Professional Documents
Culture Documents
A Rapid and Simple Liquid-Chromatography-Tandem Mass Spectrometry Method For Measuring 25-Hydroxyvitamin D2 and 25-Hydroxyvitamin D3 in Human Serum: Comparison With Two Automated Immunoassays
A Rapid and Simple Liquid-Chromatography-Tandem Mass Spectrometry Method For Measuring 25-Hydroxyvitamin D2 and 25-Hydroxyvitamin D3 in Human Serum: Comparison With Two Automated Immunoassays
A Rapid and Simple Liquid-Chromatography-Tandem Mass Spectrometry Method For Measuring 25-Hydroxyvitamin D2 and 25-Hydroxyvitamin D3 in Human Serum: Comparison With Two Automated Immunoassays
org
Annals of Clinical & Laboratory Science, vol. 46, no. 6, 2016 645
Key words: 25-Hydroxyvitamin D2, 25-Hydroxyvitamin D3, Liquid chromatography-tandem mass spec-
trometry, Automated immunoassays
Linearity Nominal Intra-assay (n=5) Inter-assay (n=5) LOQ Recovery Stability Matrix
conc.a Accuracy CV(%) Accuracy CV(%) (ng/mL) (%) (%) effect(%)
(ng/mL) (%) (%)
Table 2. Performance of the assay methods using NIST SRM 972a with certified values.
a Each level was measured in triplicate by each assay, and the measured values are expressed in ng/mL.
bThe target value and the uncertainty listed in the NIST certificates.
c,dThe values refer to the concentrations of 25(OH)D and 25(OH)D measured by LC-MS/MS.
2 3
eThe values refer to the sum of concentrations of the two metabolites because 25(OH)D3 and its C-3 epimer coelute under the LC-MS/MS conditions used in this study.
f The biases were assessed excluding the value for 3-epi-25(OH)D assuming a null reactivity for this metabolite.
3
A fast and simple LC-MS/MS method for measuring 25(OH)D2 and 25(OH)D3 647
of 0.3 mL/min. The total (QC) samples by comparison of the measured value with
0.876
0.949
chromatographic run time the calculated one. The QC samples were prepared in
ra
was 3 min. The ESI source 4% BSA prepared in PBS to give concentrations of 1, 5,
was operated in the posi- 25, and 100 ng/mL for the two 25(OH)D metabolites
tive ion mode. 25(OH) using the stock solution that were independent of those
CCC
0.928 (0.904-0.946)
0.737 (0.672-0.792)
D2, 25(OH)D3, and IS used to prepare the calibrators. 25(OH)D2 and 25(OH)
were monitored in the D3 were measured on 5 consecutive days (inter-assay)
multiple-reaction moni- with 5 replicates per day (intra-assay) using fresh samples
toring (MRM) mode us- daily. Recovery was determined by spiking known
ing the following transi- amounts (5, 25, and 100 ng/mL) of 25(OH)D2 and
CCC
potential, entrance poten- lyzed. To investigate the matrix effect [22], the blank
7.90 (-4.7-20.4)
Bland-Altman
tial, collision energy, and samples spiked with 25(OH)D2 and 25(OH)D3 to give
collision cell exit potential the concentrations of 5, 25, and 100 ng/mL were ex-
were optimized at 56, 10, tracted alongside samples prepared in the mobile phase
13, and 14 V for 25(OH) and spiked to the same concentrations. Carryover was
D2, 66, 10, 13, and 10 V determined by repeated injections of blank sample after
for 25(OH)D3, and 66, the highest standard (100 ng/mL) was measured.
