Professional Documents
Culture Documents
Manual Fabian Ncpap Evolution 1.07 GB
Manual Fabian Ncpap Evolution 1.07 GB
fabian + nCPAP
evolution
Ventilator for Premature Babies and Children
OM Version: 1.07 GB
SW Version: 1.0.0.n
Warning
2 Operating Manual................................................................................................................................ 8
2.1 Notes and Warnings................................................................................................................... 9
18 Accessories........................................................................................................................................ 89
19 Annex ................................................................................................................................................... 90
19.1 Explanation of electromagnetic compatibility ..................................................................... 91
This operating manual describes the device components and their operation. They are structured to
enable you to become familiar with the operation of your ventilator step by step.
Read the operating manual through carefully before you begin to work with the
ventilator.
Once you are familiar with the basic design and operation of the ventilator, you can use the operating
manual as a reference guide.
The operating manual are intended solely for customer information and are only refreshed or replaced
in response to customer requests.
NOTE
DANGER
DANGER indicates a non-imminent, but latent danger, which if not avoided can
lead to death or physical danger.
DANGER
PLEASE NOTE
PLEASE NOTE indicates a danger, which if not avoided can result in damage to
the device.
All handling of the ventilator presumes precise knowledge and compliance with this
operating manual. The device is only intended for the described use.
The ventilator may only operate with accessories recommended by Acutronic Medical
Systems AG.
The ventilator may only be operated by qualified expert staff or under their supervision
so that a remedy may be found immediately in case of a malfunction.
An alternative ventilation system (e.g. air-mask bag unit) must always be available
during use of the ventilator.
The ventilator may not be used in combination with inflammable gases or narcotic
agents – there is otherwise an acute risk of fire and explosion!
An acoustic alarm indicates a system or patient alarm and requires action by medical
specialists in all cases.
If a defect is identified during a device check, the ventilator may not be connected to a
patient under any circumstances!
Electric coupling with devices not mentioned in this operating manual may only be
possible after consultation with the manufacturers or an expert.
Note: the safety and health of the users is guaranteed by the fact that the products for
example do not contain any allergenic or mutagenic materials such as phthalates.
In all cases, liability for the function of the to observe the above remarks. Guarantee and
device passes to the owner or operator if the liability conditions of the terms of sale and
device is improperly maintained or repaired by delivery of ACUTRONIC Medical Systems AG
individuals who do not belong to ACUTRONIC are not extended by the above remarks.
Medical Systems AG Service or if handling that
does not correspond with the intended use
occurs. ACUTRONIC Medical Systems AG
Ventilator for premature babies, the newly-born continuous positive airway pressure
and children up to 30 kg bodyweight. (CPAP)
pressure support ventilation (PSV)
The fabian evolution is intended for in-patient
pressure support ventilation combined
use in hospitals, medically-used rooms and with SIMV (SIMV + PSV)
intra-hospital patient transport.
NIV ventilation
The ventilator is intended for the following nCPAP / duoPAP
ventilation types: Oxygen therapy
In the event of ventilation for more than several hours, care must be taken for optimal
conditioning of the respiratory gases (warmth, moisture) to guarantee secretion
mobilization and prevent damage to mucous membranes.
Detailed studies and experience can be requested from the following address:
Fabrik im Schiffli
CH-8816 Hirzel / Switzerland
Tel: +41 44 729 70 80
Fax: +41 44 729 70 81
E-mail: info@acutronic-medical.ch
www.acutronic-medical.ch
4-1
1 Inspiration tube.
4 Expiration tube
4-2
Symbol Description
USB connection
Nurse call
Information on disposal
Do not cover
1 Touch screen
Manual breath
O2-flush
Cursor buttons
These buttons serve as control in several menus
Menu / calibration
Opening setting and calibration menu. By pressing once you reach the
calibration menu, by pressing again the configuration menu is opened.
Curves/loops
… switches to the curve-/loop display.
Home
… switches to the main display
Alarm limits
Mute alarm
… for acknowledging and muting alarms for a maximum duration of 120
seconds.
Subsequent alarms with higher priority are only displayed optically during
muting.
1 Screen
2 LED display
5-2
5.2.2 Screen
1 information bar
battery status
date-/time display
status information,
System information
alarm information 5-3
2 graphic field
3 function fields
4 selection or display of operating mode
5 numeric field / alarm limits
Depending on the display mode selected,
individual fields are faded in or out.
5-4
5-5
1 - Neonatal or pediatric mode
- patient present YES / NO
- Nurse call ON / OFF
- Charging status of battery (remaining life) 5-6
Pediatric mode
5-8
Neonatal mode
5-9
5-12
Numeric field
5-13
Alarm limits
5-14
Button A: presentation with three curves
As an alternative, you can switch to the loop
presentation.
