Professional Documents
Culture Documents
J Wneu 2019 05 148
J Wneu 2019 05 148
J Wneu 2019 05 148
- BACKGROUND: The posterior ventriculoperitoneal shunt - CONCLUSIONS: Using the methods and devices
(VPS) placement procedure is technically efficient and described in this series, posterior VPS operations can be
cosmetically appealing. The main limitations of the posterior performed safely with a high degree of ventricular catheter
approach relate to the technical challenges associated with placement accuracy.
accurately placing the ventricular catheter. In this report, we
describe methods and simple devices used for posterior VPS
surgery that have evolved over a >25-year period to enhance
catheter placement accuracy and reduce complication rates.
INTRODUCTION
- OBJECTIVE: We describe the combination of methods
and customized devices used at a single institution to
perform posterior VPS surgery. Results are presented for
the most recent 11-year epoch, along with a description of
sources of technical errors and plans for further meth-
V entriculoperitoneal shunt (VPS) procedures can be per-
formed using either a frontal or posterior ventricular
catheter approach. In patients with sufficiently large
ventricles, the posterior approach is an attractive option due to
the technical efficiency and cosmetic appeal of making a single
odologic refinements. scalp incision. The primary disadvantage of the posterior
approach is that it can be more challenging to accurately and
- MATERIALS AND METHODS: The medical records and safely place a ventricular catheter into the anterior horn of the
imaging studies of 468 patients undergoing posterior VPS, lateral ventricle using a posterior approach compared with a
from 2007 to 2018 were reviewed. Ventricular catheter frontal approach.1 With the frontal approach, easily recognized
placement accuracy data were collected and complica- visual landmarks can be used to localize the appropriate frontal
tions were identified and recorded. burr-hole entrance site, the appropriate trajectory angle can be
determined using visual landmarks or a simple tripod guide tube,
- RESULTS: Optimal ventricular catheter placement was the distance separating the entrance point and the target is short,
achieved in 98.29%. Of the remaining 1.71%, one half and the planned trajectory is not in close proximity to critical
(0.85%) required acute revision surgery. Four patients neural structures. In contrast, it is more difficult to select a
posterior catheter burr-hole entrance location due to the absence
(0.85%) developed new neurologic deficits following sur-
of nearby visual landmarks. Also, the poster catheter trajectory is
gery; 2 were related to intraparenchymal hemorrhages and longer and passes in close proximity to key periventricular
2 (0.43%) as a result of a misplaced catheter. The deficits structures. For example, a misplaced posterior catheter that
resulting from poor catheter placement were transient. The damages the corticospinal tract can cause significant contralat-
complication rates due to causes other than catheter eral motor deficits.
placement accuracy compared favorably with those re- In the early 1990s, a simple device and surgical method was
ported previously in the literature. developed to address these challenges and improve the accuracy
of the posterior ventricular catheter placements.2,3 The device complications and technical errors observed in the course of
(Caroline Guide [CG]) is a C-shaped instrument with a catheter more than 2 decades of routine clinical use. Data describing
guide tube that is designed in such a way that a catheter passed the catheter placement accuracy and clinical utility of the CG
through the tube will follow a straight-line trajectory between the system for the most recent 468 patients will be presented in
burr-hole entrance site and a frontal scalp target site.2,3 In the the Results section. Patients were included if preoperative
original 1995 report, the clinical utility and accuracy of the CG brain imaging and postoperative computed tomography scan
method was described in the first 38 patients treated at the were available for review, if they had a follow-up of greater
University of Washington and University of Iowa. The CG has than 1 year, and if the CG device was used.