10, 13, and 14 V for IS,
respectively. Collision gas, Assay method comparison. The LC-MS/MS method
curtain gas, ion source was compared with the two automated immunoassays,
75.84
52.90
Max
gases 1 and 2 were 6, 20, DiaSorin LIAISON® 25-OH Vitamin D TOTAL Assay
60, and 45 psi, respective- (Stillwater, MN, USA) and Siemens ADVIA Centaur®
ly. Ionspray voltage and XP Vitamin D Total (VitD) assay (Tarrytown, NY,
11.85
Min
4.70
temperature were 5500 V USA), widely used for the measurement of 25(OH)D,
and 350°C, respectively. measuring sera collected from 150 individuals in dupli-
The data were analyzed us- cate. Each immunoassay system was calibrated and per-
ing Analyst® 1.5 data pro- formed according to the manufacturer’s instruction at
0.043 (-0.852-0.811)
Passing and Bablok regression
7.087 (5.787-7.920)
Intercept (CI 95%)
cessing software from Seoul Clinical Laboratories, Seoul, Korea. Method com-
Applied Biosystems/MDS parison was performed according to Clinical Laboratory
SCIEX. ' Standards Institute (CLSI) Evaluation Protocol 9 (EP9)
specifications. The two immunoassays were performed
Method validation. The according to manufacturer’s instructions. All serum sam-
LC-MS/MS assay method ple testing on the automated immunoassays were per-
was validated by measur- formed within the same day. Measuring ranges are 4.0 -
ing a limit of quantifica- 150 ng/mL for the LIAISON assay and 4.2–150 ng/mL
tion (LOQ), linearity, ac- for the ADVIA assay. Sensitivities are 4.0 and 4.2 ng/mL
1.109 (1.058-1.176)
1.071(0.977-1.165)
curacy, precision for the LIASON and the ADVIA assays, respectively.
(expressed coefficient of Certified SRM 972a (levels 1-4) from NIST as indepen-
Slope (CI 95%)
correlation coefficient.
variation, %CV), recov- dent controls were also used to evaluate the accuracies of
ery, processed sample sta- all the assays including the LC-MS/MS method.
bility, matrix effect, and
carryover. The LOQ was Statistical analysis. The statistical analysis of method
determined by serial dilu- validation results was performed using Microsoft Office
tion of the standard solu- Excel® 2013 (Microsoft Inc., Seattle, WA, USA). For
tion (n=9 per dilution) method comparison studies, the 25(OH)D measure-
LIAISON
a Pearson’s
and at the lowest concen- ments of serum samples by the assay methods were ana-
Methods
ADVIA
Figure 2. Comparison of the immunoassays against the LC-MS/MS method using Passing-Bablock regression plots (left pan-
els) and Bland-Altman plots (right panels). (A and B) LC-MS/MS vs. LIAISON. (C and D) LC-MS/MS vs. ADVIA. In left
panels, the solid lines, dashed lines, and dotted lines are the regression lines, the confidence intervals for the regression lines,
and identity lines, respectively. In right panels, the solid lines, dashed lines, and dotted lines are the mean differences, 95%
limits of agreement, and identity lines, respectively.
Among all serum samples tested, the proportions of to monitor clinical vitamin D status. Generally, treat-
those classified as insufficiency (<20 ng/mL) were ment of patients depend on assay results measured by
64.0% by the LC-MS/MS, 54.7% by the LIAISON clinical laboratories. Therefore, the reliability of assay
and 42.7% by the ADVIA, whereas the ADVIA, the methods may have an influence on clinician’s decision
LIAISON and the LC-MS/MS classified 58%, on treatment. The LC-MS/MS method is capable of
45.3% and 36% as sufficient vitamin D status, re- measuring 25(OH)D2 and 25(OH)D3 simultaneous-
spectively. As evidenced by Passing-Bablok regres- ly, which may be useful to monitor the efficacy of sup-
sion and Bland-Altman difference plot analyses, in- plementation especially in countries where vitamin D2
ter-rater agreement analysis using κ value also is used as a supplement [27,28]. Although a result from
demonstrated that the LC-MS/MS method exhibit- vitamin D external quality assessment scheme
ed a stronger agreement with the LIAISON assay (DEQAS) survey proposed that major assay methods
(κ=0.808, 95% CI=0.715-0.902) than with the may provide results within about 10% of the consensus
ADVIA assay (κ=0.579, 95% CI=0.463-0.694) for average [29], evidence of measurement accuracy is not
all the serum samples tested (Table 4). given by comparability between assay methods if agree-
ment between assays is not tested using SRMs or the
Discussion assays performance is not traced to reference methods.
Therefore, the performance of 25(OH)D assay meth-
LC–MS/MS has emerged as the latest technology to ods is a real challenge in most clinical laboratories and
be used to measure the metabolites of vitamin D and is an important criterion for the choice of a method.
A fast and simple LC-MS/MS method for measuring 25(OH)D2 and 25(OH)D3 651
Table 4. Agreement between LC-MS/MS method and each immunoassay on vitamin D status in 150 serum samples.
