Professional Documents
Culture Documents
BR40 64 Software Installation V2.0.2 V2.0.2
BR40 64 Software Installation V2.0.2 V2.0.2
40 / 64-Channel Configuration
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems (“Philips”) and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department. Use of this
document and the information contained in it is strictly reserved for current Philips personnel and Philips customers who have a current and valid license from Philips
for use by the customer’s designated in-house service employee on equipment located at the customer’s designated site. Use of this document by unauthorized persons
is strictly prohibited. Report violation of these requirements to the Philips Legal Department. This document must be returned to Philips when the user is no longer
licensed and in any event upon Philips’ first written request.
Warranty Disclaimer
PHILIPS PROVIDES THIS DOCUMENT WITHOUT WARRANTY OF ANY KIND, IMPLIED OR EXPRESSED,
INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
Limitation of Liability
PHILIPS HAS TAKEN CARE TO ENSURE THE ACCURACY OF THIS DOCUMENT. HOWEVER, PHILIPS ASSUMES
NO LIABILITY FOR ERRORS OR OMISSIONS AND RESERVES THE RIGHT TO MAKE CHANGES WITHOUT
FURTHER NOTICE TO ANY PRODUCTS HEREIN TO IMPROVE RELIABILITY, FUNCTION, OR DESIGN. PHILIPS
MAY MAKE IMPROVEMENTS OR CHANGES IN THE PRODUCT(S) OR PROGRAM(S) DESCRIBED IN THIS
DOCUMENT AT ANY TIME.
Symbol Descriptions
Electrical-grounding Instructions
The equipment must be grounded to an earth ground by a separate conductor. The neutral side of the line is not to be considered the earth ground. On equipment
provided with a line cord, the equipment must be connected to properly grounded, three-pin receptacle. Do not use a three-to-two pin adapter.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover i
Contents
Host Software Installation Instructions ..................................................................................................................... 40
Before Installing the Host Software..................................................................................................................... 40
Host OS Installation CD ...................................................................................................................................... 40
Restoring Computer Name and Domain / Workgroup ........................................................................................ 43
Drive Letter Correction........................................................................................................................................ 43
Installing the Host Application Software ................................................................................................................... 45
Setting Time and Date ........................................................................................................................................ 45
Installing the Host Patch 6457 CD ........................................................................................................................... 46
Installing the CD for LNA and Emphysema Options ................................................................................................ 47
Setting the K2D Visa ................................................................................................................................................ 48
Registering System Components............................................................................................................................. 50
Configuring the Display Settings .............................................................................................................................. 51
Acrobat Reader and CesarFTP Initialization ............................................................................................................ 54
Configuring Administrative Tools Security Features ................................................................................................ 55
Restoring System Files ............................................................................................................................................ 56
Installing the Options License Key ........................................................................................................................... 57
Installing the License Key from the CD and Updating Protocols ........................................................................ 57
Installing the License Key Manually and Updating Protocols ............................................................................. 59
New User-Defined Protocol Restore Procedure....................................................................................................... 60
Image Threshold Verification.................................................................................................................................... 62
Phantom Definition ................................................................................................................................................... 63
Update the Gantry Position ...................................................................................................................................... 65
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover ii
Contents
Installing OS (Windows XP) for Server S1 ............................................................................................................... 80
Set IP addresses for Server S1: ......................................................................................................................... 83
Windows Activation for Server S1....................................................................................................................... 84
Setting the Date and Time Properties................................................................................................................. 85
Activate RAID disk (S1 server only).................................................................................................................... 86
Before CIRS Application Software Installation ......................................................................................................... 91
Remote Desktop connection Via the CIRS folder ............................................................................................... 91
Remote Desktop Connection to the Server Setup .............................................................................................. 91
Loading Application Software for Server S3............................................................................................................. 94
Loading Software Application for Server S2............................................................................................................. 95
Updating the 3DBP Device Driver....................................................................................................................... 96
Loading Software Application for Server S1............................................................................................................. 97
Updating the Acquisition Device Driver .................................................................................................................... 98
Enabling Server S1 Services .............................................................................................................................. 98
Installing the CIRS Add-on.................................................................................................................................. 99
Installing the CIRS/Host Add-On V2.0.2.6457 CD.............................................................................................. 100
Installing the CIRS Test Raw Data ..................................................................................................................... 100
Installing the CIRS Test Raws #1 ....................................................................................................................... 101
Installing the CIRS Test Raws #2 ....................................................................................................................... 101
Installing the CIRS Test Raws #3 (for Brilliance 64 only) ................................................................................... 102
Verifying Hyper-threading Technologies............................................................................................................. 103
Disabling Hyper-threading Technologies ............................................................................................................ 104
Final Steps for Completing the Installation ......................................................................................................... 104
Appendix B: Procedure for Reburning CPM Boot and Application Flash ......................................... 128
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover iii
Contents
Appendix C: Virus Scan Instructions ................................................................................................ 130
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover iv
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions V2.0.2 Software Installation Overview
IMPORTANT Read this section first before beginning the Upgrade Procedure.
NOTE Make sure to save the images from the Host hard drive D: to avoid loss of images in case D: is
accidently formatted during the Host software upgrade.
This installation procedure installs the Brilliance 64-channel configuration software for Host and CIRS. Read and
understand this document in its entirety before proceeding.
This document describes the upgrade procedures for Brilliance 40/64-channel release V2.0.2.
For systems with the Lung Nodule Assessment and/or Lung Disease Quant - Emphysema options, an addtional CD
should be installed.
All the necessary instructions for performing the upgrade are included in the following chapters:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 1
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions V2.0.2 Software Installation Overview
Safety
Figure 1:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 2
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions V2.0.2 Software Installation Overview
CAUTION Follow all guidelines for ESD Anti-static protection before performing any service
procedure.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 3
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions V2.0.2 Software Installation Overview
Kit Contents
The V2.0.2 software installation DVD and CDs are listed according to
Table 1 for Brilliance 40 and Table 2 for Brilliance 64.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 4
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions V2.0.2 Software Installation Overview
For systems with Lung Nodule Assessment & Lung Disease Quant - Emphysema options installed the following CDs
must be installed.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 5
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions V2.0.2 Software Installation Overview
Prerequisites
1 Verify the kit contents according to the Kit Contents Check List.
2 Internet or phone access to CT Help Desk and authorization is required for obtaining the K2D Visa Entry.
(see Setting the K2D Visa ). This step must be done; the upgrade cannot be completed without K2D Visa Entry.
