Professional Documents
Culture Documents
Service Manual FDR MS-1000 Main
Service Manual FDR MS-1000 Main
AMULET
FDR MS-1000
Service Manual
1 mR = 0.258 μC/kg
FDR MS-1000 Service Manual This volume Used when minor or medium levels of physical injury may be incurred if the
instruction is not observed.
FDR MS-1000 Service Manual Troubleshooting (MT) Also used when the machine may suffer serious trouble (such as unrecoverable
FDR MS-1000 Service Manual Maintenance Utility (MU) or difficult-to-recover trouble).
Precautions for Handling of this Manual Used to indicate the matters that need attention during steps of the procedure.
DRMS0085.ai
DRMS0091.ai
Stand jig -
Wrench
- -
(13x13 mm)
DRMS0086.ai
Handle -
DRMS0087 ai
-
circuit cable DRMS0089.ai
Temperature
adjustment unit relay CAUTION -
cable DRMS0090.ai
Acrylic board
(40 mm thick)
DRMS0093.ai
Mammographic
compression gauge
DRMS0094.ai
Dosimeter
DRMS0095.ai
Safety Precaution
1. Safety Precautions................................................................... 1
1.1 General Precautions.........................................................................1
1.2 Precautions on Radiation.................................................................5
1.3 Laser................................................................................................6
1.4 Declaration of Conformity.................................................................6
1.5 Precautions on Infection...................................................................6
1.6 Safety and Various Symbols............................................................7
1.7 Precautions on Patient Environment................................................8
2. Labels........................................................................................ 9
2.1 Laser Precaution Labels...................................................................9
2.1.1 Laser Precaution Label Attachment Locations.............................................9
2.1.2 List of Labels................................................................................................9
2.2 Other Labels...................................................................................10
2.2.1 Attachment Locations for Ratings Indication Label and Other Labels.......10
2.2.2 List of Ratings Indication Label and Other Labels.....................................13
2.2.3 Handling Instruction Labels and Attachment Locations.............................20
3. CLASSIFICATION................................................................... 21
Product Specifications
1. Summary of the Machine.................................................. MD-1 3.4 Startup Process of the Controller Unit and FPD......................MD-30
1.1 Features of the Machine............................................................MD-1 3.5 FDR-1000DRSZ/FDR-1000DRLZ Shutdown Process............MD-31
1.2 System Configuration Examples...............................................MD-2
4. Electrical-Related Information....................................... MD-32
1.3 Overall Machine Configuration and External View....................MD-3
4.1 Information on Board Locations..............................................MD-32
1.3.1 Machine External View and Functions.................................................. MD-3
1.4 X-Ray Irradiation Field Size.......................................................MD-9 5. Interlock Mechanisms and Operations......................... MD-35
1.5 I/O Parts Information...............................................................MD-10
1.6 System Block Diagram............................................................MD-12
1. Precautions for Check, Replacement, and Adjustment... MC-1 3.1.3 Replacing the FPD.............................................................................. MC-47
3.1.4 Unpacking the FPD............................................................................. MC-54
2. Covers................................................................................ MC-4 3.1.5 Temporarily Installing the FPD............................................................ MC-56
2.1 Swivel Arm Covers....................................................................MC-4 3.1.6 FPD Inner Cover................................................................................. MC-57
2.1.1 Swivel Arm Top Cover........................................................................... MC-6 3.1.7 Board Cooling Fan.............................................................................. MC-59
2.1.2 Swivel Arm Lower Cover....................................................................... MC-8 3.1.8 PTC Board.......................................................................................... MC-61
2.1.3 Swivel Arm Left-Hand Side Cover....................................................... MC-10 3.2 Temperature Adjustment Unit..................................................MC-62
2.1.4 Swivel Arm Right-Hand Side Cover.................................................... MC-11 3.3 Collimator................................................................................MC-70
2.1.5 FPD Lower Cover................................................................................ MC-12 3.3.1 Checking the Collimator Mounting Angle............................................ MC-72
2.1.6 FPD Left-Hand Cover (Left-Hand Armrest)......................................... MC-13 3.3.2 Adjusting the Collimator Mounting Angle............................................ MC-72
2.1.7 FPD Right-Hand Cover (Right-Hand Armrest).................................... MC-14 3.3.3 Checking the Collimator Lamp............................................................ MC-73
2.1.8 Carbon Cover...................................................................................... MC-16 3.3.4 Checking the Irradiation Field Size..................................................... MC-73
2.1.9 Compression Unit Left-Hand Side Cover............................................ MC-19 3.4 Single Tank Unit......................................................................MC-74
2.1.10 Swivel Arm Left-Hand Front Cover..................................................... MC-21 3.4.1 Removing the Single Tank Unit........................................................... MC-75
2.1.11 Compression Unit Right-Hand Side Cover......................................... MC-22 3.4.2 Reinstalling the Single Tank Unit......................................................... MC-79
2.1.12 Swivel Arm Right-Hand Front Cover................................................... MC-23 3.4.3 Replacing the Single Tank Unit........................................................... MC-81
2.1.13 Compression Unit Front Cover............................................................ MC-24
3.5 Compression Unit....................................................................MC-83
2.2 Elevation Stand Covers...........................................................MC-25 3.5.1 Checking the Compression Unit......................................................... MC-88
2.2.1 Elevation Stand Top Cover.................................................................. MC-26 3.5.2 Correcting the Compression Thickness Indication.............................. MC-88
2.2.2 Elevation Stand Left-Hand Side Cover............................................... MC-26 3.5.3 Compression Plate (Plastic Plate Part)............................................... MC-90
2.2.3 Elevation Stand Right-Hand Side Cover............................................. MC-27
2.2.4 Elevation Stand Rear Cover............................................................... MC-27
3.6 CBA Board (D905)...................................................................MC-91
2.2.5 Display Cover...................................................................................... MC-28 3.7 Grid Assembly.........................................................................MC-93
2.2.6 Base Plate Cover................................................................................ MC-28 3.7.1 Removing the Grid Assembly.............................................................. MC-93
2.2.7 Elevation Stand Inner Cover (Lower).................................................. MC-29 3.7.2 Installing the Grid Assembly................................................................ MC-96
2.2.8 Elevation Stand Inner Cover (Upper).................................................. MC-29 3.7.3 Checking the Grid Position................................................................. MC-97
3.7.4 Adjusting the Grid Sensor................................................................... MC-97
3. Swivel Arm Unit............................................................... MC-30
3.8 MFI Board (D910) Assembly.................................................MC-101
3.1 FPD.........................................................................................MC-30
3.9 XIF51A Board........................................................................MC-103
3.1.1 Removal of the FPD............................................................................ MC-30
3.1.2 Installing the FPD................................................................................ MC-36 3.10 Fan (For Cooling the Single Tank Unit).................................MC-104
5.1 Controller Unit Covers...........................................................MC-137 6.3 Fuse Location of the Swivel Arm...........................................MC-161
6.3.1 XIF51A Board Fuse........................................................................... MC-161
5.2 MIF51A Board.......................................................................MC-138
5.3 MPI51A Board.......................................................................MC-139 7. Updating Software Versions........................................ MC-162
5.4 CPS51B/CPS51C Board.......................................................MC-139 7.1 Updating the RU Software Versions......................................MC-162
5.5 PAT51A Board.......................................................................MC-140
5.6 CFI51A Board (Memory Board).............................................MC-141
8. Backup and Restore of the Configuration Data......... MC-166 13. Jigs................................................................................. MC-219
8.1 Configuration Data of the Mammographic Stand..................MC-166 13.1 List of Jigs.............................................................................MC-219
8.1.1 Backing Up the Configuration Data of the Mammographic Stand.... MC-166 13.2 How to Use the Water Supply Jig..........................................MC-222
8.1.2 Restoring the Configuration Data of the Mammographic Stand....... MC-169 13.2.1 Connecting the Water Supply Jig...................................................... MC-224
8.2 RU Data.................................................................................MC-172 13.2.2 Filling the Coolant............................................................................. MC-226
8.2.1 Backing Up the RU Data................................................................... MC-172 13.2.3 Draining the Coolant......................................................................... MC-231
8.2.2 Restoring the RU Data...................................................................... MC-172 13.2.4 Connecting the Closure Hose (FPD and Temperature
8.2.3 Installing the Machine-Specific Data................................................. MC-173 Adjustment Unit)................................................................................ MC-234
Installation (IN)
1. Installation Work Flowchart............................................... IN-1 11. Installing the Mammographic Stand............................... IN-42
11.1 Marking and Drilling on the Installation Position........................IN-42
2. Installation Preparations.................................................... IN-3
11.2 Securing the Mammographic Stand..........................................IN-43
2.1 Preinstallation Work....................................................................IN-3
2.2 Tools and Measuring Instruments Necessary for Installation......IN-5 12. Connecting the Cables to the Controller Unit................ IN-44
3. Precautions in Installation................................................. IN-7 13. Connecting the Cables to the Mammographic Stand.... IN-46
8. Installing the Controller Unit............................................ IN-23 18. Installing the FPD.............................................................. IN-58
18.1 Removing the Carbon Cover.....................................................IN-58
9. Connecting the Power Supply Cable ............................. IN-26 18.2 Removing the Inner Cover and the Cable Retaining Bracket....IN-60
9.1 Connecting the Power Supply Cable of the Controller Unit.......IN-26 18.3 Mounting the FPD.....................................................................IN-61
9.2 Connecting the Power Supply Cable of 18.4 Connecting the Cable Connectors and Attaching the Cable
the Mammographic Stand.........................................................IN-30 Retaining Brackets....................................................................IN-68
9.3 Checking the Wiring Connections and Connecting to the 18.5 Putting on the Inner Cover........................................................IN-70
Distribution Switchboard............................................................IN-34
18.6 Attaching the Grid Guide...........................................................IN-70
10. Removing the Packing Material of the Swivel Arm (1).... IN-36 18.7 Changing the DIP Switch Setting on the CBA (D905) Board....IN-72
18.8 Putting on the Swivel Arm Right- and Left-Hand Front Covers...IN-73 21.5 Installing the Machine-Specific Data.......................................IN-106
18.9 Putting on the Compression Unit Right- and Left-Hand Side 21.6 Replacing the Controller Unit AEC File...................................IN-108
Covers.......................................................................................IN-74 21.6.1 18x24 (only for Use in Europe)........................................................... IN-108
18.10 Checking the Grid Position (1)..................................................IN-75 21.6.2 24x30 (for Use Outside Europe)......................................................... IN-108
18.10.1 Checking the Grid Wire......................................................................... IN-75 21.7 Replacing and Merging the AEC File in the AWS....................IN-109
18.10.2 Checking the Grid Park Position........................................................... IN-76 21.7.1 Replacing the AEC File....................................................................... IN-109
18.10.3 Checking the Grid IN Position............................................................... IN-77 21.7.2 Merging the AEC File.......................................................................... IN-109
18.10.4 Checking the Gap between the Grid and the Exposure Plane............. IN-78
21.8 Backing Up the RU Data......................................................... IN-111
18.11 Putting on the Carbon Cover.....................................................IN-78
18.12 Checking the Grid Position (2)..................................................IN-80 22. Time Setting for the Mammographic Stand.................. IN-114
18.12.1 Checking the Oscillation....................................................................... IN-80
23. FPD Calibration............................................................... IN-116
19. Installing the Temperature Adjustment Unit................... IN-81
24. Checking the Exposure Recognition Function............ IN-117
20. Putting on the Covers....................................................... IN-86
25. Checking the AEC........................................................... IN-118
20.1 Putting on the FPD Right- and Left-Hand Covers
(Right- and Left-Hand Armrests)...............................................IN-86 25.1 Confirmation of AEC Accuracy................................................ IN-118
20.2 Putting on the FPD Lower Cover...............................................IN-87 25.2 Confirmation of Dose per Exposure Mode..............................IN-121
20.3 Putting on the Swivel Arm Right- and Left-Hand Side Covers....IN-88
26. Checking the Variation in X-Ray Output....................... IN-123
20.4 Putting on the Swivel Arm Top and Lower Covers....................IN-89
20.5 Putting on the Elevation Stand Covers......................................IN-90 27. Checking for Image Problems and Checking the
S Value............................................................................. IN-127
21. Installing the RU Software................................................ IN-94 27.1 Checking for Image Problems.................................................IN-127
21.1 Installing the RU PC-TOOL.......................................................IN-94 27.2 Checking the S Value..............................................................IN-127
21.2 Setting the IP Address...............................................................IN-96
21.3 Setting the AWS and Checking Connection............................IN-100 28. Checking the Operation of the Mammographic Stand.. IN-128
21.4 Installing the RU Software.......................................................IN-100 28.1 Checking the Elevation of the Swivel Arm...............................IN-128
28.2 Checking the Rotation of the Swivel Arm................................IN-129 34. Measurement of Installation Environment.................... IN-150
28.3 Checking the Emergency Stop Switches................................IN-130 34.1 Measurement of Installation Environment
28.4 Checking the Compression Unit..............................................IN-131 (Temperature and Humidity)....................................................IN-150
28.4.1 Operation and Operation Check of the Compression Plate................ IN-131 34.2 Measurement of Installation Environment
28.4.2 Checking Decompression................................................................... IN-132 (Power Supply Voltage) [Only Inside Japan]...........................IN-150
28.4.3 Checking the Optimum Compression Pressure Function................... IN-134
28.4.4 Checking the Automatic Decompression Function............................. IN-136
Appendix 1. Requirements Regarding Locally
Obtained Parts......................................... Appx IN1-1
28.4.5 Checking the Compression Thickness................................................ IN-137
28.4.6 Checking the Compression Pressure Indication Accuracy................. IN-138 Appendix 2. Measurement of Power Source
28.4.7 Checking Continuity of the Compression............................................ IN-139 Impedance [Only Inside Japan].............. Appx IN2-1
28.5 Checking the Collimator..........................................................IN-140
28.5.1 Checking the Collimator Mounting Angle............................................ IN-140
Appendix 3. Installation Environment Measurement
28.5.2 Checking the Collimator Lamp............................................................ IN-141
(Power Supply Voltage)
[Only Inside Japan]................................. Appx IN3-1
28.5.3 Checking the Irradiation Field Size..................................................... IN-141
Appendix 3.1 Measurement Flow................................................Appx IN3-1
29. Clearing the Error Log.................................................... IN-142 Appendix 3.2 Setting Up the Working Environments...................Appx IN3-1
29.1 Clearing the RU Error Log.......................................................IN-142 Appendix 3.2.1 Installing the Waveform Viewer........................................Appx IN3-1
29.2 Clearing the Error Log of the Mammographic Stand...............IN-144 Appendix 3.3 Preparation for Measurement................................Appx IN3-3
Appendix 3.3.1 Measurement Overview...................................................Appx IN3-3
30. Protective Grounding Test (Outside Japan)................. IN-146 Appendix 3.3.2 Measurement Period........................................................Appx IN3-3
Appendix 3.3.3 Setting the Measurement Conditions...............................Appx IN3-3
31. Measurement of Leakage Current and Patient Appendix 3.3.4 Storing the Measurement Data........................................Appx IN3-5
Leakage Current (Outside Japan).................................. IN-147 Appendix 3.4 Measurement.........................................................Appx IN3-6
Appendix 3.4.1 Measurement...................................................................Appx IN3-6
32. Visual Inspection and Cleaning of the Machine........... IN-148
Appendix 3.5 Transfer of the Data to the PC and Data Display...Appx IN3-7
33. Measurement of Power Source Impedance Appendix 3.6 Data Analysis.........................................................Appx IN3-7
[Only Inside Japan]......................................................... IN-149 Appendix 3.6.1 Printing the Data...............................................................Appx IN3-7
Appendix 3.6.2 Data Analysis....................................................................Appx IN3-7
Appendix 3.6.3 Example of Reporting the Measurement Data.................Appx IN3-8
Appendix 3.7 Power Supply Voltage Specifications....................Appx IN3-9 Appendix 5. Unloading................................................. Appx IN5-1
Appendix 3.7.1 Power Supply Voltage Specifications for
the FDR MS-1000............................................................Appx IN3-9
Appendix 5.1 Unloading and Preparation for Transfer.................Appx IN5-1
Appendix 5.1.1 Unpacking the Mammographic Stand..............................Appx IN5-2
Appendix 4. Measurement of Installation Environments Appendix 5.1.2 Unpacking the FPD, the Controller Unit and
Accessories.....................................................................Appx IN5-4
(Temperature and Humidity)................... Appx IN4-1
Appendix 5.1.3 Unpacking the Operation Desk and
Appendix 4.1 Measurement Flow................................................Appx IN4-1 the Equipment Rack........................................................Appx IN5-5
Appendix 4.2 Setting Up the Working Environments...................Appx IN4-1 Appendix 5.1.4 Unpacking the Elevation Stand Cover..............................Appx IN5-6
Appendix 4.2.1 Name of Each Part of the Data Logger............................Appx IN4-1 Appendix 5.2 Transferring the Mammographic Stand.................Appx IN5-7
Appendix 4.2.2 Installing the Manager Program.......................................Appx IN4-2 Appendix 5.2.1 Using the Dedicated Carrier.............................................Appx IN5-7
Appendix 4.3 Preparing for Measurement...................................Appx IN4-4 Appendix 5.2.2 When not Using the Dedicated Carrier...........................Appx IN5-10
Appendix 4.3.1 Measurement Conditions.................................................Appx IN4-4 Appendix 5.3 Transferring Other Packed Items.........................Appx IN5-11
Appendix 4.3.2 Recording Interval and Measurable Days........................Appx IN4-4
Appendix 4.3.3 Setting the Measurement Conditions...............................Appx IN4-4 Appendix 6. Installing the Optional Monitor Arm....... Appx IN6-1
Appendix 4.4 Measurement Start................................................Appx IN4-7
Appendix 7. Installing the Optional X-Ray
Appendix 4.5 Measurement End.................................................Appx IN4-7 Protective Board...................................... Appx IN7-1
Appendix 4.5.1 Terminating the Measurement on the Data Logger
Main Body........................................................................Appx IN4-7
Appendix 8. Installing the 1.8-Fold Magnification
Appendix 4.5.2 Terminating the Measurement from the Personal
Computer.........................................................................Appx IN4-8
Exposure Stand Kit S.............................. Appx IN8-1
Appendix 4.6 Transferring Data to the Personal Computer.........Appx IN4-8 Appendix 8.1 Putting the Positioner Retainer..............................Appx IN8-1
Appendix 4.7 Data Analysis.........................................................Appx IN4-9 Appendix 8.2 Adjustment of the Position of the Positioner..........Appx IN8-2
Appendix 4.7.1 Charting the Data.............................................................Appx IN4-9
Appendix 4.7.2 Data Analysis.................................................................. Appx IN4-11
Appendix 9. Installing the Front Cover Kit................. Appx IN9-1
Appendix 4.7.3 Example of Reporting the Measurement Data............... Appx IN4-11 Appendix 9.1 Checking the Items Supplied.................................Appx IN9-2
Appendix 4.8 Measurement Conditions for Temperature and Appendix 9.2 Removing the Covers............................................Appx IN9-3
Humidity...............................................................Appx IN4-12
Appendix 9.3 Replacing the Bracket under the Swivel Arm.........Appx IN9-3
Appendix 4.8.1 Measurement Conditions for the FDR MS-1000............Appx IN4-12
Appendix 9.4 Internal Wiring and Putting on the Stereo Cover...Appx IN9-4
FDR MS-1000
03.14.2011 03 Revision (FM5842) All pages
Service Manual
Safety Precaution
1. Safety Precautions - Electrical charge (residual charge) accumulated by a capacitor might remain in a
circuit after the circuit breaker of the distribution switchboard is turned OFF.
Be sure to proceed with operations after you are sure that the charge has been
Warnings, operating precautions, and instructions should be observed to avoid completely discharged, due to fears of electric shock hazards.
possible physical hazards and serious accidents that may occur during installation and
You can check the discharge by the following indications.
servicing.
“PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
Labels and name plates that describe relevant precautions are attached to the
machine. The instructions on such labels and name plates should also be observed
while installation/servicing procedures are performed.
electric shock and short circuit, as instructed in this service manual. LED on the MFI board: The yellow LED is unlit.
MFI board
[Check] LED
DRMS0074 ai
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -1
Safety Precaution -2
- Wait 5 seconds or longer after the circuit breaker of the power supply unit (elevation - To restart the machine, wait twenty seconds or longer after turning OFF the power.
stand) is turned OFF before starting operation, when the terminal block cover of the If you turn the power back ON shortly after power OFF, the machine may fail to start
power supply unit (elevation stand) is to be removed, to avoid electrical shocks due up normally.
to the residual charge. - When servicing a board, be sure to wear an antistatic wristband to ground your
Terminal block cover body. If your body is not grounded, static buildup on your body may damage
electronic components on the board.
Wait 5 seconds or longer after
the circuit breaker of
the power supply unit (elevation stand) is
turned OFF before removing the cover.
- Press the system OFF button of the control pad without turning OFF the main circuit
breaker of the mammographic stand, to turn OFF the power of the mammographic
stand.
This is because your fingers might touch the terminals of the power supply cable
and get an electric shock as the main circuit breaker of the mammographic stand is
close to the terminal block of the power supply unit.
If you need to turn OFF the main circuit breaker of the mammographic stand for
some reason, turn OFF the circuit breaker of the distribution switchboard, and make
sure that the charge has been completely discharged, before you turn OFF the main
circuit breaker of the mammographic stand.
Mammographic stand
Terminal of the
power supply cable
DRMS0075.ai
- The mammographic stand is kept energized unless the system OFF button of the
control pad is pressed, even if the circuit breaker of the controller unit is turned
OFF. Press the system OFF button of the control pad to turn OFF the power of the
mammographic stand.
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -2
Safety Precaution -3
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -3
Safety Precaution -4
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -4
Safety Precaution -5
m
m
0
50
1,
Patient
45°
Protective board
(optional)
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -5
Safety Precaution -6
X-ray focus
1,500 mm or more
Operation
desk
Control pad
(exposure switch)
Exposure chamber
DRMS0045 ai
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -6
Safety Precaution -7
Stop
DRMS0076.ai Swivel arm up/down movement icon
DRMS0077.ai
Power-OFF (partial)
DRMS0055.ai
DRMS0078.ai
Swivel arm angle adjustment icon
Power-ON (partial)
DRMS0056.ai
DRMS0079.ai
Grid mounting/dismounting icon
Alternating current
DRMS0058.ai DRMS0081.ai Operation-off icon
Type B application
DRMS0059.ai
DRMS0082.ai
Exposure icon
DRMS0061.ai
High voltage
Year of manufacture
DRMS0083.ai
DRMS0084.ai
EFUP (Environmentally Friendly Use Period)
DRMS0064.ai
Light field lamp button icon
DRMS0065.ai
This button does not function
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -7
Safety Precaution -8
2.5 m
2.5 m 2.5 m
FR9H0001 .EPS
Medically-used room
Room equipped with protective grounding (medical use outlet or medical use
grounding terminal) implemented by the medical grounding method. The protective
grounding inside the medically-used room is equipotential, and the protective
grounding of this medically-used room is equipotential to that in the other medically-
used room.
Generally, a portion of the medically-used room is the patient environment.
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -8
Safety Precaution -9
DRMS0034 ai
l Collimator
Collimator
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -9
Safety Precaution -10
Controller unit
FDR MS-1000
Elevation system label (*)
A stand label
Pharmaceutical
system label
Collimator
Ratings indication label UL authentication
(FDR-1000DRSZ label
(FDR-1000DRS or
or FDR-1000DRLZ)
FDR-1000DRL)
Pharmaceutical
notification label
Protective ground mark label
Mammographic stand
C
Protective ground mark label
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -10
Safety Precaution -11
Ratings
indication
label
Ratings indication
label
DRMS0042.ai
Power supply (mammographic stand)
n Control Pad
Caution label
Continuous operation
Warning label condition label
Stop label
DRMS0041.ai
n FPD
High voltage High voltage
power supply label caution label
FPD
Caution label
Control pad
DRMS0044 ai
DRMS0043 ai
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -11
Safety Precaution -12
Motor inverter
Controller unit
n Foot Switches
Foot switch label n Cover
l Mammographic stand
Stop label
Continuous
operation condition DRMS0068.ai
label
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -12
Safety Precaution -13
2.2.2 List of Ratings Indication Label and Other Labels l FDR-1000DRLZ
<For use in Japan>
n System Label System label is not applied to a machine.
DRMS0088 ai
DRMS0035.ai
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -13
Safety Precaution -14
l FDR-1000DRSZ
DRMS0002 ai
DRMS0039.ai
l FDR-1000DRLZ
DRMS0003 ai
DRMS0085.ai
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -14
Safety Precaution -15
DRMS0036.ai DRMS0037.ai
l FDR-1000DRLZ l FDR-1000DRL
DRMS0086.ai DRMS0087.ai
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -15
Safety Precaution -16
DRMS0004 ai
DRMS0038.ai
l FDR-1000DRL
DRMS0005 ai
DRMS0089 ai
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -16
Safety Precaution -17
Barcode indication
Lot symbol Serial number
DRMS0020 ai DRMS0048.ai
DRMS0021.ai
DRMS0047.ai
DRMS0019 ai
DRMS0049.ai
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -17
Safety Precaution -18
DRMS0012.ai
DRMS0027.ai
DRMS0030 ai
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -18
Safety Precaution -19
DRMS0072.ai
FRAH0021.EPS
DRMS0031.ai
DRMS0071 ai
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -19
Safety Precaution -20
2.2.3 Handling Instruction Labels and Attachment
Locations
REFERENCE
No handling instruction labels are attached to the machine.
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -20
Safety Precaution -21
3. CLASSIFICATION n PC
IEC60950-1 approved product
n FDR-1000DRSZ/FDR-1000DRLZ
1. According to the type of protection against electrical shock
1. According to the type of protection against electrical shock CLASS 1 EQUIPMENT
CLASS 1 EQUIPMENT
n RX210 (Monitor)
2. According to the degree of protection against electrical shock
Type B 1. According to the type of protection against electrical shock
CLASS 1 EQUIPMENT
3. Protection against harmful ingress of water
Main unit: IPX0 2. According to the degree of protection against electrical shock
Foot switch: IPX1 Not applicable (N/A)
4. According to the degree of safety of application in the presence of 3. Protection against harmful ingress of water
a flammable anesthetics mixture with air or with oxygen or nitrous IPX0
oxide
4. According to the degree of safety of application in the presence of
Equipment not suitable for use in the presence of a flammable anesthetics mixture
with air or with oxygen or nitrous oxide. a flammable anesthetics mixture with air or with oxygen or nitrous
oxide.
5. According to the mode of operation Equipment not suitable for use in the presence of a flammable anesthetics mixture
Continuous use with intermittent load with air or with oxygen or nitrous oxide.
6. Continuous use 5. According to the mode of operation
Take the shortest exposure interval indicated in the graph (input energy of tube vs. Continuous operation
shortest exposure interval) for continuous exposures, and make less than 60 shots
per hour.
{Product Specifications:1.2.1_FDR-1000DRSZ/FDR-1000DRLZ Specifications_
l Heating curve and cooling curve of the X-ray tube}
n HUB
IEC60950-1 approved product
1. According to the type of protection against electrical shock
CLASS 1 EQUIPMENT
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -21
Safety Precaution -22
4. Cautions on Electromagnetic 4.1.1 Further Information for IEC 60601-1-2 (EN 60601-1-2)
Model name FUJIFILM Digital Mammography System FDR MS-1000 is referred to as
Waves FDR MS-1000 in this section.
- Medical electrical equipment needs special precautions regarding EMC and needs
4.1 Electromagnetic Compatibility (EMC) to be installed and put into service according to the EMC information provided in the
accompanying documents.
This equipment has been tested and found to comply with the limits for medical - Portable and mobile RF communications equipment can affect medical electrical
devices to IEC 60601-1-2:2001+A1:2004 (EN 60601-1-2:2001+A1:2006). equipment.
These limits are designed to provide reasonable protection against harmful - Information regarding the cable affecting EMC is as follows.
interference in a typical medical installation.
Name General Specification
This equipment generates, uses and can radiate radio frequency energy and, if not
Network cable
installed and used in accordance with the instructions, may cause harmful interference (FDR-1000AWS)
Cat5e or more, UTP type and straight cable
to other devices in the vicinity. Network cable
However, there is no guarantee that interference will not occur in a particular (FDR-1000DRSZ/ Cat5e or more, STP type and straight cable
installation. FDR-1000DRLZ)
If this equipment does cause harmful interference to other devices, which can be - The use of accessories, transducers and cables other than those specified, with the
determined by tuning the equipment off and on, the user is encouraged to try to exception of transducers and cables sold by FUJIFILM Corporation as replacement
correct the interference by one or more of the following measures; parts for internal components, may result in increased emissions or decreased
- Reorient or relocate the receiving device. immunity of the FDR MS-1000.
- Increase the separation between the equipment. - The FDR MS-1000 should not be used adjacent to or stacked with other equipment.
- Connect the equipment into an outlet on a circuit different from that to which the If adjacent or stacked use is necessary, the FDR MS-1000 should be observed to
other device(s) are connected. verify normal operation in the configuration in which it will be used.
Consult a FUJIFILM dealer for help as necessary. - Items to be tested (see the table below)
This Service Manual contains the names of standards applied in IEC 60601-1-2. Guidance and manufacturer’s declaration - electromagnetic emissions
The names of standards applied in EN 60601-1-2 are replaced as follows. The FDR MS-1000 is intended for use in the electromagnetic environment
specified below. The customer or the user of the FDR MS-1000 should assure that
IEC 60601-1-2 it is used in such an environment.
Standards applied in EN 60601-1-2
(descriptions in this Service Manual)
Electromagnetic environment -
CISPR 11:1997+A1:1999+A2:2002 EN 55011 Emissions test Compliance
guidance
IEC 61000-3-2:2000+A1:2001+A2:2004 EN 61000-3-2 RF emissions The FDR MS-1000 uses RF energy only
CISPR 11 for its internal function. Therefore, its
IEC 61000-3-3:1994+A1:2001+A2:2005 EN 61000-3-3
Group 1 RF emissions are very low and are not
IEC 61000-4-2:1995+A1:1998+A2:2000 EN 61000-4-2 likely to cause any interference in nearby
IEC 61000-4-3:2006 EN 61000-4-3 electronic equipment.
IEC 61000-4-4:2004 EN 61000-4-4 RF emissions
CISPR 11 Class A
IEC 61000-4-5:2005 EN 61000-4-5 The FDR MS-1000 is suitable for use in
IEC 61000-4-6:2003+A1:2004+A2:2006 EN 61000-4-6 Harmonic emissions all establishments other than domestic
IEC 61000-3-2 N/A and those directly connected to the public
IEC 61000-4-8:1993+A1:2000 EN 61000-4-8 low-voltage power supply network that
IEC 61000-4-11:2004 EN 61000-4-11 supplies buildings used for domestic
Voltage fluctuations/ purposes.
flicker emissions N/A
IEC 61000-3-3
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -22
Safety Precaution -23
Guidance and manufacturer’s declaration - electromagnetic immunity Guidance and manufacturer’s declaration - electromagnetic immunity
The FDR MS-1000 is intended for use in the electromagnetic environment specified below. The FDR MS-1000 is intended for use in the electromagnetic environment specified below. The
The customer or the user of the FDR MS-1000 should assure that it is used in such an customer or the user of the FDR MS-1000 should assure that it is used in such an environment.
environment.
Electromagnetic IEC 60601 Compliance
Immunity test Electromagnetic environment - guidance
Immunity test IEC 60601 test level Compliance level environment - test level level
guidance Conducted RF 3 Vrms 3 Vrms Portable and mobile RF communications
Electrostatic ±6 kV contact ±6 kV contact Floors should be equipment should be used no closer to any
discharge (ESD) wood, concrete or IEC 61000-4-6 150 kHz - 80 MHz part of the FDR MS-1000, including cables,
±8 kV air ±8 kV air ceramic tile. If floors than the recommended separation distance
IEC 61000-4-2 are covered with calculated from the equation applicable to
synthetic material, the frequency of the transmitter.
the relative humidity
should be at least Recommended separation distance
30%. Radiated RF 3 V/m 3 V/m d = 1.2
Electrical fast ±2 kV for power ±2 kV for power Mains power quality
transient/burst supply lines supply lines should be that of a IEC 61000-4-3 80 MHz - 2.5 GHz d = 1.2 80 MHz - 800 MHz
±1 kV for input/output ±1 kV for input/output typical commercial or
IEC 61000-4-4 lines lines hospital environment.
d = 2.3 800 MHz - 2.5 GHz
Surge ±1 kV differential ±1 kV differential Mains power quality
mode mode should be that of a where P is the maximum output power rating
IEC 61000-4-5 ±2 kV common mode ±2 kV common mode typical commercial or of the transmitter in watts (W) according to
hospital environment. the transmitter manufacturer and d is the
Voltage dips, short <5% UT <5% UT Mains power quality recommended separation distance in metres
interruptions and (>95 % dip in UT ) (>95% dip in UT ) should be that of a (m).
voltage variations on for 0.5 cycle for 0.5 cycle typical commercial or
power supply input hospital environment. Field strengths from fixed RF transmitters,
lines 40% UT 40% UT If the user of the FDR as determined by an electromagnetic site
(60% dip in UT ) (60% dip in UT ) MS-1000 requires survey,a should be less than the compliance
IEC 61000-4-11 for 5 cycles for 5 cycles continued operation level in each frequency range.b
during power mains
70% UT 70% UT interruptions, it is Interference may occur in the vicinity
(30% dip in UT ) (30% dip in UT ) recommended that of equipment marked with the following
for 25 cycles for 25 cycles the FDR MS-1000 symbol:
be powered from an
<5% UT <5% UT uninterruptible power
(>95% dip in UT ) (>95% dip in UT ) supply or a battery.
for 5 s for 5 s
Power frequency 3 A/m 3 A/m Power frequency NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
(50/60 Hz) magnetic magnetic fields NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
field should be at levels affected by absorption and reflection from structures, objects and people.
a
characteristic of a Field strength from fixed transmitters, such as base stations for radio (cellular/cordless)
IEC 61000-4-8 typical location in a telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
typical commercial or broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
hospital environment. environment due to fixed RF transmitters, an electromagnetic site survey should be
NOTE: UT is the a.c. mains voltage prior to application of the test level. considered. If the measured field strength in the location in which the FDR MS-1000 is
used exceeds the applicable RF compliance, the FDR MS-1000 should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the FDR MS-1000.
b
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3 V/m.
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -23
Safety Precaution -24
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -24
Safety Precaution -25
BLANK PAGE
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -25
Safety Precaution -26
BLANK PAGE
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -26
Safety Precaution -27
BLANK PAGE
019-201-03E
FDR MS-1000 Service Manual Safety Precaution -27
Control Sheet
Issue date Revision number Reason Pages affected
08.25.2008 01 Release for each application (FM5396) All pages
10.20.2008 01 (1) New release (FM5447) All pages
03.20.2009 02 Revision (FM5504) All pages
05.20.2009 02 (1) Revision (FM5552) All pages
FDR MS-1000
03.14.2011 03 Revision (FM5842) All pages
Service Manual
Product Specifications
COMP PLATE SPOT 9X9 #(E) Compression plate, rectangle spot FRONT COVER KIT BIOPSY
Front cover kit (18x24) Needed when using the
18×24 #(E)
biopsy positioner kit or
Compression plate for axilla FRONT COVER KIT BIOPSY the stereo exposure kit.
COMP PLATE AXILLA 8X20 #(E) Front cover kit (24x30)
radiography 24×30 #(E)
1.8-fold magnification exposure BIOPSY AC CORD 20M # Biopsy power supply cable (20 m) Only inside Japan.
TABLE KIT #(E)
stand kit (18x24) (*1)
*1: Optional only for the FDR-1000DRSZ
1.8-fold magnification exposure
TABLE KIT L #(E)
stand kit (24x30) (*2)
*2: Optional only for the FDR-1000DRLZ
019-201-03E
FDR MS-1000 Service Manual Product Specifications -1
Product Specifications -2
019-201-03E
FDR MS-1000 Service Manual Product Specifications -2
Product Specifications -3
Operating time
208/230/240 V~ (single phase two-wire or three-wire)
l Input current
13.5/12.2/11.7 A for 208/230/240 V~
Standard ambient
Spatial distance 3 mm
temperature
Percentage of change
in operating time
Ambient temperature °C DRMS1012.ai
019-201-03E
FDR MS-1000 Service Manual Product Specifications -3
Product Specifications -4
l Climate conditions l Vibration conditions on the floor (installation surface)
<Operating> Frequency Amplitude
Temperature: 20°C–30°C
Humidity: 30%–75% RH (Without moisture condensation) XY direction 10–55 Hz 0.0075 mm or less
Atmospheric pressure: 750–1060 hPa 10–20 Hz 0.0024G or less
Z direction
<Non-operating> 20–55 Hz 0.0015 mm or less
Temperature: 15°C–35°C
Humidity: 10%–80% RH (Without moisture condensation) l Variable magnetic field
Atmospheric pressure: 750–1060 hPa DC 0.3 Gauss or less
<Transit/storage>
The condition values are applicable when the FPD is not equipped.
