Professional Documents
Culture Documents
BT 500 New Catalog Merged
BT 500 New Catalog Merged
Infant Incubator
BT-500
7th Fl., A Bldg., Woolim Lions Valley 5-cha, 302, Galmachi-ro, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea
Tel : 82-31-750-0340 Fax : 82-31-750-0344 E-mail : bistos@bistos.co.kr Http://www.bistos.co.kr
BT-500 Infant Incubator
Weighing Scale
(option)
O2 Control &
Monitoring
Water Tank (1L)
X-ray Tray
Mattress Tilt
Main Display
Basket (option)
Lifting Stand (option)
• Configuration
Standard Temperature & Humidity servo control, Fixed stand, X-ray tray, Tilting,
2 pcs skin temp. sensor, 4pcs Air Filter
Options IV pole, IV plate, Lifting stand, O2 monitoring, O2 servo control, Weighing scale,
Basket, Shelf, External monitor with Masimo SpO2 & CCD camera
• Power-on Self-test
The incubator performs a self-test each time it is turned on.
This process allows the incubator to check various systems
for proper operation. When the test is successfully
completed, the BT-500 displays the main screen.
BT-500 Infant Incubator
• Double Wall
The hood is used to protect an infant from the
external environment and maintain the internal
environment of hood to best condition. BT-500
has double wall structure to minimize heat loss.
The analysis result of Air flow and temperature
distribution are shown as below figure.
Temp after 30 sec. Temp after 40 sec. Z-axis. Temp after 20 sec.
< Analysis inside incubator Air flow and Temp distribution by Fluid Analysis (Yonsei Univ. Korea) >
• Air Curtain
BT-500 has Air Curtain function to reduce temperature fluctuation in hood when the hood door
is opened for treatment of infant.
• Alarm Function
Alarm of BT-500 indicates the condition of failure.
According to each alarm condition, alarm messages
are displayed and alarm sound is caused and
sensor module’s LED blinks.
There are 27 kinds of alarms by failure situation.
System Alarms
Motor failed, Heater system failed, Humidity heater failed, Stuck key, Humidity sensor failure, Air
sensor failure, Fan power failure, Air circulation failure, Skin probe fail, Remove skin 2 probe,
Sensor module disconnected, Sensor module not in position
Temperature Alarms
Low air temperature, High air temperature, Low skin temperature, High skin temperature, High
skin 1 temperature, High skin 2 temperature, High temperature cutout, Over skin temperature
Humidity Alarms
Low humidity, Water empty
Oxygen Alarms
O2 module power failure, O2 module detached, O2 sensor fault, Low oxygen, High oxygen
Weighing Scale Alarms
Too much weight
BT-500 Infant Incubator
• Quick Guide
BT-500 has Help function which provides the simple information about the equipment. The
information is as below :
- Part of full set BT-500
- Main control panel and main display
- How to use and change mode and parameters
• External Monitor
BT-500 uses external 7” color TFT LCD monitor that displays measured values from the control
and video of infant inside the hood. Cam mode, Graph mode and Masimo mode are shown as
below.
Cam Mode (a)
Air temperature, skin temperature, humidity, oxygen concentration and weighing scale are
displayed on external monitor with real-time video of infant.
Graph Mode (b)
Air temperature, skin temperature, humidity, oxygen concentration and weighing scale, SpO2
and PR are displayed on external monitor with real-time video of infant. Also, graphs of each
parameters are displayed.
Masimo Mode (c)
Air temperature, skin temperature, humidity, oxygen concentration and weighing scale are
basically displayed. SpO2 and PR are mainly displayed.
