Professional Documents
Culture Documents
BA-A-S120Mini Software Operating Instructions
BA-A-S120Mini Software Operating Instructions
FULLY AUTO
BIOCHEMISTRY ANALYZER
Version 2023.07.01
User Manual
BA-A-S120Mini Automatic Chemistry Analyzer
Notes
1. Please read through this operating instruction and
keep it properly;
2. Please keep this operating instruction in accessible
place for reading;
3. Please operate this instrument according to operating
manual, otherwise, it may cause severe consequences.
User Manual of BA-A-S120Mini Analyzer
Catalogue
1.2.1 Symbol...........................................................................................................................................6
1.2.3 Picture...........................................................................................................................................7
-I-
User Manual of BA-A-S120Mini Analyzer
- II -
User Manual of BA-A-S120Mini Analyzer
7.7 QC ...................................................................................................................................................................96
Chapter 8 Abnormal Model and Cause Analysis of Common Results ................................ 103
8.4 Normal Reaction Curve, Results Is Zero or Even Lower ...................................................................... 104
8.6 Normal Reaction Curve, Results are Higher Or Lower Generally .......................................................105
- III -
User Manual of BA-A-S120Mini Analyzer
Chapter 1 Introduction
1.1Product overview
1.1.1Product introduction
Product name: Automatic chemistry analyzer;
Model: BA-A-S120Mini;
Safety classification: Electric shock resistance grade I, overvoltage grade II, pollution grade 2;
1.1.2Basic performance
Central wavelength
No more than ±2nm
deviation
Error no bigger than ±0.020 when absorbance is 0.5, Error no bigger than
Absorbance accuracy
±0.04 when absorbance is 1.0
Absorbance
Variable coefficient no bigger than 1.5%
repeatability
Dimension L×W×H:350×605×330mm
Net weight 22 Kg
-4-
User Manual of BA-A-S120Mini Analyzer
1.1.3Main structure
BA-A-S120Mini automatic chemistry analyzer consists of analysis unit and operation unit. The analysis unit
includes sample disk, sample absorbing mechanism, reagent disk, reagent absorbing mechanism, reaction disk,
reaction tank, mixing mechanism, cuvette cleaning mechanism, and photometer; operation software is installed
in computer or touch screen, it drives and controls analysis unit to perform all analysis actions.
1.1.4Scope of application
It is applicable to biochemical detection on human body serum, plasma and urine specimen performed by
medical institution.
Note:
This operation manual may be modified without prior notice.
1.2.1Symbol
Symbols listed below are applicable to this operation manual.
Symbol Meaning
-5-
User Manual of BA-A-S120Mini Analyzer
In vitro diagnosis
~ Alternating current
Power on
Power off
PURIFIED WATER
Purified water inlet
Purified water inlet
1.2.3Picture
All pictures in this operating manual are for instruction or example only but not for other purposes.
-6-
User Manual of BA-A-S120Mini Analyzer
1.4Training
For proper use of BA-A-S120Mini automatic chemistry analyzer and give full play to its performance,
our company holds free user training regularly.
For training schedule, please visit our company website or contact local office.
Note: this operation manual may be revised without prior notice which may cause slight difference from the
real instrument. Specific parameters and pictures should subject to real object and operating software.
1.5.5 Absorbance
It refers to the common logarithm obtained from taking the negative number after the transmitted light
intensity divides the incident light intensity. In the SL 120, the incident light intensity is the AD value obtained
when the reaction cup is filled with distilled water, and the displayed absorbance value is the calculated
absorbance value ×104.
-7-
User Manual of BA-A-S120Mini Analyzer
1.5.8 Calibration
Also refers to calibration. For the sample (also known as “calibrator”) with one or more known concentration
(or activity), measure its reaction magnitude; by depending on the selected calibration mode (linear or
nonlinear), carry out fitting by using an optimal curve-paired data set (concentration, reaction magnitude), and
calculate the mathematical expression of this curve. By using this curve, measure the reaction amplitude of the
sample with an unknown concentration (or activity), which can calculate the concentration (or activity) of this
sample.
-8-
User Manual of BA-A-S120Mini Analyzer
1.6 Safety
Electric shock
(1). When power on, unauthorized maintenance personnel are not allowed to open
instrument panel, side cover or rear cover;
(2). Do not spill liquid on the table. If liquid flows inside the instrument, please power off
immediately
(3). High voltage maintained inside computer and printer. Do not touch the inside of
computer or printer.
Injury
(1). Please pay attention to sharp parts of the analyzer, such as reagent tip, sample tip,
mixing rod, or steel tube of automatic cleaning mechanism, otherwise it may cause
injury;
(2). When analyzer is in movement, do not touch its moving parts, such as sample tip,
reagent tip, mixing rod, automatic cleaning mechanism, fan etc.
Hot
(1). Before replacing bulb, please turn off power switch and wait for at least 30 minutes
until bulb cools down totally;
(2). Touching the printer head or metal around the printer head may cause burn injury.
Strong light
(1). Do not look at light beam of bar code scanner directly. These light beams will cause
injury to the eyes;
(2). Do not look at the bulb light directly. These light beams will cause injury to your
eyes.
Biological contamination
(1). All test samples, calibrated products and quality control products are considered
infectious. Wear gloves when you are handling it;
(2). All waste liquid, reagent bottles, overdue reagent and disposable test tubes will be
considered infectious. Discharge or disposal of such object should follow local
sanitation and environmental protection and administration rules, laws and
regulations.
(3). All objects exposed to test samples such as reagent tip, sample tip, mixing rod,
cleaning mechanism, waste liquid tube and bucket will be considered infectious.
Wear gloves while handling them.
Corrosion
(1). Cleaning agent and some reagents have strong acidity or alkaline. Do not touch
with hands or clothes. Once exposed to such material, clean using water and soap;
(2). If it gets into eyes, rinse thoroughly with clean water and consult
ophthalmologist.
-9-
User Manual of BA-A-S120Mini Analyzer
Burn
(1). Do not use flammable hazardous article around analyzer, such as ethyl alcohol
and diethyl ether.
Explosion
(1). Do not use explosive articles around analyzer.
Precautions
Scope of application
(1). BA-A-S120Mini automatic chemistry analyzer is applicable to chemistry
detection on human body serum, plasma and urine specimen performed by
medical institution. When making clinical decisions based on test result, please
correlate with clinical conditions or other test results.
Operator
(1). BA-A-S120Mini automatic chemistry analyzer should only be operated by
personnel trained
Application environment
(1). Please install in an environment designated by the operating instruction.
Installation or use of the instrument in other conditions may produce unreliable
results and cause damage to the analyzer;
(2). If you need to change operating environment of the analyzer,
please contact local distributor.
Electromagnetic interference
(1) Analyzer will receive electromagnetic interference during operation, which may
affect measurements and cause malfunction. Do not use devices which may
produce electromagnetic wave during use, such as electric drill, mobile phone and
walkie-talkie near the analyzer.
(2) During operation, analyzer will send electromagnetic wave externally. Do not
install or use electromagnetic sensitive device around the analyzer.
Imperfect earth
(1). Power supply must be earthed correctly, otherwise it may cause electric shock;
(2). Ground impedance must be smaller than 10Ω. Imperfect earthing will cause
inaccurate measurements or leakage and result in electric shock.
Liquid leakage
(1). Before testing, please check each pipe joint carefully for any leakage. Liquid
leakage will cause inaccurate absorption or discharge amount;
(2). Do not place reagents or samples on analyzer table to prevent liquid spill or
- 10 -
User Manual of BA-A-S120Mini Analyzer
leakage.
Blockage
(1). Check reagents and samples carefully. There should be no insoluble floats, such
as cellulose and fibrous protein, otherwise it may cause reagent/sample blockage.
Water quality
(1). Water quality should meet the requirements below, otherwise it may cause
damage to valve or pump, or unable to clean.
Particle diameter smaller than 200 micrometer;
Electrical resistivity higher than 0.5MΩcm;
Clump count fewer than 10cfu/ml;
Dissolved silicate fewer than 0.1mg/L.
Analysis parameter
(1). Incorrect analysis parameter will cause wrong measurements.
Please consult Bioevopek Ltd.
Attention:
BA-A-S120Mini automatic chemistry analyzer meets with requirement on emission and noise immunity
in GB/T 18268.26, see table below.
User is responsible to make sure electromagnetic compatibility environment of the device so that device
functions well.
We suggest assess electromagnetic environment of the device before use.
Warning:
BA-A-S120Mini automatic chemistry analyzer is designed and detected according to category A device in
GB 4824. In domestic environment, this device may cause radio interference. Protective measures should
be taken.
Do not use this device nearby strong radiation source (e.g., unshielded RF source), otherwise it may
disturb device function.