now been used routinely at the University of Iowa for more than
24 years, and this experience has led to further refinements in
the CG method. In this report, we describe the current CG Surgical Devices
method, explain the rationale for methodologic modifications, CG Body. The C-shaped body of the CG is designed to orient a
and provide clinical utility and catheter placement accuracy data catheter guide tube, so that a ventricular catheter passed
for the patients treated using the device in the most recent through the tube will follow a straight trajectory toward the
11-year epoch (2007e2018, 468 patients). anterior (frontal) target point (Figure 1). The dimensions of the
CG body accommodate the cranium of all patients. A blunt-
tipped conical piece delineates the anterior target. The pos-
MATERIALS AND METHODS terior bend of the body is fashioned as a “pistol grip” that
In this section, the CG device and posterior VPS method will enables the surgeon to securely position the device by
be described in detail, along with a description of how the grasping the grip with the left hand. The posterior portion of
device and method has evolved since the original report. the body contains a grooved section that accommodates a
These modifications were carried out to reduce the risks of slotted guide tube. The guide tube is reversibly secured to the
Figure 1. Components of the single-handed soft passer that intersects the frontal target post. The posterior
(SHSP) and Caroline Guide (CG) system. The bend of the body is fashioned as a “pistol grip” that
ventricular catheter stylet cylinder (A) is controlled enables the surgeon to securely position the device by
using a thumb ring. An external cylinder (B) fits over the grasping the grip with the left hand. The posterior
stylet cylinder and is controlled with finger rings. The portion of the body contains a grooved section that
CG (C) orients a slotted guide tube so that a catheter accommodates a slotted guide tube. (D) Fully
passed through the guide tube will follow a trajectory assembled SHSP positioned within the CG guide tube.
CG body by a thumb screw. When secured, the guide tube is Frontal Target Template. The original CG method involved using
oriented such that a line passing through the center of the manual measurement techniques to delineate the anterior
guide tube lumen will intersect the anterior target marker. catheter trajectory target on the patient’s forehead. The
intent is to select a point that is in the midline and 2 cm su-
CG Slotted Guide Tube. The position of the CG guide tube rela- perior to a line connecting the left and right orbital rims. In
tive to the CG body is adjustable, with the optimal position rare instances, the manual technique proved inaccurate. To
being one whereby the anterior opening of the guide tube is address this problem, a triangular-shaped template was
approximately 3 cm away from the burr hole during the designed (Figure 3). The template is secured to the forehead
catheter insertion procedure. Slots were added to the guide when the patient is in the straight prone position. Left and
tube to facilitate visualization of cerebrospinal fluid (CSF) flow right orbital tabs are positioned flush with the superior
from the catheter as soon as the apertures of the catheter tip orbital rims and the template is centered using the bridge
enter the ventricle. This visual cue, along with the tactile of the nose as an orienting visual landmark. A circle in the
feedback associated with a drop in mechanical resistance, template delineates the place on the scalp where a pen is
enables the surgeon to recognize when the tip of the catheter used to mark the anterior target.
has entered the ventricle, at which point the surgeon transi-
tions from a “hard-pass” to a “soft-pass” technique, using Posterior Burr-Hole Localization Devices. The most technically
the single-handed soft passer (SHSP) described to follow. demanding aspect of the CG procedure, and the most common
source of procedural error, relates to selecting the location of the
Single-Handed Soft Pass. The original CG used a rigid stylet to posterior burr hole. Localization errors can occur in the ros-
advance the ventricular catheter. With this design, the cath- trocaudal and mediolateral axis. Manual measurement tech-
eter was constrained to the precise linear path connecting the niques using the palpated external protuberance as an orienting
burr-hole entrance site and anterior target. When an incorrect landmark are associated with high error rates.1,5 To address this
trajectory was selected, the catheter could enter the poste- problem, we first developed a headband-based device to
rior portion of the lateral ventricle but then advance further delineate a burr-hole site 3 cm to the right of the posterior
into periventricular brain tissue. This complication can largely midline and an appropriate distance above the external occipital
be avoided by using a “soft-pass” technique, whereby the protuberance.2 This device proved cumbersome to use in
stylette is withdrawn from the tip of catheter after the cath- practice, particularly in the setting in which significant portions
eter has entered the ventricle.4 When using the CG, the of the patient’s hair is preserved. With both the manual
surgeon uses the left hand to hold the device in the proper measurement and headband localization methods, we
position while advancing the catheter with the right hand observed that the most significant source of error was in the
(Figure 2). The SHSP was designed to enable catheters to mediallateral targeting dimension. To address this problem,
be soft passed using only one hand to control the position we designed and now use routinely a new caliper-like device
of the stylette relative to the catheter, while also advancing (Wishbone) that attaches reversibly to the left and right external
the catheter. As shown in Figure 2, after the catheter has meatus and directs a localizing laser to the cranial midline. A
entered the ventricle, the surgeon pulls back on the SHSP technical report describing this “Wishbone” device is currently
thumb ring to retract the stylette and then continues to under review. The Wishbone was used in the most recent 40
advance the “soft” catheter into the final position. patients in this clinical series.