BT-500 Infant Incubator
Specifications : BT-500 Infant Incubator
Functional Characteristics
Temperature Control Humidity Control
Air Control Range Normal : 23.0 ~ 37.0°C Control Range 30 ~ 95%
Override : 37.1 ~ 39.0°C Measurement Range 15~99% ±5% RH(non-condesing)
Air Measurement range 20.0 ~ 45.0°C ±0.5°C Control System Steam(Boiling at 100°C)
Skin Control Range Normal : 35.0 ~ 37.5°C Water Tank 1,000ml (Autoclavable)
Override : 37.6 ~ 39.0°C Display
Skin Measurement range 25.0 ~ 45.0°C ±0.3°C Type 7" TFT Color LCD
Function Hood
Microprocessor Controlled & Double Safety Feature Mattress tilt up to 12°
Trend 9~72hours (Weighing Scale 9days) Noise Level < 45dB
Alarm(Visual & Sound) 27 Events Alarms Air Filter Particle Size 0.3micron
Warm-up Time Silent Mode : 60min, <45dB Air Filter Efficiency 99.8%
Normal Mode : 45min, <55dB Double Wall
Fast Mode : 30min, <66dB Air Curtain
Alarm Sounds for 30min when Power Outage X-ray Tray
Multi-language Masimo SpO2 with CCD & External Monitor (Option)
Quick Guide Display 7" TFT Color LCD
Power On Self Test SpO2 Range 1~ 100%
O2 Measurement (Option) SpO2 Accuracy 70 ~ 100% ±3digits
Measurement Range 18 ~ 100% ±5% 0 ~ 69% : Unspecified
O2 Servo Control (Option) Pulse Rate Range 25 ~ 240bpm ±3digits
Control Range 21 ~ 65% ±5% CCD Camera 510 x 492 pixels
Weighing Scale (Option) Video Recording
Measurement Range 0 ~ 10.0Kg ±10g
Shelf (Option) Basket (Option)
Weight Limit Approx. 3Kg Weight Limit Approx. 10Kg
IV Plate (Option) PC Interface
Weight Limit Approx. 11Kg RS232C
Power
Input AC 110/220, 50/60Hz Consumption 950VA
Output AC 110/220, 50/60Hz, Max.400VA
Standard Configuration
Skin Temperature Sensor 2ea Air Filter 4ea
Power Cord 1ea Operation manual 1ea
Options
Lifting Stand Ringer Pole
IV Plate Weighing Scale
Basket Shelf
O2 Monitoring O2 Servo Control (O2 Monitoring Included)
Masimo SpO2 with CCD & External Monitor
Warranty
Main Unit 2years
Physical Characteristics
Dimension Weight
Standard 1,024(W) x 690(D) x 1,354(H)mm Standard / Full Option 99.3Kg / 117.5Kg
Stand Fixed / Lifting 645(H) mm / 651 ~ 851(H)mm
Mattress 730(W) x 380(D) x 27(H)mm
Packing 1,150(W) x 800(D) x 1,530(H)mm Packing 126Kg(Full Option)
Environmental Conditions
Operating Temperature 20 ~ 30°C (68 ~ 86°F) Storage Temperature –20 ~ 60°C (–4 ~ 140°F)
Operating Humidity 0 ~ 95% non-condensing Storage Humidity 0 ~ 95% non-condensing
Certificates
Certificates KFDA, CE, FDA Standard IEC60601-1, IEC60601-1-2
IEC60601-2-19
* All specifications can be different by country and are subject to change without notice.
Neonatal Care
System
Infant Incubator
Infant Warmer
Phototherapy
BT-500 Infant
Incubator
Environment designed for premature baby
ouble Wall and air curtain maintain constant temperature
D
control in the hood and help the infant keep their
temperature.
L ow operation noise and the part that makes noise is designed
to make only minimal noise to ensure quiet environment
In-built x-ray tray and weighing scale minimize the necessity
for premature baby leave the incubator for additional care.
Double wall
Swivel head
Adjustable position
Tilting*
Removable wall Highly effective blue LED light Extensive long LED life time up to 100,000 hours
Fan-less design for noiseless treatment Displays treatment time and total used timeb
Intensity control (high & low) Built-in clamp provides various installation options
Model BT-550
Category Infant Warmer
Temp. Control Mode Prewarm / Baby / Manual
Control 34 - 38 °C
Model BT-500 Skin Temp.
