Professional Documents
Culture Documents
QS-550 Service Manual
QS-550 Service Manual
QS-550 Service Manual
Free-Standing Tube
Stand
Service Manual
J 2012-08-29 Corrected Tables 2-1 & 2-2 per ECO 2686 updates
i - xiv M
1-1 thru 1-6 M
2-1 thru 2-28 M
3-1 thru 3-6 M
4-1 thru 4-10 M
5-1 thru 5-6 M
SAFETY NOTICES
GENERAL SAFETY INFORMATION ............................................................ v
WARNINGS, CAUTIONS, NOTES ............................................................... v
REGULATORY COMPLIANCE ................................................................... vii
CLASSIFICATION ............................................................................viii
COMPATIBILITY ..............................................................................viii
INTENDED OPERATOR ....................................................................viii
TRAINING ....................................................................................... ix
ACCOMPANYING DOCUMENTATION .................................................. ix
Applicable Standards ........................................................................ ix
ENVIRONMENTAL PROTECTION .............................................................. ix
ELECTROMAGNETIC COMPATIBILITY
(EN 60601-1-2:2007/IEC 60601-1-2:2007) .............................................. ix
ABBREVIATION DEFINITION .................................................................xiv
CHAPTER 1 - SPECIFICATIONS
PHYSICAL SPECIFICATIONS .................................................................1-3
SYSTEM SPECIFICATIONS ..............................................................1-3
HANDGRIP SPECIFICATIONS ..........................................................1-3
PERFORMANCE SPECIFICATIONS ....................................................1-3
ELECTRICAL SPECIFICATIONS ..............................................................1-4
SYSTEM OPTIONS ................................................................................1-4
SYSTEM OPERATING ENVIRONMENT ....................................................1-5
NON-OPERATING ENVIRONMENT .........................................................1-5
COMPATIBILITY STATEMENT ................................................................1-5
CHAPTER 5 - DIAGRAM
Model QS-550 Tube Stand Sample Configuration Diagram ...................... 5-3
Model QS-550 Tube Stand Wiring Diagram - Manual Collimator ............... 5-4
Model QS-550 Tube Stand Wiring Diagram - Automatic Collimator
(Progeny Linear II) .............................................................................. 5-5
Model QS-550 Tube Stand Wiring Diagram - Progeny Linear IV
Automatic Collimator ........................................................................... 5-6
The purpose of safety icons, such as those shown below, is to indicate at a glance the
type of caution, warning or danger.
WARNING
Quantum Medical Imaging, LLC disclaims all responsibility from any injury
resulting from improper application of this equipment.
This equipment is sold to be used exclusively under the prescribed direc-
tion of a person who is licensed by law to operate equipment of this nature.
This equipment must be used in accordance with all safety procedures
described in this manual and must not be used for purposes other than
those described herein. In the United States, Federal law restricts this
device to sale, distribution, and use by or on order of a licensed physician.
Quantum Medical Imaging, LLC cannot assume responsibility for any mal-
functioning of this equipment resulting from improper operation, mainte-
nance, or repair, or from damage or modification of its components.
Failure to observe these warnings may cause serious injuries.
X-RAY PROTECTION
WARNING! X-rays are hazardous to both patient and operator unless
established safe exposure factors and operating instructions are observed.
Only qualified and authorized personnel shall operate this system. In this
context, qualified means those legally permitted to operate this equipment
in the jurisdiction in which the equipment is being used, and authorized
means those authorized by the authority controlling the use of the equip-
ment. Full use must be made of all radiation protection features, devices,
systems, procedures and accessories.
It is important that everyone having anything to do with x-radiation be
properly trained and fully acquainted with the recommendations of the
National Council on Radiation Protection and Measurements as published
in NCRP Reports available from NCRP Publications, 7910 Woodmont Ave-
nue, Suite 800, Bethesda, Maryland 20814-3095 (www.ncrp.com), and of
the International Commission on Radiological Protection (www.icrp.org),
and take adequate steps to protect against injury.
