1. A technical service report was generated for Sri Sai Hospital in Moradabad regarding an issue with their Automated Clinical Chemistry Analyzer Model EM200 CS.
2. Upon investigation, it was determined that the SGOT reagent was deteriorated. The engineer replaced the SGOT reagent kit and performed calibration and quality control testing, finding that the new reagent resolved the issue.
3. Details of the service call such as the engineer name, instrument model and serial number, customer information, observations, actions taken, and site conditions are documented in the report.
1. A technical service report was generated for Sri Sai Hospital in Moradabad regarding an issue with their Automated Clinical Chemistry Analyzer Model EM200 CS.
2. Upon investigation, it was determined that the SGOT reagent was deteriorated. The engineer replaced the SGOT reagent kit and performed calibration and quality control testing, finding that the new reagent resolved the issue.
3. Details of the service call such as the engineer name, instrument model and serial number, customer information, observations, actions taken, and site conditions are documented in the report.
1. A technical service report was generated for Sri Sai Hospital in Moradabad regarding an issue with their Automated Clinical Chemistry Analyzer Model EM200 CS.
2. Upon investigation, it was determined that the SGOT reagent was deteriorated. The engineer replaced the SGOT reagent kit and performed calibration and quality control testing, finding that the new reagent resolved the issue.
3. Details of the service call such as the engineer name, instrument model and serial number, customer information, observations, actions taken, and site conditions are documented in the report.
ZONE: CZ II 19:38:31 CUSTOMER DETAILS INSTRUMENT DETAILS SERVICE STATUS ENGINEER NAME: MR. GAURAV MAURYA MODEL: AUTOMATED CLINICAL PRR-CMC CHEMISTRY ANALYZER MODEL EM200 CS WITH CPU, MONITOR AND PRINTER SR NO.: B200208 Sri Sai Hospital, Moradabad CALL DETAILS TYPE OF CALL DATE TIME Courtesy Call ADDRESS: COMPLAINT 08-Mar-2024 (A Unit of Medinova Health Care Pvt Ltd.), RECD. 19:38:19 Mansarovar Colony, Delhi Road Near Loco Shed Bridge RESPONSE 08-Mar-2024 19:38:31 TEL. NO.: JOB COMPLETED Closed NAME OF THE OPERATOR: Sri Sai Hospital, RESPONSE TIME 08-Mar-2024 DOWN TIME: (HH:MM) Moradabad 22:51:41 TRAVEL TIME 00:30 PROBLEM REPORTED: Result issue OBSERVATIONS: Reagent is deteriated. ACTION TAKEN: Calibrate 2 different lots SGOT and run QC along with sample found QC result is out with tech high flag. Check the SGOT Reagent reaction found reagent is deteriated. SGOT Reagent is deteriorated Need to be replaced. SGOT (AST) (No. Of Kit 1) 6 X 44 ML/6 X 12.5 ML KIT CONTENTS. R1- SGOT (AST) REAGENT/6 X 44 ML R2- SGOT (AST) REAGENT: 6 X 12.5 ML LOT S042347 MFG_04/2023 EXP 09/2024. SGOT (AST) (No. Of Kit 1) 6 X 44 ML / 6 X 12.5 KIT CONTENTS R1- SGOT (AST) REAGENT/6 X 44 ML R2- SGOT (AST) REAGENT: 6 X 12.5 ML LOT_T082306 MGF_08/2023 EXP_01/2025 SITE CONDITION: LINE NEUTRAL VOLT: 220 NEUTRAL-EARTH VOLT.: 2 LINE-EARTH VOLT.: 218 BRAND OF REAGENT USED: Erba TO BE FILLED IN BY CUSTOMER:
PREVENTIVE MAINTENANCE CARRIED
OUT SATISFACTORILY:
FAULT RECTIFIED & INSTRUMENT IS
WORKING SATISFACTORILY:
WE HEREBY APPROVE RS.__________
FOR PARTS: NO. DESCRIPTION QTY COST TOTAL COMMENTS (IF ANY):
TOTAL Rs.: Name of the Operator: Sri Sai Hospital,
Moradabad
INVOICE NO.: 0 DATE BRANCH H.O
RECEIVED ON: CHECKED BY: ENGINEER'S/APPLICATION SPECIALIST'S TIME: 08-Mar-2024 19:38:31 NAME: MR. GAURAV MAURYA NOTE: Parts replaced are chargeable except during warranty period. Consumables like printer head lamp, tubing, paper rolls etc. & breakable parts are not covered by warranty and hence are chargeable. Parts replaced due to negligence in operation will also be charged in every case. AT TRANSASIA, CUSTOMER SATISFACTION IS OUR PRIME CONCERN. IN CASE YOU HAVE ANY SUGGESTIONS PLEASE CONTACT:GENERAL MANAGER (TECHNICAL SERVICE), MUMBAI TEL. : 4030 9000