3 EOD or USB memory drive should be available for backup storage.
Required Tools
• Standard FSE tool kit
Time Requirements
The time required for the upgrade is approximately as follows:
• Host software installation: 1.5 hours
• CIRS software installation: 3 hours
TOTAL TIME: Approximately 4.5 hours
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 6
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions V2.0.2 Software Installation Overview
Figure 2:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 7
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions V2.0.2 Software Installation Overview
This document includes the procedures for the Host and CIRS software installation for a new or formatted computer.
These procedures should be used in the following cases:
• If some problem exists in the Host or CIRS software
• When replacing Host or CIRS hard disk
• When replacing Host or CIRS or any server
Based on these possibilities some different scenarios for the software installation are possible:
• Host software installation after replacing/formatting hard disk.
• CIRS all servers software installation (after replacing/formatting hard disk or application only)
• Any of the individual servers software installation (after replacing/formatting hard disk or application only)
The document contains the procedures for the complete Host and CIRS (with three servers) software installation.
Appropriate chapters of the document may be skipped according to the scenario.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 8
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
Release Notes
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 9
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
Service Utilities
• Automatic A-Plane calibration
• Various fixes in calibrations
CAUTION
• During Bolus tracking, pay attention to the CT values on images.
To start the clinical scan manually, click “Start Clinical scan”.
• When using SAS injector – start the injection only after the message “Start Injection”
appears on the scanner screen.
• Perform daily and monthly scanner checks to ensure correct CT numerical values.
• Keep the Mylar clean at all times. Any contrast material left on the Mylar surface, Patient
Table, and accessories will create image artifacts. (See instructions in the section
Cleaning the System).
• When using Automatic Dose-Right ACS, always review the mAs value before scanning.
• When using a Decubitus position, review Surview scan angle. When using a Decubitus
position use 90° for AP and PA and 180° for Lateral.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 10
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
WARNING
• While connecting the patient to different Pulmonary detection devices, ensure that the
cables do endanger CT personnel working in the Gantry room.
• While working in the scanner room, be aware of cables that may be lying on the floor in
the Gantry room.
• Pulmonary bellows may cause allergic reaction; do not place them directly on the
patient’s body.
• A doctor must sign reports used for Interventional operations.
• The doctor must avoid infecting the patient; do not touch the panel or any other surface
that is not sterile.
• Use disposable sterile gloves.
• Ask the assistant to operate the panel.
• Perform operations on the console only according to the biopsy operator’s instructions.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 11
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
Scanning
• When scanning without Surview – Don’t use * as start position; enter a number instead.
• If the Start Study button does not respond, wait 30 seconds. If it still does not respond, try clicking the Home and
then click Start Study again.
• Brilliance 40 & 64: When using collimations 40x and 64x, use rotation time 1sec or less (don’t use rotation time 1.5
and 2).
Bolus Tracking
Don’t move the patient table after clicking Go for the Locator scan.
Interventional CCT
• Lasers are not turned off in “end study”. Make sure to turn the Lasers off manually.
• In order to view previous CCT images, click on the series list.
• Start pressing the pedal only after the “Press Pedal to Scan” message is displayed on the screen.
• When using CCT Single mode, move the bed using the Gantry panel or the Scan Control Box only.
NOTE If the table height & tilt are changed during the scan, image parameters will not be updated.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 12
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
Cardiac
• Do not use the Continue Current Series option in Cardiac scans.
• Do not use the ECG button on the Tool Box.
• Use an Adaptive filter in all cardiac scans as follows:
— Large protocol: Adaptive filter + CB filter
— Routine CCTA protocol: Adaptive filter + CC filter
— Stent protocol: Adaptive filer + CD filter
TLCT
The actual cycle time may differ from the planned cycle time by up to 0.5 seconds.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 13
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
Pulmonary Gating
• Do not use the combination of Pulmo scans and Bolus tracking or timed scans or cardiac.
• Always verify that the Respiratory wave is present before scanning. For Bellows, make sure that the signal is stable
and the peaks and are successively marked in red and blue.
• Use only the intercom for instructing the patient to control his breath. (It is not recommended to use automatic voice-
messages).
• In Retrospective Pulmo, start offline recon only after the Pulmo study is completed and closed.
• The minimum pitch for Retrospective pulmonary scans is 0.04.
Offline Reconstruction
Sometimes it takes up to 20 seconds until the scan protocol is opened after selecting “View Recon”.
General
• Restart the host console every day.
• Before using Generate Protocols or Air calibration, click Start Study and make sure the Patient Data fields are
empty.
• Make sure that an EOD media is always inserted into the EOD slot, even if it is not used for backup.
• Stereotaxis: use only Spiral protocols with SP reconstruction or Axial protocols. If you are using Axial protocols,
select only the following collimations:
— For 16 slices configuration: 4x0.75, 4x1.5 or 4x3.
— For 40 channel configuration: 16x0.625 and thickness 2.5mm, 12x1.25 and thickness 3.75mm or 40x0.625 and
thickness 6.25mm.
• “Unhandled Exception” error message– In case this window appears, click Quit and logout. Do not continue
working after an “unhandled exception” was observed.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 14
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
• Dental CT Application: only images with matrix 512x512 can be loaded to the application. In case the Dental images
have higher matrix, perform offline recon with matrix 512.
• Cursor coordinates value on Surview may not be accurate; don’t use them for measuring or positioning.
Quick Review
Unable to open RF or CR modality images
Workaround: Open the images in the Viewer application
States
State doesn't include the following features:
• User graphics (including QCTA tools)
• Sculpting
• Bone removal
• Batch/cine
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 15
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
• Loading large matrix images: the Preview application can safely open up to 5000 images at a resolution of 512x512.
When opening larger matrix images, for example 1024x1024, the Preview application can safely open only 1/4, or
1250 images.
Solution: Only open a number of images proportional to the allowed limit. For example open only 1250 images of
1024x1024 resolution.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 16
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 17
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
Common Calibration
• UHR Z-alignment: After completing the automatic UHR calibration, perform manual mechanical alignment (for
details, see calibration manual). Continue with the remaining automatic calibration procedures.
IMPORTANT: Once mechanical alignment is performed, do not repeat the automatic UHR calibration.
• To avoid calibration application crash, (eg. “Calc error”) and reduce the risk of Matlab memory leakage, reboot the
host computer before starting the entire calibration procedure.
• There are two type of eye phantoms: Broad and Narrow. In the Preferences/Scan Parameters tab, check the type of
phantom before positioning it on the phantom holder. Failure use the proper setting will result in errors and bad
calibration.
• Calibrations require zero in the couch so the laser markers fall on the phantom holder and embedded help file to the
calibration process.
• The Phantom positioning refers to system and “eye” phantom.