WARNING
X-ray equipment may cause injury if used improperly. The instructions con-
tained in this manual must be read and followed when operating this unit.
Personal radiation monitoring and protective devices are available. You are
urged to use them to protect against unnecessary X-ray exposure.
CLASSIFICATION
This product has been classified as Class I, Type B by Underwriters Laboratories,
Inc. Equipment not suitable for use in the presence of a flammable anesthetic
mixture of air with oxygen or with nitrous oxide. Protection against Harmful
Ingress of Water (Ordinary), enclosed equipment without protection against
ingress of liquids.
The following symbols may be used for marking on this equipment or equipment
documentation:
COMPATIBILITY
The equipment described in this manual must only be used in combination with
other equipment or components if these are expressly recognized by Quantum
Medical Imaging, LLC as compatible.
INTENDED OPERATOR
This equipment is intended to be installed, used and operated only in accordance
with the safety procedures given within this manual for the purpose for which it
was designed. Before attempting to work with this equipment, read, understand,
note and strictly observe all warnings, cautions and safety markings on the equip-
ment.
Users include those persons who actually handle the equipment and those who
have authority over the equipment.
TRAINING
Users of this equipment shall have received adequate training on its safe and
effective use before attempting to work with the equipment. Training require-
ments may vary from country to country. The User shall make sure that training is
received in accordance with local laws or regulations that have the force of law.
ACCOMPANYING DOCUMENTATION
The documentation consists of a Model QS-500 Floor to Ceiling/Wall Tube Stand
and QS-550 Free-Standing Tube Stand Operator Manual (DC30-034).
The documentation shall be kept with the system for easy reference.
APPLICABLE STANDARDS
This equipment complies with the following regulatory standards:
• EN 60601-1: 1990 + A1:1993 + A2:1995 + A3:1996
• IEC 60601-2-32: 1994
• CAN/CSA-C22.2 No. 601.1-M90, 2005 (Medical Electrical Equipment, part 1:
General Requirements for Safety)
• UL 60601-1, 1st Edition, 2006-04-26 (Medical Electrical Equipment, part 1:
General Requirements for Safety)
• IEC 60601-1 Medical electrical equipment, Part 1: General requirements for
safety
• IEC 60601-1-2: 2007
ENVIRONMENTAL PROTECTION
This equipment contains certain materials and chemical compounds incidental to the
manufacture of electrical and electronic equipment, and improper “end-of-life” dis-
posal of such equipment can result in environmental contamination. Therefore, this
equipment should not be disposed of as ordinary household waste, but should instead
be delivered to a designated electrical and electronic waste disposal or recycling cen-
ter. For further information on disposing of electrical and electronic waste, contact
the cognizant authority within the jurisdiction.
Use of accessories, transducers and cables, other than those specified in the accom-
panying documents, may result in increased emissions or decreased immunity of the
equipment.
The Tube Stand is intended for use in the electromagnetic environment specified below. The
customer or the user of the Tube Stand should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment -
guidance
The Tube Stand is intended for use in the electromagnetic environment specified below. The customer or the
user of the Tube Stand should assure that it is used in such an environment.
Immunity test IEC 60601 Compliance Electromagnetic environ-
test level level ment - guidance
Electrical fast ±2 kV for power supply ±2 kV for power supply Mains power quality should be
transient/burst lines lines that of a typical commercial or
IEC 61000-4-4 ±1 kV for input/output ±1 kV for input/output hospital environment.
lines lines
NOTE: UT is the A.C. mains voltage prior to application of the test level.
The Tube Stand is intended for use in the electromagnetic environment specified below. The customer or
the user of the Tube Stand should assure that it is used in such an environment.