Temperature: -10°C–50°C
Humidity: 10% (10°C)–90% (50°C) RH (Without moisture condensation)
Atmospheric pressure: 750–1060 hPa
l Attenuation equivalent
Exposure table: 0.1 mmAl
Magnification exposure stand: 0.15 mmAl
Inner cover: 0.01 mmAl
l Operating mode
Continuous use with intermittent load
019-201-03E
FDR MS-1000 Service Manual Product Specifications -4
Product Specifications -5
l Heating curve and cooling curve of the X-ray tube l Load curve of tube
Heating and cooling curves of the X-ray tube assembly Focus 0.1 Mo Focus 0.1 W
DRMS1016 ai
400
Shortest exposure interval [sec]
350
300
250
200
150
100
50
0
0 2000 4000 6000 8000 10000 12000 14000 16000 18000 20000
Input for exposure [kV x m x As] DRMS1015.ai
019-201-03E
FDR MS-1000 Service Manual Product Specifications -5
Product Specifications -6
l X-ray tube (Mo/W) l High voltage generator
<Type> <Tube voltage waveform>
P40Mo/W Multi-pulse type (high frequency: 500 kHz)
Molybdenum or molybdenum/tungsten rotary anode with beryllium window
<Tube voltage range>
<Focal size> 23 kV to 35 kV (in 1 kV steps)
Large focus: 0.3 mm
Small focus: 0.1 mm <mAs value set range (28 kV Mo target)>
* based on the start pattern test Large focus: 2 mAs ─ 500 mAs (±10%, +0.2 mAs)
Small focus: 2 mAs ─140 mAs (±10%, +0.2 mAs)
<X-ray tube current (Mo/W)>
<Exposure time>
- Maximum X-ray tube current with molybdenum anode
Large focus: 10 ms ─ 4 s
Large focus: 150 mA with 25 kV
Small focus: 60 ms ─ 6 s
Small focus: 28 mA with 25 kV
- Maximum X-ray tube current with tungsten focus <Voltage set value accuracy>
Large focus: 188 mA with 25 kV ±5% (measured in a high voltage circuit)
Small focus: 34 mA with 25 kV
<mAs value setting accuracy>
<Rated maximum X-ray tube voltage> ±5% (measured in a high voltage rectifying circuit)
40 kV
<Nominal maximum power (25 kV to 35 kV)>
<Combination of anode and filter Mo/W tube> Molybdenum: Large focus 3.75 kW, small focus 0.7 kW
Mo/30 µmMo Tungsten: Large focus 4.5 kW, small focus 0.85 kW
Mo/25 µmRh
W/50 µmRh
019-201-03E
FDR MS-1000 Service Manual Product Specifications -6
Product Specifications -7
l X-ray conversion mode 1.2.2 FPD Specifications
Direct conversion
l Frequency
l Conversion substance Single phase 50 Hz–60 Hz
a-Se
l Input voltage
l Effective image area 200–240 V~ (single phase two-wire or three-wire)
FDR-1000DRSZ: 177 mm (main) ×237 mm (sub)
FDR-1000DRLZ: 236.4 mm (main) ×296.4 mm (sub) l Input current
14.0–11.7 A for 200–240 V~
l Pixel resolution
50 x 50 µm/pix l Climate conditions
<Operating>
l Number of pixels Temperature: 20°C–30°C
FDR-1000DRSZ: 3,540 (main) ×4,740 (sub) = 16,779,600 pixels Humidity: 15%–80% RH (Without moisture condensation)
FDR-1000DRLZ: 4,728 (main) ×5,928 (sub) = 28,027,584 pixels Atmospheric pressure: 750–1060 hPa
l Concentration resolution <Non-operating>
16 bit Temperature: 15°C–35°C
Humidity: 10%–80% RH (Without moisture condensation)
l Threshold spatial frequency Atmospheric pressure: 750–1060 hPa
10.0 lp/mm <Transit/storage and unpacking>
l Calibration frequency Temperature: 10°C–30°C
Humidity: 10%–80% RH (Without moisture condensation)
Once/day or more
Atmospheric pressure: 750–1060 hPa
l Overload protection
Controller unit: Circuit breaker built in
FPD: Circuit breaker in the controller unit
l I/F cable
{INAppx:1._Requirements Regarding Locally Obtained Parts}
019-201-03E
FDR MS-1000 Service Manual Product Specifications -7
Product Specifications -8
1.3 Dimensions and Weight of the Machine 1.3.2 Dimensions and Weight of FDR-1000AWS
Refer to the AWS Service Manual for the dimensions and the weight of the FDR-
1.3.1 Dimensions and Weight of FDR-1000DRSZ/FDR- 1000AWS.
1000DRLZ
n Dimensions
Mammographic stand: W1000 x D1200 x H1980 (mm)
Controller unit: W300 x D550 x H750 (mm)
Mammographic stand
Controller unit
550
750
300
1980
1200
FPD
1100
1000
Unit: mm
DRMS1001.ai
n Weight
Mammographic stand: Approx. 320 kg
FPD: Approx. 20 kg (FDR-1000DRSZ)
Approx. 23 kg (FDR-1000DRLZ)
Controller unit: Approx. 22 kg
019-201-03E
FDR MS-1000 Service Manual Product Specifications -8
Product Specifications -9
2,690 or more
2,400 or more (*1)
Mammographic stand •
•
FPD •
2,100 or more (*2)
1,500 or more 1,500 or more
990 or more
3,000 or more
Unit: mm
(*1): The machine can be installed with no clearance Unit: mm (*1): Required space when the machine is used with the defaults as
(0 mm) on the rear of the controller unit. DRMS1002.ai set in the factory without repositioning the swivel arm elevation
upper-limit switch.
(*2):Required space when the machine is used with the swivel arm
elevation upper-limit switch lowered by 300 mm. DRMS1003.ai
019-201-03E
FDR MS-1000 Service Manual Product Specifications -9
Product Specifications -10
n Operation Desk
350 350
or more or more
Operation desk
1,000 or more
Unit: mm
DRMS1010.ai
019-201-03E
FDR MS-1000 Service Manual Product Specifications -10
Product Specifications -11
REFERENCE
Although the machine is set in the factory so that the swivel arm stops at 2,320 mm
when it is moved up to its uppermost position, the uppermost position of the swivel
arm can be adjusted within a range of 2,050 to 2,320 mm by repositioning the swivel
arm elevation upper-limit switch. 90° 90°
The corresponding adjustment range of the uppermost position of the FPD upper
surface is 1,200 to 1,500 mm.
Swivel arm
Swivel arm elevation
upper-limit switch
FPD
90° 90°
Uppermost position of Lowermost position
the swivel arm: 2,320 of the swivel arm:
(adjustable in a range 1,510 (*)
of 2,000 to 2,320) Unit: mm
DRMS1007 ai
Unit: mm
DRMS1060.ai
(*): The lowermost position (the position where the swivel arm is lowered to its lowest
position) does not change as the elevation lower-limit switch is not repositioned.
019-201-03E
FDR MS-1000 Service Manual Product Specifications -11
Product Specifications -12
60
147.4
465
670
705
643
780
Unit: mm
DRMS1004.ai
019-201-03E
FDR MS-1000 Service Manual Product Specifications -12
Product Specifications -13
n Operation Desk
700
400
200
150
400
Operation desk
Unit: mm
DRMS1009.ai
019-201-03E
FDR MS-1000 Service Manual Product Specifications -13
Product Specifications -14
1.7 System Configuration Example n Installing the Operation Desk in the Exposure Chamber
CAUTION
WARNING
Do not use a multi-socket/extension cord for the quality and safety of the When the operation desk is to be installed in the exposure chamber, put the
product. optional protective board for radiation protection. Place the operation desk at a
distance of 1,500 mm or more from the X-ray focus.
u NOTES u
- The FDR-1000DRSZ/FDR-1000DRLZ can be installed in an exposure chamber
(patient environment).
- Appropriately furnish the exposure chamber and the operation chamber to visually 500 or more
and acoustically check the patient conditions.
Control pad
(exposure switch)
Controller
unit
Operation desk
FDR-1000AWS
790 790
FDR-1000DRSZ/
FDR-1000DRLZ 1,500 or more
Operation
desk
Control pad
DRMS1005 ai (exposure switch)
Patient chamber
Unit: mm
DRMS1008 ai
019-201-03E
FDR MS-1000 Service Manual Product Specifications -14
Product Specifications -15
n Lead Solder
- Lead in electrical components and printed boards: Approx. 0.5 kg
019-201-03E
FDR MS-1000 Service Manual Product Specifications -15
Control Sheet
Issue date Revision number Reason Pages affected
08.25.2008 01 Release for each application (FM5396) All pages
10.20.2008 01 (1) New release (FM5447) All pages
03.20.2009 02 Revision (FM5504) All pages
05.20.2009 02 (1) Revision (FM5552) All pages
FDR MS-1000
03.14.2011 03 Revision (FM5842) All pages
Service Manual
FDR MS-1000
FDR-1000DRSZ FDR-1000AWS
(FDR-1000DRLZ)
Personal
Controller computer
unit Monitor
1.3.1 Machine External View and Functions Controller unit Refer to "n Controller Unit".
AWS - Registers and acquires
Machine Main Body patient ID information;
checks, processes and
Mammographic stand delivers images.
- Sets the X-ray exposure
Control pad conditions.
Foot switches (right and - Moves up/down the swivel
left) arm and compresses/
decompresses the
compression plate.
Foot switch
Control pad Refer to “n Control Pad”.
Controller unit
DRMS2007.ai
Name Description Remarks Biopsy positioner kit Controller unit, Face guard,
Attachment, Needle holder,
Compression plate, rectangle Used for spot exposure. Option key for the software.
spot Stereo exposure kit Stereo exposure plug, Software
Compression plate for axilla Used in axilla radiography. key CD, Instruction manual.
radiography Face guard W Needed when using the stereo
1.8-fold magnification exposure Used for magnification exposure. Only for FDR- exposure kit.
stand kit (18x24) 1000DRSZ. Front cover kit (18x24) Needed when using the biopsy
1.8-fold magnification exposure Used for magnification exposure. Only for FDR- Front cover kit (24x30) positioner kit or the stereo
stand kit (24x30) 1000DRLZ. exposure kit.
1.8-fold magnification exposure Used for magnification exposure. Biopsy power supply cable (20 m) For biopsy controller unit. Only inside Japan.
stand kit S
Compression plate for Used for magnification exposure. Options (FDR 1000-AWS)
magnification
Refer to the AWS Service Manual for the detailed information.
Compression plate, rectangle Used for magnification spot
spot magnification exposure.
Exposure switch (remote type) Exposure button. Separately Available Accessories
Exposure switch (foot switch Exposure button.
type) Name Description Remarks
X-ray protective board Used with the operation desk 16 kg
installed in the exposure chamber. (grass-made) Equipment rack For accommodating the equipment.
Can be installed in combination with
Chest wall pad supplement Spare chest wall pad. the operation desk.
Left axilla pad supplement Spare left axilla pad.
Anchor plate For retaining the operation desk and
Right axilla pad supplement Spare right axilla pad. the equipment rack (*1)
Monitor arm Retains the AWS monitor. FCR 1 shot Phantom M Plus
15 m power supply cable For power supply to the controller Only inside Japan.
unit. Booster transformer Raises the supply voltage if the Only inside Japan.
standard is not satisfied.
25 m power supply cable For power supply to the controller Only inside Japan.
unit. Compression plate wall-hanging Accommodates the compression
18x24 compression plate (Low) Only for FDR- holder (*2) plate.
1000DRLZ.
*1: The operation desk or the equipment rack must be fixed to the floor with the anchor bolts.
18x24 compression plate (High) Only for FDR- If anchor bolt holes cannot be made on the floor as requested by the customer, however,
1000DRLZ. you need use the optional anchor plate to fix the operation desk and the equipment rack.
18x24 compression plate (Flex) *2: Attach the compression plate wall-hanging holder to a sufficiently rigid wall.
24x30 compression plate (Low) Only for FDR-
1000DRLZ.
24x30 compression plate (High) Only for FDR-
1000DRLZ.
24x30 compression plate (Flex) Only for FDR-
1000DRLZ.
2D biopsy compression plate For use of exposures for biopsy.
Mobile kit Fixed components to be mounted Only for use in
on the trailer. Europe and Australia.
Control panels
(right and left)
Grip handles
(right and left)
Exposure table
DRMS2045.ai
DRMS2006.ai
Exposure button
Circuit breaker
Compression Compression
X-ray irradiation field size
plate
16.5 cm
plate for
16.5 cm
magnification
With a grid Without a grid
24 cm
0.5 cm
18x24
0.5 cm
24 cm Chest wall edge Chest wall edge
compression DRMS2042.ai DRMS2046.ai
plate (High)
16.5 cm
18 cm
18x24
compression
plate (Low)
9 cm
Compression
0.5 cm
0.5 cm
9 cm
rectangle spot
30 cm
0.5 cm
30 cm Chest wall edge
24x30 DRMS2043.ai
compression
plate (High)
16.5 cm
24 cm
12.4 cm
24x30
compression Compression
plate (Low) plate,
0.5 cm
0.5 cm 12.4 cm
0.5 cm
2D biopsy
16.5 cm
18 cm
compression
plate Compression 20 cm
8 cm
DRMS2041.ai DRMS2042.ai
0.5 cm
Chest wall edge
DRMS2044.ai
Sensors/Switches FPD
Subscanning motor (MZ1)
Internal cooling air
temperature sensor Internal cooling fan (FAN6)
(TH2)
(TH3) DRMS2030.ai
TH2 Internal cooling air Thermistor Detects the temperature of the FAN6 Internal cooling fan DC sirocco Circulates cooling air in the FPD.
temperature sensor internal cooling air. fan
TH3 External atmospheric Thermistor Detects the external atmospheric FAN8, 9 Board chamber cooling fan DC fan Cools the inside of the board
temperature sensor temperature. chamber.
PI: Photointerrupter FAN10 Board chamber cooling fan DC fan Cools the inside of the board
chamber.
Not incorporated in the FDR-
1000DRSZ.
24V/12V
Power Service power CF card
Read light source unit TCP (Tape line) Power supply source switch
distribution 1GB
Bucky drive ASIC (ChargeAmp) PAT51A
M CFI51A
Image transfer board
rotating swivel arm and controlling Bucky
TRC51A
Bucky unit Image data
Video
(SERDES) IDE
board High-speed image Image reception
Image data transmission circuit unit
ADC51A rearrangement
High voltage X-ray detector
CPU51A
Sensor and AD timing generation
PCI
power MPI51A
source
Electromagnetic
AEC51A CPU: Pentium4-3G
Fan pump Mechanical
AutoExposure OS: VXworks
Ctrl Temperature controller unit
control PTC51A
Erasure unit (ERS51A) SZ1 device 24V DIMM
temperature- CPS51B (*)
TDB51A regulating circuit 1GB
M
JCT51A Image (for single window
Subscanning Temperature sensor
correction unit + system
drive 24V (up to 4 images) + correction)
High voltage
(actuator driver controller unit Load control board control circuit; GbE
Display circuit) SND51B Control Air-
IO expansion FPGA MIF51A
unit communication cooling
fan
Exposure stand control system XIF51A READY, PREP, STATE, X-RAY, power source synchronization signals (etc.), Network
(parallel communication); SIZE, AEC dose information, etc. (serial communication)
Acquisition
Workstation
AWS
24V 400V
Emergency- Lift Rotation Single Tank
Display Stand Display Arm
Stop Button Main Filament Unit
DYS DYA
Switches LRB Inverter STU
MFI
24V CAN
Control Box
CAN CAN Control Board
CTB Ctrl
Ctrl
Collimator
Control Board Stand Control Board Arm CBC
CBS CBA
5V
Foot-Switches 24V 24V Compression
Biospy Unit
Unit
BYU
CNU
Control
24V DC/DC Detector
Unit
Power Carbinet
Ctrl
L1 L1 48V
L2 / N L2 / N 24V
PE PE Power Supply Unit 5V
PSU 400V
24V
PSU PSU_400V_Good
Control Board
PSU_400V_Good_Rtn
CBS
Stand
Power Supply Unit PSU_400V_Enable
PSU_400V_Enable_Rtn
Main Filament
AC-Fail / Konverter
Control Board Stand
Inverter
PSU_AC_Fail_Temp
MFI
UDC_P
PSU_AC_Fail_Temp_Rtn Temperatur 400V DC
UDC_N
CBS
PSU_Plug_In
PSU_Plug_Out PSU_48V_Sense
CBS
PSU_48V_Good_Rtn
PSU_48V_Enable
PSU_48V_Enable_Rtn
Motor Lift
Rotation
Konverter
MLR
L1
Sicherungen Netzfilter PFC-Stufe 48V_P
L2 / N 48V DC
GND_48V
PE
PSU_24V_Sense
Control Board
PSU_24V_Sense_Rtn
CBA
Arm
CBS
PSU_24V_Good
PSU_24V_Good_Rtn
PSU_24V_Enable
PSU_24V_Enable_Rtn
Main Filament
24V_P
Inverter
GND_24V
MFI
Konverter
24V_P
24V DC
GND_24V
PSU_5V_Sense
Control Board
PSU_5V_Sense_Rtn
CBS
Stand
PSU_5V_Good
PSU_5V_Good_Rtn
CBS
Control Board Stand
PSU_On
Konverter
CBS
DRMS2010.ai
CAN_24V
X6 / 8 X6 / 8 X29 / A1
CAN_GND
X6 / 6 X6 / 6 Lift-Stop Top
X6 / 1 X29 / B1
CAN_L
X6 / 2 X6 / 2
CAN_H
X6 / 7 X6 / 7
CAN_GND
X6 / 3 X6 / 3
X29 / A2
Collision-Switch
CBS X29 / A4
MLR_24V_P
X25 / A1 X2 / 6
MLR_GND_24V
X25 / B1 X2 / 4 Lift-Stop Bottom
MLR_Lift_ENPO_24V
X25 / A2 X2 / 12 X3 / U X29 / A3
Bremse
MLR_Lift_3H_24V
X25 / B2 X2 / 10 X3 / V
MLR_Lift_Qstop_24V
X25 / A3 X2 / 13 X3 / W
MLR_Lift_Up_24V
X25 / B3 X2 / 14 X3 / PE X29 / A8
MLR_Lift_Down_24V
X25 / A4 X2 / 15 Emergency-Stop Bottom
X25 / B4
MLR_Lift_RSHO_24V
X2 / 16
Lift-Motor X29 / B8
MLR_Lift_RSHO_24V
30VAC
X25 / A6 X2 / 17
MLR_Lift_Brake_24V
X25 / B6 X2 / 1
X28 / 1 MLR_Lift_Brake
X28 / 3 MLR_Brake_Rth
DRMS2011.ai
X3 / 2 48V_GND X1 / L- X28 / A7
Emergency-Stop
Left / Right
X28 / B7
SSI_5P0
X6 / 3
SSI_GND
X6 / 8
CBS X6 / 4
SSI_Data_H
CAN_24V SSI_Data_L
X13 / 9 X5 / 9 X6 / 5
CAN_GND SSI_Clk_H
X6 / 14
X13 / 6 X5 / 6
SSI_Clk_L
CAN_L X6 / 15
X13 / 2 X5 / 2
CAN_H
X13 / 7 X5 / 7
CAN_GND
X13 / 3 X5 / 3
CAN_SYNC_H
X13 / 8 X5 / 8
CAN_SYNC_L
X13 / 4 X5 / 4
X5 / 5
MLR_24V_P SSI_5P0
X25 / A10 X2 / 5
MLR_GND_24V SSI_GND
Bremse
X25 / B10 X2 / 4 X3 / U
MLR_Rot_ENPO_24V SSI SSI SSI_Data_H
X25 / A6 X2 / 12 X3 / V
Geber Geber
MLR_Rot_SH_24V SSI_Data_L
X25 / B6 X2 / 16 X3 / W
MLR_Rot_Qstop_24V SSI_Clk_H
X25 / A7 X2 / 13 X3 / PE
MLR_Rot_Go_24V SSI Tomo SSI_Clk_L
X25 / B7 X2 / 14
MLR_Rot_RSHO_24V Geber SSI
X25 / B8 X2 / 18
Rot-Motor Geber
MLR_Rot_RSHI_24V
X25 / A9 X2 / 17 30VAC
MLR_Rot_Brake_24V
X25 / B9 X2 / 1
MLR_Rot_Brake
X27 / 2
MLR Brake Rtn
X27 / 3
DRMS2012.ai
MFI
Main and Filament Inverter
X1
n_MFI_Plug_In
n_Out_A_Main_H
n_Out_A_Main_L n_Lg_Foc n_Err_Main
n_Out_B_Main_H n_Foc_W n_Err_Fil
n_Out_B_Main_L n_Fan_MFI n_Frr_Fil_Open
CPLD
CPLD
n_Out_Fil_H n_Fan_Arm MFI_Data_Out n_Frr_50kHz
MFI_Data_In
n_Out_Fil_L n_STU_Rel n_Frr_5V
n_Out_Bias_H n_An_Drv_Rel n_Frr_15V
Power_Out_A_Main
MFI_Data_In_L
Power_Out_B_Main
MFI_Data_Out_H
CBA
STU
I_Load_Main Power_Foc_3
50kHz auxiliary
I_Load_Main_Rth Power_Foc_4
voltage
IF_act Power_Bias
IF_act_Rth GND_Fil_Bias
UDC_act 24VDC with Supply X5
U_Bias line filter voltage AN_0
n_MFI_Plug_Out +5VDC AN_I
GND +15DC Main inverter AN_II
Gate driver
-15DC X9
CAnode
24V_P intermediate
GND_24 circuit with line
PSU
X3 filter / capacitors
UDC_P
UDC_N
Filament inverter
Foci switching
DRMS2013.ai
STU
Single Tank Unit
Power_Out_A_Main
Power_Out_B_Main
X-Ray Tube
Main Filament Inverter
Power_Foc_1
Tube
Durchfűhrungsplatine
Power_Foc_2
Temperatur
Power_Foc_3
MFI
Power_Foc_4 T
Stator
Power_Bias
GND_Fil
AN_0
1 2 0
AN_I Tube
AN_II Pressure
P
Control Board Arm
UT_Act
IT_Act
CBA
STU_Temp
STU_Pres
DRMS2014.ai
(CBS)
Interface Interface Interface
Reserved1_Rx_RS232
Reserved1_Tx_RS232
Opdima_Rx_RS232
Opdima_Tx_RS232
CAN_GND
MLR_Normal_Out
MLR_Tomo_Out
MLR_Normal_In
DSI_Rx_RS232
DSI_Tx_RS232
DYS_Plug_Out
MLR_Tomo_In
MLR_Motor_N
MLR_Motor_P
24V_Rth
DYS_Plug_In
5V_Rth
ENET_Rx_H
ENET_Tx_H
GND
GND
ENET_Rx_L
ENET_Tx_L
CAN_Ext_H
CAN_Ext_L
DYS_Rx_H
DYS_Tx_H
DYS_Rx_L
DYS_Tx_L
DYS_GND
DYS_24V
24V
5V
Schalter
Em_Stop_Stand_Out GND
Not-
Halt
Em_Stop_Stand_In VCC_24V
Power CAN
VCC_5V RS232 RS232 RS232 ENET
Distribution Interface DYS Interface Mach Interface
VCC_3.3V Interface Interface Interface Interface
2
VCC_1.2V
IOS_Xray_Led_In
IOS_Xray_Led_Out
IOS
IOS_Safety_In
Input Output Stand
Interface
IOS_Safety_Out PSU_On
IOS
GND PSU_Off
SDRAM PSU_On_Off_Rth
VCC
FPGA
CAN_L CAN PSU_AC_Fail_Temp
CAN_H Interface SRAM PSU_AC_Fail_Temp_Rth
1
CAN_GND PSU_400V_Good
Flash Application Specific Module (ASM) PSU_400V_Good_Rth
CAN_L PSU_400V_Enable
Nios Processor System (NPS)
PSU
1 PSU
CBA_Em_Stop_In Interface PSU_48V_Good_Rth
CBA_Em_Stop_Out
Watchdog Xray Release PSU_48V_Enable
CPLD PSU_48V_Enable_Rth
CPLD_Config_Rx_H Clock RTC Emergency Stop
Temperature PSU_24V_Good
Sensor
CPLD_Config_Rx_L PSU_24V_Good_Rth
Interface CPLD Interface FPGA Interface ASM
Control Board Arm
CPLD_Config_Tx_H PSU_24V_Enable
CPLD_Config_Tx_L PSU_24V_Enable_Rth
CBA
CBA_Plug2_In PSU_5V_Good
CBA_Plug2_Out / GND PSU_5V_Good_Rth
CBA PSU_Plug_In
n_Drive_Protect_H Interface PSU_Plug_Out
n_Drive_Protect_L
XC_H
XC_L
Xray_On_H CAN CAN MLR Interface
MCP Interface
Xray_On_L Interface 1 Interface 1
CBA_Plug1_In
CBA_Plug1_Out
MLR_Em_Stop_Bottom_Out
MLR_Em_Stop_Bottom_In
MLR_Em_Stop_Right_Out
MLR_Em_Stop_Top_Out
MLR_Em_Stop_Left_Out
MLR_Em_Stop_Right_In
MLR_Rot_Memory_Out
MLR_Lift_Collision_Out
MLR_Em_Stop_Top_In
MLR_Em_Stop_Left_In
MLR_Lift_Bottom_Out
MLR_Rot_Memory_In
MLR_Lift_Collision_In
MCP_Xray_Led_Out
MLR_Lift_Bottom_In
MCP_Em_Stop_Out
MCP_Decomp_Out
MLR_Lift_Top_Out
MCP_Xray_Led_In
MCP_Em_Stop_In
MLR_Lift_Rot_Rtn
PSU_On_Off_Rtn
MCP_Decomp_In
MLR_Lift_Top_In
MLR_Status_Out
MCP_Xray_Out
MCP_Plug_Out
MLR_Status_In
MCP_Xray_In
MCP_Plug_In
MLR_Lift_Rot
MLR_ENPO
MCP_Rx_H
MCP_Tx_H
MCP_Rx_L
MCP_Tx_L
MLR_GND
MCP_24V
MLR_24V
MLR_SH
PSU_On
PSU_Off
CAN_GND
CAN_H
CAN_H
CAN_L
CAN_L
MCP_GND
CAN_GND
VCC
RCI
Reserved CAN
Interface
MLR MCP
Motor Lift Rotation Main Control Panel
DRMS2015.ai
(CBA)
Interface
Rot_CCW
24V_Rtn
Lift_Down
DSI_Rx_RS232
DSI_Tx_RS232
Rot_CW
Lift_Up
PERSRVED
RESERVED
24V
STU_Temp
GND
GND
GND
VCC
STU_Pres
COM_RX
COM_TX
UT_Act
READY
STATE
IT_Act
X-RAY
PREP
GND
VCC_24V
Power
VCC_5V RS232 LRB
Distribution XIF Interface STU Interface
Interface Interface
VCC_3.3V
Control Board Collimator
VCC VCC_1.2V
GND
n_MFI_Plug_In
CBC
Out_A_Main_H
CAN_H Out_A_Main_L
CAN_L Out_B_Main_H
CAN_GND Out_B_Main_L
Out_Fil_H
CAN SDRAM Out_Fil_L
Interface
FPGA
Out_Bias_H
SRAM Out_Bias_L
CAN_L MFI_Data_In_H
Flash Application Specific Module (ASM)
MFI
MFI MFI_Data_Out_L
Emergency Stop Interface n_MFI_Err_H
CPLD_Config_Rx_H n_MFI_Err_L
Watchdog Xray_On UT_Max
Control Board Stand
n_CBU_Plugged_out
CUB_Em_Stop_Out
BYU_Em_Stop_Out
n_BYU_Plugged_In
n_CBS_Plugged_In
CUB_Em_Stop_In
BYU_Em_Stop_In
OT_Cass_Loaded
Comp_Thickness
FW_Sel_Wing_1
FW_Sel_Wing_2
BYU_Rx_RS232
BYU_Tx_RS232
Comp_Plate_ID
RDB_VCO_H
OT_Table_ID
RDB_VCO_L
Comp_Force
Decompress
GD_Go_Out
RDB_Rx_H
RDB_Rx_H
RDB_Tx_H
Comp_Dec
RDB_Tx_L
GD_Power
GD_Go_In
Comp_Inc
Compress
GD_Front
GD_Rear
GD_Turn
GND
VCC
GND
GND
VCC
VCC
GND
VCC
Device
Light source unit
Subscanning HP
sensor (SZ1)
Sliding shaft
Shaft slide guide
Drive
Flexible cable
transmission
Transmission belt shaft
Decelerating pulley
Subscanning motor (MZ1) Decelerating belt
DRMS2033.ai
Guide rail Lower-limit switch (rear) Stops the down-movement of the swivel arm when the lower
limit switch (front or middle) does not work. If the switch
works, the power circuit breaks, and the machine stops.
DRMS2039.ai
Sprocket
Reduction gear/
sprocket
Swivel arm
motor rotate
Compression
plate mount
DRMS2034.ai
Compression plate
motor lift
u NOTES u
- The IP address of the RU (RU IP ADDR) can only be set from PC-TOOL.
- As a rule, the IP addresses of the AWS and FTP server are used as the same
setting.
However, if the FTP server is centrally managed at a single location by distinguishing
between the IP addresses of the FTP server and AWS, you should gain full
understanding of this setup before use.
REFERENCE
The following abbreviations are used:
IP: Internet Protocol
HD: Hard Disk
FTP: File Transfer Protocol Used as a data sharing area.
AWS: Acquisition Workstation PC-Console
M/B: Motherboard Motherboard of the controller unit
AEC board
RTOS
RU application
EPROM
AEC firm LUT
AEC firm LUT
Machine-
specific data PTC firm LUT
CD-R
PTC board
EEPROM
Configuration setting PTC firm LUT
Log/history
DRMS2023.ai
HD (FTP server)
Image data
Log data
DRMS2024.ai
DRMS2026.ai
DRMS2025.ai
*1: The configuration data set in the AWS is written into the “HD (FTP server)” and then into
the “Compact Flash Card”.
*2: The RU application (RU-AP) is written from the "installation CD-ROM" into the "HD (FTP
server)" and "Compact Flash Card" in order named.
Log data *2
M/B SDRAM Never turn OFF the power of the controller unit while you restore the data.
Otherwise, the program on the Compact Flash Card gets damaged, and cannot
Log/history
restart.
DRMS2027.ai u NOTE u
Restored data takes effect when you turn the controller unit power OFF and then back
*1: The configuration data (CONFIGURATION) and the machine-specific data (INDIVIDUAL ON.
DATA) are transferred from the “Compact Flash Card”, the “HD (FTP server)” and the
“Backup/restore media” in order named.
*2: The log data (I/O trace log, trace log (control and correction), temperature regulation log
and operation log) are transferred from the “SDRAM” to the “HD (FTP server)” and the
“Backup/restore media” in order name. The ERROR LOG is transferred from the “HD (FTP
server)” to the “Backup/restore media”.
*1 Machine-specific data
Backup/
Machine-specific data
restore
media
DRMS2028.ai
*1: The configuration data (CONFIGURATION) and the machine-specific data (INDIVIDUAL
DATA) are transferred from the “media having the data backed up”, the “HD (FTP server)”
and the “Compact Flash Card” in order named.
Upon turning ON the power of the controller unit, the controller unit and the FPD are (6) Perform initial checks of the AEC firmware (communication check,
started up in the following sequence. version check and AEC sensor check).
(1) The M/B BIOS boots up, and sets various controller registers. (7) Perform initial checks of the PTC firmware (communication check,
version check, power failure check, and time setting).
(2) Upon completion of the BIOS processing, the software program is
loaded from the Compact Flash Card. (8) Check the PTC temperature adjustment conditions.
Controller unit
- Check if an error occurs while the controller unit power is turned OFF.
- Retrieve the temperature regulation log of the EEPROM of the PTC board from
Compact Flash Card M/B DRAM the controller unit, and store in the HD (FTP server).
IPL IPL
(9) Make network communication with the AWS.
RTOS
RTOS
Configuration setting
Network setting
(IP address, etc.)
RU application
*1
AWS
HD (FTP server)
Log file
DRMS2029.ai
*1: If a log file cannot be stored in the HD (FTP server) due to a failure in communication
or the like at the previous bootup, the file is transferred to the HD (FTP server).
(3) Upon completion of the startup and initialization of the software, the
controller unit and FPD hardware are initialized.
- Check mounting of the MPI51A board.
- Check mounting of the SND51B board.
- Check mounting of the CPS51B board (*).
- Initialize the CPS51B board (*) (load the micro program from the Compact
Flash Card to the DSP memory on the CPS board).
- Initialize the ASIC/ADC board/TRC board.
(*): CPS51C board for use in the FDR-1000DRLZ
CBS board
DRMS2017.ai
DRMS2021.ai
Part name Description
Part name Description Power supply Supplies the DC power to the exposure unit/controller unit.
CBS board Refer to "n CBS Board". MIF51A board Communicates with the X-ray source.
CBA board Refer to "n CBA Board". MPI51A board - Control interface between mechanics and sensors in the FPD.
- Interface for acquiring image data from the FPD.
MFI board Supplies the power to the X-ray tube.
CPS51B board (*) Image processing by DSP
XIF51A board I/F communication relay board between the X-ray device and the RU
CFI51A board
CPU51A board RU control
(motherboard)
PAT51A board Distributes the DC power to the FPD/controller unit.
NFB5A board Power supply noise filter
IND51A board Power indication lamp
(*): The FDR-1000DRSZ is equipped with two CPS51B boards, and the FDR-
1000DRLZ is equipped with three CPS51C boards.
LED (V1040)
CBS board
DRMS2047.ai
V1040/bit 2 Description
1 Initialization
2 Initialization
3 Initialization ready LED (V1008)
4 Reset
Switch (S2)
5 Exposure ready, 48 V and 400 V switches are ON.
6 Preparing
7 Exposure
C Failure
mAs measurement
Name Description Value 1: ON
Open/close switch for the mAs meter measurement 2: OFF (default)
circuit. Set to ON when the mAs value is to be
mA test switch measured in the CBA board by means of the OFF CBA board DRMS2048.ai
Actuator
NC
NO
COM Swivel arm elevation
FRAH1606.EPS lower-limit switch
Actuator
NC
NO
COM
FRAH1607.EPS
Swivel arm motor lift
DRMS2036.ai
Interlock target
Swivel arm motor rotate
Interlock switch
- Swivel arm rotation limit switch
DRMS2037.ai
FDR MS-1000
03.14.2011 03 Revision (FM5842) All pages
Service Manual
Terminals of power
supply cable
[Check] DRMS4259 ai
DRMS4229 ai
u INSTRUCTION u
Do not put items which might scratch the carbon cover (exposure stand), (such as
tools, screws and removed components). A scratched carbon cover might result in
abnormal images.
u NOTE u
When the mammographic stand is not used for a week or longer, carry out aging. The
aging can stabilize the output of X-rays and prevent deterioration of the tube.
{Operation Manual}
START A B C
(1) Swivel arm top cover (4) Swivel arm right-hand (7) FPD right-hand cover (10) Swivel arm left-hand
side cover (right-hand armrest) front cover
(1) (10)
(7)
(4)
(11)
(5)
(2)
(8)
(6) FPD left-hand cover (9) Compression unit left-hand (12) Swivel arm right-hand
(3) Swivel arm left-hand side cover (left-hand armrest) side cover front cover
(9)
(6)
(3) (12)
A B C END
DRMS4138.ai
#1
[Remove]
Face guard
DRMS4139.ai
DRMS4140.ai
Swivel arm
top cover
DRMS4141.ai
#1 [Disconnect]
Protective ground
wire terminal
Swivel arm lower cover
#2 DETAIL B
[Remove]
B M5x60 (x3)
#1 [Disconnect]
Protective ground
wire terminal
CHECK1 #3
Control panel [Release]
DRMS4304.ai Swivel arm
*: The cover which need be removed beforehand is colored. lower cover
Control panel
#4 #4
[Disconnect] [Disconnect]
Connector Connector
(MBL X1 W105) (MBL X1 W106)
DRMS4266.ai
Pin
X2_24V DC cable
Collimator rear
Collimator
DRMS4340.ai
DRMS4303.ai
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
DETAIL A
#1
[Remove]
M5x10 (x2)
W5 (x2)
DRMS4302.ai
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
Removal Procedures
(1) Remove the FPD lower cover.
A
DETAIL A
[Remove]
M3x6 (x3)
Bind screw
DRMS4181.ai
B
DETAIL A DETAIL B FPD left-hand cover
(left-hand armrest)
CHECK1
#2. Swivel arm
lower cover
#3
[Remove]
TP4x6 (x2)
#2
[Remove]
BR4x25
DRMS4301.ai
Carbon cover
#2. Swivel arm
lower cover #4. Swivel arm
right-hand side cover
No gap
DRMS4341.ai
#5. FPD lower cover
DRMS4300.ai
Check/Adjustment Procedures
Be sure to leave no gap between the FPD right- and left-hand covers (right- and left-
hand armrest) and the FPD carbon cover when putting on the FPD right- and left-hand
covers (right- and left-hand armrests). Otherwise, dust might enter the FPD, resulting
in abnormal images.
A
Carbon cover
#1 [Remove]
Q4x12/ Cover (right)
B W4
Carbon cover
#2
[Remove]
BR4x25
#3
[Remove] No gap
TP4x6 (x2)
DRMS4160.ai
DRMS4341.ai
chest wall plane. *: The covers which need be removed are colored.
Do not
touch the shock absorbers.
DRMS4351 ai
DETAIL A
Do not touch the
exposure plane.
Filter cover/
filter
B
A
DETAIL B
[Remove]
Do not touch the Bind screw (M3x6)
chest wall plane.
Do not DRMS4182.ai
TP3x12
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
Collar
Collar
#2
Collar [Remove]
Carbon cover TP3x12
DRMS4162.ai
l CHECK 1 Preparation
Before retaining the carbon cover, make sure that: You need remove the following covers before removing the compression unit left-hand
- Marked positions on the carbon cover and the FPD align with each other. side cover.
- There is no gap between the carbon cover and the FPD frame.
Swivel arm top cover
[Items to be prepared]
- Large nippers capable of cutting the SUS
- Hexagon socket head cap screw (Q4x12)
- *The new compression unit cover has no screws. The screw is necessary to retain
the right- and left-hand side covers.
Hexagon socket
head cap screw 2
(Q4x12)
[Action to cope with]
For retaining covers (Nos.4-11 and - Cut off the portion indicated in the figure below by the nippers or the like to remove
4-18
-12) the retaining screw part of the old compression unit. This modification allows the
Plain washer (W4) 2 new compression unit cover to be put on.
[State of interference]
- If the new compression unit cover is put on the old compression unit, the “retaining
screw part on the old compression unit” interferes with the “retaining screw part on
the new compression unit”, and the cover cannot be put as a result.
* Since the new compression unit cover is designed to retain the right and left
covers by together tightening the screws, no problem occurs if the retaining screw
part of the old compression unit is removed.
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
Reinstallation Procedures
#8. Carbon cover Reverse the removal procedures for reinstallation.
Swivel arm top cover (1) Remove the compression unit right-hand side cover.
Swivel arm lower cover
Compression unit
Swivel arm left-hand side cover right-hand side cover
Swivel arm right-hand side cover
DETAIL A
FPD lower cover #2
[Disconnect]
FPD left-hand cover (left-hand armrest) Protective
ground wire
FPD right-hand cover (right-hand armrest) terminal
A
Carbon cover
{MC:2.1.8_Carbon Cover} DETAIL B
#1. Swivel arm top cover #4
#3. Swivel arm [Disconnect]
left-hand side cover Connector
B
(MBL X1 W108)
#3 [Remove]
Protective cover
#2. Swivel arm C
lower cover #4. Swivel arm DETAIL C
right-hand side cover
#5
#6. FPD left-hand cover [Disconnect]
(left-hand armrest) #1
Connector
[Open]
(D904 X6
Compression unit
W143)
right-hand side cover
Reinstallation Procedures
#7. FPD right-hand cover
#5. FPD lower cover (right-hand armrest) Reverse the removal procedures for reinstallation.
DRMS4297.ai
#9. Compression
unit right-hand
side cover
#2. Swivel arm [Remove]
lower cover #4. Swivel arm right-
Q5x8
hand side cover
#6. FPD left-hand cover
(left-hand armrest)
DRMS4165.ai
Reinstallation Procedures
#8. Carbon cover Reverse the removal procedures for reinstallation.
Removal Procedures
(1) Remove the compression unit.
{MC:3.5_Compression Unit}
(2) Remove the compression unit front cover.
#2
[Remove]
Compression unit
front cover
#1
[Remove]
M4x6 (x2)
Compression unit
DRMS4097.ai
Reinstallation Procedures
(1) Reinstall the compression unit front cover.
Reverse the removal procedures for reinstallation.
(2) Reinstall the compression unit.
{MC:3.5_Compression Unit}
5
3
6
DETAIL A
DRMS4062.ai
#3
[Disconnect]
Connector (emergncy stop switch)
DETAIL A
[Remove]
Q5x8 (x4)
DRMS4261.ai
Elevation stand
top cover Reinstallation Procedures
Reverse the removal procedures for reinstallation.
