Professional Documents
Culture Documents
ARESII User Manual V1.3en
ARESII User Manual V1.3en
ARESII User Manual V1.3en
3
ARES II
Contents
1
ARES II
2.5.5 Adapter for passive multi-conductor cables II ....................... 55
2.5.6 Electrodes .............................................................................. 56
2.5.7 Parallel connection of several ARES II units ......................... 57
3 MEASUREMENT .................................................. 62
2
ARES II
3
ARES II
4
ARES II
7.9 AC / DC converter .................................................................. 173
5
ARES II
1 Safety Rules
During the measurement there is a high voltage on electrodes!
In case of necessity any person can press the red main switch and thus
disconnect the measuring electrodes from the power supply!
6
ARES II
2 Basic Instructions
2.1 Theory
V
a k
I
where ΔV [mV] is measured voltage, I [mA] is current and k is
geometrical factor depending on the individual distance between the
electrodes (fig. 2).
2
k
1 1 1 1
rC1P1 rC 2 P1 rC 2 P 2 rC1P 2
7
ARES II
P1 P2
P1 r C1P2
r C1
rC
C1 2P
2
1
2P
rC
C2
fig. 2 General four-electrode array
4
k
1 1
...
rC1P1 r 'C1' P1
8
ARES II
Several positive and negative current pulses are measured for
elimination of noise and drift (e.g. influence of artificial and telluric
currents, electrode polarization)
The standard deviation of the measurement δ (St-dev) is calculated
as follows:
1 ( R R) 2
100%
R N 1
U
U Uip
i
pulse 12 3 4 5
IP-windows
9
ARES II
10
ARES II
11
ARES II
12
ARES II
13
ARES II
14
ARES II
15
ARES II
16
ARES II
17
ARES II
18
ARES II
Accessories necessary for GPS continual measurement:
19
ARES II
VES, RP and SP
VES, RP and SP measurement passes in a grid of points (profiles,
stations).
You are asked to set P-step, S-step and the position of the first
measured point before beginning of the measurement.
y [m] measured
point
16 15 14 13
9 10 11 12
9
8 7 6 5
P-Step
Profile 1 2 3 4
S-Step
Station
20
ARES II
2D and USD
The position of 2D and USD measurement is determined by the
position of the beginning of the profile and by the direction of the profile.
The direction can be chosen from X, Y, -X and -Y.
Other important parameter is the electrode distance and the profile
length. If we are planning to measure several 2D (resp. USD) profiles to
obtain one pseudo 3D profile (2D, resp. USD profiles will be combined) it
is necessary to choose the same electrode distances for all profiles in X
direction, to choose the same electrode distances for all profiles in Y
direction (may vary from electrode distance in X direction) and also to
choose the beginnings of all profiles to fit uniform grid with electrode
positions. The same rule must be kept if several 2D or USD profiles
measured in one line are to be interpreted in Res2DInv software together
(e.g. the second 2D profile can be a continuation of the first one).
Profiles can be numbered for better orientation.
y [m]
Profile direction
90° (Y)
3
270° (-Y)
2
Measured
profile
Y-position
1
Profile No.
5 6 7 8
X-position x [m]
fig. 13 Position of 2D / USD profile
21
ARES II
Electrode spacing
multicable
section
Measured
profile
Profile length
3D
The position of 3D measurement is determined by the position of
the beginning of the profile.
Other important parameter is the electrode distance in X and Y
direction and the profile length in X and Y direction. If we are planning to
interpret several 3D profiles together in Res3DInv software it is necessary
to choose the same electrode distance in X and Y direction and also to
choose the beginnings of all profiles to fit uniform grid with electrode
positions.
Profiles can be numbered for better orientation.
Electrode spacing X
y [m] multicable
section
Electrode spacing Y
Profile beginning
x [m]
X-position
Profile width
22
ARES II
Boreholes distance
23
ARES II
Borehole-Surface (BS)
The position of Borehole-Surface measurement is determined by
the position of the borehole (i.e. the profile beginning) and by the direction
of the profile. The direction can be chosen from X, Y, -X and -Y.
Other important parameters are borehole depth, surface profile
length, borehole electrode distance and surface electrode distance. If a more
detailed resolution is required (than the distance of borehole multi-electrode
cable outlets allows) it is possible to reach result adequate to smaller
distance of electrodes by the measurement procedure similar to procedure of
roll-along.
Example of more detailed measurement: The cable outlet’s
distance is 5 m and the required resolution is 1 m. Measurement is done
with Cable spacing = 5 m, Electrode spacing = 1 m. First we will hang the
multi-electrode cable to the 1 m depth and do the measurement then we will
hang the multi-electrode cable to the 2 m depth and do the measurement,
and so on up to the last depth 5 m.
Profiles can be numbered for better orientation.
Profile direction
Profile 90° (Y)
X,Y position direction
180° (-X) 0° (X)
Borehole depth
Surface length
24
ARES II
Beginning of the
multi-el. cable
GPS position
(X,Y = 0, 0)
(Electrode spacing)
(Direction of movement)
X
fig. 18 Parameters of GPS cont. measurement (reverse = No)
25
ARES II
RP
Several preprogrammed electrode arrays can be chosen for
resistivity profiling or also new electrode configuration can be defined.
Basic arrays are preprogrammed as single-channel ones, ie. for
measurement of them 4 electrodes are needed. Each array is given by its
type and parameter a, resp. na. The result of the measurement of basic array
is a curve of apparent resistivity and IP on measured profile (1D), or a map
(2D).
26
ARES II
a a a
C1 P1 P2 C2
Station
fig. 19 Wenner Alpha
a a a
C2 C1 P1 P2
Station
a a a
C1 P1 Station C2 P2
fig. 21 Wenner Gamma
na a na
C1 P1 P2 C2
Station
fig. 22 Wenner-Schlumberger
a na a
C2 C1 P1 P2
Station
fig. 23 Dipole-Dipole
27
ARES II
C2 P2
a na
a
Station
C1 P1
fig. 24 Equatorial Dipole-Dipole
na a
Infinite
electrode
C1 P1 P2
Station
C2
(x, y)
fig. 25 Pole-Dipole
a na
Infinite
electrode
P2 P1 C1
Station
C2
(x, y)
Infinite
electrode
a
Infinite
electrode P2
C1 P1 (x, y)
Station
C2
(x, y)
fig. 27 Pole-Pole
28
ARES II
Infinite
electrode
a
Infinite
electrode C2
P1 P2 (x, y)
Station
C1
(x, y)
Profile
a a a a na
Station
Pn P3 P2 P1 C1
1
Infinite
electrode Station
C2
(x, y)
29
ARES II
Pn
C2 P2 (x, y)
(x, y)
(x, y)
30
ARES II
31
ARES II
Other prescriptions for single-channel measurement known from single-
channel device ARES are stored only on the installation CD:
Dip-Dip N4.mcs Pole-Dip N4A1.mcs,
Pole-Pole S1.mcs Schlumberger N6.mcs
32
ARES II
VES measurement
VES measurement is done according to a prescription, which is
loaded into device memory. There are all necessary configurations of
electrodes for measurement of selected method in the prescription. Each
prescription contains a definition for measurement optimized for one-
channel and multi-channel measurement. ARES II is delivered with the
prescription for measurement:
Schlumberger.vss (Wenner-Schlumberger)
a a a a a
last electrode 1st electrode
Pn P3 P2 P1 C1
1
Infinite
electrode
C2
(x, y)
fig. 31 10-ch Pole-Dipole for GPS continual measurement
33
ARES II
34
ARES II
electrode cable, but due to roll-along procedure the measurement is divided
into three parts.
