.2 µmol/l Media 99.7% 98%
Litio 2.2 mg/dl 3.2 mmol/l Intervallo 97% – 102% 92% – 102%
Lovastatina 16 µg/ml 39.6 µmol/l Sensibilità analitica: ≤10 pg/ml
L-tiroxina 60 µg/dl 0.77 µmol/l
Metildopa 2.5 mg/dl 118 µmol/l La sensibilità analitica rappresenta la concentrazione più bassa di PBNP che possa essere distinta dallo zero.
Milrinone lattato 2.4 µg/ml 11.4 µmol/l La sensibilità è definita come il valore medio (n = 20) più due deviazioni standard del Calibratore PBNP di
Nicotina 0.10 mg/dl 6.2 µmol/l livello 1 (0 pg/ml).
Nifedipina 6.0 mg/dl 173 µmol/l Sensibilità funzionale: ≤30 pg/ml
Nitrofurantoina 40 µg/ml 168 µmol/l
Nitroglicerina 0.16 µg/ml 0.577 µmol/l La sensibilità funzionale rappresenta la concentrazione più bassa di analita che possa essere misurata
Oxazepam 12 µg/ml 41.8 µmol/l con una precisione intra-laboratorio di 20% CV.
Ossitetraciclina 100 µg/ml 217.2 µmol/l Prodotto su licenza di Roche Diagnostics.
Penicillina G 25 U/ml 25000 U/l Brevetto USA 5786163 ed equivalenti stranieri
Pentobarbital 8.0 mg/dl 354 µmol/l Domanda di brevetto USA: 09/890442 ed equivalenti stranieri
Fenobarbital 10 mg/dl 431 µmol/l
Fenitoina 5.0 mg/dl 198 µmol/l Interpretazione simboli: Vedere la sezione aggiunta.
Primidone 4.0 mg/dl 183 µmol/l Bibliografia: Vedere la sezione aggiunta.
Probenecid 200 µg/ml 700.9 µmol/l Dimension® e Flex® sono marchi di Siemens Healthcare Diagnostics.
Propossifene 0.16 mg/dl 4.91 µmol/l ©2009 Siemens Healthcare Diagnostics
Propranololo 0.15 mg/dl 5.1 µmol/l Tutti i diritti riservati.
Proteine, albumina umana 6 g/dl 60 g/l
Proteine, IgG umane 5 g/dl 50 g/l
Proteina, totale alto 12 g/dl 120 g/l
Proteina, totale basso 3 – 4 g/dl 30 – 40 g/l
Chinidina 20 µg/ml 61.6 µmol/l
Fattori reumatoidi 500 IU/ml 500,000 IU/l
Acido salicilico 60 mg/dl 4.34 mmol/l
Simvastatina 32 µg/ml 76.5 µmol/l
Sulfametossazolo 320 µg/ml 1263.3 µmol/l
Teofillina 4.0 mg/dl 222 µmol/l
Trimetoprim 64 µg/ml 220.7 µmol/l
Urea 500 mg/dl 83 mmol/l
Acido urico 20 mg/dl 1.2 mmol/l
Acido valproico 50 mg/dl 3467 µmol/l
Verapamil 16 mg/dl 0.33 µmol/l
Warfarina 4 µg/ml 13.0 µmol/l
2019-04-08 H PN 755423.002 - IT
Intersección
Muestra comparativa Pendiente pg/mL Coeficiente de correlación n
Heparina de litio 1.02 1.7 0.991 63
Heparina de sodio 1.04 -4.7 0.996 58
El modelo de ecuación mediante la regresión lineal de Passing-Bablok es:
Resultados en suero = [pendiente x (resultados de heparina de Li)] + intersección
Resultados en suero = [pendiente x ( resultados con heparina de sodio)] + intersección
Antes de la centrifugación debe dejarse que se forme completamente el coágulo. El suero o el plasma deben Control de calidad
separarse físicamente de las células lo antes posible, con un límite máximo de dos horas desde el momento Al menos una vez por día de uso, analice dos niveles de un material de control de calidad (CC) con
de la obtención de la muestra.20 Las muestras no deben contener partículas sólidas. concentraciones conocidas de NT-proBNP. Siga los procedimientos internos de CC de su laboratorio si los
Las muestras son estables durante 3 días a 20 – 25 °C, 3 días a 2 – 8 °C y durante 12 meses si se congelan resultados obtenidos no se encuentran dentro de los límites aceptables.
a -20 °C o menos.21 Evite congelar y descongelar las muestras varias veces. Asegúrese de que las muestras Resultados: El instrumento calcula la concentración de NT-proBNP en pg/mL según el esquema de cálculo
de pacientes, los calibradores y los controles estén equilibrados a temperatura ambiente (22 – 28 °C) antes ilustrado en el Manual del usuario del sistema Dimension®.
del análisis. Las muestras congeladas que se hayan descongelado y las muestras que contengan precipitados
deben centrifugarse antes de realizar el análisis. Los resultados de esta prueba deberán interpretarse siempre de acuerdo con la historia clínica del paciente,
la sintomatología clínica y otras observaciones.
