Professional Documents
Culture Documents
9852 2164 01d Maintenance Instructions Epiroc Tunnel Profiler
9852 2164 01d Maintenance Instructions Epiroc Tunnel Profiler
Profiler
Maintenance
Original Instructions
2018-09-18 | No: 8663980683.1 en
Epiroc Tunnel Profiler Table of Contents
Table of Contents
1 Assembly and dismantling of scanner unit........................................................ 5
1.1 General........................................................................................................................................ 5
1.2 Dismantling the scanner unit.................................................................................................... 5
1.3 Assembling the scanner unit .................................................................................................... 5
2 Calibration ............................................................................................................. 7
2.1 Safety .......................................................................................................................................... 7
2.2 Introduction ................................................................................................................................ 7
2.3 Measuring in the scanner.......................................................................................................... 7
2.4 Verification of scanner in RCS system .................................................................................. 10
3 Troubleshooting.................................................................................................. 13
3.1 Safety ........................................................................................................................................ 13
3.2 Troubleshooting the scanner unit .......................................................................................... 13
3.2.1 Visual inspection..................................................................................................................... 13
3.2.2 Check distance sensor accuracy............................................................................................ 13
3.2.3 Checking main axis zero ........................................................................................................ 13
3.2.4 Checking scanner axis zero ................................................................................................... 14
3.3 Troubleshooting the control lever.......................................................................................... 15
3.3.1 Checking the control lever in menu S01.2.1........................................................................... 15
3.4 Troubleshooting the touchscreen/computer......................................................................... 16
3.5 Self-test of scanner.................................................................................................................. 16
iv No: 8663980683.1 en
Epiroc Tunnel Profiler 1 Assembly and dismantling of scanner unit
1.1 General
It is always best to dismantle the scanner unit from Epiroc Tunnel Profiler when transport-
ing the rig.
2. Screw eye bolts into the threaded holes in the crossmember (1).
3. Fasten a suitable lifting device to the eye bolts and lift the scanner unit and arm.
5 No: 8663980683.1 en
Epiroc Tunnel Profiler 1 Assembly and dismantling of scanner unit
6 No: 8663980683.1 en
Epiroc Tunnel Profiler 2 Calibration
2 Calibration
2.1 Safety
WARNING
Serious injury
Laser beam.
u Danger of injury to the eyes.
u Do not look into the beam.
u Do not put optical instruments in the way of the beam.
2.2 Introduction
This chapter covers the following control and calibration measures for Epiroc Tunnel Pro-
filer:
• Measuring in the scanner
• Verifying the scanner precision
Both measures can be performed independently of each other.
! NOTE: As this calibration uses the rig laser it must be repeated if rig laser position or
direction is changed.
Required space
In order to measure in the scanner, at least 16-20 m of free space is needed in front of the
rig.
Precondition p Equipment
p The following equipment will be needed for measuring in:
2. Start the RCS system and then the jumbo laser by checking the Light up laser box in
the D2 Settings menu.
7 No: 8663980683.1 en
Epiroc Tunnel Profiler 2 Calibration
Menu D2 Settings
3. Place one of the profiler's calibration plates 16-20 m in front of the jumbo laser. Adjust
the plate so that the rig laser lights as close to the middle of the plate as possible. In
order for the scanner to be able to measure correctly, the point of contact must be at
least 10 cm from the edge of the plate.
4. Mark where the rig laser hits the plate with a pen.
5. Fasten the profiler's second calibration plate onto the rig's calibration plate and mount
the two plates in the feeder at least 8 m from the jumbo laser.
6. Adjust the boom and feeder so that the jumbo laser hits the middle plate (2) while the
plate is in line with the scanner as illustrated below.
1 Scanner
2 Plate
3 Plate
4 Jumbo laser
5 Distance at least 8 m
8 No: 8663980683.1 en
Epiroc Tunnel Profiler 2 Calibration
6 Distance at least 16 m
7. Set the distance sensor so that it measures from the back edge and hold it against
the jumbo laser. Measure the distance to the plate (2).
10. Enter the distance in box F5 Distance sight plate (m) and press Apply.
NOTE: If the rig is version 3.4 (not 3.4.1 or later) the distance from
! the laser distance meter must be reduced by 2 x the value X in
Jumbo calib. laser before it is entered.