<20 ng/mL (%) ≥20 ng/mL (%) <20 ng/mL (%) ≥20 ng/mL (%)
In this study, serum sample was first pretreated by on average 11.2% lower than those measured by
protein precipitation with ethanol including a deu- routine LC-MS/MS assays. These results imply that
terated IS, followed by a liquid-liquid extraction routine LC-MS/MS methods being used in clinical
with hexane. These steps were simple, rapid, and laboratories possibly overestimate total 25(OH)D
were necessary to clean up the samples, to minimize in human serum. In this regard, LC-MS/MS meth-
matrix effect and to maximize sensitivity. ods traceable to certified materials such as NIST
Additionally, we tested effectiveness of protein pre- SRM 972, 2972 and 972a are needed to minimize
cipitation with methanol, ethanol, and acetonitrile, the variations.
and ethanol was chosen as an optimal solvent. Our
method was distinct from a recent report in which We demonstrated in this study that the present LC-
methanol and heptane were used as a protein pre- MS/MS method can measure 25(OH)D2 and
cipitation and a liquid-liquid extraction solvent, 25(OH)D3 accurately in serum, as evidenced by
respectively [30]. The variability of LC-MS/MS the measurements of SRM level 1 to 3 (Table 2).
methods between laboratories may result from dif- Although the present LC-MS/MS method could
ferent ways of preparing samples. Furthermore, not resolve 3-epi-25(OH)D3 from 25(OH)D3, it
these sample preparation methods are very impor- showed a smallest mean difference (+0.9%) for
tant because they can affect the accuracy and recov- SRM level 1 to 3 when compared to the two im-
ery of the vitamin D metabolites. munoassays (-26.2% and -14.5% for the LIAISON
and the ADVIA, respectively). Interestingly, the
In this study, we validated a fast, accurate, sensitive present LC-MS/MS method showed the smallest
LC-MS/MS method for measuring serum levels of bias (-0.5%) compared to the LIAISON and the
25(OH)D2 and 25(OH)D3 at nanomolar concen- ADVIA in level 3 with a substantial amount of
trations using an IS material. Method validation 25(OH)D2 and a product from pools of human
demonstrated that the LC-MS/MS method meets serum with endogenous concentrations of 25(OH)
all technical and quality requirements for simulta- D, indicating that it has its capacity to detect
neous quantification of 25(OH)D2 and 25(OH) 25(OH)D2 and 25(OH)D3 selectively. In contrast,
D3 in human serum (Table 1). BSA is being used the LIAISON in level 3 showed a remarkable nega-
for the quantification of 25(OH)D2 and 25(OH) tive bias (-19.3%), may indicate lower recovery of
D3 and proved to be an alternative matrix for cali- 25(OH)D2, independent of 25(OH)D3 in human
bration [21,30,31]. serum [20].
However, how the LC-MS/MS methods are cali- Recently, it has been reported that the mean con-
brated is a major factor to variation between-labo- centration of 3-epi-25(OH)D3 in adult samples is
ratories or between-assays. Recently, the DEQAS generally lower (5.5%) than that in newborns
has shown differences in 25(OH)D results depend- (27.3%) which is known to include high percent-
ing on assay methods, arousing importance for the ages of 3-epi-25(OH)D3 [32]. For this reason, low
agreement and accuracy of different assays [29]. As concentrations of the C3-epimer in adults are un-
a part of this, 20 DEQAS samples were analyzed by likely to be clinically significant, although some
an LC-MS/MS method developed by as a candi- samples with significant amounts of the 3-epimer
date reference measurement procedure by the may be overestimated. More recently, it has been
NIST, revealing that the NIST results were also reported that a LC-MS/MS assays that do not
652 Annals of Clinical & Laboratory Science, vol. 46, no. 6, 2016
separately measure the C3-epimer from other vita- result in higher 25(OH)D measurements by im-
min D metabolites may potentially overestimate munoassays. Additionally, variations in vitamin
total 25(OH)D levels in newborn cord blood sam- D-binding protein concentrations between clinical
ples but the C3-epimer concentration is unlikely to subject groups may be a possible cause for the ob-
contribute to diagnostic misclassification even in served differences between automated immunoas-
the newborn samples [33]. Despite these disputes, says and LC-MS/MS methods. For example, of the
further work will be needed to investigate the im- pregnant women, 67% were vitamin D sufficient
pact of the C3-epimer on the 25(OH)D assays and when measured with a LC-MS/MS method, where
to obtain information about the physiological role ADVIA categorized only 24% as vitamin D suffi-
of the C3-epimer. cient [38].
