XG.1.DM Model
PulseVacuum Sterilizer
Introduction
The operation and maintenance manual is an essential part of the equipment, which can offer users
The operation and maintenance manual must be kept well. No matter where the equipment
damage.
damage.
1
Shandong Xinhua Medical Instrument Co.,Ltd.
Chapter 1. General
1.1 Manufacture identification materials
Address: No. 7 Taimei Road, Zibo New & Hi-tech Zone, Shandong, P.R. China.
Post Code: 255086
Technical parameter:
In the warranty period, if there is something wrong with the machine, our company will provide
After the warranty period, our company will provide high quality service at a price just cover the
cost.
2
Shandong Xinhua Medical Instrument Co.,Ltd.
Commitment: Once received your application for service, we will offer you satisfactory solution
in 24 hours.
1. Before open the package of the equipment, please check whether the package is integrity.
We advice you to keep the packing material well because you may reuse it later. After unfold the
package, please check whether the autoclave has somewhere damaged. If any, please judge and
record the damage degree, for example, panel scratch. At last, please check the packing list and
examine whether the accessories are complete. If you find some problems, please immediately
2. Please operate the machine strictly obey the procedure introduced in the operation and
maintenance manual. If the autoclave cannot operate normally, please first refer the fault
analysis table in the manual and try to solve the faults; if you cannot solve the faults, please
immediately feed back the faults to after-sale service department of our company. If you can
provide us the detailed description about the breakdowns of the autoclave and the model and
quantity of the parts probably need to change, our company will appreciate it very much.
3
Shandong Xinhua Medical Instrument Co.,Ltd.
,
P1 P2
unlock lock
Touch screen display, powerful communication function between man and machine, easy
Temperature, pressure and time can be displayed dynamically on the touch screen,
operating process status and alarm information can also be shown on the touch screen.
Independent steam generator equipped, heating rapidly, uniform and high thermal
efficiency
Adopting pulse vacuum technology to exhaust the air in the chamber thoroughly, can ensure
the saturation level of the sterilizing medium and also ensure the autoclave up to effective
sterilization in a shot time.(The C series autoclaves without vacuum pump use positive
Vacuum drying ensures the perfect drying effect for the dressing and instrument(The C
series autoclaves without vacuum pump adopt high temperature drying method )
4
Shandong Xinhua Medical Instrument Co.,Ltd.
Having BD test, PCD test and leakage test program; Vacuum leakage rate ≤
0.22μm air filter ensures the air turn back to the chamber is asepsis air, which can avoid
configuration)
It should be placed in clean, dry, shady room with ventilation and low temperature difference.
Safety device
Over temperature auto protection device: the power will be cut off when the temperature over
rated value.
Door safety interlock device: the sterilizer start the program and turn on the power only when
the door closed correctly; the door can not open when there’s pressure in the chamber.
Over pressure safety valve: the safety valve open when the pressure is too high.
Over pressure alarm: The touch screen display alarm when the main body pressure is too high
Low water position alarm device: when the water position lower than rated, the power will be
cut off and the alarm information will be shown.
5
Shandong Xinhua Medical Instrument Co.,Ltd.
Chapter 3 Installation
Machine unloading:
Pay attention to the following:
Don’t stand under the crane;
Checking:
After unpacking,check every parts of the machine and make a record, if any problem contact
us by phone.
Check carefully the links and fastening parts after a long distance transportation, if there is
Ground space required: Considering the operation and maintenance of the equipment,
distance between the wall and the sides of the autoclave should be no less than 300mm. The
height of room should be no less than 2.9m. At least 1.5 meters for the operation end.
Groundsill: If installed on the ground , the ground surface must be tamped and planished. If
on the second floor, users must consider whether to strengthen the floor or not according to
idiographic condition.
6
Shandong Xinhua Medical Instrument Co.,Ltd.
Airiness and dispel heat: In order to assure the equipment work normally and be in
comfortable working condition ,Users must fix a suited airiness system that controlled the
temperature in the range of 15 ~35℃ and humidity below 85% around the autoclave.
Electrical source: 3 phases with five wires, 3 of them are active line, one is zero line (light
blue) and one is ground line (yellow and green), power source capacity and mains load should
be more than that required for the machine. Also a power box is required on the right or back
wall of the equipment, in which a 3 phase breaker and a one phase breaker are needed.