DRMS4260.ai
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
#3 Elevation stand
[Disconnect]
Connector [Remove]
(emergncy stop switch) Q6x25 (x6)
Rear cover
DRMS4262.ai
Reinstallation Procedures
Reverse the removal procedures for reinstallation. DRMS4342.ai
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
#1
[Disconnect]
Connector Display cover
(DYS1 X1 W122) DRMS4263.ai
[Remove]
Reinstallation Procedures Base plate cover
DRMS4343.ai
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
[Loosen]
M4x6 (x14)
[Loosen]
Elevation stand M4x6 (x14)
inner cover (lower)
Elevation stand
inner cover (upper)
DRMS4264 ai DRMS4265.ai
DETAIL A
Cable retaining
bracket (large)
DETAIL A Cable retaining [Remove]
bracket (small) TP3x6 (x5)
[Remove]
TP3x6 (x6)
Cable (SND5)
Cable (TRC7)
DRMS4066.ai
DRMS4065.ai
#1 #4
[Disconnect] [Disconnect]
Connector Connector
(SND5) (CN10)
#3
[Disconnect]
Grid guide Grid Grid Grid guide
Connector
#2 (SND9) #5
[Disconnect] [Disconnect]
Connector Connector
(TRC7) (TRC1)
DRMS4067.ai
#1 #2
[Remove] [Remove]
Spacer x2 Spacer x2
TP3x6 (x2) TP3x6 (x2)
DRMS4070 ai
(1) Loosen the bolts on the upper surface of the FPD (inside the sheet
metal). B
REFERENCE
The bolts to be accessed in this procedure are located inside the sheet metal.
Use an allen wrench to tighten or loosen the bolts.
Allen wrench
Round hole of the sheet metal
A
FPD
DETAIL A DETAIL B
FPD
DRMS4103.ai
A #1 [Remove] #2 [Remove]
A [Tighten] Q6x25 Q6x25
TP3x6 (x6) SW6 SW6
W6 W6
Handle
Handle
DETAIL A [Tighten]
TP3x6 (x3)
FPD
FPD
DRMS4071.ai
Handle
DRMS4169 ai
[Remove]
FPD
#1
[Remove]
Clamp
DRMS4170.ai Joint
Joint
#2
[Release]
Coolant hose DRMS4171.ai
DETAIL A
Hold here
Do not touch the
Stand jig
exposure plane.
FPD
DETAIL A
B
A
DETAIL B
Joint
Exercise care not to break
the aluminum tape.
DETAIL A DETAIL B
Stand jig
DRMS4307.ai
#1 [Loosen] #2 [Loosen]
M6 M6
DRMS4101.ai
Route the thicker coolant hose outside when retaining the hoses. If the FPD is
mounted on the swivel arm while the hose is not correctly routed, the hose might
be pressed by the frame of the swivel arm to degrade the circulation function of
the coolant.
FPD
A
[Route]
DETAIL A Coolant hose
DETAIL A
#1 A
[Route]
Coolant hose Hose (thin)
Hose (thick)
Hose (thin)
DRMS4309.ai
Hose (thin)
#2
#2 [Retain] Hose (thick)
Hose (thin) [Retain] Clamp
Clamp
DRMS4308.ai
#1
DETAIL A DETAIL B [Verify]
No gap between the frames of the swivel arm and the FPD
B
FPD
#2 [Tighten] #2 [Tighten]
Q6x25/SW6/W6 Q6x25/SW6/W6
(Required torque value: 10 N・m) (Required torque value: 10 N・m)
DRMS4311.ai
Handle
DRMS4312.ai
(8) Fix the right- and left-hand sides of the FPD with the bolts. Hole of the sheet metal
Hexagon socket head cap screw
B
DRMS4168.ai
[Tighten] [Tighten]
M4x40 M4x40
A
FPD
DETAIL A DETAIL B
[Tighten]
[Tighten]
Q4x12
Q4x12
SW4/W4
SW4/W4
DRMS4313 ai
DRMS4314.ai
#2 [Push] #2 [Push]
#3 [Tighten] #3 [Tighten]
Grid assembly Grid assembly
M6 M6
#1 #1
Grid [Insert] [Insert]
assembly Grid guide Grid guide
A
B
#1 [Check] Half-punch of the FPD #1 [Check] Half-punch of the FPD
#2 #2
[Tighten] Grid Grid Spacer x2
[Tighten]
TP3x6 (x2) Spacer x2 TP3x6 (x2)
DRMS4320.ai
DETAIL A
[Tighten]
DETAIL A TP3x6 (x5)
Hold the copper foil.
[Connect] Bracket (large)
Cable
connector
Cable
(TPC1, CN10, SND9)
#5 #3 #2
(SND5) (SND9) (CN10) DRMS4317.ai
#4 #1
(TRC7) (TRC1)
DRMS4316 ai
A #1
[Tighten]
TP3x6 (x28)
DETAIL A
Bracket (small) Hold the copper foil.
Cable
(SND5)
Cable
(TRC7)
[Tighten]
NFI3
TP3x6 (x6)
DRMS4318.ai
NFI7
NFI3
NFI5
NFI5
MFI51A board
Do not connect the connector of the
FDR-1000DRSZ to the NFI7
(center connector).
DRMS4319.ai
(1) Check the grid wire. (7) Press the system OFF button on the control pad.
{IN:18.10.1_Checking the Grid Wire} (8) Turn OFF the circuit breaker of the controller unit.
(9) Turn OFF the circuit breaker of the distribution switchboard.
n Putting on the Swivel Arm Right- and Left-Front Covers
WARNINGS
REFERENCES
- You need put on the swivel arm right- and left-front covers before putting on the l Be sure to turn OFF the circuit breaker of the distribution switchboard to
carbon cover. The swivel arm right- and left-front covers cannot be put on after the avoid electric shock hazards.
carbon cover is put on. l Electrical charge (residual charge) accumulated by a capacitor might remain
- Set the swivel arm to the 0° position before putting on the swivel arm right- and left- in a circuit after the circuit breaker of the distribution switchboard is turned
front covers. OFF.
Be sure to proceed with operations after you are sure that the charge has
(1) Turn ON the circuit breaker of the distribution switchboard. been completely discharged, due to fears of electric shock hazards.
You can check the discharge by the following indications.
(2) Turn ON the circuit breaker of the controller unit. - “PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
(3) Press the system start button on the control pad. - LED on the MFI board: The yellow LED is unlit.
Swivel arm
right-hand front cover
Swivel arm
left-hand front cover
Exposure plane
of the FPD
DRMS4337.ai Do not wipe with a cloth moistened Do not wipe with a cloth moistened
with ethanol. with ethanol.
DRMS4321.ai
u NOTE u
Check the grid, if necessary, before reinstalling the carbon cover.
REFERENCE
Set the swivel arm to the 0° position before putting on the carbon cover.
A #2 [Connect]
Carbon cover
cable connector
Collar
#1 [Insert]
#4 [Tighten] #3 [Tighten]
Carbon cover
TP3x12 (x2) TP3x6 (x10) DRMS4323.ai
Carbon cover A
(2) Put the filter and the filter cover on the carbon cover.
DETAIL A
Carbon cover
GOOD NO GOOD
Carbon cover mark Carbon cover mark
Filter cover/filter
[Tighten]
Bind screw
(M3x6)
DRMS4324.ai
END
2. Installing the new FPD
Handle
[Remove]
TP3x6 (x6)
DETAIL A
#3 #1
[Mount] [Mount]
Cable retaining Cable retaining
bracket (small) bracket (large)
Handle
DRMS4173.ai
#4 #2
[Tighten] [Tighten]
TP3x6 (x6) TP3x6 (x5)
DRMS4174 ai
A #2 Collar #3
[Put] [Put]
TP3x12 TP3x6 (x10)
#1
[Put] Collar
FPD inner cover
TP3x6 (x28) Collar
#1
[Put]
Collar
TP3x12
DETAIL A #1 [Connect] Connectors Carbon cover
<FDR-1000DRSZ> <FDR-1000DRLZ> DRMS4176.ai
FAN x2 MFI51A board FAN x3
(6) Put on the filter and the filter cover.
[Tighten]
NFI3
Bind screw
NFI7 (M3x6) Filter/filter cover
NFI3
NFI5
NFI5
MFI51A board
DRMS4175 ai
DRMS4177.ai
(7) Remove the FPD from the stand jig, and temporarily place it on a
shock absorber or a corrugated fiberboard sheet.
Filter cover/
filter
(2) Temporarily place the FPD which has been taken out from the package
box.
{MC:3.1.5_Temporarily Installing the FPD}
TP3x12 (x2)
#3
[Pull out]
Carbon cover
DRMS4326.ai
FPD
DETAIL A [Tighten]
DETAIL A #1 [Disconnect] Connectors TP3x6 (x3)
<FDR-1000DRSZ> <FDR-1000DRLZ>
FPD
FAN x2 MFI51A board FAN x3
NFI3
NFI7
NFI5
Handle
DRMS4338.ai
MFI51A board
NFI5 NFI3
DRMS4430.ai
- When temporarily installing the FPD, use the dedicated stand jig.
DRMS4332.ai
DRMS4331.ai
#2
[Take out]
FPD FPD package
Shock absorber or
corrugated fiberboard sheet
DRMS4334.ai
#1
[Break]
4-corner of the package
#3
[Place]
FPD
(1) Release the coolant hoses routed to the rear of the FPD downwardly
from the joint with the stand jig.
u INSTRUCTION u
You need release the coolant hoses from the joint with the stand jig before
mounting the FPD on the stand jig. Otherwise, the coolant hoses might be caught
between the FPD joint and the stand jig, resulting in damage.
DETAIL A
Joint
Joint
#1
[Remove] Exercise care not to have
Clamp the hoses caught.
DRMS4336.ai
Joint
Joint
#2
[Release]
Coolant hose DRMS4171.ai
[Install]
Compression unit
left-hand side cover
DRMS4166.ai
Inner cover
A
#3
[Remove]
TP3x6 (x28)
<FDR-1000DRSZ> <FDR-1000DRLZ>
FAN x2 MFI51A board FAN x3
NFI3
NFI7
NFI5
MFI51A board
NFI5 NFI3
DRMS4063.ai
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
#1 [Unclamp]
Clamp DRMS4038.ai
n Check/Adjustment Procedures
l CHECK 1
Check that the board cooling fan is attached in its correct orientation.
Engraved
mark
DRMS4039.ai
DRMS4295.ai
DETAIL A
[Insert]
A
A Screwdriver
Temperature
Screw adjustment unit
cable
Screw
Temperature adjustment unit
DETAIL A
Temperature adjustment unit
DRMS4146.ai
A
Temperature
Temperature adjustment unit
adjustment unit
DRMS4147.ai
Stand jig
DETAIL A [Install]
Temperature
adjustment unit
Stand jig
DRMS4339 ai
B
Temperature
adjustment unit cable
DETAIL A DETAIL B
#2
[Connect] Temperature adjustment
Relay cable connector unit cable connector
#1
NFI51A board FPD [Connect]
Relay cable connector
DRMS4150.ai
Front Rear
#1
[Disconnect]
Closure hose
A370Y100306
DRMS8426.ai
DRMS8427.ai
REFERENCE
Connect the FPD coolant hoses (thick and thin) to the temperature adjustment
unit and water supply jig in the following procedures.
#1
[Insert]
Coolant hose
(thick)
FPD
A
DETAIL A
Coolant hose
(thin)
Closure hose
DRMS8251.ai
#2
[Remove]
Stand jig
FPD
DRMS4157.ai
Temperature
adjustment unit cable
DRMS4156 ai
It is recommended that you insert a screwdriver into a hole beside the screw for
retaining the temperature adjustment unit before installing the unit. Inserting the NFI51A board
screwdriver can prevent the temperature adjustment unit from dropping. Remove
the screwdriver after completion of the operation. FPD
DETAIL A
[Insert] A
A Screwdriver
Screw Temperature
adjustment unit
cable
Screw
Temperature adjustment unit
DETAIL A
Temperature adjustment unit
DRMS4146.ai
[Tighten]
TP3x12 (x3)
[Connect]
Connector
NFI1
FPD
Temperature DRMS4252 ai
CAUTION WARNING
The collimator is attached via a snap-fit jig. The collimator, if rotated clockwise Electrical charge (residual charge) accumulated by a capacitor might remain in
40 degrees, may drop. a circuit after the circuit breaker of the distribution switchboard is turned OFF.
Make sure that the collimator is kept at such position that it will not drop. Be sure to proceed with operations after you are sure that the charge has been
completely discharged, due to fears of electric shock hazards.
You can check the discharge by the following indications.
INSTRUCTIONS - “PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
- Always adjust the irradiation fields (X-ray irradiation field and light irradiation field) - LED on the MFI board: The yellow LED is unlit.
after the collimator is dismounted and then again mounted.
- After the collimator is removed, cover the opening with a paper towel or the like to (7) Remove the swivel arm lower cover.
prevent dust from entering the collimator.
{MC:2.1.2_Swivel Arm Lower Cover}
(8) Disconnect the cable connectors (X2 and X3) and the protective
Removal Procedures ground wire.
NOTE [Disconnect]
Protective ground wire
Replace the collimator as a whole unit. You cannot replace a bulb or a filter disk, for
example.
(1) Rotate the swivel arm to the 0° position, and move it up or down to an
easy-to-work height.
(2) Turn OFF the power of the AWS.
(3) Press the system OFF button on the control pad. [Disconnect]
Connector (X3)
(4) Turn OFF the circuit breaker of the controller unit.
(5) Turn OFF the circuit breaker of the distribution switchboard. [Disconnect]
Connector (X2)
WARNING Collimator
To avoid electric shock hazards, be sure to turn OFF the circuit breaker on the DRMS4209.ai
distribution switchboard.
[Remove]
Collimator
DRMS4211.ai
{IN:28.5.1_Checking the Collimator Mounting Angle} See below for the standard of the collimator mounting angle.
u INSTRUCTION u DETAIL A
Always adjust the irradiation fields (X-ray irradiation field and light irradiation field) |a-b| should be ≤ 1 mm
Exposure frame
after checking the collimator mounting angle if the collimator has been dismounted
and then again mounted. a b
Exposure stand
Exposure frame
DRMS4425.ai
Copper ring
#2
[Adjust]
Collimator
DRMS4426.ai
6. FPD calibration
END
n Checking the Exposure Count (4) Press the system OFF button on the control pad.
(5) Turn OFF the circuit breaker of the controller unit.
NOTE
(6) Turn OFF the circuit breaker of the distribution switchboard.
Check the exposure count before removing the single tank unit, and record it.
WARNING
(1) Start up the Zigma Webserver.
To avoid electric shock hazards, be sure to turn OFF the circuit breaker on the
{MU:2._Zigma Webserver_n Starting Up and Exiting the Zigma Webserver} distribution switchboard.
(2) Click “Generator”.
(3) Check the Exposure Counter value. (7) Wait for approx. 2 to 7 minutes until the machine is discharged.
Record the indicated value. WARNING
Electrical charge (residual charge) accumulated by a capacitor might remain in
a circuit after the circuit breaker of the distribution switchboard is turned OFF.
Be sure to proceed with operations after you are sure that the charge has been
completely discharged, due to fears of electric shock hazards.
You can check the discharge by the following indications.
- “PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
- LED on the MFI board: The yellow LED is unlit.
DRMS4051.ai
MFI board
DRMS4212.ai
[Disconnect]
Q6x12
W6 CBA board
#2
[Remove] #3
N3 [Remove]
Cable tie
DRMS4214.ai
DRMS4213.ai
#2
[Remove]
Cover retaining Q6x12
bracket
#1
[Remove]
[Mark] Q6x36 (special screw)
Bracket attachment
DRMS4216.ai
DRMS4215.ai
[Remove]
M4x6 (x4) DRMS4217.ai
[Remove]
Single tank unit
DRMS4218.ai
Cable
Single tank unit
DETAIL A
DRMS4219.ai
#2
[Press]
Positioning ring
DRMS4221.ai
n Replacement Flowchart
15. Checking for image problems
START
END
1. Recording the exposure count
4. Changing the serial No. of the single tank unit registered in the
AWS
7. FPD calibration
Serial No.
DRMS4370.ai
distribution switchboard.
Compression
unit
A
CBA (D905) board
[Remove]
Q6x12 (x4)
DETAIL A CBA (D905) board
W6 (x4) DRMS4057.ai
[Disconnect] [Disconnect]
Cable connector Cable connector
(CBA D905 X5) (CBA D905 X18)
DRMS4056.ai
INSTRUCTION
Disconnect also the cable connector X7 when the biopsy positioner kit or the
stereo exposure kit is mounted.
#2
[Tighten]
Q6x12 (x4)
W6 (x4)
DETAIL A
Compression plate
#1
[Butt] End surface
of the compression
plate and end
surface of the
carbon cover
Carbon cover
DRMS4222.ai
#2
[Tighten]
M4x6 (x7)
Inner cover
A
CBA (D905) board
#1
[Route]
Cable connector
(MBL X1 W107)
DRMS4352.ai
INSTRUCTION
Disconnect also the cable connector X7 when the biopsy positioner kit or the
stereo exposure kit is mounted.
(7) Press the system start button on the control pad. Once remove the carbon cover to attach all the components. The following
components cannot be attached unless the carbon cover is removed.
(8) Turn ON the power of the AWS. - Swivel arm front covers right and left
(9) Record a uniform image and check it. - Compression unit side covers right and left
{IN:27.1._Checking for Image Problems}
(16) Put on all of the covers.
INSTRUCTION
Take the removal procedures (9) and those follow reversely for reinstallation.
Always check that the compression plate is not recorded in the image. If the
(17) Check the compression unit.
compression plate is recorded in the image, the compression unit might tilt.
Reposition the unit. {IN:28.4_Checking the Compression Unit}
[Compress]
Acrylic board Compression Unit window
DRMS4282.ai
DRMS4225.ai
(6) #1
[Input]
“-10 to 10”
(7) #1
CNU Parameters window [Click]
“Set”
DRMS4227.ai
DETAIL A
Frame
[Install]
Compression
plate
DRMS4344 ai
#1 #1
[Open] [Open]
Lock Lock
#2
Plastic plate part
[Pull out]
Do not open the locks excessively. Plastic plate part
DRMS4345.ai
n Reinstallation procedures
Reverse the removal procedures for reinstallation.
CAUTION WARNING
When servicing a board, be sure to wear a wristband to ground your body. If Electrical charge (residual charge) accumulated by a capacitor might remain in
your body is not grounded, static buildup on your body may damage electronic a circuit after the circuit breaker of the distribution switchboard is turned OFF.
components on the board. Be sure to proceed with operations after you are sure that the charge has been
completely discharged, due to fears of electric shock hazards.
You can check the discharge by the following indications.
INSTRUCTIONS
- “PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
- Be sure to back up the configuration data of the mammographic stand when a board - LED on the MFI board: The yellow LED is unlit.
is to be replaced. Also restore the configuration data of the mammographic stand
after the board is replaced. If the data is not backed up/restore, adjustments on the
generator, compression unit, collimator and grid must be carried out. (8) Remove the swivel arm left-hand front cover.
- Record the exposure count before removing the board when the board is to be {MC:2.1.10_Swivel Arm Left-Hand Front Cover}
replaced. The exposure count is not recorded in the configuration data.
(9) Remove the swivel arm right-hand front cover.
{MC:2.1.12_Swivel Arm Right-Hand Front Cover}
Removal Procedures (10) Remove the inner cover (CBA board).
(1) Back up the configuration data of the mammographic stand. #1
{MC:8.1.1_Backing Up the Configuration Data of the Mammographic Stand} [Remove]
M4x6 (x2)
(2) Rotate the swivel arm to the 0° position, and move it up or down to an
easy-to-work height. #2
(3) Turn OFF the power of the AWS. [Loosen]
M4x6 (x7)
(4) Press the system OFF button on the control pad.
(5) Turn OFF the circuit breaker of the controller unit.
(6) Turn OFF the circuit breaker of the distribution switchboard.
WARNING
To avoid electric shock hazards, be sure to turn OFF the circuit breaker on the
distribution switchboard.
Inner cover DRMS4048.ai
CBA board
A
[Remove]
CBA (D905) board M4x6 (x7)
X13
DRMS4050.ai
X15
n Reinstallation Procedures
X9 (1) Put on the CBA board (D900).
Reverse the removal procedures for reinstallation.
(2) Re-write the firmware of the mammographic stand with the same
X4 X11 version as that before replacement.
u NOTE u
The procedures are similar to those for updating the firmware to be referred to.
Do not update now, however.
{MT:3.11_The 15-74 Error Occurs}
X2
The configuration data backed up in the procedures is also restored.
X1 X16
X3 X5 X18 X17
DRMS4049.ai
INSTRUCTION
Disconnect also the cable connector X7 when the biopsy positioner kit or the
stereo exposure kit is mounted.
[Click]
“Out”
DRMS4230.ai
[Remove]
Q6x10 (x2)
DETAIL A
DETAIL A
[Remove]
Q6x10 (x2)
[Remove]
Q6x10 (x2) DRMS4233.ai
DRMS4232.ai
B
Spacer x4
Grid assembly
Grid guide
A
DRMS4234.ai
FPD exposure
(14) Disconnect the cable connectors (X16 and X17) from the CBA board plane
(D905).
CBA board
DETAIL A DETAIL B
[Disconnect]
Connector (X16)
[Disconnect]
Connector (X17)
[Remove] [Remove]
M6 M6
DRMS4235.ai
DRMS4231.ai
Grid assembly
#1 #1
B [Insert] Grid Grid [Insert]
Grid guide Grid guide
FPD
A
DRMS4362.ai
#2 [Push] #2 [Push]
Grid assembly Grid assembly
DRMS4245.ai
1 2 3 4 5 6 7 8
Remove the compression unit and set the DIP switch (S1004-8) on the CBA board to
ON before adjusting the grid position.
Upon completion of the check and adjustment of the grid position, be sure to return
the DIP switch (S1004-8) on the CBA board to OFF.
ON
n Preparation
(1) Remove the compression unit.
{MC:3.5_Compression Unit}
DRMS4236.ai
NOTES
- Make preparations before adjusting the grid sensor.
{MC:3.7.4_Adjusting the Grid Sensor_n Preparation}
- Be sure to return the CBA board DIP switch (S1004-8) to OFF after completion of
the grid sensor adjustment.
Table window
[Click] [Adjust]
“Out” Grid park position
sensor
DRMS4242.ai
DRMS4230.ai
(5) Set the CBA board DIP switch (S1004-8) to OFF.
NOTES INSTRUCTION
- Make preparations before adjusting the grid sensor. Adjust the position of the grid IN sensor so that the distance between the tip of
the grid and the FPD end surface is 2 ± 1 mm.
{MC:3.7.4_Adjusting the Grid IN Sensor_n Preparation}
- Be sure to return the CBA board DIP switch (S1004-8) to OFF after completion of Grid end surface
the grid sensor adjustment. FPD end
surface
(1) Start up the Zigma Webserver.
{MU:2._Zigma Webserver_n Starting Up and Exiting the Zigma Webserver}
(2) Click [Table] and enter the Table window.
(3) Click [In] in “Move Grid”.
The grid moves to the grid IN position.
Table window
[Check]
Distance
2±1 mm
DRMS4239.ai
[Click]
“In”
[Adjust]
DRMS4238.ai
Grid IN sensor
DRMS4243.ai
NOTES NOTE
- Make preparations before adjusting the grid sensor. Mark the current positions of the grid oscillation sensors (right and left) with a
{MC:3.7.4_Adjusting the Grid Sensor_n Preparation} pencil or the like before adjustment, so that you can see their original positions.
- Be sure to return the CBA board DIP switch (S1004-8) to OFF after completion of
the grid sensor adjustment. INSTRUCTION
- Check the adjustment of the grid oscillation sensors (right and left) while the carbon
Adjust the position of the grid oscillation sensors (right and left) so that the grid
cover is put on.
oscillation times (right and left) indicated on the display come within a range of
(1) Start up the Zigma Webserver. 200 to 250 msec.
{MU:2._Zigma Webserver_n Starting Up and Exiting the Zigma Webserver} <Display example>
(2) Click [Table] and enter the Table window. Display
(3) Click [Start] in “Adjust grid oscillation sensors”.
Grid Times
Transfer
The grid starts to oscillate.
Table window
729 714
Reverse
209 206
[Check] [Check]
Grid oscillation time Grid oscillation time
(left) (right) DRMS4241.ai
[Adjust]
[Click] Grid oscillator sensor
“Start” (left)
DRMS4240.ai
[Adjust]
Grid oscillator sensor
(right)
DRMS4244.ai
INSTRUCTION #1
Be sure to carry out generator adjustment after the MFI board (D910) is replaced. [Loosen]
M4x6 (x10)
{MC:9._Generator Adjustment}
To avoid electric shock hazards, be sure to turn OFF the circuit breaker on the
distribution switchboard.
- “PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
- LED on the MFI board: The yellow LED is unlit.
#1
[Remove]
M4x12 (x4)
DRMS4053.ai
n Reinstallation Procedures
Reverse the removal procedures for reinstallation.
INSTRUCTION
Be sure to carry out generator adjustment after the MFI board (D910) is replaced.
{MC:9._Generator Adjustment}
n Removal Procedures
(1) Rotate the swivel arm to the 0° position, and move it up or down to an
easy-to-work height.
(2) Turn OFF the power of the AWS.
(3) Press the system OFF button on the control pad. #1
(4) Turn OFF the circuit breaker of the controller unit. [Disconnect] #4
Connector (XIF3) [Remove]
(5) Turn OFF the circuit breaker of the distribution switchboard. TP3x6 (x3)
WARNING #2
[Disconnect]
To avoid electric shock hazards, be sure to turn OFF the circuit breaker on the Connector (XIF1) #3
distribution switchboard. [Disconnect]
Connector (XIF2)
(6) Wait for approx. 2 to 7 minutes until the machine is discharged. DRMS4100 ai
(9) Disconnect the cable connector (X8) from the MFI board (D910).
(6) Wait for approx. 2 to 7 minutes until the machine is discharged.
WARNING
Electrical charge (residual charge) accumulated by a capacitor might remain in
a circuit after the circuit breaker of the distribution switchboard is turned OFF. [Disconnect]
Be sure to proceed with operations after you are sure that the charge has been Connector
completely discharged, due to fears of electric shock hazards. (X8)
You can check the discharge by the following indications.
- “PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
- LED on the MFI board: The yellow LED is unlit.
DRMS4268.ai
DRMS4247.ai
n Reinstallation Procedures
Fan (for cooling the
Reverse the removal procedures for reinstallation. single tank unit)
Label surface
DRMS4248.ai
(5) Wait for approx. 2 to 7 minutes until the machine is discharged. CDF LIFT CDF LIFT
X2 W126 X2 W126
WARNING DRMS4040.ai
DRMS4041.ai
CDF X3 When the motor inverter rotate is to be replaced with a new one, you need to remove
ROTATION the retaining brackets from the motor inverter rotate which has been removed, and
reposition them on the new motor inverter rotate.
[Reposition]
Bracket
DETAIL B
CDF ROT
X1 W129 B
#2
[Remove]
M4x12
Motor inverter rotate
#1
[Loosen]
M4x12
DRMS4044.ai
CAUTION WARNING
When servicing a board, be sure to wear a wristband to ground your body. Electrical charge (residual charge) accumulated by a capacitor might remain in
If your body is not grounded, static buildup on your body may damage a circuit after the circuit breaker of the distribution switchboard is turned OFF.
electronic components on the board. Be sure to proceed with operations after you are sure that the charge has been
completely discharged, due to fears of electric shock hazards.
You can check the discharge by the following indications.
INSTRUCTIONS
- “PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
- Be sure to back up the configuration data of the mammographic stand when a board - LED on the MFI board: The yellow LED is unlit.
is to be replaced. Also restore the configuration data of the mammographic stand
after the board is replaced. If the data is not backed up/restore, adjustments on the
generator, compression unit, collimator and grid must be carried out. (7) Remove the elevation stand inner cover (lower).
- Record the exposure count before removing the board when the board is to be {MC:2.2.7_Elevation Stand Inner Cover (Lower)}
replaced. The exposure count is not recorded in the configuration data. (8) Disconnect the cable connectors (17 pcs.) from the CBS board (D900).
To avoid electric shock hazards, be sure to turn OFF the circuit breaker on the X22
distribution switchboard.
CBS board
X17
X16 X11 X1 X3
DRMS4046.ai
#2 Check/Adjustment Procedures
[Remove]
M4x6 (x6)
l CHECK 1
Check that the cable connector (X5) is not connected with the cable connector (X24)
by mistake.
DRMS4047.ai
X5
X24
DRMS4363.ai
CAUTIONS WARNING
- If you improperly replace the battery, an explosion may occur. Electrical charge (residual charge) accumulated by a capacitor might remain in
Use our designated lithium battery when replacing the battery. Be sure to a circuit after the circuit breaker of the distribution switchboard is turned OFF.
return the old lithium battery to the parts center. Be sure to proceed with operations after you are sure that the charge has been
completely discharged, due to fears of electric shock hazards.
- When servicing a board, be sure to wear a wristband to ground your body.
If your body is not grounded, static buildup on your body may damage You can check the discharge by the following indications.
electronic components on the board. - “PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
- LED on the MFI board: The yellow LED is unlit.
u INSTRUCTIONS u
(7) Remove the elevation stand inner cover (lower).
- Be sure to back up the configuration data of the mammographic stand before
removing the lithium battery. Also restore the configuration data of the {MC:2.2.7_Elevation Stand Inner Cover (Lower)}
mammographic stand after the battery is detached/attached. If the data is not (8) Remove the lithium battery from the CBS board.
backed up, adjustments on the generator, compression unit, collimator and grid
must be carried out.
- When the lithium battery is removed, the exposure count is reset. Be sure to
record the exposure count before removing the battery. The exposure count is not
recorded in the configuration data of the mammographic stand.
(4) Turn OFF the circuit breaker of the controller unit. Reinstallation Procedures
(5) Turn OFF the circuit breaker of the distribution switchboard. NOTE
WARNING Model: CR2032
FUJIFILM Part No.: W11G3082146
To avoid electric shock hazards, be sure to turn OFF the circuit breaker on the
distribution switchboard. (1) Reinstall the lithium battery.
Reverse the removal procedures for reinstallation.
(2) Restore the configuration data of the mammographic stand.
{MC:8.1.2_Restoring the Configuration Data of the Mammographic Stand}
Removal Procedures
(1) Turn OFF the power of the AWS.
(2) Press the system OFF button on the control pad.
(3) Turn OFF the circuit breaker of the controller unit.
(4) Turn OFF the circuit breaker of the distribution switchboard.
WARNING
To avoid electric shock hazards, be sure to turn OFF the circuit breaker on the
distribution switchboard. A
(5) Wait for approx. 2 to 7 minutes until the machine is discharged. DETAIL A
WARNING Terminal block cover
[Remove]
(6) Remove the elevation stand inner cover (lower). TP3x6
{MC:2.2.7_Elevation Stand Inner Cover (Lower)} DRMS4250.ai
B Power supply
unit (PSU)
DETAIL A DETAIL B CHECK1
X3 X4 X5 X2
N
L1 DRMS4356.ai
E E
X7 X6
DRMS4054.ai
#2
[Remove]
Power supply unit
(PSU)
DRMS4055.ai
#2 #2 DETAIL B L1
Terminal
[Connect] block
Ground cable (black) N (white)
(Mammographic stand) DRMS4355 ai
B
W6 (x3) Na6
FG
Ground cable
DRMS4251 ai
[Tighten]
DETAIL A TP3x6
DRMS4358.ai
X3
X6
X2
X7
X4 X5
DRMS4357.ai
Elevation stand
Power supply
cable
Power supply
unit (PSU)
DETAIL A
[Verify]
The power supply cable is surely fixed and
FPD signal cable does not shift in position when pulled.
FPD power
supply cable
18 mm or less
(recommended value)
DRMS4359 ai
REFERENCE CAUTION
The removal/reinstallation procedures for the emergency stop switch are common When servicing a board, be sure to wear a wristband to ground your body.
both to the right and left switches. The procedures for the left switch are mentioned If your body is not grounded, static buildup on your body may damage
below as an example. electronic components on the board.
#2
Reinstallation Procedures [Remove]
Reverse the removal procedures for reinstallation. M2.5x6 (x4)
(self-tapping screw)
Display cover
#1 assembly
[Disconnect]
Connector Display board (D903)
(DYS1 X1 W122) DRMS4076 ai
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
[Disconnect]
Connector
D900 (X17)
DRMS4078 ai
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
Reinstallation Procedures
Electrical charge (residual charge) accumulated by a capacitor might remain in
a circuit after the circuit breaker of the distribution switchboard is turned OFF. Reverse the removal procedures for reinstallation.
Be sure to proceed with operations after you are sure that the charge has been REFERENCE
completely discharged, due to fears of electric shock hazards.
The cable connectors of the two foot switches can be connected with either the
You can check the discharge by the following indications. connector X22 or the connector X23 of the CBS board.
- “PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
- LED on the MFI board: The yellow LED is unlit.
DRMS4456.ai
(3) Remove the stand top and side covers. Attach the emergency stop
short-circuit cables.
DRMS4455.ai
DRMS4457.ai
DRMS4458.ai DRMS4460.ai
(5) Turn on the stand and AWS and move the swivel arm its lowest limit. (7) Remove the swivel arm left and right covers.
DRMS4459.ai DRMS4461 ai
DRMS4464 ai
#4 Rotate the swivel arm back to 0 degrees again and turn off the stand and
DRMS4462 ai
room power supply breakers.
#2 Fix two bolts and washers on left side by hand.
(9) Open the elevation stand inner covers (upper and lower).
DRMS4463 ai
DRMS4465 ai
DRMS4468.ai
(13) Disconnect the motor control cable from connector X26 on the CBS
board.
DETAIL A
DRMS4466.ai
(11) Mark the holes where the cable harness containing the motor cables
(see the right picture) is attached to the stand with a marker pen and
then cut and remove all the cable ties holding this harness in place.
DRMS4469.ai
DRMS4467.ai
DRMS4470.ai
(15) Pay attention to the other cables (See right picture) and remove both
power and control motor cables.
DRMS4472 ai
#1 Mark the desired area with a marker pen. Cut the frame to a depth of around
10 mm using a hacksaw. (Cut the top then the bottom sides.)
DRMS4471.ai
u NOTE u
If it is difficult, remove the top shutter guides again after labeling correctly with a
marker pen TL and TR and try again.
DRMS4473.ai
DRMS4474.ai
DRMS4476.ai
DETAIL A
DRMS4475.ai
DRMS4477 ai
DRMS4480.ai
#3 Once the bolt and washer have been removed push the spindle and sprocket
as far as possible and out of the motor housing toward the front of the stand.
REFERENCE
DRMS4478 ai
This will give access to the four bolts holding the motor in place.
(3) Remove the motor. #4 Remove the four bolts holding the motor in place.
#1 Fill the bottom of the stand and underneath the lift motor with the bubble
wrap, which came with the motor. u NOTE u
REFERENCE Three of these bolts can be accessed from the backside of the stand and one
from the front.
This is to help hold the motor in place when the fixing bolts are removed later.
DRMS4481 ai
DRMS4479 ai
DRMS4482 ai
(2) Place the bubble wrap across the bottom of the stand making sure
you have access to the cables at the right hand back side of the stand.
DRMS4483.ai
DRMS4485 ai
#2 Replace the four bolts then slide the spindle and sprocket fully through the
motor housing.
DRMS4486.ai
(4) Loosely attach a cable tie round the new motor cables and the original
cable harness to hold the new cables in place temporally while the DRMS4488 ai
motor is fitted. #3 Fix the spindle back in place with the bolt and washer.
DRMS4487.ai DRMS4489.ai
DRMS4492 ai
(2) Overtighten slightly passed the line then back off the bolt.
DRMS4490.ai
(7) Re-position the chain tensioning spring device back into place. Then REFERENCE
replace the two fixing bolts securing the chain and tensioning spring
This procedure makes the tensioning spring device sit exactly back on the lines
device back into the original positions. in the original position and at the original tension.
u NOTE u
Be careful to keep the cable ties tight at all times.
DRMS4493.ai
DRMS4491.ai
(8) When the bolts are fully secured and tightened carefully, cut the cable
ties and remove from around the tensioning spring device.
DRMS4494.ai DRMS4496.ai
(4) Replace the acrylic chain guard back onto the front of the motor
positioning bracket.
REFERENCE
This is easier done by locating the bottom hole first and then fixing the top bolt.
DRMS4495.ai
DRMS4499.ai
#6 Remove the two locking bolts on the right hand side of the swivel arm.
DRMS4497 ai
#2 Temporally reconnect the motor control cable to the CBS board on connector
X26.
DRMS4500.ai
DRMS4503 ai
DRMS4501.ai
#4 Make sure the cable lengths are correct and the earth strap will fit on to the
motor power cable exposed braid. Firmly secure the cables in place with
REFERENCE cable ties when the correct positions have been found.
If required hold the cable harness caterpillar track (E-Chain) out of the way in
order to fix the cables correctly.
DRMS4504.ai
DRMS4502.ai
DRMS4505 ai DRMS4507 ai
DRMS4506 ai
DRMS4508 ai
DRMS4509 ai DRMS4511 ai
#3 Replace the swivel arm top and bottom (tube head) covers.
DRMS4510 ai
(1) Cover the detector with bubble wrap and then place the old motor on (5) Replace the left and right hand side covers and the stand top cover.
it. (6) Restart the stand and perform final quality checks.
REFERENCE
To test to prove the motor stability, the old motor is used as a test weight to
simulate additional load (15 kgf=150 N).
DRMS4512.ai
(2) Turn on the stand from the breaker and the console control pad. Make
sure the CU and AWS are powered on.
(3) Conduct the test below.
Test procedure;
#1 Move the swivel arm to its lowest limit position.
#2 Wait 60 sec.
#3 Press the lift up button for 1 sec.
#4 Release the lift button then immediately repress the lift up button for a further
1 sec. Repeat steps #3. and #4. in the above procedure four times.
#5 Lower the swivel arm back to its lowest limit and repeat from step #2.
#6 Repeat this procedure 5 times.
u INSTRUCTION u
Please remember to wait 60 sec between each repetition and check the AWS for
errors.
#2
[Remove]
Cover
DETAIL A
GOOD NO GOOD
DRMS4001.ai
DRMS4360.ai
#3
[Remove]
MIF51A board
assembly
#1 #2
[Disconnect] [Remove]
Connector (NIF4) TP3x6 (x2) DRMS4003.ai
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
CAUTION CAUTION
When servicing a board, be sure to wear a wristband to ground your body. When servicing a board, be sure to wear a wristband to ground your body.
If your body is not grounded, static buildup on your body may damage If your body is not grounded, static buildup on your body may damage
electronic components on the board. electronic components on the board.
#3 #3
[Remove] [Remove]
MPI51A board CPS51B/CPS51C
assembly board assembly
#1 DRMS4005.ai
[Disconnect]
Connector (MPI2) Reinstallation Procedures
DRMS4004.ai
CAUTION
#1
When servicing a board, be sure to wear a wristband to ground your body. #1 [Unclamp]
If your body is not grounded, static buildup on your body may damage [Unclamp] Clamp
electronic components on the board. Clamp
Removal Procedures
#2
(1) Remove the controller unit cover and the inner cover.
[Remove]
{MC:5.1_Controller Unit Covers} TP3x6 (x4) DRMS4008.ai
(2) Disconnect the cable connectors. (4) Remove the PAT51A board.
#1
A [Remove]
TP3x6 (x6)
PAT6 PAT2
PAT9 PAT5
DRMS4007.ai
CAUTION CAUTION
When servicing a board, be sure to wear a wristband to ground your body. When servicing a board, be sure to wear a wristband to ground your body.
If your body is not grounded, static buildup on your body may damage If your body is not grounded, static buildup on your body may damage
electronic components on the board. electronic components on the board.