multi-cable
section
The rolled
ME section measured
profile
(rolovaná)
profile
beginning
profile length
ME cable length
rolled part of cable
35
ARES II
ME cable
section
measured
profile
1st
electrode
3D measurement
Current position of the multi-electrode cable is defined as the
distance of its first electrode from the profile beginning in X and Y
directions, and by the direction of installation of the multi-electrode cable
(+x / -x direction).
ME cable
section
measured profile
x
x
1st electrode
36
ARES II
BB measurement
Current position of the multi-electrode cable is defined as the
distance of first electrodes in the 1st borehole and the 2nd borehole from the
surface and by the number of electrodes placed into the 1 st borehole. If the
depth of a borehole is lower than the length of the borehole multi-electrode
cable, the negative distance of the 1st electrode from the surface can be set.
Only electrodes below the surface will be used for measurement
(the electrode with the position = 0 will not be used for measurement).
ARES II
adapter
-2
2
BS measurement
Current position of the multi-electrode cable is defined as the
distance of first electrode in the borehole from the surface, by the distance
of the first electrode on the surface from the borehole and by the number of
electrodes placed into the borehole. If the depth of the borehole is lower
than the length of the borehole multi-electrode cable, the negative distance
of the 1st electrode from the surface can be set.
Only borehole electrodes below the surface will be used for
measurement (the electrode with the position = 0 will not be used for
measurement). The first electrode of the surface multi-electrode cable can
be placed on the wellhead.
37
ARES II
1st electrode
ARES II
adapter
1st electrode
S-min
S-max
38
ARES II
Full Reduced
39
ARES II
S-Min
S-Max.
1
2
3
4
5
6
7
8
9 Measured
10 layers
layers 11
12
13
14
15
16
17 Measured zone
18
19
40
ARES II
It is possible to use several preset values of IP windows.:
In
Ui
pulse 1 2 3 4 5
IP-windows 1~10
41
ARES II
42
ARES II
If the measured potential and the standard deviation are
good, the minimum number of pulses is used for the measurement.
If the measured potential is less than the optimum value or the
standard deviation is higher than the selected maximum value, the
number of pulses is increased.
Setting of this parameter has a major influence on the total
measuring time.
43
ARES II
Stop button
Power output indicator Data transfer
Keyboard computer,
USB flash disc,
GPS
44
ARES II
T-piece AC adapter
45
ARES II
VES adapter
(Only for single-channel device)
46
ARES II
47
ARES II
48
ARES II
The attachable battery pack is useful for the SP, RP and VES
measurements, where the instrument’s easy portability is important. Thus an
undesirable random disconnecting of the supply during measurement can be
avoided. The attachable battery pack is also suitable for shorter 2D/3D
measurements.
fig. 47 AC adapter
49
ARES II
50
ARES II
Lat: 4805.272N
Lon: 01139.258E
Alt: 46.8M
Time: 132440
Satellites: 6
H. Dilution 1.5
Press key Enter if everything is OK. You can also go back to the
screen of setting GPS communication by pressing No key. GPS data
receiving is signalized by label “GPS” in the up right corner of the main
menu. If connection is established it is signalized by short beep, connection
interruption is signalized by long beep.
After establishing of the connection with GPS receiver the window
with device time and with the time from GPS receiver appears, if no
measurement or another operation is not just processed. Press Enter to set
device time according to GPS receiver time, or press No to cancel
operation.
The GPS receiver can be connected using the connector on the
remote control set if the ARES is operated via RS232 or via radio (see the
picture below).
51
ARES II
52
ARES II
C1 C2
P1 P2
pair 1
pair 2
pair 3
pair 4
pair 5
Profile Profile
beginning end
fig. 54 Connecting of multi-electrode cable
53
ARES II
channel multi-electrode cable and single channel multi-electrode cable
cannot be combined. A special single-channel T-piece, which can be
connected directly to the device, is delivered with single-channel device.
Profile
beginning Profile
end
54
ARES II
Profile Profile
beginning end
55
ARES II
2.5.6 Electrodes
Stainless steel electrodes should be hammered into the ground deep
enough to make the electrical resistance between ground and electrode as
low as possible. To decrease the resistance it is possible to water soil around
electrodes or to use more electrodes, which are placed next to each other
(perpendicular to the profile in the distance from 25 to 50 cm) and
connected by short cable. Achieving low grounding resistance is important
for obtaining high quality data especially in IP measurement or in very deep
measurement (measuring depth over 100 meters).
Use O-rings for attaching of multi-electrode cable box.
56
ARES II
Each ARES II unit must have its own power supply. All used
power supplies must be of the same type – either the electronic converters
or the car batteries (12 V and the same capacity). For VES and RP
measurement you can power two ARES II units from one car battery with
the capacity at least 60 Ah.
For maximum power the car batteries can be connected in parallel
by means of strong cables with clamps. Never connect electronic converters
together!
Connect each ARES II unit to T-piece and interconnect all T-
pieces. Then connect measuring accessories to the outside T-pieces (as
usual for single ARES II). If you need to connect electrodes for VES, RP or
infinite, connect cables to the plugs on the T-piece in the middle.
57
ARES II
2.6.1 Switching ON
The instrument is switched on by pressing the green Yes key. At
first the logo, serial number and firmware version are displayed. A moment
afterward the main menu is displayed. There is also battery status and time
shown in the right upper corner of the display.
58
ARES II
GPS batt:100% 13:55:54
New measurement
Continue measurement
File tools
Show Info
Settings
Press Enter key to run main menu, otherwise the unit will
automatically turn off (after 30 seconds).
