BD FACSCount CD4 Reagents
Accuracy
CD4 absolute count and percentage
results were compared with results from Parameters n R2 Slope Intercept Range
BD Tritest™ CD3 FITC/CD4 PE/CD45 Absolute CD4 (cells/µL) 101 0.981 0.971 12.695 59 – 3,404
PerCP with BD Trucount™ tubes on the
CD4 Percentage 99 0.99 0.999 -0.391 5.51 – 64.69
BD FACSCalibur™ flow cytometer using
BD Multitest™ software.
5000
5000 70 70
BD FACSCount CD4 absolute count
BD FACSCount CD4 absolute count
60 60
50 50
3000
3000
40 40
2000
2000 30 30
20 20
1000
1000
10 10
0 0 0 0
0 0 1000
1000 2000
2000 3000
3000 4000
4000 5000
5000 0 0 10 10 20 20 30 30 40 40 50 50 60 60 70 70
BD
BDTritest
TritestCD4
CD4absolute
absolutecount
count BD
BDTritest
TritestCD4
CD4percentage
percentage
Regression plot of test versus predicate for CD4 absolute counts Regression plot of test versus predicate for CD4 percentages
User-Reportable Ranges
Absolute counts: 50 to 5,000 CD4 cells/μL
Percentages: 5% to 65%
Stability
Staining: Within 24 hours of draw
Analysis: Within 48 hours of staining