REFERENCE
Removal Procedures
When the Compact Flash Card is replaced with a new one, install the RU software
(1) Remove the controller unit cover and the inner cover. and restore the data.
{MC:5.1_Controller Unit Covers}
(2) Remove the Compact Flash Card. Removal Procedures
(3) Remove the CFI51A board (memory board). (1) Remove the controller unit cover and the inner cover.
(3) #4 {MC:5.1_Controller Unit Covers}
[Disconnect]
Connector (CFI1) (2) Remove the Compact Flash Card.
(3) #2 (3) #3
[Remove]
Reinstallation Procedures
[Remove]
TP3x6 (x3) CFI51A board Reverse the removal procedures for reinstallation.
(memory board)
u INSTRUCTION u
When the Compact Flash Card is replaced with a brand-new one, perform the
(2) #1 following procedures:
[Remove]
l Installing the RU software
Compact Flash
Card {IN:21._Installing the RU Software}
l Restoring data
(3) #1 - Machine-specific data
[Disconnect] - Configuration data
Connector (CFI3)
DRMS4006.ai - AEC file
{MC:8.2.2_Restoring the RU Data}
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
Removal Procedures
#2
(1) Remove the controller unit cover and the inner cover. [Remove]
{MC:5.1_Controller Unit Covers} TP3x6 (x9)
{MC:5.2_MIF51A Board}
(3) Remove the MPI51A board assembly.
{MC:5.3_MPI51A Board}
(4) Remove the CPS51B/CPS51C board assembly.
{MC:5.4_CPS51B/CPS51C Board}
(5) Remove the PAT51A board assembly.
{MC:5.5_PAT51A Board}
Reposition the
memory board.
UP
DRMS4011.ai
DRMS4012.ai
To avoid electric shock hazards, be sure to turn OFF the circuit breaker on the
distribution switchboard.
CAUTION
When servicing a board, be sure to wear a wristband to ground your body.
If your body is not grounded, static buildup on your body may damage
electronic components on the board.
Removal Procedures
(1) Remove the controller unit cover and the inner cover.
{MC:5.1_Controller Unit Covers}
(2) Remove the PAT51A board assembly.
{MC:5.5_PAT51A Board}
(3) Remove the memory board.
[Remove]
Memory board
(PRIMARY) (*)
DRMS4080.ai
*: For the FDR-1000DRLZ, remove the two memory boards, PRIMARY and
SECONDARY.
CAUTION CAUTION
When servicing a board, be sure to wear a wristband to ground your body. When servicing a board, be sure to wear a wristband to ground your body.
If your body is not grounded, static buildup on your body may damage If your body is not grounded, static buildup on your body may damage
electronic components on the board. electronic components on the board.
#2
[Remove]
TP3x6 (x2)
#1
DRMS4081.ai
#1
Reinstallation Procedures [Disconnect] [Disconnect]
Connector (NFB1) Connector (NFB2) DRMS4022.ai
Reverse the removal procedures for reinstallation.
Reinstallation Procedures
Reverse the removal procedures for reinstallation.
(1) Remove the controller unit cover and the inner cover. #1
#3 [Unclamp]
{MC:5.1_Controller Unit Covers}
[Remove] Clamp
(2) Remove the controller unit cooling fan assembly. Fan (FAN3)
TP3x12 (x2)
DETAIL B
#3
DRMS4016.ai
DETAIL A
#1
[Disconnect]
Connector
(CN5, CN6, CN7)
DRMS4015.ai
(5) Remove the controller unit cover and the inner cover.
Label {MC:5.1_Controller Unit Covers}
Bracket
DRMS4017.ai
A
A
[Disconnect]
Connector
(PAT3)
[Disconnect]
[Disconnect] Connector
Connector (ACE) (PAT2)
DRMS4253.ai
DRMS4018.ai
#1
[Remove]
TP4x6 (x4)
DRMS4020.ai
DETAIL A
#1
[Remove]
Clamp
DRMS4019.ai
Slot 4-GND
Slot 4-24V
Slot 3-GND Slot 3
Slot 3-12V
Slot 2-GND Slot 2
Slot 2-5V
Slot 1-GND Slot 1
Slot 1-24V
DRMS4021.ai
Check/Adjustment Procedures
CHECK 1
Check that the connector is connected with its correct position.
CHECK 2
CHECK1 Make sure that the circuit breaker is attached in their correct orientations.
#3
[Remove] #1
Circuit [Disconnect]
breaker Connector
(NFB-OUT) Circuit breaker
#2
CHECK2 [Disconnect]
Connector (NFB-IN)
DRMS4082.ai
OFF side
DRMS4083.ai
Removal Procedures
(1) Turn OFF the power of the AWS.
(2) Press the system OFF button on the control pad.
(3) Turn OFF the circuit breaker of the controller unit.
(4) Turn OFF the circuit breaker of the distribution switchboard.
WARNING (2) #1
(3) #3
[Loosen]
[Remove] Cable retaining screw
To avoid electric shock hazards, be sure to turn OFF the circuit breaker on the TP3x6 (x2) x6
distribution switchboard. Terminal block
(TB1) (2) #2
[Remove]
(5) Remove the controller unit cover and the inner cover. TB-N/L2
(white) Cable x6
{MC:5.1_Controller Unit Covers}
(6) Disconnect the cables from the terminal block (TB1). TB-L1 TB-NOUT
(black) (white)
TB-LOUT
(black)
TB-E (green)
CHECK1 TB-EOUT
(yellow/green)
DRMS4084.ai
(5) Remove the controller unit cover and the inner cover.
L1 N L2 L1 N L2
{MC:5.1_Controller Unit Covers}
TB-NOUT TB-NOUT
Terminal block
(TB2) side TB-LOUT TB-LOUT
TB-EOUT TB-EOUT
DRMS4085.ai
(3) #1
TB1-N (white) CHECK1 [Remove]
[Remove] TB2-N (white) Cable retaining screw x6
Terminal block cover
TB2-NOUT
TB1-L (black) (white)
TB2-L (black)
TB2-LOUT
(black)
TB1-E TB2-EOUT
(yellow/green) (yellow/green)
TB2-E
(yellow/green)
(3) #2
(4) #1 [Remove]
Terminal block [Remove] Cable x9
DRMS4086.ai
(TB1) TP3x6 (x2)
DRMS4087.ai
Check/Adjustment Procedures
CHECK 1
Do not make an erroneous combination of the power supply cables on the terminal
block (TB1) side and the cables on the machine side.
Terminal block
(TB1) side Machine side
TB1-N (white)
TB2-N (white)
TB2-NOUT
TB1-L (black) (white)
TB2-L (black)
TB2-LOUT
(black)
TB1-E
(yellow/green) TB2-EOUT
TB2-E (yellow/green)
(yellow/green)
DRMS4088.ai
Locations CAUTIONS
- When servicing a board, be sure to wear a wristband to ground your body.
If your body is not grounded, static buildup on your body may damage
Fuse Classification electronic components on the board.
- When replacing the fuse, check the rated amperage of the fuse to be replaced,
How to understand fuse notations and replace it with a fuse of the same rated amperage.
An example of a fuse classification symbol and rated amperage is shown below: At that time, check the rated amperage (A) silk-screened on the board as well.
- To remove the fuse, pull it straight up and off.
- When attaching the fuse, exercise care not to bend the pins of the fuse.
FRAH3D01.eps
#1
How to identify rated amperage [Remove]
Fuse
Fuse notation Rated amperage
(A)
LM 03 (D) 0.3
05 (D) 0.5 Pins
10 (D) 1
13 (D) 1.3
16 (D) 1.6
20 (D) 2
32 (D) 3.2
40 (D) 4 FRAH3D02.EPS
50 (D) 5
Fuse Locations
F208
F203
F201
F206
F204
F403 F205
F402 F202 F207
F401
DRMS4024.ai
F404
Fuse Information
DRMS4023.ai
Board Rated voltage Rated
Type
Fuse Information indications (V) amperage (A)
F201 LM 48 3.2
Board Rated voltage Rated
Type
indications (V) amperage (A) F202 LM 48 3.2
F401 LM 48 1 F203 LM 48 1
F402 LM 48 2 F204 LM 48 1
F403 LM 48 2 F205 LM 48 2
F404 (*) LM 48 1 F206 LM 48 1
*: Fuses are soldered with the board. Therefore, a fuse cannot be replaced by itself. F207 LM 48 1
F208 LM 48 3.2
Fuse Locations
F101
F105 F102
F104 F103
DRMS4025.ai
Fuse Information
Board Rated voltage Rated
Type
indications (V) amperage (A)
F101 LM 48 1.6
F102 LM 48 3.2
F103 LM 48 3.2
F104 LM 48 3.2
F105 LM 48 3.2
Fuse Locations
F301 F74
DRMS4027.ai
*: Fuses are soldered with the board. Therefore, a fuse cannot be replaced by itself.
Fuse Locations
F1
F3
F9
F7
F6
F8
F10
F2
F4
DRMS4028.ai
Fuse Information
Board Rated voltage Rated
Type
indications (V) amperage (A)
F1 LM 48 2
F2 ST 125 20
F3 ST 125 10
F4 LM 48 1
F6 LM 48 5
F7 LM 48 5
F8 LM 48 2
F9 LM 48 2
F10 ST 125 10
Fuse Location
F600
DRMS4029.ai
Fuse Information
Board Rated voltage Rated
Type
indications (V) amperage (A)
F600 LM 48 1
EACH RU SETT NG
CONNECTION TES
RU NAME IP ADDRESS RU TYPE VERSION
MARS 192.168.0.101 FDR-1000D *.*
RU NAME LIST OF EX STING RU
RU IP DDR
[Click] #3
“START” [Click]
“X”
DRMS4424.ai
DRMS4414.ai
DRMS4417.ai
#2
[Click]
“OK”
DRMS4418.ai
#2
[Click]
“OK” DRMS4415.ai
(5) Select the RU to be updated from the “LIST OF EXISTING RU”, and
click [VERSION UP].
#1
[Select ]
<Display example>
RU NAME IP ADDRESS RU TYPE VERSION
Con igu at on( )
MARS 192.168.0.101 FDR-1000D *.*
EACH RU SETTING
CONNECTION TEST
RU NAME L ST OF EXISTING RU
RU IP ADDR
#2
BACKUP RESTORE
#2
[Click]
“OK”
#2
[Press]
Enter key DRMS4420.ai
[Click]
“OK”
DRMS4421.ai
DRMS4419.ai
EACH RU SETTING
CONNECTION TEST
RU NAME LIST OF EXIST NG RU RU NAME IP ADDRESS RU TYPE VERSION (15) Press the system start button on the control pad.
MARS 192.168.0.101 FDR-1000D
RU IP ADDR
#1
[Check] - The machine normally boots up.
● GOOD indication - There is no error occurrence.
Character color is "black". - No error is found after checking for image problems and checking the S value.
● NO GOOD indication {IN:27_Checking for Image Problems and Checking the S Value}
Character color is "red".
#3
[Click]
“EXIT” DRMS4422.ai
u INSTRUCTION u
When NO GOOD, return to the procedure (1) and repeat the software version
updating procedures.
DRMS4189.ai
[Click]
“Configuration Data”
Configuration Data window
[Click]
“Backup” DRMS4192.ai
[Click]
“Configuration Data”
Configuration Data window
#2
[Select]
Backup file
#3
[Click]
“Restore” #1
[Click]
“Reference”
DRMS4193.ai
DRMS4423.ai
[Click]
“Configuration Data”
Configuration Data window
[Click]
“Back to Main-Menu”
DRMS4194.ai
DRMS4423.ai
#1
[Click]
“Set Valid”
#2
[Click]
“Back to Configuration”
DRMS4196.ai
DRMS4197.ai
RU PC-TOOL window
[Click]
“EXECUTE”
DRMS4454.ai
u NOTE u
Each exposure in this procedure takes approx. 30 seconds. Keep pressing the Generator main window
exposure button until the LED of the button begins to flash in green. If you release
the exposure button before the LED begins to flash in green, you can resume the
operation by clearing the error on the AWS.
(2) Place lead (protective apron or the like) on the exposure stand.
u NOTE u
Place the lead or the like to protect the detector from the X-rays to be irradiated
during the exposure. Otherwise, the detector might deteriorate.
(3) Start up the Zigma Webserver and enter the main menu window. DRMS4186.ai
[Click]
“Gen Adjust”
[Click]
“Enable Reset”
DRMS4201.ai
DRMS4200.ai
[Press] [Click]
Decompression “Start Adjust”
button
DRMS4202.ai
(8) Click [Gen Adjust] and return to the Generator Adjust window.
Enable Reset window
[Click]
“Gen Adjust”
DRMS4204.ai
DRMS4203.ai
#1
#1 [Click]
[Click] “Large Wolfram”
“Large Moly”
#2
#2 [Click]
[Click] “Start Adjust”
“Start Adjust”
DRMS4206.ai
DRMS4205.ai
(16) Press the exposure button.
(13) Press the exposure button. Keep pressing the exposure button until the LED of the button begins to flash in
green.
Keep pressing the exposure button until the LED of the button begins to flash in
green. (17) Click [Gen Adjust] and return to the Generator Adjust window.
(14) Click [Gen Adjust] and return to the Generator Adjust window.
#1 #1
[Click] [Click]
“Small Moly” “Small Wolfram”
#2
#2
[Click]
[Click]
“Start Adjust”
“Start Adjust”
DRMS4207.ai DRMS4208.ai
(19) Press the exposure button. (22) Press the exposure button.
Keep pressing the exposure button until the LED of the button begins to flash in Keep pressing the exposure button until the LED of the button begins to flash in
green. green.
(20) Click [Gen Adjust] and return to the Generator Adjust window.
A
[Put]
Lead sheet
[Fix]
Dosimeter
Mo/Rh 28 kV
35 kV
Chest wall plane
23 kV
Dosimeter
W/W 28 kV
35 kV
DETAIL B X-ray detecting position Dosimeter
40
Exposure
stand Unit: mm
DRMS4431.ai
<Display example>
(2) Make exposure.
EACH RU SETTING
CONNECTION TEST
RU IP ADDR
(3) Take a note of the value indicated on the dosimeter (dose value). MARS 192.168.0.101 FDR-1000D 1.3
BACKUP RESTORE
(4) Set the tube voltage to 28 kV and make exposures. ALL RUs SETTING
(3) Click “Launcher” from the Program of the menu bar, and enter the
(5) Take a note of the value indicated on the dosimeter (dose value).
Launcher menu window.
(6) Set the tube voltage to 35 kV and make exposures. [Click]
Change only the tube voltage without changing other settings. RU PC-TOOL window Program\Launcher
(7) Take a note of the value indicated on the dosimeter (dose value).
(8) Change Target/Filter to Mo/Rh, and follow the procedures of (1) to (7).
Measure the dose values when making exposures with the Target/Filter set to Mo/Rh
with voltages of 23 kV, 28 kV and 35 kV. Take notes of the respective values.
(9) Change Target/Filter to W/W, and follow the procedures of (1) to (7).
Measure the dose values when making exposures with the Target/Filter set to W/W
with voltages of 23 kV, 28 kV and 35 kV. Take notes of the respective values.
(10) Dismount the dosimeter and the lead sheet from the exposure stand.
DRMS4434.ai
#2 [Click]
“AecExpDoseCalc”
#1 [Input]
Measurement
value
#3 [Click]
“EXECUTE” DRMS4435 ai
#2 [Click]
“Start Calc” DRMS4436.ai
[Click]
“Start Calc”
DRMS4437 ai
n Preparations
Settings for storing the AEC data for every exposure are made in the procedures
below.
(1) Start up the RU PC-TOOL.
{MU:3._RU PC-TOOL_n Starting Up and Exiting the RU PC-TOOL}
(2) Select the RU from the LIST OF EXISTING RU.
Launcher menu window
Con igu at on(C)
EACH RU SETTING
CONNECTION TEST
<Display example>
RU NAME IP ADDRESS RU TYPE VERSION
RU NAME OF EXIST NG RU
RU IP ADDR
DRMS4433.ai
DRMS4434.ai
#2 [Click] [Click]
“MntTool” Connection\
Connect
#3 [Click]
“EXECUTE” DRMS4438 ai
DRMS4439 ai
#1
(8) Start up the MUTL.
[Click]
AEC/PTC tab {MU:4._MUTL_n Starting Up and Exiting the MUTL}
(9) Click [Set Aec Sens Disp Calc] of the Calibration (Field).
REFERENCE
#2 Clicking [Set Aec Sens Disp Calc], the AEC sensitivity variation data
[Click] measurement starts when Calib\Field Calibration takes place on the AWS
“AEC dump mode ON” window.
#1 [Click]
Calibration (Field)
MUTL window
#2 [Click]
“Set Aec Sens
Disp Calc”
DRMS4440E ai
DRMS4441E.ai
A
[Place]
Acrylic board [Place]
Compression plate
Exposure
Exposure stand
stand
DETAIL A
Acrylic board
Exposure (20 mm) DRMS4443.ai
stand
(3) Click [Calib] on the inspetion window of the AWS, and click “Field
Calibration”.
Chest wall
plane Acrylic board
DRMS4442.ai #2
[Click]
“Field Calibration”
#1
[Click]
“Calib”
DRMS4444.ai
Compression unit
#3 [Click]
#2 “EXECUTE” DRMS4446 ai
[Put] Use AecDispCalc (16 ch) for the 1824FPD, and AecDispCalc (32 ch) for the
Compression plate 2430FPD.
Exposure
stand
#1 [Replace]
Acrylic board
(40 mm) DRMS4445.ai
#2 [Click]
“MntTool”
#3 [Click]
“EXECUTE” DRMS4438 ai
[Click]
“Start Calc” DRMS4447.ai
[Click]
“Start Calc”
DRMS4437.ai
[Click]
Connection\
Connect
DRMS4448 ai
DRMS4439 ai
[Click]
Mnt Mode”On” [Click]
EEPROM
”Initialize”
DRMS4449 ai DRMS4450 ai
(6) Click [OK] when “Success!!” appears. (8) Click [OK] when “Success!!” appears.
[Click] [Click]
EEPROM ”CANCEL” DRMS4452.ai
”Sum Calc” (14) Click , and exit from the RU PC-TOOL window.
(15) Click the system OFF button of the mammographic stand.
(16) Click the system start button of the mammographic stand.
DRMS4451 ai
Mentioned in this section is the method of adjusting the X-ray irradiation field and the
START
light irradiation field. When both of the X-ray irradiation field and the light irradiation
field need be adjusted, make adjustment of the X-ray irradiation field first.
The irradiation field is to be adjusted when:
- the single tank unit is removed/reinstalled or replaced; Preparation
- the collimator is removed/reinstalled or replaced.
u INSTRUCTION u
Set the swivel arm to the 0° position before adjusting the irradiation field. Adjustment of Mo large focus
11.1 Adjusting the X-Ray Irradiation Field
Adjustment methods of the X-ray irradiation field include adjusting the irradiation field
while checking images by means of a film (cassette) larger than the exposure area. Adjustment of W large focus
However, it is sometimes assumed that the film image cannot be checked depending
on the installation conditions. Described below are the procedures for adjusting the
X-ray irradiation field without using the film.
c’ d’
Left side Right side
c d
Left side Right side
a’
Chest wall side Mo large focus W large focus Mo small focus W small focus
DRMS4392 ai
a' 5 5 5 5
Mo large focus W large focus Mo small focus W small focus
b' 5 5 2.5 2.5
a 3<a<9 3<a<9 3<a<9 3<a<9
c' 5 5 5 2.5
b 3 < b < 13 3 < b < 13 0<b<5 0<b<5
d' 5 5 5 2.5
c 3 < c < 13 3 < c < 13 3 < c < 13 0<c<5
Unit: mm
d 3 < d < 13 3 < d < 13 3 < d < 13 0<d<5
a+b < 19.5
c+d < 19.5
a+b+c+d < 26
Unit: mm
The way of viewing the image is as follows when a raw image is displayed by the
Image J.
○ Measurement
<Image displayed by the Image J>
position of
coordinate
Right side
Origin (0,0)
of the Image J
A B
X-ray irradiation × Do not measure at this point.
area ×○
○ Measurement
position of
Left side
coordinate
DRMS4405.ai
#5 #3 #1
[Check] [Check] [Click]
X-coordinate Y-coordinate Line button DRMS4408.ai
Right side
#4
[Align]
Cursor
[Click]
“Magnifying glass”
DRMS4409.ai
#2
[Align]
Cursor
Left side
DRMS4407.ai
#2 #1
[Align] [Align]
Cursor Cursor
#4 #3
[Input] [Input]
X-coordinate value X-coordinate
<Excel sheet (input example)> <Excel sheet (input example)> value
DRMS4410.ai DRMS4411.ai
Right side
#1
[Align]
Cursor
#1
[Align]
Cursor
Left side
#3 #3
[Input] [Input]
X-coordinate X-coordinate
value value
DRMS4412.ai DRMS4413.ai
“Change category”
<Correct window>
#2 DRMS4372E.ai
[Click]
“Correct”
DRMS4371E.ai
Upon completion of the irradiation field adjustment, uninstall the “Image J” which has - The marker appears only on the opposite chest wall side.
been installed in the AWS PC. - The lower end coordinate of the first marker measured in this procedure is used as
analysis data.
(1) Install the “Image J” in a local disk (the C drive for example) of the
AWS PC. (1) Register the exposure conditions via the AWS.
Exposure menu: RMI156 (any exposure menu will do)
(2) Double-click the “Image J” icon on the desk top to start the program.
AEC: Manual
Target/Filter: Mo/Mo
n Starting the Excel File for Analysis mAs: 50
Focus Spot: H-mode
REFERENCE
Grid: IN
The Excel program is used for data analysis necessary for adjustment of the X-ray
irradiation field. It is recommended that you keep the Excel file in your maintenance
(2) Make exposures.
PC or the like for convenience of workability. (3) Click [File] →[Import] →[Raw] in the Image J menu bar.
[Click]
Raw
DRMS4373.ai
REFERENCE
First marker
The raw image file is stored in the following folder:
C\Program Files\FujiFilm\FCR\FDR-1000DRSZ\SYSTEM\192.168.0.101\LOG\
CRCT
(6) #2
File name: (Example) DG001.tdt
Opposite [Align]
chest wall side Cursor
Second
marker
Left side
DRMS4375.ai
DRMS4376.ai
DRMS4377.ai
DRMS4378.ai
<Collimator window>
[Check]
“Moly large”
DRMS4379.ai
#2 #1
[Click] [Select]
“in” of each “Select step size”
edge
#2
[Click]
“in” of each edge
DRMS4380.ai
DRMS4382.ai
[Input]
Large Moly set value
<Excel sheet>
DRMS4381.ai
<Collimator window>
[Input]
Large Moly set value
<Excel sheet>
DRMS4384.ai
DRMS4385.ai
[Click]
Tab “(figure)”
(8) Repeat the procedures (1) to (7) until the coordinate measurement
data is n=5.
DRMS4387.ai
If the measurement data largely deviates from the approximation curve, take the
following procedures and check the accuracy of the data.
- Again display the image data of interest and make another measurement.
- Set the irradiation field set values the same as those of the conditions where
the data deviates, make exposures, and again make measurement.
- Change the set values of the irradiation field, make exposures and again make
measurement (to increase the measurement data).
<Display example>
<Display example>
Approximation curve
[Check]
Measurement
data
[Check]
Calculation result (vact)
DRMS4389.ai
DRMS4388.ai
(2) Click [Collimator] of the Zigma Webserver, and enter the Collimator
window.
#5 #4
[Click] [Check]
“Back to Collimator” Set value
DRMS4390.ai
DRMS4377.ai
DRMS4394.ai
<Collimator window>
[Check]
“Moly small”
DRMS4395.ai
#2 #1
[Click] [Select]
“in” of each “Select step size”
edge
#2
[Click]
“in” of each edge
DRMS4380.ai
DRMS4397.ai
[Input]
Small Moly set value
<Excel sheet>
DRMS4396.ai
u NOTE u
Change each set value as much as possible to improve the data accuracy for
position adjustment.
<Collimator window>
[Input]
Small Moly set value
<Excel sheet>
DRMS4398.ai
DRMS4385.ai
[Click]
Tab “(figure)”
(8) Repeat the procedures (1) to (7) until the coordinate measurement
data is n=5.
DRMS4387.ai
If the measurement data largely deviates from the approximation curve, take the
following procedures and check the accuracy of the data.
- Again display the image data of interest and make another measurement.
- Set the irradiation field set values the same as those of the conditions where
the data deviates, make exposures, and again make measurement.
- Change the set values of the irradiation field, make exposures and again make
measurement (to increase the measurement data).
<Display example>
<Display example>
Approximation curve
[Check]
Measurement
data
[Check]
Calculation result (vact)
DRMS4400.ai
#2
[Click]
“in” or “out”
#3 of each edge
[Click]
“Details” b2
a2
#5 #4
[Click] [Check]
“Back to Collimator” Set value
DRMS4401.ai
(1) Mount the 18x24 compression plate (24x30 compression plate for the
FDR-1000DRLZ).
(2) Remove the compression plate (plastic plate part). [Check]
{MC:3.5.3_Compression Plate (Plastic Plate Part)} “Light”
DRMS4377.ai
REFERENCE
#2 Clicking [in] or [out], the light irradiation field lamp automatically lights up.
[Click]
“in” or “out” of
each edge
#3
[Check]
Light irradiation field
Exposure stand
Exposure frame
DRMS4404.ai
Closure joint
(FPD) *1 Used to fill or drain the coolant.
Power supply cable length: Approx. 5 m
DRMS4347.ai The water supply jig does not include the coolant.
{MC:13.2_How to Use the Water Supply Jig}
*1: If the closure joint (FPD) is broken or lost, purchase a
Water supply jig Closure hose closure hose (REF No.3) as a substitute.
Closure joint (FPD) *1 (temperature
adjustment unit)
Transformer Adapter
DRMS4348.ai
3-2
REF No.3-1
370N100155 Used to close the coolant hose of the FPD. The closure
hose (FPD) is handled as a substitute for the closure joint
Closure hose FUJIFILM (FPD).
3 {MC:13.2.4_Connecting the Closure Hose (FPD
(FPD) Corporation
and Temperature Adjustment Unit)}
REF No.3-2
3-1 316S2094
DRMS4123.ai
DRMS4124.ai
3 4
Closure hose FUJIFILM
5 370Y100224 REF Nos. 3 and 4 can be purchased as a set.
assembly Corporation
DRMS4349.ai
357Y100257
(*1)
Used to temporarily place the FPD. Also used when
FUJIFILM attaching/detaching the temperature adjustment unit.
6 Stand jig
Corporation *1: For FDR-1000DRSZ
*2: For FDR-1000DRLZ
357Y100270
(*2)
DRMS4255.ai
DRMS4256.ai
FUJIFILM Used to boot up the system while the right- and left-hand
8 short-circuiting 136Y101179
Corporation side covers are not put on the elevation stand.
cable DRMS4350.ai
Temperature
FUJIFILM Used to circulate the coolant while the temperature
9 adjustment unit 136Y101389
Corporation adjustment unit is removed.
relay cable DRMS4257.ai
NOTE
Fill or drain the coolant on a sheet to prevent the coolant from splashing on the Tapped joint
machine or the floor.
If the coolant accidentally attaches to the machine or the floor, immediately wipe it off.
Components DETAIL A
{MC:13.1_List of Jigs}
Power supply
switch
: OFF
: ON
DRMS4104.ai
Joint Hose
Joint [Insert]
Hose
DRMS4105.ai
Open ring
#1
[Push]
Open ring Hose
Joint Joint
DRMS4106.ai
DRMS4108.ai
Cooling unit
Hose disconnecting position
Temperature
adjustment unit
Reserve tank
Hose (blue)
FPD hose (thin)
Hose (white)
DRMS4346 ai #1
[Disconnect]
Hose (white)
Tapped joint
DRMS4109.ai
Connect the FPD hose (thin), which is disconnected in this step, to the tapped (4) Connect the FPD hose (thin) which is removed in the procedure (2) to
joint of the reserve tank. the tapped joint of the water supply pump.
(1) #1
[Connect]
Water supply jig hose
(white)
Reserve tank
Reserve tank
[Disconnect]
FPD hose (thin)
Tapped joint of
the water supply jig
DRMS4111.ai
(2) #1
[Connect]
FPD hose (thin)
DRMS4110.ai
Hose
[Remove] [Open]
Tapped joint Tap x2
Empty coolant bottle
DRMS4113.ai
[Attach]
Coolant bottle
DRMS4112.ai
[Check]
Coolant liquid level
Hose
[Close]
Tapped joint Tap x2
DRMS4125.ai
Hose
DRMS4114.ai
REFERENCE
When the FPD hose is to be disconnected from the tapped joint of the water
supply jig, it is recommended that you disconnect the hose above the joint of the
[Disconnect] reserve tank, to prevent the coolant from dripping.
Hose (white) of
the water supply jig
Tapped joint of
the water supply jig
DRMS4116.ai
The coolant may hardly drip
if the reserve tank joint and
the tapped joint of the water
supply jig are at the
approximately same height.
[Disconnect]
FPD hose (thin)
DRMS4115.ai
Hose (blue)
Hose (white)
Reserve tank
[Insert]
FPD hose (thin)
#1
DRMS4134.ai
[Insert]
Hose (white)
Tapped joint
DRMS4135.ai
CAUTION
If the liquid level which has been located in the vicinity of the upper surface of
the reserve tank gets lower than the joint of the reserve tank and the liquid level
further continues to lower, leakage is suspected. Immediately turn OFF the
circuit breaker of the controller unit, and check if the leakage occurs.
GOOD NO GOOD
The liquid level of the coolant is higher The liquid level of the coolant is lower
than the joint of the reserve tank. than the joint of the reserve tank.
DRMS4117.ai
INSTRUCTION
After the coolant is drained, attach the closure jig (closure hose) according to the
procedures in the manual. The coolant remains in the FPD and the temperature
adjustment unit after the coolant is drained via the water supply jig. To prevent the
coolant remaining in the unit from leaking, the closure jig (closure hose) need be
attached.
Hose
[Attach] DRMS4113.ai
Empty coolant bottle (5) Turn ON the power of the water supply pump.
(6) When drainage of the coolant in the reserve tank (lowering of the
liquid level) stops, turn OFF the power of the water supply pump.
Water supply jig
REFERENCE
DRMS4118.ai
Tapped joint of
the water supply jig
Hose
[Close]
Tapped joint Tap x2
[Disconnect]
FPD hose (thin)
Hose
DRMS4136.ai
DRMS4114.ai
Hose (blue)
Hose (white)
#1
Joint of the [Insert]
reserve tank Hose (white)
[Disconnect]
Hose of the water supply jig
(white) DRMS4119.ai
Tapped joint
DRMS4135.ai
(11) Connect the closure hose to the FPD and the temperature adjustment
unit.
{MC:13.2.4_Connecting the Closure Hose (FPD and Temperature Adjustment
Unit)}
[Disconnect]
Coolant hose (thick)
DRMS4361.ai
REFERENCE
Put on two hose clamps on the closure hose (FPD) before using the hose.
Hose clamp
Joint
DRMS4091.ai
DETAIL A
#1
[Unclamp] Closure hose
Clamp
#3
[Pull out]
Coolant hose
(thick)
DRMS4092.ai
A
Temperature
adjustment unit
DETAIL A DETAIL B
Reserve tank #3
Joint of the [Connect]
reserve tank Closure hose
#2
[Connect] Joint on the rear of the
Closure hose temperature adjustment unit
DRMS4122.ai
Qty.: 1 Qty.: 1
NOTE Remarks:
Remarks:
To service the machine, use the servicing instruments and tools that have been
inspected and calibrated.
If the machine were serviced using servicing instruments and tools that have not
been inspected and calibrated, proper performance of the machine could not be Thin-blade screwdriver Phillips stubby screwdriver
guaranteed. Reference type: NO.110-4 Reference type: NO.135-4
Reference manufacturer: Reference manufacturer:
PB PB
Adjustable wrench 150 mm Nippers 150 mm Qty.: 1 Qty.: 1
Remarks: Remarks: Jeweler's screwdriver set Combination box-open-end wrench (5.5 mm)
Reference type: NO.146 Reference type: M41-5.5
Reference manufacturer: Reference manufacturer:
ENGINEER KTC
Water pump pliers Tweezers 125 mm Qty.: 1 Qty.: 1
Remarks: Remarks: Combination box-open-end wrench (7 mm) Combination box-open-end wrench (8 mm)
Reference type: M41-7 Reference type: M41-8
Qty.: 1 Qty.: 1
Remarks: Remarks:
DRMS4127.ai
Combination box-open-end wrench (10 mm) Combination box-open-end wrench (13 mm) Steel rule (150 mm) Steel rule (300 mm)
Reference type: M41-10 Reference type: M41-13 Reference type: Type C Reference type: Type C
Reference manufacturer: Reference manufacturer: Reference manufacturer: Reference manufacturer:
KTC KTC YAMAYO YAMAYO
Qty.: 1 Qty.: 1 Qty.: 1 Qty.: 1
Allen wrench set Midget ratchet set Steel tape measure (5.5 m) Soldering iron (30W)
Reference type: BLX-9 Reference type: 7331 Reference type: WS5005 Reference type: Dash15
Digital multimeter Crimp pliers Solder Desoldering tool with a vacuum attachment
Reference type: CD731 Reference type: P-73 Reference type: NO.66 Reference type: US140
Reference manufacturer: Reference manufacturer: Reference manufacturer: Reference manufacturer:
SANWA HOZAN HAKKO EDSYN
Qty.: 1 Qty.: 1 Qty.: 1 Qty.: 1
DRMS4128.ai DRMS4129.ai
DRMS4130.ai
Dust-free paper
Reference type: Kimwipe
Reference manufacturer:
CRECIA
Qty.: 1
Remarks:
DRMS4131.ai
Qty.: 1 Qty.: 1
Remarks: Remarks:
Remarks: Remarks:
Reference manufacturer:
EM-30L Magnetized Phillips screwdriver Closed wrench (17 mm)
Three Bond
Reference manufacturer: Reference type: Reference type:
Qty.: 1
Qty.: 1 Reference manufacturer: Reference manufacturer:
Remarks:
Remarks: Qty.: 1 Qty.: 2
Remarks: Remarks:
DRMS4132.ai
DRMS4133.ai
INDEX
INDEX 01
● カバー
COVER
INDEX 02
● マンモユニット
MAMMO UNIT
INDEX 03
● 制御部
CONTROLLER UNIT
INDEX 04
● その他
OTHERS
INDEX 05
● 治具
JIG
INDEX 06
● ケーブル
CABLE
INDEX 07
● 回路図
CIRCUIT DIAGRAM
INDEX 08
● 部品番号検索表
PARTS NOS. SEARCH TABLE
INDEX 09
● 締結用および配線用のサービス部品一覧表
List of Service Parts for Securing and Wiring
01A カバー 1
COVER 1
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
Elevation Stand
昇降スタンド
E 1 W11G10415588 Right-Hand Side 1
右側面カバー
Cover
昇降スタンド Elevation Stand Left-
E 2 W11G10415589 1
左側面カバー Hand Side Cover
昇降スタンド Elevation Stand Top
E 3 W11G10140032 1
上カバー Cover
昇降スタンド Elevation Stand Rear
E 4 W11G10140033 1
背面カバー Cover
D 5 W11G10140036 ディスプレイカバー Display Cover 1
D 6 W11G10140034 ベースカバー Base Plate Cover 1
ディスプレイボード Display Board
C 7 W11G10139907 1
アセンブリ(D903) Assembly (D903)
Emergency Stop
D 8 W11G10139846 緊急停止スイッチ 2
Switch
D 9 W11G06343144 フェイスガード Face Guard 1
01B カバー 2
COVER 2
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
FPD Right-Hand
FPD 右カバー
D 1 350Y100950A Cover 1
(右アームレスト)
(Right-Hand Armrest)
D 2 350N100868C FPD 下カバー FPD Lower Cover 1 For FDR-1000DRSZ
D 3 376N100016A エアフィルタ Air Filter 1
D 4 360N100074G フィルタカバー Filter Cover 1
C 5 W11G10415617 操作パネル Control Panel 2
FPD 左カバー FPD Left-Hand Cover
D 6 350Y100949A 1
(左アームレスト) (Left-Hand Armrest)
D 7 376N100021A エアフィルタ Air Filter 1 For FDR-1000DRSZ
D 8 376N100023B エアフィルタ Air Filter 1
D 9 350N100930B カバー(左) Cover (Left) 1
D 10 350N100931B カバー(右) Cover (Right) 1
スイベルアーム
D 11 W11G10415612 Swivel Arm Top Cover 1
上カバー
スイベルアーム Swivel Arm Right-
D 12 350Y100861B 1
右側面カバー Hand Side Cover
スイベルアーム Swivel Arm Right-
D 13 350Y100931D 1 For FDR-1000DRSZ
右前面カバー Hand Front Cover
スイベルアーム Swivel Arm Lower
D 14 W11G10415613 1
下カバー Cover
スイベルアーム Swivel Arm Left-Hand
D 15 W11G10415615 1 For FDR-1000DRSZ
左前面カバー Front Cover
スイベルアーム Swivel Arm Left-Hand
D 16 W11G10415616 1
左側面カバー Side Cover
D 17 405N101115 銘板 Label 1
D 18 350N101267 FPD 下カバー FPD Lower Cover 1 For FDR-1000DRLZ
D 19 376N100039 エアフィルタ Air Filter 1 For FDR-1000DRLZ
スイベルアーム Swivel Arm Right-
D 20 350Y100930C 1 For FDR-1000DRLZ
右前面カバー Hand Front Cover
スイベルアーム Swivel Arm Left-Hand
D 21 350Y100929C 1 For FDR-1000DRLZ
左前面カバー Front Cover
02A マンモユニット 1
MAMMO UNIT 1
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
For FDR-1000DRSZ,
クーラント(250ml)
Flat Panel Detector
CR 1 848Y100054F 平面検出器(FPD) 1 含む/
(FPD)
Coolant (250ml) is
included.
ボードアセンブリ Board Assembly
CR 2 113Y1970C 1
XIF51A XIF51A
クーラント(250ml)
Temperature 含む/
CR 3 379Y100034E/F 温度調整ユニット 1
Adjustment Unit Coolant (250ml) is
included.
A 4 359Y100048 クーラント Coolant 2
For FDR-1000DRSZ,
C 5 350Y100820L カーボンカバー Carbon Cover 1 スイッチ含む/
Switch is included.
For FDR-1000DRSZ,
スイッチ含まない/
D 6 350Y100595L カーボンカバー Carbon Cover 1
Switch is not
included.
D 7 119Y100028B ファン Fan 2 *1
For FDR-1000DRLZ,
クーラント(250ml)
Flat Panel Detector
CR 8 848Y100065B 平面検出器(FPD) 1 含む/
(FPD)
Coolant (250ml) is
included.
For FDR-1000DRLZ,
C 9 350Y101006A カーボンカバー Carbon Cover 1 スイッチ含む/
Switch is included.
For FDR-1000DRLZ,
D 10 350Y101005 カーボンカバー Carbon Cover 1 スイッチ含まない/
Switch is not included.
