Professional Documents
Culture Documents
IFU - PM Philips Goldway GS20 - Revisi A.1 August 2019
IFU - PM Philips Goldway GS20 - Revisi A.1 August 2019
Lorem ipsum
GS20
Patient Monitor
Release A.01
English
English
Product Information
Product Model: GS20
Product Name: Patient Monitor
Manufacturer
Philips Goldway (Shenzhen) Industrial Inc.
Registered/Production address: No.2 Keji North 3rd Road, Nanshan District, Shenzhen, P.R.
China 518057
Tel: +86 755 26980999
Fax: +86 755 26980222
Edition
Revision A
August 2019
Philips Goldway (Shenzhen) Industrial Inc.
All rights reserved.
Explanation of Symbols
The following symbols appear on the monitor and its packaging.
Table 1 Monitor and Packaging Symbols
Caution, consult
Temperature limitation accompanying
documents
Atmospheric pressure
Keep dry
limitations
ii GS20 Patient Monitor Instructions for Use
NBP cuff hose
Main Menu key NBP
connector
SpO2 SpO2 probe connector ECG/RESP ECG/RESP connector
TEMP1 TEMP1 connector TEMP2 TEMP2 connector
Refer to instruction
Dangerous Voltage
manual/booklet
ECG analog output
Protective grade
port
Protective grounding Ethernet port
Safety Standards
The following table describes the safety standards of the GS20 Patient Monitor.
Table 2 Safety Standards
Parameter Specification
Class I, anti-shock, externally and internally
Protection class
powered equipment.
Degree of protection Type CF applied part
Mode of operation Continuous
Degree of noxious-liquid proof IPX1
Degree of safety of application in the presence The equipment is not suitable for use in the
of a flammable anesthetic mixture with air or presence of a flammable anesthetic mixture with
with oxygen or nitrous oxide air or with oxygen or nitrous oxide.
Sterilization and disinfection As recommended by manufacturer.
Management class Class II
After-Sales Service
When you need product support, contact Philips Customer Care Center or your local representative.
Before calling for service, note the following information:
Model and serial number of the monitor
Monitor problem
Conventions
This section describes the conventions used in this guide.
Text Formatting
The following typographical conventions are used in this guide:
Decimal Points
Because the GS20 monitors use a period (.) to indicate a decimal point in decimal numbers (for example,
10.0), all decimal numbers in this manual use a period as a decimal point. Commas are not used as a decimal
point.
Notes, Cautions, and Warnings
The manual uses the following conventions for Notes, Cautions, and Warnings.
Note — A Note calls attention to an important point in the text.
Warning A Warning calls attention to a condition or possible situation that could cause
injury to the user and/or patient.
Safety Requirements
This monitor is suitable for a case in which electro-surgery unit is used. Properly connect the electro-surgery
circuit to prevent burns and life risk therefrom.
When using in conjunction with high-frequency electro-surgery unit, the monitor restores to the previous
operation mode within 10s after the elimination of high-frequency signal and electromagnetic field, without
losing any permanently stored data.
For safe grounding of the monitor, the hospital must provide live wire, zero-line, and fully protected power
socket. If any external protective grounding conductor is in doubt during installation or wiring, the monitor
must be powered by an internal battery.
Ensure that the patient monitor is in good working condition before use. If there is a problem with the
accuracy of any measurement item, use alternative method to check the patient's vital signs first, and then
check the monitor with the results therefrom to see if the monitor is operating properly.
Equipment connected to the monitor through signal input/output port should comply with the corresponding
standards (for example, data processing equipment should comply with GB4943 and medical devices with
GB9706.1). Do not connect any permanently installed equipment to the signal input/output ports of the
monitor, please consult the manufacturer for correct usage if you still have questions.
Other equipment connected to the monitor should comply with the corresponding standards (for example, data
processing equipment should comply with GB4943 and medical devices with GB9706.1). The entire system
should comply with the latest and valid standards in GB9706.15. Do not connect components not specified in
this manual to the system.
If multiple devices are connected to a patient, the sum of the leakage currents may cause a dangerous situation.
Therefore, qualified service representatives should always perform current leakage tests before setting up the
connections. This ensures that leakage is within limits and prevents personal injury and environmental damage.
If any question remains, consult the manufacturer for correct usage.
Warning The monitor is not for home use. It is expected to be used exclusively by
healthcare professionals. The monitor may cause wireless interference or disrupt
the operation of nearby equipment. Mitigation measures may be necessary, such
as reorienting or relocating the monitor, or shielding the corresponding site.
The monitor is only for use on one patient at a time.
The monitor is not an apnea monitor.
This monitor is not suitable for use in the presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide. In the absence of other oxidants,
the oxygen concentration must be less than 25% and the partial oxygen pressure
must be less than 27.5 kPa.
Never use the monitor and the SpO2 probe during MRIs or CT scans, to avoid burns
to the patient. The monitor may affect the MRI image, and the MRI unit may affect
the accuracy of the monitor’s measurements.
Do not connect the monitor to any other equipment or device not clearly listed in
this manual.
Do not open the monitor case, to prevent electric shock. The device does not
contain user-serviceable parts.
Use the power modules and application parts manufactured or recommended by
the Company. Otherwise, personal injury and machine failure can occur.
To avoid personal injury, use only accessories and parts produced or
recommended by the company. Otherwise, damage to the patient, operator or
monitor can occur.
GS20 Patient Monitor Instructions for Use v
The monitor is resistant to electromagnetic interference. Interference takes place
when electromagnetic energy is extremely high. Ensure that nearby instruments
are also in compliance with EMC standards. Never turn on or use portable
communication devices like mobile phones or portable dual-channel radios near a
monitor. If there is an unknown EMI, check the location of the monitor to
determine the source of interference and take measures for elimination.