Figure 2. Single-handed soft passer (SHSP) catheter controlled by means of a thumb ring. An external
insertion method. When using the Caroline Guide, the cylinder, mounted over top of the internal stylette
surgeon uses the left hand to hold the device in the cylinder, is controlled by the surgeon using 2 finger
proper position while advancing the catheter with the rings. After the ventricular catheter enters the ventricle,
right hand. The SHSP was designed to enable catheters the surgeon withdraws the stylette from the tip of the
to be soft passed using only one hand to control the catheter by pulling back on the thumb ring. This
position of the stylette relative to the catheter, while “softens” the tip of the catheter as it is advanced into
also advancing the catheter. Thus, the ventricular the final insertion position by the pushing force of the
catheter is mounted to the stylette extension of the external cylinder controlled by the surgeon’s index and
attached internal cylinder. The position of the stylette is middle fingers.
Figure 3. Schematic drawing depicting the triangular right superior orbital rims. This will serve as the area
shaped frontal-target localizer. The localizer is centered whereas a modified Bovie pad will be placed
on the midline and small lateral tabs (A) are positioned (Lennarson technique, see Figure 7). The Bovie pad will
flush with the patient’s superior orbital rim. The center serve as the frontal target where the anterior part of
of the localizer aperture (B) delineates the midline the CG would rest.
target site, 2 cm above a line connecting the left and
Shunt Scissors. The ventricular catheter placement portion of Patient Selection. Patients must have sufficiently large ventricles
the CG procedure is just one aspect of the VPS procedure, to be candidates for the CG procedure. A cross-section image is
and a range of other complications and technical errors can selected that best demonstrates the lateral ventricles. A trajec-
occur. One such complication can occur when the surgeon tory line is then traced between an anterior and posterior target
creates the subgaleal pocket for the shunt valve. If that point. The anterior point is at the midline and the posterior point is
pocket is too small, the valve-tubing assembly can be kinked 3 cm to the right of the posterior superior sagittal sinus. The
resulting in shunt malfunction. In adult patients, the trajectory line must traverse the ventricles with a clear margin for
galeaperiosteum interface caudal to the posterior burr-hole lateral displacement error (Figure 5). Since the ventricular
location can be particularly dense and difficult to dissect us- anatomy was the main criteria for posterior shunt placement,
ing the stand blunt dissection method. To address this the majority of the patients were adults with normal-pressure
problem, we designed shunt scissors that the surgeon uses hydrocephalus (>85%). Other indications were pediatric hy-
to easily create the subgaleal pocket using a sharp dissection drocephalus, and adults with secondary hydrocephalus, or
technique (Figure 4).6 adults with congenital hydrocephalus who underwent shunt
revision via the posterior approach. Since the study focuses on
Surgical Method technical aspects to avoid intraoperative complications, we
Permission for this study was provided by University of Iowa accepted the diverse etiology of hydrocephalus, as long as the
Hospitals and Clinics institutional review board. ventricular anatomy was suitable for posterior shunting.