• You will need to Zero in the phantom only once at the beginning of the calibrations.
• During the calibration, there is automatic Couch movement.
Ensure the Couch Vertical adjustment is within spec. A couch that is not calibrated or is malfunctioning will generate
a position error during the calibrations.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 18
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
• Due to the phantom’s weight, there may be a slight tilt which will affect the Calibrations, constancy and performance
tests. While tightening the fastening screw at the back, support the phantom mechanically by placing a temporary
support underneath the phantom.
• For more convenient laser alignment for couch positioning, align the laser marker with the phantom holding bracket
before zeroing the couch (see Figure 3).
• Air calibration checks the existence of obstacles in the FOV. Should the Air calibration generate an error specifying
the presence of an object in the FOV and the FOV is vacant, Bad detector program must be run.
holding bracket
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 19
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
— When the test runs, a red WAIT message appears in the right hand corner of the test menu.
Do not use
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 20
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
— After choosing the test to be run, press the GO button in the main application menu (see Figure 5).
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 21
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
— The test will generate a Fee Bias Test Report (see sample in Figure 6).
Reason
Module # Result Thresholds for Failure
List of Failures
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 22
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
Activation of Tests
• Physics analysis, impulse response, slice width shall be done only from the Brilliance 2.0 Viewer | application |
service application. Do not perform physics analysis via the Diagnostics menu in the main screen.
• The desired series must be selected from the archive before activation (see Figure 7 and Figure 8).
Figure 7: Brilliance 2.0 Viewer - Selecting Series for Activation of Physics Analysis
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 23
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 24
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
DMS Diagnostics
• Before using the application, download the DMC diag file as instructed below.
— Each time you cycle power the gantry, reload the DMCdiag.
— To return to normal operation, cycle power to the Gantry.
— During a full test, note the CAN connection status at the bottom left corner of the application.
Green status - Valid connection. Red status - connection lost.
— If connection with the CAN is lost during the test, cycle power and reload the DMCdiag. do the following:
a Verify the system is E-stop opened and the application host is not running under the Philips service user.
b Open the controller utilities and select Service Mode.
c Browse to Utilities | File Configuration | RAM download | DMC.
d In the dialog box, choose file usr/diamont.root/gantry CPM/DMCdiag.bin and click Open.
e Perform RAM download to the DMC.
New Tools
• New Gantry tool: Universal log Viewer: Allows the user to view any log in the system with a single viewer. The tool
displays all logs in a common format. In addition, the tool allows the user to filter and analyze log data in a uniform
way. The tool allows viewing of the following system logs:
— Gantry log files
— Host (usp) log files
— Tamar log files
— CanTrace files
— TubeHistory files
— Windows Event logs
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 25
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
Refer to http://incenter.medical.philips.com/doclib/fetch/2000/4504/4396/4401/95541/
brilliance_ct_diagnostics_manual.pdf?nodeid=356890&vernum=0 for instructions of use
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 26
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 27
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
UltraVNC Overview
UltraVNC is the Look Over The Shoulder (LOTS) utility available on the Brilliance line of scanners.
The UltraVNC server is configured in a “View only” mode. Remote control is disabled. This is done to ensure that no
motion or X-Rays can be done remotely.
The technician must be instructed to operate the scanner. The remote users can watch the console display to assist
with troubleshooting. This utility cannot be used unless the technician is at the console to accept the remote
connection. A connection should not be attempted until you receive a call from the hospital technician and you inform
them that you will be remotely connecting to the scanner.
Follow HIPAA and other regulations involving patient confidentiality for the country where the scanner is installed when
using this tool.
HIPAA Information
http://www.medical.philips.com/us/company/connectivity/hipaa/
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 28
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
3 Use the RSN Launch Pad tool to execute the UltraVNC Viewer for the selected scanner.
4 The operator at the Host console will see the following prompt appear. The technician must Accept the remote
connection before the display will appear. On a Dual Monitor configuration the prompt window will appear in the
center split between the two monitors.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 29
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
5 After the remote connection has been accepted and established enter the Service password for the Host when
prompted to:
Validation Tables
Table 3: Service Tools Diagnostics
Comments
Tasks Pass / Fail
(write your comments/results here)
DIAGNOSTIC TOOLS
Common Calibration:
Auto constancy Pass
Re-run air calibration if:
• More than a day has passed between last air calib. and Auto performance run.
Auto performance Pass • If several other tests were done between last air calibration and performance.
• If room temp. is not stable and last air calib. were done under different
conditions.
Auto Reliability
GENERAL TOOLS
CPM Software Version Pass
CPM Reset Pass
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 30
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
Comments
Tasks Pass / Fail
(write your comments/results here)
Real Time Data Pass
CAN Trace Pass
CAN Send Pass
CAN Trace Analyzer Pass
Tube History Pass
Remote Service Pass
Log Book Pass Appears twice.(*)
System Integrity Fail Files Parameters are not updated for the released version
CONGIFUGRATION TOOLS
Setting Pass
Controller Utilities Pass
CALIBRATION TOOLS
Tilt Calibration Pass
Couch vertical calibration Pass
Spindle Block
Pass Appears twice.(*)
Commissioning
DIAGNOSTIC TOOLS
View Logs Pass
Universal Log viewer Pass
Analysis Tool Pass
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 31
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
Comments
Tasks Pass / Fail
(write your comments/results here)
Manual exposure Pass
No Load Test Pass
DMS TOOLS
DMCDiag RAM download needed including new tests
DMS Diagnostic Pass
DMP FPGA Version fails due to Test looking at the wrong file.
DMS temp. correct, but shows the wrong APDM; DMC Utility OK (temp.
DMC Utility Pass not working), A-Plane tester OK. A-Plane Reader password protected,
although password is blank.
DMS Diagnostics Pass
A Plane Reader Pass
Temp. Monitor Pass
ROTOR TOOLS
Coast Down Test (Balance) Pass
Rotor Profile Pass
BugReport Pass
Bugview Pass
Visual Diagnostic Pass
Log Parser Pass
Predictive Analysis Pass
Ibox Diagnostics Pass
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 32
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
Comments
Tasks Pass / Fail
(write your comments/results here)
Audio Test Pass
CIRS DIAGNOSTIC
ACQ diagnostic Pass
BP diagnostic Pass
CIRS Error Log Pass
Comments
Tasks Pass / Fail
(write your comments/results here)
Auto Reliability Pass
Auto Constancy Pass
Physics Analysis Fail Was moved to Application icon; in IRS, select Series.