To reduce the hazard of burns in high-frequency surgical neutral connections, the
leads and connectors must be located away from the surgical site. Keep cables
away from other devices.
Never rely exclusively on the alarm system for patient monitoring. You must
periodically check that monitor alarms are working properly. The most reliable
method of patient monitoring combines close, personal surveillance along with
the correct operation of the monitor.
Never dispose of this product (or any part thereof) as industrial or domestic waste.
The system may contain hazardous substances that cause serious pollution to the
environment. The system also contains private information. Philips recommends
that you contact Philips Service Organization before disposal.
To avoid contamination or infection with personnel, the environment, or other
equipment, please disinfect and clean the monitor before disposal in accordance
with the laws of the country/region where you locate.
Caution Immediately stop using the monitor if it becomes wet by chance, and ask a
qualified professional to conduct safety test. Use the monitor again unless it
passes the safety test.
Never simultaneously touch the patient and the input or output of the monitor.
Carefully place power cords and other cables to prevent patient contact, tangling
or electrical interference.
Periodically check all reusable accessories for damage. Replace damaged
accessories when necessary.
Measurement accuracy may decrease temporarily while performing
electro-surgery or defibrillation. This does not affect patient or equipment safety.
The monitor restores to normal measurement within 10s after the electro-surgery
or defibrillation is completed.
Clean and sterilize the monitor and accessories according to Philips requirements.
Turn off the monitor and disconnect the power cords before cleaning or
sterilization.
It is not recommended to sterilize the monitor and its accessories and
consumables, unless it is specified in the Instructions for Use of such
accessories and consumables.
The monitor should be calibrated and maintained regularly by a qualified
professional.
Keep all monitor packing materials away from children, or dispose of them in
accordance with your local environmental regulations.
Ensure that no water condenses into or on the monitor. Condensation can occur
from changes in temperature or exposure to humidity.
The device and accessories are to be disposed of according to local regulations
after their service life. Dispose of batteries according to your local regulations.
Never incinerate batteries or expose to high temperatures. If in doubt, please
contact Philips headquarter or local office.
Chapter 1 Overview
The GS20 Patient Monitor (hereinafter referred to as the “Monitor”) automatically recognizes the
physiological monitoring units and monitors the physiological parameters of patients. You can display,
review, store and print the measured data.
1.3 Contraindications
NBP measurement should not be performed on patients with sickle-cell disease, or any condition
where skin damage has occurred or is expected to occur. When measuring IBP,
Avoid subclavian puncture on patients with severe coagulation disorder.
Choose a different puncture site for patients with local skin infection.
Avoid internal jugular and subclavian puncture on patients with hemopneumothorax.
Alarm Pause
6 Please refer to 3.9.2 “Enabling or Disabling Alarms”.
key
Item
Icon Meaning Description
Number
Press to freeze the waveforms when monitoring a
Waveform patient.
7 Freeze key
Press again to unfreeze the waveforms.
Main Menu
10 Press to display or exit the Main Menu.
key
Navigation
11 None Used for selecting on the screen.
wheel
Battery Compartment
Door Latch
Figure 1-5 Bottom View
Warning Never attempt to open the monitor case. Only qualified personnel should service
the monitor.
2.4.1 Powered by AC
To connect to AC power supply:
1. Plug the power cord to the power socket on the back of the monitor. Push it into position and
lower the power cord lock catch to secure it.
2. Make sure power socket is properly grounded and provides specified voltage and frequency.
When the power LED on front panel illuminates green, it indicates that AC power is
connected.
When the monitor is powered by AC, the battery status pane on the screen shows it’s
powered by AC. When AC is disconnected, the monitor automatically switches to be powered by
battery.
Note — To ensure that the monitor can be powered by the battery at any time when the AC
power is disconnected, we recommend you to keep the monitor plugged in to AC power
when it is not in use.
The battery pane on the screen indicates battery status. More bars in the pane indicate more
charge remains on the battery.
When the battery power is low, the monitor gives battery alarm:
When there are only two bars in the battery pane, the monitor gives low power alarm. The
battery pane flashes, indicating that the battery power is low. If you don’t charge it within 20
minutes, the monitor will shut down automatically.
When there is only one bar in the battery pane, the monitor gives extreme low power alarm.
The battery pane turns red and flashes, indicating that the battery power is very low. If you
don’t charge it within 5 minutes, the monitor will shut down automatically.
Any time the monitor is connected to AC power, the battery is being charged, and no extra adapter
is needed. The charging LED on front panel shows the status of the current charge.
The charging LED shows:
Off, when no battery is installed.
Yellow, when the battery is being charged.
Green, when the battery is fully charged.
Warning Do not throw the battery into fire or short it. Burning, exploding or leaking
batteries can cause personal injury.
Note — If you do not connect any patient or perform any operation to the monitor for 30
minutes, a message appears on the bottom of the screen, indicating that no monitoring is
going on, you should press the On/Standby key to enter Standby mode.
When you press the On/Standby key in the Standby mode, the monitor enters On mode. The
screen displays Patient Information window. You can start monitoring a patient. See 2.11.1
“Admitting a New Patient”.
Shows the current system date and time, can also be used as
5 Date/Time pane
timer.
9 Tool bar Shows icons for quick operation, see Table 2-5.
The following table provides a brief description of each button in the tool bar.
Table 2-2 Buttons in the Tool bar
The screen display depends on the configuration you choose, refer to 2.7 “Changing Screen
Layout”.
2. Select the Other Bed Monitoring window to open the Other Bed View Setup menu.
3. Select Monitor Other Bed ID, choose the ID of the bed that you want to view.
4. Close the Other Bed View Setup menu.
The following illustration shows data of another bed in the Other Bed View window.