2007 to 2018. The procedures were performed by multiple follows the straight line trajectory from a posterior burr hole to an
neurosurgical attendings in patients with a broad range of hy- anterior target site.2 However, the CG alone does not address
drocephalus etiologies (age range: 11e90 years). The age distri- errors in burr-hole localization. Other devices and methods,
bution was the following: 412 (88.03%) patients were older than including most recently a caliper-based system, have been
60 years; 47 (10.04%) patients were between 18 and 60 years; designed and tested to address the burr-hole localization chal-
and 9 (1.92%) patients were younger than 18 years of age. All the lenge. The present series shows that using these devices and
patients had significantly large ventricles with large posterior methods the tip of the ventricular catheter was placed in the
horns. The majority of our patients had normal-pressure hydro- anterior horn of the ipsilateral or contralateral lateral ventricle in
cephalus (>85%). 98.3% of patients. This compares favorably with the reports of
catheter placement accuracy and revision rates using the FHT
Postoperative imaging showed that in 98.29% of cases (n ¼ 460)
and shows similar results to the frameless stereotactic (FS)
the tip of the ventricular catheter was positioned within either the
methods.5,7 Jung and Kim5 compared FS with FHT in posterior
anterior horn of the ipsilateral (n ¼ 393) or contralateral (n ¼ 67)
VPS. They reported a 28% rate of suboptimal positioning using
lateral ventricle. Thus, optimal position (ipsilateral) was achieved
the FHT compared with 7% using FS. In the Jung and Kim
in 85.43% and suboptimal (contralateral) in 14.31% of the cases.
series of 72 patients, the rate of revision from proximal
In 8 cases (1.71%), the catheter tip was positioned in either
obstruction was 31.11% and 11% for FHT and FS,
periventricular white matter, or the atrium or inferior horn of the
respectively, where 4% of the FHT cases had to be revised
lateral ventricle. Four of these patients (0.85%) underwent im-
due to catheter malposition. In our series, 0.9% had to be
mediate revision of the poorly placed ventricular catheter. In this
revised due to catheter malposition, and the cumulative rate of
series, 4 patients (0.85%) sustained new neurologic deficits as a
proximal obstruction during the period of follow-up was 9.0%.
result of the shunt placement procedure due to intraparenchymal
Our overall incidence of shunt revision was 21.7%, which is
hemorrhage (n ¼ 2) or errant placement of the ventricular cath-
similar to rates reported in the literature (15%38.5%).8-11
eter (n ¼ 2). In the 2 cases in which a misplaced ventricular
Similarly, our postoperative infection rate compares favorably
catheter caused neurologically injury, patients recovered fully
with those reported in the literature (2.5% vs. 2%10%).9,11,12
within 6 months of surgery based on the results of neurologic
exams performed in the outpatient clinic. A recent meta-analysis of 11 studies (adult and pediatric patients)
showed an overall shunt functional survival rate of 55.8% to
Other delayed complications included shunt system disconnec-
84.1% for the FHT and from 67.2% to 97.1% for shunts placed
tion at the level of the valve (1.07%), shunt infection (2.56%),
using image guidance.13 The rate of accurate catheter placement
proximal failure (8.97%), distal obstruction (4.91%), and back-out
using image guidance was generally greater in case series
of the peritoneal catheter (1.50%). The proportion of patients
(range, 95.2%100%) than in cohort studies (range, 55.3%
who subsequently required shunt ligation due to acute or chronic
100%). The authors concluded that use of intraoperative
refractory symptomatic subdural hematoma was 1.71%. A pro-
navigation will reduce the risk of proximal malfunction due to
grammable valve was used in all cases. Our overall incidence of
catheter malposition but may increase the infection rate.13 This
shunt revision during the entire follow-up period was 21.34%.
meta-analysis, however, did not describe or delineate whether
With minimal training and experience, the CG method for ven-
posterior or frontal surgical approaches were used to place the
tricular catheter placement becomes a simple, efficient and safe
ventricular catheter.
