ANALYTICAL PROCESS INSTRUMENTATION

Thermal Conductivity Analyser


for binary gas mixtures

Meets European standard Directives:


- Directive 89/336/CEE
- EN 61000-6-2: 2001
- EN 61000-6-3: 2001
- EN 50270: 1999
- Directive 73/23/CEE – 93/68/CEE

- Directive 94/9/CE ATEX

A5151700

− High speed of response and excellent stability and sensitivity.

− Max accuracy also with suppressed ranges


Thanks to 4 ports version with reference flow.

− Easy maintenance thanks to modular design.

− Signal transmission from Sensing Unit up to 500 mt. over shielded wires.
− General purpose or explosion proof housing (EEx-d) for Sensing Unit.

− Microprocessor-based powerful control unit, keyboard configurable.

− Optional autocalibration routine keyboard


selectable with Zero and Span output contacts.

ADEV srl Via S.Eurosia n. 27/A – 20031 Cesano Maderno (MI) - Italy Internet: http://www.adev.it 09/07/07
Tel./ Phone: +39 - 0362/641684 - 574775 Fax. +39 – 0362/575058 e-mail: info@adev.it ds 8866 / REV 5 ATEX
ADEV srl Analytical Process Instrumentation

Sensing Unit
The 8866 Thermal Conductivity Analyzer Sensor Assembly can be conduct heat away from their respective thermistors to the cell wall,
supplied in a general purpose or explosion-proof housing. Both types which is held at a constant temperature. The reference gas chamber,
of housing are rugged cast aluminum construction which allows with inlet and outlet openings drilled into the chamber from the base,
reliable operation under adverse ambient conditions. The Cell Block may be open or sealed. All zero-ended ranges and suppressed H2
assembly is made of stainless steel construction with two identical ranges from 20 to 90% have air-filled sealed reference cell. It’s
internal cells: the measuring cell and the reference cell. A highly necessary, for suppressed ranges from 95%, using reference cell
stable thermistor is mounted in each cell. (4 ports) with a constant flow of 100% of the measuring gas with a
flow rate between 100cc/min. and 1 lit/min. For intermediate ranges
1 2 (50-94%) it’s alternative the utilization of the first or second solution.
1 Reference Thermistor The measuring chamber is open to the continuous sample-gas flow.
3
2 Cell block The cells in which the thermistors are mounted are deadened so the
sample gas enters only by diffusion, minimising the effect of sample
3 Reference cell (open or flow variations. In addition, the entire cell-block assembly is
sealed)
maintained at a constant optimum temperature through two heaters
4 Sample outlet and a control thermistor which are located in the cell-block assembly.
4
5 Sample inlet The Sensing Unit Electronic Assembly incorporates some of the solid
6 Measuring thermistor
state electrical circuits. These circuits include:
5 − Current Regulator: supplies constant current to the thermistor cell
7 Connections
A51519 bridge unit.
6 7 − Action Temperature Controller: Maintains the entire cell block at a
constant temperature.
These matched thermistors form the active arms of a bridge circuit: − Voltage-to-Current Converter/Amplifier: Its current output is
the unbalanced current of the bridge provides the means of
transmitted to the analyzer control unit.
measuring the relative ability of the sample and reference gases to

Specification...
...Performance Pneumatic Connections:
(compression fittings supplied)
¼” or 6 mm OD tubing

Accuracy: ± 1% of span (output signal). Wiring Connections:


Repeatability: ± 0.3% of span (short term). Standard housing: 2 openings for G 3/8" (PG 13 cable grip).
Explosion proof housing: 2 openings for GK 1/2" (cable grip
Reproducibility: 24 hours: ± 1% of span. or conduit).
Response time: maximum for 2000 cc/min. flow rate: For H2:
initial: less than 1 sec.; 60%: 13 sec.; 90%: 23 sec; 99%: 40 sec.
For CO2: initial: less than 2 sec.; 63%: 24 sec.; 90%: 45 sec.; 99%: ...Physical
80 sec. Material Contacting Sample Gas: AISI 316, Viton, Glass
Drift: Zero: max. ± 2% of span per week. Span: max. ± 1% of span Weight: 7 kg in Standard housing; 8,5 Kg. EEx-d housing
per week (without self calibration). Dimensions: 150 x 150 x 290 mm (standard housing);
Ambient Temperature Influence: depends on range. 180 x 155 x 325 (explosion proof housing)
Typically less than 1% of full scale over entire temperature range. Finish: Epoxy grey textured enamel
Atm. Pressure Influence: ±0.1% of reading per hPa Protection: IP65 (heavy duty)
Sample flow rate influence: less than 0.5% of span over Positioning: must be mounted within 30° of vertical
flow range of 100 to 2000 cc/min.
Line Voltage Influence:
of voltage.
0.02% of span for each 1% change
Power Supply for DIN or OMEGA rail mounting;
Weight: 1,2 Kg.; Dimensions: 140 x 100 x 80 mm.

