AccuSim-BP Desktop

NIBP Si m u lat o r

Not many manufacturers of NIBP simulation devices

Innovation by design
can make the claim that they have worked as closely
with the actual NIBP monitoring manufacturers, or at
least not to the same level of involvement, that has
resulted in the AccuSim NIBP simulation technology
platform. Even fewer can make the claim that they
have physiologically-correct waveforms in their NIBP
simulations. And the smallest group would be those
that can claim 0.5% of setting accuracy on dynamic
NIBP simulations. But Datrend Systems can make all
three of these claims relative to the AccuSim NIBP
technology platform for NIBP simulation.

■ Accurate NIBP Simulation: Finally, NIBP simulation AccuSim-BP Desktop

numbers that match on your simulator and monitor!


AccuSim-BP Desktop's innovative CalTables™ deliver accurate testing
of NIBP monitors, typically within 1-2mmHg or better, of the setting.
■ Manufacturer-Certified NIBP Simulation: The AccuSim-BP Desktop’s CalTables™
were developed with, and certified by, NIBP monitor manufacturers. Everyone can
now test using the same accurate and certified NIBP Simulator. AccuSim-BP Desk
top’s innovative CalTables™ deliver what we consider to be the most accurate,
repeatable, and reliable NIBP simulation on the market today!
■ Upgradeable CalTables™: The AccuSim-BP Desktop includes CalTables™ for
major manufacturers. CalTables™ updates ensure that your investment in the
AccuSim-BP Desktop NIBP Simulator is protected well into the future.
■ Internal Pump and Battery: The AccuSim-BP Desktop has both an internal pump
and rechargeable battery to maximize on the flexibility and portability of your
NIBP simulation and testing.
■ Built-In Adult and Neonate Cuff Supports: With the AccuSim-BP, there is no
need for external cuff mandrels to support your NIBP simulation.

Perhaps best of all, is the fact that the AccuSim-BP Desktop NIBP Simulator is a part of
the Datrend ES601 Plus test automation family. This means that you can connect your
AccuSim-BP Desktop NIBP Simulator to your ES601 Plus for totally automated testing
of your medical devices, with storage of test results.

Accur acy M att er s • www.datr end.com


AccuSim-BP Desktop – Performance Specifications
TM
AccuSim-BP NIBP CalTabIes Digital Manometer: Display:
Datascope Accutorr Plus Pressure Range: 0.0 to 400.0 mmHg Non-Glare Graphic LCD
GE Medical Dash 4000 Accuracy: ± 0.5 mmHg, traceable to NIST CCFL Backlight
GEiMedical Dinamap Pro. Resolution: ± 0.1 mmHg Resolution: 240x320 (1/4 VGA)
Philips Medical M3 Thumbwheel contrast control
Welch-Allyn Propaq Encore Leak Test:
Philips Medical VS1 Start Pressure: Power:
Colin Medical 200 mmHg nominal, 100-240 VAC, 50 watts, 50-60 Hz,
Criticare 10 to 370 mmHg via Remote Control Desktop Switcher
CAS Medical Pressure Drop: 0 to 400 mmHg Output: 24VDC @ 2.1A, 6 foot cable
SunTech 6 Elapsed Time: 60 Seconds Internal Battery: NiMH
(Please contact Datrend for the High Resolution Manometer: 0.0 to Voltage: 19.2V
availability of other calibration tables] 400.0mmHg Capacity: 2.5A-Hr, 150 NIBP Simulations
Charge Time: 16 Hours
BP Presets: Over-Pressure:
Systolic/Diastolic (MAP): Automatically tests operation of the Weight/Dimensions:
Adult Neonatal monitor's relief valve 8.0 pounds with internal battery
250/190 (212) 200/150 (169) Pop-Off Pressure: 0 to 400 mmHg (3.63 kg)
150/120 (129) 150/100 (119) Instantaneous Pressure: 0 to 400 mmHg 8" Wide x 5" High x 12" Long
120/90 (99) 120/80 (95) (20.3 x 12.7 x 30.5 cm)
100/70 (79) 100/70 (82) Standard Accessories:
80/50 (59) 80/50 (62) Firmware: All specifications subject to change
60/30 (39) 60/30 (42) • Autosequences
without notice.
35/15 (21) • User Defined Envelopes (45)
(Exact Diastolic, Systolic and MAP • Cal Tables™
values are a function of the specific Hardware:
CalTables™. The BP Presets above are • Cuff Supports:
taken from the Welch Allyn Propaq • Plastic and Aluminum Cylinders
Encore CalTable™) • Adult (3.5" OD, 7.25" width)
• Neonate (1.25" OD, 2.25" width)
Accuracy: • Adapter Hoses:
<± 0.5% of selected target value Adapter hoses insert between the
NIBP device, cuff and analyzer.
Oscillometric Pulse: These adapter are compatible with
Pulse Amplitude @ MAP: 2.0 mmHg
oscillometric NIBP monitors.
Pulse Volume Range: 4.4 ml nominal
• Male/Female Luer
(0-150%)
Pulse Rise Time: 80 msec (minimum) • Male/Female Clippard (GE Medical,
Pulse Rate Range: 15-330 bpm Draeger/Siemens)
Pulse Rate Accuracy: ±1 bpm, 15-200 • Colder/CPC (GE Medical, Protocol
bpm; ±2 bpm 201-330 bpm Systems)
• OBAC Quick Release (Philips
Displayed/MeasuredParameters: Medical)
Dynamic NIBP Cuff PressureWaveform • Universal 5/32" I.D. Hose
Measurement Time: 0 to 180 seconds • Pressure Bulb Assembly
Max. Inflation Pressure: 0 to 400 mmHg
• Self system leak test hose
Inflation Time: 0 to 180 sec
• Communication Ports:
Inflation Rate: • 1 RS-232
0 to 999.9 mmHg/sec • 1 USB Port
Minimum Pressure: 0 to 400 mmHg • Other:
Deflation Time: 0 to 180 sec • External Power Supply and Cord
Deflation Rate: -0.0 to -999.9 mmHg/sec • Operation Manual
• Internal Pump
• Internal Battery

5355 Parkwood Place, Richmond, BC Canada V6V 2N1


Phone: 604.291.7747 • Fax: 604.294.2355
Toll-free (North America only): 800.667.6557
Email : customerservice@datrend.com © DATREND SYSTEMS INC. JAN 2013

Accur acy M att er s • www.datr end.com