Verifikasi LAF Karawang Foods
Verifikasi LAF Karawang Foods
Jl. Kapuk Blok F19 No. 9, Kawasan Industri Delta Silikon 3, Lippo
Alamat Pemilik
Cikarang, Bekasi, 17530 - Indonesia
Order Code : LBF 032024/0002 Start Temperature Room :
Costumer Name : PT. Karawang Foods Lestari End Temperature Room :
Equipment Name : Laminar Air Flow Start RH Room :
Cabinet Model : Horizontal End RH Room :
Verification Status : Re - Verification Location Test :
Merk : JISICO Engineer Verification :
Type : J-CBWV2
S/N : 2101-01 Thermohygrometer :
Receipt Date : 3/12/2024
Verification Date : 3/12/2024
Additional ID : Lab Mikrobiologi
Air Velocity ; Light Intensity ;
Test Sampling :
Ultraviolet ; Noise Level
Acceptance Criteria
1. Average Air Velocity : 0.40 m/s ~ 0.50 m/s
2. All Individual Velocities : The Individual point readings that do not vary more than ± 20% of the
Acceptance Criteria
Average light intensity shall be no less than 480 LUX greater than backgroung levels, where backgrou
Acceptance Criteria
1. The measured lamp intensity should be greater than the minimum intensity of 100 micro watts/cm
2. The ultraviolet lamp should be replaced when the lamp's intensity drops to 40 micro watts/cm²
TEST 4. NOISE LEVEL TEST Table N-1.1 Correction chart for sound level r
ter : TH-002
5 6
0.5 0.4
0.4 0.4
0.4 0.5
6
10
1238
On - Site
In - Lab
TH-001
TH-002
TH-003
TH-004
TH-005
TH-006
TH-007
TH-008
TH-009
TH-010
ORDER CODE
FORM VERIFICATION : Laminar Air Flow (Lab Mikrobiologi) Bio Safety Cabinet Cabinet Model New-Verification
COSTUMER NAME : PT. Karawang Food Lestari
✘ Laminar Air Flow Horizontal Vertical ✘ Re-Verification
MERK : JISICO Fume Hood
TYPE : J-CBWV2
S/N : 2101-01
DATE VERIFICATION : 3/12/2024
ROOM TEMPERATURE : 22,0 - 22,5 °C
TEST 1. AIR VELOCITY ROOM HUMIDITY : 58 - 63 %
Velocity Test Site Test Data
1 2 3 4 5 6
• • • • • •
Position Left Side
m/s m/s m/s m/s m/s m/s
• • • • • •
Rear 0.40 0.40 0.40 0.40 0.50 0.40
Acceptance Criteria
1. Average light intensity shall be no less than 480 LUX greater than backgroung levels, where background light levels average a max. of 160 LUX
Meet Acceptance Criteria ✘ YES / PASS
( According to NSF 49 ) ✘ NO / FAIL
Acceptance Criteria
1. The measured lamp intensity should be greater than the minimum intensity of 100 micro watts/cm² for a new lamp
2. The ultraviolet lamp should be replaced when the lamp's intensity drops to 40 micro watts/cm²
Meet Acceptance Criteria ✘ YES / PASS
( According to NSF 49 )
NO / FAIL
Accepted Criteria 3 3