You can change the waveforms of the other bed shown in the Other Bed View window.
To change the displayed the other bed waveforms:
1. Open the Other Bed View window.
2. Select the Other Bed View window to open the Other Bed View Setup menu.
3. Select Channel 2 to display the contents of channel 2.
4. Select what you want to display in channel 2.
5. Close the Other Bed View Setup menu.
Besides viewing another bed on your monitor, you can control whether other beds can view the
data on your monitor.
To control Other Bed View monitoring of your monitor:
1. Open the Other Bed View window.
2. Select the Other Bed View window to open the Other Bed View Setup menu.
The Monitor displays waveforms in single and/or bold lines. By default, waveforms are in a single
line. You can choose to display a waveform as a bold line.
To enable or disable enhanced waveform:
1. Open the Configure tab.
2. Select the Enhanced Waveform check box to select the desired setting:
– Check Enhanced Waveform, the waveforms are drawn in a bold line.
– Uncheck Enhanced Waveform, the waveforms are drawn in a single line.
3. Select OK. The settings take effect right away.
4. Close the menu.
In the dynamic trend view, the monitor displays the most recent trends (1h, 2h or 4h).The
parameter values are plotted vertically along the y-axis of the graphical display. You can set a time
range in the Dyn Trend Time menu and that range appears along the horizontal x-axis. The most
recent measurements appear on the right side of the graph and the data scrolls to the left, so the
oldest data appears on the far left side of the graph.
To enable or disable the dynamic trend:
1. Open the Configure tab.
2. Select the Dyn Trend On/Off check box to select the desired setting.
– Check Dyn Trend, dynamic trend is enabled.
– Uncheck Dyn Trend, dynamic trend is disabled.
3. If you want to change the time range of the dynamic trend, select one of the options:
– 1 hour
– 2 hours
– 4 hours
The monitor is preloaded with 4 user customized displays in its System (password-protected). You
can load one of them.
To load a user display:
1. Open the Configure tab.
2. Select Load User Display, select one of the options:
– Layout 1
– Layout 2
– Layout 3
– Layout 4
3. The new layout is loaded.
Reducing the waveform speed compresses the waveform display so that waveform segments
display for longer periods of time.
To change the waveform speed:
1. Select the waveform channel that you want to change the speed to open the setup menu for
this parameter.
2. Select Waveform Speed to open the Waveform Speed menu.
3. Select a waveform speed. Select OK to confirm.
4. Close the setup menu.
Select the Vol/Light button in the tool bar to open the Vol/Light menu.
In the Vol/Light menu, you can:
Adjust alarm tone volume
Adjust heart rate volume
Adjust screen brightness
Enter night mode
Enable or disable key sound
Set up HR sound/PR sound priority
2. Select Alarm Volume and then choose a volume. The maximum volume you can choose is 7.
By default the minimum volume is 1.
3. Select OK to confirm.
4. Close the menu.
Note — To avoid misoperation, you cannot enable Night Mode when an alarm is active.
To enable night mode:
1. Open the Vol/Light menu.
2. Select the Night Mode check box to select the desired setting:
– Check Night Mode to enter Night Mode.
– Uncheck Night Mode to exit Night Mode.
The monitor exits Night Mode automatically when
You press any key one the front panel (except the On/Standby key)
An alarm event occurs.
Adult
Pediatric (Pedi.)
Neonatal (Neo.)
Once you input a new patient ID or change the patient type, the alarm settings are restored to the
default values for the specified patient type, and all trend data will be cleared.
Select the Patient pane on the screen to open the Patient Setup menu. Use the Patient Setup
menu to:
Admit a patient
Edit a patient
Discharge a patient
Note — If you change the pace detection setting when editing a patient, the ECG algorithm
will relearn.
Select the System button in the tool bar to open the System menu.
Note — Please contact your system administrator if you need to use the User
Configuration (password protected) in the System menu. The User Maintenance
(password protected) in the System menu is not intended for use by the user.
To check the software version:
1. Open the System menu.
2. Select About..., the software version is shown.
3. Close the menu.
Chapter 3 Alarms
Alarms sound when the monitor senses an irregular condition either in monitor operation or alarm
limits.
Alarms include:
Visual alarms
Audible alarms
Alarm
Description
Icon
Audio Pause icon (white icon with black background), indicating that audio
signal of the alarm is suspended for 60 seconds.
Alarm Pause icon (red icon with white background), indicating that the alarm
of the monitor is suspended for a specified period.
Alarm Off icon (white icon with red background), the alarm icon with solid
lines indicates that this alarm is disabled.
The auditory alarm signal sound pressure is in the range from 45dB to 85dB. The sound pressure
for medium priority alarms is the highest and that for the low priority alarms lowest. If alarms of
more than one priority are triggered, the monitor sounds the most severe.
The reminder signal sounds 1 set of 2 beeps every 60 seconds at a low pitch, in order to be
distinguishable from those of the alarm signals.
You can change:
Alarm volume - increase or decrease the alarm volume. Refer to 2.9.1 “Adjusting Alarm Tone
Volume”.
Alarm tones - the monitor offers two sets of alarm tones. Only your administrator can change
the alarm tone (password protected).
Suspend alarm tones - refer to 3.6 “Suspending Audible Alarms”.
Note — Do not rely exclusively on the audible alarm system for patient monitoring.
Changing the alarm volume to a low level or turning it off during patient monitoring could
result in patient danger. The most reliable method of patient monitoring combines close
personal surveillance with correct operation of monitoring equipment.
