Rendered PDF File Page 1 of 2 DOC0843742, Rev:1

Electronic Signature Information

Name DOC0843742

Revision 1

Type Controlled Document

Title TB LSS 10 050 Aladin Cassette Agent Cassette Board

Originator 212030429_anthony_s_bean

Name Reason For Change File Size (Bytes)


Errors generated in the ADU and Aisys with certain Aladin II
TBLSS10050_DOC0843742.doc 202240
Cassettes

Route Signer Status Comments Completion Date


R-4315002 212062879_dave_t_clark Approve Vaporizer IB LPI 20 Sep 2010 21:51:07 GMT
R-4315002 212030167_richard_l_dodd Approve QA 20 Sep 2010 21:10:22 GMT
R-4315002 212030251_lawrence_r_carlson Approve 21 Sep 2010 12:43:15 GMT
R-4315002 100014553_michael__foulis Approve 21 Sep 2010 01:33:28 GMT

* Printed versions are For Reference Only *


Released
Rendered PDF File Page 2 of 2 DOC0843742, Rev:1

GE Healthcare Page 1 of 1

Technical Bulletin
Life Support Solutions/Global Service Engineering

Ref: TB LSS 10 050 Issue Date: October 1, 2010


Ref. Note: DOC0843742
Product Group: Anesthesia Delivery Priority: Normal
Product(s): Aladin II Cassette Classification: Information Only
Affected Part(s): Agent Delivery Board
Subject: Errors generated in the ADU and Aisys with
certain Aladin II Cassettes

The Service Engineering Team has received inquiries regarding machine errors generated when
certain Aladin II Cassettes are used in the ADU or Aisys machines.

An investigation into the returned samples revealed that certain Agent Cassette Boards (1011-
3170-000) are more susceptible to generate level sense error messages. The root cause of the
issue was traced to a specific manufacturer of an infrared emitter / detector chip on the Agent
Delivery Board when paired with agent level windows that contained small imperfections
(yellowed/scratched). The combination of a yellowed/scratched agent window with this specific
Agent Cassette Boards can result in “AA Liquid Sensor Failure” messages on the ADU and
“Cassette Overfill Detected” log entries in the Aisys.

Our records indicate the use of this specific manufacture of the infrared emitter / detector chip was
discontinued from use at the request of the board manufacture in July 2009.

If you encounter Aladin II Cassettes that exhibits the above described symptoms in ether the ADU
or the Aisys machine, the recommended action is to replace the Agent Cassette Board, part
number 1011-3170-000, in the Aladin II Cassette. The available stock should no longer exhibit this
issue.