Table 1:
- 11 -
User Manual of BA-A-S120Mini Analyzer
Electromagnetic emission
Table 2:
Radio frequency
GB/T 17626.3 3V/m, 80MHz~2.0GHz, 80%AM A
electromagnetic field
Power line:±1kV(5/50ns,5kHz)
Pulse group GB/T 17626.4 B
I/0 signal line: ±0.5kV(5/50ns,5kHz)
Line-to-earth: ±2kV
Surge GB/T 17626.5 B
Line-to-line: ±1kV
Power frequency
GB/T 17626.8 3A/m, 50/60Hz A
magnetic field
1 period 0%; B
Performance determination:
A. During test, device functions well within standard limit.
B. During test, function or performance compromises or loses but it recovers automatically.
C. During test, function or performance compromises or loses. It requires operator to intervene or reset
system.
- 12 -
User Manual of BA-A-S120Mini Analyzer
- 13 -
User Manual of BA-A-S120Mini Analyzer
- 14 -
User Manual of BA-A-S120Mini Analyzer
2. Click “Test settings” in the main menu, and select “Calibration settings” from the drop-down menu to
display the calibration setting interface;
3. Click “New” to enter the following contents according to the meanings of parameter items: name, the
standard value of items, and position on the calibration tray;
4. Click “Save”.
- 15 -
User Manual of BA-A-S120Mini Analyzer
tray.“Release a single reagent position” can be activated only by right-clicking on an occupied position.
Note:
If a reagent tray needs to be changed over, it can be performed only after testing on
the current tray is finished.
(4) Click the “Reagent blank” column. Select a reagent to be tested and make a confirmation (Clicking
once indicates selection; clicking once more indicates cancellation). When “Automatically test a
reagent blank” is checked, the reagent blank can be automatically tested only once a day. If reagent
blank test results of this item are not ideal, you can use the button “Reagent blank test>>” to select a
test item that will be used as a reagent blank, and perform reagent blank testing on the selected test
item.
(5) Click the button “OK”.
Note:
If the pail icon at the top left corner of the main interface becomes red and
constantly flashes during testing, add purified water in the purified water pail
immediately, otherwise it will lead to abnormal results.
- 16 -
User Manual of BA-A-S120Mini Analyzer
Note:
If a micro cup is used, the liquid level of serous must be ensured to be not less than
8mm, otherwise it is liable to liquid level detection errors, affecting calibration
results.
(5) Recheck reagent positions, and ensure that all reagent bottle caps are open;
(6) Click the shortcut button “Start” in the main interface. Enter start conditions in the displayed interface;
(7) Select reagent tray numbers (1-5) to be tested;
Note:
If a reagent tray needs to be changed over, it can be performed only after testing on
the current tray is finished.
(8) If you want to need to automatically test a reagent blank, tick before “Automatically test a reagent
blank”, otherwise proceed to the next step;
Note:
Automatically performing reagent blank testing may eliminate the effect of reagent
change on results to some extent. This function is equivalent to the blank calibration
of the Hitachi and Olympus machine types.
(9) If you do not select “Automatically test a reagent blank” but want to separately measure a reagent
blank of an individual item, click the “Reagent blank” column. Select a reagent to be tested and make a
confirmation (Clicking once indicates selection; clicking once more indicates cancellation);
Note:
For some unstable reagents such as ALP, AMY, GGT and Cr, it is recommended to
test their reagent blanks once a day.
(10) Click the “Calibration test” column (Clicking once indicates selection; clicking once more indicates
cancellation);
(11) Click the button “OK”.
Note:
If the pail icon at the top right corner of the main interface becomes red and
constantly flashes during testing, add purified water in the purified water pail
immediately, otherwise it will lead to abnormal results.
Note:
Calculation items are incorporated in item combinations. The system can
automatically calculate corresponding calculation items such as globulin only after
item combinations such as liver function are clicked.
- 17 -
User Manual of BA-A-S120Mini Analyzer
3. Repeat the previous step until all samples are applied for. After application is finished, return to the
main interface.
4. Click the sample tray. In the displayed new interface, click samples on the sample tray. Carefully check
sample positions as well as items applied for to ensure consistency with the checklist.
Note:
(1) If you want to delete a sample, after clicking this sample in the sample tray
interface, right-click and then click “Delete a single sample”.
(2) If the application for a sample is in error or you want to increase or reduce test
items of a sample, delete this sample first. After returning to the main interface,
click the shortcut button “Test application” or “High-speed test” at the top left
of the main interface (It is necessary to be consistent with the last application
status of the deleted sample). Enter the sample code of the deleted sample in the
test application interface, and just reapply.
Note:
(1) If a blood sampling tube is directly used, check carefully and ensure that serum
has been completely separated and there is no fibrous protein in serum,
otherwise the needle is liable to blockage.
(2) If a micro cup is used, the liquid level of serous must be ensured to be not less
than 8mm, otherwise it is liable to liquid level detection errors, affecting
calibration results.
6. Recheck reagent positions, and ensure that all reagent bottle caps are open;
7. Click the shortcut button “Start” in the main interface. Enter start conditions in the displayed interface;
(1) Select reagent tray numbers (1-5) and sample tray numbers (1-4) to be tested.
Note:
If a reagent or sample tray needs to be changed over, it can be performed only after
testing on the current tray is finished.
(2) If you want to need to automatically test a reagent blank, tick before “Automatically test a reagent
blank”, otherwise proceed to the next step;
Note:
Automatically performing reagent blank testing may eliminate the effect of reagent change
on results to some extent. This function is equivalent to the blank calibration of the
Hitachi and Olympus machine types.
(3) If you do not select “Automatically test a reagent blank” but want to separately measure a reagent
blank of an individual item, click the “Reagent blank” column. Select a reagent to be tested and make a
confirmation (Clicking once indicates selection; clicking once more indicates cancellation);
Note:
For some unstable reagents such as ALP, AMY, GGT and Cr, it is recommended
to test their reagent blanks once a day.
(4) If quality control testing is included in this lot of tests, click the “Quality control” column (Clicking
- 18 -
User Manual of BA-A-S120Mini Analyzer
Note:
If the pail icon at the top left corner of the main interface becomes red and constantly
flashes during testing, add purified water in the purified water pail immediately, otherwise
it will lead to abnormal results.
Note:
(1) If you want to delete a quality control test, after clicking this quality test on the
sample tray interface, right-click and then click “Delete a single sample”.
(2) If the application for a quality control solution is in error or you want to increase
or reduce test items of a quality control, delete this quality control test first. After
returning to the main interface, click the shortcut button “Test application” or
“High-speed test” at the top left of the main interface (It is necessary to be consistent
with the last application status of the deleted sample). Just reapply in the test
application interfce.
5. Place the quality control solution in the sample disk according to the applied position.
Note:
If a micro cup is used, the liquid level of serous must be ensured to be not less than
8mm, otherwise it is liable to liquid level detection errors, affecting calibration
results.
- 19 -
User Manual of BA-A-S120Mini Analyzer
6. Recheck reagent positions, and ensure that all reagent bottle caps are open;
7. Click the shortcut button “Start” in the main interface. Enter start conditions in the displayed interface;
(1) Select the number (1-5)of reagent disks and the number (1-4)of sample disk that intended to start test;
Note:
If a reagent tray needs to be changed over, it can be performed only after testing on
the current tray is finished.
(2) If you need the reagent blank can be tested automatically, please tick in the box in front of
“Automatically test reagent blank”, otherwise directly enter into next step.
Note:
Automatically performing reagent blank testing may eliminate the effect of reagent
change on results to some extent. This function is equivalent to the blank calibration of the
Hitachi and Olympus machine types.
(3) If you do not select “Automatically test reagent blank” and want to test the reagent blank of individual
item, click “Reagent Blank” and select the reagent intended to test, then click “OK” (clicking one time
for selected, and one time again for cancel);
Note:
For some unstable reagents such as ALP, AMY, GGT and Cr, it is recommended
to test their reagent blanks once a day.
(4) Click the “Quality control test" column (Clicking once indicates selection; clicking once more indicates
cancellation);
(5) If general testing is included in this lot of tests, click the “General test” column (Clicking once
indicates selection; clicking once more indicates cancellation). Enter an initial number of a desired
general test in the following input box. In case of no input, the system starts all general tests by default;
(6) Click the button “OK”.
- 20 -
User Manual of BA-A-S120Mini Analyzer
Note:
f the pail icon at the top right corner of the main interface becomes red and constantly
flashes during testing, add purified water in the purified water pail immediately, otherwise
it will lead to abnormal results.