Set the alarm volume based on the environment and ambient noise levels. Alarm volume
should be tested and verified during installation. The alarm volume should be loud enough
to be heard in the intended environment.
01:51
Displays alarm off icon and the text “OFF” in the alarm icon pane, and gives the reminder
sound if the alarm off reminder is on, if alarms are disabled.
OFF
Note — The alarm off reminder is configurable in the password-protected User
Maintenance menu.
Disables audible alarm and alarm indicators, but still displays alarm messages for all technical
alarms.
Disables all alarm signals (both audible and visible) for physiological alarms, does not save
the alarm messages or transfer them to central monitoring system or nurse call system.
Does not break alarm pause or alarm off even when a new alarm event occurs.
If an alarm is still active when the alarm pause or alarm off ends, the monitor gives normal alarm
indicators.
Note — When the one or more alarm sounds are disabled for a parameter, the icon “ ”
appears in its numeric pane.
5. If you want to apply the auto set alarm limits for this parameter, select Auto Alarm Limit.
Select OK to apply the auto alarm limits, select Cancel to cancel the operation.
6. Close the menu.
Note — Alarm limits remain in the monitor until you change them next time, even after shut
down or a power loss.
2. Select Default Alarm Setup. A message asks if you are sure to restore default alarm limits.
Select OK to confirm, or select Cancel to cancel the operation.
3. Close the menu.
For more information on default alarm limits, see A.3 “Default Alarm Limits”.
Note — Both poor quality and incorrect attachment of the electrodes will cause signal loss
or noise increasing.
To connect ECG cables:
1. Attach the clips or snaps to the electrodes before placing them. If you are not using pregelled
electrodes, apply electrode gel to the electrodes before placement.
2. Place the electrodes on the patient according to the selected lead placement. Refer to 4.3.3
“Placing ECG Leads”.
3. Plug the patient cable into the ECG connector on the monitor. ECG waveforms and value are
shown on the screen.
AAMI IEC
Lead Color Lead Color
RA White R Red
LA Black L Yellow
LL Red F Green
RL Green N Black
V Brown C White
AAMI IEC
The following figure shows the 5-lead ECG electrode locations:
AAMI IEC
Once pace pulse detection is enabled, a pace detection icon " " appears in the ECG waveform
channel on the screen. Pacemaker spike signal and pace pulse mark (a short yellow line) are also
shown in the ECG waveform, indicating the position where the pace pulse was detected.
If you disable pace pulse detection, the pace detect icon " " disappears.
Warning When monitoring patients with pacemakers, if pace pulse detection is not
disabled, the pace pulse will be counted as regular QRS complexes.
Some pace pulses are difficult to reject. When this happens, the pace pulse is
counted as a QRS complex, which can result in an incorrect heart rate count and
failure to detect cardiac arrest. Healthcare professionals should closely observe
patients with the pacemaker.
– ×2
– ×4
– Auto
Once the gain is changed, ECG waveform size changes accordingly.
Note — If you set the gain to X4, the 1mV calibration bar disappears. If you set the gain to
Auto, the monitor chooses the best gain for the current waveform.
4. Close the menu.
Warning Make sure both the monitor and the defibrillator are working safely and
effectively before using them for cardiac synchronization.
Note — Your administrator can enable or disable ECG analog signal output
(password-protected). You can enable or disable ECG analog signal output only when your
administrator enables it.
To enable or disable analog signal output:
determination of the RESP value and execute pacing with the maximum
programmed rate. Switching off the RESP measurement can prevent this.
During complete heart block or pacemaker failure (to pace or capture), tall
P-waves (greater than 1/5 of the average R-wave height) may be erroneously
counted by the arrhythmia algorithm, resulting in missed detection of cardiac
arrest.
When arrhythmia monitoring paced patients who exhibit only intrinsic rhythm,
the monitor may erroneously count pace pulses as QRS complexes when the
algorithm first encounters them, resulting in missed detection of cardiac arrest.
For patients who exhibit intrinsic rhythm only, the risk of missing cardiac arrest
may be reduced by monitoring these patients with the low heart rate limit at or
slightly above the basic/demand pacemaker rate. A low heart rate alarm alerts
you when the patient begins pacing. Proper detection and classification of the
paced rhythm can then be determined.
Instruments such as defibrillators or telemetry units produce filtered ECG
signals. Such signal will be filtered again if used as the output signal of the
patient monitor. This may result in failure to detect pace pulse, non-capture or
cardiac arrest, causing damage to the monitoring performance on patients with
pacemaker.
Using the monitor in conjunction with a pacemaker or other electric generator
on the same patient or when the patient has arrhythmia may result in
mismeasurement of HR by the monitor. The result is that the monitor may
continue displaying ECG signals even after the patient's death. Healthcare
professionals should observe the patient closely to make correct diagnosis in
time.
When an external pacemaker is being used on a patient, arrhythmia monitoring
is severely compromised due to the high energy level in the pacer pulse. This
may result in the arrhythmia algorithm’s failure to detect pacemaker
non-capture or asystole.
Note — Do not completely immerse the ECG cable in water, solvents or cleaning solutions
as the connectors are not waterproof.
The ECG cable connected to the patient during defibrillation may be damaged. For cables
that have been attached to the patient during defibrillation, carefully check that they are
functioning properly before using them again.
We recommend that you use Ag-Agcl ECG electrodes recommended by the manufacturer
or with quality guarantee, and replace the ECG electrodes at least every 24 hours to
ensure good ECG signal quality during long term monitoring, regardless of the age of
patient monitored. The reason is that, after 24 hours of use, the gel in the electrode starts
to dry, with adhesive strength weakened, besides, after long term monitoring, the patient
will feel uncomfortable with sensitive skin stimulated by the gel. ECG electrodes are for
single use and should be disposed of properly as medical waste.
