User Manual EnbioServiceTool v1 - 0

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 12

SERVICE TOOL

INSTALATION AND USER MANUAL

Desktop application ver.: 1.0


Manual ver.: 1.0
Create date: 27.11.2022
by Enbio Technology. Sp. Z o.o.
Requirments:
• operating system: Windows 7, 8,10 & 11
• connection interface: port USB 2.0
• display settings for best appeareance (control panel ->
display settings):
• USB cable 2.0 type A 2x male
INSTALLATION
1. Download EnbioServiceTool.zip
2. Unzip to your computer
3. Install driver coresponding to your operating system:
Operating system File
Windows 7 32 bits VCP_V1.5.0_Setup_W7_x86_32bits
Windows 7 64 bits VCP_V1.5.0_Setup_W7_x64_64bits
Windows 8/10/11 32 bits VCP_V1.5.0_Setup_W8_x86_32bits
Windows 8/10/11 64 bits VCP_V1.5.0_Setup_W8_x64_64bits

4. Double click on Enbio Service Tool Installer.exe


5. After installation you can find shortcut on your
desktop.
6. Application is ready to use.
CONNECTION PROCEDURE
1. Turn off device (Enbio S / Pro
autoclave).
2. Connect device to your computer by
usb cable.
3. Power on device.
4. Click Find Device.

HINT 1
To check if driver is proper installed and
device is recognized by your computer
open Control Panel -> Device Manager.

HINT 2
Some operating system cannot recognized
device if bluetooth is turned on. Switch it
off and try again.
Counters tab - READING INFORMATION

Counters: Serial Number


• Total process Date & time on device
• HEPA DIP Switch state
• Service Firmware version
• First process

Click Refresh to read


current values
Counters tab – CHANGING VALUES

Clear log history


- Clear error history
(visible in service Change serial number: write
screen) value to save and click Change

Reset Service Date & time: click Update to


- Reseting days and change time to time from your
cycles to service computer
Counters:
Clear all Write Change value
- Writing 0 to Total to save in device
process / HEPA / then click Change.
Service

Change All: save Change value (Total process / HEPA / Service) to device
Counters tab – Replacement device

The limit is automatically reset


when EnbioServiceTool connects
to the device.

To change the cycle limit for the


Replacement Device enter the
value and click Write.

Option only avaiable if connected


device is recognized as
Replacement Device.
Control tab – control output & read sensors

Click ON to open valve


Click OFF to close valve

Locking door – DO NOT OPEN


Control heaters and DURING STERILIZATION
pumps – ON / OFF PROCESS

Read live data from


temperatures and
pressure sensors
Log tab – read log and decode errors

Choose file: select log Save to file: save


to read decoded log to file

Show basic
information from Open chart:
log: visualize values from
- Serial number log on chart. Select
- Firmware ver. data to be printed
- Process nb. (pressure, process /
- Process type chamber / steam
- Error number generator temp.) –
see next slide

Decode errors: convert code to list of errors


Zoom or save chart to image file
Other tab – send log to Enbio service
By application you can directly send uploaded log
to Enbio Service Department

Fill:
• Service name or service
technician first & last name
• Service address
• Phone number
• Email contact address
• Short description of problem or
failure

Click Send.
In case of problems with the application,
please report them to: serwis@enbio.com