LOOP presentation:
pressure / volumes
volumes / flow
5-15
Button B: loop presentation
Warning LED
This LED is illuminated or flashes in red if a system
alarm occurs.
Battery LED
This LED shines yellow during battery operation and
green if the battery is fully charged.
The LED flashes in green when the battery is being
charged.
LED mute alarm
This LED lights up in yellow when the alarm is
muted.
The remaining time for the alarm muting is displayed
on the screen.
Mains LED
This LED is shines green during mains operation and
switches off during battery operation.
5-18
5-22
Alarms which are active before the mode change are automatically reset.
5-23
List of locks
You reach the graphic menu via the curve button of the keypad.
1 curves
2 loops
3 trends
5-25 curves
5-26 loops
5-27
Trends are loaded when the trends key is activated.
pressure
flow
volumes
5-29
5-30
5-31
5-32
Save loop:
freeze loop
press “Save” key for three seconds Signal
tone-> loop has been saved
The loop remains saved until a new loop is
saved.
5-33
5-34
5-35
5-36
1 Parameters / selection
2 Graphics
3 Parameter values
4 Time axis
5 Modification of the time axis (max. five days – min. 30 minutes)
6 Refresh data
The device can be operated with 100 VAC to 240 VAC and automatically adjusts to particular voltage
without manual switch-over being necessary.
PLEASE NOTE
The fuses used must correspond with the value and type approved for the fabian.
The particular fuse type and value is printed on the back of the fabian.
Replacement of the fuses may only be carried out by trained staff.
The pressure gradient via the breathing system of the ventilator measured at the patient
connection opening can be increased by the attachment of accessory parts or other
components to the breathing system.
The labelling
TOP
6-3
1 Ventilation tubes
2 Y-piece
3 Flow sensor
4 Flow sensor connection cable
5 Pressure measurement tube proximal
6-6
6-8
6-9
A device check must be carried out before each use of the ventilation device!
Gas supply Screw on compressed air and oxygen supply cables Tubes screwed on tight, plug
at the back, insert plug. inserted
Breathing system Membrane holder and membrane Membrane holder and membrane
Tubes correctly installed.
Water traps (if needed) Ventilation system assembled and
Breathing gas humidifier and tube heating connected according to
Flow sensor manufacturer information.
Connect test lung
Switch on test switch on fabian Self-test successful
Alarm tone (loudspeaker) Alarm tone audible during switch-
on test.
Calibration Calibrate flow sensor Calibration successful
Calibrate O2 sensor (occurs automatically after
leaving calibration menu)
Tightness Start CPAP mode Pressure of 80 ± 4 mbar is achieved
CPAP 5 mbar
P manual 80 mbar
You reach the configuration menu by activating the menu / calibration key of the keypad and
subsequent selection of the button “Open menu”.
8-1
1 Calibration ( Flow und O2 ) 6 Date / time
2 Display 7 Tools (e.g. save trend files)
3 Ventilation parameters 8 System information
4 Patient data 9 Service mode
5 Language 10 Close menu (back to previous menu)
8-2 Calibration
In the event of the failure of one gas type (oxygen or compressed air), no balancing of
the oxygen sensor is possible.
Manual calibration:
8-3
8-4 Calibration
Access to the calibration menu also directly via the calibration key:
“Calibration running”
8-6 Display
8-7
1 Lock touchscreen or
2 automatic lock of the touchscreen after pre-set time.
The screen can also be locked by pressing the incremental encoder for > three seconds.
The lock is automatically cancelled in the event of alarm or by pressing the incremental
encoder.
3 Curve presentation filled or line presentation.
4 Automatic scaling ON / OFF
8-8 Ventilation
1 Patient area Before setting of the ventilation parameters can begin, the
appropriate area setting must be selected.
2 Alarm delay After falling below the Pmin value, the alarm only occurs after the
delay set here.
5 Automatic oxygen calibration Only with 21 vol% or with 21 vol% and 100 vol%
7 Maximum time for manual breath Input of inspiration time for manual breath
The next page of the ventilation parameters is retrieved with the arrow keys on the lower screen edge
2 Termination criterion PSV The PSV breaths are terminated if the inspiratory flow has fallen
by the percentage value of the value previously reached.
4 Hospital settings The button “Hospital settings” is only active if special settings
were previously saved in the fabian.
Hospital settings can be loaded via the service menu or copied in
service mode.
Only carry out resetting to hospital settings or factory settings if no patient is connected
to the fabian.
