Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

IMAGE PILOT Service tool and monitoring tool

Service tool: 5678 , Operation Tool

Go to Manual: ImagePilot Ver 1.50 Installation-Service MANUAL


0604YF320L_110425_Fix

Chapter 12

Operation settings (12.2.5)

1. Uncheck [ ] Strong password


2. Check [x] add exam button in search result
3. [x] Add physician name and technician name to the pull down for manual
input
4. [x] Automatically move diagnosis completed patient to last line
5. [x] Manage unconfirmed status User-Wise
6. [x] Clear patient Info on returning from Viewer
7. [x] Enable the remote maintenance function
8. Image processing parameter [Edit] MultiView Image [x] On

Shutter transparency = 30%

9. [x] Chest AP/PA Auto flip


10. Overlay Items [edit] [x] Display OK button
11. Dicom standard violation [settings] [x] Media retrieval
12. [x] Frame output of Multi Frame images
13. [x] Create data folder even a single file is written
14. [x] Batch output anchor settings
15. [x] Combine the search item in batch output screen & Normal search
16. [x] Dicom data editor
17. [x] Print compose mode
18. [x] Edit other patient on Print composer
19.[x] NG Reason [x] Save NG reason
Modality configuration: Up to 4 modalities can be configured to send to the IP.
One license is needed for each modality.

Service tool(5678)

19. Option Settings  Modality  [x] M2

20. System configuration Modality [x} display name Ex. US

[x] Modality name Ex. US

21. Communication settings  Modality settings (here enter the modality


information). If modality sends Cine: [x] Series join.

 Local port: Ex 104, 52102, etc

22. Operation settings  Configure batch output for modalities

 Edit: Ex [x] US > Output: PACS

23. Operation settings [x] Dicom image reception…Select modality (Ex: US)
and Uncheck the others three.

24. Client settings  Viewer screen [Edit] On display name move the modality,
Ex: US
Service tool: 2923, Reader

Reader settings

Process for AEC adjustment [Settings]

AEC01  [x] Menu display

Export  Backup settings, logs (System settings, Database and Logs)

Image This is different from the user backup. Only can be seen in the same
Image Pilot with the same SW version.

Monitoring tool: Right click, Login: 5678

Monitoring tool: Right click, Login: 9012  This is for reject analisis.