instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Mar.17/ 055001


Date: March, 2017

HISTORY OF INSTRUMENT:

Customer : FEROZSONS Laboratories Ltd.


Equipment : Autoclave
Equipment I.D : LID # QC-117
Manufacturer : Hirayama
Supplier : NA
Model / Type : HV-50
Serial # : 30321113048
Located at : Quality Control Lab (Micro Lab)

CALIBRANT: : Pre-calibrated Digital Temperature Indicator certificate # CAL/JU16/5716 which is


traceable to certificate # CC-6274 of NPSL (PAK)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3

Conditions RANGE OBSERVED


Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

30C 31.23 29.47 ± 2.0C 1.24


35C 36.11 34.54 ± 2.0C 1.11
40C 41.1 40 ± 2.0C 0.77

STATUS: PASSED

Validation on: 01-March-2017 Next Thermal Validation due on: 01-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Mar.17/ 055002


Date: March, 2017

HISTORY OF INSTRUMENT:

Customer : FEROZSONS Laboratories Ltd.


Equipment : Universal Incubator
Equipment I.D : LID # QC-66
Manufacturer : Memmert
Supplier : NA
Model / Type : IN 260
Serial # : D614.0164
Located at : Quality Control Lab (Micro Lab)

CALIBRANT: : Pre-calibrated Digital Temperature Indicator certificate # CAL/JU16/5716 which is


traceable to certificate # CC-6274 of NPSL (PAK)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3

Conditions RANGE OBSERVED


Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

30C 30.99 30.00 ± 2.0C 0.70


35C 35.13 34.54 ± 2.0C 0.79
40C 40.49 39.97 ± 2.0C 0.36

STATUS: PASSED

Validation on: 03-March-2017 Next Thermal Validation due on: 03-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Mar.17/ 055003


Date: March, 2017

HISTORY OF INSTRUMENT:

Customer : FEROZSONS Laboratories Ltd.


Equipment : Universal Incubator
Equipment I.D : LID # QC-50
Manufacturer : Memmert
Supplier : NA
Model / Type : IN 200
Serial # : 910899
Located at : Quality Control Lab (Micro Lab)

CALIBRANT: : Pre-calibrated Digital Temperature Indicator certificate # CAL/JU16/5716 which is


traceable to certificate # CC-6274 of NPSL (PAK)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3

Conditions RANGE OBSERVED


Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

20C 25.36 24.79 ± 2.0C 0.40


30C 30.63 30.01 ± 2.0C 0.43
35C 35.99 34.10 ± 2.0C 1.33

STATUS: PASSED

Validation on: 07-March-2017 Next Thermal Validation due on: 07-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Mar.17/ 055004


Date: March, 2017

HISTORY OF INSTRUMENT:

Customer : FEROZSONS Laboratories Ltd.


Equipment : Hot Incubator
Equipment I.D : LID # QC-25
Manufacturer : Memmert
Supplier : NA
Model / Type : MB6
Serial # : 4585923
Located at : Quality Control Lab (Micro Lab)

CALIBRANT: : Pre-calibrated Digital Temperature Indicator certificate # CAL/JU16/5716 which is


traceable to certificate # CC-6274 of NPSL (PAK)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3

Conditions RANGE OBSERVED


Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

30C 30.81 29.54 ± 2.0C 0.89


35C 35.77 34.61 ± 2.0C 0.82
40C 40.73 39.19 ± 2.0C 1.08

STATUS: PASSED

Validation on: 08-March-2017 Next Thermal Validation due on: 08-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Mar.17/ 055005


Date: March, 2017

HISTORY OF INSTRUMENT:

Customer : FEROZSONS Laboratories Ltd.


Equipment : Dry Heat Sterilizer
Equipment I.D : LID # QC-68
Manufacturer : Memmert
Supplier : NA
Model / Type : UN 260
Serial # : B614.0022
Located at : Quality Control Lab (Micro Lab)

CALIBRANT: : Pre-calibrated Digital Temperature Indicator certificate # CAL/JU16/5716 which is


traceable to certificate # CC-6274 of NPSL (PAK)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3

Conditions RANGE OBSERVED


Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

175C 176.97 175.27 ± 2.0C 1.20

STATUS: PASSED

Validation on: 07-March-2017 Next Thermal Validation due on: 07-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Mar.17/ 055006


Date: March, 2017

HISTORY OF INSTRUMENT:

Customer : FEROZSONS Laboratories Ltd.


Equipment : Dry Heat Oven
Equipment I.D : LID # QC-74
Manufacturer : Memmert
Supplier : NA
Model / Type : UN 55
Serial # : B214.3756
Located at : Quality Control Lab

CALIBRANT: : Pre-calibrated Digital Temperature Indicator certificate # CAL/JU16/5716 which is


traceable to certificate # CC-6274 of NPSL (PAK)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3

Conditions RANGE OBSERVED


Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

70C 70.98 70.17 ± 2.0C 0.57


106C 106.34 105.06 ± 2.0C 0.90
100C 101.46 100.28 ± 2.0C 0.83
110C 112.37 111.03 ± 2.0C 0.94

STATUS: PASSED

Validation on: 08-March-2017 Next Thermal Validation due on: 08-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Mar.17/ 055007


Date: March, 2017

HISTORY OF INSTRUMENT:

Customer : FEROZSONS Laboratories Ltd.