FEE Detector Tests Pass Bug corrected from V2.0
Bad Detector Pass
XRT DFS Width Pass
XRT DFS Axes Pass
PHANTOM Calibration Pass
HCOR Calibration Pass
I-Box Diagnostic Pass
Visual Diagnostic Pass Bug corrected from V2.0
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 33
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 34
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Release Notes
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 35
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
Before Beginning
1 Log on as philips_service (password: service_only).
2 Double-click on the philips_service (scanner application) icon on the Desktop to launch the scanner application.
3 Perform a few scans in the different modes to ensure that the system is properly functioning.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 36
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 37
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
a On the keyboard, press the Windows button + D together to display the Desktop.
b Right-click My Computer and select Properties.
c Select the Computer Name tab. Write down the Computer Name and the Workgroup or Domain.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 38
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
11 Record the network card Link Speed & Duplex setting as follows:
a Click the Configure button and then click Yes.
b Select the Advanced tab and in the Property field, select Link Speed & Duplex; write down the value as it
appears in the Value field.
c Click Cancel to close all windows.
12Obtain the LAN setting as follows:
a Click Start | Settings | Control Panel | Internet Options. Click the Connections tab.
b Click LAN Settings and write down the proxy server settings:
Address Port
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 39
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
Host OS Installation CD
Host OS installation procedure takes approximately 45 minutes.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 40
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
2 Insert the CD labeled: Host OS V2.0.0.10 into the DVD-ROM drive and restart the system; the boot menu appears.
3 Select: Option 2) Custom Installation (see Figure 11).
NOTE Typical Installation formats drives C: and D:
Custom Installation formats drive C: only. All images on drive D: are retained.
Figure 11:
The next Windows XP Installation menu opens.
NOTE The drive D:\ will be formatted in the next step. Ensure that all images on the drive D: are backed up.
4 Select option 2 (Format Drive C only) ONLY if you did not backup the images on Drive D: (see Figure 12).
Figure 12:
Windows XP installation now copies the installation files to the hard drive. This takes approximately 12 minutes. The
system automatically reboots when complete.
5 When prompted, eject the Host OS V2.0.0.10 CD from the drive. Press CTRL+ALT+DEL to restart the computer.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 41
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
The Host computer restarts and the Windows XP installation continues automatically for about 35 more minutes.
During this process, the Host computer is automatically restarted several times.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 42
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
3 Enter the computer name as recorded earlier in the section Backing Up System Files.
4 In the Member of area, choose Domain or Workgroup. Type in the Domain or Workgroup name that you recorded
earlier in the section Backing Up System Files and click OK.
5 If you joined a Domain, do the following:
a Enter user and password supplied by the hospital administrator. Wait until you are admitted into the local domain.
b Restart the computer.
1 Double-click My Computer.
2 Check that your drives are defined as shown as follows:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 43
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 44
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
CAUTION It is critical that the time is set correctly at this point and not modified after the
application CD has been installed. Under certain conditions, modifying the date and
time after the application CD has been installed will result in the option and service keys
being damaged. The only way to correct this corruption is a full reinstallation.
Follow the steps in the order listed.
1 Double-click the clock in the System Tray (bottom right-hand corner of the Desktop); the Date / Time Properties
dialog opens. Verify that the following are correct:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 45
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
a time zone
b time and date
c In the Internet Time tab, verify that Automatically Synchronize with an Internet Time Server Marker is
unchecked.
2 Click OK.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 46
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
Table 5:
2 In the Philips Install Supervisor, click Start and wait until the message “Installation Completed” appears.
3 Verify that all Exit codes for all components are 0x00000000.
4 Click Close.
5 Eject the CD from the drive.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 47
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
Local Proprietary diagnostics reside on the Brilliance system and are currently enabled by the use of the Philips K2D
Visa Software key, which can be obtained by contacting your local SSR service key administrator. Contact the
appropriate SSR representative for details:
• CT RTAC - North America
• Erik Visser - EMEA
• Ravindra Sukthankar - ASPAC
• Bernardino Alonso - LATAM
In addition, in case of demand service on a non-contract site or when a system is down and the K2D Visa has expired,
a web page has been developed to allow the FSE access to immediately obtain a 5 day key. The URL for this web page
is http://pww.hfa.ms.philips.com/ServiceKey/
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 48
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
When development is complete the Philips Integrated Security Tool (Smart Card, Or security certificate) will replace the
current K2D Visa software key. Since Philips personnel will always be in possession of a valid smart card, any trained
Philips Field Service Engineer can perform demand service. No special provisions or processes need to be established.
The base level of AIAT (Assemble, Install, Adjust, Test) software required to meet FDA 21CFR Regulations (O-Level) is
resident on the system and does not require the use of a K2D Software key for access. O-Level service documentation
and phantoms are available for purchase by any interested party outside of PMS. Interested parties should be directed
to contact PMS local representative.
NOTE Either the K2 Visa or the K2000B Visa key can be used when obtaining the key from the web site.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 49
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
b Double-click RegisterGS.bat and press any key to continue. Click OK or Yes in any dialog boxes that appear.
c At the prompt, press any key to close the command window.
d Double-click RegisterServiceTool.bat and press any key to continue. Click OK or Yes in any dialog boxes that appear.
e At the prompt, press any key to close the command window.
f Double-click GSE.exe.
NOTE No activity is displayed in the interface.
4 Register MATLAB:
a Open the Explorer window.
b Navigate to c:\matlab\utils.
c Double-click Register.bat; registration executes.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 50
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
1 From the Start menu, click Settings | Control Panel and double-click the Display icon.
2 Set Dual Monitor configuration as follows:
a Select the Settings tab and then click Advance.
b If the left (1a) monitor is not highlighted as primary, click on the 1a monitor and then select the “Make this the
primary display” check box.
c Click the Quadro4 980 XGL (Figure 15) or FX-1400 (Figure 16) tab (depending on your computer model). A menu
opens on the left.
NOTE Choose the Horizontal Span setting in the next step; this is the only allowed option. Other options will
cause system instability.
d Click nView Display Settings. Select Horizontal span from the nView Display Mode drop-down list.
Figure 15: nView Display Settings - Quadro4 980 XGL Figure 16: nView Display Settings - FX-1400
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 51
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
e Click OK.
f In the Display Properties dialog, click the 1 icon (Monitor 1). Click Advanced.
g From the Monitor tab, set the Refresh rate to 75 Hz. Click Apply.
h In the Display Properties dialog, click the 2 icon (Monitor 2). Click Advanced.
i From the Monitor tab, set the Refresh rate to 75 Hz. Click Apply.
