VALIDATION
REPORT
Revision 6, April 2022
Veratox
for Almond
®
Summary
Veratox® for Almond (NEOGEN® item 8440) is intended for the
full quantitative analysis or simple screening of almond protein
residues in food products such as cookies, crackers, chocolate
bars, ice cream, beverages, cereals, and clean-in-place rinses.
Quantitation with Veratox for Almond ranges from 2.5–25 ppm,
and can be extended by dilution of positive extracts. The
validation report details the findings of the experimental
evaluation to establish product claims for the Veratox for
Almond test.
Limit of detection: The limit of detection was found to be <1.0 ppm
and the limit of quantitation was found to be 2.5 ppm.
Ruggedness: Inter (CV 7.6%) and intra (CV 3.3%) assay variability
was tested across operator, day, lot, and concentration. Results
met acceptance criteria.
Cross-reactivity: Cross reacts with sunflower seed, apple seed,
and common Prunus genus foods (ie. apricot, cherries, peaches,
nectarines, and plums). Cross-reactivity has been identified
with the seed (stone); however the flesh (seedless fruit) does not
cross react.
Market basket screening: A series of various foods were purchased
at local markets and screened for almond allergen content.
Spike and recovery: The Veratox for Almond test kit demonstrated
excellent recovery in a variety of matrices.
Robustness: Variations in the extraction and ELISA procedure
were investigated with no significant effect on the functionality
of the test kit.
2
© NEOGEN Corporation, 2022. NEOGEN and Veratox are registered trademarks of NEOGEN Corporation. All rights reserved. 8440_Veratox-for-Almond_Validation-Report_V1
Materials and Methods
All tests were conducted on standard quality control (QC) approved lots of Veratox for Almond test kits. All assays
were performed in accordance with the test kit insert. Spiked samples were spiked and held for 24 hours to allow
spike to fully incorporate into the sample.
Limit of Detection
The mean of ten known negative samples was calculated and the limit of detection (LOD) was calculated by adding
the mean to 3x standard deviation (StDev).
Replicate ppm
1 0.28
2 0.28
3 0.27
4 0.25
5 0.27
6 0.26
7 0.29
8 0.27
9 0.28
10 0.26
Mean 0.27
StDev 0.01
NEOGEN.com
3
© NEOGEN Corporation, 2022. NEOGEN and Veratox are registered trademarks of NEOGEN Corporation. All rights reserved. 8440_Veratox-for-Almond_Validation-Report_V1
Ruggedness
Inter-assay Variability
0.272 0.284 0.592 0.572 0.887 0.880 1.452 1.375 2.213 2.160
0.295 0.344 0.566 0.587 0.801 0.848 1.205 1.331 1.944 2.081
0.256 0.264 0.551 0.572 0.787 0.813 1.216 1.092 1.919 1.995
0.261 0.287 0.580 0.571 0.838 0.860 1.364 1.298 2.036 1.900
0.287 0.285 0.609 0.576 0.840 0.841 1.102 1.284 2.042 1.780
0.302 0.270 0.535 0.538 0.726 0.791 1.135 1.201 1.831 1.820
0.286 0.265 0.586 0.472 0.795 0.789 1.233 1.239 1.880 1.815
Intra-assay Variability
0.247 0.249 0.524 0.539 0.755 0.752 1.321 1.237 1.974 1.972
0.238 0.241 0.518 0.516 0.728 0.744 1.335 1.227 1.985 1.990
0.240 0.262 0.521 0.521 0.750 0.768 1.240 1.243 2.032 2.057
0.251 0.245 0.431 0.523 0.746 0.721 1.234 1.211 1.955 1.966
0.245 0.244 0.535 0.537 0.765 0.736 1.344 1.245 2.086 2.088
0.245 0.251 0.532 0.517 0.762 0.736 1.256 1.247 2.009 2.025
0.246 0.251 0.519 0.510 0.752 0.698 1.358 1.231 1.893 1.896
Mean OD 0.248 0.517 0.744 1.260 1.989
NEOGEN.com
4
© NEOGEN Corporation, 2022. NEOGEN and Veratox are registered trademarks of NEOGEN Corporation. All rights reserved. 8440_Veratox-for-Almond_Validation-Report_V1
Cross-reactivity
Cross reacts with sunflower seed, apple seed, and common Prunus genus foods (ie. apricot, cherries, nectarines, peaches,
and plums). Cross-reactivity has been identified with the seed (stone); however the flesh (seedless fruit) does not cross
react. A panel of 39 additional samples including nuts, grains, proteins, and seeds were also tested and did not
demonstrate any cross-reactivity with the Veratox for Almond test.
NEOGEN.com
5
© NEOGEN Corporation, 2022. NEOGEN and Veratox are registered trademarks of NEOGEN Corporation. All rights reserved. 8440_Veratox-for-Almond_Validation-Report_V1
Market Basket
Market basket samples with and without almond listed as an ingredient were collected from local markets. All results
above the limit of quantitation (LOQ) were not considered quantitative until further dilutions were made, which is
reflected in the data below.
1:500 0 0
Straight 0 0
1:500 0 0
NEOGEN.com
6
© NEOGEN Corporation, 2022. NEOGEN and Veratox are registered trademarks of NEOGEN Corporation. All rights reserved. 8440_Veratox-for-Almond_Validation-Report_V1
Spike and Recovery
Veratox for Almond with recovery from spiked samples.