La sangre obtenida en presencia de EDTA, citrato u oxalato y las muestras de plasma congelado con una
cantidad insuficiente de anticoagulante podrían causar agregación de las partículas de cromo, por lo que no Rango de medición analítico (AMR): 10 – 30000 pg/mL
deben utilizarse. Se trata del rango de valores del analito que puede medirse directamente a partir de la muestra sin requerir
dilución ni tratamiento previo que no sea parte del proceso analítico habitual.
Procedimiento Las muestras con resultados que superen los 30000 pg/mL deben repetirse con dilución.
Materiales suministrados Dilución manual: D iluya con el nivel 1 del calibrador de PBNP o con el diluyente de muestras Dimension®
Cartucho de reactivos Flex® de PBNP, ref. RF423A para obtener resultados dentro del intervalo informable. Introduzca el factor de dilución en el
Materiales necesarios pero no suministrados instrumento. Repita el análisis. La lectura resultante se corregirá en función de la dilución.
Calibrador de PBNP de Dimension®, ref. RC423A Autodilución (AD): No recomendada.
Vasos de reacción, ref. RXV1A
Lavado químico, ref. RD701 Las muestras con resultados inferiores a 10 pg/mL deberán registrarse como “inferiores a 10 pg/mL” en lugar
del valor numérico.
Limpiador de sonda de reactivos, ref. RD702
Limpiador de sonda de muestras, ref. RD703 Limitaciones del procedimiento
Diluyente de muestras, ref. 791092901 El sistema de información del instrumento contiene alarmas y comentarios que proporcionan al usuario
información relativa a los errores de procesamiento del instrumento, información del estado de éste y posibles
Productos de control de calidad como: errores en los resultados de NT-proBNP. Consulte el Manual del usuario del sistema Dimension® para conocer
el significado de las alarmas y los comentarios de los informes. Cualquier informe que contenga alarmas y/o
MAS® CardioImmune® proBNP Liquid Assayed Cardiac Marker Control comentarios se debe tratar siguiendo el manual de procedimiento de su laboratorio y no se debe informar sobre él.
Bio-Rad Liquichek™ Cardiac Markers Control LT
Las concentraciones de péptidos natriuréticos pueden estar elevadas en los pacientes con infarto agudo de
MAS® y CardioImmune® son marcas registradas de Medical Analysis Systems, Inc. Camarillo, CA, USA. miocardio, en pacientes que sean candidatos a diálisis renal y en pacientes que se hayan sometido a diálisis renal.
Liquichek™ es una marca comercial de Bio-Rad Laboratories, Irvine, California 92618, USA.
Las muestras de paciente pueden contener anticuerpos heterófilos que podrían reaccionar en los inmunoensayos
Proceso del análisis y dar resultados falsamente elevados o reducidos. Este análisis se ha diseñado para reducir al mínimo la
El sistema Dimension® con el módulo de inmunoensayo heterogéneo realiza de manera automática el muestreo, interferencia causada por anticuerpos heterófilos.
la dispensación de reactivos, la mezcla, el procesamiento y la impresión de resultados. Para más detalles sobre Sin embargo, no es posible garantizar la completa eliminación de esta interferencia de todas las muestras de
este proceso, consulte el Manual del usuario del sistema Dimension®. paciente. Si un resultado de la prueba se contradice con el cuadro clínico y la historia del paciente, deberá
interpretarse con precaución.22,23
Condiciones del análisis
Existe la posibilidad de un funcionamiento incorrecto del sistema si se obtiene la siguiente precisión en
Vaso de reacción 5 pruebas consecutivas:
Volumen de la muestra 50 µL
Anticuerpo-CrO2 25 µL Concentración DE
Anticuerpo-ALP 40 µL 300 pg/mL >31 pg/mL
Volumen de diluyente 135 µL 1500 pg/mL >128 pg/mL
Temperatura de incubación 42 °C
Período de incubación 6.4 minutos Valores esperados
En las tablas siguientes se muestran las concentraciones de NT-proBNP en el Grupo de estudio de referencia.
Cubeta Reacción Blanco Cada laboratorio debe establecer su propio intervalo de referencia para el método PBNP según se determine en
Volumen de transferencia 40 µL 0 µL el sistema Dimension®. Los umbrales de decisión recomendados de estas distribuciones son:
Volumen de reactivo FADP 30 µL 30 µL
Volumen de reactivo APO 30 µL 30 µL Pacientes <75 años: 125 pg/mL
Volumen de diluyente 350 µL 390 µL Pacientes ≥75 años: 450 pg/mL
Temperatura 37 °C
Longitud de onda 510 y 700 nm Grupo de estudio de referencia
Tipo de medición Tasa bicromática Se determinó la concentración de NT-proBNP en 308 sujetos sin insuficiencia cardíaca congestiva
Unidades pg/mL (163 mujeres y 145 hombres). Esta población incluyó a sujetos aparentemente sanos y a sujetos con
Calibración diabetes, hipertensión y enfermedad pulmonar. En la tabla siguiente se muestran los datos estadísticos para las
concentraciones de NT-proBNP en el grupo de estudio de referencia.
