Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

E18 − 19

A1.4.3.1 The results of an indirect verification shall comply ranges that are available. Also, the testing machine shall be
with the current values specified in Table A1.3 regardless of verified on another Rockwell scale which uses the same test
any maximum error E value marked on the test block (see Note forces and for which three blocks are available. In this case, the
8). testing machine is considered verified for the entire Rockwell
A1.4.4 The indenters to be used for the indirect verification scale.
shall meet the requirements of Annex A3 (see Note 3). (2) Alternative Procedure 2—This procedure may be used
when standardized blocks from two ranges are available. The
A1.4.5 The testing cycle to be used for the indirect verifi- testing machine shall be verified using the standardized blocks
cation shall be the same as is typically used by the user. from the two available ranges. In this case, the testing machine
A1.4.6 Prior to performing the indirect verification, ensure is considered verified for only the part of the scale bracketed by
that the testing machine is working freely, and that the indenter the levels of the blocks.
and anvil are seated adequately. Make at least two hardness A1.4.9.2 On each standardized test block, make five mea-
measurements on a suitable test piece to seat the indenter and surements distributed uniformly over the test surface. Deter-
anvil. The results of these measurements need not be recorded. mine the error E and the repeatability R in the performance of
the testing machine using Eq 2 and Eq 3 for each hardness level
A1.4.7 As-found Condition: of each Rockwell scale to be verified.
A1.4.7.1 It is recommended that the as-found condition of A1.4.9.3 The error E and the repeatability R shall be within
the testing machine be assessed as part of an indirect verifica- the tolerances of Table A1.3. The indirect verification shall be
tion. This is important for documenting the historical perfor- approved only when the testing machine measurements of
mance of the machine in the scales used since the last indirect repeatability and error meet the specified tolerances using at
verification. This procedure should be conducted prior to any least one of the user’s indenters.
cleaning, maintenance, adjustments, or repairs. A1.4.9.4 In the case that the testing machine cannot pass the
A1.4.7.2 When the as-found condition of the testing ma- repeatability and error verifications with the user’s indenter, a
chine is assessed, it shall be determined with the user’s number of corrective actions may be attempted to bring the
indenter(s) that are normally used with the testing machine. At testing machine within tolerances. These actions include clean-
least two standardized test blocks, each from a different ing and maintenance, replacing the anvil or using another of
hardness range as defined in Table A1.3, should be tested for the user’s indenters. The indirect verification procedures shall
each Rockwell scale that will undergo indirect verification. The be repeated after making the allowed corrective actions.
difference in hardness between any of the standardized test
NOTE A1.3—When a testing machine fails indirect verification, it is
blocks shall be at least 5 hardness points for each Rockwell recommended that the testing machine be verified again using a Class A
scale. (or better) indenter for those scales and hardness levels that failed the
A1.4.7.3 On each standardized test block, make at least two indirect verification with the user’s indenter. If the testing machine passes
measurements distributed uniformly over the test surface. the repeatability and error tests with a Class A indenter, it is an indication
that the user’s indenter is out of tolerance. A new indenter may be acquired
A1.4.7.4 Determine the repeatability R and the error E (Eq by the user as a corrective action (see A1.4.9.4) allowing the indirect
2 and Eq 3) in the performance of the testing machine for each verification procedures to be repeated without having to perform a direct
standardized test block that is measured. verification. If the testing machine continues to fail the repeatability or
A1.4.7.5 The error E and the repeatability R should be error tests of an indirect verification with the Class A indenter, it is an
within the tolerances of Table A1.3. If the calculated values of indication that there is a problem with the machine and not the user’s
indenter.
error E or repeatability R fall outside of the specified
tolerances, this is an indication that the hardness tests made A1.4.9.5 If the testing machine continues to fail the repeat-
since the last indirect verification may be suspect. ability or error tests following corrective actions, the testing
machine shall undergo adjustment and/or repair followed by a
A1.4.8 Cleaning and Maintenance—Perform cleaning and direct verification.
routine maintenance of the testing machine (when required) in
A1.4.10 Qualifying Additional User’s Indenters—In cases
accordance with the manufacturer’s specifications and instruc-
where the testing machine passes indirect verification using
tions.
only one of the user’s indenters, only that one indenter is
A1.4.9 Indirect Verification Procedure—The indirect verifi- considered verified for use with the specific testing machine for
cation procedure requires that the testing machine be verified the Rockwell scales that were indirectly verified using that
using one or more of the user’s indenters. indenter. Before any other indenter may be used for testing the
A1.4.9.1 One standardized test block shall be tested from same Rockwell scales, it must be verified for use with the
each of the hardness ranges (usually three ranges) for each specific verified testing machine. This requirement does not
Rockwell scale to be verified, as given in Table A1.3. The apply to changing an indenter ball. The indenter verifications
difference in hardness between any of the standardized test may be made at any time after the indirect verification, and
blocks shall be at least 5 hardness points for each Rockwell may be performed by the user as follows.
scale. The user may find that high, medium and low range test A1.4.10.1 The testing machine and indenter shall be verified
blocks are unavailable commercially for some scales. In these together using the indirect verification procedures of A1.4.9
cases, one of the following two procedures shall be followed. with the following exception. The verification shall be per-
(1) Alternative Procedure 1—The testing machine shall be formed on at least two standardized test blocks (high and low
verified using the standardized blocks from the one or two ranges) for each Rockwell scale that the indenter will be used.

Copyright ASTM International


Provided by IHS Markit under license with ASTM
12
No reproduction or networking permitted without license from IHS

You might also like