E 11 370Y100306 ホースアセンブリ Hose Assembly 1
48V/1A
A 12 137S1417 ヒューズ Fuse 1
For XIF51A F600
ボードアセンブリ Board Assembly
CR 13 113Y1952F 1
PTC51A PTC51A
*1: FDR-1000DRLZ には 3 個使用/ Use 3 pcs for FDR-1000DRLZ
02D マンモユニット 4
MAMMO UNIT 4
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
C 1 W11G10141960 シングルタンク Single Tank Unit 1
C 2 W11G10139621 コリメータ Collimator 1
ボ ー ド ア セ ン ブ リ Board Assembly MFI
C 3 W11G10139931 1
MFI (D910) (D910)
ボ ー ド ア セ ン ブ リ Board Assembly CBA
C 4 W11G10501291 2
CBA (D905) (D905)
E 5 W11G07061989 ファン Fan 1
02E マンモユニット 5
MAMMO UNIT 5
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
D 1 W11G10415592 昇降モータドライバ Motor Inverter Lift 1
D 2 W11G10139779 回転モータドライバ Motor Inverter Rotate 1
ボードアセンブリ Board Assembly CBS
C 3 W11G10501292 1
CBS (D900) (D900)
Power Supply Unit
C 4 W11G10139644 電源部(PSU) 1
(PSU)
C 5 W11G03082146 電池 LITHIUM BATTERY 1 CR2032 3V
E 6 356Y100803 ブラケット Bracket 1
D 7 FW11G10415593 昇降モータ Motor Lift 1
03A 制御部 1
CONTROLLER UNIT 1
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
D 1 345N1823D ルーバ Louver 6
D 2 376N0238 エアフィルタ Air Filter 3
D 3 317N100022B キャップ Cap 6
D 4 351Y100004F カバー Cover 1
03B 制御部 2
CONTROLLER UNIT 2
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
ボードアセンブリ Board Assembly
CR 1 113Y1980A 1
MIF51A MIF51A
ボードアセンブリ Board Assembly
CR 2 113Y1963D 1
MPI51A MPI51A
ボードアセンブリ Board Assembly
CR 3 113Y1964B 2 For FDR-1000DRSZ
CPS51B CPS51B
メモリカード
Memory Card
C 4 114N100013 (コンパクトフラッ 1 For FDR-1000DRSZ
(CF Card)
シュカード)
ボードアセンブリ Board Assembly
C 5 113Y1977A 1
CFI51A CFI51A
*1
C 6 113N100091/A/B メモリボード Memory Board 1
DIMM 1GB
ボードアセンブリ Board Assembly
CR 7 814Y100132A/B/C CPU51A(マザー CPU51A 1
ボード) (Motherboard)
ボードアセンブリ Board Assembly
CR 8 113Y1965C 1
PAT51A PAT51A
ボードアセンブリ Board Assembly
C 9 113Y1962D 1
IND51A IND51A
48V/0.5A
A 10 137S1456 ヒューズ FUSE 1
for MPI51A F302
48V/2A
A 11 137S1420 ヒューズ FUSE 3
for PAT51A F1, F8, F9
125V/20A
A 12 137S1374 ヒューズ FUSE 1
for PAT51A F2
125V/10A
A 13 137S1436 ヒューズ FUSE 2
for PAT51A F3, F10
48V/1A
A 14 137S1417 ヒューズ FUSE 1
for PAT51A F4
48V/5A
A 15 137S1423 ヒューズ FUSE 2
for PAT51A F6, F7
ボードアセンブリ Board Assembly
CR 16 113Y100273C 3 For FDR-1000DRLZ
CPS51C CPS51C
メモリカード
Memory Card
C 17 114N100026 (コンパクトフラッ 1 For FDR-1000DRLZ
(CF Card)
シュカード)
*1: FDR-1000DRLZ には 2 個使用/ Use 2 pcs for FDR-1000DRLZ
03C 制御部 3
CONTROLLER UNIT 3
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
D 1 119S0043 ファン FAN 3
E 2 350N2774C カバー Cover 4
ナット
E 3 305N100033 Nut (Anchor Bolt) 4
(アンカーボルト)
CR 4 125Y100032 電源 Power Supply 1 Alpha II 650
D 5 135S0829 端子台 Terminal Block 1 TB2
D 6 135Y100008 端子台 Terminal Block 1 TB1
D 7 128S1088 スイッチ Switch 1
ボードアセンブリ Board Assembly
C 8 113Y1979A 1
NFB51A NFB51A
E 9 304S1000820 ボルト Bolt 4
E 10 309S0220008 ばね座金 Spring Washer 4
E 11 309S0120008 平座金 Washer 4
E 12 *** ケーブル Cable * Refer to INDEX06
E 13 370N100272 スペーサ Spacer 4
04A その他(付属品)
OTHERS (ACCESSORIES)
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
1 DETAIL A D 1 W11G10415606 コントロールパッド Control Pad 1
コントロールパッド
E 2 W11G10139656 Control Pad Cable 1
ケーブル
1 セット=右+左/
D 3 W11G10139755 フットスイッチ Foot Switch 1
1set=right+left
D 4 357Y100226A 操作卓 Operation Desk 1
D 5 350Y100935B サイドカバー Side Cover 2
D 6 350Y100933C フロントカバー Front Cover 1
4
C
DETAIL B DETAIL C
5
5
7
8
9
10
6
12
11
DRMS605.pdf
04B その他(オプション:ステレオ撮影キット)
OTHERS (OPTION: STEREO EXPOSURE KIT)
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
D 1 350Y101608 カバー Cover 1 For 18x24
D 2 350Y101609 カバー Cover 1 For 24x30
For 24x30
D 3 350Y101550 アームレスト L Arm rest L 1 ( 軽量版/
Lightweight version)
For 24x30
D 4 350Y101551 アームレスト R Arm rest R 1 ( 軽量版/
Lightweight version)
E 5 356N104378B ブラケット Bracket 1
E 6 136Y102382C ケーブル Cable 1
E 7 356N104569 ブラケット Bracket 1
D 8 350Y101589 カバー Cover 1
E 9 356N104614C ブラケット Bracket 1
E 10 136Y102395B ケーブル Cable 1
E 11 136Y102826A ケーブル Cable 1
E 12 136Y102825A ケーブル Cable 1
E 13 350N101834A カバー Cover 1
E 14 136N100143 導電板 Bracket 2
E 15 356N105032 ブラケットナット Bracket 2
E 16 316S0256 クランプ Clamp 1
A 17 316S1082 束線部品 Tie 1
E 18 308S0414 六角頭TPねじ Screw 1 TP3x6
E 19 308S9780612 六角頭ねじ Screw 1 BR6x12
E 20 301S7000310 ねじ (SUS) Screw 1 S3x10(SUS)
Stereo Exposure
D 21 358Y100111A ステレオ撮影プラグ 1
Plug
4.1* 6
4.2*
4.3*
4.4*
DRMS605B.ai
05 治具
JIG
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
クーラントは含まな
給水治具 Water Supply Jig
E 1.1 133Y100033A 1 い/ Coolant is not
(AC100V 用) (For AC100V)
included.
・クーラントは含ま
ない/ Coolant is
not included.
給水治具 Water Supply Jig ・アダプタ、トラン
E 1.2 133Y100036A 1
(AC200V 用) (For AC200V) スフォーマを含
む/ Adaptor and
transformer are
included.
E 2 357Y100257 スタンド Stand Jig 1 For FDR-1000DRSZ
・左右 1 セット/
One set includes
the left and right
handles.
E 3 356Y100836 取手 Handle 1
・取付ネジ(TP3 x 6)
1.1 含む/ Retaining
screws (TP3 x 6)
are included.
緊急停止ショート Emergency Stop
E 4 136Y101179 1
治具ケーブル Short-Circuit Cable
Temperature
温度調整ユニット
E 5 136Y101389 adjustment unit relay 1
中継治具ケーブル
cable
E 6 370N100155 閉栓ホース(FPD) Closure Hose (FPD) 1
E 7 316S2094 クランプ Clamp 2
Closure Hose
閉栓ホース
E 8 370N100156 (Temperature 1
(温度調整ユニット)
1.2 Adjustment Unit)
Closure Hose
E 9 370Y100224 閉栓ホースアセンブリ 1
Assembly
E 10 357Y100270 スタンド Stand Jig 1 For FDR-1000DRLZ
2,10
3
CAUTION
CAUTION
4 5
6
8
7 7
DRMS609.pdf
06 ケーブル
CABLE
CIRCUIT
RANK REF. PART NO. PART NAME QTY. REMARKS SERIAL NO. REFER TO
DIAGRAM *2
E 1 136Y101145 ケーブル Cable 1 5/11,9/11
E 2 136Y101146 ケーブル Cable 1 5/11,8/11
E 3 136Y101152 ケーブル Cable 1 3/11,8/11,9/11,11/11
E 4 136Y101153 ケーブル Cable 1 5/11
E 5 136Y101180A ケーブル Cable 1 2/11
E 6 136Y101181A ケーブル Cable 1 2/11
E 7 136Y101183 ケーブル Cable 1 For Alpha II 650 2/11,3/11
E 8 136Y101184A ケーブル Cable 1 For Alpha II 650 2/11,3/11
E 9 136Y101185A ケーブル Cable 1 For Alpha II 650 2/11,3/11
E 10 136Y101186A ケーブル Cable 1 For Alpha II 650 2/11,3/11
E 11 136Y101187 ケーブル Cable 1 4/11
E 12 136Y101189 ケーブル Cable 1 3/11
E 13 136Y101190A ケーブル Cable 1 4/11,5/11
E 14 136Y101191 ケーブル Cable 1 4/11,5/11
E 15 136Y101192 ケーブル Cable 1 4/11,5/11
E 16 136Y101193 ケーブル Cable 1 4/11,6/11
E 17 136Y101194 ケーブル Cable 1 4/11,6/11
E 18 136Y101195A ケーブル Cable 1 4/11
E 19 136Y101196 ケーブル Cable 1 5/11
E 20 136Y101197 ケーブル Cable 1 5/11
E 21 136Y101198 ケーブル Cable 1 6/11
E 22 136Y101199C ケーブル Cable 1 6/11
E 23 136Y101200B ケーブル Cable 1 11/11
E 24 136Y101202B/C ケーブル Cable 1 2/11
E 25 136Y101382A ケーブル Cable 1 2/11
E 26 136Y101383 ケーブル Cable 1 2/11
E 27 136Y101384 ケーブル Cable 1 2/11
E 28 136Y101385B ケーブル Cable 1 5/11
E 29 136Y101386 ケーブル Cable 1 11/11
E 30 136Y101387D ケーブル Cable 1 11/11
E 31 136Y101533 ケーブル Cable 1 2/11
*1 ケーブルは FDR-1000DRSZ と FDR-1000DRLZ で共通
The cables for FDR-1000DRSZ and FDR-1000DRLZ the same.
*2 FDR-1000DRSZ の場合は「07A 回路図」を参照し、FDR-1000DRLZ の場合は「07B 回路図」を参照すること
Refer to the “07A CIRCUIT DIAGRAM” for FDR-1000DRSZ / the “07B CIRCUIT DIAGRAM” for FDR-1000DRLZ.
シス ム電
設置時 冶具ハーネス Power supply 温調電源PTC基板用 136Y101183
A AlphaⅡ650 A
Slot4 Slot4_GND 1 +24V_GND 橙 ORN
136Y101179 S_Slot4_24VG1 3
2 +24V_GND 橙 ORN
S_Slot4_24VG2 3
STOP_L 1 1 STOP_R
2 2 Slot4_24V 1 +24V 赤 RED
S_Slot4_24V1 3
2 +24V 赤 RED
S_Slot4_24V2 3
オプション(25m) S_Slot1
136Y101389
B
1 1 EXT_1 Slot1 Slot1_GND 1 +24V_GND 橙 ORN
B
136Y101178 ※組図No.136Y101532 EXT_2 S_Slot1_24VG1 3
10
9
8
7
6
5
4
3
2
1
6 6 2 +24V_GND 橙 ORN
S_Slot1_24VG2 3
2 2
1 L 1 7 7
L1 Slot1_24V 1 +24V 赤 RED
1 N 1 3 3 S_Slot1_24V1 3
N2 2 +24V 赤 RED
1 E 1 8 8 S_Slot1_24V2 3
FG2
4 4
9 9 システム電源各基板用
5 5
C
10 10 C
システム電源CPU基板用
136Y101184
Slot2 Slot2_GND
S_Slot2 1 +5V_GND 橙 ORN
S_Slot2_5VG1 3
2 +5V_GND 橙 ORN
S_Slot2_5VG2 3
オプション(15m)
10
9
8
7
6
5
4
3
2
1
Slot2_5V 1 +5V 赤 RED
136Y101201 ※組図No.136Y101531 S_Slot2_5V1 3
TB1 2 +5V 赤 RED
S_Slot2_5V2 3
(MBK6/E-Z-4P)
D L 黒 BLK L D
L 1 1 1
1 N 1 1 白 WHT N 136Y101185
N
1 E 1 1 緑/黄 E Slot3
FG Slot3_GND 1 +12V_GND 橙 ORN
S_Slot3_12VG1 3
2 +12V_GND 橙 ORN
S_Slot3_12VG2 3
S_Slot3
136Y101382 Slot3_12V 1 +12V 赤 RED
S_Slot3_12V1 3
136Y101383 2 +12V 赤 RED
S_Slot3_12V2 3
10
9
8
7
6
5
4
3
2
1
136Y101384
E システム電源CPU基板用 E
TB2 136Y101186
丸端子 端子台
SLOT1_5VSB 灰 GRY
S_Slot1_5VSB 3
ブレーカー SLOT1_CNT 灰 GRY
TB2-LOUT TB1-L S_Slot1_CNT 3
1 L 1 1 L 黒 BLK (SER-F-11-62F-3A-BWT) SLOT2_5VSB 灰 GRY
L_X S_Slot2_5VSB 3
SLOT2_CNT 灰 GRY
NFB_IN S_Slot2_CNT 3
TB2-L SLOT3_5VSB 灰 GRY
1 L 黒 BLK 黒 BLK L 1 NFB51A S_Slot3_5VSB 3
SLOT3_CNT 灰 GRY
N2 136Y101181 113N1979 S_Slot3_CNT 3
F F
TB2-NOUT TB1-N NFB_OUT NFB1 NFB2 136Y101202
1 N 1 1 N 白 WHT
N_X 1 L 黒 BLK 1 1
TB2-N 2 2 3 AC_IN
1 N 白 WHT 白 WHT N白 WHT 3 2 L 黒 BLK 1 SCN1 1
4 N 白 WHT 2 2
TB2-EOUT TB1-E
E E 黄/緑 3
FG_X 1 1 1 E 緑/黄 1 +5Vスタンバイ電源用
AC_E 4
TB2-E 5
1 E 緑/黄 E 6
G ファストン端子 G
1
136Y101381 ※組図No.136Y101533 FG_OUT
機種 5010 ユニット -
制御部背面中継コネクタ
部
A A A
N
PAT51A 4
-
136Y101183
136Y101152
橙 ORN +24V_GND 1 PAT1 136Y101189
2 S_Slot4_24VG1
赤 RED +24V 5
2 S_Slot4_24V1 CN1-P CN1-J
橙 ORN +24V_GND 2
2 S_Slot4_24VG2 PAT11 1 橙 ORN +24V_P1 1 1 白 WHT +24V_P1
赤 RED +24V 6 SND_+24V1 8
B 2 S_Slot4_24V2 6 橙 ORN GND_P1 6 6 青 BLU GND_P1 B
橙 ORN +24V_GND 3 SND_GND1 8
2 S_Slot1_24VG1 2 橙 ORN +24V_P2 2 2 赤 RED +24V_P2
赤 RED +24V 7 SND_+24V2 8
2 S_Slot1_24V1 7 橙 ORN GND_P2 7 7 灰 GRY GND_P2
橙 ORN +24V_GND 4 SND_GND2 8
2 S_Slot1_24VG2 3 橙 ORN +24V_A 3 3 黒 BLK +24V_A
赤 RED +24V 8 TRC_+24V 9
2 S_Slot1_24V2 8 橙 ORN GND_A 8 8 茶 BRN GND_A
TRC_GND 9
4 橙 ORN +24V_T1 4 4 緑 GRN +24V_T1
PTC_+24V1 11
9 橙 ORN +24V_T2 9 9 黄 YEL +24V_T2
PTC_+24V2 11
5 橙 ORN GND_T1 5 5 橙 ORN GND_T1
PTC_DGND1 11
10 橙 ORN GND_T2 10 10 桃 PIK GND_T2
PTC_DGND2 11
C C
136Y101184
136Y101185
G G
機種 5010 ユニット -
136Y101191
C C
136Y101192
136Y101194
PAT9
1 +5V 橙 ORN
CFIPWR_5V 6
2 +5V_GND 橙 ORN
CFIPWR_GND 6
E E
IND51A
113Y1962
136Y101190
PAT12 1 +5V_L 灰 GRY 1 IND1
2 +5V_DGND 灰 GRY 2
D_GND 1 PAT10
5 Fuse_GND1 3 +5V_L 灰 GRY 3
+24V_P1_Det 2
5 Fuse_+24V_P1_Det 4 +5V_DGND 灰 GRY 4
+24V_P2_Det 3
5 Fuse_+24V_P2_Det 5 +5V_L 灰 GRY 5
D_GND 4
5 Fuse_GND2 6 +5V_DGND 灰 GRY 6
+24V_A_Det 5
5 Fuse_+24V_A_Det
F +24V_F_Det 6 F
5 Fuse_+24V_F_Det 136Y101195
D_GND 7
5 Fuse_GND3 パワーオン表示灯
+5V_L_Det 8
5 Fuse_+5V_L_Det
+5V_SB_Det 9
5 Fuse_+5V_SB_Det
D_GND 10
5 Fuse_GND4
+24V_T_Det 11
5 Fuse_+24V_T_Det
+24V_TF_Det 12
5 Fuse_+24V_TF_Det
D_GND 13
5 Fuse_GND5
Reserve1 14
5 Fuse_Reserve1
Reserve2 15
5 Fuse_Reserve2
G PER_OFF 16 G
5 Fuse_PWR_OFF
D_GND 17
5 Fuse_GND6
Zigma_ON_C 18
5 Zigma_ON_C
Zigma_ON_A 19
5 Zigma_ON_A
D_GND 20
5 Fuse_GND7
機種 5010 ユニット -
136Y101190 Image_signal[1:37] 9
P A MPI2
1
D_GND 1 MPI5 Y Image_signal2
4 Fuse_GND1 2 136Y101145
+24V_P1_Det 2
4 Fuse_+24V_P1_Det
A +24V_P2_Det 3 Image_signal17 A
4 Fuse_+24V_P2_Det 17
D_GND 4 Image_signal18
4 Fuse_GND2 18
+24V_A_Det 5 Image_signal19
4 Fuse_+24V_A_Det 19
+24V_F_Det 6 Image_signal20
4 Fuse_+24V_F_Det 20
D_GND 7
4 Fuse_GND3
+5V_L_Det 8 Image_signal36
4 Fuse_+5V_L_Det 36
+5V_SB_Det 9 Image_signal37
4 Fuse_+5V_SB_Det 37
D_GND 10
4 Fuse_GND4 136Y101146
+24V_T_Det 11
4 Fuse_+24V_T_Det
+24V_TF_Det 12
4 Fuse_+24V_TF_Det MPI3 1 茶 BRN MNb_A 1 MIF3 1 黒 BLK MNb_A
B D_GND 13 MIF4 MNb_A 8 B
4 Fuse_GND5 2 赤 RED MNb_B 2 14 緑 GRN MNb_B
Reserve1 14 MNb_B 8
4 Fuse_Reserve1 3 橙 ORN MDin_A 3 2 赤 RED MDin_A
Reserve2 15 MDin_A 8
4 Fuse_Reserve2 4 黄 YEL MDin_B 4 15 橙 ORN MDin_B
PER_OFF 16 MDin_B 8
4 Fuse_PWR_OFF 5 緑 GRN MDout_A 5 3 白 WHT MDout_A
D_GND 17 MDout_A 8
4 Fuse_GND6 6 青 BLU MDout_B 6 16 黄 YEL MDout_B
Zigma_ON_C 18 MDout_B 8
4 Zigma_ON_C 7 紫 SND_INTb_A 7 4 青 BLU SND_INTb_A
Zigma_ON_A 19 SND_INTb_A 8
4 Zigma_ON_A 8 灰 GRY SND_INTb_B 8 17 桃 PIK SND_INTb_B
D_GND 20 SND_INTb_B 8
4 Fuse_GND7 9 白 WHT GND 9 5 緑 GRN GND
MPI_GND1 8
10 黒 BLK GND 10 18 青 BLU GND
MPI_GND2 8
11 茶 BRN MCLK_A 11 6 白 WHT MCLK_A
C MCLK_A 8 C
12 赤 RED MCLK_B 12 19 緑 GRN MCLK_B
MCLK_B 8
13 橙 ORN MPI_IN1_A 13 7 赤 RED MPI_IN1_A
MPI_IN1_A 8
14 黄 YEL MPI_IN1_B 14 20 黄 YEL MPI_IN1_B
MPI_IN1_B 8
15 緑 GRN MPI_OUT1_A 15 8 黒 BLK MPI_OUT1_A
MPI_OUT1_A 8
16 青 BLU MPI_OUT1_B 16 21 橙 ORN MPI_OUT1_B
MPI_OUT1_B 8
17 紫 MPI_OUT2_A 17 9 桃 PIK MPI_OUT2_A
MPI_OUT2_A 8
18 灰 GRY MPI_OUT2_B 18 22 黒 BLK MPI_OUT2_B
MPI_OUT2_B 8
19 白 WHT N.C 19 10 赤 RED N.C
MPI_NC1 8
C 1A 20 黒 BLK N.C 20 23 桃 PIK N.C
MPI_NC2 8
21 茶 BRN N.C 21 11 白 WHT N.C
D 9 6 MPI_NC3 8 D
22 赤 RED N.C 22 24 橙 ORN N.C
MPI_NC4 8
CPU51A 6 23 橙 ORN N.C 23 12 青 BLU N.C
136Y101191 - MPI_NC5 8
24 黄 YEL N.C 24 25 黄 YEL N.C
MPI_NC6 8
25 緑 GRN N.C 25 13
+3V 1
CPU28
4 CPU_3V1 26 青 BLU N.C 26 X線源部
+3V 11
4 CPU_3V2
4 CPU_3V3
+3V 2 136Y101153 XIF
-12V 12 136Y101196 Y 7 136Y101385
4 CPU_-12V
GND 3
4 CPU_GND1 MPI4 1 茶 BRN Zigma_ON_C 1 MIF1 MIF2 1 黒 BLK Zigma_ON_C 1 XIF2 XIF1 1 1 X11
GND 13
4 CPU_GND2 2 赤 RED Zigma_ON_A 2 14 緑 GRN Zigma_ON_A 14 2 2
+5V 4
E 4 CPU_5V1 3 橙 ORN PREP_C 3 2 赤 RED PREP_C 2 3 3 E
PS_ON#14
4 CPU_PSON 4 黄 YEL PREP_A 4 15 橙 ORN PREP_A 15 4 4
GND 5
4 CPU_GND3 5 緑 GRN STATE_C 5 3 白 WHT STATE_C 3 5 5
GND 15
4 CPU_GND4 6 青 BLU STATE_A 6 16 黄 YEL STATE_A 16 6 6
+5V 6
4 CPU_5V2 7 紫 X-RAY_C 7 4 青 BLU X-RAY_C 4 7 7
GND 16
4 CPU_GND5 8 灰 GRY X-RAY_A 8 17 桃 PIK X-RAY_A 17 8 8
GND 7
4 CPU_GND6 9 白 WHT READY_C 9 5 緑 GRN READY_C 5 9 9
GND 17
4 CPU_GND7 10 黒 BLK READY_A 10 18 青 BLU READY_A 18 10 10
PWR_OK 8
4 CPU_PWROK 11 茶 BRN RSV_TX_C 11 6 白 WHT RSV_TX_C 6 11 11
-5V 18
4 CPU_-5V 12 赤 RED RSV_TX_A 12 19 緑 GRN RSV_TX_A 19 12 12
+5VSB 9
F 4 CPU_5VSB 13 橙 ORN RSV_RX_C 13 7 赤 RED RSV_RX_C 7 13 13 F
+5V 19
4 CPU_5V3 14 黄 YEL RSV_RX_A 14 20 黄 YEL RSV_RX_A 20 14 14
+12V 10
4 CPU_+12V 15 緑 GRN COM_TX_C 15 8 黒 BLK COM_TX_C 8 15 15
+5V 20
4 CPU_5V4 16 青 BLU COM_TX_A 16 21 橙 ORN COM_TX_A 21 16 16
17 紫 COM_RX_C 17 9 桃 PIK COM_RX_C 9 17 17
136Y101192 18 灰 GRY COM_RX_A 18 22 黒 BLK COM_RX_A 22 18 18
19 白 WHT FPD_ON_C 19 10 赤 RED FPD_ON_C 10 19 19
GND 1
CPU29
20 黒 BLK FPD_ON_A 20 23 桃 PIK FPD_ON_A 23 20 20
4 CPU_PWRGND1
+12V 3 21 茶 BRN RSV_TX_NC_C 21 11 白 WHT RSV_TX_NC_C 11
4 CPU_PWR12V1
GND 2 22 赤 RED RSV_TX_NC_A 22 24 橙 ORN RSV_TX_NC_A 24
4 CPU_PWRGND2
G +12V 4 23 橙 ORN RSV_RX_NC_C 23 12 青 BLU RSV_RX_NC_C 12 G
4 CPU_PWR12V2 XIF3 1
24 黄 YEL RSV_RX_NC_A 24 25 黄 YEL RSV_RX_NC_A 25
PCI_SLOTx MPI1 緑 GRN N.C 2
25 25 13 13
青 BLU N.C 3
26 26
27 紫 N.C 27
28 灰 GRY N.C 28
29 白 WHT N.C 29
30 黒 BLK N.C 30 制御部背面
機種 5010 ユニット -
継 クタ
P
A A
CPU51A 5
-
136Y101198
F5 CPU19
CFI1 1 1 1 1
136Y101194 2 2 2 2
B 3 3 3 3 B
CFI3 4 4
+5V 1 4 4
4 CFIPWR_5V
GND 2 フラットケーブル
4 CFIPWR_GND 37 37
37 37
38 38 38 38
39 39 39 39
40 40 40 40
CFI51A-CFI1 ←→ CPU51A-CPU19
C Signal name № Signal name C
PHDDRST# 1 2 GND
PDD(7) 3 4 PDD(8)
PDD(6) 5 6 PDD(9)
PDD(5) 7 8 PDD(10)
PDD(4) 9 10 PDD(11)
PDD(3) 11 12 PDD(12)
PDD(2) 13 14 PDD(13)
PDD(1) 15 16 PDD(14)
PDD(0) 17 18 PDD(15)
GND 19 20 KEY(NC)
PDDRQ 21 22 GND
PDIOW# 23 24 GND
D D
PDIOR# 25 26 GND
PIORDY# 27 28 CSEL(PD)
PDDACK# 29 30 GND 136Y101199
IRQ14 31 32 IOCS16#(NC) CN4
PDA1 33 34 PDMA66DET CPU5 MDI0+
PDA0 35 36 PDA2 1 1 N
PDCS1# 37 38 PDCS3* 2 MDI0- 2
HDDLED* 39 40 GND 3 MDI1+ 3
4 MDI1- 4
1 CPU30
5 MDI2+ 5
2 MDI2-
6 6
3 MDI3+
E 7 7 E
4 MDI3-
8 8
5
136Y101193 6
7 136Y101388
PWR_ON 灰 GRY 8
4 PWR_ON
9
PWR_COM 灰 GRY 10 1 MDI0+ 1
4 PWR_COM
11 2 MDI0- 2
12 3 MDI1+ 3
13 4 MDI1- 4
F 14 5 MDI2+ 5 F
15 6 MDI2- 6
16 7 MDI3+ 7
CPU51A-CN30 (Board Side)
17 8 MDI3- 8
Signal name № Signal name
PWR LED+ 1 14 HDD LED+ 18
PWR LED+ 2 15 Reserve 19 制御部 - HUB間LANケーブル
PWR LED- 3 16 HDD LED- 20
Error LED+ 4 17 P-SW+ (O/R)
Error LED+ 5 18 P-SW- GND 21
Error LED- 6 19 P-SW+ 22
EXT_SMI 7 20 P-SW- SHORT 灰 GRY 23
G +5V 8 21 RST-SW+ 24 G
GND 9 22 RST-SW-
SPEAKER- 10 23 P-SW+ 25
NC 11 24 +5VCNT 26
Internal SPEAKER 12 25 STANBY LED+
SPEAKER+ 13 26 STANBY LED-
機種 5010 ユニット -
D
A N A
SND51B 8
136Y101154 - 136Y101156
136Y101188
MZ1 CN9-2 CN9-1
1 青 BLU MZ1_BLUE 5 1 MZ1_BLUE 青 BLU 1 SND6 SND1 1 灰 GRY SZ1_VCC
SZ1_VCC 10
2 青 BLU MZ1_RED 4 2 MZ1_RED 青 BLU 2 2 茶 BRN SZ1_VOUT
SZ1_VOUT 10
3 青 BLU MZ1_ORANGE 3 3 MZ1_ORANGE 青 BLU 3 3 黒 BLK SZ1_GND
副走査モータ SZ1_GND 10
MZ1 4 青 BLU MZ1_GREEN 2 4 MZ1_GREEN 青 BLU 4 4 ツイスト
5 青 BLU MZ1_BLACK 1 5 MZ1_BLACK 青 BLU 5 5
B B
6
136Y101157
G G
機種 5010 ユニット -
D
A
N A
SND51B 7
136Y101146 -
136Y101158
MNb_A 黒 BLK 1 SND5
5 MNb_A SND3 1 灰
GRY TEN_N
MNb_B 緑 GRN 14 TEN_N 9
5 MNb_B 2 黒
BLK GND
MDin_A 赤 RED 2 TRC_DGND1 9
5 MDin_A 3 灰
GRY TDATA_O
MDin_B 橙 ORN 15 TDATA_O 9
5 MDin_B 4 黒
BLK GND
MDout_A 白 WHT 3 TRC_DGND2 9
5 MDout_A 5 灰
GRY TDATA_I
MDout_B 黄 YEL 16 TDATA_I 9
5 MDout_B 6 黒
BLK GND
SND_INTb_A 青 BLU 4 TRC_DGND3 9
B 5 SND_INTb_A 7 灰
GRY TCLK B
SND_INTb_B 桃 PIK 17 TCLK 9
5 SND_INTb_B 8 黒
BLK GND
GND 緑 GRN 5 TRC_DGND4 9
5 MPI_GND1 9 灰
GRY Start_N
GND 青 BLU 18 Start_N 9
5 MPI_GND2 10 黒
BLK GND
MCLK_A 白 WHT 6 TRC_DGND5 9
5 MCLK_A 11 灰
GRY R_LEDON
MCLK_B 緑 GRN 19 R_LEDON 9
5 MCLK_B 12 黒
BLK GND
MPI_IN1_A 赤 RED 7 TRC_DGND6 9
5 MPI_IN1_A 13 灰
GRY TRCINT_N
MPI_IN1_B 黄 YEL 20 TRCINT_N 9
5 MPI_IN1_B 14 黒
BLK GND
MPI_OUT1_A 黒 BLK 8 TRC_DGND7 9
5 MPI_OUT1_A 15 灰
GRY TRSV_OUT
MPI_OUT1_B 橙 ORN 21 TRSV_OUT 9
5 MPI_OUT1_B 16 黒
BLK GND
MPI_OUT2_A 桃 PIK 9 TRC_DGND8 9
C 5 MPI_OUT2_A 17 灰
GRY TRSV_IN C
MPI_OUT2_B 黒 BLK 22 TRSV_IN 9
5 MPI_OUT2_B 18 黒
BLK GND
N.C 赤 RED 10 TRC_DGND9 9
5 MPI_NC1 19
N.C 桃 PIK 23 ツイスト
5 MPI_NC2 20
N.C 白 WHT 11
5 MPI_NC3
N.C 橙 ORN 24
5 MPI_NC4
N.C 青 BLU 12
5 MPI_NC5
N.C 黄 YEL 25
5 MPI_NC6
13 136Y101159
136Y101152 高圧
136Y101152 ッ
D D
SND7 1 HigPow_+24V 灰 GRY 1 VH1
白 WHT +24V_P1 +24V_P1 1 SND9
3 SND_+24V1 2 GND 灰 GRY 2
+24V_P1 2
3 HV_EN_A 灰 GRY 3
青 BLU GND_P1 GND_P1 3
3 SND_GND1 4 25KV_SEL 灰 GRY 4
GND_P1 4
赤 RED +24V_P2 +24V_P2 5 5
3 SND_+24V2 5 GND 灰 GRY
+24V_P2 6 6
6 PLY1_A 灰 GRY
灰 GRY GND_P2 GND_P2 7 7
3 SND_GND2 7 PLY2_A 灰 GRY
GND_P2 8 8
8 PLY3_A 灰 GRY
9 9
10 GND 灰 GRY 10
E E
TNC1
1
2 136Y101147
1 SND2 HV_AG
閉端子 3
2 1 HV_AG 9
4
3
5
4
5 HV_OUT 1 HV_OUT 9
TNC2 6
1 7
F
2 8 F
閉端子 3 9
4 10
5 11
12
TNC3
1 136Y101160
2 1 SND11
SND13
3 2 1 橙 ORN GND
閉端子 ADC_GND 9
4 3 2 橙 ORN -24V
ADC_-24V 9
G 5 4 G
5
TNC4 6
1 7
2 8
閉端子 3 9
4 10
5 11
12
機種 5010 ユニット -
13
H 名称 回路図 補助名称 MARS[18x24] H
バイス
136Y101158
A A
TEN_N 灰 GRY 1 TRC2
8 TEN_N
GND 黒 BLK 2
8 TRC_DGND1 デバイス
TDATA_O 灰 GRY 3
8 TDATA_O
GND 黒 BLK 4
8 TRC_DGND2 1 DB1
TDATA_I 灰 GRY 5 8 HV_OUT
8 TDATA_I
GND 黒 BLK 6
8 TRC_DGND3
TCLK 灰 GRY 7
8 TCLK
GND 黒 BLK 8
8 TRC_DGND4
Start_N 灰 GRY 9
8 Start_N
GND 黒 BLK 10
B 8 TRC_DGND5 AD A B
R_LEDON 灰 GRY 11
8 R_LEDON
GND 黒 BLK 12 1 9 6
8 TRC_DGND6
TRCINT_N 灰 GRY 13 136Y101147
8 TRCINT_N ADC2
GND 黒 BLK 14 茶 BRN 1
8 TRC_DGND7 8 HV_AG
TRSV_OUT 灰 GRY 15
8 TRSV_OUT
GND 黒 BLK 16
8 TRC_DGND8
TRSV_IN 灰 GRY 17
8 TRSV_IN
GND 黒 BLK 18
8 TRC_DGND9
ツイスト 19 TRC4 ADC5
1 1
C
20 C
FPC51B 113Y1969 DB101
ADC制御信号 ADC101 1 1
100 80
71 71
TRC5 1 1 ADC3
136Y101152
FPC51B 113Y1969
+24V_A 黒 BLK TRC1 胸壁側信号 ・
3 TRC_+24V 1 ・
GND_A 茶 BRN 100 80
3 TRC_GND 2 ・
D 3 ・ D
・
TRC6 1 1 ADC4 ・
FPC51B 113Y1969 ・
反胸壁側信号 ・
100 80 ・
・
DB116
ADC116 1 1
136Y101160
E E
5 Image_signal[1:37] 71 71
TRC3 1 GND 橙 ORN GND 1 ADC1
TRC7 2 +3.3V 橙 ORN +3.3V 2
1
Image_signal2 3 GND 橙 ORN GND 3
2
4 +3.3V 橙 ORN +3.3V 4
Image_signal17 5 GND 橙 ORN GND 5
17
Image_signal18 6 +3.3V 橙 ORN +3.3V 6
18
Image_signal19 7 GND 橙 ORN GND 7
19
Image_signal20 8 -3.3V 橙 ORN -24V 8
20
9 橙 ORN GND 9
F F
Image_signal36 橙 ORN -3.3V 10
36
Image_signal37 37
136Y101145
TRC51A-TRC7 ←→ MPI51A-MPI2
G Signal name № Signal name G
GND 1 16 GND
DATA3+ 2 17 DATA7+
DATA3- 3 18 DATA7-
GND 4 19 GND 8 ADC_GND
CLK0+ 5 20 CLK1+
CLK0- 6 21 CLK1-
GND 7 22 GND 8 ADC_-24V
DATA2+ 8 23 DATA6+
DATA2- 9 24 DATA6-
GND 10 25 GND
DATA1+ 11 26 DATA5+ 機種 5010 ユニット -
DATA1- 12 27 DATA5-
H GND 13 28 GND 名称 回路図 補助名称 MARS[18x24] H
DATA0+ 14 29 DATA4+
DATA0- 15 30 DATA4- コード Z22N5010000E Rev. -
頁
富士フイルム株式会社 9
11
1 2 3 4 5 6 7 8 9 10 11
デハ イス 内
T A
A
136Y101156 A
136Y101161
SZ1_VCC 灰 GRY 1 JCT8 JCT7 1 灰 GRY SZ1_VCC 1
7 SZ1_VCC SZ1
SZ1_VOUT 茶 BRN 2 2 茶 BRN SZ1_VOUT 2
7 SZ1_VOUT 副走査HP検出
SZ1_GND 黒 BLK 3 3 黒 BLK SZ1_GND 3
7 SZ1_GND
ツイスト 4 4 ツイスト
5 5
6 6
B B
136Y101162
ER A
4
ERS_V1 灰 GRY 1 JCT4 JCT3 1 灰 GRY ERS_V1 1 ERS1
7 ERS_V1
ERS_V2 灰 GRY 2 2 灰 GRY ERS_V2 2
7 ERS_V2
ERS_GND1 灰 GRY 3 3 灰 GRY ERS_GND 3
7 ERS_GND1
ERS_GND2 灰 GRY 4 4 灰 GRY ERS_GND 4
C 7 ERS_GND2 C
ERS_LED_CHK 灰 GRY 5 5 灰 GRY ERS_LED_CHK 5
7 ERS_LED_CHK
LED_ADJ 灰 GRY 6 6 灰 GRY ERS_ADJ 6
7 LED_ADJ
ERS_GND3 灰 GRY 7 7 灰 GRY ERS_NC 7
7 ERS_GND3
D
136Y101163 AE A D
136Y101151
136Y101157 LE
7
F GND1 灰 GRY 1 JCT2 JCT1 1 FG 1 LED1 F
7 LED_GND1
LED_V1 灰 GRY 2 2 GND 2
7 LED_V1
GND2 灰 GRY 3 3 LED_V1 3
7 LED_GND2
LED_V2 灰 GRY 4 4 GND 4
7 LED_V2
GND3 灰 GRY 5 5 LED_V2 5
7 LED_GND3
LED_V3 灰 GRY 6 6 GND 6
7 LED_V3
GND4 灰 GRY 7 7 LED_V3 7
7 LED_GND4
8 8 GND 8
9 FG 9
金メッキFFCケーブル
G G
機種 5010 ユニット -
部 136Y101152 P A N I A
温調ユニット
136Y101164 136Y101168
N 136Y101167
CN10-J CN10-P CN11-1 CN11-2 PE1
緑 GRN +24V_T 1 1 橙
ORN +24V 1 PTC5 PTC1 1 青 BLU PE1_+24V 1 NFI2 NFI1 1 青
BLU PE1_+24V 1 1
赤 PEL_+24V
A 3 PTC_+24V1 ペルチェ A
黄 YEL +24V_T 3 3 橙
ORN +24V 2 3 青 BLU PE1_GND 6 6 青
BLU PE1_GND 2 2
黒_PEL_GND
3 PTC_+24V2
橙 ORN GND_T 2 2 橙
ORN
3 2 青 BLU PE2_+24V 2 2 青
BLU PE2_+24V
3 PTC_DGND1
桃 PIK GND_T 4 4 橙
ORN GND 4 4 青 BLU PE2_GND 7 7 青
BLU PE2_GND
3 PTC_DGND2
GND 5 青 BLU PO_+24V 3 3 青
BLU PO_+24V サーモスタット
6 青 BLU PO_GND 8 8 青
BLU PO_GND CN22-1 CN22-2 136Y101200
TN1 TN2
青 BLU PO_IL 4 4 青
BLU PO_IL 1 1 1 1
灰 GRY FAN2_+24V 9 9 灰
GRY FAN2_+24V 2 2
136Y101155
灰 GRY FAN2_FAIL 5 5 灰
GRY FAN2_FAIL
灰 GRY FAN2_GND 10 10 灰
GRY FAN2_GND
灰 GRY PTC_TxD 1 PTC4 PTC2 1
B 7 PTC_TxD B
灰 GRY PTC_RxD 2 2 CN12-J CN12-P
7 PTC_RxD
灰 GRY GND 3 3 PO_+24V 1 1 赤 P1_+24V 1 .
