Professional Documents
Culture Documents
Ais-Mr15e91022-01-00 Mission Foods Malaysia SDN BHD
Ais-Mr15e91022-01-00 Mission Foods Malaysia SDN BHD
Ais-Mr15e91022-01-00 Mission Foods Malaysia SDN BHD
AIS-MR15E901022-01-00
X-RAY INSPECTION SYSTEM
Date February 5, 2016
SAMPLE TEST REPORT Approved by K.Sawano
Checked by M. Yagawa
We would like to report on our sample test with Tested by CS Eng. N. Arikawa
your product.
This report was ordered by ANRITSU INFIVIS CO., Ltd.
Test result:
Product Wrap
Dimensions (mm)/ Mass (g) 310L×250W×70H/1.9
Tube voltage, Tube current 80 kV, 4.3 mA
Test pieces Sensitivity (mm)*1*2
SUS wire (SUS304: 7.93 g/ cm3) Dia.0.3×2(6) to Dia.0.4×2(10)
Dia.0.6(6) to Dia.0.7(10)
SUS ball (SUS304: 7.93 g/ cm3)
Sensitivity within the mask: Dia.1.5(8) to Dia.2.0(10)
Glass ball (SK-C: 2.49 g/ cm3) Dia.2.0(2) to Dia.3.0(10)
Ceramics ball (Al2O3: 3.9 g/cm3) Dia.2.0(8) to Dia.3.0(10)
PTFE ball (PTFE: 2.18 g/cm3) Dia.4.0(10)
Rubber ball (EPDM-90: 1.3 g/ cm3) Dia.4.0(8) to Dia.5.0(10)
Nylon ball (Nylon: 1.14 g/cm3) JIMA card not detectable-estimate cannot be provided
Aluminum ball (Aluminum: 2.78 g/cm3) Dia.2.0(3) to Dia.3.0(10)
Missing deoxidizer detection Detectable( (10)
)
Example 1) Fe Dia.. A to B
Detectable in any part
Possible at a certain part of product
-1-
Document No. AIS-MR15E901022-01-00
Comment:
1. As for the products, Wrap, we used Deoxidizer Mask function to minimize the effect of the
deoxidizer. The deoxidizer may not be masked properly (Masked Area is shown in blue)
depending on the variation of positions and how the deoxidizer overlaps the product. Only metal
contaminants in the Masked Area (blue part in the image) can be detected. The detection
sensitivity in the Masked Area is lower than the outside of the masked area.
Note:
This report represents the test result of your sample product that was tested with the test pieces
shown in the test result column and other contaminants. Please be advised that this may not
guarantee the sensitivity which would be achieved in an actual production line. The detection
sensitivity changes with the variation of product characteristics (package material, configuration,
density etc.), the direction of contaminants, and the environmental conditions of an X-ray.
With the irraDia.tion time limitation, errors (photocell problem) may occur depending on the
relationship between belt speed and product length. In our tests, we may have tested at “bulk
flow setting” to circumvent the limitation.
-2-
Document No. AIS-MR15E901022-01-00
PASS product
Detectable
Concluded
-3-