3 Set Exceed Configuration – nVidia Driver as follows:
a Select the Quadro4 980XGL (Figure 17) or FX-1400 (Figure 18) tab (depending on your computer model).
b Click on the Open GL Settings on the field of Custom OpenGL application settings and select Exceed.
c Verify that in the field of the default color depth the Use desktop color depth is chosen.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 52
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
Figure 19:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 53
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 54
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 55
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
NOTE If LAN Restore is selected, restart the computer at the end of the Restore procedure.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 56
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
Figure 20:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 57
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
Figure 21:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 58
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
NOTE When copying the key, use ALL CAPS for letters.
4 Select the Install option and click Update Key; the message
“Update Key will change the key of the system, are you
sure?” appears.
5 Click OK and wait until the following message appears:
“Update Key finished successfully”.
6 Select the mode as follows (see Figure 22):
— For upgrade from V1.2.3 select the Install mode
— For upgrade from V2.0 select Upgrade mode
7 Click Update Protocols; a message appears: “Update
Protocols will change the Protocols of the system, are
From V1.2.3
you sure?”
8 Click OK and wait until the following message appears: From V2.0
“Update Protocols finished successfully”.
9 Click OK and then click Exit.
10 Log off and restart the Host computer.
Figure 22:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 59
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
Figure 23:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 60
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
Figure 24:
12 In the Source.dir field click the (...) browse button and navigate to c:/usr/diamond.root/spr/.
13 Select the same directory (c:/usr/diamond.root/spr/) for the Destination.dir field.
14 Click Convert and wait until the message “Finished” appears.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 61
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
Verify Value = 0
Figure 25:
Figure 26:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 62
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
Phantom Definition
Figure 27:
NOTE Always select the BROAD option in the next step for Brilliance 64.
3 From the Eye Phantom drop-down list, select broad/narrow (see Figure 28).
NOTE Improper selection of phantom type will result in error or bad calibration.
Measure the eye phantom to determine if it is broad or narrow.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 63
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 64
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Brilliance 40/64 Host Software Installation
1 Click Preferences (see Figure 26); the Preferences window opens (see Figure 29).
2 Select Scanner Options. Select Set Scanner Position.
3 Update the Gantry position according to the physical position of the system in the scanner room and click Yes and
then click OK.
Figure 29:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 65
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
The CIRS software installation is composed of two parts: the operating system installation and the CIRS application
installation. The CIRS Software Installation procedure takes approximately six hours.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 66
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
5 Close all windows and shut down the server by pressing Esc, Esc again, and the press Enter.
6 Perform Step 2 to Step 5 for the S2 server.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 67
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
NOTE The following instructions were written for OS install with a monitor, keyboard, and mouse attached
to the server.
2 Insert CD labeled: BR40/64 CIRS V4.1.14.13 into the CD-ROM drive of the server and reboot the Server S3. Booting
from the CD starts the OS installation.
4 Answer Y to the question above to install the operating system. This will partition the disk and reboot the PC.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 68
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
6 Since Dell systems are not being used, answer N to the question above.
7 Using a screwdriver, release the screws connecting the server to the server rack.
8 Slide the server slightly out from its location to view the Product Key tag (see Figure 30).
9 When prompted, type the Windows XP product key as it appears on the tag
(including dashes and spaces) and press Enter.
10Windows XP Professional Setup starts (this takes about fifteen minutes).
11 Eject the BR40/64 CIRS V4.1.14.13 CD when prompted and press CtrL+Alt+Del
to continue with the installation of Windows XP which takes approximately 30
minutes.
CAUTION Do not re-insert the CD until instructed to do so in the Before CIRS Application
Software Installation section.
12When the Windows XP installation is complete, the server reboots and displays the login prompt.
13Logon as philips_service (scanner application) (password: service_only).
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 69
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
14When login completes, the system settings change box with the following message appears:
15Click OK.
16The system reboots; Logon as philips_service (password: service_only).
Important! After OS installation using CIRS CD, or if server came directly from Granite Microsystems, each server's
IP address is preset to 192.168.0.250 on one NIC. It is absolutely critical to login to each server (directly
or by Remote Desktop Connections) right after the OS installation of the server has been completed, and
modify each server's IP addresses for both NICs to the appropriate values. If you install the OS on all
three servers simultaneously, it will create an IP address conflict since all servers are configured with
the same IP address during OS install. In this case, you must re-configure the IP address for each of the
three servers. Also, make sure that both NICs (data and diagnostic) on each server have ethernet cables
plugged in before attempting to connect to the server remotely. If any of the S2/S3 servers report an IP
address conflict, boot to the service partition, run the System Setup Utility, and check the LAN
configuration to make sure that the correct IP address is assigned the BMC NVRAM of the server.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 70
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
10On the Network Connections Window, double-click the connection named Local Area Connection.
11 On the Local Area Connection Status page, click Properties to open the Properties page.
12On the Local Area Connection Properties page, click Internet Protocol (TCP/IP) to select it, then click Properties.
13On the TCP/IP Properties page, select Use the following IP address and enter 192.168.0.152 for Server S3
14In the Subnet Mask field, enter 255.255.255.0.
15Click OK on the TCP/IP Properties page.
16Click Close on the Local Area Connection Properties page.
17Click Close on the Local Area Connection Status page.
18Close the Network Connections Window.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 71
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
1 Click Start | All Programs | Activate Windows. The following activation screen appears (see Figure 31).
2 Select the second option (telephone activation) and follow the activation instructions on the screen.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 72
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
1 Double-click on the clock in the system tray (bottom right-hand corner of the Desktop) to open the Date / Time
Properties dialog.
2 Set the correct time zone and click Apply.
3 Type the correct time and date and click Apply.
4 In the Internet Time tab, verify that Automatically Synchronize with an Internet Time Server Marker is
unchecked.
5 Click Apply and OK.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 73
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
NOTE The following instructions were written for OS install with a monitor, keyboard, and mouse attached
to the server.
1 Insert the CD labeled: BR40/64 CIRS V4.1.14.13 into the CD-ROM drive of the server and reboot the Recon PC.
Booting from the CD will start the OS installation.
3 Answer Y to the question above to install the operating system. This will partition the disk and reboot the PC.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 74
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
5 Since Dell systems are not being used, answer N to the question above.
6 Using a screwdriver, release the screws connecting the server to the server rack.
7 Slide the server slightly out from its location to view the Product Key tag (see Figure 30).
8 When prompted, type the Windows XP product key as it appears on the tag
(including dashes and spaces) and press Enter.
9 Windows XP Professional Setup will start (this takes about fifteen minutes).
10Remove the BR40/64 CIRS V4.1.14.13 CD when prompted and press
CtrL+Alt+Del to continue with the installation of Windows XP which will take
approximately 30 minutes.
CAUTION Do not re-insert the CD until instructed to do so in the Before CIRS Application
Software Installation section.
