Professional Documents
Culture Documents
Pasco
Pasco
the PASCOscientific
Model SE-9638
e/m APPARATUS
0 PASCO S5.oo
Table of Contents
Section Page
Copyright, Warranty, and Equipment Return ii
Introduction
Equipment
Operation
Measuring e/m
Analysis of elm Measurement
Deflections of Electrons in an ElectricField.......
Two Simple Demonstrations
Diagram of Connections
Improving Experimental Results
s c tent/ftc
e/m Apparatus
012-034710
Copyright Notice
The PASCO scientific Model SE-9638 e/m Apparatus will repair or replace, at its option, any part of the product
manual is copyrighted and all rights reserved. However, which is deemed to be defective in material or workman-
permission is granted to non-profit educational institutions ship. This warranty does not cover damage to the product
for reproduction of any part of this manual providing the caused by abuse or improper use. Determination of whether
reproductions are used only for their laboratories and are not a product failure is the result of a manufacturingdefect or
sold for profit. Reproduction under any other circum- improper use by the customer shall be made solely by
stances, without the written consent of PASCO scientific, is PASCO scientific. Responsibility for the return of equip-
prohibited. ment for warranty repair belongs to the customer. Equip-
ment must be properly packed to prevent damage and
shipped postage or freight prepaid. (Damage caused by
Limited Warranty improper packing of the equipment for return shipment will
PASCO scientific warrants this product to be free from
not be covered by the warranty.) Shipping costs for
defects in materials and workmanship for a period of one returning the equipment, after repair, will be paid by
year from the date ofshipment to the customer. PASCO PASCO scientific.
Equipment Return
ii
snønt't'C
012-03471 D elm Apparatus
Introduction
Equipment
The e/m Tube—- The e/m tube (sce Figure 2) is filled with
helium at a pressure of 100 mm Mg, and contains an electron Helium Filled
gun and deflection plates. The electron beam leaves a visible Vacuum tube
trail in the tube, because some of.the electrons collide with
helium atoms, which are excited and then radiate visible
light.
Thc heater heats the Electron Gun
The clcctron gun is shown in Figure 3.
cathode, which emits electrons. The clcctrons arc accelerated
applied between the cathode and the anode.
by a potential Deflection Plates
positive with respect to the cathode and
The grid is held
to the anode. It helps to focus the
negative with respect
electron beam.
octont///C
elm Apparatus
012-034710
Operation
the Helmholtz
Measuring e/m 6. Slowly turn the current adjust knob for
take care that
l. If you Willbe working in a lighted room, place the coils clockwise. Watch the ammeter and
hood over the elm apparatus, the current does not exceed 2 A.
up. When
2. Flip the toggle switch up to the e/m MEASURE 7. Wait several minutes for the cathode to heat
from
position. it does, you will see the electron beam emerge
field from
the electron gun and it will be curved by the
3, Turn (he current adjust knob for the Helmholtz beam is
coils the HelmholtzcoiJs. Check that the electron
to the position. the
parallel to the Helmholtz'coils. If it is not, turn
4, Connect your power supplies and meters
to the front tube until it is. Don't take it out of its socket. As you
panel of the e/jn apparatus, os sJiown in rotate the tube, the socket will turn.
Figure 4.
5, Adjust fhc power supplies to the following
levels: 8. Carefully read the current to the Helmholtz coils from
your ammeter and the accelerating voltage from your
voltmeter. Record the values below.
J fearer 6.3 or VOC
Llectrodes: ISOto goo VOC Current to I Iclmholtz coils I
Jlelmholf/, Coils: VDC jpplc be
less than Yo) Accelerating voltage —V
+ IJpper
Lower
O O
DC
Ammeter
(0-2 A) Voltmeter Power Supply
(0-300 (Heater
VOC) 6.3 VOC or
VAC)
Power Supply
(Helmholtz
Coils 6-9 VOC, Power Supply
ripple < 1%) (Accelerating
Voltage
150-300VOC)
on the mirrored scale. Measure the radius of the beam as you elm v/Br (3)
see it on both sides of the scale, then average the results.
Record your result below. Therefore, in order to detemine elm. it is only necessary to
know the velocity of the electrons, the magnetic field
produced by the Helmholtz coils, and the radius of the
Electron Beam Radius = r =
electron beam.
The electrons are accelerated through the accelerating
potential V, gaining kinetic energy equal to their charge
Analysis of elm Measurement times the accelerating potential. Therefore cv = 1/2 me,
The velocity of the electrons IStherefore:
The magnetic force (Fm)acting on a charged particle of
v = (2eV/m)L1 (4)
charge q moving velocity v in a magnetic field (B) is
given by the equation Fn = qv X B, (where F, v, and B are The magnetc field produced near the ans of a parr of
vectors and X is a vector cross product). Since the electron
beam in this experiment is perpendicular to the magnetic
(5)
field. the equation can be in scalar form as:
Fm= evB (1)
He:jn+'— by the equation:
where e is the charge of the electron.
A = -ur.for this forr.ula can be found tn most Introduc-
Since the electrons are moving in a circle, they must be v. t:xtä and magnetism.
experiencing a centripetal force of magnitude
Equations 4 and S can be plugged Tito equation 3 to get a
F, = mv2/r (2) final formula for e'm:
where m is the mass of the electron, v is its selocity, and r
the radius of the circular motion. Since the only force acung
on the electrons is that caused by the magnetic field, F = F ,
so equations I and 2 can be combined to give evB =
mv 2/r or
3
s ctent/ttc
e 'm Apparatus
012-03471
D
DC Ammeter HelmholtzCoils
(0-2 A)
BLK
Power Supply
(6-9'IDC.
ripple < 1%)
50
BLK
Current Adjust
for Helmholtz Coils
Slide/ToggIe Switch
(elmMEASURE—ELECTRICAL DEFLECT)
Upper YEL WHT
Voltmeter
(0-300voc) WHT
Lower
ORG
YEL Deflectio
ORG elm Tube Plates
BLK
pin 11 pin 10
RED pin 6 Anode
ORG 15K Grid
15W
Focus Adjust Cathode
150-300 VOC Variable pin 2
Resistor —-Heater
5K
BLK pin 12 pin 1
BLK
or VAC
BLU
Power
Supply
sctont//;c 5
elm Apparatus
012-034710
Measurement of e/m
Notes
l) The greatest source of error in this experiment is the velocity of the electrons. First, the non-uniformity of the
accelerating field caused by the hole in the anode causes the velocity of the electrons to be slightly less than
their theoretical value. Second, collisions with the helium atoms in the tube further rob the electrons of their
velocity. Since the equation for e/m is proportional to I/N, and r is proportional to v, experimental values for
e/m will be greatly affected by these two effects.
2) To minimize the error due to this lost electron velocity, measure radius to the of the beam path.
3) To minimize the relative effect of collisions, keep the accelerating voltage as high as possible. (Above 250V
for best results.) Note, however, that if the voltage is too high, the radius measurement will be distorted by
the curvature of the glass at the edge of the tube. Our best results were made with radii of less than 5cm.
1.6A
1.25
— 1.2
o
1.1
4) Your
be higher than
theoretical, due to
the fact that both
major sources oferror
6
Feed-Back Contacting Technical Support
If you have any comments about this product or this Before you call the PASCO Technical Support staff it
manual please let us know. If you have any sugges- would be helpful to prepare the following information:
tions on alternate experiments or find a problem in the
manual please tell us. PASCO appreciates any cus- • If your problem is computer/software related, note:
1.7 tomer feed-back. Your input helps us evaluate and Title and Revision Date of software.
Improve our product.
Type ofComputer (Make, Model, Speed).
To Reach PASCO Type of external Cables/Peripherals.
For Technical Support call us at 1-800-772-8700(toll- • If your problem is with the PASCO apparatus, note:
free within the U.s.) or (916) 786-3800.
Title and Model number (usually listed on the label).
Approximate age of apparatus.
A detailed description of the problem/sequence of
events. (In case you can't call PASCO right away,
you won't lose valuable data.)
If possible, have the apparatus within reach when
calling. This makes descriptions of individual parts
much easier.
• If your problem relates to the instruction manual,
note:
Part number and Revision (listed by month and year
on the front cover).
Have the manual at hand to discuss your questions.
scjent'f/C
Instruction Manual and 012-046950
03/99
Experiment Guide for the
PASCOscientific
Model TD-8553/8554A/8555
THERMAL
RADIATION SYSTEM
TDZ5S4A Radiation Cube
(Leslie's Cube)
O
o
TO-8555Stefan
Boltzman Lamp
e PASCO
CA 95747-71"
S C ent ' C Pha.-g • FAX 786-8905
012aut,9SD
Thermal Radiation System
Table of Contents
Section Page
Copyright and Warranty, Equipment Return ii
Introduction
Radiation Sensor
Thermal Radiation Cube (Leslie's Cube) ..
Stefan-Boltzmann Lamp
Experiments:
Experiment l: Introduction to Thermal Radiation
Experiment 2: Inverse Square Law
Experiment 3: Stefan-Boltzmann Law (high temperature) 13
Experiment 4: Stefan-Boltzmann Law (low temperature) 17
Teacher's Guide 19
Technical Support Inside Back Cover
i
Thermal Radiation System
Introduction
The PASCO Thermal Radiation System includes three In addition to the equipment in the radiation system,
items: the TD-8SS3 Radiation Sensor, the TD-8554A several standard laboratory items, such as power
Radiation Cube (Leslie's Cube), and the TD-8555 supplies and meters are needed for most experiments.