8.4 Patient data
Note:
8.5 Language
Note:
8.8 Info
Among other things, the hospital settings and nurse call can activated or deactivated.
Please contact your responsible distributor or Acutronic Medical Systems AG for more information.
9-1
The following settings can be carried out in the alarm limits menu:
9-2
9-3
After the alarm limit has been adjusted or no active alarm still exists, the bell symbol is shown grey.
The alarm history can be retrieved via the alarm limits menu
date / time
report
priority color highlighted
Exspiration flow The continuous exspiratory flow can be varied irrespective of the inspiration
flow.
Volume guarantee The mechanical breaths are volume-controlled with this additional function.
Changes in the ventilatory system are compensated.
As a result, the patient is provided with a guaranteed volume with each
breath.
Battery life with fully charged battery in IPPV mode: approx.. 3 hours.
In the event of failure of the power supply, the pneumatic system opens
against atmospheric pressure. As a result, no pressure can build up in the
breathing system and spontaneous breathing of the patient is facilitated.
This information signal can be confirmed with the mute alarm button.
30 minutes
15 minutes
60 minutes, only if upon failure of the power
supply the residual life is less than 60 11-1
minutes.
The fabian evolution can likewise be operated with an external power supply.
An external supply with a voltage of 12 – 24 volts must be connected to the back of the device in this
case.
11-2
11-3
IPPV
Controlled ventilation with prescribed ventilation patterns
Breath volume can be pre-set or limited
S – IPPV
Controlled ventilation with pre-set ventilation patterns.
Breath volume can be pre-set or limited
The inspiration is synchronized with each spontaneous breath of the patient.
S - IMV
Controlled ventilation with pre-set ventilation patterns.
Breath volume can be pre-set or limited
The inspiration is synchronized with each spontaneous breath of the patient.
The patient can breathe spontaneously between the mandatory mechanical breaths.
PSV
Synchronized ventilation form with pre-set inspiration pressure.
S-IMV + PSV
Controlled ventilation with pre-set ventilation patterns.
Breath volume can be pre-set or limited
The inspiration is synchronized with each spontaneous breath of the patient.
The patient can breathe spontaneously between the mandatory mechanical breaths; these
mechanical breaths are supported by the PSV.
CPAP
Spontaneous breathing under positive airway pressure
nCPAP / duoPAP
NIV-ventilation with special nCPAP systems.
The following nCPAP systems are supported (as of 07/2012):
- Medijet ®
- Infantflow ®
- Infantflow LP ®
O2 therapy
Oxygen supplementation via oxygen masks, nasal cannula or hood.
12-1
In addition, the ventilator is synchronized with the ventilation pattern of the patient.
If the patient is not breathing on their own, mechanical ventilation occurs with the pre-set ventilation
parameters. If the patient is breathing spontaneously, the device synchronizes with a PSV breath.
12-2
The inspiratory pressure for the PSV breath is set with the key P PSV.
In relation to PEEP pressure.
The Pressure Support Ventilation – option is for the airway pressure exceeds a pre-set
pressure support of insufficient spontaneous pressure P INSP
breathing among the triggered ventilation the inspiratory flow has fallen to a set
types. The breathing frequency is determined % value of the previously achieved
by the spontaneously breathing patient; the maximum value (this indicates an
ventilator assumes an adjustable proportion of almost full lung)
the breathing work.
As a result, the inspiration is determined by the If the patient is no longer spontaneously
patient and occurs according to the breathing, the ventilator assumes substitute
parameters set in the ventilation type SIMV or ventilation wit the ventilation parameters set.
ASSIST. But the ventilator now decides on the
exspiration. This occurs as soon as one of the
following criteria is met:
In the case of CPAP ventilation the patient breathes spontaneously. The ventilator does not perform
any breathing work. All that happens is that both during inspiration as well as exspiration positive
pressure is generated as a result of which the breathing work of the patient is greatly reduced. The
maximum automatic leak compensation is 50% of the inspiration flow.
If the spontaneous breathing of the patient stops, the ventilator takes a pre-set number of mechanical
breaths after expiry of the apnoea period in order to restimulate spontaneous breathing. When the
latter begins, the stimulation is terminated and only restarts after changed apnoea.
In nCPAP / duoPAP mode the patient breathes spontaneously through a mask or nasal cannula. Since
no flow sensor is connected, additional apnoea monitoring must be used, e.g. impedance
measurement.
Additional monitoring of SpO2 and PCO2 is recommended.
nCPAP: Provides positive airway pressure with automatic leak compensation. The maximum flow
compensation is 25%.
duoPAP: Like nCPAP, but with the option of positive pressure ventilation with adjustable frequency and
inspiration time.