11 When the Windows XP installation is complete, the server will reboot and display the login prompt.
12Logon as philips_service (password: service_only).
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 75
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
13When login completes, you will see the system settings change box with the following message:
14Type OK.
15The system reboots; Logon as philips_service (password: service_only).
Important! After OS installation using CIRS CD, or if server came directly from Granite Microsystems, each server's
IP address is preset to 192.168.0.250 on one NIC. It is absolutely critical to login to each server (directly
or by Remote Desktop Connections) right after the OS installation of the server has been completed, and
modify each server's IP addresses for both NICs to the appropriate values. If you install the OS on all
three servers simultaneously, it will create an IP address conflict since all servers are configured with
the same IP address during OS install. In this case, you must re-configure the IP address for each of the
three servers. Also, make sure that both NICs (data and diagnostic) on each server have ethernet cables
plugged in before attempting to connect to the server remotely. If any of the S2/S3 servers report an IP
address conflict, boot to the service partition, run the System Setup Utility, and check the LAN
configuration to make sure that the correct IP address is assigned the BMC NVRAM of the server.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 76
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
10In the Network Connections Window, double-click the connection named Local Area Connection.
11 On the Local Area Connection Status page, click Properties to open the Properties page.
12On the Local Area Connection Properties page, click Internet Protocol (TCP/IP) to select it, then click Properties.
13On the TCP/IP Properties page, select Use the following IP address and enter 192.168.0.151 for Server S2.
14In the Subnet Mask field, enter 255.255.255.0.
15Click OK on the TCP/IP Properties page.
16Click Close on the Local Area Connection Properties page.
17Click Close on the Local Area Connection Status page.
18Close the Network Connections Window.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 77
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
1 Click Start | All Programs | Activate Windows. The following activation screen appears (see Figure 33).
2 Select the second option (telephone activation) and follow the activation instructions on the screen.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 78
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
1 Double-click on the clock in the system tray (bottom right-hand corner of the Desktop) to open the Date / Time
Properties dialog.
2 Set the correct time zone and click Apply.
3 Type the correct time and date and click Apply.
4 In the Internet Time tab, verify that Automatically Synchronize with an Internet Time Server Marker is
unchecked.
5 Click Apply and OK.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 79
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
NOTE The following instructions were written for OS install with a monitor, keyboard, and mouse attached
to the server.
1 Insert the CD labeled: BR40/64 CIRS V4.1.14.13 into the CD-ROM drive of the server and reboot the Recon PC.
Booting from the CD will start the OS installation.
3 Answer Y to the question above to install the operating system. This will partition the disk and reboot the PC.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 80
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
5 Since Dell systems are not being used, Answer N to the question above.
6 Using a screwdriver, release the screws connecting the server to the server rack.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 81
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
7 Slide the server slightly out from its location to view the Product Key tag (see Figure 34).
8 When prompted, type the Windows XP product key as it appears on the tag
(including dashes and spaces) and press Enter.
9 Windows XP Professional Setup will start (this takes about fifteen minutes).
10Remove the BR40/64 CIRS V4.1.14.13 CD when prompted and press
CtrL+Alt+Del to continue with the installation of Windows XP which will take
approximately 30 minutes.
CAUTION Do not re-insert the CD until instructed to do so in the Before CIRS Application
Software Installation section.
11 When the Windows XP installation is complete, the server will reboot and display the login prompt.
12Logon as philips_service (password: service_only).
13When login completes, you will see the system settings change box with the following message:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 82
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
three servers simultaneously, it will create an IP address conflict since all servers are configured with
the same IP address during OS install. In this case, you must re-configure the IP address for each of the
three servers. Also, make sure that both NICs (data and diagnostic) on each server have ethernet cables
plugged in before attempting to connect to the server remotely. If any of the S2/S3 servers report an IP
address conflict, boot to the service partition, run the System Setup Utility, and check the LAN
configuration to make sure that the correct IP address is assigned the BMC NVRAM of the server.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 83
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
1 Click Start | All Programs | Activate Windows. The Activation screen appears (see Figure 36).
2 Select the second option (telephone activation) and follow the activation instructions on the screen.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 84
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
1 Double-click on the clock in the system tray (bottom right-hand corner of the Desktop) to open the Date / Time
Properties dialog.
2 Set the correct time zone and click Apply.
3 Type the correct time and date and click Apply.
4 In the Internet Time tab, verify that Automatically Synchronize with an Internet Time Server Marker is
unchecked.
5 Click Apply and OK.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 85
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 86
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
3 On the Computer Management menu, double-click Disk Management; the following screen appears (see
Figure 37).
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 87
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
4 Right-click on the first disk (Disk 1) and select Import Foreign Disk; the following screen appears (see Figure 38).
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 88
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 89
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
9 Disable paging and set virtual memory to zero by doing the following:
a Go to Start, right-click My Computer and select Properties.
b Click on the Advanced tab, and select Settings under Performance.
c Click the Advanced tab, and select the Change button under Virtual Memory.
d Select the No Paging File option, and click Set.
e Click Yes to any warnings.
f Click No if you are asked to restart the computer, you will do this later in the installation process.
g Click OK to exit all windows.
10Restart S1. If using Remote Desktop Connection: right-click on the Windows Taskbar and select Task Manager. On
the Task Manager Menu bar, click Shutdown | Restart. This will also work if you connected directly to the server with
a Monitor, Keyboard and Mouse.
11 Continue with the next section, Before CIRS Application Software Installation.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 90
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
• By using the CIRS folder icon on the Host. This is the default method for the remote connection.
• By using the Remote Desktop Connection application. This method could be used if the CIRS folder icon is
accidentally deleted.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 91
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
Figure 42:
NOTE If the IP address of the server is different from the address listed below, use the address defined for
LAN2 of the appropriate server instead.
— 192.168.0.200 for S1
— 192.168.0.210 for S2
— 192.168.0.220 for S3
4 For User Name, enter philips_service (scanner application).
5 For Password, enter service_only.
6 Click the Local Resources tab, and under Local Devices, click on Disk Drives to select it. Then click on Printers to
clear it.
NOTE During software loading, you will need to restart the computer. During restart, the CIRS servers lose
connection and it takes about 3 minutes for them to reconnect. When this occurs, wait 3 minutes and
click Connect again, until connection is made.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 92
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
1 On the Host PC, click Start | Run | mstsc.exe. The Remote Desktop Connection dialog opens (see Figure 42)
2 Type in the IP address for the server you want to connect to.
3 Click Connect at the bottom of the window to initiate RDP to the Server.
4 When the Login window opens, type in the password service_only.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 93
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 94
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 95
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 96
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 97
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 98
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
1 Insert the CD labeled: CIRS Add-on into the CD-ROM drive of the S1 server:
2 Click My Computer icon on the Desktop, and double-click on the CD icon to launch the Philips Install Supervisor
application.