Stefan-Boltzmann Lamp. This manual contains Check the experiment section of this manual for
operating instructions for each of these items plus infomation on required equipment.
instructions and worksheets for the following four
experiments: If you don't have all the items of the radiation system,
read through the operating instructions for the equip-
0) Introduction to Thermal Radiation, ment you do have, then check the experiment section
Inverse Square Law, to determine which of the experiments you can per-
(S' Stefan-Boltzmann Law* (at high temperatures), form. (A radiation sensor is required for all the
@ Stefan-Boltzmann Law* (at low temperatures). experiments.)
• The Stefan-Boltzmann law states that the radiant
energy per unit area is proportional to the fourth
power of the temperature of the radiating surface.
Radiation Sensor
The PASCO TD-8553 Radiation Sensor (Figure l) The two posts extending from the front end of the
measures the relative intensities of incident thermal Sensor protect the thermopile and also provide a
radiation. The sensing element, a miniature thermo- reference for positioning the sensor a repeatable
pile, produces a voltage proportional to the intensity of distance from a radiation source.
the radiation. The spectral response of the thermopile
is essentially flat in the infrared region (from 0.5 to 40 Specifications
pm), and the voltages produced range from the micro- Temperature Range: -65 to 85 oc.
volt range up to around 100 millivolts. (A good Maximum Incident Power: 0.1 Watts/cm2.
millivolt meter is sufficient for all the experiments Spectral Response: .6 to 30prn.
2
described in this manual. See the current PASCO Signal Output: Linear from 104 to 10-1Watts/cm .
catalog for recommended meters.)
The Sensor can be hand held or mounted on its stand Thumbscrew: Loosen to
for more accurate positioning. A spring-clip shutter is Shutter Ring: Slide reposition Sensor or to
opened and closed by sliding the shutter ring forward forward to open remove Sensor from stand
shutter
or back. During experiments, the shutter should be
closed when measurements are not actively being
taken. This helps reduce temperature shifts in the
thermopile reference junction which can cause the
Shutter
sensor response to drift.
s ctent'ttc
Cube)
Thermal Radiation Cube (Leslie's
may be HOT!
The TD-SSS4A Radiation Cube (Figure 2) provides CAUTION: cube
four different radiating surfaces that can be heated
from temperature to approximately 120 The
cube is heated by a 100 light bulb. Just plug in
the cord. flip the toggle switch to then
tum the knob clockmse to vary the pou•r.
Turn00b
Measure the cube temperature by plugging your Socov•se to
to
ohmmeter into the banana plug connectors labeled •ON' to tun'
THER-MISTOæ The in one on power.
temperature,
comer of the cube. Measure the resistance, then
Table l, below, to translate the resistance reading into a
temperature measurement- An abbrevated versionof Use Oble on
this table is printed on the base of the Radiation Cube. back to
NOTE: For best results, a digital ohmmeter To
should be used (See the current PASCOcatalog
for recommended meters) VAC
Table t
Resistance versus Temperature for the Thermal Radiation Cube
Temp. Temp. Therm Temp. Tbenw Tcnp. Themr Ternp Therm Temp
Res (Q) CC)
-'07.sso 10 130
2X4S3 g.CS.O 107 131
1S'.s.ao 12 60 4339.7 2-157.6 132
1moso 13 21.736 61 9.121S 4209.1 logs-7 133
SS.6SS so 110 2041.7
53.297 39 63 s-'227
S I .04S 40
3±1.1 111 I.es€4 135
16 64 s.-vs SS 3.S43_g 112
I ".s—€so 4S.90S I S.66S 7.969.1 3.729.'
1.932S
17.•so
113 1.009
19 I s-;os 138
43.062 4-4
91
lis 1>1.7 139
3.4110 116 1.734.1
22 39.605 93
Is-so: IS.s:.Y4 141
23 3-'tss IIS 142
24 segss 1+10 i 123.0
l.sm.6
I CQOO)
s-;-us )
143
120
so 97
32>3 sss€l 121
12 l.cs.6
s-.022 s:
29.'SS 12043 100
3.422 30 s-vss 101 1.40.3.0
31
72-seo
so
103 2 CSI I-so
33 104
12s
129
2
012 SD
Thermal Radiation System
Stefan-Boltzmann Lamp
R-R
For large temperature differences,therefore,deter-
ref mine the temperature of the tungsten filament as
For small temperature changes, the temperatureof follows:
the tungsten filament can be calculated using a, the
(D Accurately measure the resistance (R ) of the tung-
temperature coefficient of resistivity for the filament:
sten filament at room temperature (about 300 0K).
where, Accuracy is important here. A small error in Rref
I — Temperature will result in a large error in your result for the fila-
R — Resistance at temperature T ment temperature.
T — Reference temperature (usually room temp.)
@ When the filament is hot, measure the voltage and
R = Resistance at temperature T current into the filament and divide the voltage by
— Temperature coefficient of resistivity for the the current to measure the resistance (R-r).
filament(a = 45 x 100 for tungsten)
is not @ Divide R.r by Rrfto obtain the relative resistance
For large temperature differences, however, a (RT/Rrf).
constant and the above equation is not accurate.
@ Using your measured value for the relative resistiv-
ity of the filament at temperature T, use Table 2 on
the following page, or the associated graph, to de-
termine the temperature of the filament.
3
s c •ent'f•c
2
,、
5 7
6
、
8, ,0
3
9
0
R 0
180
Temperature(Kelvin)
EQUIPMENT NEEDED:
— Radiation Sensor, Thermal Radiation Cube — Window glass
— Millivoltmeter — Ohmmeter.
NOTES:
@ If lab time is short, it's helpful to preheat
the cube at a setting of 5.0 for 20 minutes before
the laboratory period begins. (A very quick method is to preheat
the cube at full power
for 45 minutes, then use a small fan to reduce the temperature
quickly as you losser the
power input- Just be sure that equilibrium is attained with the fan
off.)
Part I and 2 of this experiment can be performed simultaneously.
Make the measure-
ments in Part 2 while waiting for the Radiation Cube to
reach therrnal at each
of the settings in Part l.
@ When using the Radiation Sensor, always shield it
from the hot object except for the few
seconds it takes to actually make the measurement- This prevents
heating of the thermo-
pile which will change the reference temperature and alter the reading.
5
s c 'ent'ftc
cjnittcd
radiationsummarizing your
Use the Radiation Sensor to cxatnine the relative magnitudes table
a
ofpapcr. make listed below.
sonous objects around the mom. On a separate sheet questions
answ er the
obscrsations. Make measurements that will help you to
and record
Radiation cube window
Place the Sensor S cm from black sensor and bulb. Does
0Cwindow glass between the
thtiereading. place a
glxss effectively block thermal radiation? repeat the
Lamp) and
e the Stefan-Boltzmann with other
black surface. Repeat
instead octhe
measurements ocstep l. but using the bare bulb
matcnals.
SC'enti/tc
0124u69SD Thermal Radiation System
Questions (Part 1)
CDList the surfaces of the Radiation Cube in order of the amount of radiation emitted. Is the order
independent of temperature?
@ It is a general rule that good absorbers of radiation are also good emitters. Are your measure-
ments consistent with this rule? Explain.
Questions (Part 2)
(D Do different objects, at approximately the same temperature, emit different amounts of radiation?
@ Can you find materials in your room that block thermal radiation? Can you find materials that
don't block thermal radiation? (For example, do your clothes effectively block the thermal
radiation emitted from your body?)
7
s c lent'/tc
0124U69SD
Thermal Radiation System
EQUIPMENT NEEDED:
— Radiation Sensor
— Stefan-BoltzmannLamp, Millivoltmeter
— Power Supply (12 VDC; 3 A), meter stick.
Top View
x
Power Supply
(13 V MAX!)
9
s ctent'/tc
012-046950
AdJust the distance between the Sensor and the lamp to each of the settings listed in Table 2.2.
At each setting. record the reading on the millivoltmeter.
JMPORTANT: Make each reading quickly. Between readings, move the Sensor away
from the lamp, or place the reflective heat shield between the lamp and the Sensor, so that
the temperature of the Sensor stays relatively constant.
SO 4.0
4.5
70
5.0
80
90
6.0
7.0
Average Ambient 8.0
Radiation Level
9.0
18.0
20.0
25.0
30.0
35.0
40.0
45.0
50.0
60.0
70.0
80.0
2.2 900
Radiation Lovel vorsus Distanco
10
012-0469SD
Thermal Radiation System
Calculations
(D For each value of X, calculate l/X 2. Enter your results in Table 2.2.
@ Subtract the Average Ambient Radiation Level from each of your Rad measurements in
Table 2.2. Enter your results in the table.
@ On a separate sheet of paper, make a graph of Radiation Level versus Distance from Source,
using columns one and four from Table 2.2. Let the radiation level be the dependent (y) axis.
@ If your graph from part 3 is not linear, make a graph of Radiation Level versus l/X 2, using
columns three and four from table 2.2.
Questions
(D Which of the two graphs is more linear? Is it linear over the entire range of measurements?
@ The inverse square law states that the radiant energy per unit area emitted by a point source
of radiation decreases as the square of the distance from the source to the point of detection.
Does your data support this assertion?
@ Is the Stefan-Boltzmann Lamp truly a point source of radiation? If not, how might this
affect your results? Do you see such an effect in the data you have taken?
s c jent/f'c
012-0469.SD Thermal Radiation System
EQUIPMENTNEEDED:
— Radiation Sensor —Stefan-Boltzmann Lamp
— Ohmmeter — Ammeter (0-3 A)
— Voltmeter (0-12 V) —Millivoltmeter
— Ohmmeter — Thermometer.
Introduction
The Stefan-Boltzmann Law relates R, the power per unit area radiated by an object, to T, the
absolute temperature of the object. The equation is:
4.
