ACQUI SI T I ON I NF ORMAT I ON
The test passes if the Average ISM Retention Time Offset is less than or equal to 5.0 seconds and
the standard deviation for the ISM retention time offset is less than or equal to 1.0 seconds.
The test passes if the Caffeine sample peaks of the Optical Detector have a %RSD less than or equal
to 1.0% for Area, a %RSD less than or equal to 1.5% for Height and a Standard deviation less than or
equal to 1.0 seconds for Retention Time.
0.60
Caffeine - 0.591
0.40
AU
0.20
0.00
0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00
Minutes
Sample Name: Caffeine; Injection: 1
Sample Name: Caffeine; Injection: 2
Sample Name: Caffeine; Injection: 3
Sample Name: Caffeine; Injection: 4
Sample Name: Caffeine; Injection: 5
Sample Name: Caffeine; Injection: 6