3 Click Start in the Philips Install Supervisor window.
NOTE Do not touch the mouse or the keyboard until the message “Installation completed” appears. Verify
that the exit code for all components is 0x00000000.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 99
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
4 When the” Installation completed” message appears at the bottom of the Philips Install Supervisor window,
verify that all the items are “Done, (exit code: 0x00000000)”. Click Close in the Philips Install Supervisor window.
5 Eject the CIRS Add-on CD.
1 Click Start | My Computer and right click on the C: drive icon and choose Properties.
2 Ensure that there is at least 2.5 GB of free disk space on Drive C:
If the free disk space is less than 2.5 GB free some disk space to ensure the required minimum.
3 Repeat the procedure for Drive F:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 100
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
1 Insert the CD labeled CIRS Test Raws #1 into the CD-ROM drive of the S1 server.
NOTE If the Philips Install Supervisor does not automatically run, click Start | My Computer and double-click
on the CD icon to launch the Philips Install Supervisor application.
2 Click Start.
NOTE Do not touch the mouse or the keyboard until the message “Installation completed” appears. Verify
that the exit code for all components is 0x00000000.
3 When the Installation completed message appears at the bottom of the Philips Install Supervisor window, verify
that all the items are “Done, (exit code: 0x00000000)”. Click Close in the Philips Install Supervisor window.
4 Eject the Test Raws #1 CD.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 101
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
NOTE Do not touch the mouse or the keyboard until the message “Installation completed” appears. Verify
that the exit code for all components is 0x00000000.
4 When the Installation completed message appears at the bottom of the Philips Install Supervisor window, verify
that all the items are “Done, (exit code: 0x00000000)”.
NOTE A command window runs in the background and will automatically close upon completion.
4 When the Installation completed message appears at the bottom of the Philips Install Supervisor window, verify
that all the items are “Done, (exit code: 0x00000000)”.
NOTE A command window runs in the background and will automatically close upon completion.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 102
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 103
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CIRS Software Installation
4 From the Host computer remotely connect to each of the CIRS servers (S1, S2, S3) and verify that the time and date
is identical on all the computers. If not, perform the following on each of the servers:
a Open Windows Explorer and navigate to C:\Program Files\Philips\CIRS.
b Double-click on TimeSync.
c Close Windows Explorer and disconnect from the CIRS server.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 104
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
40-Channel Configuration
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 105
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
Gantry Components
Gantry Display
Device # on chain File name Check Sum Date
scp_gcp_rev_b.jed df24 Sept. 16, 2004
1
gcp_ctb_U103.hex Aug. 30, 2004
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 106
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
GHost
(453567042301)
Device # on chain File name Check Sum Date
1 453567042301_12_04_03.jed B603
2 TRANS_1_REV_A.mcs 00283e84c Feb. 25, 2004
3 XC2S150_FG256.bsd N/A
GHost
(453567110321)
Device # on chain File name Check Sum Date
1 453567135171_Rev_B.jed Mar. 24, 2005
2 TRANS_1_REV_B.mcs Mar. 14, 2005
3 XC2S150_FG256.bsd Jan. 12, 2005
GMP
Device # on chain File name Check Sum Date
1 453567010351a_RevE_dev1.mcs B603
2 453567010351b_RevE_dev3.mcs 00283e84c May 5, 2005
3 453567010351dev2_4.bsd N/A
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 107
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
CGMP
Device # on chain File Name Check Sum Date
453567084501_DEV1_REV_B.mcs Apr. 21, 2005
453567084501_DEV2_4.bsd Mar. 16, 2005
453567084501_DEV3_REV_B.mcs Apr. 21, 2005
IBox
Device # on chain File Name Check Sum Date
1 I_BOX_VER1_5.mcs Sept. 7, 2004
2 Not for programming
3 72050021.jed Aug. 31, 2004
MDP
Device # on chain File name Check Sum Date
1 362496_U5_rev_D.mcs 002218fbf
Aug. 30, 2004
2 362496_U16_REV_D.bsd N/A
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 108
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
R2D
Device # on chain File name Check Sum
1 362296_U9_REV_C.jed 1531 Aug. 30, 2004
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 109
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
RHost
Device # on chain File name Check Sum
1 cpm_01_27_05-rev_f.mcs 00082234b Feb. 15, 2005
2 dspcpld_rst_update.jed 578b
3 BYPASS Aug. 31, 2004
4 rhostdsp.obj 1f31181
Spellman
Device # on chain File name Check Sum Date
1 407104c2.mcs 001da91af Aug. 30, 2004
RCOM/DPU
Device # on chain File Name Check Sum Date
1 dpu3B_ver_56_0.mcs Feb. 20, 2005
2 dpu3B_ver_56_1.mcs Feb. 20, 2005
3 dpu3A_ver_56_0.mcs Feb. 20, 2005
4 dpu3A_ver_56_1.mcs Feb. 20, 2005
5 dpu2A_ver_56_0.mcs Mar. 9, 2005
6 dpu2A_ver_56_1.mcs Mar. 9, 2005
7 dpu2B_ver_56_0.mcs Mar. 9, 2005
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 110
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
RCOM/DPU
Device # on chain File Name Check Sum Date
8 dpu2B_ver_56_1.mcs Mar. 9, 2005
9 dpu_ext_b_ver_56_0.mcs Mar. 9, 2005
10 dpu_ext_b_ver_56_1.mcs Mar. 9, 2005
11 Skip This Device
12 cyp401.bsd Aug. 21, 2003
13 Skip This Device
14 DPU_EXT_A_ver_56_0.mcs Mar. 8, 2005
15 DPU_EXT_A_ver_56_1.mcs Mar. 8, 2005
16 STD_B4_421_0.mcs Mar. 9, 2005
17 STD_B4_421_1.mcs Mar. 9, 2005
18 dpu_ext_b_ver_56_0.mcs Mar. 9, 2005
19 dpu_ext_b_ver_56_1.mcs Mar. 9, 2005
20 Skip This Device
21 cyp401.bsd Aug. 21, 2003
22 Skip This Device
23 DPU_EXT_A_ver_56_0.mcs Mar. 8, 2005
24 DPU_EXT_A_ver_56_1.mcs Mar. 8, 2005
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 111
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
RCOM/DPU
Device # on chain File Name Check Sum Date