In this mode a special nCPAP patient set is needed with nCPAP generator.
The appropriate system must be selected in the configuration menu before using
the nCPAP / duoPAP modes.
Medijet ®
Infantflow ®
Infantflow LP ®
The pressure measurement connection “Prox.” can be used for pressure measurement.
The pressure measured is displayed in the curve menu.
The following additions can be combined with the mode selected to optimize ventilation.
The volume limit function switches automatically to exspiration if the set tidal volume is reached.
If the ventilation pattern is changed or the volume limit adjusted, the tidal volume is automatically
limited.
If the flow measurement fails (e.g. sensor change), “V limit” is automatically deactivated.
As soon as the flow measurement is again available, “V limit” is automatically reactivated.
The volume guarantee is a pressure-limiting additional function with guided or targeted tidal volume
application.
A red line in the graphic display shows the upper pressure limit.
If the pre-set P insp. for the application of the ventilation volume is insufficient, the message
appears: “Volume not reached”
Disconnecting the plug of the flow sensor is not enough to prevent this.
Consequently, remove the flow sensor before medicine nebulization.
Switch on fabian.
Calibrate the oxygen sensor at 22 vol. % and 100 vol.
%.
When installing a new factory sensor changed calibration can be necessary after about
30 minutes.
If possible, expose the new sensor for about 30 minutes to the ambient air BEFORE start-
up.
The exhausted O2-sensors must be disposed of professionally. The following must be borne in mind in
this connection:
Maintenance and the safety check may only be carried out by specialists trained by
Acutronic, who have suitable measuring equipment and test facilities.
Only wet disinfection, e.g. with Buraton 10 F or Terralin (Schülke & Mayr, Norderstedt, Germany) may
be carried out. The application regulations of the manufacturer must be observed in this regard.
In general, the tube system must be changed after each patient treatment to avoid contamination of
succeeding patients.
Sterilization of accessories:
Flow sensor
Rinse sensor immediately after use with warm water and then in cleaning solution, like: :
Insert Neodisher Mediclean forte ( Dr. Weigert Germany ), since otherwise encrusting may occur as
a result of which the function of the sensor is no longer guaranteed.
Never connect an uncalibrated sensor to a patient. Always calibrate the sensor before connecting
to a patient.
Clean and disinfect wet thermally (93°C / 10 min) in the automatic cleaning/disinfection machine just
with detergent.
In general, sterilize the membrane holder to ensure that residual water in the channels and control
cables dries out.
Sterilization of accessories
Ventilation tubes
Heating wire
Water trap
Temperature sensor
Y-piece
Mode
SIMV SIMV+PSV
Mode
NCPAP DUOPAP O2 Therapy
Airway measurement
Tube leak
Range: 10 – 50%
Precision: ±10%
Resolution: 1% ml
Dynamic compliance
Resistance
Inspiratory O2 concentration
Airway pressure
Inspiratory O2 concentration
Breath frequency
Volume monitoring
Alarm lower threshold value: If the breath volume set was not supplied
Apnoea alarm
Operating time if no mains supply with fully charged at least three hours
internal battery
Acoustic pressure ≤ 42 dB ( A )
PIN
1 DCD
2 RXD
3 TXD
4 DTR
5 GND
6 DSR
7/8 RTS / CTS
The air/oxygen blender provides a gas blend with adjustable proportions of oxygen and air.
Gas from the central gas supply enters the device via the gas input connections.
Blending of the gases occurs via two proportional valves.
The proportional valves, which also act as return valves, prevent the return flow of a gas into the
supply cable of the other gas.
You can obtain a complete accessories catalogue from your distributor or direct from Acutronic
Medical System AG
Note
The “fabian” device is an ELECTRICAL MEDICAL DEVICE, which is subject to special precautionary
measures regarding EMC and which must be installed and put into operation according to the
information included in this document.
Warning
Portable, mobile HF communication facilities can influence ELECTRICAL MEDICAL DEVICES!
Warning
The “fabian” device may not be stacked, arranged or used directly next to or with other devices.
However, if this is required, it is necessary to monitor this in order to check operation as intended in
these instructions.
Warning
The use of other accessories, cables or transformers with the “fabian” device can lead to increased
disturbance output and a reduction of the disturbance resistance of the “fabian.”
Note
The main features of the “fabian” ventilator are:
- The fabian must function within the defined specification and its medical purpose. If the
specifications are not complied with, ventilation is stopped.
- For this reason a second, independent ventilation facility must be kept ready, e.g. air-mask bag unit.
The “fabian” device uses HF energy exclusively for its internal functions.