Enabling pacemaker detection at the same time does not affect the monitoring on patients
with pacemaker. However, it is best to disable this function when the patient does not wear
a pacemaker, to avoid that the monitor counts the artifacts in the waveform as pacemaker
signal.
Thorough skin cleansing is very important role for minimizing static interference. Before
using high-frequency surgical equipment, carefully clean the skin and properly place the
electrodes to reduce the risk of burns when the neutral electrode connection fails.
The algorithm expects a heart rate and therefore needs at least 3 electrodes attached to the
patient. If you are monitoring respiration with only two electrodes, the detection algorithm
becomes less sensitive which may result in reduced breath detection performance.
Use Auto mode for situations where:
The respiration rate is not close to the heart rate.
Breathing is spontaneous, with or without continuous positive airway pressure (CPAP).
Patients are ventilated, except patients with Intermittent Mandatory Ventilation (IMV).
Warning Position the sensor cable and connector away from power cables to avoid
electrical interference.
For neonatal patients, make sure that all sensor connectors and adapter cable
connectors are outside the incubator. The humid atmosphere inside can cause
inaccurate measurements.
Caution The sensor consists of sensitive electronic components which may be damaged
if misused. Keep sharp-edged objects away from the cable.
Improper cleaning of the sensor could compromise the SpO2 sensor life span.
Note — Use only the specified sensor and cables. Before using the sensor, you need to
verify the compatibility between the monitor and the probe and cables, otherwise, damage
to the patient may be caused.
The sensor may cause skin irritation or rupture after prolonged application. Inspect the
application site for skin behavior and optical alignment every 2-3 hours. If the skin texture
changes, moves the sensor to another site. Change the application site at least every 4
hours.
Do not apply the sensor too tight, which can cause venous pulse. This could result in
circulation block and inaccurate measurement.
Follow with the Instructions for Use of the sensor and all Warnings and Notes therein.
Ensure that the light emitter and the photo-detector are directly opposite each other. All
light from the emitter must pass through the patient’s tissue.
On an edematous extremity.
In the process of CPR (Cardiopulmonary Resuscitation).
If the NBP measurement interval is set to a value, the NBP measurement icon " " in the NBP
numeric pane flashes to remind you to start NBP measurement. Press the NBP key on the front
panel to begin the first interval measurement. The Interval icon " " and the remaining time
before the next measurement starts appear in the NBP numeric pane. When the remaining time
comes down to 0, the next measurement starts automatically.
After the monitor enters Stat mode, the Stat measurement icon " stat" in the NBP numeric pane
flashes to remind you to start NBP measurement. Press the NBP key on the front panel to start Stat
measurements. The monitor takes as many measurements as possible within a 5-minute period. The
remaining time is shown after the Interval icon " ".
Press the NBP key on the front panel to stop Stat measurements.
Note — When using a skin surface probe, ensure good contact between the probe and the
patient skin. It is recommended to place the probe on the artery under the armpit, and ask
the patient to hold the arm tight to the body.
To select and connect the temperature probe:
1. Select a temperature probe with proper type and size for your patient. See Chapter 13
"Accessories" for compatible temperature probes.
2. If you are using a disposable probe, connect the probe to the temperature cable.
3. Plug the probe or the temperature cable into the temperature connector on the monitor.
4. Attach the probe on an appropriate part of the patient (e.g. mouth, rectum), apply a rubber
sheath on the probe.
5. Check if the alarm settings (on/off, alarm limits) are suitable for the patient and temperature
measurement.
6. Start the temperature measurement, see 8.3 “Enabling or Disabling Temperature Monitoring”.
7. Temperature numeric values appear on the monitor screen.
The following factors can affect temperature measurements:
Improper position of the temperature probe.
Poor heat balance, heat dissipation.
Patient temperature fluctuation.
When the probe is disconnected or loose, the monitor stops measuring the
temperature with no audible alarm. To ensure patient’s safety, check the probe
connection and position periodically.
Use disposable sheath to reduce the risk of cross-infection. Sheaths are
intended for single use and should be disposed of properly as medical waste
after use.
Do not use a probe if you see any signs of damage.
Select the Trend button in the tool bar to open the Trend menu.
Use the Trend menu to:
View ECG waveform
Analyze heart rate variation
View graphical trend
View tabular trend
View and analyze NBP history data
View alarm/event history
View “data management center”
Export patient data
The trend database can store up to 120 minutes of ECG waveform, 50 days of graphical trend data
and 1600 sets of NBP list, at least 200 alarm/event records (including ARR).When the memory is
full, the oldest data is deleted to make room for newer data.
The time of each ECG waveform that the monitor saves is no longer than 12 seconds.
Continuous measurements (for example, SpO2 and Respiration) are captured and stored in the trend
database every 18 seconds; aperiodic measurements (for example, NBP) are recorded at the time
that the measurement is completed.
If you want to print the trends, refer to Chapter 11 "Using the Recorder".
4. Select Position to choose the start of this ECG segment for analysis (background of the
Position turns pink).
5. Choose an end position (background of the Position button turns blue). The Time and
Amplitude under the ECG waveform indicate the relative time and relative amplitude
between the start and end of the ECG waveform.
6. Close the menu.
– 8 hours
– 24 hours
– 48 hours
– 72 hours
– 96 hours
– 120 hours
Select OK, the time scale is changed.
3. Close the menu.
4. Select End Time. Choose a time where you want to stop the dynamic monitoring.
The Dynamic Blood Pressure Analysis window shows the dynamic monitoring result for this time
scale.
5. Close the window.
Chapter 10 Calculation
Calculations are patient data that is not directly measured, but calculated by the monitor when
appropriate information is provided.