To ensure person and equipment safety, users must join a wire with the ground. Users
must reliably join the wire marked “ground “in the control cable with ground.
3.2 Cautions:
1. Installation is not allowed in a place of conductive particles, corrosive gas, inflammable
gas.
2. Installation should avoid places with possible electrical shock and vibration.
3. Places of high temperature, high humidity or accessible to rain are not suitable for
installation.
4. Places of high magnetic intensity are not allowed, either.
5. Autoclave uses absolute pressure control, please adjust the relevant parameters when it is
7
Shandong Xinhua Medical Instrument Co.,Ltd.
3.3.2 Apply to XG.1.DM series
3 phase electric motor is used on the vacuum pump, dismatching phase consequence with
power source may occur for the first installation or reinstallation of the machine.
Solution: Turn off power, exchange any two active lines, make connection again and try till
Tap water is cycling for the vacuum pump as the working medium, pressure 0.05-0.10 MPa,
temperature: ≤25℃. Start the machine and measure current value on each line of the three from the
pump with clip-on ammeter. If the measured current is much different from the rated value for the
pump, circulation water flow should be adjusted (speed and pressure) usually decreased.
3.3.3 In the first time of running the autoclaves, the injection pump must be filled( if equipped),
to avoid burning out the pump.
图3
8
Shandong Xinhua Medical Instrument Co.,Ltd.
The system has been set administrator, process technician and 8 operators, the password
of administrator and process operator is 149, the administrator’s Operating authority is the
highest.
P4 P5
P6
Picture 4-6 are the logging in pictures of administrator, process technician and 8 operators
(2)Parameters setting:
Operators: Can modify the passwords of operator, for example log in operator 1#, then
modify the new password,. Operator can only choose and start the programs.
9
Shandong Xinhua Medical Instrument Co.,Ltd.
Process technician: can modify the passwords of process technician, can modify the
parameters of sterilization procedure (as picture 7, 8), can modify the sterilization
temperature, pulse times, pressure limit inside the chamber, cooling temperature and drying
图7 图8
Administrator: has the highest operating authority, can modify the parameters (as picture 9),
can modify the printing parameters: print interval, F0 print, language, report printing.
Administrator can also modify the password both of process operator and operators.
P9 P10
(3)Option program and start
Choose the correct program ( as picture 10) XG.1.DM series do not have the
10
Shandong Xinhua Medical Instrument Co.,Ltd.
vacuum programs, choose the program then enter into the start parameters (as picture
11), then click start.
P11 P12
(4)System maintenance:
After landing as administrator, click the button of maintenance, enter into time set (picture 12),
modify the time and save, click the button of back. MAC、IP have to set only monitor under the
(5)Manual operation:
Manual operation only can be used by process operator and administrator (as picture 13). Click
the button “auto” system switch to manual operation; click each part to modify the status. “B” is
vacuum pump as the picture shows, but XG.1.DM series do not have vacuum pump valve, recycling
water valve and Circular valve. Please notice heating pipe DR1 and heating film DR2 are forbidden
11
Shandong Xinhua Medical Instrument Co.,Ltd.
图 13
(6)If the user’s area is not the standard a standard atmospheric pressure, please contact service
engineer to adjudge the parameters.
The being sterilized articles must be washed before put into the sterilizer, in order to avoid the
rested things on the articles damaging the sterilizer and themselves. For example: blood and other
impurity.
① The rested things on the instruments that need sterilize should be washed firstly.
②We suggest use washer, purified liquid and distilled water to wash the instruments.
③ Pay attention to the following before placing instruments into the sterilizer tray:
Instruments should be placed uniformly in the tray, away from each other, overlapping is not
Make sure that the instruments not of the same metal (stainless steel, carbon steeled.) are
In case carbon steel instruments are placed in stainless steel trays, the tray should be lined
with a towel or paper-wrap before placing the instrument. There should be no direct contact
Wrapped instruments should be packed in material, which promotes drying, with good gas
Clean tubing and rinse with pyrogen free water. Ensure that both ends of the tubing are open, without any
13
Shandong Xinhua Medical Instrument Co.,Ltd.
Place packs upright on trays. They should not touch the chamber walls
Water preparation
PH value:5—7
Attention:
Water should be in accordance with requirement to be used like: distilled, purified
water, otherwise blockade in the pipe may occur and damaging on the solenoid valve
Work medium “steam” must be generated from pure water thus to prevent steam
source contamination.