7 PTC_GND1
灰 GRY GND 4 PO_GND 2 2 黒_P1_GND 2 .
7 PTC_GND2 ポンプ
灰 GRY PTC_RESET_N 5 PO_IL 3 3 白_P1_IL 3 .
7 PTC_RESET_N
灰 GRY PTC_IN1 6
7 PTC_IN1
灰 GRY PTC_IN2 7
7 PTC_IN2
灰 GRY PTC_OUT1 8
7 PTC_OUT1
灰 GRY PTC_OUT2 9
7 PTC_OUT2 CN13
FAN2_+24V 灰 GRY 4 1 +24
C 3 2 NC C
FAN2_FAIL 灰 GRY 2 3 ALARM FAN7
FAN2_GND 灰 GRY 1 4 GND
デバイス室内
1 PTC7 PTC3 1 FAN4_+24V 灰 GRY 4 NFI6 NFI5 1 1 .
136Y101170 2 2 FAN4_GND 3 2 2 .
TDB51A 136Y101182 FAN9
113Y1961 3 3 FAN4_FAIL 灰 GRY 2 3 3 .
CN14-2 CN14-1 4 4 FAN3_+24V 灰 GRY 1 4
TDB1-2 4赤 TH2_+5V 1 4赤 TH2_+5V 赤 5 5 FAN3_GND
D
リード線 D
3黄 TH2_OUT 2 3黄 TH2_OUT 黄 6 6 FAN3_FAIL
2白 TH2_GND 3 2白 TH2_GND 白 7 7 FAN2_+24V
1黒 TH2_FG 4 1黒 TH2_FG 黒 8 8 FAN2_GND
TH3_+5V 赤 9 9 FAN2_FAIL
機内冷却風 TH3_OUT 黄 10 10 FAN1_+24V 灰 GRY 4 NFI4 NFI3 1 1 .
TH3_GND 白 11 11 FAN1_GND 3 2 2 . FAN8
TH3_FG 黒 12 12 FAN1_FAIL 灰 GRY 2 3 3 .
TDB51A 灰 GRY 1 4
113Y1961
リード線
E TDB1-3 4赤 TH3_+5V
E
3黄 TH3_OUT
2白 TH3_GND
1黒 TH3_FG
機内雰囲気側
PTC8 1 4 NFI8 NFI7 1 1 .
2灰
GRY
3 2 2 . FAN10
3灰
GRY
2 3 3 .
4 1 4
5灰
GRY
リード線
6灰
GRY
F F
136Y102110
デバイス室内
CN16-1 CN16-2
灰 GRY 3 1 1 .
灰 GRY 2 2 2 . 機内循環用ファン
G FAN6 (シロッコファン) G
灰 GRY 1 3 3 .
136Y101386 136Y101387
X20 CN30-R CN30-P RSW
1黒 黒1 1 1
2白 白2 2 2 機種 5010 ユニット -
シス ム 源
PSU51A 125N100011 温調電源PTC基板用 136Y101183
A A
Slot4 Slot4_GND 1 +24V_GND 橙 ORN
S_Slot4_24VG1 3
2 +24V_GND 橙 ORN
S_Slot4_24VG2 3
S_Slot1
オプション(25m)
B
136Y101178 ※組図No.136Y101532
Slot1 Slot1_GND 1 +24V_GND 橙 ORN
B
S_Slot1_24VG1 3
10
9
8
7
6
5
4
3
2
1
2 +24V_GND 橙 ORN
S_Slot1_24VG2 3
1 L 1
L1 Slot1_24V 1 +24V 赤 RED
1 N 1 S_Slot1_24V1 3
N2 2 +24V 赤 RED
1 E 1 S_Slot1_24V2 3
FG2
システム電源各基板用
C C
システム電源CPU基板用
136Y101184
Slot2 Slot2_GND
S_Slot2 1 +5V_GND 橙 ORN
S_Slot2_5VG1 3
2 +5V_GND 橙 ORN
S_Slot2_5VG2 3
オプション(15m)
10
9
8
7
6
5
4
3
2
1
136Y101201 ※組図No.136Y101531
Slot2_5V 1 +5V 赤 RED
S_Slot2_5V1 3
TB1 2 +5V 赤 RED
S_Slot2_5V2 3
(MBK6/E-Z-4P)
D L 黒 BLK L D
L 1 1 1
1 N 1 1 白 WHT N 136Y101185
N
1 E 1 1 緑/黄 E Slot3
FG Slot3_GND 1 +12V_GND 橙 ORN
S_Slot3_12VG1 3
2 +12V_GND 橙 ORN
S_Slot3_12VG2 3
S_Slot3
136Y101382 Slot3_12V 1 +12V 赤 RED
S_Slot3_12V1 3
136Y101383 2 +12V 赤 RED
S_Slot3_12V2 3
10
9
8
7
6
5
4
3
2
1
136Y101384
E システム電源CPU基板用 E
TB2
(BTB50C3) 136Y101186
丸端子 端子台
SLOT1_5VSB 灰 GRY
S_Slot1_5VSB 3
TB2-LOUT TB1-L
ブレーカー SLOT1_CNT 灰 GRY
S_Slot1_CNT 3
1 L 1 1 L 黒 BLK (SER-F-11-62F-3A) SLOT2_5VSB 灰 GRY
L_X S_Slot2_5VSB 3
SLOT2_CNT 灰 GRY
TB2-L
NFB_IN S_Slot2_CNT 3
SLOT3_5VSB 灰 GRY
1 L 黒 BLK 黒 BLK L 1 NFB51A S_Slot3_5VSB 3
SLOT3_CNT 灰 GRY
N2 136Y101181 113Y1979 S_Slot3_CNT 3
F F
TB2-NOUT TB1-N NFB_OUT NFB1 NFB2 136Y101202
1 N 1 1 N 白 WHT
N_X 1 L 黒 BLK 1 1
TB2-N 2 2 3 AC_IN
1 N 白 WHT 白 WHT N白 WHT 3 2 L 黒 BLK 1 SCN1 1
4 N 白 WHT 2 2
TB2-EOUT TB1-E
E E 黄/緑 3
FG_X 1 1 1 E 緑/黄 1 +5Vスタンバイ電源用
AC_E 4
TB2-E 5
1 E 緑/黄 E 6
G ファストン端子 G
136Y101180 FG_POW
1
1
136Y101381 ※組図No.136Y101533 FG_OUT
機種 5011 ユニット -
頁
富士フイルム株式会社 2
11
1 2 3 4 5 6 7 8 9 10 11
制御部背面中継コネクタ
A A A A
Y
PAT51A 4
-
136Y101183
136Y101152
橙 ORN +24V_GND 1 PAT1 136Y101189
2 S_Slot4_24VG1
赤 RED +24V 5
2 S_Slot4_24V1 CN1-P CN1-J
橙 ORN +24V_GND 2
2 S_Slot4_24VG2 PAT11 1 橙 ORN +24V_P1 1 1 白 WHT +24V_P1
赤 RED +24V 6 SND_+24V1 8
B 2 S_Slot4_24V2 6 橙 ORN GND_P1 6 6 青 BLU GND_P1 B
橙 ORN +24V_GND 3 SND_GND1 8
2 S_Slot1_24VG1 2 橙 ORN +24V_P2 2 2 赤 RED +24V_P2
赤 RED +24V 7 SND_+24V2 8
2 S_Slot1_24V1 7 橙 ORN GND_P2 7 7 灰 GRY GND_P2
橙 ORN +24V_GND 4 SND_GND2 8
2 S_Slot1_24VG2 3 橙 ORN +24V_A 3 3 黒 BLK +24V_A
赤 RED +24V 8 TRC_+24V 9
2 S_Slot1_24V2 8 橙 ORN GND_A 8 8 茶 BRN GND_A
TRC_GND 9
4 橙 ORN +24V_T1 4 4 緑 GRN +24V_T1
PTC_+24V1 11
9 橙 ORN +24V_T2 9 9 黄 YEL +24V_T2
PTC_+24V2 11
5 橙 ORN GND_T1 5 5 橙 ORN GND_T1
PTC_DGND1 11
10 橙 ORN GND_T2 10 10 桃 PIK GND_T2
PTC_DGND2 11
C C
136Y101184
136Y101185
G G
機種 5011 ユニット -
頁
富士フイルム株式会社 3
11
1 2 3 4 5 6 7 8 9 10 11
136Y101191
C C
136Y101192
136Y101194
PAT9
1 +5V 橙 ORN
CFIPWR_5V 6
2 +5V_GND 橙 ORN
CFIPWR_GND 6
E E
IND51A
113Y1962
136Y101190
PAT12 1 +5V_L 灰 GRY 1 IND1
2 +5V_DGND 灰 GRY 2
D_GND 1 PAT10
5 Fuse_GND1 3 +5V_L 灰 GRY 3
+24V_P1_Det 2
5 Fuse_+24V_P1_Det 4 +5V_DGND 灰 GRY 4
+24V_P2_Det 3
5 Fuse_+24V_P2_Det 5 +5V_L 灰 GRY 5
D_GND 4
5 Fuse_GND2 6 +5V_DGND 灰 GRY 6
+24V_A_Det 5
5 Fuse_+24V_A_Det
F +24V_F_Det 6 F
5 Fuse_+24V_F_Det 136Y101195
D_GND 7
5 Fuse_GND3 パワーオン表示灯
+5V_L_Det 8
5 Fuse_+5V_L_Det
+5V_SB_Det 9
5 Fuse_+5V_SB_Det
D_GND 10
5 Fuse_GND4
+24V_T_Det 11
5 Fuse_+24V_T_Det
+24V_TF_Det 12
5 Fuse_+24V_TF_Det
D_GND 13
5 Fuse_GND5
Reserve1 14
5 Fuse_Reserve1
Reserve2 15
5 Fuse_Reserve2
G PER_OFF 16 G
5 Fuse_PWR_OFF
D_GND 17
5 Fuse_GND6
Zigma_ON_C 18
5 Zigma_ON_C
Zigma_ON_A 19
5 Zigma_ON_A
D_GND 20
5 Fuse_GND7
機種 5011 ユニット -
頁
富士フイルム株式会社 4
11
1 2 3 4 5 6 7 8 9 10 11
136Y101190 Image_signal[1:37] 9
MPI2
1
D_GND 1 MPI5 Y Image_signal2
4 Fuse_GND1 2 136Y101145 (他社製造)
+24V_P1_Det 2
4 Fuse_+24V_P1_Det
A +24V_P2_Det 3 Image_signal17 A
4 Fuse_+24V_P2_Det 17
D_GND 4 Image_signal18
4 Fuse_GND2 18
+24V_A_Det 5 Image_signal19
4 Fuse_+24V_A_Det 19
+24V_F_Det 6 Image_signal20
4 Fuse_+24V_F_Det 20
D_GND 7
4 Fuse_GND3
+5V_L_Det 8 Image_signal36
4 Fuse_+5V_L_Det 36
+5V_SB_Det 9 Image_signal37
4 Fuse_+5V_SB_Det 37
D_GND 10
4 Fuse_GND4 136Y101146 (他社製造)
+24V_T_Det 11
4 Fuse_+24V_T_Det
+24V_TF_Det 12
4 Fuse_+24V_TF_Det MPI3 1 茶 BRN MNb_A 1 MIF3 F A 1 白 WHT MNb_A
B D_GND 13 MIF4 MNb_A 8 B
4 Fuse_GND5 2 赤 RED MNb_B 2 14 黒 BLK MNb_B
Reserve1 14 Y MNb_B 8
4 Fuse_Reserve1 3 橙 ORN MDin_A 3 2 赤 RED MDin_A
Reserve2 15 MDin_A 8
4 Fuse_Reserve2 4 黄 YEL MDin_B 4 15 黒 BLK MDin_B
PER_OFF 16 MDin_B 8
4 Fuse_PWR_OFF 5 緑 GRN MDout_A 5 3 緑 GRN MDout_A
D_GND 17 MDout_A 8
4 Fuse_GND6 6 青 BLU MDout_B 6 16 黒 BLK MDout_B
Zigma_ON_C 18 MDout_B 8
4 Zigma_ON_C 7 紫 SND_INTb_A 7 4 黄 YEL SND_INTb_A
Zigma_ON_A 19 SND_INTb_A 8
4 Zigma_ON_A 8 灰 GRY SND_INTb_B 8 17 黒 BLK SND_INTb_B
D_GND 20 SND_INTb_B 8
4 Fuse_GND7 9 白 WHT GND 9 5 茶 BRN GND
MPI_GND1 8
10 黒 BLK GND 10 18 黒 BLK GND
MPI_GND2 8
11 茶 BRN MCLK_A 11 6 青 BLU MCLK_A
C MCLK_A 8 C
12 赤 RED MCLK_B 12 19 黒 BLK MCLK_B
MCLK_B 8
13 橙 ORN MPI_IN1_A 13 7 灰 GRY MPI_IN1_A
MPI_IN1_A 8
14 黄 YEL MPI_IN1_B 14 20 黒 BLK MPI_IN1_B
MPI_IN1_B 8
15 緑 GRN MPI_OUT1_A 15 8 橙 ORN MPI_OUT1_A
MPI_OUT1_A 8
16 青 BLU MPI_OUT1_B 16 21 黒 BLK MPI_OUT1_B
MPI_OUT1_B 8
17 紫 MPI_OUT2_A 17 9 桃 PIK MPI_OUT2_A
MPI_OUT2_A 8
18 灰 GRY MPI_OUT2_B 18 22 黒 BLK MPI_OUT2_B
MPI_OUT2_B 8
19 白 WHT N.C 19 10 水 N.C
MPI_NC1 8
C A 20 黒 BLK N.C 20 23 黒 BLK N.C
MPI_NC2 8
21 茶 BRN N.C 21 11 紫 N.C
D Y MPI_NC3 8 D
22 赤 RED N.C 22 24 黒 BLK N.C
MPI_NC4 8
CPU51A 6 23 橙 ORN N.C 23 12 赤 RED N.C
136Y101191 - MPI_NC5 8
24 黄 YEL N.C 24 25 白 WHT N.C
MPI_NC6 8
25 緑 GRN N.C 25 13
+3V 1
CPU28
4 CPU_3V1 26 青 BLU N.C 26 X線源部
+3V 11
4 CPU_3V2
+3V 2 X A
4 CPU_3V3 136Y101153 (他社製造)
-12V 12 136Y101196 Y 7 136Y101385
4 CPU_-12V
GND 3
4 CPU_GND1 MPI4 1 茶 BRN Zigma_ON_C 1 MIF1 MIF2 1 白 WHT Zigma_ON_C 1 XIF2 XIF1 1 1 X11
GND 13
4 CPU_GND2 2 赤 RED Zigma_ON_A 2 14 黒 BLK Zigma_ON_A 14 2 2
+5V 4
E 4 CPU_5V1 3 橙 ORN PREP_C 3 2 赤 RED PREP_C 2 3 3 E
PS_ON#14
4 CPU_PSON 4 黄 YEL PREP_A 4 15 黒 BLK PREP_A 15 4 4
GND 5
4 CPU_GND3 5 緑 GRN STATE_C 5 3 緑 GRN STATE_C 3 5 5
GND 15
4 CPU_GND4 6 青 BLU STATE_A 6 16 黒 BLK STATE_A 16 6 6
+5V 6
4 CPU_5V2 7 紫 X-RAY_C 7 4 黄 YEL X-RAY_C 4 7 7
GND 16
4 CPU_GND5 8 灰 GRY X-RAY_A 8 17 黒 BLK X-RAY_A 17 8 8
GND 7
4 CPU_GND6 9 白 WHT READY_C 9 5 茶 BRN READY_C 5 9 9
GND 17
4 CPU_GND7 10 黒 BLK READY_A 10 18 黒 BLK READY_A 18 10 10
PWR_OK 8
4 CPU_PWROK 11 茶 BRN RSV_TX_C 11 6 青 BLU RSV_TX_C 6 11 11
-5V 18
4 CPU_-5V 12 赤 RED RSV_TX_A 12 19 黒 BLK RSV_TX_A 19 12 12
+5VSB 9
F 4 CPU_5VSB 13 橙 ORN RSV_RX_C 13 7 灰 GRY RSV_RX_C 7 13 13 F
+5V 19
4 CPU_5V3 14 黄 YEL RSV_RX_A 14 20 黒 BLK RSV_RX_A 20 14 14
+12V 10
4 CPU_+12V 15 緑 GRN COM_TX_C 15 8 橙 ORN COM_TX_C 8 15 15
+5V 20
4 CPU_5V4 16 青 BLU COM_TX_A 16 21 黒 BLK COM_TX_A 21 16 16
17 紫 COM_RX_C 17 9 桃 PIK COM_RX_C 9 17 17
136Y101192 18 灰 GRY COM_RX_A 18 22 黒 BLK COM_RX_A 22 18 18
19 白 WHT FPD_ON_C 19 10 水 FPD_ON_C 10 19 19
GND 1
CPU29
20 黒 BLK FPD_ON_A 20 23 黒 BLK FPD_ON_A 23 20 20
4 CPU_PWRGND1
+12V 3 21 茶 BRN RSV_TX_NC_C 21 11 紫 RSV_TX_NC_C 11
4 CPU_PWR12V1
GND 2 22 赤 RED RSV_TX_NC_A 22 24 黒 BLK RSV_TX_NC_A 24
4 CPU_PWRGND2
G +12V 4 23 橙 ORN RSV_RX_NC_C 23 12 赤 RED RSV_RX_NC_C 12 G
4 CPU_PWR12V2 XIF3 1
24 黄 YEL RSV_RX_NC_A 24 25 白 WHT RSV_RX_NC_A 25
PCI_SLOTx MPI1 緑 GRN N.C 2
25 25 13 13
青 BLU N.C 3
26 26
27 紫 N.C 27
28 灰 GRY N.C 28
29 白 WHT N.C 29
30 黒 BLK N.C 30 制御部背面
機種 5011 ユニット -
頁
富士フイルム株式会社 5
11
1 2 3 4 5 6 7 8 9 10 11
面 クタ
A 4 0 A
CPU51A 5
-
136Y101198
1 1 CPU19
1 Y 77 CFI1 1 1 C
136Y101194 2 2 2 2
3 3 PCI_SLOTx CN1 Y 0
B 3 3 B
CFI3 4 4
+5V 1 4 4
4 CFIPWR_5V
GND 2 フラットケーブル
4 CFIPWR_GND 37 37
37 37
38 38 38 38
39 39 39 39
40 40 40 40
CFI51A-CFI1 ←→ CPU51A-CPU19
PCI_SLOTx CN1 Y 0
C Signal name № Signal name C
PHDDRST# 1 2 GND
PDD(7) 3 4 PDD(8)
PDD(6) 5 6 PDD(9)
PDD(5) 7 8 PDD(10)
PDD(4) 9 10 PDD(11)
PDD(3) 11 12 PDD(12)
PDD(2) 13 14 PDD(13)
PDD(1) 15 16 PDD(14)
PDD(0) 17 18 PDD(15)
GND 19 20 KEY(NC) C
PDDRQ 21 22 GND
PDIOW# 23 24 GND PCI_SLOTx CN1 Y 0
D D
PDIOR# 25 26 GND
PIORDY# 27 28 CSEL(PD)
PDDACK# 29 30 GND
IRQ14 31 32 IOCS16#(NC)
PDA1 33 34 PDMA66DET
PDA0 35 36 PDA2
PDCS1# 37 38 PDCS3*
HDDLED* 39 40 GND
E
1 CPU30 E
2
3
4 136Y101199
5 CN4
136Y101193 6 CPU5 1 MDI0+ 1 AN
7 MDI0-
PWR_ON 灰 GRY 2 2
4 PWR_ON 8 MDI1+
3 3
9 MDI1-
PWR_COM 灰 GRY 4 4
4 PWR_COM 10 MDI2+
F
5 5 F
11 MDI2-
6 6
12 MDI3+
7 7
CPU51A-CN30 (Board Side) 13 MDI3-
8 8
Signal name № Signal name 14
PWR LED+ 1 14 HDD LED+ 15
PWR LED+ 2 15 Reserve 16
PWR LED- 3 16 HDD LED- 17
Error LED+ 4 17 P-SW+ (O/R)
Error LED+ 5 18 P-SW- GND 18
Error LED- 6 19 P-SW+ 19
EXT_SMI 7 20 P-SW- 20
G +5V 8 21 RST-SW+ 21 G
GND 9 22 RST-SW-
SPEAKER- 10 23 P-SW+ 22
SHORT 灰 GRY 23
NC 11 24 +5VCNT
Internal SPEAKER 12 25 STANBY LED+ 24
SPEAKER+ 13 26 STANBY LED- 25
26
機種 5011 ユニット -
頁
富士フイルム株式会社 6
11
1 2 3 4 5 6 7 8 9 10 11
影
ND
A Y A
SND51B 8
136Y102000 - 136Y101156
136Y101188
MZ1 CN9-2 CN9-1
1 青 BLU MZ1_BLUE 5 1 MZ1_BLUE 茶 BRN 1 SND6 SND1 1 灰 GRY SZ1_VCC
SZ1_VCC 10
2 青 BLU MZ1_RED 4 2 MZ1_RED 茶 BRN 2 2 茶 BRN SZ1_VOUT
SZ1_VOUT 10
3 青 BLU MZ1_ORANGE 3 3 MZ1_ORANGE 茶 BRN 3 3 黒 BLK SZ1_GND
副走査モータ SZ1_GND 10
MZ1 4 青 BLU MZ1_GREEN 2 4 MZ1_GREEN 茶 BRN 4 4 ツイスト
5 青 BLU MZ1_BLACK 1 5 MZ1_BLACK 茶 BRN 5 5
B B
6
136Y101157
G G
機種 5011 ユニット -
頁
富士フイルム株式会社 7
11
1 2 3 4 5 6 7 8 9 10 11
ND
A
Y A
SND51B 7
136Y101146 (他社製造) -
136Y101158
MNb_A 白 WHT 1 SND5
5 MNb_A SND3 1 灰
GRY TEN_N
MNb_B 黒 BLK 14 TEN_N 9
5 MNb_B 2 黒
BLK GND
MDin_A 赤 RED 2 TRC_DGND1 9
5 MDin_A 3 灰
GRY TDATA_O
MDin_B 黒 BLK 15 TDATA_O 9
5 MDin_B 4 黒
BLK GND
MDout_A 緑 GRN 3 TRC_DGND2 9
5 MDout_A 5 灰
GRY TDATA_I
MDout_B 黒 BLK 16 TDATA_I 9
5 MDout_B 6 黒
BLK GND
SND_INTb_A 黄 YEL 4 TRC_DGND3 9
B 5 SND_INTb_A 7 灰
GRY TCLK B
SND_INTb_B 黒 BLK 17 TCLK 9
5 SND_INTb_B 8 黒
BLK GND
GND 茶 BRN 5 TRC_DGND4 9
5 MPI_GND1 9 灰
GRY Start_N
GND 黒 BLK 18 Start_N 9
5 MPI_GND2 10 黒
BLK GND
MCLK_A 青 BLU 6 TRC_DGND5 9
5 MCLK_A 11 灰
GRY R_LEDON
MCLK_B 黒 BLK 19 R_LEDON 9
5 MCLK_B 12 黒
BLK GND
MPI_IN1_A 灰 GRY 7 TRC_DGND6 9
5 MPI_IN1_A 13 灰
GRY TRCINT_N
MPI_IN1_B 黒 BLK 20 TRCINT_N 9
5 MPI_IN1_B 14 黒
BLK GND
MPI_OUT1_A 橙 ORN 8 TRC_DGND7 9
5 MPI_OUT1_A 15 灰
GRY TRSV_OUT
MPI_OUT1_B 黒 BLK 21 TRSV_OUT 9
5 MPI_OUT1_B 16 黒
BLK GND
MPI_OUT2_A 桃 PIK 9 TRC_DGND8 9
C 5 MPI_OUT2_A 17 灰
GRY TRSV_IN C
MPI_OUT2_B 黒 BLK 22 TRSV_IN 9
5 MPI_OUT2_B 18 黒
BLK GND
N.C 水 10 TRC_DGND9 9
5 MPI_NC1 19
N.C 黒 BLK 23 ツイスト
5 MPI_NC2 20
N.C 紫 11
5 MPI_NC3
N.C 黒 BLK 24
5 MPI_NC4
N.C 赤 RED 12
5 MPI_NC5
N.C 白 WHT 25
5 MPI_NC6
13 136Y102001 高圧
136Y101152 (他社製造) ト
136Y101152 (他社製造)
D
125N100010 D
SND7 1 HigPow_+24V 茶 BRN 1 VH1
白 WHT +24V_P1 +24V_P1 1 SND9
3 SND_+24V1 2 GND 茶 BRN 2
+24V_P1 2
3 HV_EN_A 茶 BRN 3
青 BLU GND_P1 GND_P1 3
3 SND_GND1 4 25KV_SEL 茶 BRN 4
GND_P1 4
5 1.7mA_SEL 茶 BRN 5
赤 RED +24V_P2 +24V_P2 5
3 SND_+24V2 6 GND 茶 BRN 6
+24V_P2 6
7 PLY1_A 茶 BRN 7
灰 GRY GND_P2 GND_P2 7
3 SND_GND2 8 PLY2_A 茶 BRN 8
GND_P2 8
9 PLY3_A 茶 BRN 9
10 GND 茶 BRN 10
E E
TNC1
1
2 136Y101147
1 SND2
閉端子 3
2 HV_AG 1 HV_AG 9
4
3
5 136Y101148 (デバイス附属ハーネス)
4
5 HV_OUT 1 HV_OUT 9
TNC2 6
1 7
F
2 8 F
閉端子 3 9
4 10
5 11
12
TNC3
1 136Y101160
2 1 SND11
SND13
3 2 1 橙 ORN GND
閉端子 ADC_GND 9
4 3 2 橙 ORN -24V
ADC_-24V 9
G 5 4 G
5
TNC4 6
1 7
2 8
閉端子 3 9
4 10
5 11
12
機種 5011 ユニット -
13
H 名称 回路図 補助名称 MARS[24x30] H
頁
富士フイルム株式会社 8
11
1 2 3 4 5 6 7 8 9 10 11
バ ス 内
C A
Y 7
136Y101158
A A
TEN_N 灰 GRY 1 TRC2
8 TEN_N
GND 黒 BLK 2
8 TRC_DGND1 デバイス
TDATA_O 灰 GRY 3
8 TDATA_O 136Y101148 (デバイス附属ハーネス)
GND 黒 BLK 4
8 TRC_DGND2 1 DB1
TDATA_I 灰 GRY 5 8 HV_OUT
8 TDATA_I
GND 黒 BLK 6
8 TRC_DGND3
TCLK 灰 GRY 7
8 TCLK
GND 黒 BLK 8
8 TRC_DGND4
Start_N 灰 GRY 9
8 Start_N
GND 黒 BLK 10
B 8 TRC_DGND5 A C B
R_LEDON 灰 GRY 11
8 R_LEDON
GND 黒 BLK 12 Y
8 TRC_DGND6
TRCINT_N 灰 GRY 13 136Y101147
8 TRCINT_N ADC2
GND 黒 BLK 14 茶 BRN 1
8 TRC_DGND7 8 HV_AG
TRSV_OUT 灰 GRY 15
8 TRSV_OUT
GND 黒 BLK 16
8 TRC_DGND8
TRSV_IN 灰 GRY 17
8 TRSV_IN
GND 黒 BLK 18
8 TRC_DGND9
ツイスト 19 TRC4 ADC5
1 1
C
20 C
FPC51B 113Y1969 DB101
ADC制御信号 ADC101 1 1
100 80
71 71
TRC5 1 1 ADC3
136Y101152 (他社製造)
FPC51B 113Y1969
+24V_A 黒 BLK TRC1 胸壁側信号 ・
3 TRC_+24V 1 ・
GND_A 茶 BRN 100 80
3 TRC_GND 2 ・
D 3 ・ D
・
TRC6 1 1 ADC4 ・
FPC51B 113Y1969 ・
反胸壁側信号 ・
100 80 ・
・
DB120
ADC120 1 1
136Y101160
E E
5 Image_signal[1:37] 71 71
TRC3 1 GND 橙 ORN GND 1 ADC1
TRC7 2 +3.3V 橙 ORN +3.3V 2
1
Image_signal2 3 GND 橙 ORN GND 3
2
4 +3.3V 橙 ORN +3.3V 4
Image_signal17 5 GND 橙 ORN GND 5 DB121
17
Image_signal18 6 +3.3V 橙 ORN +3.3V 6 ADC6 1 1
18
Image_signal19 7 GND 橙 ORN GND 7
19
Image_signal20 8 -3.3V 橙 ORN -24V 8
20
9 橙 ORN GND 9 5 5
F F
Image_signal36 橙 ORN -3.3V 10
36
Image_signal37 37
136Y101145 (他社製造)
TRC51A-TRC7 ←→ MPI51A-MPI2
G Signal name № Signal name G
GND 1 16 GND
DATA3+ 2 17 DATA7+
DATA3- 3 18 DATA7-
GND 4 19 GND 8 ADC_GND
CLK0+ 5 20 CLK1+
CLK0- 6 21 CLK1-
GND 7 22 GND 8 ADC_-24V
DATA2+ 8 23 DATA6+
DATA2- 9 24 DATA6-
GND 10 25 GND
DATA1+ 11 26 DATA5+ 機種 5011 ユニット -
DATA1- 12 27 DATA5-
H GND 13 28 GND 名称 回路図 補助名称 MARS[24x30] H
DATA0+ 14 29 DATA4+
DATA0- 15 30 DATA4- コード Z22N5011000 Rev.
頁
富士フイルム株式会社 9
11
1 2 3 4 5 6 7 8 9 10 11
バ ス 内
J T5
136Y101156
A
95 A
136Y102002
SZ1_VCC 灰 GRY 1 JCT8 JCT7 1 灰 GRY SZ1_VCC 1
7 SZ1_VCC SZ1
SZ1_VOUT 茶 BRN 2 2 茶 BRN SZ1_VOUT 2
7 SZ1_VOUT 副走査HP検出
SZ1_GND 黒 BLK 3 3 黒 BLK SZ1_GND 3
7 SZ1_GND
ツイスト 4 4 ツイスト
5 5
6 6
B B
136Y102003
ER
Y 7
ERS_V1 灰 GRY 1 JCT4 JCT3 1 茶 BRN ERS_V1 1 ERS1
7 ERS_V1
ERS_V2 灰 GRY 2 2 茶 BRN ERS_V2 2
7 ERS_V2
ERS_GND1 灰 GRY 3 3 茶 BRN ERS_GND 3
7 ERS_GND1
ERS_GND2 灰 GRY 4 4 茶 BRN ERS_GND 4
C 7 ERS_GND2 C
ERS_LED_CHK 灰 GRY 5 5 茶 BRN ERS_LED_CHK 5
7 ERS_LED_CHK
LED_ADJ 灰 GRY 6 6 茶 BRN ERS_ADJ 6
7 LED_ADJ
ERS_GND3 灰 GRY 7 7 茶 BRN ERS_NC 7
7 ERS_GND3
D
136Y102004 E 5 D
12 AEC1
AEC_FG 灰 GRY 11 JCT6 JCT5 茶 BRN AEC_FG Y 7
7 AEC_FG 11 11
AEC_TxD 灰 GRY 10 10 茶 BRN AEC_TxD 10
7 AEC_TxD
AEC_RxD 灰 GRY 9 9 茶 BRN AEC_RxD 9
7 AEC_RxD
AEC_GND1 灰 GRY 8 8 茶 BRN AEC_GND1 8
7 AEC_GND1
AEC_IN1 灰 GRY 7 7 茶 BRN AEC_IN1 7
7 AEC_IN1
AEC_IN2 灰 GRY 6 6 茶 BRN AEC_IN2 6
7 AEC_IN2
AEC_OUT1 灰 GRY 5 5 茶 BRN AEC_OUT1 5
7 AEC_OUT1
AEC_OUT2 灰 GRY 4 4 茶 BRN AEC_OUT2 4
7 AEC_OUT2
AEC_+5V 灰 GRY 3 3 茶 BRN AEC_+5V 3
7 AEC_+5V
E AEC_GND2 灰 GRY 2 2 茶 BRN AEC_GND2 2 E
7 AEC_GND2
AEC_-5V 灰 GRY 1 1 茶 BRN AEC_-5V 1
7 AEC_-5V
136Y102005
136Y101157 D
Y 0 7
F GND1 灰 GRY 1 JCT2 JCT1 1 FG 1 LED1 F
7 LED_GND1
LED_V1 灰 GRY 2 2 GND 2
7 LED_V1
GND2 灰 GRY 3 3 LED_V1 3
7 LED_GND2
LED_V2 灰 GRY 4 4 GND 4
7 LED_V2
GND3 灰 GRY 5 5 LED_V2 5
7 LED_GND3
LED_V3 灰 GRY 6 6 GND 6
7 LED_V3
GND4 灰 GRY 7 7 LED_V3 7
7 LED_GND4
8 8 GND 8
9 FG 9
金メッキFFCケーブル
G G
機種 5011 ユニット -
頁
富士フイルム株式会社 10
11
1 2 3 4 5 6 7 8 9 10 11
影 A A
温調ユニット
136Y101152 (他社製造) 136Y101164 136Y102006
Y Y 136Y101167
CN10-J CN10-P CN11-1 CN11-2 PE1
緑 GRN +24V_T 1 1 橙
ORN +24V 1 PTC5 PTC1 1 青 BLU PE1_+24V 1 NFI2 NFI1 1 青
BLU PE1_+24V 1 1
赤 PEL_+24V
A 3 PTC_+24V1 ペルチェ A
黄 YEL +24V_T 3 3 橙
ORN +24V 2 3 青 BLU PE1_GND 6 6 青
BLU PE1_GND 2 2
黒_PEL_GND
3 PTC_+24V2
橙 ORN GND_T 2 2 橙
ORN
3 2 青 BLU PE2_+24V 2 2 青
BLU PE2_+24V リード線
3 PTC_DGND1
桃 PIK GND_T 4 4 橙
ORN GND 4 4 青 BLU PE2_GND 7 7 青
BLU PE2_GND
3 PTC_DGND2
GND 5 青 BLU PO_+24V 3 3 青
BLU PO_+24V サーモスタット
6 青 BLU PO_GND 8 8 青
BLU PO_GND CN22-1 CN22-2 136Y101200
TN1 TN2
青 BLU PO_IL 4 4 青
BLU PO_IL 1 1 1 1
茶 BRN FAN2_+24V 9 9 灰
GRY FAN2_+24V 2 2
136Y101155
茶 BRN FAN2_FAIL 5 5 灰
GRY FAN2_FAIL
茶 BRN FAN2_GND 10 10 灰
GRY FAN2_GND
灰 GRY PTC_TxD 1 PTC4 PTC2 1
B 7 PTC_TxD B
灰 GRY PTC_RxD 2 2 CN12-J CN12-P
7 PTC_RxD
灰 GRY GND 3 3 PO_+24V 1 1 赤 P1_+24V 1 .
7 PTC_GND1
灰 GRY GND 4 PO_GND 2 2 黒_P1_GND 2 .
7 PTC_GND2 ポンプ
灰 GRY PTC_RESET_N 5 PO_IL 3 3 白_P1_IL 3 .
7 PTC_RESET_N
灰 GRY PTC_IN1 6
リード線
7 PTC_IN1
灰 GRY PTC_IN2 7
7 PTC_IN2
灰 GRY PTC_OUT1 8
7 PTC_OUT1 136Y102006
灰 GRY PTC_OUT2 9
7 PTC_OUT2 CN13
FAN2_+24V 灰 GRY 4 1 +24
C 3 2 NC C
FAN2_FAIL 灰 GRY 2 3 ALARM FAN7
FAN2_GND 灰 GRY 1 4 GND
デバイス室内
1 PTC7 PTC3 1 FAN4_+24V 茶 BRN 4 NFI6 NFI5 1 1 .
136Y101170 2 2 FAN4_GND 3 2 2 .
TDB51A 136Y101182 FAN9
113Y1961 3 3 FAN4_FAIL 茶 BRN 2 3 3 .
CN14-2 CN14-1 4 4 FAN3_+24V 茶 BRN
1 4
TDB1-2 4赤 TH2_+5V 1 4赤 TH2_+5V 赤 5 5 FAN3_GND
D
リード線 D
3黄 TH2_OUT 2 3黄 TH2_OUT 黄 6 6 FAN3_FAIL
2白 TH2_GND 3 2白 TH2_GND 白 7 7 FAN2_+24V
1黒 TH2_FG 4 1黒 TH2_FG 黒 8 8 FAN2_GND
TH3_+5V 赤 9 9 FAN2_FAIL
機内冷却風 TH3_OUT 黄 10 10 FAN1_+24V 茶 BRN 4 NFI4 NFI3 1 1 .
TH3_GND 白 11 11 FAN1_GND 3 2 2 . FAN8
TH3_FG 黒 12 12 FAN1_FAIL 茶 BRN 2 3 3 .
TDB51A 茶 BRN 1 4
113Y1961
リード線
E TDB1-3 4赤 TH3_+5V
E
3黄 TH3_OUT
2白 TH3_GND
1黒 TH3_FG
機内雰囲気側
PTC8 1 FAN5_+24V 茶 BRN 4 NFI8 NFI7 1 1 .
2 FAN5_GND 3 2 2 . FAN10
3 FAN5_FAIL 茶 BRN 2 3 3 .
4 茶 BRN 1 4
5 136Y102006 リード線
6
F F
136Y102006
デバイス室内
CN16-1 CN16-2
茶 BRN 3 1 1 .
茶 BRN 2 2 2 . 機内循環用ファン
G FAN6 (シロッコファン) G
茶 BRN 1 3 3 .