R=oT =5.6703 x 10 4
m 2K
In this experiment, you will make relative measurements of the power per unit area emitted
from a hot object, namely the Stefan-Boltzmann Lamp, at various temperatures. From your
data you will be able to test whether the radiated power is really proportional to the fourth
power of the temperature.
Most of the thermal energy emitted by the lamp comes from the filament of the lamp. The
filament temperature can be determined using the procedure given on pages 3 and 4 of this
manual.
Ammeter
Power Supply
(13 V MAX!)
Millivoltmeter
Voltmeter
6 cm
sctont//tc
Thermal Radjaoon System
012-046950
Procedure
IMPORTANT: The voltage into the lamp should NEVER exceed 13 V. Higher voltages
will burn out the filament.
LMPORTANT: Make each Sensor reading quickly. Between readings, place both sheets
of insulating foam between the lamp and the Sensor, with the silvered surface facing the
lamp, so that the temperature of the Sensor stays relatively constant.
14
012-0469SD Thermal Radiation System
= 4.5 x 10-3K-'
R (filament resistance at =
Table 3.1
Data Calculations
1 Rad
(Volts) (Amps) (Ohms) (K)
I .00
2.00
3.00
4.00
5.00
6.00
7.00
8.00
9.00
10.00
11.00
12.00
scjent/ftc 15
012-0469SD
Thermal Radiation System
EQUIPMENT NEEDED:
— Radiation Sensor — Thermal Radiation Cube
—Millivoltmeter — Ohmmeter.
Introduction
In experiment 3, you investigated the Stefan-Boltzmann Law = ST4) for the high
n temperatures attained by an incandescent filament. At those high temperatures (approxi-
mately 1,000 to 3,000 K), the ambient temperature is small enough that it can be neglected
in the analysis. In this experiment you will investigate the Stefan-Boltzmann relationship at
much lower temperatures using the Thermal Radiation Cube. At these lower temperatures,
the ambient temperature can not be ignored.
If the detector in the Radiation Sensor were operating at absolute zero temperature, it would
produce a voltage directly proportional to the intensity of the radiation that strikes it. How-
ever, the detector is not at absolute zero temperature so it is also radiating thermal energy.
According to the Stefan-Boltzmann
law, it radiates at a rate, Rdet
ST 4. The voltage produced by the
sensor is proportional to the radia-
tion striking the detector minus the
radiation leaving it. Mathemati-
cally, the sensor voltage is propor-
tional to R -R rad-Rae, -
T 4). As long as you are careful to Ohmmeter
shield the Radiation Sensor from
the Radiation Cube when measure-
ments are not being taken, TJetwill
be very close to room temperature Millivoltmeter Heat Shield
(reflectiveside
toward cube)
Procedure Figure 4.1 Equipment Setup
(D Set up the equipment as shown in
Figure 4.1. The Radiation Sensor should be pointed directly at the center of one of the
better radiating surfaces of the cube (the black or white surface). The face of the Sensor
should be parallel with the surface of the cube and about 3 to 4 cm away.
With the Thermal Radiation Cube off, measure R , the resistance of the thermistor at room
temperature. Enter this data in the space on the following page.
O) Shield the sensor from the cube using the reflecting heat shield, with the reflective side of
the shield facing the cube.
O) Turn on the Radiation Cube and set the power switch to 10.
6) When the thermistor resistance indicates that the temperature is about 12 c o above room
temperature, turn the power down so the temperature is changing slowly. Read and record
R, the ohmmeter reading, and Rad, the millivoltmeter reading. The readings should be
taken as nearly simultaneously as possible while briefly removing the heat shield. Record
these values in Table 4.1.
17
s c /ont/"c
ThcrmaJ Radiation Systcm
012-0469SD
IMPORTANT: Make each reading quickly, removing the heat shield only as long as it
takes to make the measurement. Take care that the position of the sensor with respect to the
cube is the same for all measurements.
6) Replace the heat shield, and turn the cube power to 10. When the temperature has risen an
additional 12-15 c o, repeat the measurements of step 5. Repeat this procedure at about 12-150
intervals until the maximum temperatureof the cube is reached.
Table 4.1
Data Calculations
Rad T4-T
(Y) (K)
s c 'ent'f/c
0120469SD
Thermal Radiation System
Teacher's Guide
Experiment 1: Introduction to Thermal Radiation
Notes on Questions
Black 100
Absorbtion and Transmission of Thermal Radiation
White 96.86
(D Heat loss through (closed) windows is primarily
conductive. Although the glass tested transmitted
Dull 20.23
some infrared, most was blocked.
Polished 738 A greenhouse allows light in, but does not allow
much heat to escape. This phenomenon is used to
@ Measurements are consistent with the rule, The bet- grow tropical plants in cold climates.
ter reflectors (poorer absorbers) are poor emitters.
Calculations
(J (3)
30 30
2.0(.0229L.2• (I A-1 )
Je2 - 9.822989L-J
25
s 20 s 20
15
10 c 10
5 5
o
o 10 20 30 50 00 o 0 02 0.04 0 06 o.oa 0.1 0.12 O.14 0.16
Dictanco (cm) 1/Ø2
19
oC/ont/'tc
Thermal Radiation System 012-046950
temperatures)
Experiment 3: Stefan-Boltzmann Law (at high
200 400 600 000 1000 1200 1400 1600 1800 2000
Temperature (K) 0.1
100 1000 10000
Temperature (K)
20
s c'ønt't'c
012-046951) Thermal Radiation System
The lamp filament is not a true black body. If it (3) Any other thermal source in the room would influ-
were, it would be completely and totally black at ence the results, including the warm body of the ex-
room temperature. It is a fairly good approximation, perimenter and the room itself. These introduce
though, as long as the temperature is high enough some error, but it is small as long as the tempera-
that the emitted light is much greater than the inci- ture of the lamp is high compared to the tempera-
dent light. ture of these other sources.
Difference in
Notes on Questions
0) The linearity of this graph indicates that the Stefan-
Boltzmann equation is correct, even at low tempera-
tures.
@ The graph should be straight, with some statistical
variations.
21
s c / enti/tc
Technical Support
For Technical Suppon call us at 1-800-772-8700(toll- • If your problem is with the PASCO apparatus, note:
free within the U.s.) or (916) 786-3800. Title and Model number (usually listed on the label).
Approximate age of apparatus.
A detailed description of the problem/sequence of
events. (In case you can't call PASCO right away,
you won't lose valuable data.)
If possible, have the apparatus within reach when
calling. This makes descriptions of individual parts
much easier.
• If your problem relates to the instruction manual,
note:
Part number and Revision (listed by month and year
on the front cover).
Have the manual at hand to discuss your questions.
Includes
Teachers Notes
and
Instruction Manualand 012-061230
Typical
Expenment Results Experiment Guide for the
PASCOscientific
Model AP-8210
Table of Contents
Section
Page
Copyright, Warranty, and Equipment Return
ii
Introduction
1-2
Equipment
Equipment Setup
Aligning the Optical System
Functions of Controls
Adjusting and Measuring the Voltage.............
Determining the Temperature of the Droplet Viewing Chamber .
Experimental Procedure .... . . ..
Computation of the Charge of an Electron................
Using a Projection Microscope with the Millikan Oil Drop Apparatus
Historical 11-15
Maintenance Notes
Cleaning ................................. .16
Replacing the Halogen Bulb .. .16
Adjusting the Vertical Reticle and Viewing Scope Alignments 16
Adjusting the Horizontal Reticle Alignment.......... 17
Touching Up the Black Painted Surface on the Plastic Spacer 17
Appendix
A: Viscosity of Dry Air as a Functionof Temperature 19
B: Themistor Resistanceat VariousTemperatures .20
INTRODUCTION
The electric charge carried by a particle may be calculated equal to the charge of the electron multiplied by the
by measuring the force experienced by the particle in an number of molecules in a mole. Through electrolysis
14
electric field of known strength. Although it is relatively experiments, the faraday has been found to be 2.895 x 10
easy to produce a known electric field, the force exerted electrostatic units per gram equivalent weight (more 7
by such a field on a particle carrying only one or several commonly expressed in the mks system as 9.625 x 10
excess electrons is very small. For example, a field of coulombs per kilogram equivalent weight). Dividing the
1000 volts per cm would exert a force ofonly 1.6 IO- faraday by the charge of the electron,
dyne on a particle bearing one excess electron. This is a
2.895 x 1014 e.s. uJgm equivalent weight
force comparable to the gravitational force on a particle
with a mass of IO- (one million millionth) gram. e-s.u.,
yields 6.025 x 10 molecules per gram equivalent weight,
The success of the Millikan Oil Drop experiment depends or Avogadro's number.
on the ability to measure forces this small. The behavior
of small charged droplets ofoil, having masses ofonly
10 12gram or less, is observed In a gravitational and an EQUATION FOR CALCULATING THE
electric field Measuring the velocity of fall of the drop in CHARGE ON A DROP
air enables, with the use of Stokes' Law, the calculationof An analysis of the forces acting on an oil droplet will
the mass of the drop. The observation of the velocity of yield the equation for the determination of the charge
the drop rising in an electric field then permits a carried by the droplet.
calculation of the force on, and hence, the charge carried
by the oil drop. Figure I shows the forces acting on the drop when it is
falling in air and has reached its terminal velocity.