En el Manual del usuario del sistema Dimension® se describe el procedimiento general de calibración.
Todos <55 años 55 – 64 años 65 – 74 años ≥75 años
El PBNP requiere ajustes de escala de calibración específicos del lote. Consulte la etiqueta del cartucho Flex® Media 32.1 39.1 83.1 353.6
de PBNP. Estos ajustes de escala se deben introducir en la pantalla de la calibración antes de proceder a la DE 38.2 46.2 54.6 775.8
misma. Si no se introducen estos ajustes de escala, los resultados serán inexactos.
Mediana 17.9 18.8 59.5 125.6
Percentil 95 81.9 163.1 202.6 1372.8
Intervalo del ensayo 10 – 30000 pg/mL % <125 pg/mL 97% 93% 88% -
Material de calibración Calibrador de PBNP de Dimension®, ref. RC423A % <450 pg/mL - - - 83%
Esquema de calibración Nivel 1 n = 4, niveles 2, 3 y 5 n = 3, nivel 4 n = 2 N 163 15 8 122
Unidades pg/mL
Niveles habituales de calibración0, 300, 1500, 12000, 30000 pg/mL
Frecuencia de calibración Cada 30 días para cualquier lote Hombres <55 años 55 – 64 años 65 – 74 años ≥75 años
Se requiere una nueva calibración• Para cada lote nuevo de cartuchos de reactivos Flex® Media 29.7 37.9 76.0 414.3
• Después de la realización de importantes tareas de mantenimiento o DE 42.5 41.6 - 889.0
servicio, si los resultados de control de calidad así lo indican Mediana 13.4 13.4 76.0 111.8
• Tal como se indica en los procedimientos de control de calidad del Percentil 95 134.4 114.3 76.0 1475.7
laboratorio % <125 pg/mL 95% 100% 100% -
• Cuando es obligatorio según las reglamentaciones gubernamentales % <450 pg/mL - - - 77%
Coeficientes de inicio C0 -2826.0 N 76 6 1 62
C1 16014.0
C2 -2.2
C3 18293.0 Mujeres <55 años 55 – 64 años 65 – 74 años ≥75 años
C4 0.5 Media 34.2 39.9 84.1 291.0
Ajustes de escala de calibración Véase la etiqueta del envase. Los ajustes de escala del lote de Flex® deben DE 34.0 51.6 58.9 639.7
incluirse en la pantalla de calibración A, B, C y D antes de la calibración. Mediana 23.7 18.8 59.5 131.3
Percentil 95 75.8 163.1 202.6 1080.7
% <125 pg/mL 99% 89% 86% -
% <450 pg/mL - - - 88%
N 87 9 7 60
Grupo de estudio de enfermedad Interpretación de los resultados
Se obtuvieron muestras de sangre de 227 pacientes con diagnóstico de insuficiencia cardíaca congestiva La curva de características operativas del receptor (ROC) presenta la sensibilidad y especificidad clínicas en
(ICC). La población incluyó a 69 mujeres y a 158 hombres. A continuación se muestran los datos estadísticos diversos valores límite para los 227 pacientes diagnosticados de ICC y los 308 sujetos sin ICC. A continuación,
descriptivos y las clases funcionales de la New York Heart Association (NYHA). Cada laboratorio debe establecer se muestra la curva ROC para los análisis de NT-proBNP con el sistema Dimension®. El área bajo la curva
un intervalo de referencia que represente la población de pacientes que se va a evaluar. Además, los laboratorios (AUC) para los análisis de NT-proBNP del sistema Dimension® es 0.921 con un intervalo de confianza del
deben conocer la práctica actual de su centro para la evaluación de la ICC. 95% de 0.898 a 0.943.
Población con ICC – Todos
<55 años 55 – 64 años 65 – 74 años ≥75 años 1
Media 2400.1 3120.7 4167.4 4579.3
DE 3855.9 8809.1 8550.7 8721.2
Mediana 660.5 533.3 1506.7 2513.4 0.9
Percentil 95 10961.2 16817.2 10889.3 11398.6
% >125 pg/mL 79% 73% 90% - 0.8
0.6
Población con ICC – Hombres
<55 años 55 – 64 años 65 – 74 años ≥75 años 0.5
Media 2748.2 3475.1 4427.8 5773.0
DE 4179.3 10332.0 9669.3 10238.5
Mediana 1148.2 821.7 1273.4 3086.6 0.4
Percentil 95 12127.1 16817.2 10889.3 14118.8
% >125 pg/mL 77% 64% 89% - 0.3
% >450 pg/mL - - - 92%
N 34 32 45 47 0.2
6LHPHQV+HDOWKFDUH'LDJQRVWLFV,QF
*%&'ULYH
1HZDUN'(86$
2019-04-08 H PN 755423.002