HF emissions
Group 1 Consequently, its HF emissions are very low and it is unlikely that neighboring
CISPR 11
devices will be disturbed.
HF emissions
Class A
CISPR 11
Emission of overtones
Class A
IEC 6100-3-2 The “fabian” device is suitable for use in all other areas apart from the
residential area and apart from those areas that are connected directly to a
public supply grid which also supplies buildings used for residential purposes.
Emission of voltage fluctuations /
flicker Met
IEC 6100-3-3
HF emission
Class A
CISPR 11
Emission of
overtones IEC N/a
61000-3-2
Emission of voltage
fluctuations / flicker N/a
IEC 61000-3-3
Surge voltage ± 1 kV Cable to The mains quality should correspond to the typical
cable ± 1 kV Cable to cable surroundings of businesses or hospitals.
IEC 61000-4-5
± 2 kV Cable to ± 2 kV Cable to ground
ground
Voltage dips, temporary The mains quality should correspond to the typical
interruptions and surroundings of businesses or hospitals. If the user of the
fluctuations of supply <5% UT “fabian” device requires continuous use during a mains
voltage (>95% Dip of UT) <5% UT interruption, it is recommended that the “fabian” device is
for (>95% Dip of UT) for supplied via an uninterruptible power supply or a battery.
IEC 61000-4-11 ½ period ½ period
40% UT 40% UT
(60% Dip of UT) for (60% Dip of UT) for
5 periods 5 periods
70% UT 70% UT
(30% Dip of UT) for (30% Dip of UT) for
25 periods 25 periods
<5% UT <5% UT
(>95% Dip of UT) (>95% Dip of UT) for
for 5 seconds
5 seconds
Magnetic field in supply 3 A/m 3 A/m Magnetic fields in the mains frequency should have typical
frequency values that comply with what is to be found in business
(50/60 Hz) and hospital surroundings.
IEC 61000-4-8
Note UT is the mains alternating voltage before application of the test level.
Portable, mobile radio devices should not be used at any distance closer to
the “fabian” device, including the cables, than the recommended
protective distance calculated according to the equation appropriate for
the transmission frequency.
10 Veff
150 kHz to 80 MHz
in the ISM bands b
NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2 These guidelines may not be applicable in all cases. The propagation of electromagnetic factors is influenced by absorption and
reflection of buildings, objects and people.
a
Die ISM bands (industrial, scientific and medical) between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz;
26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.7 MHz.
b
The compliance level in the ISM frequency bands between 150 kHz and 80 MHz and between 80 MHz and 2.5 GHz were established to
ensure that the probability was reduced that mobile/portable transmitting equipment would produce disturbance in case it was
brought into the patient’s surroundings by mistake. For this reason an additional factor of 10/3 was used in calculating the
recommended protective distance for transmitters in this frequency range.
c
The field strength of stationary transmitters, such as base stations of radio telephones (mobile/cordless) and mobile terrestrial radio
equipment, amateur radio stations, AM and FM radio and TV broadcasters cannot be precisely predicted in theory. A study of the
location should be considered to determine the electromagnetic surroundings with regard to stationary transmitters. If the field
strength measured at the location where the “FABIAN” is used exceeds the compliance level above, the “FABIAN” device should be
observed to determine function as intended. If unusual features are observed, additional measures may be necessary, such as altered
orientation or another location of the “FABIAN” device.
d
The field strength should be less than 10 V/m across the frequency range from 150kHz to 80MHz.
The “fabian” device is intended for operation in an electromagnetic environment in which HF disturbance factors are controlled. The
customer or user of the “fabian” device can help to avoid electromagnetic disturbances by complying with the minimum distance between
portable, mobile HF telecommunications devices (transmitters) and the “fabian” device – depending on the output power of the
communications device as stated below.
100 4.0 12 12 23
For transmitters whose maximum nominal power is not stated in the above table, the recommended protective distance d in meters (m) can
be determined using the equation belonging to the respective column with P being the maximum nominal power of the transmitter in Watt
(W) according to the information of the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz the protective distance for the higher frequency range applies.
NOTE 2 The ISM bands (industrial, scientific and medical) between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567
MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.7 MHz.
NOTE 3 An additional factor of 10/3 was used in calculating the recommended protective distance for transmitters in the ISM frequency
range between 150KHz and 80 MHz and between 80MHz and 2.5GHz to reduce the probability that mobile/portable transmitting equipment
would produce disturbance if they were brought into the patient’s surrounding by mistake.
NOTE 4 These guidelines may not be applicable in all cases. The propagation of electromagnetic factors is influenced by absorption and
reflection of buildings, objects and people.