Although this experiment will allow one to measure the (Terminal velocity is reached in a few milliseconds for the
total charge on a drop, it is only through an analysis of the droplets used in this experiment.) In Figure l, v is the
data obtained and a certain degree ofexpenmental skill velocity of fall, k is the coefficient of friction between the
that the charge ofa single electron can be determined. By
air and the drop, m is the mass of the drop, and g is the
selecting droplets which nse and fall slowly, one can be
acceleration of gravity. Since the forces are equal and
certain that the drop has a small number ofexcess
opposite:
electrons. A number ofsuch drops should be observed
and their respective charges calculated. If the charges on (1)
these drops are integral multiples ofa certain smallest
charge, then this is a good indication of the atomic nature
of electricity. However, since a different droplet has been
used for measunng each charge, there remains the
question as to the effect of the drop itself on the charge.
This uncertainty can be eliminated by changing the charge
on a single drop wllile the drop is under obsenation. An
mg mg
iomntion source placed near the drop will accomplish
this. In fact, it is possible to change the charge on the
same drop several times. If the results of measurements
on the same drop then yeld charges Inch are tmegral Figure 1 Figure 2
multiples of some smallest charge, then this is proof of the
atomic nature ofelectncity. Figure 2 shows the forces acting on the drop when
it is
nsmg under the influence of an electric field. In Figure
The measurement of the charge of the electron also 2,
E is the electnc Intensity, q is the charge camed
permits the calculauon of Avogadro's number. The by the
amountof currentrequired to electrodeposit one gram drop, and v, ts the velocity of rise. Adding the forces
equivalent ofan element on an electrode (the faraday) is vectonally yields.
Millikan 011 Drop Cxrcnrnent
012-0612JD
2gp (5) The definitions of the symbols used, together with their
proper units for use in equation ( 9 ) are* **
Stokes' Law, however, becomes incorrect when the
velocity of fall of the droplets is less than 0.1 cm/s. q—charge, in e.s.u. , can-ied by the droplet
(Droplets having this and smaller velocities have radii, d—separation of the plates in the condenser in cm
on
the order of 2 microns, comparable to the mean free path p —density ofoil in gm/cm 3
of air molecules, a condition which violates one of the g— acceleration of gravity in cm/s 2
assumpuons made in deriving Stokes' Law.) Since
the n —viscosityof air in poise ( dyne s/cm )
velocities of the droplets used in this experiment
will be in b—constant, equal to 6. 17 x 10 (cm ofHg) (cm)
the rowe of0.0J to crn/s, the viscosity must be
rnultjpJJed by a correction factor. The resulting p —barometricpressure in cm ofmercury.
effective a —radius of the drop in cm as calculated by equation
Viscosity is:
(5)
vy—velocity of fall in cm/s
(6) vr—velocity of rise in cm/s
where b is a constant, p is the atmospheric pressure, V—potentialdifference
and a across the plates in volts
IS the radius of the drop as calculated by the
uncorrected Note: The accepted value
form of Stokes' Law, equatjon ( 5 e.s.u.,
fore is 4.803 x 10-10
or 1.60x 10-1
Substituting jn equation (6) into equation coulombs.
(5), and
then soJvjng for the radius a gives:
*For additional
information about Stokes' Law, the
student is referred
to Introduction to Theoretical Physics,
by L. Page (New York, Van
Nostrand), Chapter 6.
(7) A derivation may
Substituting equations (4), (5), ond (6) info be found in The Electron by R. A.
equation (3) Millikan (Chicago,
yjclds The University of
Chapter 5. Chicago Press),
Modern calculations
units. (See Evperimenta/ of q are usually conducted in SI
(8) Charge of an Electron, Procedure, Computation of the
page 7.)
012-061230 Millikan Oil Drop Expenment
EQUIPMENT
Included equipment:
plato charging switch
• apparatus platform and plate charging switch (see
detailed description below and on page 4)
atomizer
• 12 volt DC transformer for the halogen lamp
• non-volatile oil (Squibb #5597 Mineral Oil, density =
oil
SS6
• atomizer
thermlstor connectors
halogen lamp housing
convex lens
filament adjustment knob
(vertical)
droplet viewing chamber housing
lamp power jack
Componcntsof platform:
It is recommended that you store the
• droplet saewqngchamber (see details below)
equipment in the original packing material.
• oeuing scope (30X, bright-field, erect image) with After unpacking, remove the foam insert
reticle (line separation: 0.5 mm major divisions, 0.1 from the droplet viewing chamber. Store
mm mmor divisions), reticle focusing ring, and the plate charging switch on the velcro tabs
droplet focusing ring located on the platform.
• halogen lamp (12 V, 5 W halogen bulb and dichroic,
infrared heat-absorbing window, horizontal and Required equipment, not included:
vertical filament adjustment knobs) • high voltage, well regulated power supply that
• focusing wire (for adjusting viewing scope) delivers up to 500 V DC, 10 mA minimum (for
example, the PASCO SF-9585 High Voltage Power
• plate voltage connectors Supply)
• thermistor connectors (thermistor is mounted in the digital multimeter (to measure voltage and
resistance)
bottom plate)
(for example, the PASCO SB-9599A Universal
Digital Multimeter)
WARNING: Do not apply voltage to the connectors (4) (for
• patch cords with banana plug
thermistor connectors. Banana Plug Patch
example, the PASCO SE-9415
Cord)
• therrmstor table (resistance versus temperature) PASCO SE-8702A
• stopwatch (for example, the
• mintion source lever (with three positions: Digital Stopwatch)
IonizationON, Jonintion OFF, and Spray Droplet
Additional recommended equipment:
Posmon)
stand
• PASCO ME-8735 Large Rod
• bubble level 45 cm (2)
• PASCO ME-8736 steel Rods,
• suppnlrod mountsand screws (to pennit mounting
of
platform on a PASCO ME-8735 Large Rod
Stand, so lid
vjewjng uope can be raised to a comfortable
eye
level)
• 3 leveling feet
• plate charpng switch (on a I meter
cord to prevent
vibranon of platform during switching housing
activity)
Components of dropjct viewing
cjj;unbej• (Figure 5)
• housing
• droplet holc cover
dropjot holo covor upper capacitor
support
rods
rod stand
tip
Although the dichroic window reflects
much of the heat generated by the halogen
bulb, the temperature inside the droplet shaft
viewing chamber may rise after prolonged
exposure to the light. Therefore, the
temperature inside the droplet viewing Figure 9. Correct position of the atomizer tip
chamber should be determined periodically
(about every 15 minutes).
3. Move the ionization source lever to the Spray Droplet
Position to allow air to escape from the chamber
during the introduction of droplets into the chamber.
4. Place the nozzle of the atomizer into the hole on the lid
of the droplet viewing chamber.
5. While observing through viewing scope, squeeze the
atomizer bulb with one quick squeeze. Then squeeze
it slowly to force the droplets through the hole in the
droplet hole cover, through the droplet entry hole in
the top capacitor plate, and into the space between the
two capacitor plates.
6. When you see a shower of drops through the viewing
scope, move the ionization source lever to the OFF
position.
7
Millikan 0/1 Drop Cmpcnment
012-06123D
Selection of Drop
L [•rornthe drops In View, 2. Calculate the charge on the droplet. If the result of this
select a droplet that
slowly (about 002-0.05 both falls first determination for the charge on the drop is greater
mm/s) when the plate
chargjng switch ISin 'Jit: than 5 excess electron, you should use slower moving
"Plates Grounded"
aruj can be dnven up position droplets in subsequent determinations.
and down by
voltage, turnjng on the
3. Introducemore oil
droplets into the chamber using the
Jlinf: A drop that rcqujres
procedure previously described and select another
about 15 droplet
seconds to fall the distance
between the major
telic le Jtnes (0 5 mm) 4. Measure the
Willrise (be same nse and fall velocities of the selected
distance. under the influence droplet about 10-20
of an electnc times or until the charge changes
V/cm), jn the followjng field
tunes With'lie spontaneously or the droplet
moves out of view.
8
012-061230 Millikan Oil Drop Expenment
S. Bring the droplet to the top of the field of view and Suggested Procedure for Computation of the Charge
move the ionization lever to the ON position for a few of an Electron:
secon&s as the droplet falls.
1. Using equation (7) (page 2), calculate the radius (a) of
6. If the rising Nelocity of the droplet changes, make as the oil drop:
many measurements of the new rising velocity as you
can (10 to 20 measurements). b 2
¯
7. If the droplet is still in view, attempt to change the
2g(p)
charge on the droplet by introducing more alpha
2. Substitute a from the above equation to find the mass
particles, as described previously, and measure the
(m) of the oil droplet:
new rising velocity 10—20times, if possible.
8. Repeat step (7) as many times as you can.
m = —rip
9. Record the plate potential, the oil density, the viscosity
of air at the temperature of the droplet viewing 4 2
chamber, (see appendix A), and the barometric 3 2gp
pressure for each set of velocity measurements.
9
Millikan Oil Drop L.xrcnment 012-061230
Using a Projecting Microscope with b. Proceed with step I of Aligning the Optical
the Millikan Oil Drop Apparatus
System.
2. Align and focus the projecting microscope:
To demonstrate Millikan 's experiment for an entire a. Set up the projecting microscope as directed in
classroom on a TV screen or computer monitor, use a the manufacturer's instructions.
projecting mjcroscope, such as the Ken-A-Vision b. After completing the setup and focusing of the
VideoFlex Microscope (PASCO Model No. SE-7227). Millikan Apparatus on the oil droplets, carefully
slide the eyepiece of the projecting microscope
Procedure: over the eyepiece of the Millikan Apparatus.
1. Puta white paper screen in the droplet viewing c. Focus the projecting microscope as directed in
chamber to reflect more light: the manufacturer's instructions.
I
a. Prior to focusing the viewing scope (step
under the Equipment Setup section, Aligning
screen
the Optical System), place a white paper
chamber as illustrated in
in the droplet viewing
Figure 10.
o
Cut a 32 x 6 mm
strip of white
paper and bend
back tabs of
6 x 6 mm.