...Operating
Sample Requirements Explosion Proof Housing
Sample Flow rate: 100 ÷ 2000 cc/min.
Sample Pressure: 3000 Pa minimum (with filter and flow meter).
(optional)
Linearization: within 1% of span accuracy Protection Mode:
Measuring Range: see suffix C on the ordering information. II GD EEx-d IIC T6 IP65 T85°C
Ambient Requirements Meets European Standard
Relative Humidity: 90% maximum.
Temperature Range: -10 ÷ +50°C (14 ÷ 122°F); directive ATEX 94 / 9 / CE
Storage Temperature: 70°C (158°F) max.
Power Requirements: 24 ÷ 30 Vcc 40 VA directly or from Certificate Number:
power supply energized with 115 Vac ± 10 Vac or 230 Vac ± 20 Vac;
50/60 Hz, 40 VA CESI 03 ATEX 130 A5160500

ds 8866 / REV 5 ATEX -2- 09/07/07


ADEV srl Analytical Process Instrumentation

Standard M702 Control Unit


Description
Powerful microprocessor-based system (keyboard configurable) allowing selection of
range, type of alarms, set point and it can receive 3 analogical input signals (2 of them
can be used for cross sensitivity compensation or special functions. On request, a system
to perform zero and/or full scale autocalibration can be provided. Optional RS232C output
can directly drive a printer with selectable timing and baud rate. Isolated current output is
standard. The unit can be housed in a compact Noryl DIN case for panel mounting
(standard), in a IP 65 case with clear front door for wall mounting or in explosion proof
housing for mounting in classified area.

Display
It provides a continuous readout indication of the requested variable in engineering units
(ex. %), of alarms set point and alarms condition.

A5150800

Single or dual alarm


A single alarm (high or low) or dual alarms (one high and one low, two
high or two low) can be provided as option. Each alarm consists of: 1)
a keyboard configurable alarm threshold; 2) a LED, which is lit when
an alarm is detected; 3) a relay contact that can be used to actuate an
external signal or to start a shutdown process device.

Auxiliary functions
Other options with diagnostic and faulty calibration contacts are
available (contact Adev technical office).

Background compensation
If interfering gases are present in the background, it’s possible to
Standard Control Unit dimensions compensate their effects on the analysed gas measurement
A5103600 (automatically through the 2 auxiliary inputs from the relative
analysers, or manually setting up their average value into control unit).
The error can be ignored when interfering gases are present in low
percentage.

Technical Specification
Inputs Analogical output Clock
Input signals: 3 adjustable and Output: 4-20 mA isolated proportional to Clock: Gregorian
linearizable from 10 mV with accuracy 100% of range on maximum load of 500 Ω Back-up: by means of lithium battery
better than 1:10000 Total Range (over range): 3.6 - 24 mA Battery life: 1 year in case of power supply
Scanning time: 0,5 seconds Resolution: 1/3800 absence. Precision: 1 second/month
Conversion Type: double ramp Isolation: 1500 V
Resolution: 1/20000 Uploading time: 1 second
Input impedance: 100 Mohm typical
Isolation between channels: none Message printing M702 in explosion proof housing
Periodical printing: programmable in hours,
Protection mode: II 2 GD EEx d
Alarms (rating) minutes
−IIC T6 IP65 T85°C Tamb -20 ÷ +60°C
1 Contact: N.O./N.C. 1 A @ 250 Vac (on Alarm printing: automatic printing
Message printing: year, month, day, hour, for classified area ZONE 1 / ZONE 21
order, define the alarm contact condition -
soldering type -. Refer to suffix E). minutes, % (Ch1), temperature (Ch2, Ch3), For wall or panel mounting
Set: programmable on 100% of range alarm 1 status, alarm 2 status Dimensions: 240 x 240 x 300 mm
Relay status: normally triggered / not Function buttons outside the box
triggered Ranges of variables Equipped with 3 Gk 3/4” holes
Number of alarms: 2 on concentration. Read out in engineering units, according to
Refer to suffix D and note 1. the customer specification.
Threshold: to choose on order; field
adjustable. Refer to suffix D, H and note 1. Ambient requirements
Working temperature: 0 ÷ 50 °C
Serial interface Storage temperature: -10 ÷ +75°C
Humidity: 10 ÷ 90% without condense
Standard: RS 232 C
Power Supply: 220/110 Vca; 50/60 Hz; 5 VA
Check lines: Cts
Velocity: 9600, 4800, 2400, 1200. 600, 300 Physical specification
baud/sec. Dimensions: 96 x 96 x 185 mm.
Parity: even, odd, none Weight: about 0,5 Kg.
Isolation: 1500 V Mounting: on front panel A51594