59
ARES II
Switching off... _
Device is set
to automatically start.
Next start:
hh:mm:ss dd/mm/yyyy
20:05:00 01/11/2012
60
ARES II
61
ARES II
3 Measurement
3.1 New Measurement Creation
Choose New measurement from the main menu. Choose proper
measurement method from the next menu. Some measurement methods
may not be offered according to the type of the device.
62
ARES II
The following shows the input of all parameters for the
measurement. For each measurement it's a different set of parameters.
After entering of all the parameters the file for storage of measured
data is created in the memory of the device. File generation may take a
while but it is possible to interrupt it by pressing the No key. After creating
the file for 2D, 3D, BB, BS, USD and GPS measurements a summary table
with some data about the measurement is displayed.
529 measurements
1348 points
27 kB
_ _
63
ARES II
Estimated measurement time – Estimated time of measurement
of the whole profile. Considered as an ideal time to measure the
whole profile at once, without roll-along or other interruption.
A time for each separate set of measured points is displayed for
time monitoring measurement; a time for one cycle of
measurement is displayed for GPS continuous measurement
with adapter.
Note: Only data for one cycle measurement are created in the data
file for GPS continual measurement with adapter for passive multi-
conductor cables. This data will serve as measuring prescription for
individual measuring cycles during the measurement
When the file for time monitoring measurement is created you are
asked whether you want the device to Switch off now. Confirm it by Enter
key or cancel it by No key. To automatically start the measurement the
device must be turned off.
When the file for manual measurement is created, the question
whether the measurement is to be started right now is displayed. Confirm it
by Enter key or cancel it by No key. The measurement file is already stored
in memory and its measurement can be started at any time using Continue
measurement in the main menu.
64
ARES II
65
ARES II
Electrode spacing: 5m
Profile length: 235 m
multiple channel
66
ARES II
S-min: 0m
S-max: 200 m
full
67
ARES II
Yes
C2 250 975 m
68
ARES II
Potential: 200 mV
Stacking min: 4
Stacking max: 8
St. dev. max: 2%
Output power: 850 W
69
ARES II
70
ARES II
IP – type of IP measurement (None / Newmont’s st. /
Full decay 200 ms / Full decay 500 ms / Full decay 1000 ms /
Full decay 2500 ms / User defined IP windows)
IP windows length – only if User defined IP windows are
measured. By setting of zero length IP window the
editing of the parameter is done, only non-zero length
windows will be measured.
Do you want to use positions of infinite electrodes to calculate
exact geometric factor? – yes for calculation of the exact
geometric factor, no for calculation of the approximate
geometric factor – only if any infinite electrode is used for
measurement.
Inf. electrode positions – If exact geometric factor is
selected to be calculated the x, y position of infinite
electrodes is set. The position is relative to the same
point as the position of the beginning of the profile x, y.
Potential – the required level of potential to be measured
Stacking max – maximum count of pulses on one measured point
Stacking min – minimum count of pulses on one measured point
St. dev. max – maximum standard deviation that is still accepted
Output power – maximum output power
71
ARES II
72
ARES II
S-min(y) – minimum distance between the first and the last used
electrodes of the measuring array in the y direction in meters
S-max(y) – maximum distance between the first and the last used
electrodes of the measuring array in the y direction in meters
Measurement density – (full, reduced)
Pulse length – current pulse length (0.2 – 30 s)
IP – type of IP measurement (None / Newmont’s st. /
Full decay 200 ms / Full decay 500 ms / Full decay 1000 ms /
Full decay 2500 ms / User defined IP windows)
IP windows length – only if User defined IP windows are
measured. By setting of zero length IP window the
editing of the parameter is done, only non-zero length
windows will be measured.
Do you want to use positions of infinite electrodes to calculate
exact geometric factor? – yes for calculation of the exact
geometric factor, no for calculation of the approximate
geometric factor – only if any infinite electrode is used for
measurement.
Inf. electrode positions – If exact geometric factor is
selected to be calculated the x, y position of infinite
electrodes is set. The position is relative to the same
point as the position of the beginning of the profile x, y.
Potential – the required level of potential to be measured
Stacking max – maximum count of pulses on one measured point
Stacking min – minimum count of pulses on one measured point
St. dev. max – maximum standard deviation that is still accepted
Output power – maximum output power
73
ARES II
74
ARES II
Inf. electrode positions – If exact geometric factor is
selected the x, y positions have to be set. The x, y
positions are related to the same coordinate system as
the position of the measured profile.
Potential – the required level of potential to be measured
Stacking max – maximum count of pulses on one measured point
Stacking min – minimum count of pulses on one measured point
St. dev. max – maximum standard deviation that is still accepted
Output power – maximum output power
75
ARES II
76
ARES II
IP windows length – only if User defined IP windows are
measured. By setting of zero length IP window the
editing of the parameter is done, only non-zero length
windows will be measured.
Do you want to use positions of infinite electrodes to calculate
exact geometric factor? – yes for calculation of the exact
geometric factor, no for calculation of the approximate
geometric factor – only if any infinite electrode is used for
measurement.
Inf. electrode positions – If exact geometric factor is
selected to be calculated the x, y position of infinite
electrodes is set. The position is relative to the same
point as the position of the beginning of the profile x, y.
Potential – the required level of potential to be measured
Stacking max – maximum count of pulses on one measured point
Stacking min – minimum count of pulses on one measured point
St. dev. max – maximum standard deviation that is still accepted
Output power – maximum output power
77
ARES II
78
ARES II
Measurement mode – the mode of the measurement. Choose the
single-channel mode, if the single-channel multi-electrode
cable will be used for the measurement or if only single-
channel measurement is to be performed. Choose multiple
channel - up to 5 / multiple channel - up to 10 mode if multi-
channel measurement will be done with five-channel or ten-
channel multi-electrode cable. The measurement mode may not
be editable according to the type of the device.
79
ARES II
80
ARES II
Measurement mode – the mode of the measurement. Choose the
single-channel mode, if the single-channel multi-electrode
cable will be used for the measurement or if only single-
channel measurement is to be performed. Choose multiple
channel - up to 5 / multiple channel - up to 10 mode if multi-
channel measurement will be done with five-channel or ten-
channel multi-electrode cable. The measurement mode may not
be editable according to the type of the device.
81
ARES II
82
ARES II
Pulse length – current pulse length (0.2 – 30 s)
IP – type of IP measurement (None / Newmont’s st. /
Full decay 200 ms / Full decay 500 ms / Full decay 1000 ms /
Full decay 2500 ms / User defined IP windows)
IP windows length – only if User defined IP windows are
measured. By setting of zero length IP window the
editing of the parameter is done, only non-zero length
windows will be measured.