Place the
paper screen lens
so the light
path Is not
blocked and
so the light is path of
reflected light
towards the source
lens of the
plastic
spacer.
paper screen
scopo
Top View
10
012-061231) Millikan Oil Drop Experiment
Historical
HISTORICAL NOTES acid, positively charged hydrogen and oxygen gasses
were produced (although there were one million mil-
The Greeks were the first to report the effects of lion neutral molecules to every charged one). This
elect-icity when they recorded that rubbed amber method was used to produce an ionized gas that was
attracted light objects. However, theories explaining this then bubbled through water to form a cloud. For his
phenomenon did not emerge until 1747, when Benjamin determination Ofe Townsend proceeded in the follow-
Franklin proposed that an electrical fluid or fire existed in ing manner:
certain amounts in all matter. An excess of this fluid in
matter would produce a positive charge and a deficiency 1. He assumed that in saturated water vapor each ion
of this fluid would produce a negative charge. A slightly condensed moisture about it, so that the number of
different theory was put forth by the physicist Symmer ions was the same as the number of droplets.
twelve years later. He proposed that matter in a neutral 2. He determined with the aid of a quadrant
state shows no electrical properties because it contains electrometer the total electrical charge per cubic
equal amounts of two weightless fluids, which were centimeter can-iedby the gas.
called positive and negative electricity respectively.
3. He found the total weight of the cloud by passing it
Franklin also postulated the existence ofan electrical through drying tubes and determining the increase in
particle small enough to easily permeate matter. weight of these tubes.
Faraday's experiments in electrolysis, which
demonstrated that when a current is passed through an 4. He found the average weight of the water droplets
electrolyte, the masses of compounds deposited at constituting the cloud by observing their rate of fall
opposite electrodes are in proportion to the chemical under gravity and computing their mean radius with
equivalent weights of the compounds, also supported the aid ofa purely theoretical law known as Stokes'
Franklin's concept of an elementary electrical particle. Law.
The fluid theories, along with a theory explaining 5. He divided the weight of the cloud by the average
electricity as a state of strain in matter, were the prime weight of the droplets of water to obtain the number of
explanations of electrical phenomena until late in the droplets which, if assumption I is correct, was the
19th century. number of ions, and he then divided the total charge
per cubic centimeter in the gas by the number of ions
EARLY DETERMINATIONS OF e to find the average charge carried by each ion, that is,
to find e.
The word "electron" was first suggested in 1891 by Dr.
G. Johnstone Stoney as a name for the "natural unit of Townsend achieved results in the range of 3 x e.s.u.
electricity," namely, that quantity of electricity that must for e. J. J. Thompson, in 1900, used a method similar to
pass through a solution in order to liberate at one Townsend's and obtained a value of 6 x 1010e.s.u. In
both of these methods, however, the first assumption
electrode one atom of hydrogen or any univalent
(each droplet formed around only one ion) proved to be
substance. It would follow that the charge of the electron
only approximately correct, and the experimental methods
multiplied by the number of molecules in a gram mole
were not adequate to provide a precise determination of e.
would give the amount ofelectricity required to deposit
one gram mole by electrolysis. This quantity had been I-KS.Wilson improved upon Townsend's and
determined by Faraday to be 9650 absolute Thompson's work by adding B.vobrass plates which could
electromagnetic units of electricity. Using this method, be connected to a 2000 volt battery. A cloud was formed
Stoney obtained a value of 03 x 10 10es.u. (The Kinetic between these plates (not Charged) and the falling velocity
Theory provided the basis for Stoney's estimation of
Avogadro's number).
condensed from Robert A. Millikan's book, The
The first experimental attempt to measure the charge
Electron (University of Chicago Press, Chicago, 1993,
of an ion was made by Townsend in the late 1890's.
pp. 45-46) and used with permission of the publishers.
He had observed that during electrolysrs of sulfuric
11
Millikan Oil Drop t;spenmcnt 012-061230
of the cloud recorded. A second cloud was then formed placed from 3 to 10 cm away from the plates. A second or
and its falling velocity observed in an electric field (the so after the cloud was producect 3 the radium was removed
plates being charged). Since the two velocities are . and the field thrown on by means a double-throw
proportional to the forces acting on the drops , and the switch. If the drops were not found to be held suspended
velocity ofthe cloud with the plates uncharged determines by the field the potential difference was changed ... The
the size and mass of the drops by Stokes' Law, Wilson cross-hairs were set near the lower plate, and as soon as a
ums able to obtain a value of 3 x 10 10e.s.u. fore. Since stationaty drop was found somewhere above the upper
Wilson's measurements were always made on the top of cross-hair, it was watched for a few seconds to make sure
the cloud, or the drops with the smallest charge (the more that it was not moving and then the field was thrown off
heavily charged drops being driven downward faster in and the plates short-circuited by means of the double-
the field), the assumption of one ion per drop was throw switch, so as to make sure that they retained no
charge. The drop was then timed by means of an accurate
stop watch as it passed across the three cross-hairs, one of
MILLIKAN'S DETERMINATION OF e the two hands of the watch being stopped at the instant of
passage across the middle cross-hair, and the other at the
Millikan improved upon Wilson 's design by using a
instant of passage across the lower one. It will be seen
higher potential across the plates so that the falling
that this method of observation fumishes a double check
velocity of the cloud could not only be impeded, but
upon evaporation; for if the drop is stationary at first, it is
actually reversed. Some charged drops moved upward,
not evaporating sufficiently to influence the reading of the
some moved rapidly downward, while the uncharged
rate of fall, and if it begins to evaporate appreciably before
drops were unaffected and continued to drift downward.
the reading is completed, the time required to pass
A few drops, which carried a charge of the proper through the second space should be greater than that
magnitude so that the force of gravity on the drop almost required to pass through the first space. It will be seen
equaled the force of the electric field on the drop, from the observations which follow that this was not, in
remained in view. By varying the potential of the plates, general, the case.
Millikan couldjust balance these drops. This situation
proved to be a significant improvement for it permitted all It is an exceedingly interesting and instructive experiment
measurements to be made on a single drop. By using this to watch one of these drops start and stop, or even reverse
balanced drop method, Millikan was able to observe the its direction of motion, as the field is thrown offand on. I
properties of individual ions and to determine whether have often caught a drop which was just too light to
different ions carry one and the same charge. remain stationaty and moved it back and forth in this way
four or five times between the same two cross-hairs,
In the following passage, taken from the "Philosophical watching it first fall under gravity when the field was
Magazine" for February, 1910,Millikan describes the thrown off and then rise against gravity when the field
actual procedure of the experiment. was thrown on ....
"The observations on the rate of fall were made with a
Furthermore, since the observations ... are all made upon
short-focus telescope placed about 2 feet away from the
the same drop, all uncertainties msto whether conditions
plates. In the eyepiece of this telescope were placed three
can be exactly duplicated in the formation of successive
equallyspacedcross-hairs.... A small section of the
clouds obviously disappear. There is no theoretical
space between the plates wcs illuminated by a narrow
uncertainty whatever left in the method unless it be an
beam from an arc light, the heat of the arc being absorbed
by three water cells in series. The air between the plates uncertainty as to whether or not Stokes' Law applies to
was ionized by 200 mg of radium of activity 20,000 the rate of fall of these drops under gravity."
Experiments with the balanced water drop produced the
2
With the plates uncharged the iörce is mg where m is value of 3.422 x 10 10e.s.u. fore. The most important
the nurss ofthe drop and g is the acceleration aspect of these experiments, however, was the observation
of grav-
ity. With the plates charged the force is mg ±
Ee by Millikan that a rising drop would suddenly change its
where E is the electric intensity between velocity. This phenomenon could easily be produced by
the plates
and en is the charge on the drop.
placing a radioactive source near the drop. This
The italicizedphrases indicate demonstrated that the drop had "captured" an ion, thus
a slight change in
wording, for purposes of clarity, changing the charge of the drop and its respecuve
from Millikan's
original work.
velocity.
12
012-OS123D Millikan Oil Drop Expenmznt
THE EXACr EVALUATIONOF e Ofe possible. The following passage is part of Millikan's
description of the apparatus, including a diagram of the
apparatus designed for the observation of single oil drops device.
for extended periods of time. Since water drops had "Accordingly, J built two years ago a new condenser
proved inadequate for prolonged observation ofthis ion having surfaces which were polished optically and made
catching phenomenon, Millikan used oil drops, which flat to within two wave-lengths of sodium light They
were not affected by evaporation. The apparatus were 22 cm. in diameter and were separated by three
consisted oftwo parallel brass plates separated by a pieces of echelon plates, 14.9174 mm. thick, and having
distance of 16 mm by ebonite blocks. Non-volatile oil optically perfect plate surfaces. The dimensions ofthe
was sprayed into the chamber above the plates, and small condenser, therefore, no longer introduced an uncertainty
drops slowly found their way into the area between the of more than about I part in 10,000.
plates through a small hole in the top plate. The drops
"Complete stagnancy of the air between the condenser
were illuminated by a beam from a carbon arc lamp and
plates was attained, first, by absorbing all the heat rays
were observed through a measuring scope. The detailsof
from the arc lamp by means of a water cell 80 cm. long,
the construction of Millikan•s final apparatus built in
and a cupric chloride cell, and secondly, by immersing the
1914 (which was basically similar to his earlier devices,
whole vessel in a constant temperature bath of gas-engine
and for the purposes of this discussion can be considered
oil (40 liters), which permitted, in general, fluctuations of
the same as the earlier pieces of apparatus) attest to the
not more than 0.02 oc during an observation-
effort expended in obtaining the most accurate evaluation
J J —A. atomzzr through *hicb thz bloun tnto thruylrndncal vessel D G, 011tank to keep the tem-
peraturec.ocstant BfaruJ,V. circulat brasl pläiZ', electncal fizlrf hy throwng or I G.OOO-voItbatteryB. Light
frum arc l•np a afia hat raj" are removed by razoge throuÉh and enters chamber through glass '.qndow g and Illu-
droplzl.p bctae.en plates grand V through the pmholc If Additional tcns are produced aboutp by X-rays from
bulb
X
'MtJ11kan,Robert X, p. J IS. 7
Millikan. Robert p. J 16.