The monitor can perform the following calculations:
Select the Calculator button in the tool bar to open the Calculator menu.
2. Calculate the infusion rate, titration rate, and medication time, see 10.1.2 “Calculating
Infusion Rate, Titration Rate and Medication Time”.
3. Select Titration Chart. Titration chart information is shown.
4. If you want to change the displayed step, select Step and choose an option from 1 to 10.
5. If you want to view more titration chart information, select Page to view other pages.
6. Close the window.
3. Place a new roll in the holder so that the end comes from under the roll, extending beyond the
recorder door, as shown in Figure 11-2.
Select the Print Setup button in the tool bar to open the Print Setup menu.
Use the Print Setup menu to:
Change the timed printing interval.
Change the print speed.
Choose the real-time printing waveforms
Print signal-channel ECG waveform
– OFF
– 15 minutes
– 30 minutes
– 1 hour
– 4 hours
– 8 hours
– 12 hours
Set it to OFF to disable Timed Printing, or choose one interval time for the recorder to print the
real-time waveforms.
4. Close the menu.
Whether to print the waveform of the parameter that triggers the alarm depends on the Print Mode
setting. See 11.3.1 “Choosing Alarm Print Mode”.
To enable print on physiological alarm:
1. Open the Print Setup menu.
2. Select the Alarm Trigger check box to select the desired setting:
– Check Alarm Trigger, enable print on physiological alarm.
– Uncheck Alarm Trigger, disable print on alarm.
3. Close the menu.
– 2 seconds
– 3 seconds
4. Close the menu.
Warning If you spill liquid on the exterior of the monitor or accessories, use a clean cloth
to dry the monitor. If you believe the liquid may have gotten inside the monitor,
power off the monitor and contact your service representative.
12.2 Cleaning
It is critical to periodically clean the monitor and accessories according to the regulation of your
facility.
Caution Never heat the probe over 100˚C (212℉). Only heat to 80˚C (176℉) - 100˚C (212℉)
for a short time.
To clean the temperature probe:
1. Wipe the probe with a clean cloth dampened with one of the following cleaning agents:
– mild detergent
2. Wipe the probe with a water-dampened cloth to remove all cleaner residues.
12.3 Disinfection
Disinfect the monitor and accessories periodically as determined by hospital’s policy. Clean the
monitor and accessories before disinfection.
Note — Never mix disinfecting solutions (such as bleach and ammonia) as hazardous
gases may result.
3. Examine all cables, the power plug and cord for damage. Make sure that the prongs of the
plug do not move in the casing. If damaged, replace it with an appropriate one.
4. Inspect all accessories (cables, transducers, sensors and so on). If any show signs of damage,
do not use.
5. Inspect the patient cables, leads and their strain reliefs for general condition. Make sure there
are no breaks in the insulation. Make sure that the connectors are properly engaged at each
end to prevent rotation or other strain.
12.6 Maintenance
Under the conditions that meet the intended use of this product and an average daily use of 6 hours,
the life expectancy is 10 years. Internal parts may need to be repaired and replaced.
The following tasks are for qualified service professionals only.
Ensure that these tasks are carried out as indicated by the monitor’s maintenance schedule, or as
specified by local laws. Clean and disinfect equipment to decontaminate it before testing or
maintaining it.
Table 12-1 Maintenance
12.7 Troubleshooting
If you suspect a problem with an individual measurement, read the related section in this guide and
double check the measurement settings. If the problem remains, call your service personnel. If you
suspect an intermittent, system-wide problem, call your service personnel.
Use the following table to troubleshoot problems with the monitor and accessories. For more
information on troubleshooting the monitor, see GS20 Patient Monitor Service Guide. If needed,
the company also provides circuit diagram and part list.
Table 12-2 Troubleshooting
Problem Solution
Inaccurate HR Check/adjust the lead position.
measurements Check/re-clean the skin.
Check/replace ECG electrode
Problem Solution
The monitor does not show Ensure that the environment is dry and that the monitor is
the ECG baseline waveform not wet. If there is dampness in the environment or monitor,
consistently on the screen run the monitor continuously for 24 hours to dry it out.
and the waveform moves Check the quality of the electrode and whether the patient
out of the display area at site where the electrode is attached is clean. Replace the
times. electrode or clean the contact site, if necessary.
The lead wires are Ensure that the electrode is in good contact with the patient.
connected but there is no Check all the extension parts of the ECG lead wires. The
ECG waveform and the circuit between the plugs and pins of the 5 ECG extension
screen displays Leads Off. cables are open. Replace the lead wire if the circuit is open.
Check the ECG electrode and replace the electrode after
long-term use.
The SpO2 values turn on and In long term monitoring or operation, patient movement may
off during SpO2 monitoring. result in SpO2 interruptions. Keep the patient stabilized.
There is no SpO2 waveform Check whether the red light of the finger probe is flashing.
or values. If there is no red light flashing, there might be poor contact.
Check whether the environmental temperature is too low.
Do not expose the patient’s arm to cold air since this can
affect readings.
There is no temperature The TEMP sensor might be off the patient or the monitor. If the
displayed. problem continues, consult your service representative.
Ensure that the cuff is correctly applied, refer to 7.1
Cuff Loose displays in the “Selecting and Applying an NBP Cuff”.
NBP numeric pane. Check the cuff for damage. If any sign of damage shows,
replace it.
Overpressure displays in There is a problem with the inflation circuit inside the monitor,
the NBP numeric pane. consult your service representative.
Chapter 13 Accessories
Note — Refer to the packaging for the accessories’ manufacture dates.
Accompanying registered accessories are for one year use.
Do not reuse disposable accessories.
