DIGITAL HEIGHT GAUGE W.

I 86
OPERATION AND MAINTENANCE MANUAL
Before operating the instrument, please read the manual thoroughly, and retain it for future reference.
IMPORTANT !
 Do not use the instrument near a high-voltage or magnetic equipment.
 Do not measure electrified body.
 Do not disassemble or modify the instrument and any excessive shock should avoided.
 The carbide- tipper scriber is very sharp. Handle it carefully to avoid injury.
 Wipe the anti-rust oil from the instrument with a soft cloth when use the instrument at the first time; and make sure
the carbide-tipper scriber, the racks, the base and other reference surface of the instrument are clean.
 Please take anti-rust oil, if do not use the instrument for a long time.
 Before taking measurements with the instrument, take sure "ORG "is ok.
NOTES ! (Non•fault phenomenon)
Status light brightness changes:Without any operation in 3 minutes, the system will turn to sleeping mode. There
are two options for turn the driving wheel or press any key to restore display.
Auto Power •Off: More than 30 minutes without any operation, pressing " " Key for restart the instrument.

 FEATURES
High accuracy Clear OLED display Auto/delay power- off
Power: 3.7V rechargeable lithium-ion battery Relative humidity:≤80%
Operating temp.:0℃~40℃ Storage temp.:-20℃~60℃

 FUNCTIONS
Language selection Absolute origin correction
Absolute/relative measurement Metric/inch conversion
Ability to hold the measured values Zero- setting of display
Preset function for entering a known value or initial value Set and determine
tolerance Bluetooth data output

 SPECIFICATION
Range Resolution Accuracy

0-300mm/0-12” 0.01mm/.0005” ±0.015mm


0-500mm/0-20” 0.01mm/.0005” ±0.025mm
0-600mm/0-24” 0.01mm/.0005” ±0.025mm

1
DIGITAL HEIGHT GAUGE W.I 86
 NOMENCLATURE
Stopper
Locking device Battery lid

Digital parts driving wheel charging port


Locking nut
Scriber clamp box Reset
Slider
Scriber Screw
Spherical contactor
Column
Plane contactor
Base

Fig.1

 CONTROL PANEL
Data hold prompt INC prompt Tolerance determine prompt
Unit
Measuring data Power prompt
2nd function prompt HOLD INC TOL Status light
Charging light
Mode conversion Bluetooth prompt
Inch,Metric setting
Tolerance determine
ABS,INC measurement
Power 2nd function
Origin correction Zero set
Data hold
Bluetooth output Tolerance prompt Fig.2
 INSTALL BATTERY
Unscrew the battery lid screw, remove the battery lid, pull out the interface plug from the battery box,and connect battery. Push the battery
with the belt and cable into the battery box then cover the battery lid and tighten the screw (Fig.3).

Battery lid Plug(in the battery box)


Battery lid

Screw
Screw Belt Plug Battery Fig.3

When charging?
Change the battery, when the instrument not function and the power light twinkle, Insert
the charger plug to power, the charging begins when the red light is on. It shows charging complete when the red light is off.

 USING ATTACHMENT

Fig.4

 KEY FUNCTION
 About " ", "in/mm", "2ndF", " ZERO" and "HOLD" key
Pressing " " Key for power on or off.
Pressing "in/mm " key for changeover between metric and inch.

2
DIGITAL HEIGHT GAUGE W.I 86
Pressing "2ndF " key for changeover function mode with other key.
Pressing "ZERO " key for zero-setting of display anywhere within the measuring range.
Pressing "HOLD" key at any position, the OLED will display and hold the measured value of this location.
 About "ORG " key
"ORG" mode is used to correct the origin of the instrument. Do not measure in “ORG” state, should be correct the origin
after to start measuring.
Before using this instrument for the first time, change the battery, after the battery ran out and press the reset button to restart the
system, please re-correct the origin.
Suggest correction the origin on a regular.
Operating steps:
(1) Pressing"2ndF"and"ORG"key on the control panel, the OLED will display the interface: (When change the battery, after the
battery ran out and restart the system, will automatically enter the interface.)
(2) Turn the driving wheel slowly and make the slider move down until make the spherical contactor contact with the plane contactor
on the base (Fig.5),

If the OLED display this interface:

then pressing "2ndF" and "ZERO" key,

Fig.5
the OLED will display ,it shows correction complete. BASE

Pressing "2ndF" and" ORG" key to exit then you can start making measurements.