C
M11hkan, Robert A, J 10.
13
Millikan Oil Drop Lxpenmcnt 012-0612M)
With this new apparatus hundreds of measurements on charge. The column headed 4.917 x n contains simply the
different drops were made, for the purpose of both making whole series of exact multiples of this number from I to
an exact evaluation Ofe and proving or disproving the 18. The column headed 'Observed Charge' gives the
atomic theory of electricity. The value Ofe that was successive observed values of the rising velocity of the
obtained Cromthese five years of work was 4.774 x 10-10 drop plus the falling velocity. Jt will be seen that during
e.s.u. This value Ofe was accepted until 1928 when a the time of observation, about four hours. this drop camed
precise determination of Avogadro's number by X-ray all possible multiples of the elementary charge from 4 to
diffraction measurements on crystals permitted the 17, save only 15. No more exact or more consistent
calculation Ofe to be 4.803 x 10 10e.s.u. The discrepancy multiple relationship is found in the data which chemists
was later traced to Millikan's too low value for the have amassed on the combining powers of the elements
viscosity of air. and on which the atomic theory ofmatter rests than is
found in the foregoing numbers.
ATOMIC NATURE OF ELECTRICITY Such tables as these—and scores of them could be
given—place beyond all question the view that an
The atomic nature of electricity is best exemplified by the electricalcharge wherever it is found, whether on an
following table taken from Millikan's data: insulator or conductor, whether in electrolytesor in
metals, has a definite granular structure, that it consists of
n 4.917 x n Observed an exact number of specks of electricity (electrons) all
Charge exactly alike, which in static phenomena are scattered
4.917 over the surface of the charged body and in current
phenomena are drifting along the conductor. Instead of
2 9.834 giving up, as Maxwell thought we should some day do,
3 14.75 the 'provisional hypothesis of molecular charges,' we find
ourselves obliged to make all our interpretationsof
19.66 19.66
electrical phenomena, metallic well as electrolytic, in
5 24.59 24.60 terns of it."
6 29.50 29.62 Although the values of the charge on a specific drop were
34.47 found to be exact multiples of a certain value (e), the value
7 34.42
Ofe varied for drops ofdilTerent masses. This
8 39.34 39.38 discrepancy was traced to the breakdown ofStokes• Law.
44.42 Through experimentation the law was found to fail when
9 44.25
the size of the drop approached the mean free path of air
49.17 49.41 molecules. When this situation occurs, the medium in
10
54.09 53.91 which the drop falls is no longer homogeneous In relation
to the drop. This contradicts one of the upon
12 59.00 59.12 which Stokes' Law is based. Through his work on the
63.92 63.68 electron, Millikan was able to determjne a correction
13
factor for Stokes' Law.
68.84 68.65
By performing the expenrncnt Withmercury drops and
15 73.75 drops of other materials. Millikan demonstrated that the
16 78.67 78.34 elementary electrical charge was the same for Insulators,
semi-conductors, and conductors. Ile also demonstrated
17 83.59 83.22
that the beta particle had the same charge msan electron
18 88.51 (indeed, it is an electron) and that positive and negative
SuggestedReading
Should the student desire a more detailed back ground in
this classic experiment, the following references are
suggested:
l. Millikan, Robert A. , The Electron, (Chicago, The
University ofChicago Press, 1917 (reprinting in
paperback form, 1963).
2. Millikan, Robelt A., "The Isolation ofan Ion, A
Precision Measurement of its Charge, and the C01Tection
of Stokes' Law," The Physical Review, Vol. 2, No. 2, pp.
109-143, June 1913.
3. Millikan, Robert A., "On the Elementaw Electrical
Charge and the Avogadro Constant," ThePhysical
Review, vol. 32, No. 4, pp. 349-397, April, 1911.
4. Shamos, M.H., Great Experinlents in Physics (Holt-
Dryden, New York, 1959), pp. 238—249.
15
012-0612M)
Millikan 011 Drop Especimcnt
Maintenance Notes
Cleaning Adjusting vertical reticle and viewing scope
alignments
1. The housing of the droplet viewing chamber, the
capacitor plates, the plastic spacer, and the droplet Ifthe alignment of the reticle or viewing scope is altered
hole cover should be cleaned with water and during rough handling, realign it using the following
detergent, with particular attention to the droplet hole procedure:
in the top capacitor plate, the glass observation port 1. Loosen the set screw in the viewing scope holder
covers on the housing, and the droplet hole cover. (Figure 12).
2. The plastic spacer should be polished with a soft, lint-
free cloth to remove any oil, finger prints, or lint.
3. The lens on the plastic spacer should be cleaned on
both sides using a Q-tip.
4. Apply a thin film of oil to the capacitor plates to help
prevent corrosion.
5. Dry all parts completely before reassembly.
sot scrow
Always handle the plastic spacer and capacitor
plates chrefully to avoid scratching them.
Replacing the halogen light bulb 2. With the focusing wire in place and while looking
through the eyepiece, rotate the viewing scope until
l. Disconnect the apparatus from all power sources. the vertical reticle lines are vertical to the focusing
2. Remove the four screws on the halogen lamp housing
and lift off the top cover. Gently pull the halogen bulb 3. Find the center of focus in the adjustment knob on the
out of its socket. viewing scope (this will be approximately half-way
3. Replace with a GE #18426 halogen bulb (12 V, 5 W, between minimum and maximum focus).
T3 type with 2-Pin G4 base, C6 straight filament). 4. Manually move the viewing scope in and out through
Carefully insert the pins at the base of the bulb into the its holder until the focusing wire comes into focus.
socket and press firmly to seat the bulb securely.
5. Recheck the reticle to assure that it is still in proper
alignment with the focusing wire (as in step 2).
Handle the new halogen bulb only with tissue
6. Lock the viewing scope into position by tightening
paper—oil on the hands may damage the
bulb.
the set screw into the viewing scope holder.
16
012-061230 Millikan Oil Drop Expenrnent
Adjusting the horizontal reticle alignment Touching up the black painted surface on the
If the horizontal alignment of the viewing scope is altered plastic spacer
during rough handling, realign it using the following After prolonged use and repeated cleaning, the black paint
procedure: (Figure 14) that absorbs refracted and reflected light on
1. Loosen one of the two socket head cap screws on the the plastic spacer may begin to wear off. In that event,
touch up the surface with a thin coat of flat black acrylic
bottom of the platform shown in Figure 13.
paint such as that available at hobby stores. Do not use a
2. With the focusing wire in place and while looking lacquer or oil-based paint.
through the eyepiece, gently tap the viewing scope
until the focusing wire is centered in the reticle.
3. Lock the viewing scope into position by tightening the black-painted area
two socket head cap screws into the viewing scope
holder.
bottom of platform
17
012-061231) Millikan (Dil Drop Experiment
Appendix
Appendix A: Viscosity of Dry Air as a Function of Temperature*
1.8840
1.8800
1.8760
1.8720
1.8680
1.8640
1.8600
1.8560
1.8520
1.8480
x 1.8440
E 1.8400
z 1.8360
1.8320
1.8280
1.8240
1.8200
1.8160
1.8120
1.8080
1.8040
1.8000
15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32
Temperature 2C
19
012-0612M)
Appendix B:
Various Temperatures
Millikan Oil Drop Apparatus Thermistor Resistanceat
20
012-061231) Millikan Oil Drop Experiment
Teachers Guide
Note: It is best that students work in pairs—one to observe the drop and one to
record the experimental data.
Note: Leveling will be most accurate if the bubble level is observed from
directly above during leveling.
Note: If more accuracy with leveling is needed to prevent the oil droplets from
gradually drifting off to one side during prolonged observations, perform the level-
ing operation using a two-dimensional level or ball bearing placed directly on the
bottom capacitor plate.
As an example of typical experimentalresults, the following pages list one teacher's data, using the
alternative method for calculating the charge on an electron that is presented on page 9. The
teacher measured the velocity of the drop moving down with the plates charged and used the
method listed below to organize the data for computation.
For each different charge event for a drop the following was done:
1. Measured the velocity of the falling drop with the plates not charged, the velocity of the rising
drop with the plates charged, and the velocity of the falling drop with the plates charged;
2. Assigned a "charge letter" to each event of differing charge for each oil droplet; for example, for
drop l: IA, 0 (velocity of the falling drop with the plates not charged), 1B, u (velocity of the
rising drop), 1B, d (velocity with the plates charged of the falling drop), IC, u, d, and ID, u, d ;
for droptlet 2: 2A0, u, d, 2B u, d, 2C u, d, and so on.
3. Averaged all measurements for a drop falling with the plates not charged, for use to determine a;
4. For each charge letter, averaged the measurementsfor the cases of the drop rising or falling with
the plates charged;
5. Calculated the average charges on each droplet for each charge letter (averaging the charges for
the cases with the droplet rising and with the droplet falling while the plates were charged);
6. Listed the average charges for the charge letters in order of increasing size and calculated the av-
erage
difference between charges;
7. Calculated the number of e for droplet under each of the charge conditions by dividing the aver-
age charge for each charge letter by the average difference in charge from step 5, above.