Philips Part
Item Description
Number
989803177341 3-Lead ECG Cable+Lead AAMI Snap style
989803177331 5-Lead ECG Cable+Lead AAMI Snap style
989803160641 3/5-Lead ECG Trunk Cable, AAMI/IEC
989803160651 3-Lead Grabber, AAMI
989803160661 3-Lead Grabber, IEC
ECG Cables 989803160691 5-Lead Grabber, AAMI
989803160701 5-Lead Grabber, IEC
989803143171 3-Leadset Grabber, General Use/ICU, IEC
989803143181 3-Leadset Grabber, General Use/ICU, AAMI
989803143191 5-Leadset Grabber, General Use/ICU, IEC
989803143201 5-Leadset Grabber, General Use/ICU, AAMI
ECG Electrode 40493D Adult Foam ECG Electrode
Philips Part
Item Description
Number
989803177501 NBP Cuff -Adult Thigh GCF1205
989803177531 Reusable NBP Cuff -Adult Large GCF1204
NBP Cuff 989803177521 NBP Cuff - Adult GCF1203
989803177511 NBP Cuff - Pediatric GCF1202
989803177481 NBP Cuff - Infant GCF1200
Blood Pressure
989803177471 Blood Pressure Interconnect Tube - Adult/Infant, 3m
Interconnect Tube
Philips Part
Item Description
Number
Esophageal/Rectal Temperature Probe (autoclave
21075A
Temperature sterilization), 12FR, reusable, continuous measuring, 3m
Probe Skin Surface Temperature Probe (autoclave sterilization),
21078A
9.5mm diameter, reusable, continuous measuring, 3m
13.5 Accessories
Table 13-5 Accessories
Philips Part
Item Description
Number
Li-Ion Battery 989803195041 Li-Ion Battery
989803172771 Roll stand
Roll stand
989803172761 Wall mount
Chapter 14 Specifications
14.1 ECG Specifications
Table 14-1 ECG Specifications
Parameter Specification
Type CF
Auxiliary current (Leads Off Active electrode: < 100nA
detection) Reference electrode: < 900nA
Excitation current parameter Current: < 250 µA
for RESP measurement Frequency: 37 kHz
The relative change in the amplitude of the displayed signal
Measured voltage deviation
caused by the application of ± 300 mV DC voltage does not
caused by polarization voltage
exceed ±5%.
Tall T-wave rejection ≥ 1.2 mV (Adult/Pediatric)
capability ≥ 0.6 mV (Neonate)
Aspect ratio 0.24 s/mV – 0.6 s/mV
Type of averaging:
In normal circumstance, the heart rate is higher than 50
bpm, the heart rate is acquired by averaging the 12 most
Type of averaging to compute
recent RR intervals.
the minute heart rate
If each of 3 consecutive RR intervals is larger than 1200
ms (means heart rate is lower than 50 bpm), the heart rate
is acquired by averaging the 4 most recent RR intervals.
Response time of heart rate HR change from 80/min to 120/min: ≤ 10s.
meter to change in heart rate HR change from 80/min to 40/min: ≤ 10s.
Heart rate updating rate of the
1s
display
Bandwidth: 1–100 Hz
Sensitivity: 0.5V/mV
Auxiliary Output
Propagation Delay Time: ≤ 35ms
No internal pacemaker pulse represented
The monitor does not reject the pacemaker pulses with
Pacemaker pulse rejection overshoot.
performance Amplitude: (aP) from ±2mV to ±700mV
Bandwidth: (dP) from 0.1ms to 2ms
Suppression of pacemaker
pulse detector to fast ECG Minimum input slew rate: 0.9V/s
signals
In the Monitoring, Operation, or Extended mode, the operator
Pacemaker pulse detector
can disable the pacemaker pulse detection (or cause it to be
failure
ignored).
The monitor shall comply with YY1079-2008, 10s after the
Overload protection application of differential mode voltage 1V (p-v) of mains
power supply frequency.
QRS amplitude (p-v RTI) 0.5mV - 5mV
QRS width (monitor-adult) 70ms - 120ms
QRS width
40ms - 120ms
(monitor-neonate/pediatric)
Parameter Specification
QRS no response amplitude
(except for neonate mode) ≤ 0.15 mV
(p-v RTI)
1mV QRS no response width
≤ 10ms
(except for neonate mode)
Power frequency voltage The maximum power frequency sine wave-to-valley
tolerance amplitude is not less than 100 μV (p-v).
Drift tolerance Compliant with 4.2.5.3 in YY1079-2008.
Adult/Pediatric: 15bpm - 300bpm
HR measurement range
Neonate: 15bpm - 350bpm Step:1bpm
HR measurement accuracy ±1 % or ±5 bpm, whichever is greater
Adult/Pediatric: 15bpm - 300bpm
HR alarm limit range
Neonate: 15bpm - 350bpm
HR alarm resolution 1bpm
No more than nominal value ±10% or ±5bpm, whichever is
HR alarm limit accuracy
greater.
Time to alarm for cardiac
≤10s
arrest
Time to alarm for low heart
≤10s
rate
Time to alarm for high heart
≤10s
rate
Time to suspend audible
1 minute, non-adjustable
alarms
Input Dynamic Range Compliant with 4.2.8.1 in YY1079-2008.
Input impedance ≥2.5MΩ
System noise ≤30µV(p-vRTI)
Multi-channel interference Compliant with 4.2.8.4 in YY1079-2008.
Gain control and stability Compliant with 4.2.8.5 a), c) and d) in YY1079-2008.
Time base selection and
Compliant with 4.2.8.6 a) in YY1079-2008.
accuracy
Output display Compliant with 4.2.8.7 a) in YY1079-2008.