- 21 -
User Manual of BA-A-S120Mini Analyzer
5.2 Power-on
Turn the power on in the order below:
1. Press the “Power” button on the right-hand side of the analyzer(The power indicator light will be on);
2. Power up the display;
3. Power up the main unit;
4. Power up the printer.
- 22 -
User Manual of BA-A-S120Mini Analyzer
- 23 -
User Manual of BA-A-S120Mini Analyzer
information on the interface. Then click the “save” button in the upper left-hand corner.
d) To redo the test of a item, select the “Re-test” option on the right-hand side of the results bar(click
to select, and click again to de-select). Or you may click “select all” to select all the tested items of
the sample, and then click “Start and Select Re-test”.
Note:
In case of serious errors during the test which requires emergency exit, you may turn
the power off on the left-hand side of the analyzer, and click the shortcut
“Emergency Exit” button on the main interface.
5.9 Power-off
Power off in the order below:
1. The instrument;
2. The printer;
3. The computer;
4. The display.
- 24 -
User Manual of BA-A-S120Mini Analyzer
Login user name The user name to log in the Select in the drop-down
instrument system list after it.
- 25 -
User Manual of BA-A-S120Mini Analyzer
Serial port selection The serial port number of the PC Select in the drop-down
connected to the instrument list after it. Select the
serial port number as it
is indicated on the
computer
Parity check bit The data check bit of Select in the drop-down
communication data in the frame list after it.”No check” is
format used in this system.
- 26 -
User Manual of BA-A-S120Mini Analyzer
Length of sending Length of the Send Buffer Zone Type in the square frame
area (number of bytes) during data after it. The default
transmissions value is “512”
Length of receiving Length of the Receive Buffer Type in the square frame
area Zone (number of bytes) during after it. The default
data transmissions value is “4096”
1 Adjust some of the above parameter items according to the actual parameters;
2 Click the “Default value” button to change all parameters to the default
values, and click the “Confirm” button to complete the changes to
parameters. Click the “Cancel” button to keep the parameters unchanged.
- 27 -
User Manual of BA-A-S120Mini Analyzer
- 28 -
User Manual of BA-A-S120Mini Analyzer
Automatic backup file The backup will take place when Designate the path for
path: the operational software is run automatic backup by
for the first time each day. means of selecting in the
Therefore, the path for “drive” and “directory”
automatic backup needs to be lists.
designated, and is preferably
located on other than system
drive. As in the case of system
failure, which requires
re-installation of the system, the
backup data are lost and cannot
be restored.
Manual backup file The manual backup allows for Designate the path for
path selective backup and restoration automatic backup by
of data. Therefore, the path for means of selecting in the
manual backup needs to be “drive” and “directory”
designated, and is preferably lists.
located on other than system
drive.As in the case of system
failure, which requires
re-installation of the system, the
backup data are lost and cannot
be restored.
1 In the area of automatic backup path, select the “drive” and “directory” for
automatic backup data.
2 Click the “Confirm” button to store the designated path in the configuration
file.
1 In the area of manual backup path, select the “drive” and “directory” for
manual backup data.
2 Click the “Confirm” button to store the designated path in the configuration
file.
- 29 -
User Manual of BA-A-S120Mini Analyzer
1 Click “Restore automatic backup data”, and confirm the restoration. Then the
backed-up data will be restored into the existing system. The progress is
indicated on the progress bar during the restoration.
Manual backup
1 Select the “Backup Start Date” and “Backup End Date” in the Manual
Backup Timeframe.
2 Click “Manual backup data”, and confirm the backup. Then data during the
designated timeframe will be backed up in the designated path. The progress
is indicated on the progress bar during the backup.
1 Select the “Backup Start Date” and “Backup End Date” in the Manual
Backup Timeframe.
2 Click “Restore manual backup data”, and confirm the restoration. Then the
backed-up data during the designated timeframe will be restored into the
existing system.
The progress is indicated on the progress bar during the restoration.
- 30 -
User Manual of BA-A-S120Mini Analyzer
Total Staff The total number of the hospital Type in the square frame
staff
Total Departments The total number of hospital Type in the square frame
departments
1 Click “Change”;
- 31 -
User Manual of BA-A-S120Mini Analyzer
Department information
Total staff Total number of department staff Type in the square frame
- 32 -
User Manual of BA-A-S120Mini Analyzer
1 Click “Add”;
3 Click “Save”.
2 Click “Add”;
Doctor information
- 33 -
User Manual of BA-A-S120Mini Analyzer
1 Select the department to which the doctor belongs in the drop-down list;
- 34 -
User Manual of BA-A-S120Mini Analyzer
2 Click “Add”;
4 Click “Save”.
1 Select the department to which the doctor belongs in the drop-down list;
2 Select in the doctor list the doctor of which the information is to be changed;
3 Click “Add”;
1 Select the department to which the doctor belongs in the drop-down list;
3 Click “Add”;
- 35 -
User Manual of BA-A-S120Mini Analyzer
1 Click “Add”;
3 Click “Save”;
1 Click the “〉〉” or “〈〈” button till the user name to be changed is shown on
the interface;
2 Click “Add”;
1 Click “〉〉” or “〈〈” till the user name to be deleted is shown on the interface;
2 Click “Add”;
- 36 -
User Manual of BA-A-S120Mini Analyzer
1 Click “〉〉” or “〈〈”till the user name to which permissions are to be set is
shown on the interface;
2 Set the operational permissions in the general setting group. Click to select,
and click again to de-select;
- 37 -
User Manual of BA-A-S120Mini Analyzer
User name Name used to log in Choose a user name from the
dropdown menu.
2 Input the original registration password in the original password input box;
4 Input the new password again in the confirmation new password input box;
- 38 -
User Manual of BA-A-S120Mini Analyzer
1 Choose the new user name which you would like to switch to from the
dropdown menu ;
2 Input the registration user password in the user password input box;
- 39 -
User Manual of BA-A-S120Mini Analyzer
2 Click “Move Forward”,then print sequence of the item will move forward 1
bit, click“move forward”continuously , until the item reach its set print
sequence;
- 40 -
User Manual of BA-A-S120Mini Analyzer
2 Click “Move backward” , then the print sequence of the item will move
backward for 1 bit, click“backward” continuously, until the item reach its set
print sequence;
Unit name Unit name of the item Type in the square frame
1 Click “Add”;
3 Click “Save”.
- 41 -
User Manual of BA-A-S120Mini Analyzer
1 Choose the Item unit you would like to amend in the item unit list;
2 Click “Add”;
4 If you would like to save the change , click “ save” ; otherwise , click
“backspace”;
1 Choose the item unit you would like to delete in the item unit list;
2 Click “Add”;
- 42 -
User Manual of BA-A-S120Mini Analyzer
2 Click “ move forward” button, then the item test sequence move forward for
1 place, click “Move forward” continuously until the item reach its set test
sequence;
2 Click “move backward”, then the item test sequence move backward for 1
bit, click “Move backward” continuously until the item reach its set test
sequence;
- 43 -
User Manual of BA-A-S120Mini Analyzer
Select export Three kinds of export form for User can choose one export form
form the test result data : in among three forms as needed.
clipboard, 、 TXT file, EXCEL
file.
Exportable The data item list which can be Select one Data item in the “item
Item exported for normal sample can be exported”,configure it to
test result data, the export item list with the
middle button“>>”.
Derived date For normal sample test result Choose one data item in
item and data , which has been the“exported item data”list, and
order configured to the data item list you can cancel this exported item
need to be exported, with middle button “<<”, you can
also change the sequence of the
data item with “move upward”
and “move downward” button in
the middle.
Configure the export method, export data item and exported array sequence.
1 User can choose one export form among three forms as needed.
2 Select one Data item in the “item can be exported”,configure it to the export
item list with the middle button“>>”.
3 Select one data item in the “export item data”list and you can cancel this
export item with middle button“<<”.
4 The data item export sequence can be changed by “move upward” and “move
downward” buttons.
- 44 -
User Manual of BA-A-S120Mini Analyzer
Full name Item full Chinese name Type in the square frame
Decimal digits The reserved decimal digits for Dropdown frame,and select from 0,1,2,3.
the test result
Result unit Test result unit Select from the dropdown frame
ItemUnit Set item unit online. The item unit set interface popup after clicking.
settings
Test method Set up the item test method dropdown frame, select from “destination
method”, “two points method”, and “ dynamics
method”.
Direction The change direction of Dropdown frame, select from ascend and
absorbance during reaction. descend.
- 45 -
User Manual of BA-A-S120Mini Analyzer
Reference Tested secondary wave length Dropdown frame , Select from the remained
wave length wave length.
Sample volume The sample volume for normal Enter in to the square frame directly , range is
testing, the unit is microlitre. 3-30,0.1 increase progressively
More Set up the added sample quantity New interface popup after clicking,enter in to
for different type sample the square frame directly.
concentration and dilute test.
R1 volume The 1st reagent quantity when Enter in to the square frame directly. The range
testing,the unit is microlitre. is 150-300,1 increase progressively .
R2 volume The 2nd reagent quantity when Enter in to the square frame directly. The range
testing,the unit is microlitre. is 20-150,1 increase progressively .