If the OLED display such as interface , and there is" Force Over" prompt,
it shows the measuring force is too large, please according to the step (2) operation again.
 About "MODE" key
"MODE" mode is used to set one of the function: selection language and set tolerance and preset H and set Bluetooth data output and so
on(Fig.6, Fig.7, Fig.17, Fig.21).
 About language selection (Fig.6):
2ndF

MODE (five times)


.

in/mm or ZERO: Selection language 2ndF MODE EXIT

Fig.6
 About ”TOL” key
" TOL" mode is used to determine tolerance.
(1) Setting the tolerance limits (Fig.7): For example ( +0.15
150 mm)
-
2ndF

MODE (twice )
.

HOLD or ABS: Moving cursor


MODE
in/mm or ZERO: Selection value

HOLD or ABS: Moving cursor


2ndF MODE EXIT
in/mm or ZERO: Selection value
Fig.7
(2) Using:

Pressing "TOL" key the OLED will display "TOL" prompt, the digital height gauge worked at determining tolerance mode at the moment

3
DIGITAL HEIGHT GAUGE W.I 86
(Fig.8). When measured values beyond the limits of tolerance, the OLED will display "+ OV "or "- OV " prompt during the measurement
(Fig.9, Fig10). If no prompt, it shows the measured values is qualified (Fig.11).
Pressing "TOL" key once more to exit and the "TOL" prompt disappear.

Fig.8
Fig.9 Fig.10 Fig.11
 About "ABS " key
"ABS" mode is default setting. For example (Fig.12):
(1) Pressing "ZERO" key at position 1 (datum plane),then complete measurement at position 2, the OLED will display measured value
(Fig.13);
(2) Pressing "ABS" key the OLED will display "INC" prompt at position 2, this moment the digital height gauge worked at "INC"
(comparison)mode, and the display be reset(Fig14.), the slider move upwards and complete measurement at position 3, the OLED
will display reading of this time(Fig.15). At this moment if pressing "ABS" key once more, then the OLED will display the sum of the
front reading and reading of this time (from datum plane to position 3) (Fig.16) .
3
("IN

2 Fig.13 Fig.14
("AB

("AB

Fig.15 Fig.16
Datum plane
Fig.12
 Preset function(Digital entry)
Preset function is used to enter a known value for a reference plane which cannot be directly contacted by the height gauge measures
distances from the indirect reference to any surface of the workpiece. For example (Fig.18):
(1) Preset H: Before taking measurement with the height gauge, enter a known value for a reference plane at first (Fig.17).
2ndF
MODE
HOLD or ABS: Moving cursor
in/mm or ZERO: Selection value 2ndF MODE EXIT
Fig. 17
(2) Using: Pressing "2ndF" and "ZERO" key, the OLED will display the preset value of the reference plane(Fig.19), this moment the

digital height gauge worked at presetting function mode. And go on measuring at the need to measure position. the OLED will display
the sum of the front reference reading and reading of this time (Fig.20).

Preset H Fig.19

Fig.20 Datum plane


Fig.18
 About " " key and Bluetooth
The mode is used to fill the measured value in the blank of the report file with WORD, EXCEL or NOTEPAD, etc.
Bluetooth data output formats (Fig.21):
2ndF

MODE (four times)


.

HOLD or ABS: Moving cursor


2ndF MODE EXIT
in/mm or ZERO: Selection value
Fig.21

4
DIGITAL HEIGHT GAUGE W.I 86
Notes: Those functions are the same as icon buttons " " , " ", " ", " ", " ", " " in the interface as the corresponding computer keyboard
keys. The cursor will move according to the selected mode.
Using: (1) Pressing "2ndF" and " " key on the digital panel, the OLED will display " " icon, it shows Bluetooth function has turned on.
(2) User pairing(Bluetooth Version 2.1 and above):
①Paired with the computer:Before using , please make sure Bluetooth drivers has installed on the computer:

Find out the" " icon in the lower right corner of computer, right-click " " and select "Add Devices", or click "Control panel" and
select "Device manager" , "Devices and printers", "Add Devices" in turn, and then waiting for showing "HG2400", double-click
"HG2400"complete matching, you can use the Bluetooth function.
②Paired with mobile phone: Find out and click " "icon on the phone then select "search equipment" and then waiting
for showing "HG2400", next to click "HG2400"complete matching, you can use the Bluetooth function.
(3) start transmitting:Open the report file, take the cursor to stays the filling position then press "2ndF"and " " key on the digital
panel, the measured value will be output and filled in the blank.
Press the " " key on the digital panel each time, a data will be output.
(4) Pressing "2ndF" and " "key again, the Bluetooth function is disabled, and the " "icon disappear.
 ACCESSORIES

NAME QUANTITY NAME QUANTITY NAME QUANTITY


Carbide‐tipper scriber 1 Screwdriver 1 Battery 1
Clamping slider 1 Needle 1 Report 1
Attachment 1 Charger 1 Operation manual 1

You might also like