In this study, the RIAISON and ADVIA assays Although there was a lack of agreement between
generally gave higher 25(OH)D measurements the ADVIA and RIAISON assays compared to the
than the LC-MS/MS method in serum specimens LC-MS/MS method, it was proven through the
tested. Especially, the mean bias of the ADVIA as- CCC and κ analysis that the agreement with the
say (+7.9 ng/mL) was higher than that of the LIAISON was better than that with the ADVIA
RIAISON (+2.4 ng/mL) over the range of concen- (Tables 3 and 4). These results were also similar to
tration measured when compared to the LC-MS/ the findings of the previous studies [25,34]. In this
MS method traceable to the NIST-SRM (Figure 2 study, the prevalence of vitamin D insufficiency was
and Table 3). However, compared to the ADVIA higher with the RIAISON and lower with the
assay (CCC=0.737), the LIAISON (CCC=0.928) ADVIA, compared with the LC-MS/MS result,
showed better concordance with LC-MS/MS re- suggesting that the ADVIA assay would report a
sults (Table 3). Recently, many of the automated lower frequency of patients with vitamin D insuf-
immunoassays turned out to have significant posi- ficiency or deficiency than the LIAISON or the
tive biases when compared to LC-MS/MS methods LC-MS/MS method. However, further studies will
[34,35]. Furthermore, it is noteworthy that most be needed to establish decision limits by different
immunoassays had difficulties measuring low lev- assays and to standardize existing assays for 25(OH)
els. Especially, the RIAISON and the ADVIA as- measurement because our results demonstrated
says which turned out to show excessive mean bi- that the use of uniform cut-off value by assays could
ases of +35% and +118%, respectively, at low levels be an issue when classifying vitamin D status.
<8 ng/mL [34]. In contrast, several studies also re-
ported that the automated RIAISON and ADVIA In conclusion, the present LC-MS/MS method
demonstrated significant negative biases compared with a traceability to SRM 972a is a rapid, simple,
to LC-MS/MS methods [30,36]. reliable, and cost effective alternative to other meth-
ods for determination of 25(OH)D in human se-
There are several possible explanations for the varia- rum. Compared to the LC-MS/MS results, the
tions between the immunoassays and the LC-MS/ agreement with the automated RIAISON assay was
MS methods. First, the variations between the better than that of the ADVIA which showed a sub-
methods might be caused by different calibrator stantial positive bias. Thus, these results should fa-
traceability. Importantly, both immunoassays and cilitate the clinical application of the LC-MS/MS
LC-MS/MS methods may experience matrix effects method for separate quantification of 25(OH)D2
that occurs between the matrix in the calibrators and 25(OH)D3.
and the clinical samples, providing lower or higher Acknowledgment
25(OH)D measurements [13]. Especially, the
25(OH)D2 and 25(OH)D3 concentrations mea- This work was supported by Seoul Medical Science Institute &
Seoul Clinical Laboratories in 2014-2015.
sured by routine LC-MS/MS methods may be dif-
ferent between laboratories and show biases when References
compared to those by NIST LC-MS/MS [37].
Second, all immunoassays show high cross-reactivi- 1. DeLuca HF. Evolution of our understanding of vitamin D.
Nutr Rev 2008;66:S73-S87.
ty with other metabolites of 25(OH)D such as 2. DeLuca HF. Overview of general physiologic features and
24,25(OH)2D, which is known to be present in the functions of vitamin D. Am J Clin Nutr 2004;80:1689S-1696S.
serum at levels of nearly 10-15 nmol/L [37], may 3. Holick MF, Binkley NC, Bischoff-Ferrari HA, Gordon CM,
Hanley DA, Heaney RP, Murad MH, Weaver CM. Endocrine
A fast and simple LC-MS/MS method for measuring 25(OH)D2 and 25(OH)D3 653
Society 2011 Evaluation, Treatment, and prevention of vitamin 25-hydroxyvitamin D2 and D3. J Clin Lab Anal
D deficiency: an endocrine society clinical practice guideline. J 2012;26:349-357.
Clin Endocrinol Metab 2011;96(7):1911-1930. 22. Matuszewski BK, Constanzer ML, Chavez-Eng CM.