Accuracy of rebuilt input
Compliant with 4.2.8.8 in YY1079-2008.
signal
Calibration voltage < ±10%.
CMRR (common mode The output signal amplitude should not exceed V (p-v) RTI
rejection ratio) within 60s, with CMRR≥89dB.
Baseline control and stability Compliant with 4.2.8.11 in YY1079-2008.
Pacemaker pulse display
Compliant with 4.2.8.12 in YY1079-2008.
capability
Interval: ≤ 35ms
Synchronous pulse of Amplitude: 500 times the input signal amplitude
cardioversion Width: No signal processing
Output impedance: ≥1.5MΩ
Electrosurgical interference
Comply with 4.2.8.14 in YY1079-2008.
suppression
ECG cables 3/5-lead
Leads off alarm Audible and visual indicators
Parameter Specification
6.25 mm/s,12.5 mm/s, 25 mm/s, 50 mm/s
Display sweep speeds
Error less than ±10%.
×1/4, ×1/2, ×1, ×2, ×4, Auto
ECG gain
Error less than ±10%, 1/4 grade error undefined.
Defibrillator protected Tested with 5 kV, recovery time< 5s
Monitoring Mode: 0.67Hz - 40Hz (Adult), 0.67Hz - 55Hz
(Pediatric)
Bandwidth
Operation Mode: 0.67Hz - 20Hz
Extended Mode: 0.05Hz - 100Hz
2. Information about wavelength ranges can be useful for clinicians performing photo dynamic
therapy.
Parameter Specification
Regardless of rise or drop in blood pressure, the maximum
NBP sensor accuracy error in measurements within the cuff should be ±0.4 kPa (±3
mmHg) at any measurement point in the range.
Adult/Pediatric:
SYS: 30 mmHg- 254 mmHg (4.0 kPa – 33.9 kPa)
DIA: 10 mmHg- 220 mmHg (1.3 kPa – 29.3 kPa)
NBP measurement and alarm MAP: 20 mmHg - 235 mmHg (2.7 kPa – 31.3 kPa)
limit range Neonate:
SYS: 30 mmHg - 135 mmHg (4.0 kPa – 18.0 kPa)
DIA: 10 mmHg - 110mmHg (1.3 kPa – 14.7 kPa)
MAP: 20 mmHg - 125mmHg (2.7 kPa – 16.7kPa)
Max. Std. Deviation:8 mmHg(1.1 kPa)
NBP accuracy
Max. Mean Error:±5 mmHg (±0.7 kPa)
Adult/pediatric: 0mmHg- 300mmHg (0 kPa- 40.0kPa)
Cuff pressure range
Neonate: 0 mmHg- 150 mmHg (0 kPa- 20.0kPa)
Adult: 160 mmHg (22.7 kPa)
Initial inflation pressure range Pediatric: 120 mmHg (21.3 kPa)
Neonate:90 mmHg (13.3 kPa)
+30 mmHg above the last measured Systolic value (only for
Subsequent cuff inflation
Auto and STAT)
NBP PR measuring range and
40bpm -240bpm
alarm limit range
PR measuring accuracy ±5bpm or ±5%, whichever is greater.
Display resolution 0.133 kPa (1 mmHg)
Static pressure measurement
0.0 kPa -34.7 kPa (0 mmHg -260 mmHg)
range
Parameter Specification
Display type Color TFT LCD
Size 26cm (10.4")
Resolution 800x600
Interface Specification
Nurse Call Alarm Output Connector: 3.5 mm phone jack, N.O and N.C contacts
Contact rating:
≤ 1A @ < 25Vac
≤ 1A @ < 60Vdc
Isolation: 1.5 kV
ECG analog output port 5 mm phone jack
USB ports Software upgrade or data export
Parameter Specification
Type Thermal array recorder
Printing mode Real-time, timed and alarm-triggered
Paper width 50mm
Print speed 12.5 mm/s , 25.0mm/s, 50.0mm/s, accuracy ±5%.
Parameter Specification
Relative humidity - storage 15% – 95%, condensation free
and transportation (15% – 80% with recorder)
Barometric pressure -
70.0 kPa - 106.0 kPa
working
Barometric pressure -
50.0 kPa - 106.0 kPa
storage and transportation
Warning Use of accessories, transducers, and cables other than those specified may
result in increased emissions and/ or decreased immunity of the GS20 Patient
Monitor.
The GS20 Patient Monitor should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the monitor should be
observed to verify normal operation in the configuration in which it is used.
Caution The following ports of the GS20 Patient Monitor are sensitive to ESD during
monitoring:
SpO2
ECG
To reduce ESD, please follow these instructions:
Do not contact the connector pins.
Ensure that the following ESD precautions are applied when connecting and
disconnecting the port cables:
— Make sure the operator is grounded before connecting the cable or
touching the grounded metal.
— Wear ESD grounded ribbon and ESD protective cloths.
— Provide ESD training for all users. Users should understand what is ESD,
how does it cause damage and how to prevent it.
d =[3.5/ E1] p
3V (effective 80MHz -800MHz
Conducted RF
value) [V1] V (effective
GB/T 17626.6 d =[7/ E1] p
150kHz-80MHz value)
800MHz - 2.5GHz
Radiated RF P — the maximum output power
3V/m [E1] V/m
GB/T 17626.3 rating of the transmitter in watts
80MHz-2.5GHz
(W) according to the transmitter
manufacturer;
d — the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey should
be less than the compliance level in
each frequency range.
Interference may occur in the
vicinity of equipment marked with
1. For transmitters rated at a maximum output power not listed above, the recommended
separation distance (d) in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
Note — At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Printed in China
*453564883131*
*A*