ASSESSMENT
Anode volt error and Cathode Volt error when using high factor. Possible problem is the Generator 2 - batte
Exposure hand switch is deffective. Collimator bulb is working, Tube - No arcing in HV cables. Table - all locks is work
Tube stand - alligned and all locks is working. Handgrip assembly - all buttons are intact and working. Wall stand - l
Generator: Suspected battery problem. All boards had voltage present. We can test the full function of the machine
replace with new one and to find if the error will clear or cause by other components.
REMARKS
For quotation of exposure hand switch (1) and 2 Battery trays.
Machine is functional and in use
VITA FLEX
CLIENT NAME San Jose District Hospital
ASSESSMENT
REMARKS
DV5700
CLIENT NAME San Jose District Hospital
ADDRESS San Jose, Occidental Mindoro
EQUIPMENT Carestream K-NUMBER 58016
MODEL DV5700 SERIAL NUMBER 57041
ASSESSMENT
REMARKS
No Parts quotation needed.
d 2 Battery trays.
se
DATE 31-Mar-22
DEPARTMENT Radiology
58014812
210817903
DATE 31-Mar-22
DEPARTMENT Radiology
58016521
57041069