リード線
136Y101387
136Y101386
X20 CN30-R CN30-P RSW
1黒 黒1 1 灰
GRY
1
2白 白2 2 灰
GRY
2 機種 5011 ユニット -
08 部品番号検索表
PARTS NOS. SEARCH TABLE
Part No. INDEX No.-Drawing No Part No. INDEX No.-Drawing No Part No. INDEX No.-Drawing No Part No. INDEX No.-Drawing No Part No. INDEX No.-Drawing No Part No. INDEX No.-Drawing No
113N100091/A/B 03B-6 136Y101202B/C 06-24 350N101272 02C-6 376N100021A 01B-7 W11G10140045 02C-2
113Y100273C 03B-16 136Y101382A 06-25 350N101744B 04C-3 376N100023B 01B-8 W11G10141960 02D-1
113Y1962D 03B-9 136Y101383 06-26 350N101834A 04B-13 376N100039 01B-19 W11G10415588 01A-1
113Y1963D 03B-2 136Y101384 06-27 350N2774C 03C-2 379Y100034E/F 02A-3 W11G10415589 01A-2
113Y1964B 03B-3 136Y101385B 06-28 350Y100595L 02A-6 386N100335A 04C-4.4 W11G10415592 02E-1
113Y1965C 03B-8 136Y101386 06-29 350Y100820L 02A-5 386N100336A 04C-4.2 W11G10415596 02C-3
113Y1970C 02A-2 136Y101387D 06-30 350Y100861B 01B-12 386N100337A 04C-4.1 W11G10415606 04A-1
113Y1977A 03B-5 136Y101389 05-5 350Y100929C 01B-21 386N100338A 04C-4.3 W11G10415612 01B-11
113Y1979A 03C-8 136Y101533 06-31 350Y100930C 01B-20 386Y100040 04C-4 W11G10415613 01B-14
113Y1980A 03B-1 136Y102382C 04B-6 350Y100931D 01B-13 405N101115 01B-17 W11G10415615 01B-15
114N100013 03B-4 136Y102395B 04B-10 350Y100933C 04A-6 814Y100132A/B/C 03B-7 W11G10415616 01B-16
114N100026 03B-17 136Y102825A 04B-12 350Y100935B 04A-5 848Y100054F 02A-1 W11G10415617 01B-5
119S0043 03C-1 136Y102826A 04B-11 350Y100949A 01B-6 848Y100065B 02A-8 W11G10415617 02B-4
119Y100028B 02A-7 137S1374 03B-12 350Y100950A 01B-1 W11G03082146 02E-5 W11G10415617 02C-4
120S5268 04A-12 137S1417 02A-12 350Y101005 02A-10 W11G06343144 01A-9 W11G10415618 02B-8
125Y100032 03C-4 137S1417 03B-14 350Y101006A 02A-9 W11G07061989 02D-5 W11G10415618 02C-8
128S1088 03C-7 137S1420 03B-11 350Y101488 04C-2 W11G10139621 02D-2 W11G10415619 02B-9
133Y100033A 05-1.1 137S1423 03B-15 350Y101550 04B-3 W11G10139644 02E-4 W11G10415619 02C-9
133Y100036A 05-1.2 137S1436 03B-13 350Y101551 04B-4 W11G10139656 04A-2 W11G10415626 02B-3
135S0829 03C-5 137S1456 03B-10 350Y101589 04B-8 W11G10139755 04A-3 W11G10501290 02B-1
135Y100008 03C-6 301S7000310 04B-20 350Y101608 04B-1 W11G10139779 02E-2 W11G10501290 02C-1
136N100143 04B-14 304S1000820 03C-9 350Y101609 04B-2 W11G10139824 02B-7.6 W11G10501291 02D-4
136Y101145 06-1 304S1000820 04A-7 351Y100004F 03A-4 W11G10139824 02C-7.6 W11G10501292 02E-3
136Y101146 06-2 305N100033 03C-3 356N104378B 04B-5 W11G10139846 01A-8 W11G14409787 04A-11
136Y101152 06-3 305N100033 04A-10 356N104569 04B-7 W11G10139907 01A-7
136Y101153 06-4 308S0414 04B-18 356N104614C 04B-9 W11G10139931 02D-3
136Y101179 05-4 308S9780612 04B-19 356N104719 04C-1.2 W11G10139940 02B-7.1
136Y101180A 06-5 309S0120008 03C-11 356N105032 04B-15 W11G10139940 02C-7.1
136Y101181A 06-6 309S0120008 04A-9 356S1138 04C-7 W11G10139942 02B-7.2
136Y101183 06-7 309S0220008 03C-10 356Y100803 02E-6 W11G10139942 02C-7.2
136Y101184A 06-8 309S0220008 04A-8 356Y100836 05-3 W11G10139943 02B-7.5
136Y101185A 06-9 315S0057 04C-2.2 356Y101132A 04C-1 W11G10139943 02C-7.5
136Y101186A 06-10 316S0256 04B-16 356Y101134B 04C-6 W11G10139944 02C-7.8
136Y101187 06-11 316S1082 04B-17 357Y100226A 04A-4 W11G10139950 02B-5.1
136Y101189 06-12 316S2094 05-7 357Y100257 05-2 W11G10139952 02B-5.2
136Y101190A 06-13 316S2106 04C-5 357Y100270 05-10 W11G10139964 02C-5
136Y101191 06-14 316S4016 04C-2.1 358Y100111A 04B-21 W11G10139970 02B-7.3
136Y101192 06-15 317N100022B 03A-3 359Y100048 02A-4 W11G10139970 02C-7.3
136Y101193 06-16 345N1823D 03A-1 360N100074G 01B-4 W11G10139972 02B-7.4
136Y101194 06-17 346N100689 04C-1.1 370N100155 05-6 W11G10139972 02C-7.4
136Y101195A 06-18 346N100690 04C-1.3 370N100156 05-8 W11G10139974 02C-7.7
136Y101196 06-19 350N100868C 01B-2 370N100272 03C-13 W11G10140032 01A-3
136Y101197 06-20 350N100930B 01B-9 370Y100224 05-9 W11G10140033 01A-4
136Y101198 06-21 350N100931B 01B-10 370Y100306 02A-11 W11G10140034 01A-6
136Y101199C 06-22 350N101267 01B-18 376N0238 03A-2 W11G10140036 01A-5
136Y101200B 06-23 350N101272 02B-6 376N100016A 01B-3 W11G10140045 02B-2
09 締結用および配線用のサービス部品一覧表
List of Service Parts for Securing and Wiring
1. 締結用および配線用のサービス部品について 1. Service Parts for Securing and Wiring
● 発注の際は部品コードと発注数量を明記すること。発注数量は 100 個単位とする。 l Clearly indicate the part code and quantity required when ordering service parts.
Service parts can be ordered in sets of 100.
2. ねじ類
2. Screws
● サービス部品一覧表の中のねじの記載方法の例を以下に示す。この記載方法はサービ
スマニュアル内の記載方法と同じである。 l The following is an example of how screws in the service parts table are described.
This description method is the same as that in this Service Manual.
09 締結用および配線用のサービス部品一覧表
List of Service Parts for Securing and Wiring
ネジ ナット
参考情報 参考情報
記号 部品コード Reference information 記号 部品コード Reference information
Symbol Parts code 形状 部品名称 材質、処理など Symbol Parts code 形状 部品名称 材質、処理など
Shape Parts name Material/treatment, etc. Shape Parts name Material/treatment, etc.
鋼 ステンレス
座金組込み ばね座金 + みがき丸座金 六角ナット 不動態化処理
Na
十字穴付六角ボルト 亜鉛メッキ + 白色 3 価クロメート Hexagon nut Stainless steel
BR4x20 308S0410 Cross recessed 色:うすい白色 Passive state treatment
BR6x12 308S9780612 hexagon head Steel
screw with spring Zinc plating/White trivalent
and plain washers chromating
Color: Light white 配線用部品
鋼
参考情報
亜鉛メッキ + 白色 3 価クロメート
六角頭 TP ネジ 記号 部品コード Reference information
TP3x6 308S0414 色:うすい白色
Cross recessed Symbol Parts code
TP3x12 308S0419 Steel 形状 部品名称
hexagon head TP
TP4x6 308S0416 Zinc plating/White trivalent Shape Parts name
screw
chromating
LWS-0711Z 316S1027
Color: Light white
LWS-1211Z 316S1028
ばね・平座金組込 コードクランプ
六角穴付ボルト LWSM-0511 316S0259 Code clamp
Hexagon socket
BQ6x16
head cap screws LWSM-0306 316S0258
with spring and
plain washer
六角穴付ボルト コードクランプ
LES-0510 316S0256
Q6x40 Hexagon socket Code clamp
head cap screws
六角穴付 EMT-3N 316S1309 クランプ
ボタンボルト Clamp
EMT-6N 316S1310
Hexagon socket
button head screws 束線バンド
PLT2S 316S1082
Cable tie
Cross recessed
S3x10 301S7000310 十字穴付サラ小ネジ
flat-head screw
FDR MS-1000
03.14.2011 03 Revision (FM5842) All pages
Service Manual
...........
(1) Turn ON the circuit breaker of the distribution switchboard.
...........
...........
(3) Press the system start button on the control pad.
Run... #1 [Click]
(4) Turn ON the power of the AWS.
.... Run ?
(5) Press the AWS's [Windows] key to open the Start menu of Windows. .... Type the name of a program, folder, document, or
Internet resource, and Windows will open it for you.
EACH RU SETTING
CONNECTION TEST
RU NAME LIST OF EXISTING RU
RU IP ADDR
BACKUP RESTORE
DRMS9004.EPS
EDIT CL NAME
BACKUP RESTORE
ERROR LOG
EXECUTE EXECUTE
(8). #1
(7). #3
[Click]
ERROR-DB [Scroll]
window <Display example>
(7). #4
[Click]
(7). #5
[Check]
Details of
the error
DRMS9005.ai
[Click]
“Error Log”
DRMS9007.ai
#1
[Scroll]
#2 Scroll bar
[Check]
Error log window “Error Log”
DRMS9008.ai
DRMS9009.ai
DRMS9010.ai
[Check]
“Exposure counter”
value
DRMS9011.ai
DETAIL A DETAIL B
[Retighten]
Anchor bolt x5
DRMS9012.EPS
DRMS9013 ai
DRMS9014.ai
#1
[Remove] Swivel arm
Face guard lower cover
#2
[Remove]
Compression DRMS9016.ai
DRMS9017.ai
DRMS9015.ai
#2
Swivel arm [Remove]
right-hand side cover Filter cover/
air filter
#1
[Remove]
Bind screw
(M3x6)
DRMS9023.ai
DRMS9018.ai
DRMS9019.ai
[Remove]
Air filter FPD lower cover
DRMS9025.ai
DRMS9024.ai
Side Cover
(1) Remove the air filter from the swivel arm right-hand side cover.
Swivel arm
right-hand side cover
Filter retaining
bracket
[Clean]
Air filter
DRMS9027.ai
#2
#3 [Remove]
[Remove] Filter retaining
Air filter bracket
DRMS9026.ai
[Remove]
Air filter
Filter cover
DRMS9025.ai
DRMS9024.ai
Swivel arm
right-hand side cover
#1
[Remove]
TP3x6
#2
#3 [Remove]
[Remove] Filter retaining
Air filter bracket
DRMS9026.ai
Reserve tank
DETAIL A
Side cover of the GOOD NO GOOD
temperature adjustment unit
Joint of the reserve tank
The liquid level of the coolant is higher The liquid level of the coolant is lower
[Remove] than the joint of the reserve tank. than the joint of the reserve tank.
TP3x6 (x5) DRMS9029.ai
A [Check]
Single tank unit joint
(whole circumference)
DETAIL A DRMS9031.ai
[Check]
Single tank unit
retaining screw
DRMS9030.ai
(1) Reinstall the filter cover and the air filter of the carbon cover.
Swivel arm
right-hand side cover
Filter cover/
air filter
[Tighten] DRMS9018.ai
Bind screw (4) Reinstall the swivel arm left-hand side cover.
(M3x6) {MC:2.1.3_Swivel Arm Left-Hand Side Cover}
DRMS9043.ai
Swivel arm
left-hand side cover
DRMS9019.ai
#2
[Reinstall]
Swivel arm Face guard
lower cover
#1
DRMS9016.ai
[Reinstall]
Compression plate
(6) Reinstall the swivel arm top cover.
{MC:2.1.1_Swivel Arm Top Cover}
DRMS9057.ai
DRMS9015.ai
(1) Check to make sure that there is no gap between the positioner
retainer and the positioner when the 1.8-fold magnification exposure
stand S is locked.
Positioner retainer
DRMS9099.ai
DRMS9098.ai
{MC:9_Generator Adjustment}
{Calibration (CA)}
DRMS9100.ai
DRMS9101.ai
Test the protective grounding as needed, in compliance with the local regulations.
Measure the leakage current and the patient leakage current as needed, in
compliance with the local regulations.
FDR MS-1000
03.14.2011 03 Revision (FM5842) All pages
Service Manual
Installation (IN)
2.2 Tools and Measuring Instruments Necessary for 14. Changing the Swivel Arm Height (Uppermost Position)
Installation
15. Removing the Covers
3. Precautions in Installation
4. Transfer 16. Removing the Packing Material of the Swivel Arm (2)
4.1 Unloading and Preparation for Transfer 17. Unpacking and Temporarily Installing the FPD
18.2 Removing the Inner Cover and the Cable Retaining Bracket
7. Installing the AWS
18.3 Mounting the FPD
8. Installing the Controller Unit
18.4 Connecting the Cable Connectors and Attaching the
Cable Retaining Brackets
9. Connecting the Power Supply Cable
18.5 Putting on the Inner Cover
9.1 Connecting the Power Supply Cable of the Controller Unit
18.6 Attaching the Grid Guide
9.2 Connecting the Power Supply Cable of the Mammographic
Stand 18.7 Changing the DIP Switch Setting on the CBA (D905) Board
9.3 Checking the Wiring Connections and Connecting to 18.8 Putting on the Swivel Arm Right- and Left-Hand Front Covers
the Distribution Switchboard
18.9 Putting on the Compression Unit Right- and Left-Hand
Side Covers
10. Removing the Packing Material of the Swivel Arm (1)
18.10 Checking the Grid Position (1)
11. Installing the Mammographic Stand
18.11 Putting on the Carbon Cover
19. Installing the Temperature Adjustment Unit 25.1 Confirmation of AEC Accuracy
20. Putting on the Covers 25.2 Confirmation of Dose per Exposure Mode
20.1 Putting on the FPD Right- and Left-Hand Covers (Right- 26. Checking the Variation in X-ray Output
and Left-Hand Armrests)
27. Checking for Image Problems and Checking the S Value
20.2 Putting on the FPD Lower Cover
20.3 Putting on the Swivel Arm Right- and Left-Hand Side Covers 28. Checking the Operation of the Mammographic Stand
20.4 Putting on the Swivel Arm Top and Lower Covers 28.1 Checking the Elevation of the Swivel Arm
20.5 Putting on the Elevation Stand Covers 28.2 Checking the Rotation of the Swivel Arm
21. Installing the RU Software 28.3 Checking the Emergency Stop Switches
21.3 Setting the AWS and Checking Connection 29. Clearing the Error Log
21.4 Installing the RU Software 30. Protective Grounding Test (Outside Japan)
21.5 Installing the Machine-Specific Data
31. Measurement of Leakage Current and Patient Leakage Current
21.6 Replacing the Controller Unit AEC File (Outside Japan)
21.7 Replacing and Merging the AEC File in the AWS
32. Visual Inspection and Cleaning of the Machine
21.8 Backing Up the RU Data
33. Measurement of Power Source Impedance [Only Inside Japan]
22. Time Setting for the Mammographic Stand
34. Measurement of Installation Environment
23. FPD Calibration
34.1 Measurement of Installation Environment (Temperature
24. Checking the Exposure Recognition Function and Humidity)
END
- The machine includes portions where components are intricately assembled. When Closure joint
DRMS8161.ai
accessing the screws for retaining those components, use of a flexible screwdriver
kit as a jig improves working efficiency. Water
Note, however, the flexible screwdriver kit is not a required tool. It is recommended supply jig Closure hose
that the flexible screwdriver kit be prepared in advance when you wish to improve
the working efficiency for installing the machine.
Flexible screwdriver shaft Used to fill or drain the coolant.
(including one set of (Cable length: approx. 5.0 m)
Water supply
screwdriver bits) The water supply jig does not
jig 1
Closure joint include the coolant.
(for AC200 V)
Torque screwdriver Parts No.: 133Y100036
Set of hexagonal bits
(2.0/2.5/3.0/4.0/5.0/6.0 mm)
DRMS8162.ai
Transformer
- The adjuster bolts attached to the pedestal are turned to adjust the tilt of the Adapter
mammographic stand. It is recommended that you prepare a wrench of 13x13 mm.
DRMS8280.ai
A DETAIL A
A Adjuster bolt
Pedestal A Used to temporarily place
13 mm the FPD. Also used when
A attaching/detaching the
A Stand jig 1
temperature adjustment unit.
13 mm
DRMS8259.ai
Parts No.: 357Y100257
DRMS8159 ai
Acrylic board
1
(40 mm thick)
DRMS8183.ai
DRMS8184.ai
DRMS8185.ai
[Check]
LED (PSU MAINS)
DRMS8192.ai Terminals of
- LED on the MFI board: The yellow LED is unlit. power supply cable
DRMS8206 ai
[Check] LED
DRMS8193 ai
CAUTIONS
- When unloading the machine, two or more persons should always cooperate Handles DRMS8170.ai
to do so.
- Wear safety shoes for operation.
- Before unloading the machine, secure a proper machine transfer route. u INSTRUCTIONS u
- The mass of the mammographic stand is approx. 320 kg. When the crate - Fully leave the machine with the plastic cover put on in the room. If the cover is put
packing the mammographic stand is to be unloaded from the platform of the off immediately after the machine is transferred into the room, the machine might
truck, confirm the gravity center for balance. get condensation.
Mammographic stand - Do not remove the protective sheets and the packing material until instructed in the
manual.
Do not remove the protective Do not remove the protective
sheet and the packing material sheet of the collimator until
of the grid until instructed. instructed.
DRMS8171.ai
WARNINGS
- The mass of the mammographic stand is approx. 320 kg. When transferring
the mammographic stand, two or more persons should always cooperate to
do so.
- If the machine is to be transferred over differing floor levels, slowly move the
machine not to shock it. Note that the carrier can come over a differing level
up to approx. 10 mm.
CAUTION
The mass of the mammographic stand is approx. 320 kg. When unloading the
mammographic stand and standing it up, four or more persons should always
cooperate to do so.
DRMS8284 ai
DRMS8281.ai
FPD right-hand
3-3 cover (right-hand 1
armrest)
DRMS8285.ai
2-1 Controller unit 1
DRMS8410.ai
DRMS8289.ai
DRMS8401.ai
(Q6x15) DRMS8303.ai
Spring washer
2
4-20 Washer For retaining the FPD (SW8)
4 DRMS8305.ai
(W6) DRMS8304.ai
Hexagon socket head For retaining the
Spring washer 4-27 cap screw 6 elevation stand rear
4 (Q6x25) DRMS8303.ai
cover
(SW6) DRMS8305.ai
CAUT ON
Performance check
4-35 1 Only inside Japan
sheet
DRMS8401.ai
Confirmation of use
4-36 1 Only inside Japan
of the AMULET
DRMS8416.ai
Mammographic
7-1 1
stand
5-1 Control pad 1
DRMS8312.ai
DRMS8318.ai
Elevation stand
8-1 right-hand side 1 Elevation stand left-
9-1 1
cover hand side cover
DRMS8320.ai
DRMS8322 ai
DRMS8323.ai
Elevation stand
8-2 1
rear cover
DRMS8321 ai
DR 1000 DR L Z MBL KIT E Mobile kit 1 - Only for use in Europe and Australia.
POSITIONER KIT BIOPSY #(E) Biopsy positioner kit 1 Includes the face guard W.
STEREO EXPOSURE KIT #(E) Stereo exposure kit 1
FACE GUARD W #(E) Face guard W 1 Needed when using the stereo exposure kit.
FRONT COVER KIT BIOPSY 18×24 #(E) Front cover kit (18x24) 1 Needed when using the biopsy positioner kit
FRONT COVER KIT BIOPSY 24×30 #(E) Front cover kit (24x30) 1 or the stereo exposure kit.
BIOPSY AC CORD 20M # Biopsy power supply cable (20 m) 1 Only inside Japan.
Options (FDR-1000AWS)
Refer to the AWS Service Manual for the detailed information.
*1: The operation desk or the equipment rack must be fixed to the floor with the anchor bolts. If anchor bolt holes cannot be made on the floor as requested by the customer, however, you need use
the optional anchor plate to fix the operation desk and the equipment rack.
*2: Attach the compression plate wall-hanging holder to a sufficiently rigid wall.
REFERENCE
Anchor plate When the operation desk is to be used in combination with the separately available
(separately available equipment rack, fix the operation desk by means of the anchor bolts and the anchor
accessory) plate only on the operation desk.
DRMS8155 ai
Couple the operation desk with the rear of the equipment rack via the bolts to fix.
[Combination with the equipment rack] <When the operation desk is to be combined with the equipment rack
via the anchor bolts.>
Equipment rack Operation desk
(separately
[When viewed from the [When viewed from the available accessory) [Retain]
Operation equipment rack] operation desk] Anchor bolt
desk [Retain] M8x20 (x2)/
Equipment rack M6x12 (x2)/ SW8 (x2)/
(separately SW6 (x2)/ W8 (x2)
W2 (x2) <standard accessory>
available Operation Operation
accessory) desk desk
+
via the anchor plate.>
Equipment rack Operation desk
(separately
[Retain]
available accessory)
Anchor bolt
M8x20 (x2)/
Equipment rack Equipment rack [Retain] SW8 (x2)/
(separately available (separately available M6x12 (x2)/ W8 (x2)
accessory) accessory) SW6 (x2)/ <standard accessory>
DRMS8156 ai W2 (x2) Anchor plate
(separately available
accessory)
DRMS8260.ai
Operation desk
Operation desk
[Remove]
Front cover
#2
DRMS8084.ai [Remove]
Lower shelf plate
DRMS8085.ai
DETAIL A
Floor
#1 [Loosen]/
0.5-1.0 mm #3 [Tighten]
Screw
A A
DRMS8009.ai
#1 [Loosen]/
#3 [Tighten]
Screw
#1 #2 #3 #4
[Insert] [Attach] [Embed] [Remove] DETAIL A
Anchor nut x4 Disposable Anchor nut Disposable #2 Spring washer (SW8)
M8x35 bolt x4 bolt <standard accessory>
[Secure]
<standard accessory> M8x15 Anchor bolt Washer (W8)
(recommended) M8x20 (x2) <standard accessory>
<standard
35 to accessory>
37 mm
Machine
main Pedestal
DRMS8010 ai body
DRMS8158 ai
(12) Reinstall the lower shelf plate which is removed in the procedure (5).
(14) Attach the optional X-ray protective board to the operation desk.
The procedure need not be taken when the optional X-ray protective board is not to
be attached to the operation desk.
{IN:Appendix 7._Installing the Optional X-Ray Protective Board}
(15) Put on the front cover of the operation desk which is removed in the
procedure (4).
[Remove]
T4x8 (x6)
DRMS8020 ai
(2) Mark the anchor nut embedded positions on the floor with an oil-
based pen, a center punch or the like.
DETAIL A
GOOD NO GOOD
A A
Hold the stay. Do not hold the cover to lift.
A A
DRMS8022 ai
DETAIL A
[Mark]
With center punch
DRMS8257.ai
Floor
0.5-1.0 mm
DRMS8009.ai
A A #1 #2 #3 #4
[Insert] [Attach] [Embed] [Remove]
Disposable
A A Anchor nut x4
bolt
Anchor nut Disposable
DETAIL A M8x35
M8x15
x4 bolt
<standard accessory>
(recommended)
35 to
[Drilling]
37 mm
4 places Drill
Drill diameter: Φ12.5
Depth: 35-37 mm DRMS8010.ai
DRMS8062.ai
A A
A A
#1
DETAIL A [Secure]
#2 Anchor bolt
[Put] (M8x20 (x2))
Anchor cover <standard
<standard accessory>
accessory> Spring washer
(SW8)
<standard
Controller accessory>
unit Plain washer
(W8)
<standard
accessory>
Pedestal
DRMS8107.ai
CAUTION
u INSTRUCTIONS u
- Use the locally obtained power supply cables for the controller unit, conforming to Do not connect the power supply cable to the distribution switchboard until
the local specifications. instructed in the manual, to avoid electric shock hazards.
- Connect the controller unit power supply cable through the conduit. The diameter of
a hole for the conduit cable tube (metallic) on the left-hand side of the controller unit u NOTE u
is Φ40 mm.
The connection procedure varies depending on whether a single-phase two-wire
system or single-phase three-wire system is employed. Refer to the procedures
DETAIL A conforming to the installation site to make power supply cable connections. The
factory default setting provides the connection to a single-phase two-wire system.
Inner cover
A Φ40 mm [Remove]
TP3x6 (x26)
Controller unit
REFERENCE DRMS8021.ai
#1
[Insert]
Power supply cable
#2 #1 Terminal
[Fix] [Connect] block (TB1)
Nut Power supply cable (x3)
L1 N L2
TB-N/L2 (white)
TB-L1 (black)
Conduit tube TB-E (green)
#2
[Secure]
Clamp
DRMS8174.ai
DRMS8173.ai
[Remove]
TP3x6 (x26)
#1
[Insert]
Power supply cable
#2
DRMS8021.ai
[Fix]
Nut
Conduit tube
DRMS8173 ai
Controller unit
Controller unit
#1
[Connect] Terminal
Power supply cable (x3) block (TB1)
[Change
connection] Terminal block (TB1)
TB-N/L2 (white)
L1 N L2
TB-N OUT_cable
L1 N L2
connected with the
N-terminal
TB-L1 (black)
TB-E (green)
#2
[Secure]
Clamp Terminal block (TB1) TB-N OUT_cable
DRMS8175.ai connected with the
L1 N L2
L2-terminal
DRMS8176.ai
Mammographic Stand
Connect the power supply cable of the mammographic stand to the controller unit and Controller unit
the mammographic stand.
(1) Insert the power supply cable of the mammographic stand into the
controller unit.
Controller unit
[Insert]
Power supply
cable
<standard
accessory>
DRMS8026.ai
Terminal block
(TB2)
(2) Retain the power supply cable of the mammographic stand.
[Tighten]
TP3x6 (x2)
<standard accessory>
Power supply
cable [Remove]
Cover
DRMS8027 ai
DRMS8028 ai
#1
[Route]
Power supply cable
x3
#3
[Secure]
Clamp x3
Power supply unit
DRMS8029.ai
DRMS8178.ai
(5) Reinstall the cover of the terminal block (TB2).
Reverse the removal procedures for reinstallation.
Ground cable
Power supply
cable
#2 #2 DETAIL B
FG [Connect]
Ground cable removed in procedure (7)
Ground cable
(mammographic stand)
#1 B
[Disconnect] W6 (x3) Na6
Na6 (x2)/W6 (x5) FG
A DRMS8179.ai
Ground cable
#3 #3 #3 DETAIL C
[Connect]
L1 cable (black)/N cable (white)
(power supply cable of
the mammographic stand)
C
Terminal
L1 block
(black) N (white)
Power supply cable DRMS8180.ai
Elevation stand
Be sure to put on the terminal block
cover. #1
Terminal block cover [Route]
[Tighten]
Power supply cable
TP3x6 (x2)
18 mm or less
(recommended value)
Cable retaining
bracket #2
[Tighten]
Na
DRMS8095.ai
CAUTION
Follow the local regulations in connecting the power supply cable of the
machine to the distribution switchboard.
(3) Connect the network cable (standard accessory) to the CBS board
(D900).
DRMS8196.ai
Connector
[Connect] (X10)
Network cable
<standard accessory of
the mammographic stand>
Insert the connector (Control Box X2) into its position before retaining with The machine cannot boot up while the elevation stand cover is not put on. The
screws. If the connector is not fully inserted, the machine cannot normally boot emergency stop short-circuit cable need be connected to the emergency stop
up. switch connector when the machine is to be booted up without the elevation
stand cover on.
Mammographic stand
Emergency stop short-
A circuit cable
Mammographic stand
DRMS8013.ai DRMS8014 ai
Mammographic stand
Terminals of
power supply cable
DRMS8206 ai
#2 [Verify]
LED lit (PSU MAINS)
DRMS8186 ai
#2 [Press]
System start button
DETAIL A DRMS8188 ai
LED lights up
DRMS8187.ai
Control panel
Left button
Right button
[Remove]
DRMS8189 ai
Packing material
DRMS8015.ai
[Press]
System OFF button
Control panel
[Press] DRMS8191.ai
Up button
(14) Turn OFF the circuit breaker of the distribution switchboard.
Packing material
DRMS8190 ai
[Check]
LED (PSU MAINS)
DRMS8192.ai
MFI board
[Check] LED
DRMS8193 ai
96 mm
DRMS8258.ai
DETAIL A
A
A
[Mark]
A
5 places #1
Center punch A [Drilling]
DRMS8059 ai
A Drill 5 places
(3) Move the mammographic stand away to a location not interfering with
the drilling of anchor holes.
DRMS8060 ai
(5) With a vacuum cleaner or like device, remove concrete chips and dust
from the drilled holes.
Level
[Adjust]
Adjuster bolt
Adjuster bolt
DRMS8092.ai
Controller unit
NIF4
MPI2
CN1
[Connect]
Network cable
A <locally obtained
Cables
part> DRMS8058.ai
(3) Connect the network cable connected with the connector of the
DETAIL A controller unit to the AWS HUB.
#2 [Retain]
Cables
[Remove/Tighten]
TP3x6
DRMS8195.ai
#1
[Tighten]
T4x8 (x6)
#2
[Install]
Cap x6
<standard
accessory>
DRMS8327.ai
DRMS8326.ai
(4) Route the connected cables to the rear of the mammographic stand
and fix there.
Connector
(1) (X23)
[Connect]
Foot switch cable x2 Connector
(X22) CBS board (D900) DRMS8197.ai
#1 [Remove]/ #2
#3 [Retain] [Route]
Na x2 Cables
Mammographic stand
DRMS8199.ai
Unit: mm
DRMS8200 ai
(*): The lowermost position (the position where the swivel arm is lowered to its lowest
position) does not change as the elevation lower-limit switch is not repositioned.
DRMS8201.ai
DRMS8202.ai
(2) Cut the cable tie which retains the swivel arm elevation upper-limit
switch cable.
[Cut]
Cable tie
DRMS8204.ai
(4) Retain the swivel arm elevation upper-limit switch cable with the cable
tie.
Reverse the removal procedures for reinstallation.
(5) Put on the elevation stand inner cover (upper).
Reverse the removal procedures for reinstallation.
(6) Turn ON the circuit breaker of the distribution switchboard.
(7) Turn ON the circuit breaker of the controller unit.
(8) Press the system start button on the control pad.
Limit switch cable
(9) Turn ON the power of the AWS.
DRMS8203.ai
(10) Press the Lift Up/Down button of the swivel arm control panel to move
up/down the swivel arm. Check to make sure that the swivel arm
stops at the uppermost position demanded by the customer.
WARNINGS
(1) Press the right or left button on the swivel arm control panel to turn
the swivel arm to the 0° position. l Be sure to turn OFF the circuit breaker of the distribution switchboard to
avoid electric shock hazards.
(2) Press the down button on the swivel arm control panel to lower the
swivel arm to the lowermost position. l Electrical charge (residual charge) accumulated by a capacitor might remain
in a circuit after the circuit breaker of the distribution switchboard is turned
Swivel arm OFF.
(5) #1 Be sure to proceed with operations after you are sure that the charge has
Control panel
[Press]
DETAIL A been completely discharged, due to fears of electric shock hazards.
Right/left button 0 You can check the discharge by the following indications.
0 196
° mm N - “PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
- LED on the MFI board: The yellow LED is unlit.
(5) #2
[Verify]
(6)
Arm angle of “0°”
[Press]
Down button
A DRMS8205.ai
#1 [Disconnect]
Protective ground
wire terminal
Swivel arm lower cover
#2 DETAIL B
[Remove]
M5x60 (x3)
Serial No. B
DRMS8344.ai
[Remove]
M5 (x3) #1 [Disconnect]
Protective ground
Swivel arm
wire terminal
top cover
Swivel arm lower cover
#3
[Release]
Swivel arm
lower cover
Control panel
#4 Control panel
[Disconnect]
DRMS8207.ai
Connector
(MBL X1 W105) #4
[Disconnect]
Connector
(MBL X1 W106)
DRMS8208.ai
DETAIL A
[Remove] [Remove]
M5 (x2) M5x10 (x2),
DRMS8209.ai W5 (x2)
DRMS8210.ai
DETAIL B
DETAIL B #4
#5
[Disconnect] [Disconnect]
B Connector
Connector
(MBL X1 W107) C (MBL X1 W108)
C #3 [Remove]
#1 [Open]
#4 [Remove] Protective
Compression unit
Protective cover cover
right-hand side cover
DETAIL C
DETAIL C
#6
[Disconnect] #5
Connector [Disconnect]
(D904 X5 Connector
W142) (D904 X6 W143)
DRMS8212.ai
DRMS8211.ai
DETAIL A
[Remove]
Q5x8
DRMS8215.ai
#2
[Remove]
Q5x8
#1
[Remove]
Q5x8
DRMS8213.ai
Grid
[Remove]
Packing
material
DRMS8221.ai
FPD
DRMS8159.ai
#2
[Take out]
FPD FPD package
[Remove]
Antistatic plastic sheet
FPD
#1
[Break]
4-corner of the package
#3
[Place]
FPD
Shock absorber or
corrugated fiberboard
DRMS8041.ai
FPD
DETAIL A
Joint
Joint
#1
[Remove]
Clamp
Joint
#2 [Release]
Hose DRMS8219.ai
DETAIL A
B
A
DETAIL B
DRMS8225.ai
DRMS8222.ai
Collar
#2
[Remove]
TP3x12 (x2)
#3
[Pull out]
Carbon cover
DRMS8223.ai
<FDR-1000DRSZ> <FDR-1000DRLZ>
FAN x2 MFI51A board FAN x3
NFI3
NFI7
NFI5
MFI51A board
NFI5 NFI3
DRMS8226.ai
DETAIL A
Hold here.
Do not touch the
DETAIL A exposure plane.
B
A
DETAIL B
DRMS8227 ai
#2
[Verify]
Arm angle of “180°”
Control panel
Handle
A
DRMS8218.ai
Swivel arm
DRMS8240.ai
(10) Turn OFF the circuit breaker of the controller unit. The position of installing the grid assembly is defined with the half-punches of the
FPD as reference and defined by pressing the entire assembly against the FPD.
(11) Turn OFF the circuit breaker of the distribution switchboard. You need loosen the retaining screws of the grid assembly beforehand.
WARNINGS
l Be sure to turn OFF the circuit breaker of the distribution switchboard to
avoid electric shock hazards.
l Electrical charge (residual charge) accumulated by a capacitor might remain
A
in a circuit after the circuit breaker of the distribution switchboard is turned
OFF.
Be sure to proceed with operations after you are sure that the charge has
been completely discharged, due to fears of electric shock hazards.
You can check the discharge by the following indications.
- “PSU MAINS” LED on the power supply unit (PSU): The green LED is unlit.
- LED on the MFI board: The yellow LED is unlit.
Grid assembly
DETAIL A DETAIL B
#1 #2
[Loosen] [Loosen]
M6 M6
DRMS8083.ai
FPD
A
DETAIL A
Stand jig
DRMS8229.ai
#1
[Route]
Coolant hose
Hose (thin)
Hose (thin)
#2
#2 [Retain] Hose (thick)
Hose (thin) [Retain] Clamp
Clamp
DRMS8230.ai
DETAIL A DETAIL B
[Route]
Coolant hose
DETAIL A
Hose (thin)
Hose (thick)
DRMS8043.ai
Handle
#1 [Verify]
No gap between the frames of the swivel arm and the FPD
DRMS8232.ai
(19) Fix the right- and left-hand sides of the FPD with the bolts.
B
FPD B
A
A
DETAIL A DETAIL B FPD
DETAIL A DETAIL B
[Tighten]
[Tighten]
Q4x12
Q4x12
SW4/W4
#2 [Tighten] #2 [Tighten] SW4/W4
Q6x25 Q6x25 DRMS8234.ai
SW6/W6 SW6/W6
(Required torque value: 10 N・m) (Required torque value: 10 N・m)
DRMS8233.ai
Round hole of the sheet metal Hexagon socket head cap screw
DRMS8235.ai
[Tighten] [Tighten]
M4x40 M4x40
Grid
assembly
A
B
#1 [Align] Half-punch of the FPD #1 [Align] Half-punch of the FPD
[Remove]
TP3x6 (x3)
DRMS8239 ai
DETAIL A
[Connect]
Cable
connectors
#1 #3 #4
(SND5) (SND9) (CN10)
#2 #5
(TRC7) (TRC1)
DRMS8238 ai
A A
FPD FPD
DETAIL A DETAIL A
Bracket (small) Hold the copper foil.
[Tighten] Hold the copper foil.
TP3x6 (x5)
Bracket (large)
Cable
Cable (SND5)
(TPC1, CN10, SND9)
Cable
(TRC7)
[Tighten]
TP3x6 (x6)
DRMS8053.ai DRMS8054.ai
#2
[Verify]
DETAIL A #1 [Connect] Connectors #1
Arm angle of “0°”
[Press]
<FDR-1000DRSZ> <FDR-1000DRLZ>
FAN x2 MFI51A board FAN x3 Right/left button
Control panel
NFI3
NFI7
NFI3
NFI5
NFI5
A DRMS8241.ai
MFI51A board
Do not connect the connector of the
FDR-1000DRSZ to the NFI7
(center connector).
DRMS8046.ai
FPD exposure
plane
DRMS8048 ai
DETAIL A
#2
[Change setting]
[Loosen] After changing
When shipped from
M4x6 (x7) #1 the factory the setting
[Remove]
M4x6 (x2)
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8
Inner cover
ON
DRMS8245.ai
ON
The DIP switches 6 All set to OFF
and 7 are set to ON.
DRMS8246.ai
Inner cover
Inner cover
Swivel arm right-hand front cover
[Tighten]
Q5x8
#1
DRMS8247.ai
[Route]
Cable connector
(MBL X1 W107)
DRMS8261.ai
DETAIL A
A
DETAIL A
#4
[Connect]
Protective
ground wire
B terminal
#1 DETAIL B
[Tighten] #2
Q5x8 C [Connect]
Connector
(MBL X1 W108)
#5 #3 [Put]
[Put] Protective cover
Compression unit
right-hand side cover
DETAIL C
#2 #1
[Tighten] [Connect]
Q5x8 Connector
(D904 X6 W143)
DRMS8262.ai
DRMS8264.ai
DETAIL B
B
B
#2
[Connect]
Connector
C (MBL X1 W107)
#3 [Put]
Protective cover
DETAIL C
A
#1
[Connect]
Connector
(D904 X5 W142)
DETAIL A
GOOD NO GOOD
DRMS8263.ai
Frame
Grid wire
DRMS8356.ai
[Check]
167±1 mm Distance
DRMS8349.ai
[Click]
“Out”
DRMS8348.ai
FPD end
surface
[Click] [Check]
“In” Distance
2±1 mm
DRMS8351.ai
DRMS8350.ai
Exposure plane
of the FPD
A #2 [Connect]
Carbon cover
cable connector
Collar
#1 [Insert]
#4 [Tighten] #3 [Tighten]
Carbon cover
TP3x12 (x2) TP3x6 (x10) DRMS8242.ai
Carbon cover A
(2) Put the filter and the filter cover on the carbon cover.