4. Lappe JM, Travers-Gustafson D, Davies KM, Recker RR, Strategies for the assessment of matrix effect in quantitative
Heaney RP. Vitamin D and calcium supplementation reduces bioanalytical methods based on HPLC-MS/MS. Anal Chem
cancer risk: results of a randomized trial. Am J Clin Nutr 2003;75:3019-3030.
2007;85(6):1586-1591. 23. Lai JK, Lucas RM, Banks E, Ponsonby AL. Variability in vita-
5. Welsh J, Vitamin D and breast cancer: insights from animal min D assays impairs clinical assessment of vitamin D status.
models. Am J Clin Nutr 2004;80:1721S-1724S. Intern Med J 2012;42(1):43-50.
6. Morris HA. Vitamin D activities for health outcomes. Ann Lab 24. Choi HS, Oh HJ, Choi H, Choi WH, Kim JG, Kim KM, Kim
Med 2014;34(3):181-186. KJ, Rhee Y, Lim SK. Vitamin D insufficiency in Korea--a
7. Cumhur CM, Cure E, Yuce S, Yazici T, Karakoyun I, Efe H. greater threat to younger generation: the Korea National
Mean platelet volume and vitamin D level. Ann Lab Med Health and Nutrition Examination Survey (KNHANES)
2014;34(2):98-103. 2008. J Clin Endocrinol Metab 2011;96(3):643-651.
8. Endres DB, Rude RK Disorders of Bone. In: Tietz 25. Moon HW, Cho JH, Hur M, Song J, Oh GY, Park CM, Yun
Fundamentals of Clinical Chemistry, 6th ed (Burtis CA, YM, Kim JQ. Comparison of four current 25-hydroxyvitamin
Ashwood ER, Bruns DE, Eds), Saunders, Missouri, 2008; pp D assays. Clin Biochem 2012;45:326-330.
723-726. 26. Thacher TD, Clarke BL. Vitamin D insufficiency. Mayo Clin
9. Jones G. Pharmacokinetics of vitamin D toxicity. Am J Clin Proc 2011;50-60.
Nutr 2008;88:582S-586S. 27. Saenger AK, Laha TJ, Bremner DE, Sadrzadeh SM.
10. Ross AC, Taylor CL, Yaktine AL, Del Valle HB(Institute of Quantification of serum 25-hdyroxyvitamin D(2) and D(3)
Medicine (US) Committee), Dietary reference intakes for cal- using HPLC-tandem mass spectrometry and examination of
cium and vitamin D. National Academy Press, Washington, reference intervals for diagnosis of vitamin D deficiency. Am J
DC, 2011. Clin Pathol 2006;125:914-920.
11. Zerwekh JE. The measurement of vitamin D: analytical as- 28. Eyles D, Anderson C, Ko P, Jones A, Thomas A, Burne T,
pects. Ann Clin Biochem 2007;41:272-281. Mortensen PB, Nørgaard-Pedersen B, Hougaard DM,
12. Herrmann M. The measurement of 25-hydroxyvitamin D - an McGrath J. A sensitive LC/MS/MS assay of 25OH vitamin D3
analytical challenge. Clin Chem Lab Med 2012;50:873-1875. and 25OH vitamin D2 in dried blood spots. Clin Chim Acta
13. Wallace AM, Gibson S, de la Hunty A, Lambert-Allardt C, 2009;403:145-151.
Ashwell M. Measurement of 25-hydroxyvitmain D in the clini- 29. Carter GD. Accuracy of 25-hydroxyvitamin D assays; con-
cal laboratory: current procedures, performance characteristics fronting the issues. Curr Drug Targets 2011;12(1):1-10.
and limitations. Steroids 2010;75:477-488. 30. Zhang S, Jian W, Sullivan S, Sankaran B, Edom RW, Weng N,
14. Phinney KW, Bedner M, Tai SS, Vamathevan VV, Sander LC, Sharkey D. Development and validation of an LC-MS/MS
Sharpless KE, Wise SA, Yen JH, Schleicher RL, Chaudhary- based method for quantification of 25 hydroxyvitamin D2 and
Webb M, Pfeiffer CM, Betz JM, Coates PM, Picciano MF. 25 hydroxyvitamin D3 in human serum and plasma. J
Development and certification of a standard reference material Chromatogr B 2014;961:62-70.