Do you want to use positions of infinite electrodes to calculate
exact geometric factor? – yes for calculation of the exact
geometric factor, no for calculation of the approximate
geometric factor – only if any infinite electrode is used for
measurement.
Inf. electrode positions – If exact geometric factor is
selected to be calculated the x, y position of infinite
electrodes is set. The position is relative to the same
point as the position of the ship with ARES II and GPS.
Potential – the required level of potential to be measured
Stacking max – maximum count of pulses on one measured point
Stacking min – minimum count of pulses on one measured point
St. dev. max – maximum standard deviation that is still accepted
Output power – maximum output power
83
ARES II
The information about the file will appear (file name, location
name, type of measurement, date and time of the measurement, note,
measurement method, file size, total number of points to be measured /
number of already measured points). Confirm it by Enter key.
84
ARES II
Name: Brno
Locality: Medlanky
Type: 2D-Multicable
Date: 19/09/2012 16:54:29
Note:
Method: Schlumberger HD
Size: 14524 B
Points/done: 276 / 0 _
Is it OK?
Measure now?
85
ARES II
3.3 Measurement
3.3.1 Resistivity profiling (RP) and self-potential (SP)
If the selected electrode array uses infinite electrodes the device
prompts you to connect them. Connect it and press Enter key.
y [m]
measured
point
Profile
x [m]
Station
fig. 89 Specification of the measured point position
86
ARES II
Profile: 15 m
Station: -258 m
Note:
Start measurement?
87
ARES II
The measurement is signalized by the message MEASURE on the
display. The current pulse is indicated by the shining of the red LED
(MEASURING) on the device panel.
When the measurement is finished the measured values (injected
current and potential, measured voltage, apparent resistivity, standard
deviation, measured point number, Self-potential – voltage offset on
potential electrodes, IP values with related standard deviations) are
displayed on one of the measured values screens.
The displayed measured values can be either deleted using the No
key (and then you return to the screen of the measured point position) or
they can be saved by pressing the Enter key (and the screen of the position
of the next measured point is displayed). The keys 1 to 8 serve for the
movement between the measured values screens.
Potential: 200 mV
Stacking min: 4
Stacking max: 8
St. dev. max: 2%
Output power: 850 W
88
ARES II
The screen with position of the VES appears. Move to the point
with shown coordinates (profile, station).
The coordinates as well as Note can by edited after pressing No
key.
y [m]
measured
VES
Profile
x [m]
Station
fig. 94 Specification of the VES position
89
ARES II
Profile: 15 m
Station: -258 m
Note:
On the next screen the information about electrode positions and also
about used potential pair of VES adapter (if VES adapter is used for
measurement) will be displayed.
C1: 61.50blue m
C2: -61.50blue m
P1: 1.000blue m
P2: -1.000blue m
P3: 3.150red m
P4: -3.150red m
_______________________________
Are positions right?
90
ARES II
C1: 61.50blue m
C2: -61.50blue m
P1: 1.000blue m
P2: -1.000blue m
VESA channel: 2
_______________________________
Are positions right?
If you wish to skip the actual spacing and to move to the next (or
preview) one, press the down or up arrow key. You can change (set) the
actual position of the electrodes after pressing the No key or finish this VES
measurement by another pressing of No key if necessary.
If the electrodes are placed in their positions you can continue by
pressing of Enter key.
Start measurement?
91
ARES II
potential electrodes, IP values with related standard deviations) are
displayed on one of the measured values screens.
Potential: 200 mV
Stacking min: 4
Stacking max: 8
St. dev. max: 2%
Output power: 850 W
92
ARES II
Connected electrodes: 24
Required (by S-max): 36
Connected electrodes: 24
Laying direction: X
X Pos. of 1st el.: 24 m
Y Pos. of 1st el.: 4m
93
ARES II
Connected electrodes: 24
Count of electrodes placed
into borehole 12
1st depth in B: 2m
1st pos. in S: 0m
Connected electrodes: 24
Count of electrodes placed
into borehole 1: 12
1st depth in B1: 2m
1st depth in B2: 2m
94
ARES II
Measurement status _
Rem. measurements: 125
Remaining points: 378
Done measurements: 15
Done points: 22
Estimated rem. T: 7 min
Temperature: 29°C
Battery status: 55 %
95
ARES II
number of electrode within multi-electrode cable is shown after entering the
position.
By entering the position -1 (resp. Exit) the editing of dummy
electrodes is finished and screen with question Start measurement? is
displayed.
6 3
-1
10 4 3
-1
Boreh.1 4 8
Exit
96
ARES II
Set max. 100 dummy positions
boreh./surf., depth/pos. [m] _
Boreh. 2 10
Surf. 0 11
Exit
Start measurement?
Unlock the STOP button for start measurement or press No key for
end of the measurement. Before the measurement is started the resistance of
the electrodes is tested. The test can be interrupted (and skipped) by holding
down the No key. To properly test of electrodes no infinite electrodes
should be connected to the device.
_______________________________
Push STOP before touching el.!
97
ARES II
_______________________________
Push STOP before touching el.!
_______________________________
Push STOP before touching el.!
_______________________________
Push STOP before touching el.!
98
ARES II
If you would like to check the connection of the electrodes
found by the test press the STOP button at first.
By selecting any of the electrode position shown in the list and
pressing the Enter key the contact resistance is re-measured. Pressing No
key cancels the test of electrode resistance and returns to editing of dummy
electrodes. Select the item All are OK, start meas. to end the electrode test
and to continue to measure data on the profile.
When more than 100 electrodes of multi-electrode cable is
connected the resistance test is carried out in steps of hundred electrodes.
99
ARES II
Press the Enter key when the section is moved. The multi-
electrode cable position window will be shown. The measurement can be
finished by pressing the No key.
100
ARES II
Potential: 200 mV
Stacking min: 4
Stacking max: 8
St. dev. max: 2%
Output power: 850 W
101
ARES II
For data viewing and their deleting use following buttons:
1 - 8 change the type of display with measured values so you can
display all information about the selected data point. The cursor is
automatically set to the selected data point on the new screen.
select + 0 - 9 allows filtering displaying of multi-channel
measurement if displaying of all channels is unclear. Only the
selected channel will be shown. Repeated pressing of the button
switches the filter off.
select + arrow up / down on the screen with measured layers
selects another layer to be displayed (it decreases / increases the
distance between the electrodes arrangement of which the
measured layer is shown).
select + arrow left / right on the screen with measured VES selects
another VES to be displayed (it decreases / increases the position
of the electrode arrays of which the measured VES is shown).
