CALIBRATION CERTIFICATE

Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-001


Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: MEDX TECHNOLOGIES
Model: SP 100
Range: 0 to 60 ml Sr. No.: 5137160917A01546
Least Count: 0.1 ml I.D. No.: GH/SYP/315/16
Condition of Instrument: Ok Location: First Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-002
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: ATHARVA
Model: N/A
Range: 0 to 60 ml Sr. No.: 14106057
Least Count: 0.1 ml I.D. No.: GH/SYP/315/17
Condition of Instrument: Ok Location: First Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
9. Total liability of our Organization is limited to the invoiced amount.
10. Calibration Certificate without signature are not valid.
11. All our certificates are metrological traceable.
12. The results relate only to the item calibrated.
13. UUC: Unit Under Calibration.
14. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-03
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: AKAS MEDICAL EQUIPMENT
Model: AKAS TERUMAX RX
Range: 0 to 60 ml Sr. No.: RZE1052222859
Least Count: 0.1 ml I.D. No.: GH/SYP/MICU/06
Condition of Instrument: Ok Location: Ground Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
15. Total liability of our Organization is limited to the invoiced amount.
16. Calibration Certificate without signature are not valid.
17. All our certificates are metrological traceable.
18. The results relate only to the item calibrated.
19. UUC: Unit Under Calibration.
20. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-004
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: BPL MEDICAL TECHNOLOGIES
Model: ACURA-S1
Range: 0 to 60 ml Sr. No.: KLTA3H18427
Least Count: 0.1 ml I.D. No.: GH/SYP/MICU/07
Condition of Instrument: Ok Location: Ground Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
21. Total liability of our Organization is limited to the invoiced amount.
22. Calibration Certificate without signature are not valid.
23. All our certificates are metrological traceable.
24. The results relate only to the item calibrated.
25. UUC: Unit Under Calibration.
26. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-005
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: AKAS MEDICAL EQUIPMENT
Model: AKAS TERUMAX RX
Range: 0 to 60 ml Sr. No.: RZE10522C3456
Least Count: 0.1 ml I.D. No.: GH/SYP/MICU/08
Condition of Instrument: Ok Location: Ground Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
27. Total liability of our Organization is limited to the invoiced amount.
28. Calibration Certificate without signature are not valid.
29. All our certificates are metrological traceable.
30. The results relate only to the item calibrated.
31. UUC: Unit Under Calibration.
32. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-006
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: AKAS MEDICAL EQUIPMENT
Model: AKAS TERUMAX RX
Range: 0 to 60 ml Sr. No.: RZE1052223054
Least Count: 0.1 ml I.D. No.: GH/SYP/MICU/09
Condition of Instrument: Ok Location: Ground Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
33. Total liability of our Organization is limited to the invoiced amount.
34. Calibration Certificate without signature are not valid.
35. All our certificates are metrological traceable.
36. The results relate only to the item calibrated.
37. UUC: Unit Under Calibration.
38. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-007
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: MEDX TECHONOLOGIES
Model: SP-100
Range: 0 to 60 ml Sr. No.: 5137160917A01582
Least Count: 0.1 ml I.D. No.: GH/SYP/MICU/10
Condition of Instrument: Ok Location: Ground Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
39. Total liability of our Organization is limited to the invoiced amount.
40. Calibration Certificate without signature are not valid.
41. All our certificates are metrological traceable.
42. The results relate only to the item calibrated.
43. UUC: Unit Under Calibration.
44. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-008
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: MEDX TECHONOLOGIES
Model: SP-100
Range: 0 to 60 ml Sr. No.: 5137160917A01530
Least Count: 0.1 ml I.D. No.: GH/SYP/MICU/11
Condition of Instrument: Ok Location: Ground Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
45. Total liability of our Organization is limited to the invoiced amount.
46. Calibration Certificate without signature are not valid.
47. All our certificates are metrological traceable.
48. The results relate only to the item calibrated.
49. UUC: Unit Under Calibration.
50. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-009
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: MEDX TECHONOLOGIES
Model: SP-100
Range: 0 to 60 ml Sr. No.: 5137160917A01531
Least Count: 0.1 ml I.D. No.: GH/SYP/MICU/12
Condition of Instrument: Ok Location: Ground Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
51. Total liability of our Organization is limited to the invoiced amount.
52. Calibration Certificate without signature are not valid.
53. All our certificates are metrological traceable.
54. The results relate only to the item calibrated.
55. UUC: Unit Under Calibration.
56. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-010
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: MEDX TECHONOLOGIES
Model: SP-100
Range: 0 to 60 ml Sr. No.: 5137160917A01538
Least Count: 0.1 ml I.D. No.: GH/SYP/MICU/13
Condition of Instrument: Ok Location: Ground Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
57. Total liability of our Organization is limited to the invoiced amount.
58. Calibration Certificate without signature are not valid.
59. All our certificates are metrological traceable.
60. The results relate only to the item calibrated.
61. UUC: Unit Under Calibration.
62. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.

CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-011
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 24-02-2024
Date of Calibration: 23-02-2024
Next Due Date (Suggested by Customer): 22-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: ATHARVA
Model: N/A
Range: 0 to 60 ml Sr. No.: 14106054Z
Least Count: 0.1 ml I.D. No.: GH/SYP/NICU/14
Condition of Instrument: Ok Location: First Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
63. Total liability of our Organization is limited to the invoiced amount.
64. Calibration Certificate without signature are not valid.
65. All our certificates are metrological traceable.
66. The results relate only to the item calibrated.
67. UUC: Unit Under Calibration.
68. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.

CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-012
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: MEDX TECHONOLOGIES
Model: SP-100
Range: 0 to 60 ml Sr. No.: 513760917A01531
Least Count: 0.1 ml I.D. No.: GH/SYP/CL/18
Condition of Instrument: Ok Location: First Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
69. Total liability of our Organization is limited to the invoiced amount.
70. Calibration Certificate without signature are not valid.
71. All our certificates are metrological traceable.
72. The results relate only to the item calibrated.
73. UUC: Unit Under Calibration.
74. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.

CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLKH20240213-013
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Lab
M/s. Gandhi Hospital Date of Receipt: 21-02-2024
(A Unit of Pawan Gandhi Healthcare Pvt.Ltd.) Service Request No.: SCCPLKHSRN20240213-001
C-50 & 51, Om Vihar, Uttam Nagar, New Delhi-110059 Certificate Issue Date: 23-02-2024
Date of Calibration: 22-02-2024
Next Due Date (Suggested by Customer): 21-02-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Syringe Pump Make: MEDX TECHONOLOGIES
Model: SP-100
Range: 0 to 60 ml Sr. No.: N/A
Least Count: 0.1 ml I.D. No.: GH/SYP/CL/19
Condition of Instrument: Ok Location: First Floor
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Weighing 0 to 220 g & SCCPL SCCPLCC20230113-002 14-01-2025
Balance SCCPL/CAL/005
2 Digital Stop Watch 0 to 24 Hrs. & ATSPL ATSPL24000000030 27-01-2025
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 022 & --------
SCCPL 03 SOP MIS 008
CALIBRATION RESULTS
Sr. No. UUC Value (ml/hr) Ref. Standard Value Error Expanded Uncertainty
(ml/hr) (ml/hr) (±) % rdg.
1 10 10.2 -0.2
2 20 20.2 -0.2
3 30 30.2 -0.2 3
4 40 40.3 -0.3
5 50 50.3 -0.3
6 60 60.4 -0.4
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
75. Total liability of our Organization is limited to the invoiced amount.
76. Calibration Certificate without signature are not valid.
77. All our certificates are metrological traceable.
78. The results relate only to the item calibrated.
79. UUC: Unit Under Calibration.
80. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.

You might also like