Equipment : Dry Heat Oven
Equipment I.D : LID # QC-76
Manufacturer : Memmert
Supplier : NA
Model / Type : UN 260
Serial # : B615.0789
Located at : Quality Control Lab

CALIBRANT: : Pre-calibrated Digital Temperature Indicator certificate # CAL/JU16/5716 which is


traceable to certificate # CC-6274 of NPSL (PAK)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3

Conditions RANGE OBSERVED


Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

70C 71.12 69.00 ± 2.0C 1.49


100C 101.44 99.95 ± 2.0C 1.05

STATUS: PASSED

Validation on: 09-March-2017 Next Thermal Validation due on: 09-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Mar.17/ 055008


Date: March, 2017

HISTORY OF INSTRUMENT:
Customer : FEROZSONS Laboratories Ltd.
Equipment : Stability Chamber
Equipment I.D : LID # QC-83
Manufacturer : Memmert
Supplier : NA
Model / Type : HPP 750
Serial # : W816.0250
Located at : Quality Control Lab

CALIBRANT: :Pre-calibrated Thermo Hygro Meter certificate # CAL/JU16/5711 which traceable


to certificate # CC-5429 of NPSL (PAK Approved Lab No.001)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3

Conditions RANGE OBSERVED


Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

30C 31.00 30.00 ± 2.0C 0.70


40C 41.00 40.00 ± 2.0C 0.70

Humidity (%RH)
SET %RH OBSERVED %RH Tolerance Percentage Error

Max. %RH Min. %RH

65 69.5 64.5 ± 5.0C 3.53


75 79.5 76.00 ± 5.0C 2.47

STATUS: PASSED

Validation on: 09-March-2017 Next Thermal Validation due on: 09-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Mar.17/ 055009


Date: March, 2017

HISTORY OF INSTRUMENT:

Customer : FEROZSONS Laboratories Ltd.


Equipment : Stability Chamber
Equipment I.D : LID # PD-04
Manufacturer : Hot Pack
Supplier : NA
Model / Type : 435304
Serial # : 79431
Located at : Quality Control Lab

CALIBRANT: :Pre-calibrated Thermo Hygro Meter certificate # CAL/JU16/5711 which traceable


to certificate # CC-5429 of NPSL (PAK Approved Lab No.001)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3
Conditions RANGE OBSERVED
Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

30C 31.00 30.00 ± 2.0C 0.70


40C 42.00 40.00 ± 2.0C 1.41

Humidity (%RH)
SET %RH OBSERVED %RH Tolerance Percentage Error

Max. %RH Min. %RH

65 62.5 59.5 ± 5.0C 3.83


75 72.5 67.50 ± 5.0C 3.53

STATUS: PASSED

Validation on: 09-March-2017 Next Thermal Validation due on: 09-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Mar.17/ 055010


Date: March, 2017

HISTORY OF INSTRUMENT:

Customer : FEROZSONS Laboratories Ltd.


Equipment : Stability Chamber
Equipment I.D : LID # QC-84
Manufacturer : Memmert
Supplier : NA
Model / Type : HPP 750
Serial # : W816.0378
Located at : Quality Control Lab

CALIBRANT: :Pre-calibrated Thermo Hygro Meter certificate # CAL/JU16/5711 which traceable


to certificate # CC-5429 of NPSL (PAK Approved Lab No.001)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3

Conditions RANGE OBSERVED


Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

30C 31.5 29.00 ± 2.0C 1.76


40C 41.00 39.5 ± 2.0C 1.06

Humidity (%RH)
SET %RH OBSERVED %RH Tolerance Percentage Error

Max. %RH Min. %RH

65 69.5 66.5 ± 5.0C 2.12


75 79.5 76.5 ± 5.0C 2.12

STATUS: PASSED

Validation on: 09-March-2017 Next Thermal Validation due on: 09-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE
MOHSIN RAZA

Certificate No. TVAL (1/1) / Mar.17/ 055011


Date: March, 2017

HISTORY OF INSTRUMENT:
Customer : FEROZSONS Laboratories Ltd.
Equipment : Stability Chamber
Equipment I.D : LID # PD-05
Manufacturer : Memmert
Supplier : NA
Model / Type : HPP 749
Serial # : X512.0018
Located at : Quality Control Lab

CALIBRANT: :Pre-calibrated Thermo Hygro Meter certificate # CAL/JU16/5711 which traceable


to certificate # CC-5429 of NPSL (PAK Approved Lab No.001)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3
Conditions RANGE OBSERVED
Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

30C 31.0 29.5 ± 2.0C 1.06


40C 40.5 39.5 ± 2.0C 0.70

Humidity (%RH)
SET %RH OBSERVED %RH Tolerance Percentage Error

Max. %RH Min. %RH

65 69.5 62.00 ± 5.0C 2.12


75 78.5 73.5 ± 5.0C 1.06

STATUS: PASSED

Validation on: 09-March-2017 Next Thermal Validation due on: 09-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Mar.17/ 055012


