Technical News
Subject: New fan assembly is released

RECORD IDENTIFIER

Ref No.: TXQ-19101- BC-5000 Relate To:


Version: V1.0 Software/Manual Upgrade
Author: Franky Wang Parts Change
Date: 2019-10-09 Appearance
Model: BC-5000 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,

Because of statutory and regulatory requirements, the direction of power switch is


changed to vertical from horizontal. Please check the details of difference as below.
(1) Power switch.

Before changing

After changing

Ref No.: TXQ-19101- BC-5000 Page 1 of 4


Technical News- For reference only
QF254.02(2.0)

(2) Fan assembly.

Before changing

After changing

The direction of power switch Order number Part name

Horizontal (Before changing) 115-065024-00 Fan assembly(FRU)

Vertical (After changing) 115-017923-00 Fan assembly

Ref No.: TXQ-19101- BC-5000 Page 2 of 4


Technical News- For reference only
QF254.02(2.0)

(3) Supporting plate.

Before changing

After changing

There is a gap on the bottom of supporting plate after changing.

Caution: Hence, if you need to replace the fan assembly of analyzer, please check the
direction of power switch firstly and then apply for corresponding spare part.

If you need download information or have any question, please feel free to contact Mindray
regional service engineer or Mindray Service Department (email: service@mindray.com).

Thank you for your kind attention and cooperation.

Sincerely yours,
Mindray Service Department
Mindray Co., Ltd.

Ref No.: TXQ-19101- BC-5000 Page 3 of 4


Technical News- For reference only
QF254.02(2.0)

Technical News Revision


Subject: New fan assembly is released Ref No: TXQ-19101- BC-5000

Version Effective Date Description


1.0 2019-10-09 Document released.