Immunity test IEC 60601 Compliance Electromagnetic environment - guid-
test level level ance
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the Tube Stand is used exceeds the applicable RF compliance level above, the Tube Stands (including Tech-
Vision option) should be observed to verify normal operation. If abnormal performance is observed, addi-
tional measures may be necessary, such as re-orienting or relocating the Tube Stand.
b
Over the frequency range 150 kHz to 80 kHz, field strengths should be less than 3 V/m.
The Tube Stand is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Tube Stand can help prevent electromagnetic interfer-
ence by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Tube Stand as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum output Separation distance according to frequency of transmitter
power of transmitter m
W
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation dis-
tance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
ABBREVIATION DEFINITION
The following abbreviations and acronyms may be found in this document. Their def-
inition is explained below.
cm Centimeters
CCW Counterclockwise
CW Clockwise
in. Inches
GHz Gigahertz
mm Millimeters
MHz Megahertz
W Watts
SPECIFICATIONS
1-1
1-2
Chapter 1 Specifications
PHYSICAL SPECIFICATIONS
Refer to Figure 5-1 in Chapter 5, DIAGRAMS for Model QS-550 Tube Stand dimen-
sions.
SYSTEM SPECIFICATIONS
• Column Height: 2336.8 mm (92.0 in.)
• Tube Mount: Fork or Trunnion
• Weight: 272.0 kg (600 lb)
• Collimator
Compatibility: Manual, Selectable or Automatic, 4.5 to 9.1 kg
(10-20 lb) typical. Mounting to fork/tube housing
requires four 1/4-20 or M6 x 16 bolts spaced on a
92 mm (3.625 in.) diameter circle
• X-ray Tube
Compatibility: 76.2 or 101.6 mm (3 in. or 4 in.) rotating anode
type, 13.6 to 18.1 kg (30 - 40 lb) typical)
• Max combined load
(Tube & Collimator) 31.8 kg (70 lb)
• Heat Dissipation: 200 BTU per hour
• Compatible X-Ray
Tube Housing Styles: Toshiba E7239, E7242, E7252 E7254, E7269; Var-
ian Diamond, Varian Sapphire, Dunlee DA10
• Compatible
Collimators Carestream MC150, MC150 Pinnacle; Linear II, Lin-
ear IV, Ralco R 221/A DHHS; Huestis 150MC,
Huestis CM32211-QMI (Programmable)
HANDGRIP SPECIFICATIONS
• Push Button
Switches: Vertical lock release, Longitudinal lock release,
Angulation lock release, Transverse lock release, All
Locks release, Auto-Stop
• LED Indicators: 100 cm (40 in.) SID, 180 cm (72 in.) SID
PERFORMANCE SPECIFICATIONS
• Tube Stand
Longitudinal Travel: 1752.6 mm (96.0 in.)
• Vertical Travel (x-ray
tube): 1524.0 mm (60.0 in.)
control panel. Refer to the table below for the appropriate documentation for
the application:
OPERATOR SERVICE
APPLICATION TYPE
MANUAL MANUAL
ASSEMBLY &
INSTALLATION
2-1
2-2
Chapter 2 Assembly & Installation
OVERVIEW
This chapter describes the steps required to assemble and install the Model QS-550
tube stand, including making all required electrical connections, and performing all
required alignments and calibrations.
ROOM CONSTRUCTION
A conventional radiographic room construction should be used in the area where
the tube stand is to be installed. Consult with your State Health Department or
local Building Codes for specific radiation shielding requirements.
INSTALLATION PRECAUTIONS
Install the tube stand before installing the radiographic table.
Check that the space location, conduit arrangements and junction boxes are pre-
pared for the permanent installation of the tube stand. Verify that the location
and space allowed is sufficient for the full range of tube stand/Transverse Arm
travel without hitting any walls or structures in the room (see Figure 5-1 for
equipment dimensions). Note that the Transverse Arm in/out adjustment and the
Vertical Column position relative to the rear wall are critical. Column rotation in
the counterclockwise (right-hand) direction could result in the Tube Housing strik-
ing the wall before the 90 degree index position is reached.
