Professional Documents
Culture Documents
Transcutaneous Energy Transfer System For Powering
Transcutaneous Energy Transfer System For Powering
net/publication/226948483
CITATIONS READS
7 268
5 authors, including:
Simon Malpas
University of Auckland
183 PUBLICATIONS 4,786 CITATIONS
SEE PROFILE
Some of the authors of this publication are also working on these related projects:
Call for Speical Issue on wireless power transfer and energy harvesting View project
All content following this page was uploaded by D. Budgett on 02 June 2014.
Abstract — Time varying magnetic fields can be used to of variation in coupling is due to posture changes of the
transfer power across the skin to drive implantable biomedical patient causing variation in the alignment between the pri-
devices without the use of percutaneous wires. However the mary and the secondary coils. The typical separations be-
coupling between the external and internal coils will vary tween the internal and external coils are in the range of 10-
according to orientation and posture. Other potential sources
of power delivery variations arise from changes in circuit
20mm. If insufficient power is delivered to the load then the
parameters and loading conditions. To maintain correct device implanted device will not operate properly. If excessive
function, the power delivered must be regulated to deal with power is delivered, then it must be dissipated as heat with
these variations. This paper presents a TET system with a the potential for causing tissue damage. Therefore it is im-
closed loop frequency based power regulation method to de- portant to deliver the right amount of power matching the
liver the right amount of power to the load under variable load demand.
coupling conditions. The system is capable of regulating power
for axially aligned separations of up to 10mm and lateral dis- Skin
Primary Coil
placements of up to 20mm when delivering 10W of power. The
Magnetic Secondary Coil
TET system was implanted in a sheep and the temperature of coupling
implanted components is less than 38.4 degrees over a 24 hour Power
period. DC Pickup Load
Converter
Supply
Keywords — Magnetic field, coupling, Transcutaneous En-
ergy Transfer (TET) Power
feedback
Controller
Chwee Teck Lim, James C.H. Goh (Eds.): ICBME 2008, Proceedings 23, pp. 235–239, 2009
www.springerlink.com
236 T. Dissanayake, D. Budgett, A.P. Hu, S. Malpas and L. Bennet
power delivered to the load. Depending on the actual power PWM signal is passed through a Digital to Analogue Con-
requirement of the pickup load, the operating frequency of verter (DAC), in order to obtain a variable reference volt-
the primary power converter is varied so the secondary age. This variable reference voltage is then used to vary the
prick-up is either tuned/detuned, thus the effective power frequency of the primary resonant converter which in turn
delivered to the implantable load is regulated [4]. The sys- varies the power delivered to the implantable system. The
tem discussed in this paper uses frequency control to control response time of the system is approximately 360ms.
power delivery to the load, and a Radio Frequency (RF) link
is used to provide wireless feedback from the implanted A. Frequency controller
circuit to the external frequency controller.
The frequency controller employs a switched capacitor
control method described in [7]. The controller varies the
II. SYSTEM ARCITECHTURE overall resonant frequency of the primary resonant tank in
order to tune/detune to the secondary resonant frequency.
The TET system is designed to deliver power in the The frequency of the primary circuit is adjusted by varying
range of 5W to 25W. Figure 2 illustrates the architecture of the effective capacitance of the primary resonant tank. This
the overall system. A DC voltage is supplied to the system is illustrated in figure 3.
with an external battery pack. A current fed push pull reso-
nant converter is used to generate a high frequency sinusoi-
dal current across the primary coil. The magnetic coupling Secondary resonant
L1 L2 Primary resonant tank tank
between the primary and the secondary systems produces a
sinusoidal voltage in the secondary coil which is rectified CS
VIN CP LP LS Load
by the power conditioning circuit in the pickup to provide a
CV1 CV2
stable DC output to the implanted load. As shown in figure
2, a DC inductor is added to the secondary pick up follow- SV1 S1 S2 SV2
ing the rectifier bridge in order to maximize the power
transfer to the load. The DC inductor aids to sustain a con-
tinuous current flow in the pick up [5].
Fig. 3 System based of primary frequency control [4]
_________________________________________ ___________________________________________
IFMBE Proceedings Vol. 23
Transcutaneous Energy Transfer System for Powering Implantable Biomedical Devices 237
III. EXPERIMENTAL METHOD also put on the site of the wound to reduce infection. Fol-
lowing the surgery the sheep transferred to a crate where it
A prototype TET system was built and tested in a sheep. was kept over a three week period. The primary coil was
The internal coil and the resonant capacitor were Parylene placed directly above the secondary coil and held on the
coated and encapsulated with medical grade silicon to pro- sheep using three loosely tied strings. A PowerLab ML820
vide a biocompatible cover. The total weight of the im- data acquisition unit and Labchart software (ADInstru-
planted equipment was less than 100g. As illustrated by the ments, Sydney Australia) was used for continuous monitor-
cross sectional view in figure 4, thermistors were attached ing of the temperature, the output power to the load and the
to the primary and the secondary coils to measure the tem- variation in input current of the system during power regu-
perature rise caused by the system in the surrounding tissue. lation. The data acquisition was carried out at a frequency
of 10 samples per second.