Reaction time The starting metering point for Enter in to both adjacent square frame directly,
calculation the numerical value in first frame is , L,the
numerical value in second frame is M,
So L and M have to meet below relational
expression:
(1) Single reagent Destination method 1≤L <
13≤M
(2) Double Reagent destination method 14≤L
<26≤M
(3) Single reagent” “Two points method” and
“Dynamics method”
14≤L<M
(4) Double reagent “Two points method” and
“Dynamics method”
35≤L<M
Display after clicking:
Help Display the purpose of each
metering point and time interval
online.
Calibration Set up item calibration mode Dropdown frame , select from linear,Log-4P,
mode Log-5P,Exponentional-5P 、 Polynominal-5P ,
spline and factors.If you select “ factors”, then
enter in the specific numerical value in the
square frame.
Detailed setting If the calibration method is not” New interface popup after clicking:
factors”, then the corresponding
button will show up, and the
detailed parameters of the
Enter in the specific numerical value to the input
calibration curve can be set after
- 46 -
User Manual of BA-A-S120Mini Analyzer
Slope Calibrate the result according to Enter in the specific numerical value to the
formula y=ax+b, therein x is the square frame, default is 1.
actual measured result , y is the
Intercept calibrated result,a is the slope in Enter in specific numerical value to the square
the calibration formula,b is the frame, default is 0.
intercept in the calibration
formula.
Linearity limit Decide the reaction curve Enter in numerical value between 0-100.
linearity degree.
Exhaust limit The set decision limit for Enter in numerical value between 0-40000.
substrate exhausted during
reaction , is absorbance value
times 10000.
Prozone check Set up whether proceed front belt radio , selected means proceed front belt
examination or not, and the time examination;
starting point and examination
limit; The next three frames are activated after it is
selected, the first frame means the front
Front belt examination can only examination time starting point, whose value is
be proceeded with “destination N; the second frame means the front
method”; examination time stop point, whose value is P;
the third frame means the front examination
limit, whose value is PC.N&P and reaction
time L&M have to meet below relational
expression:
Single reagent “Destination method” 1≤L <
13≤N<P<M
Double Reagent “Destination method” 14≤L <
26≤N<P<M
PC is integer between 0-100.
Reference Enter in the default normal Enter in specific numerical value in both
range reference range adjacent square frames.
Detailed After clicking, enter in different New Interface popup after clicking:
gender and differnt type sample,
and normal reference range of
different age group. Newly added reference range
(1) Select the gender and sample type from
the dropdown frame;
(2) Click the age upper and lower limit for
the age and reference range;
(3) Click button “save”.
- 47 -
User Manual of BA-A-S120Mini Analyzer
Linearity range Input the item linear range Input the specific numerical value in both
adjacent square frame .
Reaction Input the amplitude range of the Inputthe specific numerical value in both
amplitude range item adjacent square frame, range is 0 to 40000
R1 absorbance Input the normal range of the 1st Input the specific numerical value in both
Range reagent absorbance. adjacent square frame, range is 0-40000
Working Input the normal range of the Input the specific numerical value in both
solution absorbance. adjacent square frame, range is 0-40000
absorbance
range
1 Click “Add”;
2 Click “Correction”;
2 Click”Delete”;
- 48 -
User Manual of BA-A-S120Mini Analyzer
Name Input the name of the calibration Directly input in the square frame
solution.
Position Set the position of the calibration Click the related position on the left calibration
solution on the calibration panel. panel.
Item Set the concentration of each item Directly input in the square frame
Concentration of the calibration solution.
1 Click”New addition”;
2 Click “Correction”;
3 Input related information in each aquare frame. To change the position of the calibration
solution, click the target position (it must be empty), and click "Save".
- 49 -
User Manual of BA-A-S120Mini Analyzer
2 Click “Delete”;
3 Click "Delete" to delete the calibration solution. If you want to delete it, click "Yes"; otherwise,
click "No".
Adding a QC solution
1 Click “Add”;
- 50 -
User Manual of BA-A-S120Mini Analyzer
Correcting a QC solution
2 Click “Change”;
Deleting a QC solution
2 Click “Delete”;
Full name The full Chinese name of the item Type in the square frame
calculated.
Result unit Calculate the result unit of the Select an option from the drop-down list frame.
- 51 -
User Manual of BA-A-S120Mini Analyzer
Decimal Keep decimal of the result of the Dropdown frame,and select from 0,1,2,3.
item calculated.
Reference Input the default reference range. Enter in specific numerical value in both
range adjacent square frames.
Detailed After clicking, enter in different New interface popup after clicking:
gender and differnt type sample,
and normal reference range of
different age group. Newly added reference range
(1) Select the gender and sample type from
the dropdown frame;
(2) Click the age upper and lower limit for
the age and reference range;
(3) Click button “save”.
Correction Reference Range
(1) Select the reference range to be changed
in the reference range;
(2) Click “change” for the age upper and
lower limit and reference range;
(3) Click button “save”.
Delete reference range
(1) Select the reference range to be deleted
from the reference range;
(2) Click “Delete”;
(3) To delete the reference range, click "Yes".
Otherwise, click "No".
1 Click “Add”;
3 In the item selection zone, click the item to be selected. In the calculation button zone, click a
value and a calculation symbol to compose a formula for calculating the item.
4 To save the added calculation item, click "Save". Otherwise, click "Cancel".
1 In the calculation item list display zone, select the calculation item to be corrected.
2 Click “Change”;
- 52 -
User Manual of BA-A-S120Mini Analyzer
4 To modify the calculation formula, click "Rebuild Expression", and input a new expression.
1 In the calculation item list display zone, select the calculation item to be deleted.
2 Click “Delete”;
No. Number of the item combination System automatically generated, can not be
changed
Full name The full Chinese name of the item Type in the square frame
calculated.
1 Click “Add”;
- 53 -
User Manual of BA-A-S120Mini Analyzer
2 In each square frame, input related information. In the item selection zone, click an item. The
item is selected by one click, and it is deselected by another click.
3 To save the added item combination, click the "Save". Otherwise, click "Cancel".
1 In the item combination list display zone, select the item combination to be corrected.
2 Click “Change”;
3 In each square frame, input related information. In the item selection zone, select or deselect an
item.
1 In the item combination list display zone, select an item combination to be deleted.
2 Click “Delete”;
- 54 -
User Manual of BA-A-S120Mini Analyzer
Sample type A list of existing sample types You can select a sample type
list record in the list, and correct or
delete it as required.
1 Click “Add”;
3 Click “Save”;
- 55 -
User Manual of BA-A-S120Mini Analyzer
2 Click “Change”;
4 If you would like to save the change , click “ save” ; otherwise , click
“backspace”.
2 Click “Delete”;
6.4 Statistics/Query
6.4.1 History Query
Query the history test result of a regular sample.
Test date Start date of the samples to In the square frame, directly input a date, or
- 56 -
User Manual of BA-A-S120Mini Analyzer
Sample No. Start code of the samples Type in the square frame
to be queried
Sex Patient gender of the Select an option from the drop-down list box.
samples to be queried
Age unit Age unit of the samples to Select an option from the drop-down list box.
be queried
Bed No. Hospital bed code of the Type in the square frame
samples to be queried
Docimaster Inspection doctor of the Select an option from the drop-down list box.
samples to be queried
- 57 -
User Manual of BA-A-S120Mini Analyzer
1 In the query conditional input zone, input or select one or more query conditions.
3 Click “query”.
4 In the query result display zone, all the records meeting the query conditions are
displayed.
5 Select a query record, and click “Reaction curve”. You can see the real-time reaction
curve of this record.
6 If you only query one item, the “Profile” button is active. You can click it to view
the distribution of all query records.
7 Click “Recalculation”, the software will automatically recalculate and display all the
results in the query result list.
8 Click “Export”, and zccording to the configuration export method and format
mentioned in Section 9.2.8, the system will export all the results in the query result
list.
- 58 -
User Manual of BA-A-S120Mini Analyzer
Viewing QC status
1 In the QC type selection zone on the left, select a desired QC type from real-time QC, intra-day QC,
and inter-day QC.
3 Click “query”, in the QC solution display zone, all the used QC solutions are displayed. In the QC
status display zone, the QC status of the item is displayed.
4 Click “Detailed QC Result”, the test results of each QC solution within the selected time period are
displayed
5 Click “Multi-rule QC Chart”, “Accumulated QC Chart”, and “twin plot” , the related QC charts of
the item within the selected time period are respectively displayed.
8 Click “Return” to close the QC status interface and return to the main interface.
- 59 -
User Manual of BA-A-S120Mini Analyzer
2 In the query item selection bar, select an item to be queried. The item is selected by one click, and
it is deselected by another click.
3 Click “query”, in the query result display zone on the right, all the calibration records meeting the
query conditions will be displayed.