CDRH CERTIFICATION
The tube stand is not a CDRH certifiable product. However, it must be compatible
and operate in conjunction with other components in the x-ray system so that the
x-ray system performs in compliance with CDRH performance standards.
REQUIRED TOOLS AND MATERIALS
The following tools and materials are required to complete the assembly and
installation procedures:
• Double-bubble level - two feet minimum length
• Tape Measure - eight feet minimum length
• Standard hand tools
• Alcohol - to clean the high-tension cable ends, tube head, and high- tension
transformer sockets
• Dow-Corning DC4 vapor-proofing compound, silicon insulating grease or
equivalent - for packing x-ray tube head and high-tension cable terminals
• Special film cassette (11" x 14" cassette with scribed center line)
• Electric drill and drill bits
• Lubricant - white lithium grease
UNPACKING
Open crate or carton marked “packing list enclosed” first. Remove packing list
and use it as a guide to open the remaining cartons. Do not dispose of packing
material until packing list is matched with actual parts received. Should there be
a shortage or damage, notify Quantum Medical Imaging, LLC.’s customer service
department immediately. The manufacturer is relieved of any responsibility for
damage during shipment after unit is picked up by the carrier.
NOTE: Examine all cartons and crates at time of delivery. If
damage is apparent, have delivery driver write a “Damaged
Shipment Note” on copies of freight bill, sign it, and file appro-
priate carrier claim. Should you discover concealed damage,
immediately notify the transporting agent and ask for an
“Inspection of Damage”. Carrier will not accept concealed
damage claim if filed after 15 days from date of receipt of mer-
chandise
INSTALLATION INSTRUCTIONS
When the tube stand is used with a vertical wall frame or wall stand, care should
be taken during room planning and pre-installation to allow for the accurate align-
ment of these components. Alignment accuracy must be within 2 % of the SID
used. After the tube stand Floor Track is aligned with the table receptor, the track
cannot be adjusted transversely. Therefore, transverse alignment of the x-ray
field to the image receptor in the vertical wall frame or wall stand must be made
by adjusting the position of the image receptor within the wall stand or relocation
of the entire wall stand.
All mounting procedures must be approved before and checked after installation
by the hospital or facility engineering department and a registered, professional,
structural engineer. It is recommended that a leakage current measurement be
taken following installation to verify system meets allowable levels.
To assemble and install the tube stand, proceed as follows:
Floor Track Installation
The area of the floor on which the Floor Track is to be mounted must be
level in all directions. Shims may be used, however, grouting is recom-
mended.
1. Position the Floor Track in the proper location for installation.
2. Using a bubble-level, check that the Floor Track is level both lon-
gitudinally and transversely. If not, shims can be used to level
track, however, grout is recommended.
3. Fasten the Floor Track to the floor using appropriate fastening
hardware (capable of supporting 100 lb. pull-load) depending on
the installation site (see Figure 2-1).
4. Install five (5) Hole Plugs in holes for floor mounting screws.
MOUNTING
HOLES (5X)
RETAINING
RING
FRONT
BEARING
BRACKET
BEARING W/
ECCENTRIC
BUSHING (2X)
REAR
BEARING
BRACKET (2X)
1. Re-install the End Cap to the end of the Floor Track using four (4) 5/16-18 x
1/2” hex head bolts, lockwashers and flat washers.
2. Using a two-foot (minimum) double-bubble level, placed on the front of the
Vertical Column, check that the Vertical Column is plumb. If not, adjustment
can be made by rotating the two (2) Vertical Angle Adjustment Bearings with
eccentric bushings (see Figure 2-4). Loosen the two (2) eccentric bearings
mounted to the rear of the Base Assembly; rotate the bearings to adjust. Re-
tighten the bearings after the column angle adjustment is made.