The data and calculations for measurementson one droplet are listed on the followingpages.
21
012-0612M)
= 28.80 C
Sample Data for Millikan Apparatus: Voltage = 386 V; Temperature
Drop#/
Distance timed (mm) Time (s) Direction:
Charge Letter
0 = no field
U = going up
D = going down
0.5 18.24
0.5 18.56
0.5 19.24
0.5 18.05
0.5 17.23
0.5 15.35
0.5 16.70
0.5 17.99
0.5 15.35
0.5 1725
0.5 18.38
0.5 18.32
JD 0.5 16.56
0.5 18.70
0.5 16.56
1.0 33.63
0.5 17.30
0.5 19.06
0.5 18.33
0.5 16.21
0.5 15.36
0.5 15.70
0.5 17.10
0.5 17.30
0.5 17.80
22
012-0612M)
Miljikan Oil Drop Expcnmcnt
of
l. The temperature was T = 0.822 mQ = 28.80C, which gives the viscosity
air = n = 1.867x 10-5 kg/m•s.
2. For this droplet, a = 4.9 x IO- m ± 0.2 x IO- m.
gives the
3. The plate separation was 0.767 cm and the plate voltage was 386 V, which
electric field = E = 5.03 x 104 V/m.
4. The density of the oil was 886 kg/m , which gives the weightof the
droplet = mg = 4.26 x 10 15N.
5. The pressure was 1.01 x 105 Pa.
Table of Results
Drop# / Charge for Going Up Charge for Going Down Average Charge
Charge Letter (x 10 19C) (x 10 19C) (x 10 19C)
24
Technical Support
Feedback Contacting Technical Support
If you have any comments about the product or Before you call the PASCO Technical Support staff, it
manual, please let us know. If you have any sugges- would be helpful to prepare the following informa-
tions on altemate experiments or find a problem in the tion:
manual, please tell us. PASCO appreciates any
customer feedback. Your input helps us evaluate and If your problem is computer/softwarerelated, note:
improve our product. - Title and revision date of software;
To Reach PASCO - Type of computer (make, model, speed);
For technical support, call us at 1-800-772-8700(toll- - Type of external cables/peripherals.
free within the U.S.) or (916) 786-3800.
If your problem is with the PASCO apparatus, note:
fax: (916) 786-3292
- Title and model number (usually listed on the
e-mail: techsupp@PASCO.com label);
web: www.pasco.com - Approximateage of apparatus;
- A detailed description of the problem/sequence
of events. (In case you can't call PASCO right
away, you won't lose valuable data.);
- If possible, have the apparatuswithin reach when
calling to facilitate description of individual parts.
h/e Apparatus
and
h/e Apparatus Accessory Kit
Dener
Foothills Blvd. • Roseville, CA 95747-7100 rays to
s c / ent/ttc Phone (916) 786-3800 • FAX (976) 786-8905 • www.pasco.com teacnsoence
012-04049J We Apparatus and h/e Apparatus Accessory Kit
Tableof Contents
Section Page
Copyright, Warranty, and Equipment Return. ii
Introduction .
Background Theory
Equipment and Setup
Equipment List
Installing the Batteries .
Battery Voltage Check.....
Equipment Setup
Using the Accessory Kit Filters ..
Experiments:
Experiment l: Wave Model vs Quantum Model
Experiment 2: The Relationship of Energy, Wavelength
and Frequency ................... 11
Technical Information 13
Theory of Operation 13
Schematic Diagram......... . 14
Teacher's Guide..... ... .15
Technical Support Inside Back Cover
012-04049J
li te Apparatus and lite Apparatus Accessory Kit
Credits
This manual edited by: Dave Griffith
Teacher's guide written by: Eric Ayar
012-NNAJ We we Accesso—y
Introduction
The enussion and absorption of light w•asan early subject Experimental data for the radiation of light by a hot.
for imotigation by German physicist Max Planck As glowing body showed that the maximum Intensity of
Planck attempted to formulate a theory to explain the emitted light also departed dramatically from the clas-
spectral distribution ofemitted light based on a classical sically predicted values (Wien's Law). In order to rec-
model, he ran into considerable difficulty. Classical oncile theory with laboratory results, Planck was
theory (Rayleigh-Jeans Law) predicted that the amount of forced to develop a new model for light called the
by light emitted from a black body would increase dramati- quantum model. In this model, light is emitted in
call}'as the wavelength decreased, whereas experiment small, discrete bundles or quanta.
showed that it approached zero. This discrepancy became
The relationship between the classical and quantum theo-
known as the ultraviolet catastrophe.
ries for the emission of light can be investigated using the
PASCO scientific We Apparatus. Using Apparatus in
combination with the PASCO Mercury Vapor Light
Source (Model OS-9286) allous an accurate determina-
tion of the We ratio and thus a determination of h,
st be Planck's constant.
billty
the
Figure I. The h/eApparatus Shown With the Accessory Kit and MercuryVapor Light Source
0120049J
tVo Apparatus
h/o Apparatus Accessory
Kit
Background Theory
kinetic energy of the emit-
Planck's Quantum Theory whereKE is the maximumthe energy needed to re-
IV o is
By the late 1800's many physicists thought they had ex- ted photoerectrons,and
of the material (the work
plained all the main principles of the universe and discov- move them from the surface
E is the energy supplied by the quantum of
cred all the natural laws. But as scientists continuedwork- function).
ing, inconsistencies that couldn't easily be explainedbe- light known as a photon.
gan showing up in some areas of study. The h/e Experiment
upon an elec-
In 1901 Planck published his law of radiation.In it he A lightphotonwith energy hr is incident
stated that an oscillator, or any sijnilar physicalsystem, tron in the cathode of a vacuum tube. The electron uses a
has a discrete set of possible energy values or levels;cn- minimum of its energy to escape the cathode, leaving
ergies between these values never occur.
it witha maximumenergy of KE in the forn of kinetic
Planck went on to state that the emission and absorption energy. Nonnally the emitted electrons reach the anode of
of radiation is associated with ttansitions orjumps be- the tube, and can be measured as a photoelectric current.
tween two energy levels. The energy lost or gained by the However, by applying a reverse potential V between the
oscillator is emitted or absorbed as a quantum of radiant anode and the cathode, the photoelectric current can be
energy, the magnitude of which is expressed by the equa-
stopped.KE can be determined by measuring the mini-
tion:
mum reversepotential needed to stop the photoelectrons
E=hv and reduce the photoelectric current to zero. • Relating
where E equals the radiant energy, v is the frequencyof kineticenergyto stoppingpotential gives the equation:
the radiation, and h is a fundamental constantof nature.
The constant, h, became known as Planck'sconstant, Therefore, using Einstein's equation,
Planck's constant was found to have significancebeyond
relating the frequency and energy of light, and becajnea
When solved for V, the equation
cornerstone of the quantum mechanical view of the suba- becomes:
tornic world. In 1918,Planck was awardeda Nobelprize
Cor introducing the quantum theory of light. If we plot r vs v for different
frequencies of light, the
graph will look like Figure
The Photoelectric Effect 2. The V intercept is equal to -
and the slope is h/e.
In photoelectric emission, light strikes a material, causing Coupling our experimental de-
tennination of the ratio h/e
with the accepted value for
e, 1.602x 1009coulombs,
dicted tLat as the intensity of incident light was increased, we can determine Planck's
constant, h.
the amplitude and thus the energy of the wave would in-
crease. This would then cause more energeticphotoelec- Stoppjng
Irons to be emitted, The new quantum model, however, Potential
V
predicted that higher frequency light would produce
higher energy photoelectrons, independentof intensity,
while jncrcased 'litcnsity would only increasethe number
olelcctrons emitted (or photoelectric current). In the
early 1900s several investigators round that the kinetic Slope
= h/e
energy of lljc•photoelectrons was dependent on the
Wave-
length, or frequency, ond independent of intensity,
while
flit: magnitude of lhc•photoelectric current, or number
oc
electrons was dependent on the intensity as predicted
by Frequency v
Ille quantum model, Einstein applied Planck's
theoty
and Figure 2. The
the photoelectric effect in tenns of
model usjng his Camousequation for which he
the quantum graph of V vs. V
received *NOTE: In
tlje Nobel prjze jn 1921: paratus the
experiments with
stopping potential the PASCO Ap-
rather than is measured directly,
by monitoring
see the the photoelectric
current
motion section in the Technical Infor-
of the manual
2 for details.
h/e Apparatus and h/e Apparatus Accessory Kit
Light Shield
Base Support Rod (shown tilted to
the open position)
5
012-04049J
hfo Apparatus and
Apparatus Accessory Kit
Whito
+0
Ultraviolot
Bluo
Groon
Yellow
14. Press "PUSI 1 TO ZERO" button on the side panel Using the Filters
of the life Apparatus to discharge any accumulated po-
The (AP-9368)lb/eAppamtus
tcnfial in the unit's electronics. This will assure the Ap- includes three filters: one
paratus records only the potential oCthe light you are Green and one Yellow,plus
a Vanable Transmission Filter.
measuring. Note thal the output voltage will drill with The filter frames have
magnetic strips and mount to the out-
tljc•absence oc light on the photodiode, side ofthe White
Reflective NILSkof the h/e
Apparatus.
output voltage on your digital voltmeter. It is Use the green and
15. Read yellow filters when you're
green and yellow using the
a direct measurement of the stopping potential for the spectral lines. These
photoelectrons, (See frequencies filters limit higher
in the Tech- oflight fromentering
njcal Information section of tlje manual for an expla- prevents ambient the hie Apparatus. This
room light from
nation of' toe measurement.) lowercnergy yellow interfering with the
and green light and
results. It also masking the true
blocks the higher
NOTE: roc some apparatus, the stopping poten- from the higher frequency ultraviolet light
order spectra
lower orders of which may overlap with
yellow and green.
j to actual btoppjnc potential voltage.
