Professional Documents
Culture Documents
GH900 HbA1c Analyzer Service Menual
GH900 HbA1c Analyzer Service Menual
Service Manual
IVD
Shenzhen Lifotronic Technology Co., Ltd. (hereinafter “Lifotronic”) has the copyright
of this manual not publicly published the right to treat it as confidential materials. This
manual is merely a reference for the operation, maintenance and service of Lifotronic’s
products. This manual and all intellectual properties (including copyright) pertaining to
it shall reside with Lifotronic. Without the prior written permission of Lifotronic, no
person shall use, disclose or allow others to obtain this manual by any means in whole or
in part, and no one shall photograph, copy, duplicate or translate (without limitation to
the foregoing) this manual in whole or in part.
Lifotronic reserves the right to update product technology without prior notice;
Lifotronic reserves the right to modify product specifications without prior notice.
Lifotronic reserves the right to modify the manual without prior notice.
Statement
Lifotronic makes no guarantee for this document whatsoever, including (but not limited
to) implied merchantability and suitability for a particular purpose.
Lifotronic shall be responsible for the safety, reliability and performance of its product
only when:
The assembling, expansion, readjustment, improvement and repair are carried out
by Lifotronic authorized personnel;
The machine is used according to operation instructions;
Electrical devices are in compliance with national standards.
Lifotronic shall not be responsible for the safety, reliability and performance of its
product if:
I
The product has reached its lifetime limit;
Parts have been disassembled, stretched or readjusted;
The machine is not used properly according to User Manual.
Warranty
Repair Service
(1) Lifotronic will provide fee-based service if the product is not within the range of
Lifotronic’s warranty provisions;
(2) Within warranty period, fee-based service will be provided if the product defect is
caused by:
Misuse;
Man-made damage
Replacement with parts not approved by Lifotronic;
Repair of the machine by personnel not authorized by Lifotronic;
Grid voltage beyond equipment specifications;
Uncontrollable natural disasters.
Lifotronic shall not be responsible for direct, indirect or consequential damage or delay
caused by the foregoing (including but not limited).
II
Return of Product
(1) Obtain Return Material Authorization. Contact Lifotronic’s Customer Service Dept.
and provide the S/N of Lifotronic product. The S/N is marked on the package. Return of
product will not be accepted if the S/N is not clearly legible. Please note the type, S/N of
product and reason of return.
(2) Freight expense: Freight (including customs fees) of products returned to Lifotronic
for service shall be borne by the customer.
Contact
III
● The system is operated and used by Lifotronic or designated agents trained
professionals, doctor or lab assistant;
● If hospitals or institutions can not achieve a satisfactory repair / maintenance plan, it
may cause abnormal instrument failure, and may endanger human health.
● Ensure to run the instrument under the conditions that are written in this manual. If
not, it may fail to run, result in inaccurate results, or even damage the components
and injure humans.
IV
Content
Chapter 2 Hardware.................................................................................................................................. 9
4
3.5 Test Chromatogram of Sample with Over Dilution ............................................................ 25
5
Chapter-1 Instrument Introduction
GH-900 HbA1c Analyzer is consisted of the parts listed below, see in Picture 1
1) Liquid Line: Liquid line switch part, sample aspirating part, rinsing part,
disposing part, reagent tube
2) Hardware: Control panel, analog board, power module, display module, print
module, wire harness
3) Software: System software, data processing algorithm, liquid line sequential
algorithm, printer driver, clock & memory, HCI, temperature control
algorithm
4) Mechanical Part: Industrial design, shell, holder, sample arm, injection
transmission components, waste transmission components, cuvette, column,
mixing cup
6
Picture 1 Overall Framework
1.2 Methodology
HbA1c Analyzer adopts Lower Pressure Liquid Chromatography (LPLC)
methodology, to separate HbA1c based on the differences of electric charge quantity
between different hemoglobin. HbA1c will be eluted first at certain low concentration
eluting condition. And then measure the absorbance points to form chromatogram, get
the precise HbA1c testing result.
7
Picture 2 HbA1c Test Curve
● GH-900 Analyzer is a diagnosis device that is used for disease screening. When diagnosing disease,
clinicians should take clinical examination results or other test result into consideration.
GH-900 Analyzer adopts Low Pressure Liquid Chromatography (LPLC) to measuring the
content of HbA1c in human blood.
The GH-900 Analyzer could display as much as 5 types of blood parameters and 1 type of
chromatogram result.
8
Chapter 2 Hardware
2.1 Mainboard
The hardware of HbA1c Analyzer shows in Picture 2 and includes 5 main parts:
Power module, control module, temperature control module, cuvette testing module,
input/output module.