INSTALLING THE X-RAY TUBE, COLLIMATOR AND HANDGRIPS
The system is supplied with either a Trunnion type x-ray tube mounting bracket or
a Tube Mounting Fork. Refer to the appropriate paragraph that follows corre-
sponding with the type of x-ray tube mounting bracket used on your system. On
systems equipped with optional TechVision Hand Grips, refer to TechVision Ser-
vice Manual for complete hand grips mounting instructions.
Trunnion Configuration
1. Install the Trunnion Assembly onto the Transverse Arm using four
(4) 1/4-20 x 2" hex head bolts and lockwashers (see Figures 2-5
and 2-6). There are two sets of four (4) Trunnion mounting holes
that can be used as shown in Figure 2-5. The upper set of holes
is used when mounting 3-inch x-ray tubes. The lower set of
holes is used for mounting 4-inch x-ray tubes.
2. Remove top half of Trunnion Rings.
3. Place x-ray tube onto lower half of Trunnion Rings and re-mount
upper half.
4. Install Trunnion Handle on Trunnion Assembly using two (2) 1/4-
20 x 1/2” button head socket screws (see Figure 2-7).
TUBE SUPPORT
BRACKET
TRUNNION
ASSEMBLY
TRUNNION
HANDLE
1/4-20 X 1/2”
BUTTON HEAD
TRUNNION SOCKET SCREW
CAM (2X)
HANDGRIP
ASSEMBLY
COLLIMATOR
MOUNTING RING
CANBUS DATA
NETWORK
CABLE (AY51-123)
ETHERNET
PORT
CANBUS
POWER CANBUS NOT NETWORK
FROM DATA 1 USED ETHERNET
QS-550 PORT
DATA1
NOT
NETWORK USED
COLLIMATOR ETHERNET
(OPTIONAL) PORT CANBUS
DATA1
QS-550 POWER
AND LOCKS
TUBE FORK
COLLIMATOR
LEVELING SET
SCREW (4X)
TRANSVERSE ARM
INNER COVER
TRANSVERSE ARM
OUTER COVER
ACCESS
OPENING
Figure 2-14. Trim Weight Access Opening (Front Side of Vertical Column)
LONGITUDINAL
BRAKE
ASSEMBLY
COLUMN ROTATION
ADJUSTMENT
SCREW (2X)
HORIZONTAL
SID SWITCH
ASSEMBLY
The Horizontal SID Switch Assembly mounts inside the Base Assembly (on the left
side when facing front of column) as shown in Figure 2-16. Proceed as follows:
1. Loosely install Horizontal SID Switch Assembly on rear wall of Base Assembly
using two (2) 10-32 x 1/2” phillips pan head screws, flat washers and kep
nuts.
2. Using a Switch Actuator placed on the Floor Track as a reference (do not peel
off adhesive backing at this time), adjust the height of the switch so that
when it contacts the actuator, the switch activates (clicks).
3. When the switch height is correct, tighten the two (2) 10-32 x 1/2” phillips
pan head screws.
4. Remove terminal strip cover from back of tube stand.
5. Connect wires from Horizontal SID Switch Assembly and Longitudinal Brake
Assembly to terminal block TB1 as follows:
Table 2-1. Manual Collimator TB1 Wiring
WIRE CONNECTION
WIRE DESCRIPTION
COLOR ON TB1
LONGITUDINAL MAGNET RED TB1-7
6. If tube stand is being interfaced with automatic collimator, connect four (4)
wires from Horizontal SID Switch Assembly as follows:
WIRE CONNECTION
WIRE DESCRIPTION
COLOR ON TB1
LONGITUDINAL MAGNET RED TB1-7
VERTICAL SID
SWITCH VERTICAL
ASSEMBLY COLUMN
SID SWITCH
CABLE CLAMP CABLE
LOWER HALF
CABLE
CLAMP (2X) TERMINAL
STRIP COVER
5. Install a High Voltage Cable Strain Relief on each high voltage cable attach-
ment to x-ray tube horns. (see Figure 2-21).