5 Click “Details” to display a new interface, where the detailed information about the calibration
record is displayed.
6 To print the details of a calibration record, select it, and click "Print".
7 To set a calibration record as the current record, select it and click “Set to Current”. Then, the
concentration values of later tested samples of this item will be calculated according to the current
calibration record.
2 Click “Details” button to display the detailed information about the calibration, shown as below:
- 60 -
User Manual of BA-A-S120Mini Analyzer
Operator Operator of working volume statistics Select from the drop-down list
Sample type Sample type of working volume statistics Select from the drop-down list
Time Time of working volume statistics Click an option to select it, or directly
input a value.
All items If selected, it indicates the number of all The option is selected by one click, and
items in the statistics. it is deselected by another click.
1 From the drop-down list box, select an operator and a sample type.
3 To query all items, click the “All Items” radio. Otherwise, select an item in the query item
selection zone.
4 Click “query”.
5 In the query result display zone on the right, all the results meeting the query conditions are
displayed.
- 61 -
User Manual of BA-A-S120Mini Analyzer
6.5 Maintenance
6.5.1 Instrument Debugging and Maintenance
The maintenance scope includes the main control unit, sampling probe unit, stirring rod unit, reaction tray
photoelectric unit, and reagent sample tray unit. They are respectively described as follows:
Handshake instruction Communication handshaking between the PC and each unit chip
2 To send cyclically, select "Send Cyclically", and input the interval and number of times in the
subsequent square frames. Then click "Send".
- 62 -
User Manual of BA-A-S120Mini Analyzer
3-R1 unit
- 63 -
User Manual of BA-A-S120Mini Analyzer
Move downward to init bit The sampling probe moves downwards to the init bit.
Downward movement to the The sampling probe moves downwards to the cleaning bit.
cleaning position
Down to the reagent bottle or The sampling probe moves downwards to the reagent bottle or sample
sample cup cup, and stops until it detects the liquid level.
Sampling probe dips into The sampling probe moves downward into the reaction vessel, and
reaction vessel moves a specific number of steps.
Move downward a specific The sampling probe moves downward a specific number of steps.
number of steps
Move downward to draw The sampling probe moves downward, draws a sample, and then
sample (and back again after move upwards again.
drawing)
Move downward to reaction The sampling probe moves downward, emit a sample, and then move
vessel to emit sample upwards again.
Down to cleaning pool for The sampling probe dips into the cleaning pool for cleaning.
cleaning
Reagent residue check The sampling probe checks the remaining volume of the reagent.
command
Injector to init bit The injector moves to the initial position.
Injector draws sample The sample injector draws a sample of a preset volume.
Injector emits sample The sample injector emits a sample of a preset volume.
Always open sampling probe Open the sampling probe external wall valve.
external wall valve
Close sampling probe external Close sampling probe external wall valve
wall valve
Open/Close sampling probe Open or close the sampling probe external wall valve at a given time.
external wall valve by time
Always open sampling probe Open the sampling probe inner wall valve.
inner wall valve
Close sampling probe inner Close sampling probe inner wall valve
wall valve
Open/Close sampling probe Open or close the sampling probe inner wall valve at a given time.
inner wall valve by time
- 64 -
User Manual of BA-A-S120Mini Analyzer
4-R2 unit
R2 Probe vertical reset The stirring rod returns to the init bit in vertical direction.
R2 syringe vertical reset The stirring rod returns to the init bit in horizontal direction (init bit –
stirring bit of the reaction plate - cleaning bit)
R2 probe rotating reset The stirring rod moves to the init bit in rotation direction.
R2 tray rotating reset The stirring rod rotates to the position of the stirring rod cleaning pool.
R2 tray rotate to certain R2 The stirring rod rotates to the upper stirring bit of the reaction plate, and
Pos. dips a specific number of steps to the cuvette.
R2 probe rotate to R2 tray The stirring rod moves vertically to the init bit.
R2 probe rotate to reaction The stirring rod moves downwards to the position of the cleaning pool.
tray
R2 probe rotate to cleaning The stirring rod moves downwards into the reaction vessel.
sation
- 65 -
User Manual of BA-A-S120Mini Analyzer
R2 probe down to cleaning The stirring rod moves downwards a specific number of steps.
station
R2 probe down to reagent The stirring rod moves downwards into the reaction vessel (and back
bottle again after stirring).
Dip to the cleaning pool The stirring rod dips to the position of the cleaning pool (and back again
(and back again after after cleaning).
cleaning)
Always enable the stirring Open the stirring rod motor to let the stirring rod stir.
command
Disable the stirring Close the stirring rod motor to stop the stirring rod.
command
Enable/Disable stirring by Open the stirring rod motor and set the stirring time.
time
Always open the stirring rod Open the stirring rod external wall valve, and clean the external wall of
external wall valve the stirring rod.
Close the stirring rod Close the stirring rod external wall valve, and stop cleaning the external
external wall valve wall of the stirring rod.
Open/Close the stirring rod Open the stirring rod motor, and set the time for cleaning the external
external wall valve by time wall of the stirring rod.
2 To send cyclically, select “Send Cyclically”, and input the interval and number of times in the
subsequent square frames. Then click “Send”.
- 66 -
User Manual of BA-A-S120Mini Analyzer
Rotation reset The reaction tray goes through the init bit in a direction and stops at the
first cuvette.
Reaction tray rotates through The reaction tray rotates through the init bit to specific cup position.
init bit to specific cup position
Reaction disk directly rotation The reaction tray rotates from the current position directly to the
to the specified cup position specified cup position.
Reaction disk rotation to the The reaction tray rotates from the current position directly to the static
static collection cup position collection cup position.
Rotate the reaction disk directly Rotate the reaction disk directly
Stop the reaction disk in the Stop the reaction disk in the direct rotation
direct rotation
Static photo-electricity The reaction tray does not rotate, but collect the photoelectric data of the
collection command preset wavelength.
Liquid open pump Open the liquid pump, and maintain the preset time.
- 67 -
User Manual of BA-A-S120Mini Analyzer
Open liquid refilling pump Open the liquid refilling pump to refill.
Close liquid refilling pump Close the liquid refilling pump to stop refilling.
2 To send cyclically, select “Send Cyclically”, and input the interval and number of times in the
subsequent square frames. Then click “Send”.
5-S unit
- 68 -
User Manual of BA-A-S120Mini Analyzer
Reagent tray rotates to specified cup Reagent tray rotates to specified cup position through init bit
position through init bit
Reagent tray rotates directly to The reagent tray rotates from the current position directly to the
specified cup position specified cup position.
Sample tray rotates to specified cup Sample tray rotates to specified cup position through init bit
position through init bit
Sample tray rotates directly to The sample tray rotates from the current position directly to the
specified cup position specified cup position.
Directly rotate reagent sample disk Directly rotate reagent sample disk motor
motor
Directly stop rotating reagent sample Directly stop rotating reagent sample disk motor
disk motor
6.5.2 Cuvette Maintenance (This function is applicable only to the instrument system with
self-cleaning function)
According to maintenance requirements, conduct a thorough cleaning on all cuvettes. Click the menu to
display the following interface:
Clean all cuvettes: Clean all cups on the cuvette tray one by one.
- 69 -
User Manual of BA-A-S120Mini Analyzer
Wipe all cuvettes: Evacuate and wipe all cups on the cuvette tray one by one.
Fill all cuvettes with water: Fill water in all cups on the cuvette. Starting from cup No.1, fill water in five cups
as a group. Perform the operation 14 times. At last, there are still two empty cups. If necessary, you can
manually fill the two cups with water.
For the three functions including "Clean all cuvettes, wipe all cuvettes, and fill all cuvettes with water", you
can select one or more of them as required. Then click the "Start" button for corresponding processing.
First, select the contents of maintenance as required. Then click the "Start" button. The instrument will start
maintaining the corresponding cuvettes. The process is in this order: clean all cuvettes, wipe all cuvettes, and
fill all cuvettes with water. The process is over until the last maintenance item is completed. If you do not want
to maintain cuvettes, click "Return".
Click the "Y" to start clearing the corresponding records. Click the "N" to abort the operation and return to the
main interface.
- 70 -
User Manual of BA-A-S120Mini Analyzer
Click the "Y" to start transferring the corresponding record. Click the "N" to abort the operation and return to
the main interface.
- 71 -
User Manual of BA-A-S120Mini Analyzer
Start Date Start date of the error In the square frame, directly input a date, or select one
information to be queried from the drop-down list box.
End Date End date of the error In the square frame, directly input a date, or select one
information to be queried from the drop-down list box.
Query error List of error information Move the mouse to a record in the list. The related
information list satisfying the query period information will be displayed.
1 In the query condition input zone, input the start/end time of errors.
3 Click “query”. If any error information satisfies the query conditions, it will be written into
the "Error Information List".
