SOP for Cleaning of Autoclave


Standard operating procedure for cleaning of double door autoclave.

1.0 OBJECTIVE
To lay down the cleaning procedure of autoclave.

2.0 SCOPE
This procedure is applicable for cleaning of Autoclave

3.0 RESPONSIBILITY
3.1 Doing: Lab Asst./Assistant/Technical Assistant
3.2 Checking: Technical Assistant/Executive

4.0 ACCOUNTABILITY
Head of the Department

5.0 PROCEDURE
5.1 After the completion of cycle unload the autoclave.
5.2 Then shut the door II of the sterile side.
5.3 Open the non-sterile side door.
5.4 Check the chamber temperature. It should be below 50 °C.
5.1 After the completion of cycle unload the autoclave.
5.2 Then shut the door II of the sterile side.
5.3 Open the non-sterile side door.
5.4 Check the chamber temperature. It should be below 50 °C.
5.5 Clean the chamber & trays with tap water.
5.6 Keep the chamber drain valve open
5.6 Use brush for cleaning if required.
5.7 Remove the strainer from drain hole and clean and put it back in place.
5.8 Rinse the entire chamber and trays with distilled water.
5.9 Clean the outer surface with a clean cotton cloth.
5.10 Record the cleaning in Annexure-1.

6.0 ABBREVIATIONS
SOP: Standard Operating Procedure