In the first time of running the autoclaves, the injection pump must be filled( if
14
Shandong Xinhua Medical Instrument Co.,Ltd.
Chapter 5 Operation
screen, then open the door, load materials to be treated into the trays and close the door.
The items of liquid has to be unsealed, XG.1.DM series do not have the vacuum programs and pulse
stages
4) Choose proper program, then press “start/stop”, the door is locked automatically, the
program starts to run, the non-liquid programs have stages, they are: “preparation”, “pulse”(apply to
XG.1.DM), “Heating up”, “Venting”, “drying”, “end” and the liquid programs have stages, they are
15
Shandong Xinhua Medical Instrument Co.,Ltd.
5) During the running of program, click the button of numerical value then can Real-time monitor
the running times in the screen, Visual review on the relative parameters for different stages on the
screen is available.
图 14
6) The system will alarm and stop operation automatically when there is “overheating”, “over
pressure” or something abnormal. Make sure the pressure is in safe range ±10kPa,the buzzer loud,
When the screen shows “end” stage lights and buzzer loud, can open the door to take out items. If
the screen shows “ low temperature alarm” but the program does not stop and completely finished,
which means the program has adjusted the sterilization flow automatically and finish.
7) Close the door, turn off the switch and cut off power.
numerical value, then click the button of exit. Till there is a pressure balance in and out of the
autoclave and the temperature reach to opening requirement, then open the door.
Chapter 6 Maintenance
16
Shandong Xinhua Medical Instrument Co.,Ltd.
In order to ensure the autoclave work in the favorable condition,and reduce the trouble time to a
Caution: Before starting, ensure the electric cord is disconnected and there is no
Daily:
Use a gauze with cleansing agent and water to clean the wall and bracket of autoclave
Supplies pure water or distilled water timely, To prepare for the next Sterilization.
Weekly
◆ Clean the cover with soft cloth.
Take down the door gasket, clean seal groove, then wipe it with clean paper or gauze.
Clean the door gasket and seal groove with alcohol. Put it to the sealed groove after it is dried
Monthly
Check the open and close door. Removed door handle, add lubricate grease on the door
bolt.
Yearly
Tighten the tie-in and check communication state by professional electrician once a
year.
17
Shandong Xinhua Medical Instrument Co.,Ltd.
Faults Analyze
After switch on, when the screen display the mark of alarm and the buzzer alarmed,
the printer will print the alarm name and code, click the button of the screen to check the
information and click the button of ensure, then check the faults, can not start the program without
19
Shandong Xinhua Medical Instrument Co.,Ltd.
Steps as follows:
Tab
Safety valve
Safety valve
joint
In order to prevent the safety valve from blockage, every month, under ordinary usage, it is
necessary to allow the steam pressure to escape through it.
1. Operate a sterilization cycle according to the manual
3. Pull the ring of the safety valve using a screwdriver, and lift the safety valve for 2 seconds.
4. Switch off the power supply to interrupt the operation, and exhaust steam from the chamber.
5. Wait until the pressure goes down to zero, then open the door
2. Replace it with a new one (be sure the valve opens at a pressure between 0.25 MPa-0.26Mpa).
20
Shandong Xinhua Medical Instrument Co.,Ltd.
2. Pull out the stainless press slip on top of the electrical magnetic valve with a screwdriver.
21
Shandong Xinhua Medical Instrument Co.,Ltd.
Hexago
nal nut
Remove
Follow the lift
steps , remove the
nut then take out the
spool.
7.8 Cleaning for the water feeding filter and drainage filter
1. The water feeding filter is used for preventing dirt enter into the steam generator.
2. The drainage filter is used for preventing enter into the inside pipeline and the magnetic
valve
22
Shandong Xinhua Medical Instrument Co.,Ltd.
23
Shandong Xinhua Medical Instrument Co.,Ltd.
24
Shandong Xinhua Medical Instrument Co.,Ltd.
25
Shandong Xinhua Medical Instrument Co.,Ltd.
1537
chamber exhaust
steam generator drain
26
Shandong Xinhua Medical Instrument Co.,Ltd.
Sealing door^
wat er
1630
cover
Filter
真空泵
Fi lter
+30
1420(1670) 0
A
Ou tlet valve
900
100?
27