5 on the screen No. 5 with measured layers changes the scale of the
horizontal axis (smaller part of the profile is displayed). To return
to the default setting (the whole profile is shown) hold the button 5
down.
Yes or meas deletes the selected data point. You will be asked to
confirm the operation. A deleted data will be re-measured
immediately after measurement recovery. The deleted value is not
deleted absolutely, but it is only marked as “remeasured”, so there
are both old and new values in the data file (old value will not be
used for data export).
No exits the data editing and resumes the measurement.
fig. 117 LAYER and VES curve of the selected data point
102
ARES II
What is shown on LAYER and VES graphs (screens No. 5 and 6):
LAYER is a graph, in which there are drawn all data points with
the same distance between the first and the last electrode used for
data points measurement. Some measurement prescriptions define
more layers with the same distance of outer electrodes – those are
drawn into one graph so it is possible to compare measured values.
VES is a graph, in which there are drawn all data points with the
same position of the middle between the first and the last electrode
used for their measurement.
MC out of range means that the selected data point of the LAYER
or VES is out of current multi-electrode cable range (the data
cannot be re-measured without roll-along of the multi-electrode
cable).
spac means the distance between the first and the last electrode
used for selected data point measurement.
el.pos means the middle position of the electrodes used for selected
data point measurement.
exclamation mark warns about the data point with zero apparent
resistivity (probably measured potential was reverse polarity then
expected so the apparent resistivity would be calculated negative).
The data point drawn only with the outline with a blank center
signalizes the last measured data.
103
ARES II
Connected electrodes: 24
104
ARES II
Measurement status _
1 cycle measurements: 125
1 cycle points: 378
Done measurements: 15
Done points: 22
Estimated cycle T: 7 min
Temperature: 29°C
Battery status: 55 %
105
ARES II
If the selected measurement method uses infinite electrodes the
device prompts you to connect them. Connect it and press Enter key.
If the screen for test of GPS connection shows up, unlock the
STOP button (resp. press the Enter key if the button is already unlocked) to
start measurement. The No key opens the window for measurement
parameters editing.
Lat: 4805.272N
Lon: 01139.258E
Alt: 46.8M
Time: 132440
Satellites: 6
H. Dilution 1.5
Start measurement?
(Unlock STOP button)
106
ARES II
Potential: 200 mV
Stacking min: 4
Stacking max: 8
St. dev. max: 2%
Output power: 850 W
107
ARES II
Potential: 49.53 mV
Potential: 49.53 mV
App. res.: 14.35 m
Output potential: 964.0 V
Current: 12.99 mA
Standard dev.: 0.1 %
Self potential: 0 mV
IP1[%]: 1.03% St.dev.: 1.0 %
IP2[%]: 0.91% St.dev.: 1.3 %
IP3[%]: 0.82% St.dev.: 1.8 %
IP4[%]: 0.68% St.dev.: 1.9 %
108
ARES II
Measurement status _
Done measurements: 15
Temperature: 29°C
Battery status: 55 %
fig. 126 Measurement status window (No. 2) (SP, RP, VES, GPS)
Measurement status _
Rem. measurements: 125
Remaining points: 378
Done measurements: 15
Done points: 22
Estimated rem. T: 7 min
Temperature: 29°C
Battery status: 55 %
Battery under load: 43 %
fig. 127 Measurement status window (No. 2) (2D, 3D, USD, BB, BS)
Note: If the Battery under load is displayed with its value much
less than Battery status it means the battery voltage drops significantly
when the load is applied and the battery is at the end of its life.
109
ARES II
1.09
R4: R5: 1.12
1.07R6:1.1% 1.1%
0.95 0.93
1.2% 1.30.96
0.97 0.83
1.4 %
%0.85 1.4 % 2.20.84
0.88 0.66
2.3 %
%0.70 2.2 % 2.50.68
0.70 %3.1 %2.4 %
1.08
R2: R3: 0.9%1.0%0.941.15 0.84
1.0 %1.1 % 0.69
1.8 %1.9 % 2.1 %2.2 %
1.07
R5: R6: 1.11
1.12R7:1.1% 1.2%
0.93 0.97
0.7% 1.40.98
0.96 0.88
1.4 %
%0.83 0.9 % 2.30.89
0.84 0.70
2.2 %
%0.66 1.8 % 3.10.72
0.68 %2.4 %2.5 %
1.09
R3: R4: 1.0%1.1%0.95 0.85
1.1 %1.3 % 0.70
1.9 %2.2 % 2.2 %2.5 %
1.11
R7: 1.15
R8: 1.18
0.7%0.96
0.3% 1.41.05
0.98 0.84
0.9 %0.6 % 2.20.97
0.89 0.68
1.8 %1.5 % 2.40.80
0.72 %2.5 %2.1 %
R5: 1.12
R6: 1.1%1.2%0.97 1.4 % %0.88
R4: 1.07 1.1% 0.93 1.3 % 0.83 2.2 % 0.66 2.5 % 2.3 % %0.70 3.1 %
1.15
R8: 1.18
R9: 1.25
0.3%0.98 1.05
0.8% 1.09 0.89
0.6 % 0.97
1.0 % 1.01 0.72
1.5 % 0.80
1.9 % 0.85
%2.1 %2.7 %
1.121.11
R5: R6:
R7:
1.1%1.2%0.970.96
0.7%
0.880.84
1.4 %1.4 %
0.9 %
0.700.68
1.8
2.3 %2.2 %
%
3.1 %2.4 %
2.5
1.18
R9:
1.15
R7: R8:
1.250.98
0.7%0.3%
0.8%1.05 1.09 0.89
0.9 %0.6 %
1.0 %0.971.01 0.72
1.9 %0.80
1.8 %1.5 %
0.85 2.7 %
2.5 %2.1 %
1.181.25
R8: R9: 0.3%0.8%1.051.09 0.971.01
0.6 %1.0 % 0.800.85
1.5 %1.9 % 2.1 %2.7 %
R9: 1.25 0.8% 1.09 1.0 % 1.01 1.9 % 0.85 2.7 %
110
ARES II
Note:
LAYER is a graph, in which there are drawn all data points with
the same distance between the first and the last electrode used for
data points measurement. Some measurement prescriptions define
more layers with the same distance of outer electrodes – those are
drawn into one graph so it is possible to compare measured values.
VES is a graph, in which there are drawn all data points with the
same position of the middle between the first and the last electrode
used for their measurement.
exclamation mark warns about the data point with zero apparent
resistivity (probably measured potential was reverse polarity then
expected so the apparent resistivity would be calculated negative).
