ECL 760 Instrument IP Setting for LIS

When using the LIS function the instrument must be set an IP. The IP of instrument and the
computer must be in the same network segment.
E.g. if computer IP is 192.168.1.26, the instrument IP must be “192.168.1.*”.
The instrument IP setting steps are showing as bellow.
1. Close the “FBWF” Protect. “FBWF.exe” is in the disc C.
1.1 Press the keys on the keyboard “Ctrl”+”Alt”+”Delete” at the same time. And then Press
the “Task Manager” to open task manager.

1.2 Select click “Analyzer.exe” click button “End Process”. And then press “Yes” to close.

1.3 Press “Applications” and then press button “New Task”. Input explor.exe, clink “OK”

1/4
1.4 Enters into Disc C, find the “FBWF.exe” double-clink it. Inputs pass word: 1008

Press button “ ” to close RBWF Protect. And then press button “ ” to

restart system.

1.5 Repeat the steps 1.1 to1.3 again. Find “my network places”. Scan Network Connections,

1.5 Inserts LAN cable into the port of the instrument. You can see network card is reactive.
The hardware was connecting.

2/4
1.6 Press right clink menu, select “properties”, select Internet Protocol (TCP/IP), and then setting
IP. That in the same network segment to service computer IP.

1.7 Press the keys on the keyboard “Ctrl”+”Alt”+”Delete” at the same time. Press button “shut
down”, select “restart” to restarts the instrument system.

3/4
1.8 Repeat the steps 1.1 to1.3 again. Enter into Disc C. Find the “FBWF.exe” double-clink it. Inputs
pass word: 1008 Press button “ ” to open RBWF Protect. And then press button
“ ” to restart system.

Now the instrument IP has setting.