HIGH VOLTAGE
CABLE
STRAIN RELIEFS
(2X)
4. Clean marked area with alcohol and secure actuator in marked location using
double-sided tape.
5. Check switch height to ensure activation at 40" SID.
VERTICAL DISTANCE SCALES
The Vertical Distance Scales mount on the front side of the Vertical Column (see
Figure 2-23). The Vertical Distance Scales, calibrated in either centimeters (met-
ric system) or inches (SAE system), indicate the focal spot-to-bucky and focal
spot-to-tabletop distances
Bucky SID
1. Set tube position for 40" (100 cm) bucky SID.
2. Affix label on Vertical Column so that the 40" (100 cm) mark lines
up with the arrow in the bucky SID window of the SID Indicator
Bracket.
Table SID
1. Set tube position for 40" (100 cm) tabletop SID.
2. Affix label on Vertical Column so that the 40" (100 cm) mark lines
up with the arrow in the table SID window of the SID Indicator
Bracket.
“BUCKY”
VERTICAL SID SID INDICATOR
INDICATOR BRACKET
“TABLE”
VERTICAL SID
INDICATOR
Refer to the collimator manual for additional alignments and performance verifi-
cations associated with beam limiting device.
Collimator Light Field Versus Actual X-Ray Field
1. Using any size film with cassette, set the proper technique in the
x-ray generator.
2. Make an exposure and develop the film to verify actual x-ray field
versus collimator light field alignment is within required specifica-
tions.
3. Use collimator horizontal cross light to align the table. Rotate the
x-ray tube to check the pitch adjustment as required.
4. Verify vertical movement and alignment with bucky stand or cas-
sette holder using collimator light.
THEORY
OF OPERATION
3-1
3-2
Chapter 3 Theory of Operation
OVERVIEW
The Model QS-550 tube stand is a free-standing x-ray tube support device (i.e., does
not require wall and/or ceiling mounting) that provides a wide range of tube positions
through longitudinal, vertical, rotational, and transverse tube stand motions. The
Handgrip Assembly is designed for operator convenience with ergonomically posi-
tioned push button switches that enable tube stand movement when pressed.
The following paragraphs provide detailed theory of operation of the various tube
stand functions. See Figure 5-2, 5-3, 5-4 in Chapter 5, DIAGRAMS for system sche-
matics.
THEORY OF OPERATION
VERTICAL MOTION OPERATION
The vertical motion of the x-ray tube is controlled by a fail-safe type “vertical
brake”; when power is removed from the system, the brake remains mechanically
locked. A +24 VDC input, typically provided from the radiographic system’s x-ray
generator, is applied through connector TB1-1 and switch S1-2 to Vertical Lock
Switch S2-1. When S2 is pressed, the +24 VDC is applied to electromagnetic Ver-
tical Lock Coil L1, releasing the brake (i.e., magnetic lock) and enabling vertical
positioning of the x-ray tube. When Vertical Lock Switch S2 is released, +24 VDC
is immediately removed from L1 and the vertical brake (magnetic lock) is acti-
vated inhibiting x-ray tube vertical motion.
LONGITUDINAL MOTION OPERATION
The longitudinal motion of the x-ray tube (i.e., entire Vertical Column) is con-
trolled by a “longitudinal brake” located in the tube stand’s Base Assembly. The
+24 VDC input is applied through connector TB1-1 and switch S1 to Longitudinal
Lock Switch S3-1. When S3 is pressed, the +24 VDC is removed from electro-
magnetic Longitudinal Lock Coil L3, releasing the brake (i.e., magnetic lock) and
enabling longitudinal positioning of the x-ray tube. When push button S3 is
released, +24 VDC is immediately applied to L3 and the longitudinal brake (mag-
netic lock) is activated inhibiting x-ray tube longitudinal motion.