2 Click “Delete Current Error Information”. After confirmation, delete the selected error
information.
1 Click “Clear All Error Information”. After confirmation, delete all the error information in
the database.
- 72 -
User Manual of BA-A-S120Mini Analyzer
necessary to stop the upper computer software. By implementing the "communication connection" function,
you can set up a communication link between the upper computer and lower computer.
- 73 -
User Manual of BA-A-S120Mini Analyzer
1 In the calibration item selection zone on the left, select an item. The item is selected by one click,
and it is deselected by another click.
2 In the item calibration solution request zone, select one or more calibration solutions according to
the hints on calibration types shown below.
3 If you want to perform calibration on multiple items, repeat steps 1 and 2, until all the items are
selected.
4 Click "Request".
5 Click “Calibration Request List”. You can see the detailed calibration request list.
- 74 -
User Manual of BA-A-S120Mini Analyzer
Batch input Input samples of the same test in Input the number of samples inputted in batches.
batches
Sample Select a sample plate code. There Select an option from the drop-down list box.
plate code are four plates in all.
Cup Cup position in a sample plate. Input a value in the square frame, or click "Remaining
position There are 24 positions in all. cups view" and select a cup position from the pop-up
box.
Remaining View the empty positions on the Click this option to display a new interface. Select an
cups view current sample plate. empty cup position, and click "OK".
- 75 -
User Manual of BA-A-S120Mini Analyzer
Sample Type of the selected sample Select an option from the drop-down list box.
type
Emergency Set the current sample as Radio. If selected, it indicates emergency treatment.
treatment emergency treatment
3 If emergency treatment test, please select Eemergency treatment radio, otherwise do not select;
4 In Item test selection area, click the items needed to be tested, with clicking one time for selected,
and one time again for cancel. Or in the combination item selection area, click the combination
set up, with clicking one time for selected, and one time again for cancel. If you need to set up a
combination item online, click “Combination Setup” and setup the combination on the new
interface popped up, and then return.
1 Input the initial sample No, then input the number of batch;
3 In Item test selection area, click the items needed to be tested, with clicking one time
for selected, and one time again for cancel. Or in the combination item selection area,
click the combination set up, with clicking one time for selected, and one time again
for cancel. If you need to set up a combination item online, click “Combination
Setup” and setup the combination on the new interface popped up, and then return.
4 Click "Request".
- 76 -
User Manual of BA-A-S120Mini Analyzer
Note:
(1) When batch application, the number and the position in sample disk of the sample
demanding batch application should be increased progressively with order based on
the number and cup position of the initial sample.
(2) Batch application and a single application can be done at the same time.
1
By means of clicking the button or , the application information of the first or the
last number can be viewed; By means of clicking the button or , the application
information of the prior or the latter number can be viewed;
1
Click the button , , or into the number of samples intended to be
modified;
2 In item test selection area, click the items needed to be tested, with clicking one time for selected,
and one time again for cancel.;
3 Click "Request".
4 For samples that have been applied but not tested, if you want to delete the extra application
items, please move the mouse below the item and press the right key, and then a new interface
will pop up and you just press the “Delete” button. If you want to add the omitted items, please
directly click the item and then just press “Apply for” button. But the samples that have begun to
test cannot be modified.
Note:
Only the number of ordinary and quality control test that have been applied but not started
can be viewed and modified.
6.6.3 Start
On the left toolbar of the main interface, please click “Start” and then the following functional form that starts
test will appear. Select the type and range that intended to start test, and start test after clicking “OK”(see
below picture):
- 77 -
User Manual of BA-A-S120Mini Analyzer
Start test
1 “Cup state” indicates that the service condition of each cup in the reaction disk can be viewed.
When there are no used cups (dirty cups) in the reaction disk, click “Cup change confirm” to tell
the operating system that all cups in the reaction disk are clean. After starting test, the operating
system will start test from the first cup.
2 Select the number (1-5) of reagent disks and the number (1-4) of sample disk that intended to
start test;
3 If you need the reagent blank can be tested automatically, please tick in the box in front of
“Automatically test reagent blank”, otherwise directly enter into next step.
4 If you do not select “Automatically test reagent blank” and want to test the reagent blank of
individual item, click “Reagent Blank” and select the reagent intended to test, then click “OK”
(clicking one time for selected, and one time again for cancel);
5 In the system, the calibration test is independent. That is, if you want to conduct the calibration
test, the calibration test must be completed, then other non-calibration test can be started.
Certainly, during the calibration test, the reagent blank test also can be conducted at the same
time. In the system, the test sequence is: reagent blank test – calibration test—quality control test
– ordinary test. Prior to starting any test, tests related to reagent blank can be added.
5 If the test for this batch includes the quality control test, please click “Quality Control” (clicking
one time for selected, and one time again for cancel).
- 78 -
User Manual of BA-A-S120Mini Analyzer
6 Click “Ordinary Test” (clicking one time for selected, and one time again for cancel), and input
the starting number of ordinary test intended to start in the below input box. If you do not input
anything, it will be defaulted to start all ordinary tests.
Note:
SL 120 system can start the reagent blank test, quality control test and ordinary test at the
same time based on the below priority level.
(1) Reagent blank test
(2) Quality control(QC) test
(3) Ordinary test
Note:
(1) For the reagent blank test, calibration test and quality control test, once they are
selected, the test of all current applied items mentioned above will be started.
(2) For the ordinary test, you can randomly choose the sample number that intended to
start test. If only the starting sample number is input, all tests after the number will
be started; If only the closing sample number is input, all tests before the number
will be started; if no sample number is input, all tests that have applied will be
started.
- 79 -
User Manual of BA-A-S120Mini Analyzer
Sex Sex of the current patient Select an option from the drop-down list box.
Age Age of the current patient The first box can be input directly, and the second box can be
selected from the drop-down menu.
Department Department of the current Select an option from the drop-down list box.
of sample patient
sending
Bed No. Bed No. of the current patient Type in the square frame
Clinical Clinical diagnosis for the Chosen from the drop-down menu or input directly
diagnosis current patient
Submitting Doctor who opens the Input the doctor number or input directly
physicians inspection application form for
- 80 -
User Manual of BA-A-S120Mini Analyzer
Sample type Sample type for inspection Select from the drop-down list
Sample Sample status for inspection Select from the drop-down list
status
Auditor The name of doctor who signs Select an option from the drop-down list box.
audit report form
Delete Delete the information and Click to pop up a new interface and then click “OK”
singly relevant test items of one
sample under current non-test
status
Delete all Delete the information and Click to pop up a new interface and then click “OK”
relevant test items of one
sample under current non-test
status, also can delete sample
data which have finished test
- 81 -
User Manual of BA-A-S120Mini Analyzer
1 Select one sample number which needs to be input patient information from the left sample
number list area;
2 All test results of selected numbers can be displayed directly in the results display area in lower
right. If the test of one item has not been finished, the Result Display Area of the item will display
nothing;
3 After selecting one item that has had result, please click “Reaction Curve” to view the reaction
curve and reaction data of the result.
1 Select one sample number which needs to be input patient information from the left sample
number list area;
2 Input all relevant information in the Information Input Area in upper right and then click “Save”
on the right;
3 Switch to next sample number, and then repeat step 1 and step 2 to complete the information
input of next number.
2 Select the available print template or built a new one, and setup it as default.
Result Print
1 If you want to print all results, please click radio-button “All samples”, then click “Print”;
2 If you want to print the result with its number selected, please click radio-button “Current
Selected Sample”, then click “Print”;
3 If you want to print the results with its number appointed, please click radio-button “Select
Sample”, and input directly the appointed numbers in the activated input box behind, then click
“Print”;
Results re-calculation (when the standard curves of one or some items are changed, it is not need to
retest the items. Only their test results needs to be recomputed.)
1 If one sample on the left of the interface is selected, its test items results will be displayed in the
item results list.
2 If you click ”Re-calculation”, the software will automatically re-calculate the results of all test
items of the selected sample and display the results.
Export (according to the export ranges “current sample, all samples, selected sample” closed by users,
which is the same to the selection of sample print)
1 The sample range exported by selection results: “current sample, all samples, selected sample”,
the sample selection mode is in the interval between the independent sample number separated
with “.” and the consecutive sample number separated with “_” input in the latter box.
- 82 -
User Manual of BA-A-S120Mini Analyzer
2 If you click “Export”, the system will export the mode and format according to the configuration
in 9.2.8 and complete the appointed sample results export.
6.6.5 Stop
Click “Stop” on the left toolbar of the main interface, the below form will appear. The function of stopping all
tests in progress only can be performed in case that the user needs to stop the current work due to a variety of
reasons.
a After clicking “OK”, the analysis meter will stop immediately. If you do not want to stop, just click
“Return”.