NEOGEN.com
7
© NEOGEN Corporation, 2022. NEOGEN and Veratox are registered trademarks of NEOGEN Corporation. All rights reserved. 8440_Veratox-for-Almond_Validation-Report_V1
Robustness
Incubation Time
The recommended incubation time in minutes is 10:10:10. Studies were performed with less, more, and
variable incubations to demonstrate the effect of timing.
0 ppm Almond Allergen Added 5 ppm Almond Allergen Added 15 ppm Almond Allergen Added
Almond Recovered
Day 1 Times Replications Avg. Replications Avg. % Replications Avg. %
8:8:8 Minutes 0.0 0.0 0.1 0.0 6.2 5.4 5.7 5.8 115.3 18.1 15.3 18.1 17.2 114.4
10:10:10 Minutes 0.3 0.3 0.1 0.2 5.6 5.7 4.7 5.3 106.7 16.3 16.7 14.7 15.9 106.0
12:12:12 Minutes 0.4 0.4 0.3 0.4 5.2 5.2 5.4 5.3 105.3 15.4 16.5 16.1 16.0 106.7
8:10:12 Minutes 0.0 0.0 0.1 0.0 5.5 5.4 5.7 5.5 110.7 15.6 15.6 16.7 16.0 106.4
Day 2 Times Replications Avg. Replications Avg. % Replications Avg. %
8:8:8 Minutes 0.0 0.0 0.0 0.0 4.5 4.1 5.1 4.6 91.3 12.8 13.2 14.6 13.5 90.2
10:10:10 Minutes 0.2 0.0 0.2 0.1 4.8 4.5 4.8 4.7 94.0 14.5 13.3 14.6 14.1 94.2
12:12:12 Minutes 0.1 0.4 0.3 0.3 5.0 4.6 5.0 4.9 97.3 15.2 13.8 15.5 14.8 98.9
8:10:12 Minutes 0.1 0.3 0.0 0.1 4.9 4.6 4.8 4.8 95.3 14.4 13.4 14.0 13.9 92.9
Reagent Volume
The recommended reagent volume is 100 µL. This study was performed using 95 µL and 105 µL volumes to
demonstrate the effect of reagent volumes.
0 ppm Almond Allergen Added 5 ppm Almond Allergen Added 15 ppm Almond Allergen Added
Almond Recovered
Day 1 Volumes Replications Avg. Replications Avg. % Replications Avg. %
95 µL 0.2 0.0 0.2 0.1 5.5 5.8 6.2 5.8 116.7 15.8 17.5 18.0 17.1 114.0
100 µL 0 0 0 0 5.9 5.4 5.6 5.6 112.7 16.9 16.7 18.4 17.3 115.6
105 µL 0.4 0 0.2 0.2 5.8 5.4 4.9 5.4 107.3 15.7 15.6 14.3 15.2 101.3
Day 2 Volumes Replications Avg. Replications Avg. % Replications Avg. %
95 µL 0.1 0.2 0.0 0.1 5.4 4.9 5.0 5.1 102.0 13.2 14.7 15.6 14.5 93.7
100 µL 0.0 0.0 0.0 0.0 4.6 4.6 4.3 4.5 90.0 13.1 14.2 14.0 13.8 91.8
105 µL 0.0 0.3 0.1 0.1 4.6 4.8 4.5 4.6 92.7 14.0 12.8 13.2 13.3 88.9
NEOGEN.com
8
© NEOGEN Corporation, 2022. NEOGEN and Veratox are registered trademarks of NEOGEN Corporation. All rights reserved. 8440_Veratox-for-Almond_Validation-Report_V1
Shaking Effects
The recommended shake time at the beginning of each incubation is 20 seconds. This study compares no shaking
and a 30 second shake to the standard 20 second shake to demonstrate the effect of shaking.
0 ppm Almond Allergen Added 5 ppm Almond Allergen Added 15 ppm Almond Allergen Added
Almond Recovered
Day 1
Replications Avg. Replications Avg. % Replications Avg. %
Shake Times
0 Seconds 0.5 0.0 0.0 0.2 5.6 5.3 5.3 5.4 108.0 15.7 17.3 16.2 16.4 109.3
20 Seconds 0.2 0 0.2 0.1 4.6 5.4 5.2 5.1 101.3 14.4 15.2 14.2 14.6 97.3
30 Seconds 0.2 0.1 0.4 0.2 4.5 2.3 4.6 3.8 76.0 13.7 13.4 12.7 13.3 88.4
Day 2
Replications Avg. Replications Avg. % Replications Avg. %
Shake Times
0 Seconds 0.0 0.0 0.2 0.1 4.7 4.5 5.0 4.7 94.7 13.6 12.9 14.1 13.5 90.2
20 Seconds 0.0 0.0 0.4 0.1 4.5 4.5 4.9 4.6 92.7 13.1 12.8 12.0 12.6 84.2
30 Seconds 0.2 0.0 0.0 0.1 4.6 4.4 4.5 4.5 90.0 12.5 12.4 14.4 13.1 87.3
Conclusion
Throughout a wide range of tests, Veratox for Almond demonstrated excellent spike recovery and superior test
robustness. This, coupled with its reproducibility, make Veratox for Almond a good choice for the quick, accurate
detection of almond contamination in a range of food products.