1) Power module provides different voltage for the system to keep it working
normally;
2) Adopting MCU single chip, control module is the core part of the whole
system, which conducts data collection, chromatogram display and liquid line
control etc.
3) Temperature control module is used to provide an appropriate environment for
mixing cup and column. That will make system more stable and results more
precise.
4) Cuvette testing module is intended to test the contents of eluted components.
5) Input/output module is the main route of human-computer communication that
has functions like QC, calibration, sample testing, system setting, maintenance
etc.
9
24V Control
Sample_Tray Panel Power
X arm Optocoupler
Step Motor-X
Z arm Optocoupler
Step Motor-Z
Inject Pump
Peristaltic Pump In
SV1
SV2
10
Peltier
Heater
Fan
Printer
SV 3-9
Data Collect
LED
Scan Com PC LCD Display RFID Air Temp PET Temp Heater Temp
11
J24 +24V,GND 24VControl Panel Power Port
12
J4 SV5 Solenoid Valve 5 Port
13
2.2 Sample Loader
Follow the user manual instruction to dissemble the sample loader as seen in Picture
Sample Loader
Connection Cable
Base Board
Optocoupler Test
Sample Tray
Driving Disk
Groove
Protruding Point
1) There are two protruding points stuck with two grooves on the driving disk, to rotate
sample tray.
2) Sample Loader Positioning: There are is gap on the back of sample tray working as
sample tray optocoupler test point, it will generate a signal once the optocoupler
meets the gap. The system uses the test point as the default position and orientates
the sample loader position by this process.
3) Please check sampling needle and sample tray position every time you installed the
sample loader (whether it’s an installation or a service). if there is deviation, please
go the service menu to adjust positioning.
14
2.3 Thermal Printer
Printer Screws
2.4 2D Arm
15
2D Arm is to move the sample probe vertically and horizontally.
Sample Probe
Motion Slider
X-axis
Y-axis
Sample Probe
Driving Belt
Sample Probe
Wiper
16
Nylon Screw to Fix
Colorimetric Cell
Filter Lens
Optical Sensor
Colorimetric Cell
Disassemble Assemble
Long Tube to
Connect Column
Short Tube to
Screw Position
Connect Waste Line
When LED value rise 1, the light is getting stronger, ADC value will rise (700-800).
Thus, the optical sensor aging will lead to ADC value shrink.
ADC value within 17,000-19,000 means the colorimetric cell is filled with bubble-free eluent and
has no significant changes.
If ADC value is out of 17,000-19,000, or vary greatly to reach 20,000 or 0, it means the
instrument has faults to be fix. (Please refer to the Fault Information Section)
17
2.6 Column Warming Module
This module is to stabilize the column temperate at 25 ℃, please see below picture.
Cooling Fan
Heatsink
Column
Peltier Heater
Temperature
Sensor
Warming Cover
Peltier heater, heatsink and cooling fan compose the warming module, and temperature sensor will
monitor real time temperature through the warming cover.
Temperature too high or too low will affect the ion change and then test results.
There are 8 3-way solenoid valves in the analyzer. Each solenoid valve has two exits and one
entry, but only one exit can work with the entry to be a route at one time. For example, the entry
and the exit marked with ON work as a route once there is no electrical current. The entry and the
exit marked with OFF work as a route once there is electrical current.
It’s easy to test the solenoid valve at service menu, input the service password and then choose
“Service Test” menu to check the solenoid valve one by one. If there is “Di” sound, it means the
solenoid valve connection is good. If there is no sound, it means the solenoid valve is broken or
connection wrong.
18
ON Exit, will close once
there is electrical current
OFF
Exit
Entry
Plunger pumb is to precisely aspirate sample and hemolysin. And it’s connected to No.2 Solenoid
Valve. It’s necessary to check whether the aspirate volume is correct.
19
2.9 Mixing Cup Warming Module
Once the analyzer is powered on, the film heater will start, the heat will transfer through the
aluminum block to the mixing cup. (Temperature sensor will monitor the real time temperature
and keep it at 51℃. Blood sample and hemolysin should react at this temperature to destroy the
Erythrocyte membrane completely to release all the hemoglobin.
Mixing Cup
Aluminum Block
Film Heater
Temperature Sensor
Chromatography Column: Set the duration of mixed liquid pass through the chromatography
column, default is 2000ms.