surgical technique. Compared with a free-hand technique (FHT),
use of the CG method adds less than 5 minutes to the total Compared with the FS technique, the CG device and method are
operative time for a posterior VPS operation. The posterior burr- less expense and more time-efficient. FS systems are more
hole localization procedure took less than 3 minutes with the complicated to use because they require skillful implementation
Wishbone device. of image registration and related tasks.14 However, even though
the CG method is less complex, the surgeon must still pay
DISCUSSION meticulous attention to every technical detail of the procedure
to achieve optimal results. Common sources of technical error
The posterior VPS operation is an appealing shunt surgery option are briefly discussed to follow.
because of its technical efficiency and minimal cosmetic impact
on the patient. The primary limitation of the procedure, and the The most common and clinically significant technical error that
consideration that causes many surgeons to select the frontal occurs is linked to poor localization of the posterior burr-hole site
VPS method instead, relates to the technical challenges of due to an inaccurate localization of the posterior midline. When
accurately and safely placing a posterior ventricular catheter. In the burr hole is placed too medial, the catheter trajectory may
this report, we describe a system of devices and methods that traverse sulci of medial hemisphere cortex, rupture sulcal ves-
have been used in an evolving fashion to improve the accuracy of sels, and cause an intraparenchymal hemorrhage (2 patients in
posterior catheter placement over the last 25 years. The occipital this series, 0.43%; Supplementary Figure 1). Medial burr-hole
approach is known to have the smallest margin of errors due to placement can also increase the probability of contralateral can-
the small range of possible angles for successful ventricular nulation of the ventricle, particularly if the SHSP device is not
cannulation, as Lind et al.1 demonstrated. This, combined with deployed properly. A laterally placed burr hole can lead to errant
the lack of a clear anatomical landmark for localizing the placement of the ventricular catheter into periventricular tissue
posterior burr-hole site, makes this a more technically chal- and potentially damage the corticospinal tract (Supplementary
lenging VPS procedure and stresses the need for instrumentation Figure 2). Since the Wishbone device was invented to localize
that assists with burr-hole localization and catheter trajectory the midline, we only had 2 posterior burr holes that were not
determination. The CG method was developed to address this accurately placed but did not result in any malposition of the
challenge and is effective at insuring that a ventricular catheter ventricular catheter requiring revision. We only used the
frameless stereotaxy techniques vs. frame-based 14. Zrinzo L. Pitfalls in precision stereotactic surgery.
REFERENCES localization. J Neurooncol. 2006;76:65-70. Surg Neurol Int. 2012;3:S53-S61.
1. Lind CRP, Tsai AMC, Law AJJ, Lau H,
Muthiah K. Ventricular catheter trajectories from 8. Bir SC, Sapkota S, Maiti TK, Konar S, Bollam P,
traditional shunt approaches: a morphometric Nanda A. Evaluation of ventriculoperitoneal Conflict of interest statement: Dr. Howard and Dr. Zanaty
study in adults with hydrocephalus. J Neurosurg. shunt-related complications in intracranial me- are listed as inventors on a provisional patent application
2008;108:930-933. ningioma with hydrocephalus. J Neurol Surg B Skull filed at the University of Iowa for the “Wishbone” device
Base. 2017;78:30-36. used in this manuscript. The “Wishbone” is the only
2. Garell PC, Mirsky R, Noh MD, et al. Posterior component of the system to which the patent applies. All
ventricular catheter burr-hole localizer. Technical 9. Khan F, Rehman A, Shamim MS, Bari ME. Fac- the other patents have expired. All of the devices
note. J Neurosurg. 1998;89:157-160. tors affecting ventriculoperitoneal shunt survival described in this report were fabricated in the University
in adult patients. Surg Neurol Int. 2015;6:25. of Iowa Medical Instruments Shop. Surgeons at other
3. Howard MA 3rd, Srinivasan J, Bevering CG, institutions have fabricated Caroline Guide devices at their
Winn HR, Grady MS. A guide to placement of 10. Oka K, Yamamoto M, Ikeda K, Tomonaga M. own hospitals or used devices that were manufactured at
parietooccipital ventricular catheters. Technical Flexible endoneurosurgical therapy for aqueductal the University of Iowa and shipped to their institutions in
note. J Neurosurg. 1995;82:300-304. stenosis. Neurosurgery. 1993;33:236-242 [discussion support of collaborative device development research
242-243]. projects. None of the devices are protected by issued U.S.