- 83 -
User Manual of BA-A-S120Mini Analyzer
Start
Stop
- 84 -
User Manual of BA-A-S120Mini Analyzer
After clicking:
- 85 -
User Manual of BA-A-S120Mini Analyzer
1 Switch the bookmark of sample disks to view test status of samples in different sample disks.
2 In the sample disk position display area, different test types and test status will be indicated in
different shapes and colors.
3 After selecting one sample in the sample disks, the sample number and sample type will be
displayed in the right information display area for selected samples, the applied items and the test
status of all items will be displayed in the test results display area, and the time needed for the
sample to finish test will be displayed in the test remaining time estimation column.
4 After selecting one item that has finished test from the test results display area of the selected
samples, click “Select reaction curves of finished items” to view the reaction curve and reaction
data of the test.
5 Click “Return” to close sample disk interface and return to the main interface.
1 Switch the bookmark of the reagent disk to view the position and remaining amount of reagents
in different reagent disks
2 In the left reagent disk position display area, the occupation conditions of reagent position will be
displayed in “Occupied” and “Free”.
3 In the right item display area, gray indicates that the reagent position of the item has been setup;
and yellow indicates that the reagent position of the item has not been setup.
- 86 -
User Manual of BA-A-S120Mini Analyzer
4 In the reagent display area that has been setup, the position and remain of all reagents can be
viewed.
2 In the right item display area, please click one item in yellow, and also activate R1 or R2 if
dual-reagent, then double click one vacant position in the left reagent disk, thus the reagent can
be setup to the position.
2 Firstly click one reagent needed to change position, then double click one vacant position, thus
the reagent can be setup from original position to the current one.
2 Firstly click one reagent needed to delete position, then click right “Delete occupied cup
position”, thus the reagent position can be deleted.
2 Click “Test Reagent Remain” to popup a new interface. Select the reagent needed to test the
remain and then click button “OK”.
Note:
(1) For the dual-reagent item, its R1 and R2 should be setup in one reagent disk;
(2) Reagent remain test only can be tested in the standby mode.
- 87 -
User Manual of BA-A-S120Mini Analyzer
1 In the left reaction disk position display area, the test status of all reaction cups will be displayed
in different colors.
2 If click one reaction cup, all the items tested and in the progress of test on the same day will be
displayed in the right test item column of the day for the selected reaction cups.
3 If select one test, the sample number and sample type of the item will be displayed in the detailed
information column for the selected test items.
4 If select one finished test item, and click button “Reaction curve of selected finished items”, the
reaction curve and reaction data of the item can be viewed;
5 Click “Return” to close the reaction disk interface and return to the main interface
- 88 -
User Manual of BA-A-S120Mini Analyzer
Click “Return” to close the alarm information interface and return to the main interface. For all previous alarm
information, please see Error Log Query.
From left to right is expressed as: the first one is “Liquid Waste” (there are two liquid levels for the liquid
waste in the waste container: “Full” and “Not Full”); the second one is “Cleaning Water” (there are four water
levels: “Empty”, “Water Replenishing” and “between Water Replenishing and Full” and “Full”); the third one
is “Instrument Running” (there are two status: standby and running, with the graph in dynamic rotation
indicating the instrument is in testing, with the graph in static indicating the instrument is in standby or not
online).
- 89 -
User Manual of BA-A-S120Mini Analyzer
Reaction cycling
Double reagent test
Reaction ends
Add samples
Shut down
Max. 49 cycles
Test process
Metering spot
Return
- 90 -
User Manual of BA-A-S120Mini Analyzer
- 91 -
User Manual of BA-A-S120Mini Analyzer
7.4.1 Absorbance
for SL 120, the computation formula of absorbance is as below:
Solution Absorbance=Log (AD water-AD dark) / (AD solution-AD dark)
Thereinto:
(1) “Log” means normal 10-base logarithm operation;
(2) “AD”means the numerical value of transmitting light intensity after photoelectric conversion and
digitaltoanalog conversion ;
(3) “AD dark” means the AD value with light off, “AD water” means the AD value for reaction vessel
filled up with water, “AD solution” means the Ad value for reaction vessel filled up with solution to be
tested;
- 92 -
User Manual of BA-A-S120Mini Analyzer
For absorbance, if it is single wavelength, then use the absorbance of the dominant wavelength; if it is
dual-wavelength, then the absorbance of each point equals to dominant wavelength absorbance minus
sub-wavelength absorbance;
(3) Dynamics method
Reaction amplitude = The change rate per minute for starting point and destination absorbance
Thereinto:
Destination and starting point is set in the “computing time” of “item parameter” by the user;
For absorbance, if it is single wavelength, then use the absorbance of the dominant wavelength; if it is
dual-wavelength, then the absorbance of each point equals to dominant wavelength absorbance minus
sub-wavelength absorbance;
7.5 Calibration
- 93 -
User Manual of BA-A-S120Mini Analyzer
Calibration formula R=R0+Ke[alnC+b(lnC)2+c(lnC)3], with 5 parameters, which are R0,K,a,b and c. It is required to
provide at least 5 standard sample, the concentration(activity) of first standard sample is zero, the
corresponding R equals to R0, other parameters can be determined with iteration method.
(7) Polynomial-5P
Calibration formula LnC=a+b (R-R0)+c(R-R0)2+d(R-R0)3, with 5 parameters, which are R0, a, b, c and d. It is
required to provide at least 5 standard sample, the concentration(activity) of first standard sample is zero, the
corresponding R equals to R0, other parameters can be determined with iteration method.
(8) Spline
Calibration formula C-Ci=R0i+ai(C-Ci)+bi(C-Ci)2+ci(C-Ci)3-R, with 4i parameters, which are R0i,ai,bi and ci.
It is required to provide at least 2 standard sample, the parameters of each inter zone can be determined with
iteration method.
12S 1 point fall out of the mean value +2SD or -2SD. Warning
13S 1 point fall out of the mean value +3SD or -3SD. Out of control(random error 、
system error)
22S 2 continuous points fall out of +2SD or -2SD. Out of control(system error)
41S 4 continuous points fall out of mean value +1SD or Out of control(system error)
-1SD.
10X 10 continuous points fall in the same side of the mean Out of control(system error)
value
- 94 -
User Manual of BA-A-S120Mini Analyzer
No
Under control
Yes
No No No No No
Out of control
- 95 -
User Manual of BA-A-S120Mini Analyzer
as the ordinate, the mean value as the centerline, and mark the lines ±1SD, 2SD and ±3 SD, with the value of
the two quality control solutions measured at the same time formed one point that will be marked on the
coordinate shown as Fig. 10-6:
3SD
2SD
1SD
-1SD
-2SD
-3SD
- 96 -
User Manual of BA-A-S120Mini Analyzer
Rij Ri
i 1 j 1
' 2
SD =
Nn 2
Where, Rij, the response amplitude (effective measurement) measured for calibration solution i in one time;
Ri ' , the degree of response of calibration solution i computed based on the calibration curve; N, the number
of calibration siolution; n, times of effective repeated measurement.
Logit-4P
N n
Rij Ri
i 1 j 1
' 2
SD=
Nn 4
Where, Rij, the degree of response (effective measurement) measured for calibration solution i in one time;
Ri ' , the degree of response of calibration solution i computed based on the calibration curve; N, the number
of calibration siolution; n, times of effective repeated measurement.
Logit-5P
N n
Rij Ri
i 1 j 1
' 2
SD=
Nn 5
Where, Rij, the degree of response (effective measurement) measured for calibration solution i in one time;
Ri ' , the degree of response of calibration solution i comupted based on the calibration curve; N, the number
of calibration siolution; n, times of effective repeated measurement.
Exponential-5P and polynomial-5P
N n
Rij Ri
i 1 j 1
' 2
SD=
Nn 5
Where, Rij, the degree of response (effective measurement) measured for calibration solution i in one time;
Ri ' , the degree of response of calibration solution i computed based on the calibration curve; N, the number
of calibration siolution; n, times of effective repeated measurement.
Spline
N n
Rij Ri
i 1 j 1
' 2
SD=
Nn 4
Where, Rij, the degree of response (effective measurement) measured for calibration solution i in one time;
Ri ' , the degree of response of calibration solution i computed based on the calibration curve; N, the number
of calibration siolution; n, times of effective repeated measurement.
(5) Factor Related to Calibration Curve
Only applicable to multiple linear and nonlinear calibration curves, the computational formula is shown as
follows:
- 97 -
User Manual of BA-A-S120Mini Analyzer
Cij C Rij R
N n
2 2
i 1 j 1
R2
Cij C Rij R
N n N
2 2
i 1 j 1 i 1
Where, C, the concentration of the calibration solution; R, the response amplitude; N, the number of
calibration solution; n, times of effective repeated measurement.
Note:
SL 120 will measure each calibration solution for three times regularly and repeatedly.