x Thermistor 1: Placed on top of primary
x Thermistor 2: Placed under the skin
x Thermistor 3: Placed on the muscle side IV. EXPERIMENTAL RESULTS
x Thermistor 4: 1cm from the secondary coil
x Thermistor 5: 2cm from the secondary coil Experimental results were obtained for delivering 10W
x Thermistor 6: Near the subcutaneous tissue near the of power to the load when the system was implanted in
exit of the wound. sheep. Figure 5 illustrates the closed loop controlled power
delivered to the load over a period of 24 hours. The input
Prior to experimentation, the thermistors were calibrated voltage to the system was 23.5V. The controller is able to
against a high precision FLUKE 16 Multimeter temperature control the power to the load for axially aligned separations
sensor and a precision infrared thermometer. and lateral displacements between 10mm to 20mm. Beyond
Primary coil
this range the coupling is too low for the controller to pro-
1 vide sufficient compensation, and delivered power will drop
below the 10W set point. Evidence of inadequate coupling
can be seen at intervals in Figure 5.Variation in input cur-
rent reflects the controller working to compensate for
Subcutaneous tissue changes in coupling using frequency variation between 163
6 2 kHz (fully detuned) and 173 kHz (fully tuned). When the
4 5 coupling between the coils is good, the primary resonant
tank is fully detuned in order to reduce the power trans-
1cm ferred to the secondary. When the coils are experiencing
3
2cm poor coupling, the primary resonant tank is fully tuned to
Secondary coil Healthy tissue
increase the power transfer between the coils.
Fig. 4 The placement of the temperature sensors Graph of closed loop control power at 10 W
12.0 0.8
was shaved. Iodine and disinfectant was applied over the 6.0 0.7
skin to sterilize the area of surgery. Using aseptic tech-
4.0
niques a 5 cm incision was made through the skin on the 0.65
dorsal chest. A tunnel was created under the skin approxi- 2.0 Outp ut Vo lt ag e
inp ut current
mately 20 cm long and a terminal pocket created. The sec-
0.0 0.6
ondary coil and the thermistors were placed within this 0 200 400 600 800 1000 1200 1400
pocket. The thickness of the skin at this site was approxi- Time (mins)
mately 10mm. The secondary coil was then sutured in place
Fig. 5 Regulated power to the load and the input current to the system
and the power lead from the coil and leads of the thermis-
tors were tunneled back to the incision site and exteriorised
through the wound. The wound was stitched and Marcain Figure 6 shows the temperature recorded from the six
was injected to the area of the wound. Iconic powder was thermistors. It takes approximately 20 minutes for the tem-
_________________________________________ ___________________________________________
IFMBE Proceedings Vol. 23
238 T. Dissanayake, D. Budgett, A.P. Hu, S. Malpas and L. Bennet
34.0
32.0
The maximum temperature observed in this system is
30.0
38.10C on the thermistor placed on the muscle side under
28.0
the primary coil. The maximum temperature rise was 3.80C
26.0 on the thermistor placed under the skin.
24.0
22.0
20.0 REFERENCES
0 200 400 600 800 1000 1200 1400 1600
Time (mins) 1. Carmelo A. Milano, A.J.L., Laura J. Blue, Peter K. Smith, Adrian F.
unde r s kin m us c le s ide 2 c m fro m s e c
Hernadez, Paul B. Rosenberg, and Joseph G. Rogers, Implantable Left
ne a r wo und e xit prim a ry s urfa c e 1 c m fro m s e c Ventricular Assist Devices: New Hope for Patients with End stage
Heart Faliure. North Carolina medical journal, 2006. 67(2): p. 110-
115.
Fig. 6 Temperature profile of the thermistors
2. C. C. Tsai, B.S.C.a.C.M.T. Design of Wireless Transcutaneous En-
ergy Transmission System for Totally Artificial Hearts. in IEEE
APPCAS. 2000. Tianjin, China.
V. DISCUSSION 3. Guoxing Wang, W.L., Rizwan Bashirullah, Mohanasankar Sivapraka-
sam, Gurhan A. Kendir, Ying Ji, Mark S. Humayun and James
D.Weiland. A closed loop transcutaneous power transfer system for
Although the system performs well at delivering 10W implantable devices with enhanced stability. in IEEE circuits and sys-
over a 24 hour period, there are short intervals when this tems. 2004.
power level was not delivered. These intervals correspond 4. Ping Si, P.A.H., J. W. Hsu, M. Chiang, Y. Wang, Simon Malpas,
to times when the coupling is too low for the controller to David Budgett Wireless power supply for implantable biomedical de-
vice based on primary input voltage reglation. 2nd IEEE conference
compensate. A variety of approaches can be taken to solve on Industrial Electronics and Applications, 2007.
this problem. The first is to tighten the coupling limitations 5. Ping Si, A.P.H., Designing the DC inductance for ICPT Power pick-
to prevent coupling deteriorating to beyond the equivalent ups. 2005.
limit of 20mm of axial separation. The second approach is 6. Ping Si, A.P.H., Simon Malpas, David Budgett, A frequency control
method for regulating wireless power to implantable devices. IEEE
to allow occasional power drops on the basis that an internal ICIEA conference, Harbin, China, 2007.
battery could cover these intervals, (patient alarms would
activate if the problem persists). The third approach is to
increase the controller tolerance to low coupling. The ability Author: Thushari Dissanayake
Institute: Auckland Bioengineering Institute
to tolerate misalignments of the frequency controlled sys- Street: 70, Symonds Street
tem is mainly determined by the systems quality factor (Q City: Auckland
value), which is defined by: Country: New Zealand
Email: t.dissanayake@auckland.ac.nz
_________________________________________ ___________________________________________
IFMBE Proceedings Vol. 23
Transcutaneous Energy Transfer System for Powering Implantable Biomedical Devices 239
_________________________________________ ___________________________________________
IFMBE Proceedings Vol. 23