Measurement 26 - 42 ±0.3°C
Category Infant Incubator Type Infrared
Normal : 23 - 37°C Control 0 - 100% / 20 Level
Control
Air Temp. Override : 37.1 - 39°C Heater Power 600W
Measurement 20 - 45 ±0.5°C Life Time 5,000 Hour
Rotation 90°
Normal : 35 - 37.5°C
Control Setting 0 - 59 min. 59 sec.
Skin Temp. Override : 37.6 - 39°C APGA Timer
Beep 1, 5, 10 min.
Measurement 25 - 45 ±0.3°C Power 40W LED (10W x 4)
Control 30 - 95% Examination Control 3 Level
Measurement 15 - 99 ±5% RH Illumination > 7,000 lx
Humidity Scale * Measurement 0 - 10Kg ±10g
System Steam (Boiling at 100°C)
Measurement 1 - 100%
Water Tank 1,000ml (Autoclavable)
Masimo ± 3% (70-100%)
Measurement 0 - 10Kg ±10g Accuracy
Scale * SpO2 * Unspecified (0~69%)
Trend 216 Hour Pulse Rate Range 25 - 240bpm ±3bpm
Control 21 - 65 ±5% Tilt * < 15°
O2 *
Measurement 18 - 100 ±5% Bed Matress 80 x 49 x 2cm
Measurement 1 - 100% X-ray tray 35 x 40 cm
Lifting Stand * Height 65 - 85 cm
Masimo ± 3% (70-100%)
Accuracy IV Plate * Load limit 11 Kg
SpO2 * Unspecified (0~69%) IV Pole * Hook 2 Hook
Pulse Rate Range 25 - 240bpm ±3bpm Basket * Load limit 10 Kg
Resolution 2.5M pixel Shelf * Load limit 3 Kg
CCD Camera * Oxygen Delivery * Regulator, Humidifier, Bracket
Video Transfer Wi-Fi
Display 7” TFC Color LCD
Hand Port 6 Door
Alarm 6 Events (Visual & Sound)
Tubing Port 2 x 6 Port General
Self Test Power On
Hood Structure Double wall Leveler Bubble type
Air Curtain Support Standard (HWD) 1,890 x 1,027 x 690 mm
Dimensions
Noise < 45dB With Lifting (H) 1,780 - 1,980 mm
Tilt < 12° Standard 83 Kg
Weight
Maximum 98 Kg
Bed Matress 72 x 39 x 2 cm
Warranty Main unit 2 year
X-ray tray 35 x 40 cm
Lifting Stand * Height 65 - 85 cm Model BT-400
IV Plate * Load limit 11 Kg Category Phototherapy
IV Pole * Hook 2 Hook Light Source Blue LED
Basket * Load limit 10 Kg Wavelength Peak Between 450 - 475 nm
Shelf * Load limit 3 Kg Low 25 - 35 μW/cm²/nm
Intensity
High 35 - 55 μW/cm²/nm
Particle 0.3 Micron
Air filter Intensity Variation (over 6 hrs) ±10%
Efficiency 99.8% LED Life Time 100,000 Hour
Display 7” TFC Color LCD Effective Surface (at 40cm) Approx. 40 x 20 cm
Trend 72 Hour Heat output over 6hours < 10 ℃ than ambient
Alarm 27 Events (Visual & Sound) Audible Noise < 30 dB (Fanless)
Clamp Max. diameter 3.5cm
General Warm up < 30 Minute (Fast Mode)
Shade * Covers Incubator and Warmer
Self Test Power On
Cart * Height : 74 - 118cm
Multi Language Support Display 2.4” TFT Color LCD
Quick Guide In-built General Time Display Treatment time, Total used time
Standard (HWD) 1,354 x 1,024 x 690 mm Timer 30 min. - 999 hrs.
Dimensions Head 75 x 340 x 210 mm
With Lifting (H) 1,244 - 1,444 mm Dimensions(HWD)
Cart 1000 x 430 x 520 mm
Standard 99.3 Kg
Weight Main unit 3.6 Kg
Maximum 117.5 Kg Weight
Main unit & Cart 13 Kg
Warranty Main unit 2 year Warranty Main unit 2 year
bistos@bistos.co.kr
www.bistos.co.kr
CTD-NICU-001 / Rev.01