The data point drawn only with the outline with a blank center
signalizes the last measured data.
111
ARES II
112
ARES II
Note: Voltage and current curves are for reference only. Dotted
line shows the waveforms as they are measured. Full line shows the
waveforms after simple deducting the DC offset.
113
ARES II
Battery status: 65 %
Temperature: 23 °C
114
ARES II
115
ARES II
The information about the file will appear (file name, location
name, type of measurement, date and time of the measurement, note,
measurement method, file size, total number of points to be measured /
number of already measured points).
Confirm it by Enter key.
Name: Brno
Locality: Medlanky
Type: 2D-Multicable
Date: 19/09/2012 16:54:29
Note:
Method: Schlumberger HD
Size: 14524 B
Points/done: 276 / 0 _
Edit this file?
fig. 142 LAYER and VES curve of the selected data point
116
ARES II
For operations in data and their deleting use following buttons:
117
ARES II
The information about the file will appear (file name, location
name, type of measurement, date and time of the measurement, note,
measurement method, file size, total number of points to be measured /
number of already measured points).
Deleting of very large files may take a while. Estimated time
needed to delete the file is shown on the last line.
Confirm it by Enter key.
Name: Brno
Locality: Medlanky
Type: 2D-Multicable
Date: 19/09/2012 16:54:29
Note:
Method: Schlumberger HD
Size: 14524 B
Points/done: 276 / 0 _
Are you sure to delete the file?
Estimated deleting time: 0.2 s
118
ARES II
Number of files: 66
Total size: 2420300 B
119
ARES II
Connect your USB flash memory to the device via USB pc / flash
disc communication cable. Choose Save files to USB Flash from File tools
menu. The following window will be shown:
ARES/DATA/2012> _
..
BRNO
PRICELIST
MEDLANKY
POLE
EXPORT
Select path where new project folder will be created. Use arrows
up, down, right and left to through folders on the disc to enter directories or
to exit directories. The current path is displayed above the line. Use Enter
key to confirm selected path. Then the next screen is shown.
120
ARES II
ARES/DATA/2012> _
Enter name of new folder. After you confirm the folder name the
data saving process starts. At the end of the process you are informed about
amount of stored data.
121
ARES II
122
ARES II
4.3 Settings
Choose Settings from the main menu. The following menu will be
shown.
batt:100% 13:55:54
LCD contrast
Filtering
GPS comm speed
Date / Time
Methods priority
Power boosting
123
ARES II
In brackets is IP window
length coefficient
hh:mm:ss dd/mm/yyyy
124
ARES II
125
ARES II
126
ARES II
The program is designed for:
o Downloading data from the ARES II memory to a PC and
to export data into files open-able in a variety of programs
for data processing.
o Converting binary files (files stored directly on the USB
flash drive by the ARES II and files from remote
measurement) to ASCII form.
o Uploading of measuring methods or new firmware into
the ARES II memory.
o Remote control of the device – performing the
measurement without operator assistance at the measuring
spot (e.g. time-lapse investigations on difficult or
dangerous places).
127
ARES II
- Right upper panel shows the files stored in the selected directory.
File name and it type is displayed. You can select one or more files
to work with them. The context menu (mouse right click) as well
as File menu offers a number of functions:
o Rename file – the file can also be renamed after the
second click of the left mouse button.
o Delete file – the file can also be deleted by pressing the
Delete key on the keyboard.
o Edit file – opens the selected file in Windows Notepad. To
start editing the file you can also double-click the mouse
on selected file.
o Export – exports data from the measured data file to
a different format depending on the selected interpretation
software.
o Convert .bin to ASCII – convert binary files to ASCII
readable form. Binary file is, for example, a file stored
directly to the USB flash drive by the ARES II.
o Upload to ARES II – uploads selected measuring methods
to the device memory if that is connected. Method files
can also be uploaded simply by dragging the file to the
lower panel.
o Download all files – downloads all the files from the
device memory if that is connected.
o Create new measurement – create new binary file that
contains all information about the measurement. The
binary file can be used for measurement using remote
control.
o Continue measurement – open stored binary file and
continue in its measurement.
- Right lower panel shows ARES II memory contents. For its view it
is necessary to connect ARES II to the computer, turn it on (the
display will show PC Control) and then click Connect to methods
memory (displays measuring methods) or Connect to data memory
(displays measured data). The name of each file, its type, locality
name, date and time of measurement and the file length is
displayed. All files can be sorted by the displayed data by clicking
on the column title. The context menu (mouse right click) as well
as File menu offers the following functions:
128
ARES II
o Delete file – the file can also be deleted by pressing the
Delete key on the keyboard.
o Download file - downloads selected files from the device
memory. Files can also be downloaded simply by
dragging the selected file to the upper panel.
o Download all files – downloads all the files from the
device memory.
129
ARES II
130
ARES II
131
ARES II
Grounding resistances – In the case of measurement with multi-
electrode cable are measurement parameters followed by
measured electrode grounding resistances.
Example: Connected MC 12+24 electrodes, layed from 0 m,
resistances[Ohm] of 36 el. from el.No 0: -, 196, 190,
193, 241, 222, 238, 222, 231, 188, 218, 219, 203, 240,
264, 214, 231, 251, 235, 269, 198, 188, 212, 168, 153,
140, 122, 115, 114, 113, 101, 89, 84, 83, 76, 84
Example description: Connected multi-electrode cable –
12 electrodes on the left and 24 electrodes on the right
from the T-piece, the first electrode position is 0 m,
resistances of 36 electrodes beginning by the electrode
No. 0 are: -, 196 , 190 , 193 ...(electrode No. 0
was set as Dummy).
132
ARES II
Array – array code
Uout[V] – transmitter output potential
I[mA] – transmitter output current
U[mV] – receiver input potential
EP[mV] – receiver input DC potential (polarization)
AppRes[Ohmm] – calculated apparent resistance
St-dev[%] – standard deviation of apparent resistance
IP1[%] .. IP20[%] – measured polarization in individual IP
windows
IP1-dev .. IP20-dev – standard deviation of measured
polarization in individual IP windows [%]
Note – text note connected with measured point
Lat, Lon, Alt, Time, Satellites, Dilution – values obtained
from GPS receiver
Northing[m], Easting[m], Track[m] – UTM coordinates and
walked distance from the profile beginning – only GPS
measurement
133
ARES II
134
ARES II
ARES-II
GF Instruments
SETUP mode v1.1
PC Control
Press X key to exit
135
ARES II
The names of the data files from which will be exported are shown
in the upper part of the window. Below is shown the name of the file that
will be created. File name as well as its location can be changed by clicking
on the button on the left.