The result with maximum deviation will be deleted, and the other two results will be
taken an average value, hence n equals 2 fixedly.
- 98 -
User Manual of BA-A-S120Mini Analyzer
conducted, the insoluble antigen-antibody complex amount generated by two samples with very large different
concentration can be equal, which will cause the same test result.
Fig. 10-7 Schematic diagram of the relationship between insoluble antigen-antibody complex and the ratio of
antigen-antibody
In SL 120, the prozone check can be conducted according to the following methods:
(1) Double Reagent End Point Method
Shown as the following figuer, L is for reaction starting point, M is for reaction ending point, N and P are for
prozone check point. L, M, N and P should meet the following relationship:
13≤L<26≤N<P<M≤77
Fig.10-8 Schemtaic diagram of the prozone check for double reagent end point method
The value of the prozone check (PC) is equal to:
AM A P
PC M P 100%
AP AN
PN
If PC is less than the prozone check limit, it can be judged that the prozone phenomenon is existed;
(2) Single Reagent End Point Method
Shown as the following figuer, L is for reaction starting point, M is for reaction ending point, N and P are for
prozone check point. L, M, N and P should meet the following relationship:
- 99 -
User Manual of BA-A-S120Mini Analyzer
1≤L<13≤N<P<M≤77
Fig.10-8 Schemtaic diagram of the prozone check for single reagent end point method
The value of the prozone check (PC) is equal to:
AM A P
PC M P 100%
AP AN
PN
If PC is less than the prozone check limit, it can be judged that the prozone phenomenon is existed;
- 100 -
User Manual of BA-A-S120Mini Analyzer
The transmittance value of any wavelength is less than 40% of factory value;
The transmittance value of any wavelength falls outside of plus or minus 3 times standard deviation of
transmittance average value of all cups.
- 101 -
User Manual of BA-A-S120Mini Analyzer
- 102 -
User Manual of BA-A-S120Mini Analyzer
2. The quality control results of individual item fall outside the scope of quality control
Reagent is expired.
The regent blank test is abnormal
Reagent is misplaced
The calibration solution and quality control solution are not matched
- 103 -
User Manual of BA-A-S120Mini Analyzer
Note:
While placing new set of cuvettes, do not touch the optical surface of the
reactionreaction vessel by hand.
Do use the cuvettes specially provided by Biosino only to ensure the accuracy of
the test results.
1 Clean the analyzer table with wet towel absorbing cleaning fluid until all the stains are erased;
2 Clean the analyzer table with the wet towel absorbing disinfectant fluid;
3 Fifteen minutes later, remove the disinfectant fluid on the table with clean wet towel, which is
wring out.
- 104 -
User Manual of BA-A-S120Mini Analyzer
Note: Corrosion
The cleaning fluid is chemically corrosive. During the operation, do wear protective
gloves.
Note: Combustion
The ethanol is combustible. During the cleaning operation, make sure the analyzer is
power off. Use no more than 10ml ethanol in areas adjacent to the analyzer.
1 Clean the discharge needle tube and point with cotton swab absorbing absolute ethyl
alcohol, until there is no visible attachment;
2 Clean the take-in needle tube and point with cotton swab absorbing absolute ethyl
alcohol, until there is no visible attachment;
1 Clean the four sides, top and bottom of the nozzle with clean cotton swab absorbing
purified water, until there is no visible attachment;
2 Clean the four sides, top and bottom of the nozzle with clean cotton swab absorbing
absolute ethyl alcohol, until there is no visible attachment;
- 105 -
User Manual of BA-A-S120Mini Analyzer
Note:
Note that during the cleaning, there may be some cotton fiber from the cotton swab left
between the drainage and take-in needle tubes. In this case, clean the fiber in time.
Note: Combustion
The ethanol is combustible. During the operation, make sure the analyzer is power off.
Use no more than 10ml ethanol in areas adjacent to the analyzer.
1 Cut off the power supply of the system. Make sure that the power cord is disconnected from the
power grid
2 Open the upper shield, and clean the workbench with clean water
3 After cleaning the workbench, clean the interior part of the upper shield.
4 Then open the front door at the lower part off the instrument, and clean the inside and outside
walls of the door
5 Finally, clean both sides of the instrument, and the back side of the case
Cautions:
1. Do not plug in the power cord and start the instrument immediately after cleaning the case. Start the
instrument when you make sure that the surface of the instrument case is free of water.
2. Do not put the reagent bottle on the workbench when the instrument is working, lest the reagent bottle
should fall over and the reagent flow into the instrument. This may shorten the use life of the
instrument.
9.4.2 Automatic Flushing the Stirring Rod, Sampling Probe and Pipeline
Prepare the following stuff: the reagent bottle containing 35ml cleaning fluid, and sample cup. Clean as the
following steps:
2 Put the reagent bottle holding 35ml cleaning fluid in the designated position on the reagent tray
- 106 -
User Manual of BA-A-S120Mini Analyzer
3 Power on and the system initiation starts. When the system is in standby mode, run the
automatic cleaning program.
4 Run the cleaning program as needed. The entire process is about 5 minutes
Cautions:
1. Keep the hand and clothes clear of the cleaning fluid. In case of touching the cleaning fluid, clean with
water immediately. If, however, the fluid is in your eyes or mouth, clean with water and see the doctor
immediately.
2. Keep the instrument clear of the cleaning fluid. Or else, clean the fluid with clean cloth or paper.
Finally clean it with dry cloth 2-3 times.
3. Do not mix the cleaning fluid with other chemicals. In case of such mixture, neutralize properly and
discard the mixture.
2 Open the upper cover and sample tray lid. As shown in the Figure below, two black fasteners are
used to secure the sample tray
3 Lift the two fasteners with proper force. As shown in the Fig. below, you can see the bottom of the
tray
4 Take out the ragent bottle on the tray. Clean the reagent tray and its bottom with gauze absorbing
the cleaning fluid, until there is no visible stain, then clean with dry gauze.
5 Mount the sample tray and cover the lid and upper cover.
- 107 -
User Manual of BA-A-S120Mini Analyzer
3 Clean the inside and outside walls of the wash tank of sampling probe with cotton swab
absorbing the cleaning fluid, until there is no visible stain. Then clean with dry gauze;
5 Clean the inside and outside walls of the wash tank of the stirring rod with cotton swab
absorbing the cleaning fluid, until there is no visible stain. Then clean with dry gauze;
2 Lift the reaction plate with proper force. Clean all the parts of the inside wall (indicated in red) of
the reaction tank with gauze absorbing the cleaning fluid, until there is no visible stain. Then clean
with dry gauze.
- 108 -
User Manual of BA-A-S120Mini Analyzer
1 Power off;
2 Turn the sampling probe to an appropriate position, and open the top cover of the probe rocking
arm
4 Unplug the wire to liquid level checkout console and the joint with the tube. Loosen the holddown
spring.
6 Unclog the sampling probe with Φ0.3mm stainless steel wire from the probe point repeatedly.
7 Connect the one-shot injector with an appropriate hose on one end. The other end of the hose is
connected to the rapid acting coupling. Inject water into the probe. That the water shooting out in a
line from the probe indicates the probe has been unclogged;
8 Mount the probe and cover the rocking arm in the reverse order;
Before replacement, make sure that the sampling probe is above the wash tank, as liquid may
1
drop off during the replacement;
2 Check that the system is on, and in the warm-up or stand-by mode;
3 Remove the old sampling probe according to the steps to unclog the sampling probe;
Insert the new sampling probe from the point, connect the probe with the joint and with liquid
4 level monitoring console. Ensure that there is no leaks at the joint. Tighten the screws fastening
the sampling probe;
Start up the system, and reset twice or three times, until the tube is free of air. When the tube is
5
free of air, water will jet out from the probe point. Check that the water jets out properly;
Mount the sampling probe cover. Ensure that the cover is securely in place as indicated by a clip
6
sound
- 109 -
User Manual of BA-A-S120Mini Analyzer
2 Loosen the two screws fastening the stirring rod, and take out the stirring rod to be replaced.
3 Insert the new stirring rod into the motor shaft in the bottom-top direction. Tighten the screws
fastening the stirring rod.
4 Rotate the stirring rod lightly, and make sure that there is no marked movement.
- 110 -
User Manual of BA-A-S120Mini Analyzer
4 Make the instrument perform initialization, reset and self-check cleaning repeatedly;
5 Open the clean water barrel, and clean the bottle and pipeline with remaining cleaning agent.
Then rinse with clean water.
- 111 -
User Manual of BA-A-S120Mini Analyzer
2 Move the stirring rod, so that the drive rod is turned to an angle suitable for wiping;
3 Wipe the drive rod with clean gauze until there is no visible dust and stain. Then apply the
lubricating oil, and pull the drive rod up and down, so that the lubricating oil is evenly aplied to
the rod;
- 112 -