Date: March, 2017

HISTORY OF INSTRUMENT:
Customer : FEROZSONS Laboratories Ltd.
Equipment : Stability Chamber
Equipment I.D : LID # PD-05
Manufacturer : Memmert
Supplier : NA
Model / Type : HPP 749
Serial # : X512.0018
Located at : Quality Control Lab

CALIBRANT: :Pre-calibrated Thermo Hygro Meter certificate # CAL/JU16/5711 which traceable


to certificate # CC-5429 of NPSL (PAK Approved Lab No.001)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3
Conditions RANGE OBSERVED
Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
Temperature
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

30C 31.0 29.5 ± 2.0C 1.06


40C 40.5 39.5 ± 2.0C 0.70

Humidity (%RH)
SET %RH OBSERVED %RH Tolerance Percentage Error

Max. %RH Min. %RH

65 69.5 62.00 ± 5.0C 2.12


75 78.5 73.5 ± 5.0C 1.06

STATUS: PASSED

Validation on: 09-March-2017 Next Thermal Validation due on: 09-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE

Certificate No. TVAL (1/1) / Jan.22/01/003


Date: January, 2022

HISTORY OF INSTRUMENT:
Customer : FEROZSONS Laboratories Ltd.
Equipment : Autoclave
Equipment I.D : LID # QC-117
Manufacturer : Hirayama
Supplier : NA
Model / Type : HV-50
Serial # : 30321113048
Located at : Quality Control Lab (Micro Lab)

CALIBRANT: : Pre-calibrated Pressure Test Gauge certificate # CAL/AU 21/6309 which is traceable to
certificate # APCIC (1267)/ILO # 224/16.06.14 (PCSIR)
: Pre-calibrated Digital Temperature Indicator certificate # CAL/JU21/6381 which is
traceable to certificate # APCIC/EL (1342) of PCSIR

Environmental Conditions during Thermal Validation ISO 17025, Section 5.3

Conditions RANGE OBSERVED


Temperature 22 ± 2°C 21.5°C
Humidity 50 ± 10% 52 %

VALIDATION DATA:
Temperature
SET TEMP. OBSERVED AVERAGE TEMP. (°C) (45 Min) Average Deviation
(°C) (8 LOCATIONS) (°C) (°C)
116C 116.77/116.84/116.92/116.84/116.46/116.76/116.77/116.52 116.74 0.74

SET TEMP. OBSERVED AVERAGE TEMP. (°C) (20 Min) Average Deviation
(°C) (8 LOCATIONS) (°C) (°C)
121C 121.49/121.65/121.65/121.68/121.61/121.51/121.73/121.62 121.62 0.62

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE
SET TEMP. OBSERVED AVERAGE TEMP. (°C) (5 Min) Average Deviation
(°C) (8 LOCATIONS) (°C) (°C)
132C 132.39/132.41/132.35/132.41/132.39/132.42/132.42/132.44 132.41 0.41

Pressure
Applied Value Observed Value Deviation/Error
Unit: Kgf/CM Unit: psi Unit: psi
1.3 15 0.00
STATUS: PASSED

Validation on: 10-January-2022 Next Thermal Validation due on: 10-January-2024


……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
Fahad Tariq

Certificate No. TVAL (1/1) / Mar.17/ 055014


Date: March, 2017

HISTORY OF INSTRUMENT:

Customer : FEROZSONS Laboratories Ltd.


Equipment : Furnace
Equipment I.D : LID # QC-20
Manufacturer : Thermolyne CO.
Supplier : NA
Model / Type : F-D1520M
Serial # : 266/5358
Located at : Quality Control Lab

CALIBRANT: : Pre-calibrated Digital Temperature Indicator certificate # CAL/JU16/5716 which is


traceable to certificate # CC-6274 of NPSL (PAK)

Environmental Conditions during Thermal Validation


ISO 17025, Section 5.3

Conditions RANGE OBSERVED


Temperature 22 ± 2°C 21°C
Humidity 50 ± 10% 54%

VALIDATION DATA:
SET TEMP OBSERVED TEMP Tolerance Percentage Error

Max. TEMP Min. TEMP

100C 102.12 99.5 ± 5.0C 1.85


200C 201.44 199.7 ± 5.0C 1.23

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com
instruemends THERMAL VALIDATION CERTIFICATE
300C 302.00 298.00 ± 5.0C 2.82
400C 403.00 398.50 ± 5.0C 3.18
500C 503.50 497.7 ± 5.0C 4.10

STATUS: PASSED

Validation on: 09-March-2017 Next Thermal Validation due on: 09-March-2018

……………………… ……………………
( VALIDATED BY ) ( CHECKED BY )
MOHSIN RAZA

Office: 2-G Mian Appartments, 14 Hussain Street Old Muslim Town, Wahdat Road, Lahore.
Phone Office: 042-35912901, Fax Office: 042-35912902
Email: truemend@gmail.com