136
ARES II
137
ARES II
IP1
IP2
IP3
IP4 IP redefine – mean value
IP5
IP6
IP...
138
ARES II
Export into Res2DInv and Res3DInv software allows also setting
a simple filter to filter out unsuitable data:
o Method – select measuring electrode array to be exported.
o Minimal injected current – minimal transmitted current
o Minimal measured potential
o Resistivity range – the range of apparent resistivity to be
exported
o Maximal error – maximum allowable standard deviation
to be exported (separately for resistivity and
chargeability)
o IP range – the range of apparent chargeability to be
exported
o Density – only for BB a BS measurement, the parameter
sets the number of exported values between two
electrodes for softening results during followed data
processing. Higher number causes finer grid of the final
picture but the time of the inversion computation is
higher.
o Time monitoring – If you are exporting several identical
2D measurements (time monitoring) to Res2DInv,
checking this option causes to export them to a format
allowing simultaneous processing (see instructions for
Res2DInv, Time Lapse Inversion)
139
ARES II
When exporting GPS measurements to Res2DInv the distance from
the beginning of the profile (according to GPS coordinates) rounded to an
integer multiple of electrode configuration parameter a is used as distance
on the profile.
140
ARES II
141
ARES II
Go to PC control mode
Configure RS232 module
Connect the RS232 module to the PC. The module adds new COM
port. Its number can be found in the device manager.
Select proper COM port and proper communication speed in the
software ARES II Control Interface in the menu Options /
Connection port.
Options / Connection port / Communication test can be used for
communication checking.
142
ARES II
Go to PC control mode
Configure radio module
Transmit power
View my MAC address
Set destination MAC address
143
ARES II
144
ARES II
145
ARES II
The measurement can be run when the new file is saved: You can
confirm the question Start the measure now? or you can double click any
binary file shown in the upper right window or also you can select Continue
measurement from menu.
The measurement is done in similar way as it would be measured
directly on the ARES II unit.
146
ARES II
147
ARES II
148
ARES II
149
ARES II
150
ARES II
a a a
C1 P1 P2 C2
Station
fig. 170 Wenner Alpha
a a a
C2 C1 P1 P2
Station
a a a
C1 P1 Station C2 P2
fig. 172 Wenner Gamma
151
ARES II
na a na
C1 P1 P2 C2
Station
fig. 173 Wenner-Schlumberger
a na a
C2 C1 P1 P2
Station
fig. 174 Dipole-Dipole
C2 P2
a na
a
Station
C1 P1
fig. 175 Equatorial Dipole-Dipole
na a
Infinite
electrode
C1 P1 P2
Station
C2
(x, y)
fig. 176 Pole-Dipole
152
ARES II
a na
Infinite
electrode
P2 P1 C1
Station
C2
(x, y)
Infinite
electrode
a
Infinite
electrode P2
C1 P1 (x, y)
Station
C2
(x, y)
fig. 178 Pole-Pole
Infinite
electrode
a
Infinite
electrode C2
P1 P2 (x, y)
Station
C1
(x, y)
153
ARES II
154
ARES II
6.1.1 Example of *.vss prescription
10 channels
-1R 1R -0.2R 0.2R - - - - - - - - - WS 0; receiver 0
-6.15B 6.15B -0.2R 0.2R -1B 1B - - - - - - - WS 02; receiver 0+2
-29.1B 29.1B - - -1B 1B - - -5R 5R - - - WS 26; receiver 2+6
5 channels
-1R 1R -0.2R 0.2R - - - - - - - - - WS 0; receiver 0
-6.15B 6.15B -0.2R 0.2R -1B 1B - - - - - - - WS 02; receiver 0+2
-29.1B 29.1B - - -1B 1B -5R 5R - - - - - WS 24; receiver 2+4
1 channel
-1R 1R -0.2R 0.2R WS 0
-1.3B 1.3B -0.2R 0.2R WS 0
-4.74R 4.74R -1B 1B WS 1
-22.49R 22.49R -5R 5R WS 2
155
ARES II
156
ARES II
6.2.1 Example of *.mcs prescription
Pole-Dipole with overlapped layers ; method description
C2 ; list of infinite electrodes
0
0
10 channels
0 inf 1 2 3 4 5 6 7 8 9 10 - PD 012345678
0 inf 2 4 6 8 10 12 3 5 7 9 11 PD 012346789
0 inf 5 8 11 14 6 9 12 7 10 13 16 PD 01245789
0 inf 11 16 21 12 17 22 13 18 15 20 25 PD 0134689
0*1 inf 16 23 30 17 24 31 19 26 21 28 35 PD 0134689
0*4 inf 53 76 99 58 81 104 64 87 70 93 116 PD 0134689
5 channels
0 inf 1 2 3 4 5 6 - - - - - PD 01234
0 inf 7 8 10 12 15 19 - - - - - PD 01234
0*1 inf 16 23 30 17 24 31 - - - - - PD 0134
1 channel
0 inf 1 2 PD
0 inf 2 3 PD
0 inf 3 4 PD
0*1 inf 12 17 PD
0*1 inf 13 18 PD
0*2 inf 24 31 PD
157
ARES II
158
ARES II
Ten-channel part of the measurement prescription contains
distances of 13 electrodes (‘a’ for the first borehole or ‘b’ for the second
borehole + electrode number of C1, C2, P1, ..., P11), array code and
numbers of used receivers (0..9, individual numbers without spaces).
Five-channel part of the measurement prescription has the same
structure as ten-channel part, but it does not use electrodes P7 – P11 and
receivers 5 – 9.
Single-channel part of the measurement prescription contains
distances of 4 electrodes (‘a’ for the first borehole or ‘b’ for the second
borehole + electrode number of C1, C2, P1, P2) and array code.