ANGULATION MOTION OPERATION
The angulation motion of the x-ray tube (i.e., clockwise and counter-clockwise
rotation of the x-ray tube) is controlled by an “angulation brake” located in the
tube stand’s Transverse Arm Assembly. The +24 VDC input is applied through
connector TB1-1 and switch S1 to Angulation Lock Switch S4-1. When S4 is
pressed, the +24 VDC is removed from electromagnetic Angulation Lock Coil L2,
releasing the brake (i.e., magnetic lock) and enabling rotation of the x-ray tube.
When push button S4 is released, +24 VDC is immediately applied to L2 and the
angulation brake (magnetic lock) is activated inhibiting x-ray tube angulation
motion.
SERVICE
INSTRUCTIONS
4-1
4-2
Chapter 4 Service Instructions
OVERVIEW
This chapter is designed to assist the service technician in maintaining the smooth
operation of the tube stand. The tube stand will function reliably when maintained
according to the instructions provided in this chapter. Safe equipment performance
requires the attention of service personnel who are specifically trained and experi-
enced with medical x-ray apparatus. Applicable preventive maintenance or any repair
service should be performed by these skilled individuals. Failure to follow manufac-
turer’s or service personnel’s recommendations may result in serious injury.
SERVICE MAINTENANCE
Visual Inspection
The following inspection and adjustment procedures are recommended to
maintain the system in its original operating condition.
Perform Every Six Months:
• Check for evidence of loose hardware or loose wires
• Check all bearings and bearing surfaces for cleanliness and
corrosion
• Clean Bearing Tracks, Floor Track
• Conduct a general inspection for worn or damaged parts
• Check for proper counterbalancing and smoothness of
vertical (up/down) tube stand motion
• Verify that all ground conductors are properly and securely
installed and free of corrosion or damage
• Check all electrical cabling and wiring for wear and fraying
Perform Annually:
• Lubricate tube rotation detents, angulation plate, transverse
assembly bearing wheels, column rotation detents using
lithium grease
TROUBLESHOOTING PROCEDURE
The following table provides instructions for troubleshooting problems that could
be encountered on Model QS-550 tube stand.
40" SID or 72" SID • Loss of +24 Vdc power. 1. Check for +24 Vdc on hand-
indicators not function- grips cable connector A3P1-
• Horizontal Switch (40"
ing 24.
WALL SID or 72" WALL
2. Check Horizontal Switch (40"
SID) failure
WALL SID or 72" WALL SID).
• SID lamp failure 3. Replace SID lamp.
ORDERING INFORMATION
To order replacement parts for the tube stand, contact the Service Department at;
Quantum Medical Imaging, LLC.
2002-B Orville Drive North
Ronkonkoma, New York 11779-7661 USA
Phone: (631) 567-5800 (x2)
Fax: (631) 567-5074
e-mail: service@qmiteam.com
When ordering replacement parts, supply the following information:
• Model and serial number of equipment
• Part number
• Part description
• Quantity required
• P.O. Number
• Shipping Instructions
When ordering components or parts not listed in Table 4-2, a complete descrip-
tion of the part, including its function and location should be provided with the
model number and serial number of the unit. Use only original replacement
parts, as supplied or recommended by Quantum Medical Imaging, LLC
1
2
(ANGULATION
MAGNET)
2
(LONGITUDINAL
MAGNET)
5
NOTE: Vertical SID
switch and actuator
are shown in this
figure; the
Horizontal SID
switch, located in
the Base Assembly,
and longitudinal
actuators have the
same respective
part numbers.
6
7 10
8 11
9 12
14
13
DIAGRAMS
5-1
5-2
Chapter 5 Diagrams
61/2" [165]
1/2
18 " [470]
AUTO-TRACK
ENABLE SWITCH
72.75 MAX NOTE 10
92.00 [1848]
[2337] 40"
NOTE 8 [1016] C.G.
93.38 7.92 NOTE 7
NOTE 9
72.75 MAX.
[1848]
13.00 MIN.
32.5 [330] AUTO-TRACK
[825] ENABLE SWITCH
NOTE 10
13.00 MIN
[330]