9 No: 8663980683.1 en
Epiroc Tunnel Profiler 2 Calibration
11. Using the control lever, aim the profiler's laser at the mark where the jumbo laser hits
the plate (2). Check the display on the control lever that its distance seems plausible.
Then press F6 Accept bwd.ref..
12. Using the control lever, aim the profiler's laser at the jumbo laser's point of contact on
the plate (3). Check the display on the control lever that its distance seems plausible.
Then press F7 Accept fwd.ref..
13. After measuring both plates, a new box will open that recalculates the new values. Al-
low it to finish and do not press OK or cancel.
14. Once the New transform calculated: Ok message is shown, press OK.
15. Check the values XP(x,y,z) in the Current navigation box. They should correspond
to the scanner position according to the list below.
## UPPST [START] ##
x is the longitudinal distance from the scanner to the rig origin.
y is the transverse distance from the scanner to the jumbo laser.
z is the horizontal distance from the scanner to the jumbo laser.
## UPPST [END] ##
2. Start the RCS system and then the jumbo laser by checking the Light up laser box in
the D2 Settings menu.
10 No: 8663980683.1 en
Epiroc Tunnel Profiler 2 Calibration
Menu D2 Settings
4. Set the laser distance meter so that it measures from the back edge and hold it
against the jumbo laser. Measure the distance to the point where the jumbo laser
meets the wall as illustrated below.
1 Scanner
2 Jumbo laser
5. Go to menu M9.2.2 Profiler Measure and enter the reading from the laser distance
meter into the Laser depth box.
11 No: 8663980683.1 en
Epiroc Tunnel Profiler 2 Calibration
NOTE: The entered value will differ slightly from what the RCS sys-
! tem shows as Actuated value as the rig compensates for the dis-
tance between the boom plate and the front of the jumbo laser.
7. The profiler laser is now directed to the same point as the jumbo laser as illustrated
below.
1 Scanner laser
2 Jumbo laser
! NOTE: If the laser points do not meet each other, the profiler must
be recalibrated.
12 No: 8663980683.1 en
Epiroc Tunnel Profiler 3 Troubleshooting
3 Troubleshooting
3.1 Safety
WARNING
Serious injury
Laser beam.
u Danger of injury to the eyes.
u Do not look into the beam.
u Do not put optical instruments in the way of the beam.
2. With a measuring tape or other distance meter, measure the distance from the centre
of the scanner axis to the laser point of contact.
5. The distance sensor reading must not differ more than 2 cm from that measured
manually.
If the values for both or one of the checked distances differs more than 2 cm, get in
touch will Epiroc concerning service of the unit.
13 No: 8663980683.1 en
Epiroc Tunnel Profiler 3 Troubleshooting
Menu S01.2.1
3. Hold a spirit level on the top of the main cabinet and make sure the Inclination value
is correct.
4. Make sure the Main axis value is within the interval 86.5º - 93.5º.
5. If the value is outside the interval 86.5º - 93.5º there may be a fault in the scanner
unit. Contact Epiroc concerning service of the unit.
14 No: 8663980683.1 en
Epiroc Tunnel Profiler 3 Troubleshooting
Menu S01.2.1
2. Rotate the scanner housing so that its side is parallel to the side of the main cabinet,
i.e. the laser is pointing parallel with the direction of the main axis.
Angle Skanner axis should now be 0º.
3. If Skanner axis is greater than 1º, contact Epiroc concerning service of the unit.
! NOTE: Remember to wait for the scanner start-up time, approximately one minute.
15 No: 8663980683.1 en
Epiroc Tunnel Profiler 3 Troubleshooting
The left-hand box shows the axis value, -1 is at full deflection down and 1 is at full deflec-
tion up.
The right-hand box shows the raw value of the control lever.
• Max speed (%)
This should normally be set to 100%, which means the scanner runs at full speed at full de-
flection.
• Min speed (%)
Is normally set to 30%. This means the scanner speed is limited to max 30% of full speed.
This is the lower speed that is activated with F41 or the button on top of the control lever.
16 No: 8663980683.1 en
Epiroc Tunnel Profiler 3 Troubleshooting
Menu S01.2.2
17 No: 8663980683.1 en
Epiroc Tunnel Profiler 3 Troubleshooting
18 No: 8663980683.1 en
3 Troubleshooting
19 No: 8663980683.1 en
2018-09-18 | No: 8663980683.1
www.epiroc.com