159
ARES II
6.5.1 Example of *.bss prescription
C1C2-P1P2 Dipole-Dipole measurement ; method description
0 ; list of infinite electrodes
20000 ; geometric factor limit
inf ; allowed shift of 2nd borehole electrodes
10 channels
a0 a1 b0 b1 b5 b21 b85 - - - - - - GEN 0123
a0 a4 b0 b4 b20 b84 - - - - - - - GEN 012
a0 a16 b0 b16 b80 - - - - - - - - GEN 01
a0 a64 b0 b64 - - - - - - - - - GEN 0
5 channels
a0 a1 b0 b1 b5 b21 b85 - - - - - - GEN 0123
a0 a4 b0 b4 b20 b84 - - - - - - - GEN 012
a0 a16 b0 b16 b80 - - - - - - - - GEN 01
a0 a64 b0 b64 - - - - - - - - - GEN 0
1 channel
a0 a1 b0 b1 GEN
a0 a4 b0 b4 GEN
a0 a16 b0 b16 GEN
a0 a64 b0 b64 GEN
160
ARES II
T-piece AC adapter
161
ARES II
Standard version:
fig. 181 MCS5 Active single-channel cable section with 8 electrode boxes and 5 m spacing
162
ARES II
fig. 182 ME II/5-5 Active five-channel cable section with 8 electrode boxes, 5 m spacing
and section switch box
fig. 183 ME II/10-5 Active ten-channel cable section with 12 electrode boxes, 5 m spacing
on the reel with section switch box
163
ARES II
ME II/10-5 is active ten-channel cable on the reel serving for the
most exacting tasks like deep survey to several hundreds of meters and
advanced IP measurements. Its rugged design is typical with high resistance
against mechanical damage during transport and work.
164
ARES II
Clamp
165
ARES II
Standard version:
MCC5: Cable reel with 12 electrode boxes with 5 m spacing
MCC2: Cable reel with 12 electrode boxes with 2 m spacing
166
ARES II
The gray cable of the adapter as well as the connector on the right
side of the adapter serve for connection of 24 (optionally up to 36)
conductor cable.
The measuring multi-conductor cable is connected to the adapter
so that the lower numbers of the take-outs are situated on the left (1-24 for
the 48-channel adapter) the higher numbers of the take-outs are situated on
the right (25-48 for the 48-channel adapter).
167
ARES II
The smaller connector on the right side of the adapter serves for
connection to another adapter. Several adapters can be connected together
in a cascade for increasing the number of electrodes up to 200. In this case
the ARES must be always connected to the first adapter on the profile. For
mutual connection of adapters it is necessary to order the prolongation cable
of a proper length (for maximum used distance of adapters according to
used spacing of electrodes on cables).
1 48 49 96
For the software handling, please, follow the same rules as for
a line with active electrodes.
4 3 1 21 1 1 1
168
ARES II
6 5 14 1 31 2 11 1 1
Turn on only the switches belonging to the cable sections that are
really used – connected to the adapter.
169
ARES II
170
ARES II
171
ARES II
172
ARES II
7.9 AC / DC converter
AC / DC converter with power 1500 W is suitable especially for
very long profile measurement with multi-electrode cables. The converter is
supplied by 115 V or 230 V.
173
ARES II
174
ARES II
RS232
ARES II connection
RS232 module for PC
GPS receiver may be
connected here
175
ARES II
Antenna
176
ARES II
8 Technical Specifications
Transmitter
Power 850 W (adjustable 300 / 600 / 850 W)
Voltage 15 – 1000 V (2000 Vp-p)
Current 5.0 A
Protection full electronic protection
Precision 0.1%
Feeding 12 V / 85 A
Efficiency up to 91%
Powerful passive cooling, working internal temperature < 75 °C
Receiver
Number of channels 10 or 1 according to device type
Input impedance 20 M (higher on request)
Input voltage range 20 V (higher on request)
Mains frequency filtering 50 or 60 Hz
Precision 0.1%
Supported methods
2D/3D/4D – Multi-Electrode Resistivity and IP Tomography
VES – Vertical Electrical Sounding
RP – Resistivity Profiling
SP – Self-Potential
Borehole survey
User defined survey
Moving application with GPS (ten-layers and also more layers)
Supported arrays
Wenner Alpha / Beta / Gamma
Wenner-Schlumberger
Dipole-Dipole
Pole-Dipole
Reverse Pole-Dipole
Pole-Pole
MGM
Equatorial dipole-dipole
User defined configurations
177
ARES II
Measurement – features
Self-adapting control system
Automatic calibration
Automatic checking of measured values
Easy interruption of the measurement
Capability of profile prolongation by means of multi-electrode
cable rolling
Capability of borehole measurement resolution increase by means
of multi-electrode cable shifting
Capability of grounding resistance bypassing by means of
automatic multiple electrode switching
IP - Induced Polarization (Chargeability) available for all 1D / 2D /
3D / 4D methods, up to 20 adjustable IP-windows, each max.
30 s, step 20 / 16.66 ms
Pulse 0.2 s – 30 s, step 0.1 s
SP compensation constant and linear, time-invariant
Stacking manual and automatic (with self-adaptive setting) in the
range 3 – 60
Adjustable optimum measured voltage and maximum acceptable
measurement error
Easy check of data consistency in the field, possibility of
individual data editing and re-measure
Stored values: position of the measured point, output current and
potential, input voltage, SP, apparent resistivity, standard
deviation, chargeability with standard deviation for all 20 IP
windows, GPS coordinates
Output data format: RES2DINV / RES3DINV, Surfer (and others)
Maximum number of electrodes: 65 000 (connected in one array)
Possibility of more devices connection for output power increase
(for example up to 1000 V / 25 A / 3850 W)
Control unit
Easy-Control system, no need of PC for field measurement
Alphanumeric keyboard, 4,7“ graphic LCD display
Safety switch
Memory: 256 MB, up to 100 files, more than 5 mil. readings
PC Interface: USB
Power supply: 12 V car battery, attachable battery pack or AC/DC
converter
178
ARES II
Connectors: for PC, battery and a universal one for all measuring
accessories
Dimensions: 13 x 17 x 39 cm
Weight: 5.9 kg
Ambient conditions: -10°C to +60°C, weatherproof
Standard accessories
Transport case
T-piece (for connection of multi-electrode cable sections and
cables for current and potential electrodes) – different type for
single-channel and ten-channel device
Cable for external 12 V battery (protective – against reverse
polarity)
VES & Profiling adapter (only for ten-channel device)
USB cable
GPS communication cable
AC adapter (for all countries)
Measuring software ARES
Manual
Optional accessories
12 V / 300 W attachable battery pack with fast 3-stage battery
charger (suitable for VES and RP measurement)
AC / DC converter (12 V / 1500 W)
Multi-Electrode cable sections (for single-, five- and ten-channel
measurement)
48-channel adapter II (for ten-channel measurement) for multi-
conductor cable
VES adapter – for 5 pairs of potential electrodes (only for single-
channel device)
Cable reels
Standard electrodes, non-polarizable electrodes
Interpretation and mapping software
RS232, Radio or GSM set for remote control
179
ARES II
Service information
For warranty or after warranty services contact your local
distributor or the producer:
GF Instruments, s.r.o
Ječná 29a
621 00 Brno
Czech Republic
e-mail: info@gfinstruments.cz
tel.: +420 549 522 919
fax: +420 549 522 915
180