DETAIL A
Carbon cover
GOOD NO GOOD
Carbon cover mark Carbon cover mark
Filter cover/filter
[Tighten]
Bind screw
(M3x6)
DRMS8243.ai
[Click]
“Start”
DRMS8353.ai
Stand jig
DRMS8244.ai
#1
MFI51A board FPD [Connect]
Relay cable connector
DRMS8248 ai
Front Rear
#1
[Disconnect]
Closure hose
A370Y100306
DRMS8426.ai
DRMS8427.ai
REFERENCE
Connect the FPD coolant hoses (thick and thin) to the temperature adjustment
unit and water supply jig in the following procedures.
#1
[Insert]
Coolant hose
(thick)
FPD
A
DETAIL A
Coolant hose
(thin)
Closure hose
DRMS8251.ai
DRMS8289.ai
- Connect the coolant hose (thin) to the temperature adjustment unit (the #2
connection for routine operation) upon completion of filling the coolant in the [Remove]
procedures below. Stand jig
(8) Disconnect the connector of the FPD MFI51A board, and the DRMS8253 ai
FPD
Temperature
adjustment unit cable
DRMS8252.ai
It is recommended that you insert a screwdriver into a hole beside the screw for
retaining the temperature adjustment unit before installing the unit. Inserting the NFI51A board
screwdriver can prevent the temperature adjustment unit from dropping. Remove
the screwdriver after completion of the operation. FPD
DETAIL A
A A
Screw
Temperature
adjustment unit
cable
Screw
[Insert] DETAIL A
Temperature adjustment unit Screwdriver
DRMS8254.ai
[Tighten]
TP3x12 (x3)
<standard accessory> [Connect]
Connector
NFI1
FPD
Temperature
adjustment unit Temperature adjustment unit
<standard accessory>
DRMS8256.ai
DRMS8255.ai
u INSTRUCTION u
Be sure to leave no gap between the FPD right- and left-hand covers (right- and left-
hand armrests) and the FPD carbon cover when putting on the FPD right- and left-
hand covers (right- and left-hand armrests). Otherwise, dust might enter the FPD,
resulting in abnormal images.
#4 [Tigten]
Cover (right)
Carbon cover Q4x12/W4
<standard
<standard
accessory>
accessory>
B
DETAIL A #3 [Tigten] DETAIL B #1 [Bring close]
BR4x25 FPD right-hand cover
FP
FPD
<standard <standard accessory>
accessory>
Carbon cover
#2 [Tigten]
FPD right-hand cover TP4x6 (x2)
No gap <standard accessory>
DRMS8268.ai
DRMS8267.ai
[Tighten]
Bind screw
M3x6 (x3)
<standard accessory>
DRMS8270.ai
DETAIL A
[Tighten]
M5 (x2)
DRMS8266.ai
[Tighten]
M5x10 (x2)
W5 (x2)
DRMS8265.ai
Covers
Swivel arm
u INSTRUCTION u lower cover
When putting on the swivel arm lower cover, exercise care not to press the pin on
the swivel arm lower cover against the DC cable connected with the collimator. If
the cover is fixed while the DC cable is pressed against the pin, the connector of the
DC cable might be disconnected. If the machine is booted up while the DC cable
connector is disconnected or incorrectly connected, a failure in DC cable connection
might be recognized, and no home positioning of the collimator takes place, resulting Control panel
in no reaction. #1 Control panel
[Connect]
Connector #1
(MBL X1 W105) [Connect]
Connector
(MBL X1 W106)
Pin #2
[Tighten] DETAIL A
M5x60 (x3) A
Collimator
B M5x60 (x3)
#3 [Connect]
Protective ground
wire terminal
Swivel arm lower cover
Swivel arm lower cover DRMS8271.ai
DRMS8273.ai
Elevation
stand
[Put]
DRMS8272.ai
Base plate cover
<standard accessory>
DRMS6274.ai
Elevation stand
Elevation stand
rear cover
<standard accessory>
B
DETAIL A DETAIL B
[Disconnect] [Disconnect]
Emergency stop Emergency stop
short-circuit cable short-circuit cable
DRMS8275.ai
DRMS8276 ai
#2
[Put]
Left-hand side cover
<standard accessory>
#1
[Connect]
Connector
(Emergency stop switch)
Elevation stand
#1
[Connect]
Connector
(Emergency stop switch)
#2
[Put]
Right-hand side cover
<standard accessory>
Elevation stand
DRMS8278.ai
DRMS8277.ai
[Tighten]
Q5x8 (x4)
Top cover
<standard accessory>
Elevation stand
DRMS8279.ai
END
[Click]
“INSTALL”
[Click]
“Next”
Installation location
selection window
is selected. [Click]
“SET”
DRMS8123 ai
Inspection window
(5) #1
[Click] #1
“Install” [Select]
“System OFF”
Installation completion
window
#2
(wizard window)
[Click]
“OK”
Click while pressing the Shift key
on the keyboard. Keep pressing
(5) #2 the Shift key until the indication
[Click] of termination in process disappears.
“Finish” DRMS8111.ai
#1
[Verify]
Default value of
NetConfig window “192.168.0.101”
DRMS8112.ai
Ping check window
(3) Open the NetConfig window. C:¥WINDOWS¥system32¥cmd.exe
D:¥INITIAL¥NET_CONFIG>
#1
[Pull down] #3
“Menu” [Verify]
Result of
.\ INITIAL \ NET CONFIG \ NET CONFIG TOOL.exe
"Lost=0 (0 % loss)"
DRMS8114.ai
#3 REFERENCE:
[Click] Clicking “SHOW” in the NetConfig
“EXECUTE” window, you can view the current
#2 settings.
[Select]
“.\ INITIAL \ NET_CONFIG \ NET CONFIG TOOL.exe” DRMS8113.ai
#1
[Input] 192 168 0 101
“192.168.0.101” (arbitrary address) #1
255 255 255 0
#2 [Input] SubNet Mask
[Click] “255.255.255.0” (arbitrary address)/
“Mnt Mode” #2
[Click]
“SET”
192 168 0 2
DRMS8117.ai
DRMS8115.ai
(8) Set the Secure IP Address, Secure SubNet Mask and Default Gateway.
(6) Set the IP address of the RU.
REFERENCE
NetConfig window (input example)
The default values (factory settings) of the Secure IP Address, Secure SubNet
192 168 0 101 Mask and Default Gateway are "0.0.0.0", respectively.
#2 #1
255 255 255 0
[Click] [Input]
“SET” 0 0 0 0 Secure IP Address,
192 168 0 2
Secure SubNet Mask and
0 0 0 0
Default Gateway
DRMS8116.ai
0 0 0 0 “0.0.0.0” (arbitrary address)/
#2
192 168 0 2
[Click]
“SET”
DRMS8118.ai
NetConfig window (input example) If the NetConfig window is closed ( is clicked) with erroneous settings and the
RU is rebooted, continuity between the RU and the AWS (ping) is lost, and the
192 168 0 101
machine cannot start.
As an emergency measure in this case, set the DIP switch settings on the
MPI51A board mounted on the RU controller unit to reboot the RU. Rebooting
192 168 0 101 the RU can restore the IP address to the default value (factory setting) and
tentatively start the machine. After the machine is tentatively started, correct the
255 255 255 0 IP address as desired in the NetConfig window, restore the DIP switch setting to
the default values (factory settings), and reboot the RU.
0 0 0 0
Never carry out any operations other than setting the IP address while the
#1 machine is tentatively started.
0 0 0 0
[Input] FTP Server IP Address
0 0 0 0
IP address of the AWS [Change settings]
“192.168.0.2” (arbitrary address)/ DIP switches
192 168 0 2 #2 (Change the DIP switch settings only when the NetConfig setting fails and the
[Click] machine cannot start. The DIP switches 2 to 4 have no function assigned.)
“SET” DRMS8119.ai
DIP switch (Si) DIP switch (Si)
(10) Exit from the NetConfig window (click ), and reboot the RU. OFF OFF
Rebooting the RU, the values setting in the NetConfig window are effective. ON ON
4
3
2
1
4
3
2
1
NetConfig window Default settings Settings for restoring the machine
192 168 0 101 (factory settings)
[Click] MPI51A board
0 0 0 0
0 0 0 0
0 0 0 0
192 168 0 2
Controller unit
DRMS8120.ai
DRMS8121 ai
u NOTE u
Use the application software CD attaching to the product when the software is
installed anew or the FPD is replaced.
DRMS8130 ai
#2
[Input]
RU PC-TOOL window “RU NAME” and “RU IP ADDR”
CONNECTION TEST
RU NAME ru0
RU IP ADDR 192 168 0 101
DRMS8127 ai
DRMS8126.ai
#1
[Click]
“PING” Install disk verification window
C:¥W NDOWS¥system32¥cmd.exe
D:¥ NITIAL¥NET_CONFIG>
#2 [Click]
[Verify] “OK”
“Lost=0 (0% loss)”
DRMS8128 ai
RU INSTALL window
MARS V1.5-0201 172.16.1.10
V1.5-0300
DRMS8131.ai
#1 [Click]
“EDIT CONFIGURATION”
DRMS8133 ai
RU Configuration Setting
EQU PMENT ID (A-Z) A
DRMS8134.ai
(10) Click [OK], and exit from the RU INSTALL window. #2 [Click]
“SET”
RU INSTALL window SET
DRMS8138.ai
MARS V1.5-0201 172.16.1.10
[Click]
V1.5-0300
“OK”
DRMS8135.ai
REFERENCE
It takes approx. 30 minutes to write into the Compact Flash Card. It sometimes
takes longer depending on the network environments between the RU and the
AWS.
Install is complete.
CL NAME window
#2
[Press]
ENTER key
#3
[Click]
“OK”
DRMS8141.ai
DRMS8142.ai
GOOD indication
#2 Character color is "black".
[Click]
NO GOOD indication
Character color is "red".
#3
[Click]
DRMS8143.ai
u INSTRUCTION u
When NO GOOD, delete the RU information by the [DETAIL] button on the RU
PC-TOOL window, and repeat the installing operation from the procedure (1).
#1
[Select]
192 168 0 101 ru0 192.168.0.101 MARS V1.5-0201 Erroneously installed RU
#2
[Click]
“DELETE”
DRMS8144.ai
EACH RU SETTING
CONNECTION TEST
<Display example>
RU NAME IP ADDRESS RU TYPE VERSION
RU NAME F EXIST NG RU
RU IP ADDR
#1
[Click]
³ ´
RESTORE #1
[Click]
#2
CONFIGURATION [Select]
INDIVIDUAL DATA (HD->RU) #2
CD drive
INDIVIDUAL DATA [Select]
MANAGEMENT INFORMATION (HD->RU) Data
#3
DRMS8396.ai
[Double-click]
³0$&+,1(´ IROGHU
DRMS8398.ai
[Click]
“OK”
DRMS8399.ai
(1) Store the AEC file for use in Europe (NHS) in the following folder of
the AWS.
<Location of storage>
C:/Program Files/FujiFilm/FCR/FDR-1000DRSZ
<Files (22 files) to be stored>
EXPMODE.LUT, MASCAL1.LUT to MASCAL21.LUT
(2) Start up the command prompt on the AWS.
(3) Input “telnet, 192.168.0.101” in the command prompt, and press the
[Enter] key.
(4) Execute the following command for each of the files stored in the
procedure (1).
copy “file name”,"/LDISK/SYSTEM/AEC/LUT/file name”
(5) Restart the machine from the control pad.
AWS V3.0 For use in Japan * Needs to be replaced with the (3) Select a merge source file.
file “for use in Europe (NHS)”.
Select “\AEC\AecLut_Dosage.csv” in the machine-specific data CD-R supplied with
For use in Japan of the old For use in Japan of the old the machine.
version version
AWSV2.0 or older
* Needs to be replaced with the * Needs to be replaced with the
file for use in Japan. file for use in Europe (NHS).
- * The AEC table is modified in the AWS V3.0 to support biopsy exposure.
n Procedures
(1) Store the latest AEC file (AECLUT.csv) suitable for the destination in
the following folder of the AWS.
<Storage location>
C:/Program Files/FujiFilm/IIP/Param
(2) Restart the AWS.
D:¥AEC¥AecLut_Dosage.csv
C:¥Program Files¥FujiFilm¥IIP¥Param¥AECLUT.csv
[Click]
<Check window>
DRMS8407.ai
DRMS8406.ai
The Compact Flash Card of the controller unit is accessed when the “INDIVIDUAL Open: "C:\Program Files\FujiFilm\FCR\Tool\RuPcTool\RuPCTool.exe" (1)-3_#1
DATA (RU -> HD)” is to be backed up. Therefore, never turn OFF the RU power [Input or
(1)-3_#2 [Click] OK Cancel Browse...
and the circuit breaker of the distribution switchboard during the backup. If select]
turned OFF, the data on the Card gets damaged, and the RU cannot boot up as a
result. RU PC-Tool window
Configuration(C)
Prepare an external storage device (such as FD and USB) beforehand when the data RU IP ADDR
BACKUP RESTORE
DRMS8341.EPS
#2 EXECUTE
[Click]
“FTP”
BACKUP #1
[Click]
ll "SYSTEM/192.168.0.101/CONFIG"
History.cfg CONFIGURATION
hosts INDIVIDUAL DATA (RU->HD) #2
Irset.cfg MANAGEMENT INFORMATION (RU->HD) [Select]
Script is done. ERROR LOG Data
value = 0 = 0x0 TRACE LOGS
ALL LOGS
#3
#4 [Verify] DRMS8146 ai
RU PC-TOOL window
BACKUP [Click]
EXECUTE
“EXECUTE”
DRMS8147.ai
DRMS8150.ai
DRMS8149.ai
RU PC-TOOL
Selected Folder:
C:\Program Files\FujiFilm\FCR\TEMP
Drive:
C:
OK #2
Directory: [Click]
C:\ “OK”
#1 Program Files
[Select] FujiFilm
FCR CANCEL
Backup location TEMP
#3
[Click]
“OK”
DRMS8148.ai
[Click]
“System”
System window
[Click]
“Set Time”
DRMS8345.ai
DRMS8346.ai
#3 #4
[Set] [Click]
Hour/Minute/Second “Set”
DRMS8347.ai
Mammographic stand
[Verify]
“STANDBY” DRMS8379.ai
Control pad
#2
[Check]
LED (yellow)
AWS screen
#3
[Check]
Radiation mark (yellow)
DRMS8380.ai
A
[Place]
Acrylic board DRMS8382.ai
Exposure
stand
DETAIL A
Exposure
stand
Chest wall
plane Acrylic board
DRMS8381.ai
<When the dosimeter is to be fixed inside the compression plate (plastic plate
B
part)>
[Fix] DETAIL A
Dosimeter (80)
Exposure center
Dosimeter
Exposure
stand
Acrylic board
(130)
DETAIL A Exposure center
100
Exposure Chest wall plane “AEC sensor area”
110 Do not allow the
stand
Compression plate dosimeter to locate
Acrylic board Dosimeter within this area.
DETAIL B
100
Dosimeter
Compression
Chest wall plane plate
110 “AEC sensor area” Acrylic board
Compression plate Do not allow the
dosimeter to locate Exposure stand
(plastic plate part) Unit: mm
within this area. Unit: mm DRMS8384.ai
DRMS8383.ai
When the deviation exceeds ± 0.05, the AEC may possibly have failures.
Replace the FPD immediately.
Output
(1) Mount the compression plate of 18x24 size (24x30 compression plate
for the FDR-1000DRLZ).
(2) Place the 40 mm acrylic board on the exposure stand.
Place the acrylic board to protrude from the chest wall plane of the exposure stand. [Compress]
Acrylic board
A
[Place]
Acrylic board
DRMS8382.ai
Exposure
stand
DETAIL A
Exposure
stand
Chest wall
plane Acrylic board
DRMS8381.ai
<When the dosimeter is to be fixed inside the compression plate (plastic plate
B
part)>
[Fix] DETAIL A
Dosimeter (80)
Exposure center
Dosimeter
Exposure
stand
Acrylic board
(130)
DETAIL A Exposure center
100
Exposure Chest wall plane “AEC sensor area”
110 Do not allow the
stand
Compression plate dosimeter to locate
Acrylic board Dosimeter within this area.
DETAIL B
100
Dosimeter
Compression
Chest wall plane plate
110 “AEC sensor area” Acrylic board
Compression plate Do not allow the
dosimeter to locate Exposure stand
(plastic plate part) Unit: mm
within this area. Unit: mm DRMS8384.ai
DRMS8383.ai
If the coefficient of variation exceeds 0.05, carry out the generator adjustment. If
the coefficient of variation exceeds 0.05 after the generator adjustment, replace
the single tank unit.
{MC:9._Generator Adjustment}
(11) Change “Target Filter” of the exposure conditions to W/Rh, and take
the procedures (7) to (10).
(12) Remove the dosimeter and the acrylic board.
(3) Calculate the variance (s2). (3) Calculate the variance (s2).
A value of the sum of squares S divided by (n-1) is called variance, and <Formula: C>
expressed as s2 (square of lower case s). S =44
<Formula: C> s2=
10-1
S
s2= (4) Calculate the standard deviation (s).
n-1
<Formula: D>
(4) Calculate the standard deviation (s). s= 44 =6.6
The root of the variance is called standard deviation and expressed as s (lower (5) Calculate the coefficient of variation (CV).
case s).
<Formula: E>
<Formula: D>
s= s2 6.6
CV= =0.013
521.8
(5) Calculate the coefficient of variation (CV). _
The standard deviation (s) divided by the average value (X) is called coefficient of
variation (CV). Coefficient of variation (CV) = 0.013 < 0.05
<Formula: E> The result of the example is GOOD.
S
CV= -
X
DRMS8328.ai
u NOTE u
Check the elevation of the swivel arm for each of six buttons individually.
- Swivel arm control panel (right/left)
- Compression unit control panel (right/left)
- Foot switch (right/left)
Elevation stand
Compression unit
control panel
Swivel arm
control panel Foot switch
Elevation button
DRMS8357.ai
Compression unit (4) Keep pressing the right button from the 0° position of the swivel arm
control panel and check that the arm automatically stops at the 90° position.
Swivel arm Check that the angle indication on the display shows “90°”.
control panel (5) Keep pressing the right button from the 90° position of the swivel arm
and check that the arm automatically stops at the 180° position.
Check that the angle indication on the display shows “180°”.
(6) Keep pressing the left button from the 180° position of the swivel arm
and check that the arm automatically stops at the 0° position.
Check that the angle indication on the display shows “0°”.
REFERENCE
When the swivel arm is to be returned from the -180º (or 180º) position to the 0º
position, it does not stop at the -90º (or 90º) position.
(7) Press the left button, and release the button while the swivel arm is
rotating.
Display Check that the swivel arm stops rotating immediately after you release the button.
(8) Press the right button, and release the button while the swivel arm is
rotating.
DRMS8358.ai Check that the swivel arm stops rotating immediately after you release the button.
Control pad
DRMS8359.ai
Display
Decompression Compression
button button
DRMS8360.ai
- Check to make sure that no abnormal noise occurs during the operation and that the
compression plate smoothly works.
(1) Press the compression button of the foot switch and check that the
compression plate moves in the compressing direction.
Check that the compression plate continuously moves while the button is pressed.
(2) Press the decompression button of the foot switch and check that the
compression plate moves in the decompressing direction.
[Mount]
Check that the compression plate continuously moves while the button is pressed. Compression
(3) Press the compression button of the foot switch and release the gauge
button halfway.
Check that the compression plate stops moving immediately after you release the
button. Exposure
stand
(4) Press the decompression button of the foot switch and release the
button halfway. DRMS8361.ai
Check that the compression plate stops moving immediately after you release the (2) Turn the compression/decompression dial clockwise to compress the
button. compression gauge with an arbitrary pressure.
Compression
plate
[Compress]
Compression
gauge
DRMS8362.ai
#2
[Check]
Compression
plate
#2
[Check]
DRMS8363.ai
Compression
(4) Dismount the compression gauge from the exposure stand. plate
DRMS8364ai
[Mount]
Compression
gauge
Exposure
stand
DRMS8361.ai
[Click]
“Compression
parameters setting”
DRMS8365.ai
Display
#2
[Compress]
Compression gauge
DRMS8368.ai
DRMS8367.ai
- AEC: MANUAL
- kV: 28
[Compress]
- mAs: 2.0
Compression
- Auto Decomp: ON
gauge
(3) Press the exposure button of the control pad.
(4) Check that the compression is canceled (automatic decompression).
DRMS8362.ai
DRMS8369.ai
0 40 100 Display
° mm N
[Compress]
Acrylic board DRMS8371.ai
DRMS8370.ai
#2 [Check]
Compression plate Compression
pressure of
#1 100 N ± 10 N
[Compress] (10.2 kgf ± 1 kgf)
Compression gauge
DRMS8374.ai
0 109 100
DRMS8372.ai
0 112 50 Display
DRMS8375.ai
° mm N
DRMS8373.ai
[Check] 200 N
(Display example) Compression pressure
(3) Compress the compression gauge with the maximum compression
pressure (200 N).
0 105 150 Display (4) Keep the compression gauge compressed for approx. 1 minute.
° mm N Check that the compression pressure value indicated on the display does not vary
and that the gauge is kept compressed.
(5) Cancel the compression.
DRMS8377.ai
(6) Restore the maximum compression pressure to the original value by
means of the compression parameters setting box of the AWS.
(9) Dismount the compression gauge from the exposure stand.
{Operation Manual}
The compression gauge may be kept mounted when the subsequent procedures are
to be taken.
Light irradiation
field
Exposure stand
Exposure frame
DRMS8393.ai
...........
(1) Press the AWS's [Windows] key to open the Start menu of Windows. ...........
...........
Press ...........
...........
DRMS8385.ai
Run... #1 [Click]
.... Run ?
EACH RU SETTING
CONNECTION TEST
RU NAME LIST OF EXISTING RU
RU IP ADDR
BACKUP RESTORE
DRMS8386.EPS
EDIT CL NAME
BACKUP RESTORE
ERROR LOG
EXECUTE EXECUTE
#3
ERROR-DB [Scroll]
window <Display example>
DRMS8388.ai
#4
[Click]
DRMS8387.ai
DRMS8389.ai
DRMS8390.ai
[Click]
“Clear Log”
DRMS8392.ai
#2
[Click]
“Options”
Opsions window
DRMS8391.ai
REFERENCE
Setting and measurement examples for reference:
{IN:Appendix 3._Installation Environment Measurement (Power Supply
Voltage) [Only Inside Japan]}
- X-ray shield (lead apron with lead equivalent of 0.25 to 0.35 mm Pb, or 60 mm Circuit breaker
acrylic board) Protective ground OFF
DRMS8335 ai
DRMS8336.ai
(4) Press the system start button on the control pad.
(5) Turn ON the power of the AWS.
#2 [Verify] #1
#1 “STANDBY” [Click]
[Click] “Start inspection”
“Next” DRMS8338.ai
Set inspection item window (8) Set the parameters in the inspection window.
The settings are:
- AEC: MANUAL
- Target/Filter: W/Rh
- Tube voltage: 35 kV
- Tube current time product: 500 mAs
Set inspection item window
#2
[Click]
“Left breast CC”
“Right breast CC”
Circuit breaker
Protective ground OFF
ON Circuit breaker
terminal (for N)
Protective ground OFF
Circuit breaker
terminal terminal (for L)
DRMS8335 ai
Set the measurement conditions in the memory hicoder. (4) Input the source drive and the execution program name (A:\WV118J.
Measurement EXE).
(The drive A is selected in this documentation.)
Acquire the data from the memory hicoder to When the program is downloaded to the computer, specify SETUP.EXE in the
the PC. destination of storage. Upon input, click “OK”.
Transfer of the data to the PC
(7) Specify the drive and the folder of the installation destination.
The default setting is “C:\Program Files\HIOKI\Wv”. If the default is OK, select
“Next”.
Press the [Set trigger] and [Set] buttons to display the status (3/3).
Trigger (3/3) Memory
Trigger Continuous Pull trigger 0%
Trigger source OR Manual trigger ON
CH1
RMS Level
CH2
RMS OFF
CH3
RMS OFF If numerous files are created in the card, the CH2 setting remains effective. Set
CH4 CH2 of the trigger (3/3) to “OFF”.
RMS OFF Press the [Set function] and [Set] buttons, to display the status (*/3).
Effective value recorder function
Press the [Basic settings] and [Set] buttons, to display the status (1/3).
Status (1/3)
Basic settings
Time axis range: 100 ms
(data interval) (1.25 msec)
Recording length: 20 DIV
(2.00 s)
Frequency: 50 Hz/60Hz (select depending on the local
power supply frequency)
Window split: One window
Automatic storage: Binary
File name [AUTO]
Skip OFF
Printer recording format: Waveform
Print OFF
Instantaneous waveform record OFF
(memory record)
Instantaneous power failure was observed on the xxx power supply xxxV line on
Remarks
xx(day)/xx(month)/xxxx(year).
Measurement complete
(7) Specify the drive and the folder of the installation destination.
The default setting is “C:\Program Files\SIE\TH-101 manager”. If the default is
OK, select “Next”.
(4) Input the source drive and the execution program name (A:\SETUP.
EXE). (The drive A is selected in this documentation.) When the
program is downloaded to the computer, specify SETUP.EXE in the
destination of storage. Upon input, click “OK”.
* The number of recorded days depends on the service life of a battery, which is
approx. three months, although it is influenced by the state of the battery.
* The internal setting is kept if the battery is replaced within one minute.
(5) Select [Time]. If the same date and time as the personal computer
time is to be set, make a check on [Use PC time], and select [Set]
button.
1. POWER switch
5. COM connector
1. POWER switch 2. RECORD switch
5. COM connector 3. INTERVAL switch
2. RECORD switch 4. MODE switch
3. INTERVAL switch
Request of battery replacement
4. MODE switch
In communication Farhrenheit
Press keeping the RECORD switch for 2 or more seconds. The measurement
starts when “REC” appears on the liquid crystal display. Celcius
In communication Farhrenheit
u NOTE u
The following window appears while the data logger is measuring the temperature
and humidity. Terminate the measurement, referring to “1.5.2 Terminating the
Measurement from the Personal Computer”.
n Charting
The data read into the personal computer is charted to evaluate the installation
environments. Make a graph with unified scales of the vertical and transverse axes
as follows, in charting the data.
Graph scale
X axis: 1 day
Y axis: Temperature with the lower limit of 17°C and upper limit of 27°C
Humidity with the lower limit of 0% and upper limit of 100%
Refer to “8.1.3 How to Change the Graph Scale” for setting the scales. The data
(6) Select “File” and “Save as” in the main menu. Name the file as which is measured in the same temperature unit (Celsius) and the same recording
desired, and store the data. interval can be combined into a graph.
(2) When the setting window shown below appears, make a check in “1
day” for the X axis and “Manual” for the Y axis. Click “Details”.
(3) When the detail setting window for the Y axis appears, set the upper
and lower limits of the humidity as follows, and click “OK”. The graph
appears with the changed scales.
(3) Upon completion of the specification, select [OK]. The specified data
files are displayed in one graph.
Result report of temperature and humidity measurement Appendix 4.8 Measurement Conditions for
(example) Temperature and Humidity
Installation Model of installed Described below are the measurement conditions for temperature and humidity for the
XXX hospital FDR MS-1000
site machine FDR MS-1000. Make measurements according to the measurement conditions.
Measurement Measurement
period
06/10/2008 to 18/10/2008
room
Examination room Appendix 4.8.1 Measurement Conditions for the FDR MS-1000
Measurement - Measurement period:
Work description Remarks Weather In a week (normal operating days) after the installation of the machine.
date
- Measurement site:
06/10 (Mon) Transfer The door is kept open. Cloudy
Machine room
07/10 (Tue) Installation of the machine Air-conditioned at 24°C Sunny The machine room wall at 1 m above the floor near the cabinet.
08/10 (Wed) Assembly Without the door open Sunny The site should be free from direct contact with the wind from the air-conditioner.
The data logger for remote maintenance after the machine is put into practice should
Air-conditioner changed
09/10 (Thu) Assembly
to 23°C
Sunny be installed at the same position.
- Recording interval:
(1) 30 seconds since the installation starts until the clinical practice starts.
Be sure to read the data into the personal computer and analyze data once
a day, and change the operation conditions of the air-conditioner as the need
arises.
(2) 2 minutes for a week after the clinical practice starts.
As the memory is full of data in 11 days, read the data into the personal
computer a week later.
CAUTIONS
- When unloading the machine, two or more persons should always cooperate
to do so.
- Wear safety shoes for operation.
- Before unloading the machine, secure a proper machine transfer route.
- The mass of the mammographic stand is approx. 320 kg. When the crate
packing the mammographic stand is to be unloaded from the platform of the
truck, confirm the gravity center for balance.
Mammographic stand
(1) Move the crate in which the mammographic stand is packed to a place u INSTRUCTIONS u
where it can be disassembled. - Fully leave the machine with the plastic cover put on in the room. If the cover is
put off immediately after the machine is transferred into the room, the machine
(2) Disassemble the crate.
might get condensation.
#2
- Do not put off the protective sheet of the collimator to prevent entry of dust and
[Remove]
dirt.
Crate (right- and left-hand side faces,
Do not put off the protective
upper face and rear)
sheet from the collimator.
DRMS8011.ai
#1
[Within Japan] [Outside Japan]
[Remove]
Crate (front) DRMS8002.ai
#1 #1
[Remove] [Remove]
Pedestal Pedestal cover
cover #2
[Remove] #2
Plastic cover [Remove]
Protective
cover
#3
[Remove]
Plastic cover
DRMS8012.ai
#2
Do not remove this
[Remove]
packing material
BQ10x40
until instructed.
Pedestal
DRMS8096 ai
(5) Remove the bracket retaining the upper surface of the mammographic
stand.
Mammographic stand DRMS8004.ai
#2
[Remove] A
BQ10x40 (x2)
#1 [Remove] Pedestal
Upper surface bracket BQ10x40 (x2) Pedestal retaining bracket
DRMS8005 ai
DRMS8003.ai
CAUTION u INSTRUCTION u
Carry the handles on the packing box when lifting the box packing the FPD.
The FPD need be unpacked under such temperature and humidity conditions
that the FPD performance can be kept. Therefore, do not unpack the FPD until
Dot not unpack the FPD box
instructed in the manual.
until instructed in the manual.
{IN:17.1_Unpacking the FPD}
[Remove]
Stretch film [Mount]
Accessory packing box
(under cover, cables)
Handles
[Mount]
Controller unit
packing box
DRMS8097 ai
[Mount]
Accessory packing box
(Temperature adjustment unit,
coolant bottle, pads, cover,
manual, etc.) DRMS8098.ai
(1) Move the crate packing the operation desk and the equipment rack to
a place where it can be disassembled. [Mount]
Operation desk and
(2) Disassemble the crate. equipment rack
[Remove]
Crate
DRMS8100.ai
DRMS8099 ai
- The dedicated carrier is designed for exclusive use of the mammographic stand. It
up to approx. 10 mm. can carry the machine, as well as fix the machine to the carrier and is equipped with
adjustable feet capable of temporary installation of the machine while it is on the
u NOTE u carrier.
If the bump is 10 mm or higher, a slope should be provided.
(4) Fix the mammographic stand lowered onto the dedicated carrier.
(2) Route the rope through the eye bolt and crane up the mammographic Mammographic stand
stand.
#2
Mammographic stand
[Crane]
Mammographic stand
#1 [Tighten]
[Route] Hex head bolt x4
Rope
Eye bolt
DRMS8104.ai
Dedicated carrier
DRMS8109.ai
[Remove]
M8x120 (x6)
Mammographic stand
DRMS8110 ai
(1) Load the mammographic stand on the carrier while moving it on the CAUTION
pallet.
#1 The mass of the mammographic stand is approx. 320 kg. When unloading the
[Move] mammographic stand from the carrier and standing it up, four or more persons
Mammographic stand should always cooperate to do so.
Pallet
#2 [Stand up]
Carrier [Load] DRMS8006 ai Mammographic stand
(2) Transfer the mammographic stand on the carrier to the installation
position.
DRMS8106 ai
#1
[Transfer]
Mammographic stand
DRMS8007 ai
DRMS8008 ai
Monitor Arm
(1) Open the right-hand side cover. Shaft
assembly
Operation desk
[Open]
Right-hand side cover [Tighten]
Nut
Spacer
Operation desk
DRMS8165 ai
[Install]
[Draw out] Monitor arm
Shelf
DRMS8164.ai
DRMS8166.ai
(1) Remove the hex head bolts (x6), open the lid of the crate, and take out
the X-ray protective board together with the packing material from the
transporting box. DRMS8409.ai
Bracket
Screw
Protective board
(packing material)
Rear cover
#1
B [Temporarily fix]
Q6x30 (x2)
DETAIL A
DETAIL A DETAIL B
#2
[Temporarily fix]
TP3x6 (x5)
DRMS8411.ai
#1 [Raise] #2 [Lower]
Stopper Stopper
DRMS8412 ai
#1
[Check]
Logo mark
”MLIFIJUF“
DETAIL A #2
[Insert]
X-ray
protective
board
A
DRMS8414.ai
[Tighten]
#4 BQ6x12 (x4)
[Tighten]
TP3x6 (x5)
DRMS8169.ai
DRMS8413.ai
DRMS8417 ai
DRMS8419 ai
DRMS8418.ai
#3
A DRMS8421 ai
(3) Put the positioner on the exposure table, and adjust its position so
that the retainer of the armrest is on a flat area of the aluminum shaft
DETAIL A DETAIL B lock.
#1
Cover
Flat area
Retainer
DRMS8420 ai
DRMS8422 ai
#1
Retainer
Brass block
Flat area
DRMS8423.ai DRMS8425.ai
#3
#2
#1
Lock pin
Retainer
DRMS8424 ai
Kit START
Front cover kit (24x30) FRONT COVER KIT BIOPSY 24X30 #(E) Appendix 9.5 Putting on the Swivel Arm Lower Connector
Cover
END
DRMS8428.ai
13 Screw (BR6x12) 1
DRMS8440.ai
2 Bracket 1
DRMS8429.ai 14 Screw (S3x10 (SUS)) DRMS8441.ai 8 Only for 18x24
3 Bracket 1 356N104378*
DRMS8430.ai
350Y101550*
15 Armrest 1
Only for 24x30
4 Cable 1 136Y102382*
DRMS8442 ai
DRMS8431.ai
5 Bracket 1 356N104569*
350Y101551*
DRMS8432.ai
16 Armrest 1
Only for 24x30
6 Screw (TP3x6) 7
DRMS8433.ai
DRMS8443.ai
7 Clamp 2 316S0256*
DRMS8434.ai
17 Label 1 405N102277*
DRMS8444 ai
DRMS8435 ai
9 Bracket 1 356N104614*
DRMS8436.ai
#2
#1 #1
Stereo bracket
DRMS8446 ai
#1
DRMS8448.ai
#2 #3
DRMS8449 ai
#2 u INSTRUCTION u
The connector has two D-cut portions. Make alignment so that the protrusions in
the connector are directed as shown below.
DRMS8446.ai
DRMS8450 ai
#1
DRMS8451.ai #4 #2 #3
#5 Clamp: 2-positions
DRMS8453 ai
#5
DRMS8454.ai
DRMS8452.ai
#1
#4
DRMS8455 ai
#3
DRMS8457 ai
#2
DRMS8456.ai
#2
DETAIL A #1
DRMS8459.ai
#6 (5) Gather the cables connected with X5, X14 and X15 at the center.
#5 #7
DRMS8458.ai
DRMS8460.ai
Notch DRMS8461.ai
DRMS8463.ai
#1
#2 DRMS8462 ai
#1
DRMS8466.ai
#2
#3 (12) Fix the flat cable to the location as indicated in the figure below.
#1 Bind
DRMS8464 ai
(9) Retain the stereo inner cover (CBA board) on the swivel arm.
(10) Retain the cables extending from X5, X14 and X15 to the binding sheet
metal.
#1 Bind
#1
#1
DRMS8467.ai
DRMS8465.ai
#1
A DETAIL A
Wiring diagram
#2
#2
#1
DRMS8469.ai
#2
DRMS8468.ai
DRMS8470.ai DRMS8471 ai
DRMS8472.ai
#1
#1
Stereo cover
#1
DETAIL A
#1
DRMS8473.ai
DRMS8475.ai
Rail
DRMS8474.ai
Bracket
#1
DRMS8476 ai
(21) Put on the right and left compression unit side covers.
Connect the cable from the CBA board to the connector inside the cover.
Connect the ground wire (green) inside the cover to the compression unit frame.
u NOTE u
Do not intersect the right and left cables extending from the CBA board to the
compression unit in connecting them. Put on the inner cover (CBA board) while
identifying the right and left cables. Wrong connection might result in damage of
the tube during stereo motion.
#2
Ground wire
DRMS8477.ai
#2 #1
DRMS8478.ai
#1
DRMS8479 ai
#2
#1
DRMS8481.ai
#1
Ground
wire
DRMS8482 ai
END
The target values of adjusting the X-ray irradiation field for stereo exposure are listed
in the table below (the underscored values refer to the changes).
Open the analysis Excel file supporting the stereo exposure, which corresponds to the
target value, and make adjustment of the X-ray irradiation field.
Refer to the MC manual for the adjustment procedures.
{MC:11.1_Adjusting the X-Ray Irradiation Field}
REFERENCE
The analysis Excel file supporting the stereo exposure is separately delivered.
Check to make sure that the AWS is of the version corresponding to the stereo The direction of the stereo motion depends on the angle of the swivel arm.
exposure kit (V3.1 or higher). - From 0º to 90º: Stereo motion in the + direction.
- From -2º to -90º: Stereo motion in the - direction.
- -1º: Stereo motion in either + or - direction.
Appendix 10.6 Installing the Stereo Digital (This is because the threshold value is -1.5º.)
REFERENCES
- In the “-0” menu, exposure is made with the tube set to 0° (same as normal
exposure).
In the “+4” menu, exposure is made with the tube tilted by 4° in the right.
- While the stereo exposure plug is connected, Auto Decompression is OFF
although Auto Decompression ON is set on the AWS. Compression needs to
be canceled manually after the exposure.
DRMS8485 ai
(3) Click [Calib] on the inspection window of the AWS, and click “Field
Calibration”.
(4) When the message appears on the AWS, click [OK].
(5) Make exposure when “Perform exposure for calibration” appears.
DRMS8487 ai
- If exposures of are made 0° and 4° are made under the same exposure
conditions, the image quality at 4° might be worse than that at 0°. This is
because the X-rays are incident slantly to the grid and the dose arriving at the
device decreases.
AMULET
AMULET AMULET
Correction data item AMULET +Biopsy
+Biopsy +SDM
+SDM
Erasure of permanent dot defect
O O O O
{IN:Appendix 10.9.1_Erasure of
Permanent Dot Defect}
Exposure dot defect
[Set Maintenance (Defect-Solid)]
For AMULET O O O O
{Calibration (CA)}
Exposure dot defect
[Set Biopsy (Defect-Solid)]
For biopsy
- O - O
optional item
{FDR-1000BPY Service
Manual}
Exposure dot defect DRMS8489.ai
DRMS8488.ai
(5) Click [Calib] on the inspection window of the AWS, and click “Field
Calibration”.
(6) When the message appears on the AWS, click [OK].
(7) Make exposure when “Perform exposure for calibration” appears.
FDR MS-1000
Service Manual