Applause_DPU_1_Version_0056.xsvf May 1, 2005
Applause_DPU_4_Version_0056.xsvf May 1, 2005
From Host
Applause_DMS_Version_0421.xsvf May 1, 2005
Applaus_Rcom_Version_0421.xsvf May 1, 2005
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 112
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
DMC16
Device # on chain File name Check Sum Date
1 B0020D0020V0015.mcs
Aug. 30, 2004
4 B0020D0012V0038.mcs
2, 3, 5, 6 Not for programming
CDMP
Device # on chain File name Check Sum Date
1 CDMP_FPGA_53nf.mcs Aug. 30, 2004
2 Not for programming
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 113
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 114
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
Components on Workstation
CT Box
Device # on chain File name Check Sum Date
scp_gcp_rev_a.jed 5B98 Aug. 30, 2004
1
gcp_ctb_U103.hex Sep. 10, 2004
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 115
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
64-Channel Configuration
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 116
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
Gantry Components
Gantry Display
Device # on chain File name Check Sum Date
scp_gcp_rev_b.jed df24 Sept. 16, 2004
1
gcp_ctb_U103.hex Aug. 30, 2004
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 117
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
GHost
(453567042301)
Device # on chain File name Check Sum Date
1 453567042301_12_04_03.jed B603
2 TRANS_1_REV_A.mcs 00283e84c Feb. 25, 2004
3 XC2S150_FG256.bsd N/A
GHost
(453567110321)
Device # on chain File name Check Sum Date
1 453567135171_Rev_B.jed Mar. 24, 2005
2 TRANS_1_REV_B.mcs Mar. 14, 2005
3 XC2S150_FG256.bsd Jan. 12, 2005
GMP
Device # on chain File name Check Sum Date
1 453567010351a_RevE_dev1.mcs B603
2 453567010351b_RevE_dev3.mcs 00283e84c May 5, 2005
3 453567010351dev2_4.bsd N/A
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 118
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
CGMP
Device # on chain File Name Check Sum Date
453567084501_DEV1_REV_B.mcs Apr. 21, 2005
453567084501_DEV2_4.bsd Mar. 16, 2005
453567084501_DEV3_REV_B.mcs Apr. 21, 2005
IBox
Device # on chain File Name Check Sum Date
1 I_BOX_VER1_5.mcs Sept. 7, 2004
2 Not for programming
3 72050021.jed Aug. 31, 2004
MDP
Device # on chain File name Check Sum Date
1 362496_U5_rev_D.mcs 002218fbf
Aug. 30, 2004
2 362496_U16_REV_D.bsd N/A
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 119
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
R2D
Device # on chain File name Check Sum
1 362296_U9_REV_C.jed 1531 Aug. 30, 2004
RHost
Device # on chain File name Check Sum
1 cpm_01_27_05-rev_f.mcs 00082234b Feb. 15, 2005
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 120
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
RHost
Device # on chain File name Check Sum
2 dspcpld_rst_update.jed 578b
3 BYPASS Aug. 31, 2004
4 rhostdsp.obj 1f31181
Spellman
Device # on chain File name Check Sum Date
1 407104c2.mcs 001da91af Aug. 30, 2004
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 121
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
RCOM/DPU
Device # on chain File Name Check Sum Date
1 dpu3B_orion_3e_0.mcs Feb. 20, 2005
2 dpu3B_orion_3e_1.mcs Feb. 20, 2005
3 dpu3A_orion_3e_0.mcs Feb. 20, 2005
4 dpu3A_orion_3e_1.mcs Feb. 20, 2005
5 dpu_2A_orion_3e_0.mcs Mar. 13, 2005
6 dpu_2A_orion_3e_1.mcs Mar. 13, 2005
7 dpu2B_orion_3e_0.mcs Feb. 20, 2005
8 dpu2B_orion_3e_1.mcs Feb. 20, 2005
9 dpu_EXT_B_Orion_3e_0.mcs Feb. 2, 2005
10 dpu_EXT_B_Orion_3e_1.mcs Feb. 2, 2005
11 Skip This Device
12 cyp401.bsd Aug. 21, 2003
13 Skip This Device
14 DPU_EXT_A_Orion_3e_0.mcs Feb. 20, 2005
15 DPU_EXT_A_Orion_3e_1.mcs Feb. 20, 2005
16 Center_BIP_B400_0.mcs Apr. 6, 2005
17 Center_BIP_B400_1.mcs Apr. 6, 2005
18 dpu_EXT_B_Orion_3e_0.mcs Feb. 20, 2005
19 dpu_EXT_B_Orion_3e_1.mcs Feb. 20, 2005
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 122
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
RCOM/DPU
Device # on chain File Name Check Sum Date
20 Skip This Device
21 cyp401.bsd Aug. 21, 2003
22 Skip This Device
23 DPU_EXT_A_Orion_3e_0.mcs Feb. 20, 2005
24 DPU_EXT_A_Orion_3e_1.mcs Feb. 20, 2005
JTAG apdm_bip_42.mcs Feb. 17, 2005
All_APDMs_Ver_42.xsvf April 7, 2005
DPU_1_Version_3e.xsvf April 26, 2005
From Host DPU_4_Version_3e.xsvf April 26, 2005
Orion_DMS_Version_B400.xsvf April 12, 2005
Rcom_Version_B400.xsvf April 26, 2005
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 123
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
DMC16
Device # on chain File name Check Sum Date
1 B0020D0020V0015.mcs
Aug. 30, 2004
4 B0020D0012V0038.mcs
2, 3, 5, 6 Not for programming
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 124
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
CDMP
Device # on chain File name Check Sum Date
1 CDMP_FPGA_53nf.mcs Aug. 30, 2004
2 Not for programming
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 125
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 126
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions CPM and FPGA Software Versions
Components on Workstation
CT Box
Device # on chain File name Check Sum Date
scp_gcp_rev_a.jed 5B98 Aug. 30, 2004
1
gcp_ctb_U103.hex Sep. 10, 2004
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 127
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Procedure for Reburning CPM Boot and Application Flash
NOTE Gantries having a GMP controller do not have a TDP controller. Therefore, in the User Feedback
window, there will be no response displayed for the TDP controller.
NOTE Revisions listed in the figures below are for demonstration purposes only.
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 128
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Procedure for Reburning CPM Boot and Application Flash
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 129
BrillianceTM CT 40/64-channel Configuration V2.0.2 Installation Instructions Virus Scan Instructions
The Virus Scan application is configured by default to run automatically every night at 2:00 AM.
To run the Virus Scan application manually on the Host computer, do the following:
CSIP Level 1 © 2006 Koninklijke Philips Electronics N.V All Rights Reserved Refer to Front Cover 130