The variable
Transmission Filter
generated patterns ofcomputer-
oCd0Lsand lines
(not frequency) that vary the intensity
oflhe incident
mission percentages light The relative trans-
Te 100%,
80%, 60%.40%, and 200/0.
01204049J h/e Apparatus and h/o Apparatus Accossory Kit
Setup
Set up the equipment as shown in the diagram below. Focus the light from the Mercury Vapor
Light Source onto the slot in the white reflective mask on the WeApparatus. Tilt the Light
Shield of the Apparatus out of the way to reveal the white photodiode mask inside the Appara-
tus. Slide the Lens/Grating assembly forward and back on its support rods until you achieve the
sharpest image of the aperture centered on the hole in the photodiode mask. Secure the
Lens/Grating by tightening the thumbscrew.
Align the system by rotating the WeApparatus on its support base so that the same color light
that falls on the opening of the light screen falls on the window in the photodiode mask, with no
overlap ofcolor from other spectral lines. Retum the Light Shield to its closed positiom
Check the polarity ofthe leads from your digital voltmeter (DVM), and connect them to the
OUTPUT terminals of the same polarity on the h/e Apparatus.
Procedure
Part A mask
the opening of the
l. Adjust the h/e Apparatus so that only one Ofthe spectral colors falls upon
corresponding colored
of the photodiode. If you select the green or yellow spectral line, place the
filter over the White Reflective Mask on the h/e Apparatus
(and over the colored
2. Place the Variable Transmission Filter in front of the White Reflective Mask
reaches the pho-
filter, if one is used) so that the light passes through the section marked 100% and
todiode. Record the DVM voltage reading in the table below.
is re-
Press the instrument discharge button, release it, and observe approximately how much time
quired to return to the recorded voltage.
3. Move the Variable Transmission Filter so that the next section is directly in front of the incoming
light. Record the new DVM reading, and approximate time to recharge after the discharge button
has been pressed and released.
Repeat Step 3 until you have tested all five sections of the filter.
80
60
40
20
Color #2
0/0Transmission
(name) Stopping
Potential
Approx. Charge
100 Time
80
60
40
20
8
012-04049J h/o Apparatus and h/o Apparatus Acconsory Kit
Part B
l. You can easily see five colors in the Inercury light spectrum, Adjust the h/e Apparatus so that
only one of the yellow colored bands Callsupon the opening of the mask oc the photodiode.
Place the yellow colored filter over the White Reflective Mask on the h/e Apparatus.
2. Record the DVM voltage reading (stopping potential) in the table below.
3. Repeat the process for each color in the spectrum. Be sure to use the green jilter when measur e
ing the green spectrum.
Analysis
l. Describe the effect that passing different amounts of the same colored light through the Vari-
able Transmission Filter has on the stopping potential and thus the maxtmum energy of the
photoelectrons, as well as the charging time after pressing the discharge button.
2. Describe the effect that different colors of light had on the stopping potential and thus the
maximum energy of the photoelectrons.
3. Defend whether this experiment supports a wave or a quantum model of light based
on your
lab results.
Explain why there is a slight drop in the measured stopping potential as the light intensity
is
decreased.
NOTE: While the impedance of the zero gain amplifier is very high (A it is not
infinite and some charge leaks off. Thus charging the apparatus is analogous to filling
a
bath tub with different water now rates while the drain is partly open.
Yellow
Green
Blue
Violet
Ultravjolct
9
012-04049J h/o Apparatus and h/o Apparatus Accossory Kit
Setup
Set up the equipment as shown in the diagram below. Focus the light from the Merct11YVapor
Light Source onto the slot in the white reflective mmskon the h/e Apparatus. Tilt the Light Shield
of the Apparatus out of the way to reveal the white photodiode mask inside the Apparatus. Slide
the Lens/Grating assembly fonvard and back on its support rods until you achieve the sharpest im-
age of the aperture centered on the hole in the photodiode mask. Secure the Lens/Grating by tight-
ening the thumbscrew.
Align the system by rotating the h/e Apparatus on its support base so that the same color light that
falls on the opening of the light screen falls on the window in the photodiode mask with no overlap
of color from other spectral bands. Retum the Light Shield to its closed position.
Check the polarity of the leads from your digital voltmeter (DVM), and connect them to the OUT-
PUT terminals of the same polarity on the h/e Apparatus.
Il
012-040490
h/o Apparatus and
h/o Apparatus Accossory Kit
Procedure
l. You can see five colors in two the WeApparatus
orders of the mercury lightspectrum.Adjust of
careiillly so that only one color from the first order (the brightestorder) fallson the opening
the mask of the photodiode.
that
2. For each color in the first order, measure the stopping potentialwith the DVMand record
measurement in the table below. Use the yellow and green colored filterson the Reflective
Mask of the h/e Apparatus when you measure the yellow and green spectrallines.
3. Move to the second order and repeat the process. Record your results in the table below.
Analysis
Determine the wavelength and frequency ofeach spectral line. Plot a graph of the stopping
potential vs. frequency.
Determine the slope and yr-intercept.Interpret the results in terms of the h/e ratio and the rro/e
ratio. Calculate h and W .
In your discussion, report your values and discuss your results with an interpretationbased on
a quantum model for light.
Yellow
Green
Blue
Violet
Ultraviolet
Yellow
Green
Blue
Violet
Ultraviolet
12
01204049J hie Apparatus and h/e Apparatus Accessory Kit
Technical Information
Theory of Operation Due to the ultra high input impedance, once the capacitor
In experiments with the h/e Apparatus, monochromatic has been charged from the photodiode current it takes a
light falls on the cathode plate of a vacuum photodiode long time to discharge this potential through some leak-
tube that has a low work function, age. Therefore a shorting switch labeled "PUSH TO
Photoelectrons
ejected from the cathode collect on the anode. Zero" enables the user to quickly bleed off the charge.
However, the op-amp output will not stay at 0 volts aner
The photodiode tube and its associated electronics have a the switch is released since the op-amp input is floating.
small capacitance which becomes charged by the photo-
electric current. When the potential on this capacitance Due to variances in the assembly process, each appara-
reaches the stopping potential of the photoelectrons, the tus has a slightly different capacitance. When the zero
current decreases to zero, and the anode-to-cathode volt- switch is released, the internal capacitance along with
age stabilizes. This final voltage between the anode the user's body capacitance coupled through the switch
and cathode is therefore the stopping potential of the is enough to make the output volatge jump and/or os-
photoelectrons. cillate. Once photoelectrons charge the anode the input
voltage will stabilize.
To let you measure the stopping potential, the anode is
connected to a built-in amplifier with an ultrahigh input
impedance (> 1013Q), and the output from this amplifier
is connected to the output jacks on the front panel of the
apparatus. This high impedance, gain (Vout/Vin= l)
amplifier lets you measure the stopping potential with a digi-
tal voltmeter.
13
01-z0U43J
Dl
111314
2- 7
71
4
UI
VACUJJLI
1%314
1733
70
33._2 71—4
77—2*
Schemztic Dizgram
14
012040490 h/e Apparatus and h/e Apparatus Accessory Kit
Teacher's Guide
Part A
2
In each case, the
time to reach full 1.8
0.2
o 10 20 30 40 50 60 70 80 90 100
Part B Irierrsity
(X)
Analysis
l. The amount of light does not 2
significantly affect the f(x) —
- 4.145193E-1S4x + -1.412029E+O
1.8
stopping potential. It does R A2 = 9.942S24E-1 (Fim order)
affect the time it takes to
reach this potential. From roc) = 4222680E-iS•x + -1 4S8628E+O
RA2 = 9942518E-1 (Second order)
this we can determine that
the intensity of the light
affects the number of elec- 0.0
Irons emitted, but not the 506
maximumenergy of the o First Ordd
electrons. O Second Cr der
02
2. Different colors of light do
affect the maximum energy
of the photoelectrons. The
relationship appears to be
Frequercy
linear.
3. This experiment supports a
quantum model of light.
The slight drop in the measured stopping potential is due to the leakage of charge through
the
zero-gain amplifier. As the intensity decreases, the equilibrium point between the electrons
arriving and the electrons leaving through the amplifier becomes lower.
15
h/o Apparatus
and h'o Apparatus
Accessory Kit
012-04049J
Analysis
2
1.8 f(x) = + -1.412029E+O
First order: R A2 = 9.942524E-1 (First order)
1.6
W = 1.412ev
0.8
Second order:
0.6
h 6.7647E-34 0.4 o first Orda
(2.1% off) O Second der
02
W 1.459ev
it-j
Freq uercy (H z)
16
TechnicalSupport
Feed-Back
Contacting Technical Support
If any comments about this product or this
please let us Before you call the PASCO Technical Support staff, it
If you have any suggestions
ca exp:rzmats or find a problem in the would be helpful to prepare the follomng information:
please tell us PASCO appreciates any cus-
If your problem is '*ith the PASCO apparatus, note:
toner feed-back Your input b:lps us evaluate and
our product - Title and model number (usually listed on the
label);
To Reach PASCO - Approximateage of apparatus;
For Technical Support call at I -SOO-772-S700
- A detailed description of the probleffsequence of
(toll-free v.lthin the US.) or (916) 7S6-3SOO.
events (in case you can't call PASCO right at*ay,
(916) 7863292 you won't lose valuable data);
e-az.ll: - have the apparatus within reach when
78. calling to facilitate description of indi'ådual parts