Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

The particle question

Published: 30-Nov--0001

Design & Build Analysis

Allen Humphries, General Manager of PMT (GB) discusses particle monitoring and the
implications of the GMP Revision to Annex 1

The EC guide to Good Manufacturing Practice (GMP), covering the manufacture of sterile medicinal
products, revision to Annex 1 became operational on 1 September.

The proposed changes have focused attention on particle counting and raised questions
throughout the industry with regard to monitoring techniques and procedures.

This paper seeks to highlight some of the issues and provide information that may be of assistance
to those persons tasked with ensuring compliance.

Key issues

While ISO 14644-1 cleanroom standards are now widely recognised and understood it may not be
immediately apparent that there are major differences between this standard and the EC GMP
guideline figures for acceptable particle population levels, especially for the cleanest grades A and
B. Grades A and B 'at rest' and also Grade A 'in operation' allow no more than 3,500 particles per m3
greater than 0.5 micron and 1 (previously 0) particles per m3 greater than 5.0 microns. This
compares to ISO 14644-1 class 5 limits of 3520 p/m3 > 0.5 micron and 29 p/m3 > 5 microns.

The GMP guideline is, as can be seen, intent in maintaining exceptionally low particle populations
in the larger size range, which in itself poses fundamental questions regarding monitoring
procedures and equipment.

The outgoing guide which required 0 particles/m3 at 5 micron for these classes was statistically
impossible to achieve without sampling for infinity. 1 particle/m3 is achievable, however if the
general rule (as applied in ISO14644-1 and Fed Std 209E) is used whereby a volume of air is
sampled such that if at the maximum particle concentration allowable then a minimum of 20
particle events should be measured, using a conventional 1ft3/min sample flow rate particle
counter then sample times would need to exceed 700 minutes, bearing in mind that there are over
35ft3 in a cubic metre.
It is therefore not surprising that the guideline states that a continuous measurement system
should be used for monitoring the concentration of particles in the grade A zone and is
recommended for the surrounding grade B areas. Continuous is a term which itself has in the past
caused concern with regard to definition, as it has been used to describe both sequencing
manifold systems, which automatically and regularly revisit one of a number of sample locations,
and the more recently developed miniaturised particle sensors, which will monitor and generate
data truly continuously.

Portable instruments

Most commonly used cleanroom particle counters operate at 1ft3/min (28.3 litres/min). For routine
testing GMP requires that at least 1m3 (35.3ft3) of sample is taken for each class zone, the outcome
of this is that additional time is probably going to need to be spent conducting routine testing than
is currently done. From an instrument design stand point, the move to providing ultra high flow rate
instruments is very unlikely as sample flow rate very closely defines both the size and dimensions
of the optical system and internal air pump. In fact, industry needs have driven manufacturers to
produce lightweight, highly portable instruments, based upon the flow rate norm of 1ft3/min, to
fulfill ever more pressing Health and Safety issues relating to lifting and carrying.

For GMP compliance in monitoring (as opposed to routine testing), especially in Grade A and B
areas, portable instruments could be used, however in most manufacturing environments the
number of locations and availability of instruments would preclude this as a workable option.

Continuously monitoring

These small discrete instruments are now widely available and with their increased use across a
wide range of industries, are a feasible, cost effective option. For pharmaceutical applications they
are designed to measure continuously at 0.5 and 5 microns at an airflow rate of 1ft3/min, reporting
data directly to a PC or to an in-house data management system. By running constantly during the
defined periods (i.e. at rest or in-operation) air volume/flow rate is no longer an issue and, if
relatively short data collection periods are set, then episodic events associated with individual
particle bursts may be identified and distinguished from general overall background levels.

As with all data, whether collected via discrete sensors or portable instruments, the loss of larger
(i.e. 5.0 micron plus) particles in sample tubing must be considered. Therefore the measuring
device needs to be as close as practically possible to the sample location with the minimum
length of tubing and bends.

In the absence of recommendations either in the current standards or guidelines, it is not


unreasonable to refer to the old FED-STD-209E which states that for particles in the range 2-10
microns the transit tube should be no longer than 3m (at an airflow rate of 1.0ft3/min).

Sequencing manifold sampling


This type of system was, and in fact still is, widely used for regular automated monitoring. A single
particle counter is linked to a sequencing scanning manifold, which automatically samples up to 32
individual ports. Radiating from the manifold are lengths of special low particle shedding, low
attraction tubing, which may be up to 125ft (38m) in length.

For critical class A/B regions the main concern with this type of monitoring lies in the fact that
large particles will be 'lost' within tubing runs even when the manifold system is linked to a high
flow rate auxiliary vacuum pump, which creates turbulent flow within the sample lines.

Essentially the larger the particle, the longer the run, and the greater the number of bends: the
greater the particle loss. Small particles (< 1.0 micron) remain airborne and transport very efficiently.
The diagram indicates four particle sizes and their transportation efficiency in tubes of up to 125ft
(38m) at an (assisted) airflow velocity of 3ft3/min. For general cleanroom monitoring where
confidence in room stability is important and where a rapid indication of out of specification particle
events is required, the data generated by a manifold system for 0.5 micron and above particles is
very sound.

If we assume that the ISO 14644-1 (or old Fed Std 209E) standard approximates to a typical particle
distribution curve likely to be found in a clean environment, then the ratio of 0.5 to 5.0 micron
particles is in the order of 120:1. Therefore if all larger particles were to be lost then the
accumulative total above 0.5 would be only marginally affected.

The system becomes questionable however, if absolute numbers are required to be measured at
the 5 micron and above size, as tube losses will reduce the total count above this threshold. The
actual magnitude of loss at each point is extremely difficult to determine as there are numerous
factors creating the loss, and in-situ comparison testing is fraught with problems and open to
experimental error.

While the above comments are focused on large particle losses in manifold tubing it is worth
noting that losses will occur in the tubing of the industry standard 1ft3/min portable particle
counters, and therefore if precise counts are required, at larger sizes then the tubing length should
be minimised, or removed entirely. In many cases of course this is simply not practical, however
extending the tubing length past manufacturers recommendation should be avoided otherwise
the user will simply create a filter device.

While changes to guidelines in themselves always create concern and anxiety, the rationale
behind the changes should not be forgotten, nor should the realisation that confidence in the
operation and control within critical manufacturing environments can only be addressed by a
rigorous risk assessment from which will naturally develop a fuller understanding as to the
implications of the changes and whether or not alterations to current testing and monitoring
procedures are required.

Share 

You might also like