for vitamin D metabolites in human serum. Anal Chem 31. van den Ouweland JMW, Beijers AM, Demacker PNM, van
2012;84(2):956-962. Daal H. Measurement of 25-OH-vitamin D in human serum
15. Denimal D, Ducros V, Duprè T, Dousset B, Meunier C, Aho using liquid chromatography tandem-mass spectrometry with
S, Guilland JC, Lemaire-Ewing S. Agreement of seven 25-hy- comparison to radioimmunoassay and automated immunoas-
droxyvitamin D3 immunoassays and three high performance say. J Chromatogr B 2010;878:1163-1168.
liquid chromatography methods with liquid chromatography 32. van den Ouweland JMW, Beijers AM, van Daal H.
tandem mass spectrometry. Clin Chem Lab Med Overestimation of 25-hydroxyvitamin D3 by increased ioniza-
2014;52(4):511-520. tion efficiency of 3-epi-25-hydroxyvitamin D3 in LC-MS/MS
16. Tai SS, Bedner M, Phinney KW. Deveolopment of a candidate methods not separating both metabolites as determined by an
reference measurement procedure for the determination of LC-MS/MS method for separate quantification of 25-hy-
25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 in human droxyvitamin D3, 3-epi-25-hydroxyvitamin D3 and 25-hy-
serum using isotope-dilution liquid chromatography-tandem droxyvitamin D2 in human serum. J Chromatogr B
mass spectrometry. Anal Chem 2010;82:1942-1948. 2014;967:195-202.
17. Stepman HC, Vanderroost A, Van Uytfanghe K, Thienpont 33. D.J. Cooke, B. Cooke, D. Bell, S. Vasikaran, P. Glendenning,
LM. Candidate reference measurement procedures for serum C-3-epi-25-hydroxyvitamin D is universally present in neona-
25-hdyroxyvitamin D3 and 25-hydroxyvitamin D2 by using tal Western Australian samples but is unlikely to contribute to
isotope-dilution liquid chromatography-tandem mass spec- diagnostic misclassification. Ann. Clin. Biochem. (2015) pii:
trometry. Clin Chem 2011;57:441-448. 0004563215625693.
18. Kwak HS, Chung HJ, Cho DH, Park MH, Ku ES, Park EJ, Oh 34. Farrell CJ, Martin S, McWhinney B, Straub I, Williams P,
HJ. Efficacy of the measurement of 25-hydroxyvitamin D2 and Herrmann M. State-of-the-art vitamin D assays: A compari-
D3 levels by using PerkinElmer liquid chromatography-tan- son of automated immunoassays with liquid chromatography-
dem mass spectrometry vitamin D kit compared with DiaSorin tandem mass spectrometry methods. Clin Chem
radioimmunoassay kit and Elecsys vitamin D total assay. Ann 2012;58:531-542.
Lab Med 2015;35(2):263-265. 35. Enko D, Kriegshäuser G, Stolba R, Worf E, Halwachs-
19. Herrmann M, Harwood T, Gaston-Parry O, Kouzios D, Wong Baumann G. Method evaluation study of a new generation of
T, Lih A, Jimenez M, Jauc M, Seibel MJ. A new quantitative vitamin D assays. Biochem Med 2015;25(2):203-212.
LC tandem mass spectrometry assay for serum 25-hydroxyvita- 36. Hsu SA, Soldo J, Gupta M. Evaluation of two automated im-
min D. Steroids 2010;75:1106-1112. munoassays for 25-OH vitamin D: comparison against LC-
20. de Koning L, Al-Turkmani MR, Berg AH, Shkreta A, Law T., MS/MS. J Steroid Biochem Mol Biol 2013;136:139-145.
Kellogg MD. Variation in clinical vitamin D status by DiaSorin 37. Carter GD. 25-hydroxyvitamin D: a difficult analyte. Clin
Liaison and LC-MS/MS in the presence of elevated 25-OH vi- Chem 2012;58:486-488.
tamin D2. Clin Chim Acta 2013;415:54-58. 38. Heijboer AC, Blankenstein MA, Kema IP, Buijs MM.
21. Garg U, Munar A, Frazee C, Scott D. A simple, rapid atmo- Accuracy of 6 routine 25-hydroxyvitamin D assays: Influence
spheric pressure chemical ionization liquid chromatography of vitamin D binding protein concentration. Clin Chem
tandem mass spectrometry method for the determination of 2012;58(3):543-548.