Standard Operating Procedure (SOP) for Operation and Calibration of Melting Point
Apparatus used for analysis / identification of starting materials (Raw Material – API &
Excipient) in Quality Control Laboratory.
Table of Contents
1. Operation and Calibration of Melting Point Apparatus
1.1. 1.0 PURPOSE:
1.2. 2.0 SCOPE:
1.3. 3.0 REFERENCES – SOP FOR MELTING POINT APPARATUS:
1.4. 4.0 RESPONSIBILITY – SOP FOR MELTING POINT APPARATUS:
1.5. 5.0 PROCEDURE – MELTING POINT APPARATUS:
1.5.1. NOTE: Be sure about the settings (initial temperature and °C/min) as they cannot be
viewed or changed while heating process is ON.
1.6. 6.0 ANNEXURES – MELTING POINT APPARATUS:
1.6.1. Annexure -1 : Format for Calibration Record of Melting Point Apparatus
1.6.2. Annexure -2 : Format for Calibration Status label
https://www.pharmaguideline.net/sop-on-operation-and-calibration-of-melting-point-apparatus/ 1/8
م1:47 2023/7/30 SOP on Operation and Calibration of Melting Point Apparatus - Pharmaceutical Guidelines
The purpose of this SOP is to describe the procedure for operation, calibration
and maintenance of Melting Point Apparatus.
2.0 SCOPE:
SOP for Preparation of internal and external (Third Party) Calibration schedule
and calibration practices
To provide training to all the concerned persons before implementing the SOP.
To ensure the operation and calibration of the instrument is carried out as per
the SOP.
Follow the SOP on Instrument/Equipment usage log book, for the entry of
usage of the instrument.
Maintain the third party calibration schedule and the internal calibration
schedule for the instrument as per SOP, Preparation of internal and external
(Third Party) calibration schedule and calibration practices.
Operational procedure
Switch ON power supply by switching on the main supply and switch provided
on backside of the instrument.
Press the “FUNCTION” Key to set initial temperature. The “INI TEMP” LED blinks
and parameter is displayed in the display.
Set the initial temperature to 10°C below the expected melting point and the
initial temperature should be (15 – 25 )°C above the bath temperature to
overcome undershoot or overshoot of temperature, the maximum melting
range for the operation of the instrument is 350°C.
https://www.pharmaguideline.net/sop-on-operation-and-calibration-of-melting-point-apparatus/ 3/8
م1:47 2023/7/30 SOP on Operation and Calibration of Melting Point Apparatus - Pharmaceutical Guidelines
Press the “FUNCTION” key to set “°C/min”, “°C/min” LED blinks and the
parameter is displayed in the display.
Press “START” key. The “INI TEMP” and “°C/min” LED blinks one after another
displaying the corresponding setting with an interval of 2.5 sec.
NOTE: Be sure about the settings (initial temperature and °C/min) as they cannot be
viewed or changed while heating process is ON.
Press “START” key. The oil bath temperature is raised to the initial During this
period “INI TEMP” LED glows continuously and “HEATER” LED glows
continuously or blinks.
Press “FUNCTION” key to start Rate of heating (i.e. °C/min). During this period
“°C/min” LED glows continuously and “HEATER” LED blinks.
Press “STORE” key at the beginning of Melting point, “T1” LED glows
continuously indicating that initial melting temperature is stored.
Press “STORE” key again to store final melting point. “T2” LED glows
continuously indicating that final melting point temperature is stored.
To view initial and final melting point temperatures press “STORE” key to view
initial melting point temperature, “T1” blinks to indicate that the displayed
temperature is initial melting point.
Press “STORE” key again to view final melting point temperature, “T2” LED blinks
to indicate that the displayed temperature is final melting point
Calibration Procedure of Melting Point Apparatus
https://www.pharmaguideline.net/sop-on-operation-and-calibration-of-melting-point-apparatus/ 4/8
م1:47 2023/7/30 SOP on Operation and Calibration of Melting Point Apparatus - Pharmaceutical Guidelines
Charge the capillary, one end of which is sealed with sufficient quantity of
reference standard to form a column in the bottom of the tube 2.5 mm to 3.5
mm height when packed down as closely as possible by moderate tapping on
a solid surface. Surface. (There should not be gap in the column).
Heat the block until the temperature is about 10°C below the expected melting
point and continue heating at a rate of temperature increase of about 1° C per
minute.
Insert the capillary tube in the melting point apparatus when the temperature is
about 5° C below the lower limit of the expected melting range and continue
heating until melting is complete.
https://www.pharmaguideline.net/sop-on-operation-and-calibration-of-melting-point-apparatus/ 5/8
م1:47 2023/7/30 SOP on Operation and Calibration of Melting Point Apparatus - Pharmaceutical Guidelines
Instrument
Make/Model
Code
Calibration Monthly ± 3
Location
frequency days
Procedure
Charge the capillary, one end of which is sealed with sufficient quantity of
reference standard to form a column in the bottom of the tube 2.5 mm to 3.5
mm height when packed down as closely as possible by moderate tapping on a
solid surface. (There should not be gap in the column.)
Heat the block until the temperature is about 10° C below the expected
melting point and continue heating at a rate of temperature increase of about
1° C per minute.
Insert the capillary tube in the melting point apparatus when the temperature is
about 5°C below the lower limit of the expected melting range and continue
heating until melting is complete.
Record the temperature at which the last particle passes into the liquid phase.
Remove the capillary tubes care fully and dispose it off after cooling.
https://www.pharmaguideline.net/sop-on-operation-and-calibration-of-melting-point-apparatus/ 6/8
م1:47 2023/7/30 SOP on Operation and Calibration of Melting Point Apparatus - Pharmaceutical Guidelines
Observation Table:
81°C to
1 Vanillin
83°C
112°C to
2 Acetanilide
115°C
163°C
3 Sulfanilamide
to166°C
235°C to
4 caffeine
239°C
Remarks : The instrument is calibrated & qualified / Out of calibration & not
qualified for use.
CALIBRATION STATUS