ds 8866 / REV 5 ATEX -3- 09/07/07


ADEV srl Analytical Process Instrumentation

Rugged threaded cap

Stainless steel handle

Electronic unit

Stainless steel cell


block assembly

Inlet and outlet to


Standard Sensing Unit dimensions
analysis cell A5102401

Reliable solid state Circuit


Modular Construction
Rugged standard Background
Power supply compensation
purpose water and Constant current source
dust proof cast
Voltage/current
aluminum housing Cell-block assembly converter aplifier
Acquisition auto cal (opt)
Linearization alarms (opt)
Compensation display
Temperature controller Range selection diagnostica
Inlet / outlet ports for mA isolated output RS 232 C (opt)
sample gas
A51509 Sensing Unit M702 Control Unit

Ordering A B1+B2 C D1 D2 E F G H I O
8866
Suffix A - Line voltage 990 Dissociated Ammonia Suffix G - Sensing Unit
0 24 Vcc 999 On specification 1 IP65 housing for 2 ports sensing
2 230 V 50/60 Hz TRIPLE RANGE: refer to ds 8866TR 2 ATEX housing for 2 ports sensing
4 115 V 50/60 Hz Suffix D - Alarms 3 IP65 housing for 4 ports sensing
Suffix B - Stream composition D1 - Alarm threshold 4 ATEX housing for 4 ports sensing
B1 - Analyzed gas B2 - Background 0 None Suffix H - Autocalibration
1 H2 1 H2 1 1 low alarm 0 NO
2 CO2 2 CO2 2 1 high alarm 1 Autocalibration
3 Air 3 Air 3 1 high alarm + 1 low alarm 2 Autocal. + autocal. start inlet free contact
4 Elio 4 Elio 4 2 low alarms 3 Autocal. + autocal. start inlet free contact
5 Argon 5 Argon 5 2 high alarms with calibration in progress free contact
6 SO2 6 SO2 9 On specification
7 CH4 7 CH4 Suffix I - Control unit
D2—Wrong calibration and fault alarms 00 Without control unit and power supply
8 N2 8 N2 0 None
9 N2O 9 N2O 01 M702 standard
1 Wrong calibration alarm shared with 02 M702 + IP65 housing
S Su specifica S Su specifica an alarm threshold
Suffix C - Ranges 03 M702 + explosion proof housing
2 Fault alarm shared with an alarm 04 EC2001 (in alternative to M702)
001 0-1% (only H2) threshold
#002 0-2% (only H2) 05 EC2001 (in alternative to M702) +
3 Wrong calibration alarm on relay IP65 housing
#005 0-5% 4 Fault alarm on separated relay
010 0-10% 06 EC2001 (in alternative to M702) +
5 Wrong calibration alarm shared with explosion proof housing
015 0-15% # Autocalibration requested
for components different fault alarm 07 7876 (in alternative to M702)
020 0-20% 9 On specification
from H2 08 7876 (in alternative to M702) +
030 0-30% @ Four ports for H2, CO2 , N2O
040 0-40% Suffix E – Alarm contacts explosion proof housing
075 0-75% 0 No contacts 99 On specification
100 0-100% 1 Normally close when in alarm Suffix O - Options
080 80-100% 2 Normally open when in alarm 0 None
#195 95-100% @ Suffix F - RS232C output + internal clock 1 Automatic or manual compensation for 2
#198 98-100% @ 0 NO variable gases in the background
992 Double range 1 YES 9 On specification

ds 8866 / REV 5 ATEX -4- 09/07/07