Speed: Set “High” and “Low” speed to generate a speed difference for
certain pressure between 2 pumps.(Difference should be 60, e.g
when High speed is 50, Low speed should be 110)
2D Arm
20
Z Axis Routes: To adjust Sample Probe vertically to reach lowest 2mm to the
sample tube bottom for aspiration, distance is in mm
Sample Loader: Adjust the sample loader position to suit sample probe exactly in the
middle.
It’s very useful to diagnose the parts one by one to detect the system issues.
Solenoid Valve: There are 9 solenoid valves, press “on” and “off” to test them by sound.
Inlet Pump: Press “FWD” and “REV” to check whether it’s working properly.
Outlet Pump: Press “FWD” and “REV” to check whether it’s working properly.
Plunger Pump: Press “In” and “Out” to check whether it’s working properly.
X Axis Motor: Press “In” and “Out” to check whether it’s working properly.
Z Axis Motor: Press “Up” and “Down” to check whether it’s working properly.
Bar Code:
21
HIS Test, RFID Test, Print Test, Sampler Loader Positioning
22
Chapter 3 Abnormal Test Chromatograms
23
3.3 Test Chromatogram with Bubbles
24
3.5 Test Chromatogram of Sample with Over Dilution
25
3.7 Test Chromatogram with ADC Value over 20000
26
3.9 Test Chromatogram of Eluent B Polluted by Eluent A
27
3.11 Test Chromatogram of Eluent C Polluted by Eluent A or
28
Chapter 4 System Liquid Line Chart
Hemolysin Liquid Route: Diaphragm pump generate power to aspirate hemolysin from
29
SV2SV1Sample Mixing Cup.
Sample Probe Wiper Liquid Route: Diaphragm pump generate power to wash and wipe
sample probe with hemolysin before and after aspirate sample.
30
Sample mixed with hemolysin and heated to 51Ԩ, inlet pump and outlet pump will aspirate
proper sample volume to pass through column.
Once sample pass through column, glycated hemoglobin will be adsorbed to resin in the column
for next washing step.
31
Eluent A is adsorbed into column by 2 peristaltic pumps, and to elute HbA1a and HbA1b which
have the weakest absorption to resin. The change of absorbance will form the first crest of
diagram through colorimetric cell.
32
Eluent B is adsorbed into column by 2 peristaltic pumps, and to elute HbA1c which has the 2nd
level absorption to resin. The change of absorbance will form the 2nd crest of diagram through
colorimetric cell.
33
Eluent C is adsorbed into column by 2 peristaltic pumps, and to elute HbA10 which have the
strongest absorption to resin. The change of absorbance will form the 3rd crest of diagram through
colorimetric cell (It’s too high to display on the screen completely).
34
There will be liquid line wash every time when the operator power on/off the instrument.
And there will be liquid line wash before/after the sample test.
Procedure:
3. At last absorb eluent A to wash the column and balance the resin.
35
Eluent A is absorbed into column by 2 peristaltic pump’s reverse motion to counter-wash the
liquid line. It’s main for clean the colorimetric cell, pretreatment and chromatography column.
36
Eluent A is absorbed into column by 2 peristaltic pump’s reverse motion to wash the mixing
cup, and then diaphragm pump will absorb the waste liquid into the waste liquid bottle.
37
Chapter 5 Fault Table
The operator can use below Faults Table to diagnose possible reasons for specific fault
and take proper action to solve the issue.
38
1. Bad connection of motor 1. Reconnect the signal cable.
Peristaltic pump signal cable. 2. Change the motor.
doesn’t move 2. Motor broken. 3. Change the mainboard.
3. Mainboard broken.
1. Defective card. 1. Change new RFID card.
Can’t swipe card 2. RFID cable bad 2. Reconnect RFID cable.
or reading error connection. 3. Change RFID card reader.
3. RFID card reader broken. 4. Change mainboard.
4. Mainboard broken.
1. There are bubbles, bulk 1. Shake the blood sample
or flocculent tissue in tube several times and then
sample. test again.
Abnormal test
2. Software problem. 2. Contact your distributor for
chromatogram and 3. Pretreatment or software update.
result chromatography column 3. Change the pretreatment or
defective or blocked. chromatography column.
4. Inlet pump blocked. 4. Change inlet pump.
1. 2D arm’s X Axis screw 1. Tighten the screw.
Sample probe loosen. 2. Adjust X & Y Axis.
arrive wrong 2. 2D arm’s X Axis motion 3. Change optical sensor.
position for slider move with friction
aspiration, or 2D and noise.
arm error message 3. Optical sensor broken.
As the service engineer, you need to be familiar with the principle, structure, operation
procedure, liquid line and action sequence etc. Thus you could help you to find failure point
precisely and solve it quickly.
39