4. Garell PC, Noh D, Loftus CM, Hitchon PW,
patents. A provisional patent has been filed by the
Howard MAI. Single-handed variable stiffness
11. Park M-K, Kim M, Park K-S, Park S-H, Hwang J-H, University of Iowa for the Wishbone midline localizer.
introducer for catheters. J Clin Eng. 1998;23:
Hwang SK. A retrospective analysis of ven- Work is underway to enable hospitals to purchase
29.
triculoperitoneal shunt revision cases of a single Caroline Guide systems through a non-profit, university-
institute. J Korean Neurosurg Soc. 2015;57:359-363. based medical device company (MADE) that is part of the
5. Jung N, Kim D. Effect of electromagnetic navi-
gated ventriculoperitoneal shunt placement on University of Iowa.
failure rates. J Korean Neurosurg Soc. 2013;53: 12. Junaid M, Ahmed M, Rashid MU. An experi-
Received 16 April 2019; accepted 16 May 2019
150-154. ence with ventriculoperitoneal shunting at
Keen’s point for hydrocephalus. Pak J Med Sci. Citation: World Neurosurg. (2019).
6. Dacey RG, Flouty OE, Grady MS, Howard MA, 2018;34:691-695. https://doi.org/10.1016/j.wneu.2019.05.148
Mayberg MR. Shunt scissors: technical note. Journal homepage: www.journals.elsevier.com/world-
J Neurosurg. 2018;129:1200-1202. 13. Nesvick CL, Khan NR, Mehta GU, Klimo P Jr. neurosurgery
Image guidance in ventricular cerebrospinal fluid
Available online: www.sciencedirect.com
7. Quiñones-Hinojosa A, Ware ML, Sanai N, shunt catheter placement: a systematic review and
McDermott MW. Assessment of image guided meta-analysis. Neurosurgery. 2015;77:321-331 [dis- 1878-8750/$ - see front matter ª 2019 Published by
accuracy in a skull model: comparison of cussion: 331]. Elsevier Inc.
Supplementary Figure 1. Computed tomography (CT) sagittal sinus, instead of the planned 3 cm. The too-medial
images from a patient who sustained a complication trajectory resulted in the catheter traversing medial
related to placing the posterior burr hole too medially. The hemisphere sulci and causing an intracerebral hemorrhage
CT scan on the left shows the center of the burr hole is resulting in visual field loss.
w16 mm lateral to the center of the posterior superior
Supplementary Figure 2. Computed tomography images from a patient the right of the midline. The catheter traverses brain tissue lateral to the
who sustained a transient neurologic deficit due to poor catheter lateral ventricle in close proximity to the corticospinal tract. Proximal
placement resulting from placement of the posterior burr hole too shunt revision surgery was required.
laterally. In this case the center of the burr hole was placed 4.5 cm to
Supplementary Figure 3. Computed tomography images from a patient error, and the failure of the surgeon to properly deploy the single-
with a burr-hole localization error in the rostrocaudal dimension. In this handed soft passer when the catheter entered the ventricle, the tip of
case the burr hole was positioned too low (caudal); 4 cm above the the ventricular catheter was positioned in periventricular brain tissue
external occipital protuberance instead of 7. Due to this localization adjacent to the third ventricle.
Supplementary Figure 4. An intraoperative figure of the the center of the guide tube lumen will intersect the
surgeon using the Caroline Guide. The front part is on the anterior target marker. The surgeon is using the single-
frontal template and the posterior part is at the burr hole. handed soft passer technique.
The guide tube is oriented such that a line passing through